510(k) Data Aggregation

    K Number
    K061343

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-05-31

    (16 days)

    Product Code
    Regulation Number
    882.1320
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Inspirstar Glove Electrode is intended to for use with legally marketed electrostimulation devices such as TENS to deliver stimulation current for pain relief.

    Inspirstar Glove Electrode is cutaneous electrode to be used with legally marketed TENS stimulating device. The glove electrodes will deliver the stimulation signals generated by the stimulator to the skin with which they are contact.

    Device Description

    Inspirstar Glove Electrode is the conductive glove knitted from a continuous silver coated nylon yarn into the form of the glove. The conductivity is derived from the silver material of nylon yarn. Nylon yarn provides elasticity which can ensure skin contact. Inspirstar Glove Electrode has the same conductivity for dry or wet usage. The entire glove electrode is conductive having the resistance of less than 7 ohms per inch consistently.

    AI/ML Overview

    The provided document is a 510(k) summary for the Inspirstar Glove Electrode, which focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed study proving the device meets specific acceptance criteria. Therefore, much of the requested information regarding detailed study design, sample sizes, expert involvement, and ground truth establishment is not present in this filing.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The filing does not explicitly define "acceptance criteria" in a quantitative, measurable sense for the glove electrode's performance as a medical device. Instead, it focuses on demonstrating technological characteristics similar to a predicate device. The primary performance characteristic mentioned is conductivity.

    Acceptance Criteria (Implied)Reported Device Performance
    Conductivity (resistance per inch)Less than 7 ohms per inch consistently (for the entire glove)
    Ability to deliver stimulation currentFunctions as a conductive electrode for TENS devices
    Substantial equivalence to predicate deviceDeemed substantially equivalent in indications for use, technological characteristics, conductivity performance, connector interface, and use/care characteristics.

    2. Sample Size Used for the Test Set and Data Provenance

    The document states "Inspirstar Glove Electrode is tested to verify the device meets the design requirement. Predicate devices are also tested to support substantial equivalence." However, it does not specify the sample size of glove electrodes that were tested, nor does it detail the specific tests conducted beyond mentioning resistance measurement.

    • Sample Size (Test Set): Not specified.
    • Data Provenance: The document does not provide details on the specific data provenance (e.g., country of origin, prospective or retrospective) for the testing. It's implied the testing was conducted by or for Inspirstar Inc. for this 510(k) submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    This information is not applicable/not provided in this type of 510(k) submission for a device like a glove electrode. Establishing "ground truth" with human experts is typically relevant for diagnostic devices that interpret complex data (e.g., medical images, physiological signals), where clinical consensus or pathology reports serve as the gold standard. For a conductive glove, the "ground truth" is based on physical property measurements and functional testing.

    4. Adjudication Method for the Test Set

    This information is not applicable/not provided for the reasons stated above. Adjudication methods (e.g., 2+1, 3+1) are used to resolve discrepancies in expert interpretation, which is not relevant for the performance assessment of a conductive glove.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    This information is not applicable/not provided. MRMC studies are typically used for diagnostic imaging or equivalent interpretation devices to assess human-in-the-loop performance and usually involve multiple human readers interpreting cases with and without AI assistance. This device is a conductive electrode, not an AI or diagnostic interpretation device.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    This information is not applicable. This device is a physical electrode, not an algorithm, and therefore does not have a "standalone" algorithmic performance.

    7. The Type of Ground Truth Used

    For the Inspirstar Glove Electrode, the "ground truth" for its performance is based on physical property measurements, specifically:

    • Resistance measurement: The resistance of the silver-coated nylon yarn (less than 7 ohms per inch).
    • Functional performance: Its ability to act as a conductive electrode to deliver stimulation current when connected to a legally marketed TENS device.

    8. The Sample Size for the Training Set

    This information is not applicable/not provided. The Inspirstar Glove Electrode is a manufactured medical device, not an AI model that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no "training set" for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K060368

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-03-14

    (29 days)

    Product Code
    Regulation Number
    882.5890
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Inspirstar IS02 Microcurrent Stimulator is intended to be used for the symptomatic relief of chronic intractable pain.

    Device Description

    Inspirstar IS02 Microcurrent Stimulator is a Microcurrent TENS device intended to be used for the symptomatic relief of chronic intractable pain. This product can generate low current intensity pulses at microampere level and frequency of pulses. The unit supports five pre-defined therapy programs. The current intensity, frequency, time, etc. are programmable by doctors.

    AI/ML Overview

    The provided document describes a 510(k) premarket notification for the Inspirstar IS02 Microcurrent Stimulator, a TENS device intended for symptomatic relief of chronic intractable pain. This type of submission relies on demonstrating substantial equivalence to existing legally marketed predicate devices, rather than conducting new clinical trials to establish de novo effectiveness against specific acceptance criteria.

    Therefore, the document does not contain information on acceptance criteria for a novel device performance study, nor a study proving the device meets such criteria. Instead, its "performance data" section focuses on compliance with general safety and electromagnetic compatibility standards and software verification.

    However, I can extract the information relevant to the substantial equivalence process which implicitly defines the "acceptance criteria" for a 510(k) submission and how the device "meets" them:


    1. Table of Acceptance Criteria and Reported Device Performance (as applied to Substantial Equivalence):

    Acceptance Criteria (based on Predicate Device Equivalence)Reported Device Performance (Inspirstar IS02)
    Indications for Use: Same as legally marketed predicate device."Inspirstar IS02 Microcurrent Stimulator has the same indications for use as the legally marketed predicate device." (Section 6) Indications for Use: "symptomatic relief of chronic intractable pain." (Section 4)
    Technological Characteristics: Same or comparable to predicate device without raising new questions of safety or effectiveness."Inspirstar IS02 Microcurrent Stimulator has the same technological characteristics as all of the predicate devices. All are battery-powered, current source pulse generator devices with two output channels." (Section 7) "Inspirstar IS02 Microcurrent Stimulator has very close waveforms and output characteristics including frequency, current intensity, wave slope, etc with Precision Micro ('THERASTIM') and has comparable waveforms and output characteristics with Alpha-Stim 100('ALPHA-STIM CS')." (Section 7) "The user operating interfaces to setup values of frequency, current intensity, etc have some differences... But these differences do not affect the technology and do not affect the safety and effectiveness." (Section 7)
    Safety and Effectiveness: Demonstrated through compliance with recognized standards and software verification (where applicable)."Inspirstar IS02 Microcurrent Stimulator conforms to applicable voluntary standards IEC 60601-1 for safety requirement, and IEC60601-1-2 for electromagnetic compatibility requirement. In addition, the software verification has been carried out according to FDA software guidance." (Section 8)

    The following information is NOT applicable or NOT available in the provided document, as it pertains to a different type of device evaluation (e.g., a clinical study with novel performance claims, not a substantial equivalence submission):

    • 2. Sample size used for the test set and the data provenance: Not applicable. No test set for a performance study is described. The "test" is a comparison to predicate devices based on design and technical specifications.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No "ground truth" for a performance test set is described.
    • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a TENS unit, not an AI-assisted diagnostic tool requiring human reader performance studies.
    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithmic diagnostic device.
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable. The "ground truth" in a 510(k) is the established safety and effectiveness of the predicate device(s).
    • 8. The sample size for the training set: Not applicable. Not an AI/ML device requiring a training set.
    • 9. How the ground truth for the training set was established: Not applicable.

    Summary of the "Study" (Substantial Equivalence Demonstration):

    The "study" undertaken for the Inspirstar IS02 Microcurrent Stimulator was a 510(k) premarket notification to the FDA. This regulatory pathway demonstrates that a new device is "substantially equivalent" to a legally marketed predicate device.

    • Nature of the "Study": A comparison of the new device's indications for use, technological characteristics, and performance data (safety and EMC standards, software verification) against one or more predicate devices.
    • "Acceptance Criteria" for this "Study": The FDA's criteria for substantial equivalence, primarily:
      • Same intended use as the predicate device(s).
      • Same technological characteristics, OR
      • Different technological characteristics that do not raise new questions of safety or effectiveness and are supported by appropriate data (e.g., performance testing as per standards).
    • Conclusion: The FDA reviewed the submission and found the Inspirstar IS02 Microcurrent Stimulator to be substantially equivalent to the predicate devices (Precision Micro / THERASTIM and ALPHA-STIM CS). This determination allows the device to be legally marketed.