Search Results
Found 1 results
510(k) Data Aggregation
(29 days)
The InPulse Laser and the delivery accessories that are used with them are intended for use in surgical procedures involving open, laparoscopic and endoscopic ablation, vaporization, excision, incision, and coagulation of soft tissue in the medical specialties of general and cosmetic dentistry, otolaryngology, ENT surgery, and dermatology & plastic surgery including intraoral soft tissue dental surgery, oral maxillo-facial and cosmetic surgery, general surgery, E.N.T. surgery, podiatry, and dermatology and plastic surgery.
Podiatry (ablation, vaporization, incision, excision, and coagulation of soft tissue) including:
- Matrixectomy
- Periungual and subungual warts
- Plantar warts
- Radical nail excision
- Neuromas
The InPulse Laser is indicated for use for the temporary increase of clear nail in patients with onychomycosis (e.g., dermatophyton rubrum and T. mentagrophytes, and/or yeasts Candida albicans, etc.).
Dermatology and Plastic Surgery (ablation, vaporization, excision, and coagulation of soft tissue) including:
- Lesions of skin and subcutaneous tissue
- Telangiectasia
- Port wine lesions
- Spider veins
- Hemangiomas
- Plantar warts
- Periungual and subungual warts
- Removal of tattoos
- Debridement of decubitus ulcer
- Treatment of keloids
Oropharangeal/ Dental Surgery Indicated for:
- Abscess incision and drainage
- Aphthous ulcers treatment
- Biopsies, excisional and incisional
- Crown lengthening
- Exposure of unerupted I partially erupted teeth
- Fibroma removal
- Frenectomy
- Frenotomy
- Gingival incision and excision
- Gingivectomy
- Gingivoplasty
- Hemostasis
- Implant removal
- Lesion (tumor) removal
- Leukoplakia
- Operculectomy
- Oral papillectomy
- Pulpotomy
- Pulpotomy as adjunct to root canal therapy
- Removal of filling material such as gutta percha or resin as adjunct treatment during root canal re-treatment
- Selective ablation of enamel (first degree) caries removal
- Sulcular debridement (removal of diseased or inflamed soft tissue in the periodontal pocket) to improve clinical indices including gingival index, gingival bleeding index, probe depth, attachment loss, and tooth mobility
- Tissue retraction for impressions
- Vestibuloplasty
General Surgery Indicated for:
Open, laparoscopic, and endoscopic general surgery (ablation, vaporization, incision, and coagulation of soft tissue) including:
- Cholecystectomy
- Lymphadenectomy
- Mastectomy
- Partial nephrectomy
- Hepatectomy
- Pilonidal cystectomy
- Pancreatectomy
- Resection of lipoma
- Splenectomy
- Pelvic adhesiolysis
- Hemorrhoidectomy
- Removal of lesions
- Thyroidectomy
- Removal of polyps
- Parathyroidectomy
- Removal of tumors
- Herniorrhaphy
- Tumor biopsy
- Tonsillectomy
- Debridement of decubitus ulcers
- Appendectomy
Endonasal Surgery (ablation, vaporization, incision, and coagulation of soft tissue) including:
- Lesions or tumors of the oral, nasal, glossal, pharyngeal & laryngeal tissues
- Tonsillectomy
- Adenoidectomy
The InPulse Laser is comprised of the following main components:
- Main console containing the major electrical components, including:
- Control/ Display Panel with the:
- Keyswitch (that controls authorized access to the laser system);
- emergency Laser Stop button;
- Displays (laser emission indicator, average power, pulse energy, repetition rate)
- LCD screen user interface permitting selection of treatment emission when the footswitch is depressed and a fiber optic is properly attached):
- 1064 nm treatment laser (solid state Nd:YAG laser rod) with flashlamp and associated light regulation components and electronics;
- 630 -680 nm (red) aiming beam diode laser;
- Delivery device fiber-optic connector port;
- Remote interlock connector (External door interlock connector);
- Connector ports for the footswitch and power cord;
- Accessory holder (attached to the rear of the main console);
- Footswitch;
- Medical grade power cord;
- Delivery Devices for Non-Contact and Contact with Intact Skin/Tissue:
- Guide Tip No Standoff: Reusable, cleanable, tip is provided for noncontactuse to direct and control the placement of the laser beam) at the treatment location. The Guide tipattaches to the end of the handpiece. The optical fiber is threaded through the handpiece and fits securely into the bore of the Guide tip;
- Guide Tip - With Standoff: Reusable, cleanable, tip is provided for minimal-contact with intact skin/ tissue to direct and control the placement of the laser beam at the treatment location. The Guide tip attaches to the end of the handpiece. The optical fiber is threaded through the handpiece and fits securely into the bore of the Guide tip;
- Delivery Devices for Contact with Breached Surfaces:
- Optical Fibers Reusable, cleanable, sterilizable optical fibers (range of 200 1000 um diameter) provided non-sterile, clean and ready for sterilization (steam autoclave).
- Handpieces Reusable, cleanable, sterilizable handpieces (large and small diameter shafts) provided non-sterile, clean and ready for sterilization (steam autoclave). The optical fiber is threaded through the handpiece and secured and held in place with the handpiece locking cap;
- Handpiece Tips - Disposable single-use tips are provided in straight and curved configurations and are used to direct and control the placement of the optical fiber tip at the treatment location. The handpiece tips attach to the end of the handpiece. The optical fiber is threaded through both the handpiece and the handpiece tip;
- Accessories:
- Safety Glasses
- Tools:
- Optical Fiber Striper;
- Optical Fiber Cleaver (carbide wedge, ceramic, or equivalent scribe for cleaving the optical fibers).
The provided document is a 510(k) summary for the InPulse Laser, seeking substantial equivalence to a predicate device (K093547). This type of submission generally focuses on demonstrating that the new device is as safe and effective as a legally marketed predicate device, rather than proving performance against specific acceptance criteria through novel clinical studies.
Therefore, the document does not contain the detailed clinical study information typically found when a device needs to demonstrate its own performance against set acceptance criteria. Instead, it relies on the comparison of technological characteristics and identical indications for use to the predicate device.
Here's an analysis based on the provided text, addressing your points where possible, and noting where information is not present:
Description of Acceptance Criteria and Study to Prove Device Meets Them:
The InPulse Laser's acceptance criteria are implicitly defined by its substantial equivalence to the predicate device, the PinPointe FootLaser (K093547). The study demonstrating this substantial equivalence is a comparison of technological characteristics and identical indications for use. No formal clinical studies with specific acceptance criteria are reported for the InPulse Laser itself in this 510(k) summary, as its performance is assumed to be equivalent to the predicate.
1. Table of Acceptance Criteria and Reported Device Performance
Since the submission is based on substantial equivalence, the "acceptance criteria" are the characteristics of the predicate device. The "reported device performance" is the InPulse Laser's matching of these characteristics.
| Characteristic | Acceptance Criteria (Predicate: K093547 PinPointe FootLaser) | Reported Device Performance (InPulse Laser) |
|---|---|---|
| Product Code | General & Plastic Surgery | General & Plastic Surgery |
| Regulation | GEX, 21 CFR 878.4810 | GEX, 21 CFR 878.4810 |
| Intended Use | Intended for use in dermatologic and general surgical procedures | Intended for use in dermatologic and general surgical procedures |
| Indications for Use | See K093547 (Extensive list including Podiatry, Dermatology, Oropharyngeal/Dental, General, and Endonasal Surgery) | Exactly the same as K093547 |
| Wavelength | 1064nm | 1064nm |
| Aiming beam | 630-680 nm (< 2.5 mW) | 630-680 nm (< 2.5 mW) |
| Power Watts | 6W, 30W, 100W | 6W, 30W, 100W |
| Pulse Duration (usec) | 100-700 (6W), 350-3000 (30W), 350-3000 (100W) | 100-700 (6W), 350-3000 (30W), 350-3000 (100W) |
| Energy per pulse (mJ) | 20-200 (6W), 20-1000 (6W), 20-3500 (100W) | 20-200 (6W), 20-1000 (6W), 20-3500 (100W) |
| Output Mode | Pulsed, multi mode | Pulsed, multi mode |
| Repetition rate | 5-100Hz | 5-100Hz |
| Laser media | Flashlamp pumped, solid state laser rod | Flashlamp pumped, solid state laser rod |
| User interface | Push button panel | LCD screen |
| Laser activation | Footswitch | Footswitch |
| Delivery devices, how supplied | Non-sterile, reusable, cleanable, sterilizable | Non-sterile, reusable, cleanable, sterilizable |
| Electrical requirements | 90-130 VAC, 50/60 Hz; 200-240 VAC, 50/60 Hz | 90-130 VAC, 50/60 Hz; 200-240 VAC, 50/60 Hz |
Note on User Interface: The only noted difference is the user interface (LCD screen for InPulse Laser vs. Push button panel for predicate). The FDA's substantial equivalence determination implies this difference does not raise new questions of safety or effectiveness.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: Not applicable. No "test set" in the context of clinical data is reported in this 510(k) summary. The submission relies on comparative technological characteristics and identical indications for use.
- Data Provenance: Not applicable. No clinical data is presented for the InPulse Laser in this document. The statement "No clinical data was needed for these indications. They are identical to those on K093547" explicitly clarifies this.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- This information is not applicable and not provided. There was no "test set" requiring expert ground truth in this substantial equivalence submission.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- This information is not applicable and not provided. There was no "test set" requiring adjudication in this substantial equivalence submission.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- This information is not applicable and not provided. The InPulse Laser is a surgical laser device, not an AI-assisted diagnostic or interpretive tool, and no MRMC study was conducted or reported.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- This information is not applicable and not provided. The InPulse Laser is a physical medical device, not a standalone algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- This information is not applicable. No new ground truth was established for the InPulse Laser itself in this substantial equivalence submission. The ground truth for the safety and effectiveness of the predicate device would have been established during its own clearance process, but that information is not part of this document.
8. The sample size for the training set
- This information is not applicable. No new training set for an algorithm was used or reported in this substantial equivalence submission.
9. How the ground truth for the training set was established
- This information is not applicable. No new training set for an algorithm was used or reported in this substantial equivalence submission.
Ask a specific question about this device
Page 1 of 1