Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K091061

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-11-13

    (213 days)

    Product Code
    Regulation Number
    878.5000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Implicitguide™ Surgical Suture System is indicated for use in soft tissue approximation and elevation of subdermis and underlying muscle.

    Device Description

    The ImplicitCare Implicitguide™ Surgical Suture System combines USP size 4-0 braided nonabsorbable poly(ethylene terephthalate), or polyester, surgical suture with a set of manual instruments designed to efficiently and accurately place the suture for the purpose of performing approximation of tissue and elevation of subdermis/underlying muscle procedure. All components of the Implicitguide™ Surgical Suture System are provided sterile, and are intended for single use only. The manual instruments consist of the following: Tape (marking tape) to aid in marking the skin for needle (Suturod) point of entry locations)- if needed, Lancet to create a point of entry for needle (Suturod) and suture, Clearing device to clear subcutaneous ligamentous attachments around the puncture site, Two (2) stainless steel Suturods™ which, like suture needles, guide the suture through the tissue. Each Suturod™ has an internal light guide which, when connected to the Light Handle, illuminates the distal tip of the Suturod™M, Light Handle which contains an LED light source to connect to Suturod™.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study information for the Implicitguide™ Surgical Suture System based on the provided text. It's important to note that the document is a 510(k) summary, which focuses on demonstrating substantial equivalence to predicate devices rather than proving performance against novel acceptance criteria in a clinical trial. Therefore, some requested information (like effect size of AI, standalone algorithm performance, or detailed ground truth for training) is not applicable or not present in this type of submission.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test CategoryAcceptance Criteria (Stated or Implied)Reported Device Performance
    BiocompatibilityCompliance with AAMI/ISO 10993-1 requirements and FDA's Blue Book memorandum (G95-1) for an Implant Device contacting Tissue, Bone, or Blood permanently. Specific tests: Physico-chemical (Plastics, USP), Cytotoxicity (ISO MEM Elution), Sensitization (ISO Murine Lymph Node Assay), Irritation (ISO Intracutaneous Reactivity Test), Systemic Toxicity (ISO Acute Systemic Injection Test), Genotoxicity (ISO Salmonella typhimurium Reverse Mutation), Hemocompatibility (NIH Method), Pyrogen (ISO Materials Mediated Rabbit Pyrogen), Implantation (7 day and 90 day Muscle implant)."All results are acceptable." (for the TEVDEK® II polyester suture, which is stated to be exactly the same suture used in the Implicitguide™ system). "These tests demonstrate that the suture contained in the Implicitguide™ Surgical Suture System is as safe as the predicated devices."
    Suture MechanicalMeets all requirements of USP Monograph for Nonabsorbable Surgical Suture. Specifically USP XXXI <871> Needle attachment - Sutures."The subject device will meet all requirements of USP XXXI <871> Needle attachment - Sutures." (due to being the same as TEVDEK® II).
    Device Functionality (Cadaver Tests)Performance of brow elevation and shaping. Trans-illumination and color of light for Suturod. Tensile strength of medial platysmal border muscles and facial retaining ligaments. Demonstrated substantial equivalence to predicate device with added advantage of Suturod light trans-illumination."Test results are acceptable and test reports are on file."
    Electromagnetic Compatibility (EMC)Compliance with IEC 60601-1-2:2002/A1:2004 and EN60601-1-2:2007 for Radio Frequency Emissions and Electromagnetic Immunity Tests for medical electrical equipment."All test required passed requirements." "This report demonstrates the safety of the Suturod."
    Design VerificationDesign and USP requirements for various components, including: Clearing Tool Torque, Lancet Blade Puncture, Suture Length, Suture Attachment Force, Suturod Bending, Light Intensity, Light On-Time."All test results are acceptable." "These tests demonstrate that the device meets design and USP requirements."

    2. Sample Size Used for the Test Set and Data Provenance

    • Biocompatibility/Suture Mechanical Tests: The tests were conducted on the TEVDEK® II polyester suture, which is described as being "exactly the same suture" as used in the Implicitguide™ system. The sample size for these specific tests is not explicitly stated, but it would have followed the requirements of the standards (e.g., ISO, USP). The data provenance is implied to be from Teleflex Medical (formerly Genzyme Corp.), the manufacturer of TEVDEK® II. These are likely retrospective, pre-existing test results.
    • Cadaver Tests: Four cadavers (two females aged 92 and 84, two males aged 58). Data provenance: University of Maryland, State Anatomy Board, Baltimore, MD, USA. This was a prospective, specifically designed pre-clinical test.
    • EMC Evaluation: Not specified as a "test set" in terms of biological samples. The "test required" refers to the device itself being subjected to EMC testing in a laboratory environment.
    • Design Verification Tests: Sample sizes for each specific design verification test (e.g., number of sutures for attachment force, number of Suturods for bending) are not explicitly stated, but are implied by the nature of engineering design verification. The data provenance is ImplicitCare, LLC's internal testing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Biocompatibility/Suture Mechanical Tests: Ground truth is established by compliance with international standards (ISO, AAMI) and USP monographs, which are developed by expert consensus in those fields. No individual experts for a specific test set are mentioned as establishing "ground truth" in the way one would for clinical imaging studies.
    • Cadaver Tests: Two Board Certified plastic surgeons, Dr. Gregory Mueller and Dr. Sherrell Aston, performed the procedures and evaluated the results. Their CVs were on file, indicating their qualification as experts.
    • EMC Evaluation & Design Verification Tests: Ground truth is defined by the technical specifications of the standards (IEC, EN) and internal design requirements. Expert involvement would be in the design and interpretation of these tests, but not in establishing "truth" for a clinical test set.

    4. Adjudication Method for the Test Set

    • Cadaver Tests: No explicit adjudication method (like 2+1 or 3+1) is mentioned. The description states the procedures were "conducted by" the two plastic surgeons. It implies they jointly or individually evaluated the results, possibly through consensus, but this is not detailed.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, an MRMC comparative effectiveness study was not done. The submission explicitly states: "No clinical testing was conducted to support this submission." This type of study is typically associated with clinical trials evaluating diagnostic or prognostic devices with human readers interpreting cases.

    6. Standalone (Algorithm Only) Performance Study

    • No, a standalone (algorithm only) performance study was not done. This device is a surgical suture system, not an AI algorithm.

    7. Type of Ground Truth Used

    • Biocompatibility/Suture Mechanical: Established by compliance with recognized standards and monographs (e.g., ISO, USP) and prior regulatory approval (510(k) for predicate suture). This is a form of regulatory and scientific consensus.
    • Cadaver Tests: Expert evaluation/observation by Board Certified plastic surgeons during a pre-clinical cadaver study.
    • EMC Evaluation & Design Verification: Technical specifications and engineering requirements for the device components, validated against international standards.

    8. Sample Size for the Training Set

    • This information is not applicable. The device is a surgical suture system and does not involve machine learning algorithms that require a "training set." The submission details non-clinical tests demonstrating safety and equivalence.

    9. How the Ground Truth for the Training Set Was Established

    • This information is not applicable, as there is no training set for this device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1