510(k) Data Aggregation

    K Number
    DEN180001

    Validate with FDA (Live)

    Device Name
    IDx-DR
    Manufacturer
    Date Cleared
    2018-04-11

    (89 days)

    Product Code
    Regulation Number
    886.1100
    Type
    Direct
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    IDx-DR is indicated for use by health care providers to automatically detect more than mild diabetic retinopathy (mtmDR) in adults diagnosed with diabetes who have not been previously diagnosed with diabetic retinopathy. IDx-DR is indicated for use with the Topcon NW400.

    Device Description

    The IDx-DR consists of several components. A fundus camera is attached to a computer, where the IDx-DR Client is installed. The Client allows the user to interact with the server-based analysis software over a secure internet connection. Using the Client, users identify two fundus images per eye to be dispatched to IDx-Service is installed on a server hosted at a secure datacenter. IDx-DR Analysis, which runs inside IDx-Service, processes the fundus images and returns information on the image quality and the presence or absence of mtmDR to IDx-Service. IDx- Service then returns the results to the IDx-DR Client.

    AI/ML Overview

    Acceptance Criteria and Device Performance for IDx-DR

    This document details the acceptance criteria for the IDx-DR device and summarizes the study conducted to demonstrate its performance.

    1. Acceptance Criteria and Reported Device Performance

    The primary outcomes for the IDx-DR study were sensitivity and specificity for detecting more than mild diabetic retinopathy (mtmDR). Pre-defined performance thresholds were established, and the study results demonstrate the device met these criteria.

    MetricAcceptance Criteria (Threshold)Reported Device Performance (Full Analyzable Set)95% Confidence Interval (Reported)
    Sensitivity85.0%87.4%81.9% - 92.9%
    Specificity82.5%89.5%86.9% - 93.1%
    ImageabilityNot explicitly stated96.1%94.0% - 96.8%
    Positive Predictive Value (PPV)Not explicitly stated72.7%(Implicitly provided as 173/238)
    Negative Predictive Value (NPV)Not explicitly stated95.7%(Implicitly provided as 556/581)

    Note: The reported performance also includes enrichment-corrected sensitivity and specificity, which were also high and met the thresholds.

    2. Sample Size and Data Provenance for Test Set

    • Sample Size (Test Set): 819 participants were fully analyzable in the pivotal clinical study.
    • Data Provenance: The data was collected prospectively from 10 primary care sites across the United States. The target population was adults diagnosed with diabetes who had not been previously diagnosed with diabetic retinopathy. The study population was enriched by targeting enrollment of subjects with elevated Hemoglobin A1c (HbA1C) levels.

    3. Number and Qualifications of Experts for Ground Truth (Test Set)

    • Number of Experts: Three experienced and validated readers.
    • Qualifications of Experts: The readers were certified by the Fundus Photography Reading Center (FPRC) and had expertise in evaluating the severity of retinopathy and diabetic macular edema (DME) according to the Early Treatment for Diabetic Retinopathy Study (ETDRS) scale and Diabetic Retinopathy Clinical Research Network (DRCR) grading paradigm.

    4. Adjudication Method for the Test Set

    The adjudication method used for establishing the ground truth from the FPRC readers was a majority voting paradigm for the four widefield stereo image pairs.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No explicit MRMC comparative effectiveness study involving human readers' improvement with AI vs. without AI assistance was reported. The study focused on the standalone performance of the IDx-DR device against an expert-derived reference standard.

    6. Standalone (Algorithm Only) Performance

    Yes, a standalone performance study was conducted. The reported sensitivity, specificity, PPV, and NPV values are for the IDx-DR algorithm operating autonomously, without human-in-the-loop assistance during the diagnostic process.

    7. Type of Ground Truth Used (Test Set)

    The ground truth used was expert consensus based on comprehensive ophthalmic imaging (dilated four widefield stereo color fundus photography and macular optical coherence tomography (OCT) imaging) read by three experienced and validated readers at the Fundus Photography Reading Center (FPRC). The severity of retinopathy and DME was determined according to the ETDRS scale and DRCR grading paradigm, using a majority voting paradigm.

    8. Sample Size for the Training Set

    The document does not explicitly state the sample size used for the training set. It describes the clinical study as a pivotal clinical study with 900 enrolled patients, which formed the basis for evaluating the device's performance, but it does not specify what portion (if any) of this dataset was used for training or validation during the development phase. The language focuses on the "analyzable fraction" of participants for the primary outcomes, implying this was the test set.

    9. How the Ground Truth for the Training Set Was Established

    The document does not provide details on how the ground truth was established for the training set. It primarily describes the methodology for establishing the ground truth for the test set used in the pivotal clinical study. It mentions that IDx has provided a full characterization of the technical parameters of the software, including a description of the algorithms, and that IDx will make future algorithm improvements under a consistent medically relevant framework. However, the details of training data ground truth establishment are not discussed.