    K Number
    K020576

    Validate with FDA (Live)

    Date Cleared
    2002-05-01

    (69 days)

    Product Code
    Regulation Number
    866.1640
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For the visual or spectrophotometric detection of the vanA and vanB genes in determining vancomycin resistance in enterococci isolated from culture.

    Device Description

    The Velogene™ Genomic Identification Assay for VRE (Vancomycin Resistant Enterococci) is an in vitro, DNA probe based, diagnostic device that utilizes Cycling Probe™ Technology (CPT) to generate a spectrophotometric or visual result. Results can be generated 90 minutes after primary isolation. The Velogene™ Genomic Identification Assay for VRE (hereafter may be referred to as the Velogene™ assay) utilizes a fluorescein labeled, biotinylated DNA-RNA-DNA chimeric probe providing an RNase H cleavable linkage when bound to the complementary sequence of the van A or vanB gene. RNase H cleaves the RNA portion of the chimeric probe when it is hybridized to the target DNA. The uncleaved probe (vanA and vanB negative) is detected by binding of the fluoresceinated probe to a solid surface and attachment of an antifluorescein antibody conjugated with horseradish peroxidase, which converts a substrate to a colored end product. Cleavage of the probe (vanA or vanB positive) prevents binding of the probe-anti-fluorescein antibody enzyme complex, thus preventing formation of the colored end product. A vancomycin-resistant isolate (i.e. vanA or vanB gene is present) will produce a colorless result (or OD650 of ≤0.14). A vancomycin-sensitive isolate (i.e. vanA and vanB gene are absent) will produce a distinctly blue color (or OD650 of >0.14). Each Velogene™ assay kit contains supplies sufficient to process 48 samples and consists of two separate reagent kits: a VRE Lysis/Cycle Kit and a VRE Microwell Detection Kit.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not explicitly state pre-defined acceptance criteria (e.g., minimum sensitivity, specificity, or agreement metrics that needed to be met). Instead, it presents the results of a comparative study and implies that the observed agreement rate was sufficient for substantial equivalence.

    MetricAcceptance Criteria (Implied)Reported Device PerformanceComments
    Agreement with Predicate Device (Vancomycin Screen Agar)Sufficient for Substantial Equivalence94.6% agreementThis agreement rate was based on a comparative study. The submission implies that this level of agreement was acceptable for FDA clearance.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 518 isolates
    • Data Provenance: The study was conducted at "3 geographically distributed U.S. sites" using "routinely submitted samples for microbiological identification." This indicates the data was prospective or at least collected in a routine clinical setting.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not specify the number of experts or their qualifications for establishing the ground truth. The ground truth was based on the predicate device, Vancomycin Screen Agar, which is a growth-based test. Interpretation of this test would typically be performed by trained laboratory personnel, but no specifics are given regarding expert review or reconciliation.

    4. Adjudication Method for the Test Set

    The document does not describe any specific adjudication method (e.g., 2+1, 3+1). The "ground truth" was established by the predicate device (Vancomycin Screen Agar), and the Velogene™ assay's results were compared against this. It's implied that the results of the Vancomycin Screen Agar were considered definitive for the purpose of this comparison.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not applicable and therefore not performed. This device is a standalone diagnostic assay (an in vitro DNA probe test) that provides a direct result (colorless or blue, or OD650 reading). It does not involve human readers interpreting images or data where AI assistance would improve human performance.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance evaluation was done for the Velogene™ Genomic Identification Assay for VRE. The study compared the results of the Velogene™ assay (an "algorithm only" in the sense of a chemical/molecular assay without human interpretation of complex visuals) directly against the predicate device. The assay itself generates a direct spectrophotometric or visual result without human interpretation beyond reading the color or OD650 value.

    7. The Type of Ground Truth Used

    The ground truth used for the comparative study was the phenotypic result from a legally marketed predicate device: Vancomycin Screen Agar (Brain Heart Infusion (BHI) Agar with 6 ug/mL vancomycin). This is a growth-based test that identifies vancomycin resistance.

    8. The Sample Size for the Training Set

    The document does not provide information on a separate training set or its sample size. This type of submission (510(k)) for an IVD kit often focuses on clinical validation rather than a deep dive into the development and training of a machine learning model, which is not applicable here as it's a chemical assay. The comparison is the validation study.

    9. How the Ground Truth for the Training Set Was Established

    Since no separate training set is mentioned in the provided text, and the device is a molecular assay, the concept of establishing ground truth for a "training set" as it would apply to a machine learning algorithm is not relevant to this submission. The "ground truth" for the overall evaluation was the predicate device's results, against which the Velogene™ assay was compared.

    Ask a Question

    Ask a specific question about this device

    K Number
    K010858

    Validate with FDA (Live)

    Date Cleared
    2001-05-24

    (63 days)

    Product Code
    Regulation Number
    866.1640
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K990640

    Validate with FDA (Live)

    Date Cleared
    1999-07-09

    (133 days)

    Product Code
    Regulation Number
    866.1640
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ID Biomedical Velogene™ Rapid MRSA Identification Assay is intended as a qualitative assay for the definitive identification of methicillin resistance in presumptively identified cultures of Staphylococcus aureus by detecting the presence of the mecA gene. The presence of the mecA gene confers resistance to methicillin.

    Device Description

    The Velogene™ Rapid MRSA Identification Assay is a DNA probe based diagnostic device that is based on Cycling Probe™ Technology (CPT) to generate spectrophotometric or visual results. The assay consists of two reagent kits, MRSA Lysis/Cycle Kit and MRSA Microwell Detection Kit.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the Velogene™ Rapid MRSA Identification Assay, based on the provided text:

    Acceptance Criteria and Device Performance

    The primary acceptance criterion is substantial equivalence to the predicate device, the Oxacillin Screen Agar (NCCLS approved test for MRSA detection). This substantial equivalence is demonstrated through high agreement in identifying methicillin resistance in S. aureus samples.

    Acceptance Criteria CategoryAcceptance Criteria (Implicit)Reported Device Performance (Velogene™ Rapid MRSA Identification Assay)
    Overall Agreement with Predicate DeviceHigh agreement with Oxacillin Screen Agar in clinically relevant S. aureus samples.99.3% agreement (420/423) in routine clinical samples.
    Reproduction of MRSA ResultsHigh reproducibility for MRSA results.99.8% (449/450) reproducible for MRSA (OD650 ≤ 0.18, Visual: Clear).
    Reproduction of MSSA ResultsHigh reproducibility for MSSA results.99.8% (449/450) reproducible for MSSA (OD650 > 0.18), 100% (450/450) visual reproducibility.
    Analytical Sensitivity (MRSA)Detect MRSA at clinically relevant concentrations.Detects 93.75 x 10 to 375 x 10 MRSA CFU/reaction (OD650 < 0.18 or visual clear).
    Analytical Specificity (S. aureus)Specific for S. aureus mecA gene detection, limiting false positives from other species.Variations observed with coagulase-negative staphylococci (Staphylococcus epidermidis, S. warneri, S. sciuri), indicating the assay should only be used for coagulase-positive Staphylococcus.
    Challenge Panel AccuracyHigh accuracy in identifying known MRSA/MSSA/BORSA isolates.95.7% accurate across 4 clinical sites for a challenge panel.
    BORSA Panel IdentificationCorrectly identify BORSA samples.100% correct identification (32/32) visually and spectrophotometrically for BORSA samples.

    Study Details

    2. Sample size used for the test set and the data provenance:

    • Clinical Test Set:

      • Sample Size: 423 coagulase-positive S. aureus samples.
      • Data Provenance: Routinely submitted samples for microbiological identification at 4 geographically distributed U.S. sites. This indicates a prospective collection of real-world clinical samples.
    • Reproducibility Panel:

      • Sample Size: 800 samples (10 isolates in a panel, 2 replicates per panel member, 5 days, 2 lots, 4 clinical sites; 10 * 2 * 5 * 2 * 4 = 800).
      • Data Provenance: Tested at 4 clinical sites.
    • Challenge Panel:

      • Sample Size: 75 isolates (49 homoresistant/heteroresistant MRSA, 25 MSSA, 1 BORSA).
      • Data Provenance: Tested by each of the 4 clinical sites.
    • BORSA Panel:

      • Sample Size: 32 samples (8 samples per site across 4 clinical sites).
      • Data Provenance: Tested across 4 clinical sites.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    The text does not explicitly state the number or qualifications of experts used to establish the initial ground truth for S. aureus classification prior to testing with the predicate device. However, the ground truth for discrepant isolates was established by nuc/mecA PCR, which is a molecular, objective gold standard.

    4. Adjudication method for the test set:

    For the initial comparison, the "ground truth" was largely established by the predicate device (Oxacillin Screen Agar). For the 3 discrepant isolates between the Velogene™ assay and the Oxacillin Screen Agar, the adjudication method was nuc/mecA PCR. This serves as a definitive molecular gold standard to resolve discrepancies.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No, this is not an MRMC comparative effectiveness study in the context of human readers and AI assistance. This study evaluates a diagnostic assay's performance against a predicate device. The "readers" are the methods of interpretation (spectrophotometer vs. visual) of the Velogene™ assay, and its comparison is against a growth-based culture method. There's no AI involved.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    The Velogene™ Rapid MRSA Identification Assay can be interpreted visually or using a spectrophotometer. While the spectrophotometer provides an objective reading (OD650 ≤ 0.18 or > 0.18) similar to an "algorithm only" interpretation, visual interpretation is also an option, implying a human-in-the-loop component for some results. However, the performance data presents both spectrophotometric and visual results as highly concordant, suggesting objective interpretation is a primary mode. The PCR results validating discrepancies are also a standalone, objective measure.

    7. The type of ground truth used:

    • Primary Ground Truth for Clinical Comparison: The Oxacillin Screen Agar (a NCCLS approved test for MRSA detection) served as the primary comparative ground truth.
    • Definitive Ground Truth for Discrepancy Resolution: nuc/mecA PCR was used as the definitive ground truth for the 3 discrepant isolates, confirming the presence or absence of the mecA gene. This is a molecular/pathology-based ground truth.
    • Analytical Sensitivity/Specificity/Reproducibility: Used defined control strains (S. aureus ATCC 29213 - mecA negative, S. aureus ATCC 33592 - mecA positive) and specific isolates for analytical specificity testing. These are laboratory-defined ground truths.

    8. The sample size for the training set:

    The document describes performance evaluation studies and does not explicitly mention a dedicated training set or its sample size. This is common for traditional diagnostic assays where performance is validated against established methods and/or molecular gold standards, rather than "trained" like machine learning models. The study focuses on validation of the assay.

    9. How the ground truth for the training set was established:

    As no explicit training set is described, this question is not applicable. The assay's principles are based on known molecular biology (detection of the mecA gene).