Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K180670

    Validate with FDA (Live)

    Date Cleared
    2018-12-07

    (268 days)

    Product Code
    Regulation Number
    888.3040
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ICONN Campbell Interference Screw System is intended for fixation of tissue, including ligament or tendon to bone for the following indications:

    Shoulder: Rotator Cuff Repairs, Bankart Repair, Biceps Tenodesis. Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction.

    Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Metatarsal Ligament Repair, Flexor Hallucis Longus for Achilles Tendon reconstruction, tendon transfers in the foot and ankle.

    Knee: Anterior Cruciate Ligament Repair, Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Illiotibial Band Tenodesis.

    Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction.

    Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction, Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty), Carpal Ligament Reconstructions and repairs, tendon transfer in the hand/wrist, Lateral Epicondylitis repair.

    Device Description

    The ICONN Campbell Interference Screws (various part numbers) are sterile (gamma irradiation), single use implantable devices available in two materials: Solvay's Zeniva®ZA-600L polyetheretherketone (PEEK) or Ti 6Al-4 ELI (Grade 23) and is intended for fixation of tissue, including ligament or tendon to bone. The screws are available in multiple sizes ranging from diameters of Ø6mm - Ø12mm in 1mm increments. The Ø6mm screw is available only in 20mm long, the Ø7mm - Ø10mm screws are available in 20mm and 25mm lengths, and the Ø11mm and Ø12mm screws are available in 25mm lengths only.

    AI/ML Overview

    Here's a breakdown of the requested information based on the provided FDA 510(k) summary for the ICONN Campbell Interference Screw:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Mechanical Performance
    Axial Pullout Strength (after cyclic loading)Similar or better than the predicate device (Arthrex PEEK Interference Screw) OR within acceptable ranges for the intended use.Performed similarly or better than the predicate OR within acceptable ranges for the intended use.
    Displacement during Cyclic LoadingSimilar or better than the predicate device (Arthrex PEEK Interference Screw) OR within acceptable ranges for the intended use.Performed similarly or better than the predicate OR within acceptable ranges for the intended use.
    Torsional Properties (Insertion Testing)Device can be safely implanted.Testing results showed that the device can be safely implanted.
    Driving Torque (Insertion Testing)Device can be safely implanted.Testing results showed that the device can be safely implanted.
    Biocompatibility
    CytotoxicityMet acceptance criteria (in accordance with ISO 10993-1:2009/(R)2013).Met acceptance criteria.
    IrritationMet acceptance criteria (in accordance with ISO 10993-1:2009/(R)2013).Met acceptance criteria.
    SensitizationMet acceptance criteria (in accordance with ISO 10993-1:2009/(R)2013).Met acceptance criteria.
    Endotoxin (LAL Test)< 0.05 EU/device (in accordance with ANSI/AAMI ST72:2016).< 0.05 EU/device.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes for the mechanical testing (axial pullout, cyclic loading, torsional properties, driving torque). It mentions that "the version of the screw made out of PEEK material was tested, as that material constitutes a worst case between the two materials offered." For the LAL testing, it states "One (1) each of the test articles were placed in 10 mL of Water for Injection."

    The data provenance is not specified in terms of country of origin. The study appears to be prospective in nature, as it involves testing a newly developed device against pre-established standards and a predicate device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This device is a medical implant, and the "ground truth" for its performance is established through physical and chemical testing against engineering standards and comparison with a predicate device, not through expert human interpretation of data like in an AI/imaging study.

    4. Adjudication Method for the Test Set

    Not applicable. As noted above, the evaluation is based on objective physical and chemical test results, not subjective interpretation requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is typically used for diagnostic or imaging devices where human readers interpret medical images or data. The ICONN Campbell Interference Screw is a surgical implant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the performance evaluation described is entirely "standalone" in the sense that it pertains to the physical device itself, independent of human interaction beyond implantation. There is no AI algorithm involved.

    7. The Type of Ground Truth Used

    The ground truth used for this device is based on engineering standards (ASTM F543-2013, ISO 10993-1:2009/(R)2013, ANSI/AAMI ST72:2016) and comparative performance metrics with a legally marketed predicate device (Arthrex PEEK Interference Screw, K062466).

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/machine learning device that would require a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. There is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182264

    Validate with FDA (Live)

    Date Cleared
    2018-10-19

    (59 days)

    Product Code
    Regulation Number
    888.3040
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ICONN Revolution Knotless Suture Anchor is intended for the reattachment of soft tissue to bone for the following indications:

    Shoulder: Rotator cuff repair, Bankart repair, Biceps Tenodesis, Acromio-Clavicular separation, Deltoid repair, and Capsular shift or Capsulolabral reconstruction.

    Foot/Ankle: Lateral stabilization, Medial stabilization, Achilles tendon repair, Hallux valgus reconstruction, Mid-foot reconstruction, Metatarsal ligament repair.

    Device Description

    The ICONN Revolution Knotless Suture Anchor (P/N 350000) is a sterile (gamma irradiation), single use implantable device made from Solvay's Zeniva® ZA-500 polyetheretherketone (PEEK) material indicated for soft tissue reattachment procedures in the shoulder, foot, and ankle. The anchor is available in one size (Diameter x Length): 5mm x 23mm. The implants with attached disposable inserter are supplied sterile and the instrument is supplied Non-Sterile and is to be sterilized in the ICONN instrument tray prior to surgery. The disposable inserter is made out of 17-4 stainless steel and the reusable driver is made out of 17-4 stainless steel (shaft) and silicone (handle). All anchors are intended to be used with the supplied #2 sutures and/or 1.5mm ICONN TetherTape, both made from Ultra High Molecular Weight Polyethylene (UHMWPE). The sutures/tape are supplied sterile, are packaged individually, and are single strands measuring 36/38 inches in length.

    AI/ML Overview

    The provided text is a 510(k) summary for the ICONN Revolution Knotless Suture Anchor, a medical device. It describes the device, its intended use, and comparative studies against a predicate device to establish substantial equivalence for FDA clearance. The summary does not provide information about an AI/algorithm-based device as requested in the original prompt. Therefore, I cannot extract the specific details you've asked for regarding acceptance criteria and an AI study.

    However, I can provide the available information related to the device's performance testing and acceptance criteria as described in the document, which are related to its mechanical and biocompatibility properties, not AI performance.

    Here's a breakdown of the information that is available in the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document mentions performance testing but does not present a formal table of "acceptance criteria" with specific numerical targets alongside "reported device performance." Instead, it states the following regarding the outcomes of the performance tests:

    Test TypeAcceptance Criteria (Implicit)Reported Device Performance
    Pullout CharacteristicsCompliance with FDA's Draft Guidance Document for Testing Bone Anchor Devices (April 20, 1996)"performed similarly or better than the predicate or within acceptable ranges for the intended use."
    - Anchor displacement after cyclic loading(Implicitly, within limits acceptable for safe and effective use and comparable to predicate)(Met implicit criteria)
    - Ultimate anchor pull-out(Implicitly, within limits acceptable for safe and effective use and comparable to predicate)(Met implicit criteria)
    Insertion TestingCompliance with ASTM F543-2013 (specifically sections A1 and A2)"demonstrate the device can be safely implanted."
    - Torsional properties(Implicitly, within limits acceptable for safe implantation)(Met implicit criteria)
    - Driving torque(Implicitly, within limits acceptable for safe implantation)(Met implicit criteria)
    Biocompatibility TestingCompliance with ISO 10993/(R)2013"all testing met acceptance criteria."
    - Cytotoxicity(Met ISO 10993/(R)2013 acceptance criteria)(Met acceptance criteria)
    - Irritation(Met ISO 10993/(R)2013 acceptance criteria)(Met acceptance criteria)
    - Sensitization(Met ISO 10993/(R)2013 acceptance criteria)(Met acceptance criteria)
    LAL testing (Endotoxin)In accordance with ANSI/AAMI ST72:2006, result <0.05 EU/device"results demonstrated the test article were <0.05 EU/device."
    Pyrogenicity(Non-pyrogenic status)"device was determined to be non-pyrogenic."

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes used for the mechanical or biological testing described. It also does not provide information about data provenance (e.g., country of origin, retrospective or prospective nature) for these tests. The tests performed are laboratory bench tests and material tests, not clinical studies.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This question is not applicable. The evaluations described are mechanical and biocompatibility tests conducted in a laboratory setting, not clinical studies requiring expert interpretation of medical data or ground truth establishment by medical professionals.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This question is not applicable, as the tests described are objective laboratory and material tests, not subjective interpretations requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no mention of an MRMC study or any AI component in the provided text. The device is a physical Suture Anchor.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This question is not applicable, as the device is a physical Suture Anchor and does not involve any algorithm or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This question is not applicable in the context of the physical device's performance and biocompatibility testing. "Ground truth" in this context refers to objective measurements and established standards (e.g., precise force measurements, chemical analysis, cell viability assays according to ISO standards).

    8. The sample size for the training set

    This question is not applicable, as the device does not involve a "training set" in the context of an AI or algorithm.

    9. How the ground truth for the training set was established

    This question is not applicable, as the device does not involve a "training set" in the context of an AI or algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181680

    Validate with FDA (Live)

    Date Cleared
    2018-08-09

    (44 days)

    Product Code
    Regulation Number
    888.3040
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ICONN Answer II Suture Anchor is indicated for soft tissue reattachment procedures in the shoulder, foot, and ankle. Specific indications are as follows: Shoulder Indications: Rotator Cuff Repairs, Biceps Tenodesis Foot/Ankle Indications: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair.

    Device Description

    The ICONN Answer II Suture Anchor (P/N's 147501 and 155001) is a sterile (gamma irradiation), single use implantable device made from Solvay's Zeniva® ZA-600L polyetheretherketone (PEEK) material indicated for soft tissue reattachment procedures in the shoulder, foot, and ankle. The anchor is available in two sizes (Diameter x Length): 4.75mm x 16mm and 5.5mm x 16mm.

    AI/ML Overview

    The provided text is a 510(k) summary for the ICONN Answer II Suture Anchor, a medical device. It describes performance testing conducted to demonstrate substantial equivalence to a predicate device. However, it does not contain information about studies involving AI or human readers for diagnostic purposes.

    Therefore, I cannot fulfill your request for information related to:

    • A study proving the device meets acceptance criteria using an AI system.
    • Number of experts establishing ground truth or their qualifications.
    • Adjudication method.
    • MRMC comparative effectiveness study.
    • Standalone algorithm performance.
    • Sample size and ground truth for training set.

    The document focuses on mechanical and biocompatibility testing for a physical medical device (suture anchor), not a diagnostic AI system.

    Here's the information that can be extracted from the provided text, related to the device's acceptance criteria and performance, as appropriate for a physical implantable device:


    Description of Device Testing and Acceptance Criteria

    The ICONN Answer II Suture Anchor underwent performance testing to demonstrate its safety and effectiveness, and substantial equivalence to its predicate device (Arthrex PEEK Corkscrew FT (K061665)). The tests focused on mechanical properties, biocompatibility, and sterilization.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test TypeSpecific Test / ParameterAcceptance CriteriaReported Device Performance
    Mechanical PerformanceAnchor displacement after cyclic loadingPerformance similar or better than predicate, or within acceptable ranges. (Guidance from FDA's Draft Guidance Document for Testing Bone Anchor Devices, April 20, 1996)Performed similarly or better than the predicates or within acceptable ranges for the intended use.
    Ultimate anchor pull-outPerformance similar or better than predicate, or within acceptable ranges. (Guidance from FDA's Draft Guidance Document for Testing Bone Anchor Devices, April 20, 1996)Performed similarly or better than the predicates or within acceptable ranges for the intended use.
    Torsional properties (Insertion Testing)Demonstrate device can be safely implanted. (Followed ASTM F543-2013, section A1)Analyzed and compared to demonstrate the device can be safely implanted.
    Driving torque (Insertion Testing)Demonstrate device can be safely implanted. (Followed ASTM F543-2013, section A2)Analyzed and compared to demonstrate the device can be safely implanted.
    Biocompatibility TestingCytotoxicityCytotoxicity score of 2 or less (in accordance with ISO 10993/(R)2013).All test articles resulted in a cytotoxicity score of zero (0).
    Irritation (SC extract)Score difference of 0.2 or less (in accordance with ISO 10993/(R)2013).Score difference of 0.0.
    Irritation (SO extract)Score difference of 0.2 or less (in accordance with ISO 10993/(R)2013).Score difference of 0.1.
    SensitizationNo evidence of causing delayed dermal contact sensitization (in accordance with ISO 10993/(R)2013).No evidence of causing delayed dermal contact sensitization.
    LAL testing (Endotoxin)< 20 EU/device (in accordance with ANSI/AAMI ST72:2016).The test articles were <0.05 EU/device.
    Material-mediated PyrogenicityNon-pyrogenic.Determined to be non-pyrogenic.
    Sterilization ValidationBioburden RecoveryUsed as an adjustment factor; no specific acceptance criteria stated for this test itself.51.5% (used as an adjustment factor for bioburden data).
    Bioburden (3 Lots)< 1,000 CFU/device.<3.9 CFU/device for all three lots.
    Bacteriostasis/Fungistasis (B/F)Demonstrates no inhibition from test article; all challenge organisms show positive growth on both subject device and control.Demonstrated no inhibition from the test article with all three challenge organisms showing positive growth on both subject device and control samples.
    SterilityNot more than one (1) positive test of sterility.Zero (0) positive tests of sterility.

    2. Sample size used for the test set and the data provenance:

    The document mentions "side by side mechanical testing performed on the subject device and predicate" and "Bioburden – 3 Lots," but does not explicitly state the number of samples or devices tested for each mechanical or biocompatibility test. The data provenance is not explicitly stated as retrospective or prospective, but testing for device clearance typically occurs prospectively as part of product development. The country of origin of the data is implied to be within the US, given the FDA submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable. This device is a physical implant, and the "ground truth" for mechanical and biocompatibility testing is based on objective laboratory measurements and established biological assays, not expert interpretation of diagnostic images.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    Not applicable. This concept is relevant for studies involving human interpretation or consensus, such as clinical trials or AI performance evaluations, not for basic device performance testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This document pertains to a physical suture anchor, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This document pertains to a physical suture anchor, not an algorithm.

    7. The type of ground truth used:

    • Mechanical Testing: Objective measurements (e.g., force, displacement, torque) against established engineering standards (e.g., FDA Draft Guidance, ASTM F543-2013). The "ground truth" is the physical behavior of the device under defined conditions.
    • Biocompatibility Testing: Results of standardized biological assays (e.g., ISO 10993) against predefined acceptance criteria for cellular response, irritation, sensitization, and endotoxin levels. The "ground truth" is the biological reaction to the material.
    • Sterilization Validation: Microbiological testing results against established sterility assurance levels and bioburden limits. The "ground truth" is the confirmed absence of viable microorganisms.

    8. The sample size for the training set:

    Not applicable. This is not an AI/machine learning study.

    9. How the ground truth for the training set was established:

    Not applicable. This is not an AI/machine learning study.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172491

    Validate with FDA (Live)

    Date Cleared
    2017-11-29

    (104 days)

    Product Code
    Regulation Number
    888.3040
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ICONN Square Suture Anchor System is intended to provide secure fixation of soft tissue to bone in orthopedic procedures where fixation is needed as a result of injury or degenerative disease. The system is designed for use in the hip, shoulder, foot, ankle, elbow, wrist, hand, and knee for the following indications:

    Hip Indications: Capsule Repair, Acetabular Labrum Reattachment

    Shoulder Indications: Capsular Stabilization, Bankart Repair, Anterior Shoulder Instability, SLAP Lesion Repair, Capsular Shift or Capsulolabral Reconstruction, Acromioclavicular Separation Repair, Rotator Cuff Repair, Biceps Tenodesis.

    Foot/Ankle Indications: Hallux Valgus Repair, Medial or Lateral Instability Repair/Reconstruction, Midfoot Reconstruction, Metatarsal Ligament/Tendon Repair/Reconstruction, Bunionectomy.

    Elbow. Wrist, and Hand: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction, Lateral Epicondylitis Repair.

    Knee: Extra Capsular Repair, Medial Collateral Ligament Repair, Lateral Ligament Repair, Posterior Oblique Ligament Repair, Patellar Realignment and Tendon Repair, Vastus Medialis Obliquous Advancement, Iliotibial Band Tenodesis.

    Device Description

    The ICONN Square 2.5MM Suture Anchors (P/N 125000) are sterile (gamma irradiation), single use implantable devices made from Solvay's Zeniva® ZA-500 polyetheretherketone (PEEK) material intended to provide secure fixation of soft tissue to bone in orthopedic procedures where fixation is needed as a result of injury or degenerative disease. The anchors are available in one size (Width x Length): 2.5mm x 7.95mm and can be used as a knotted or knotless anchor.

    AI/ML Overview

    This document is related to a 510(k) premarket notification for a medical device, specifically the ICONN Square 2.5MM Suture Anchor. It primarily focuses on demonstrating substantial equivalence to predicate devices, rather than a clinical study evaluating an AI algorithm or diagnostic device. Therefore, many of the requested elements for an AI-based device's acceptance criteria and study proving its performance (like sample size for test/training sets, expert ground truth establishment, MRMC studies, standalone performance, etc.) are not applicable to this type of submission.

    The "acceptance criteria" for this device are based on its mechanical performance relative to predicate devices, ensuring it is at least as safe and effective.

    Here's the information extracted from the provided text as it relates to the mechanical device, with an explanation of why certain AI-specific questions are not applicable:

    1. Table of Acceptance Criteria and Reported Device Performance

    For this mechanical device, "acceptance criteria" are embodied by demonstrating performance that is "similarly or better than the predicates or within acceptable ranges for the intended use."

    Acceptance Criteria (Implied)Reported Device Performance
    Mechanical Performance:
    Anchor displacement after cyclic loading "acceptable ranges"Performed similarly or better than the predicates.
    Ultimate anchor pull-out within "acceptable ranges"Performed similarly or better than the predicates.
    Insertion of the device is safe.Insertion testing demonstrated the device can be safely implanted.
    Biocompatibility:
    Biocompatibility equivalent to predicate devices.Demonstrated substantial equivalence in biocompatibility to a predicate.
    Sterility:
    Sterile (gamma irradiation).Device is sterile via gamma irradiation.
    Material:
    Material (PEEK) is suitable for intended use.Made from Solvay's Zeniva® ZA-500 polyetheretherketone (PEEK) material.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not specify the exact number of units tested for "anchor displacement after cyclic loading," "ultimate anchor pull-out," or "insertion testing." It refers to "side by side mechanical testing."
    • Data Provenance: The testing was conducted in a laboratory setting as part of the premarket notification process for the manufacturer (ICONN Orthopedics, LLC). It is not clinical data from patients or a specific country of origin in the way an AI study would entail. It's not retrospective or prospective clinical data but rather engineering test data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Not Applicable. This is a mechanical device, not an AI algorithm requiring expert interpretation to establish a "ground truth" for diagnostic purposes. The "ground truth" here is objective mechanical measurements (e.g., force, displacement).

    4. Adjudication Method for the Test Set

    • Not Applicable. This concept applies to human interpretation of data where disagreements need resolution. Mechanical testing results are objective measurements.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • Not Applicable. An MRMC study is for evaluating the impact of a new diagnostic technology (often AI) on human reader performance. This device is a surgical implant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • In a sense, yes, for mechanical performance. The "standalone" performance here refers to the device's intrinsic mechanical properties (pull-out strength, displacement) in a test environment, independent of a surgical outcome. However, this is not an "algorithm" and does not relate to human-computer interaction.

    7. The Type of Ground Truth Used

    • Mechanical Measurement and Engineering Standards: The ground truth for this device's performance is derived from established mechanical testing standards and comparison to the performance characteristics of predicate devices. The FDA's "Draft Guidance Document for Testing Bone Anchor Devices (April 20, 1996)" was used for guidance.
    • Biocompatibility: Demonstrated through equivalence to a predicate device (ICONN Answer Suture Anchor, K132724) which presumably had its biocompatibility established.

    8. The Sample Size for the Training Set

    • Not Applicable. This is not an AI/machine learning device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. As above, no training set was used.
    Ask a Question

    Ask a specific question about this device

    K Number
    K132724

    Validate with FDA (Live)

    Date Cleared
    2014-04-17

    (230 days)

    Product Code
    Regulation Number
    888.3040
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1