    K Number
    K160266

    Validate with FDA (Live)

    Date Cleared
    2016-04-08

    (66 days)

    Product Code
    Regulation Number
    888.3358
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ICONACY I-Hip is indicated for the following conditions: (1) a severely painful and/or disabled hip joint as a result of osteoarthritis, traumatic arthritis, rheumatoid arthritis, or congenital hip dysplasia, (2) avascular necrosis of the femoral head, (3) acute traumatic fracture of the femoral head or neck, (4) failed previous hip surgery including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement, (5) certain cases of ankylosis, (6) nonunions, correction of functional deformity, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.

    The ICONACY I-Hip consists of femoral stem and acetabular cup (i.e. shell) porous coated components intended for cementless, press-fit fixation.

    Device Description

    The ICONACY I-Hip consists of a collarless, tapered, forged titanium allov femoral stem mated to a cobalt chrome alloy modular femoral head. Forty percent of the femoral stem is circumferentially coated with a titanium coating designed to attain a cementless, press-fit fixation. This femoral construct articulates with an acetabular device assembly.

    The acetabular device assembly consists of a hemispherical titanium alloy cup coupled with a highly cross linked ultra-high molecular weight polyethylene (HXL-UHMWPE) liner. The acetabular cup is machined from forged Ti-6Al-4V ELI alloy and is available in standard, finned and spike configurations. The acetabular cup has a threaded polar hole for insertion. The standard and finned cups have screw holes for additional fixation.

    The outer hemispheric surface of the acetabular cup has a titanium plasma spray coating for cementless, press-fit fixation. A titanium locking ring is fixed into a groove on the cup to engage a groove on the HXLUHMWPE liner. Standard instrumentation is used to implant the device.

    The current submission addresses modifying the standard and finned acetabular cups to add a third screw hole. The acetabular cup materials, porous coating and indication for use remain the same. No additional sizes are being added to the product line. No new instruments are required. The spiked cups are unchanged with no screw holes.

    The compatible femoral stems, femoral heads, acetabular liners and bone screws remain unchanged from those cleared under K121034.

    AI/ML Overview

    The provided text is a 510(k) summary for the ICONACY I-Hip, a hip joint prosthesis. The document focuses on showing substantial equivalence to a predicate device, specifically regarding a minor modification to the acetabular cup. It does not contain information about acceptance criteria and a study proving the device meets those criteria in the context of an AI/ML device.

    Therefore, I cannot extract the requested information regarding acceptance criteria, device performance, sample sizes, ground truth, expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance, as these are not discussed in this medical device submission.

    The document states: "Clinical testing was not required to establish equivalency of the device." and "Finite Element Analysis (FEA) confirmed that adding a third screw hole to the acetabular cups had little impact on cup stiffness." This indicates that the safety and effectiveness were established through non-clinical means (FEA) and comparison to a legally marketed predicate device, rather than a clinical study with performance metrics in the way your request implies for an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151307

    Validate with FDA (Live)

    Date Cleared
    2016-02-26

    (284 days)

    Product Code
    Regulation Number
    888.3353
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ICONACY I-Hip is indicated for the following conditions: (1) a severely painful and/or disabled hip joint as a result of osteoarthritis, traumatic arthritis, theumatoid arthritis, or congenital hip dysplasia, (2) avascular necrosis of the femoral head, (3) acute traumatic fracture of the femoral head or neck, (4) failed previous hip surgery including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement, (5) certain cases of ankylosis, (6) nonunions, correction of functional deformity, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.

    The ICONACY I-Hip consists of femoral stem and acetabular cup (i.e. shell) porous coated components intended for cementless, press-fit fixation.

    Device Description

    The ICONACY I-Hip consists of a collarless, tapered, forged titanium alloy femoral stem mated to a modular femoral head. This femoral construct articulates with an acetabular device assembly. The acetabular device assembly consists of a hemispherical titanium alloy cup (i.e., shell) coupled with a highly cross-linked ultra-high molecular weight polyethylene (HXL-UHMWPE) liner. Forty percent of the femoral stem is circumferentially coated with a titanium coating designed to attain a cementless, press-fit fixation. The acetabular cup is machined from forged Ti-6Al-4V ELI alloy. The cup has a threaded polar hole for insertion. The outer hemispheric surface of the cup has a titanium plasma spray coating for cementless, press-fit fixation. Titanium bone screws may be used for additional fixation. A titanium locking ring is fixed into grooves on the cup to engage grooves on the HXL-UHMWPE liner. Standard instrumentation is used to implant the device.

    The current submission adds ceramic femoral heads in 28, 32 and 36mm diameters and modified CS Acetabular Liners to the system. Compatible femoral stems, acetabular cups, locking rings, bone screws and apex hole plug were cleared in K121034, K131279, K133228 and K132572.

    AI/ML Overview

    The provided text is a 510(k) summary for the ICONACY™ I-Hip™ Total Hip System, which is a medical device. This document describes the device, its indications for use, and the basis for its substantial equivalence to predicate devices. However, it does not include information about AI/ML algorithms, studies proving device performance against acceptance criteria in the context of an AI/ML algorithm, or details about ground truth establishment, sample sizes for training/test sets, or expert adjudication.

    The "PERFORMANCE DATA" section states: "Pre-fatigue burst, fatigue, post-fatigue burst, pull-off, wear testing, wear testing under adverse conditions and biocompatibility testing confirm that the ICONACY I-Hip Ceramic Femoral Heads and CS Acetabular Liners meet pre-determined acceptance criteria and are expected to be safe and effective for the proposed indications."

    And "CLINICAL TESTING: Clinical testing was not required to determine substantial equivalence with the predicate devices."

    Therefore, I cannot provide the requested information for an AI/ML device because the document describes a mechanical medical device (hip implant) and not an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132572

    Validate with FDA (Live)

    Date Cleared
    2014-03-28

    (225 days)

    Product Code
    Regulation Number
    888.3358
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ICONACY I-Hip System is indicated for the following conditions: (1) a severely painful and/or disabled hip joint as a result of osteoarthritis, traumatic arthritis, rheumatoid arthritis, or congenital hip dysplasia, (2) avascular necrosis of the femoral head, (3) acute traumatic fracture of the femoral head or neck, (4) failed previous hip surgery including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement, (5) certain cases of ankylosis, (6) nonunions, correction of functional deformity, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.

    The ICONACY I-Hip System consists of femoral stem and acetabular cup (i.e. shell) porous coated components intended for cementless, press-fit fixation.

    Device Description

    The ICONACY I-Hip System consists of a collarless, tapered, forged titanium alloy femoral stem mated to a cobalt chrome alloy modular femoral head. This femoral construct articulates with an acetabular device assembly. The acetabular device assembly consists of a hemispherical titanium alloy cup (i.e., shell) coupled with a highly cross linked ultrahigh molecular weight polyethylene (HXL-UHMWPE) liner. Forty percent of the femoral stem is circumferentially coated with a titanium coating designed to attain a cementless, press-fit fixation. The acetabular cup is machined from forged Ti-6Al-4V ELI alloy. The cup has a threaded polar hole for insertion. The outer hemispheric surface of the cup has a titanium plasma spray coating for cementless, press-fit fixation. A titanium locking ring is fixed into grooves on the cup to engage grooves on the HXL-UHMVPE liner. Standard instrumentation is used to implant the device.

    The current submission adds angled face, lateralized and extended wall acetabular liners and an apex hole plug to the I-Hip Total Hip System. The compatible femoral heads, acetabular cups, locking rings and bone screws were cleared in K121034. Additional compatible acctabular cups and locking rings were cleared in K131279.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device, the ICONACY™ I-Hip™ System. This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, not for proving a device meets specific acceptance criteria through clinical or standalone performance studies like those common for AI/ML-enabled devices.

    Therefore, most of the requested information regarding acceptance criteria, study details, sample sizes, ground truth establishment, expert qualifications, and MRMC studies is not applicable to this document. The device is a total hip system, which is a physical implant, not an AI/ML-enabled device or software.

    Here's a breakdown of what can be extracted and what is not relevant:

    1. A table of acceptance criteria and the reported device performance

    • Not Applicable. This document does not describe specific numerical acceptance criteria typical for AI/ML performance metrics (e.g., sensitivity, specificity, AUC). Instead, it relies on demonstrating substantial equivalence to a predicate device.
    • Reported Device Performance: The only performance data explicitly mentioned is:
      • "Range of motion analyses were completed for the acetabular liners to ensure that adequate range of motion was maintained."
      • "Characterization and biocompatibility data for color additives is provided in a Master File."
      • No specific quantitative results from these analyses are provided in this summary.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable. No test set in the context of AI/ML evaluation is mentioned or performed. The demonstration of substantial equivalence is based on design, materials, function, and indications for use compared to predicate devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable. There is no "ground truth" establishment in the context of an AI/ML test set mentioned.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. No test set or adjudication method is described.

    5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. MRMC studies are relevant for evaluating AI-assisted diagnostic tools. This device is a physical orthopedic implant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This device is a physical implant, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not Applicable. Ground truth in the context of AI/ML validation is not applicable here.

    8. The sample size for the training set

    • Not Applicable. There is no AI/ML model, hence no training set.

    9. How the ground truth for the training set was established

    • Not Applicable. No training set or ground truth for it exists in this submission.

    Summary of Device Acceptance and Study:

    The ICONACY™ I-Hip™ System met the criteria for substantial equivalence to its predicate device (ICONACY I-Hip Total Hip System, K121034) by demonstrating similarities in design, materials, function, and indications for use.

    • Acceptance Criteria (for regulatory clearance): Substantial Equivalence to a legally marketed predicate device.
    • Study Proving Acceptance: The regulatory submission (510(k)) itself, which included comparisons of the device's characteristics to the predicate.
    • Performance Data Mentioned:
      • Range of motion analyses for acetabular liners to ensure adequate range of motion.
      • Characterization and biocompatibility data for color additives (referenced in a Master File).
    • Clinical Testing: Explicitly stated as "Clinical testing was not required for determining substantial equivalence with the predicate devices."

    In essence, the "study" for this type of device clearance is a comparative analysis against a predicate device based on technical specifications and existing regulatory precedents, not a performance study in the same vein as those for AI/ML-enabled devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K133228

    Validate with FDA (Live)

    Date Cleared
    2014-01-17

    (88 days)

    Product Code
    Regulation Number
    888.3358
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ICONACY I-Hip System is indicated for the following conditions: (1) a severely painful and/or disabled hip joint as a result of osteoarthritis, traumatic arthritis, theumatoid arthritis, or congenital hip dysplasia, (2) avascular necrosis of the femoral head, (3) acute traumatic fracture of the femoral head or neck, (4) failed previous hip surgery including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement, (5) certain cases of ankylosis, (6) nonunions, correction of functional deformity, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.

    The ICONACY I-Hip System consists of femoral stem and acetabular cup (i.e. shell) porous coated components intended for cementless, press-fit fixation.

    Device Description

    The ICONACY I-Hip System consists of a collarless, tapered, forged titanium alloy femoral stem mated to a cobalt chrome alloy modular femoral head. This femoral construct articulates with an acetabular device assembly. The acetabular device assembly consists of a hemispherical titanium alloy cup (i.e., shell) coupled with a highly cross linked ultra-high molecular weight polyethylene (HXL-UHMWPE) liner. Forty percent of the femoral stem is circumferentially coated with a titanium coating designed to attain a cementless, press-fit fixation. The acetabular cup is machined from forged Ti-6Al-4V ELI alloy. The cup has a threaded polar hole for insertion. The outer hemispheric surface of the cup has a titanium plasma spray coating for cementless, press-fit fixation. A titanium locking ring is fixed into grooves on the cup to engage grooves on the HXL-UHMWPE liner. Standard instrumentation is used to implant the device.

    The current submission adds stems with additional offsets to the I-Hip System. Compatible femoral heads, acetabular cups, acetabular liners, locking rings and bone screws were cleared in K121034 and K131279.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the ICONACY™ I-Hip™ System:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    Proximal Stem Fatigue (ISO 7206-6)Met pre-determined criteria
    Distal Stem Fatigue (ISO 7206-4)Met pre-determined criteria
    Range of Motion AnalysisDid not create a new worst case compared to the original I-Hip Femoral Stems.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document describes performance data conducted on the device components (femoral stems) and does not refer to a "test set" in the context of diagnostic or clinical evaluation. Instead, it refers to biomechanical testing. The sample sizes for these specific tests are not provided in the document. The data provenance is also not specified (e.g., country of origin, retrospective/prospective). It can be inferred that these were laboratory-based engineering tests.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    This information is not applicable as the described studies are biomechanical engineering tests on device components, not studies requiring expert interpretation of clinical data for ground truth establishment.

    4. Adjudication Method for the Test Set:

    This information is not applicable for the same reasons as #3.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size:

    No, an MRMC comparative effectiveness study was not done. The document explicitly states: "CLINICAL TESTING: Clinical testing was not required for determining substantial equivalence with the predicate devices."

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

    This information is not applicable. This device is a physical orthopedic implant, not an AI algorithm or software. Therefore, the concepts of "standalone performance" or "human-in-the-loop performance" do not apply.

    7. The Type of Ground Truth Used:

    The "ground truth" for the performance data in this context refers to established engineering standards and specifications (ISO 7206-6, ISO 7206-4) for mechanical properties and the performance of the predicate device for range of motion.

    8. The Sample Size for the Training Set:

    This information is not applicable as the device is a physical orthopedic implant, not a machine learning model.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable for the same reasons as #8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K131279

    Validate with FDA (Live)

    Date Cleared
    2013-08-09

    (95 days)

    Product Code
    Regulation Number
    888.3358
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ICONACY I-Hip System is indicated for the following conditions: (1) a severely painful and/or disabled hip joint as a result of osteoarthritis, traumatic arthritis, theumatoid arthritis, or congenital hip dysplasia, (2) avascular necrosis of the femoral head, (3) acute traumatic fracture of the femoral head or neck, (4) failed previous hip surgery including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement, (5) certain cases of ankylosis, (6) nonunions, correction of functional deformity, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.

    The ICONACY I-Hip System consists of femoral stem and acetabular cup (i.e. shell) porous coated components intended for cementless, press-fit fixation.

    Device Description

    The ICONACY I-Hip consists of a collarless, tapered, forged titanium alloy femoral stem mated to a cobalt chrome alloy modular femoral head. This femoral construct articulates with an acetabular device assembly. The acetabular device assembly consists of a hemispherical titanium alloy cup coupled with a highly cross linked ultra-high molecular weight polyethylene (HXL-UHMWPE) liner. Forty percent of the femoral stem is circumferentially coated with a titanium coating designed to attain a cementless, press-fit fixation. The acetabular cup is machined from forged Ti-6Al-4V ELI alloy. The cup has a threaded polar hole for insertion. The outer hemispheric surface of the cup has a titanium plasma spray coating for cementless, press-fit fixation. A titanium locking ring is fixed into a groove on the cup to engage a groove on the HXL-UHMWPE liner. Standard instrumentation is used to implant the device.

    The current submission is for modifications to the acetabular cups and locking rings cleared in K121034, and the addition of spiked and finned acetabular cups in the same sizes and materials as those cleared in K120134. The compatible femoral heads, acetabular liners and bone screws are unchanged from those cleared in K121034.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the ICONACY™ I-Hip™ System, based on the provided document:

    This 510(k) submission (K131279) is for modifications to existing components and the addition of new acetabular cups, not for an entirely new device. Therefore, the "study" described focuses on demonstrating substantial equivalence to a predicate device rather than presenting a clinical trial for novel device performance.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (What was measured)Reported Device Performance (Results)
    Mechanical Performance of Modified Designs
    Push-out testing (of modified acetabular cups/locking rings)Performed, and results contributed to a determination of substantial equivalence (specific numerical values are not provided in this summary document).
    Lever-out testing (of modified acetabular cups/locking rings)Performed, and results contributed to a determination of substantial equivalence (specific numerical values are not provided in this summary document).
    Torque-out testing (of modified acetabular cups/locking rings)Performed, and results contributed to a determination of substantial equivalence (specific numerical values are not provided in this summary document).
    Equivalence to Predicate Device (K121034)
    Similarity in designDemonstrated to be similar.
    Similarity in materialsDemonstrated to be similar.
    Similarity in functionDemonstrated to be similar.
    Similarity in indications for useDemonstrated to be similar.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set" in the context of a traditional clinical study with patient samples. The performance data is derived from in vitro mechanical testing of the device components.

    • Sample Size for Mechanical Testing: Not explicitly stated (e.g., number of cups tested, number of forces applied).
    • Data Provenance: This is an in vitro engineering study, not human or animal data. The origin would be the testing facility of ICONACY Orthopedic Implants, LLC.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. The "ground truth" for this type of submission is based on engineering principles and established mechanical testing protocols. There is no mention of external human experts establishing a ground truth for the in vitro test results in this 510(k) summary.

    4. Adjudication Method for the Test Set

    Not applicable, as this was an in vitro mechanical testing study, not a clinical study requiring adjudication of patient outcomes or imaging interpretations. The determination of "substantial equivalence" based on these tests would be made internally by the sponsor and reviewed by the FDA.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not a study involving AI, human readers, or medical imaging. It is a submission for an orthopedic implant.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithmic device.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this submission's performance data is based on mechanical testing standards and engineering specifications designed to ensure the modified components are as robust and perform similarly to the predicate device. The goal is to demonstrate that the physical properties and performance under stress (push-out, lever-out, torque-out) meet acceptable thresholds for orthopedic implants of this type and are comparable to the previously cleared predicate.

    8. The Sample Size for the Training Set

    Not applicable. There is no concept of a "training set" in this 510(k) submission, as it does not involve machine learning or AI.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for this device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121034

    Validate with FDA (Live)

    Device Name
    IOI TOTAL HIP
    Date Cleared
    2012-09-04

    (152 days)

    Product Code
    Regulation Number
    888.3358
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ICONACY I-Hip system is indicated for the following conditions: (1) a severely painful and/or disabled hip joint as a result of osteoarthritis, traumatic arthrius, theumatoid arthritis, or congenital hip dysplasia, (2) avascular necrosis of the femoral head, (3) acute traumatic fracture of the femoral head or neck, (4) failed previous hip surgery including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement, (5) certain cases of ankylosis, (6) nonunions, correction of functional deformity, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.

    The ICONACY I-Hip system consists of femoral stem and acetabular cup (i.e., shell) porous coated components intended for cementless, press-fit fixation

    Device Description

    The ICONACY I-Hip consists of a collarless, tapered, forged titanium alloy femoral stem, mated to a cobalt chrome modular femoral head. This femoral construct articulates with an acetabular device assembly. The acetabular device assembly consists of a hemispheric titanium alloy cup (i.e., shell) coupled with a highly cross linked ultra-high molecular weight polyethylene (HXL-UHMWPE) liner. Forty percent of the femoral stem is circumferentially coated with a titanium coating designed to attain a cementless, press fit fixation. The acetabular cup is machined from forged Ti-6Al-4V ELI alloy. The cup has a threaded polar hole for insertion and two screw holes to permit the use of screws for adjunct fixation. The outer hemispheric surface of the cup has a titanium plasma spray coating for cementless, press fit fixation. A titanium locking ring is fixed into a groove on the cup to engage the groove in the HXL-UHMWPE liner. Standard instrumentation is used to implant the device.

    AI/ML Overview

    The provided text describes a 510(k) submission for the ICONACY I-Hip System, a total hip replacement device. It focuses on demonstrating substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria through clinical studies. Therefore, much of the requested information about device performance, study details, and ground truth establishment is not available in the provided text.

    Here's an breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Nonclinical Testing:Outcome: Demonstrated substantial equivalency to predicate devices.
    Characterization of HXL-UHMWPEPerformed
    Testing of the locking mechanism between the cup and linerPerformed
    Range of motion analysisPerformed
    Impingement testingPerformed
    WearPerformed
    Femoral head pull-off testingPerformed
    Fatigue (endurance limits) for the femoral stemPerformed

    2. Sample Size Used for the Test Set and Data Provenance

    Not applicable. No clinical test set or human subject data was used. The evaluation was based on nonclinical (bench) testing.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Experts

    Not applicable. No expert-established ground truth for a test set was used, as no clinical testing was performed.

    4. Adjudication Method for the Test Set

    Not applicable. No clinical test set or adjudication process was described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No. A multi-reader multi-case comparative effectiveness study was not performed. The submission explicitly states "Clinical Testing: Clinical testing was not required for determining substantial equivalency with the predicate devices."

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical orthopedic implant, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth" for demonstrating substantial equivalence was derived from nonclinical (bench) testing results (e.g., measurements of material properties, mechanical performance under various loads, wear rates) compared against established performance characteristics of the predicate devices.

    8. The Sample Size for the Training Set

    Not applicable. There was no "training set" in the context of an AI/ML algorithm. The device's design and manufacturing processes are informed by engineering principles and existing knowledge of predicate devices.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. As noted above, there was no "training set" for an algorithm. The design and manufacturing relied on established engineering standards and performance data from predicate devices.