Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K172821

    Validate with FDA (Live)

    Date Cleared
    2018-06-11

    (266 days)

    Product Code
    Regulation Number
    872.3640
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This Product is to be used in mandible or maxilla-arch by dental surgery, and to support artificial teeth and maxilla to restore patient's chewing function.

    This may be accomplished by either a two-stage surgical procedure or a single surgical procedure. If a single surgical procedure is used, single or multiple implants may be inserted (type I, II or III bone) provided good initial stability (> 40 Ncm) is achieved. Intended for delayed loading.

    Device Description
    • (1) This Product is designed to be used in dental surgery in mandible or maxilla-arch, where has missing teeth, to have dental crown, fixation-bridge, over-denture installed afterwards, and to support and restore patient's chewing function.
    • (2) This implant (Ti-one 101 TS dental implants (Fixture)) is made of grade 4 pure titanium, featuring sandblasting + acid etching surface treatment (SLAffinity).
    • (3) Healing screw is one-piece abutment made of grade 4 titanium.
    • (4) Abutment is two-piece abutment made of grade 4 titanium. Its surface was anodized.
    • (5) Label attached to the surface of outer packaging indicates specification and model number.

    The range of diameter for implant (fixture) is provided as below:
    Ø Diameter: 3.5, 4.0, 4.5, 5.0 mm
    Length: 7.0, 8.0, 9.5, 11.0, 12.5, 14.0 mm

    All diameters are available in all lengths, other than the 3.5mm diameter which is not available in the 7.0mm length.

    The range of diameters and angulations for each screw model and abutment model are provided as below:
    Healing Screw: Ø 4.3 mm, G/H 1.0, 3.0, 4.0 mm, Height (H) 3.5, 5.0, 7.0 mm
    Angled Abutment: Ø 4.5, 5.5 mm, G/H 3.5 mm, V/H 11.0 mm, Length (L) 14.13, 14.15, 14.20 mm, Angulation range 12.5°, 20.0°, 27.5°

    AI/ML Overview

    The provided document is a 510(k) summary for the "Ti-one 101 TS Dental Implant System." It does not describe an AI or software as a medical device (SaMD). Instead, it details a traditional medical device (dental implant system) and its non-clinical testing for substantial equivalence to predicate devices. Therefore, many of the requested categories related to AI/SaMD studies are not applicable.

    Here's an analysis of the provided information, focusing on what is available:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test CategoryAcceptance Criteria (General)Reported Device Performance
    SterilizationCompliance with ISO11737-1:2006, ISO11737-2:2009, ISO 11137-2:2013Proposed device complies with all listed ISO standards.
    Shelf LifeCompliance with ASTM F1980:2016, ASTM F1929:2012, ISO 11607-1:2006, ASTM F-88 / F88M:2015, ISO 11737-2:2009Proposed device complies with all listed ASTM and ISO standards.
    BiocompatibilityCompliance with ISO 10993 series (various parts), USP, and OECD guidelines for predicate device; USP and AAMI for non-pyrogenic test.Predicate Device (K110425): Complies with all listed ISO, USP, and OECD requirements for cytotoxicity, intracutaneous reactivity, skin sensitization, acute systemic toxicity, pyrogenicity, in vitro mutation/aberration tests, and 14-day/90-day toxicity tests.
    Proposed Device (Non-pyrogenic): Complies with USP 40:2017 <85>, USP 40:2017 <161>, AAMI ST72:2011.
    Performance TestingCompliance with ISO 14801:2007 and ASTM F543-13:2013Reported as meeting "pre-defined acceptance criteria." (Specific details of torque, fatigue, or SEM/EDS results are not provided beyond this statement).

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size: Not specified in the document. The document refers to "test reports" for each test category but does not include the number of units tested.
    • Data Provenance: The biocompatibility tests for the predicate device (K110425) are leveraged from the same manufacturer, Hung Chun Bio-S Co., Ltd. The document does not specify the country of origin of the data or whether the tests were retrospective or prospective, though performance and biocompatibility testing of medical devices are typically prospective bench or in-vitro tests conducted in a lab setting.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. This device is a physical dental implant, not an AI or SaMD requiring expert interpretation for ground truth establishment. Its performance is evaluated through standardized mechanical, biological, and sterilization tests.

    4. Adjudication method for the test set

    • Not Applicable. As above, the device is evaluated through objective, standardized tests, not through expert adjudication of data.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This study is for a physical medical device (dental implant system), not an AI or SaMD. No human readers or AI assistance are involved in its evaluation as described.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is not an algorithm or SaMD.

    7. The type of ground truth used

    • For the non-clinical tests (sterilization, shelf life, biocompatibility, performance), the "ground truth" is defined by the specific requirements and passing criteria of the referenced international/national standards (e.g., ISO, ASTM, USP, AAMI, OECD guidelines). The device's performance is measured against these objective criteria.

    8. The sample size for the training set

    • Not Applicable. This is a physical medical device, not an AI/SaMD. There is no concept of a "training set" in the context of this 510(k) summary.

    9. How the ground truth for the training set was established

    • Not Applicable. As above, there is no training set for this device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1