Found 7 results

510(k) Data Aggregation

    K Number
    K220283

    Validate with FDA (Live)

    Date Cleared
    2022-05-02

    (90 days)

    Product Code
    Regulation Number
    882.5890
    Age Range
    0 - 21
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SM9009:
    TENS(1-7,9,10,14,16,19,21,22,24):
    To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
    PMS(8,11,12,13,15,17,18,20,23):
    It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.

    SM9021:
    TENS(3-6,8-11,15,16,19,21):
    To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
    PMS(1,2,7,12,13,14,17,18,20,22,23,24):
    It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.

    SM9128H:
    TENS(2-6,9,11,14,16,19,21,22,24):
    To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
    PMS(1,7,8,10,12,13,15,17,18,20,23):
    It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.

    SM9015:
    TENS(2-4.8.9.10.12.13.14.15.17.18.19.20.22):
    To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
    PMS(1.5,6.7,11,16,21,23,24):
    It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.

    SM9055:
    TENS(1-8,10,11,13,14,16,19,21-25,30,31,35,36):
    To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
    PMS(9,12,15,17,18,20,26,27,28,29,32,33,34):
    It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.

    SM9033:
    TENS(1-16,19,21,22):
    To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
    PMS(17,18,20,23,24):
    It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.

    Device Description

    Not Found

    AI/ML Overview

    The provided text is a 510(k) premarket notification document from the FDA for TENS & PMS devices. It outlines the regulatory clearance process and includes "Indications for Use" for several models. However, it does not contain any information regarding acceptance criteria, device performance studies, sample sizes, ground truth establishment, or expert reviews for a medical device.

    Therefore, I cannot provide the requested information. The document focuses on regulatory approval based on substantial equivalence to predicate devices, not on detailed performance study results of the new device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220152

    Validate with FDA (Live)

    Date Cleared
    2022-04-19

    (90 days)

    Product Code
    Regulation Number
    882.5890
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SM9089:
    TENS(3,4,5,6,8,15,19,21):
    To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
    PMS(1,2,7,9,10,11,12,13,14,16,17,18,20,22,23,24):
    It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.

    SM9587:
    TENS(3.4.5.6 8.11.15.16.19.21.22):
    To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
    PMS(1,2,7,9,10,12,13,14,17,18,20,23,24):
    It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.

    SM9592:
    TENS(2,3,4,5,7,9,10,19,21,22,24):
    To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
    PMS(1,6,8,11,12,13,14,15,16,17,18,20,23):
    It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.

    SM9555:
    TENS(3-6, 13,14,19-21,23):
    To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
    PMS(1,2,7-12,15-18,22,24):
    It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.

    Device Description

    Not Found

    AI/ML Overview

    The provided document is an FDA 510(k) clearance letter for Transcutaneous Electrical Nerve Stimulators (TENS) and Powered Muscle Stimulators (PMS) devices. It confirms that the devices are substantially equivalent to legally marketed predicate devices.

    However, this document does not contain information about the acceptance criteria or a study that proves the device meets specific acceptance criteria. The letter primarily addresses regulatory clearance based on substantial equivalence to existing devices, general controls provisions of the Federal Food, Drug, and Cosmetic Act, and specifies the Indications for Use for various models.

    For a device's performance to be evaluated against acceptance criteria, a clinical study or performance testing data would typically be required to be referenced in the submission. This information is usually found in the 510(k) summary or the detailed technical documentation submitted to the FDA, which is not part of this public clearance letter.

    Therefore, I cannot provide the requested details using only the provided text. The document is a regulatory approval, not a performance study report.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220150

    Validate with FDA (Live)

    Date Cleared
    2022-02-17

    (29 days)

    Product Code
    Regulation Number
    882.5890
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TENS (3-6, 13, 14, 20, 21, 23): To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities. PMS (1, 2, 7-12, 15-19, 22, 24): It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.

    Device Description

    Not Found

    AI/ML Overview

    I am sorry, but the provided text does not contain the information required to answer your request. The document is an FDA 510(k) clearance letter for a TENS & PMS device. It details the regulatory clearance for the device but does not include any study data, acceptance criteria, or information regarding performance testing, sample sizes, ground truth establishment, or expert involvement in product validation.

    Therefore, I cannot provide a table of acceptance criteria and reported device performance or any of the other requested details about a study proving the device meets those criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201354

    Validate with FDA (Live)

    Device Name
    TENS & PMS
    Date Cleared
    2020-12-11

    (204 days)

    Product Code
    Regulation Number
    890.5850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TENS(1,3,4,5,6,21): To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
    PMS(2,7,8,9,10,11,12,13,14,15,16,17,18,19,20,22,23,24): It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
    Model: SM9126
    TENS(3,4,5,6,11,12,16): To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
    PMS(1,2,7,8,9,10,13,14,15): It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
    Model: SM9186
    TENS(3,4,7,8,11): To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
    PMS(1,2,5,6,9,10,12): It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.
    Model: SM9196
    TENS(1015): To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
    PMS(1
    9): It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.

    Device Description

    TENS & PMS is a portable and DC 3.7V battery powered multifunction device with multiple models, offering both Transcutaneous Electrical Nerve Stimulation (TENS) and Powered Muscle Stimulation (PMS) qualities in one device. The device is equipped with accessories of electrode pads, electrode cables, a battery charger, and one USB cable. The electrode cables are used to connect the pads to the device; the USB cable is used to connect the charger and the built-in lithium battery. All accessories, including USB cables, electrode pads, electrode cables, chargers can only be changed or replaced by a qualified person.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a TENS & PMS device. This submission aims to demonstrate substantial equivalence to an already legally marketed predicate device (K121719). The document focuses on comparing the technical characteristics and performance of the new device models (SM9079, SM9126, SM9186, SM9196) against the predicate device.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria are implicitly defined by the safety and performance parameters of the legally marketed predicate device (K121719) and compliance with various recognized standards. The reported device performance is demonstrated through the comparative tables (Table 1, 2, 3, 4) where the new device's parameters are listed alongside the predicate device's, and through the statement of compliance with specific IEC and ANSI/AAMI standards.

    Since the tables in the document are already in a comparative format, I will summarize the key parameters and their reported performance for one of the new models (SM9079) against the predicate. The "Remark" column in the original tables already serves as a concise statement of whether the new device meets or is similar to the predicate.

    Table: Acceptance Criteria and Reported Device Performance (Example: Model SM9079 vs. Predicate K121719)

    Parameter / Acceptance Criteria (Derived from Predicate)Reported Device Performance (New Device SM9079)Meets Criteria?Notes
    General Characteristics
    ManufacturerHong Qiangxing (Shenzhen) Electronics LimitedYesSame
    Intended Use (TENS)Temporary relief of pain for specific body parts due to strainYesSame (Expanded list of programs, but comparable)
    Intended Use (PMS)Stimulate healthy muscles to improve performanceYesSame (Expanded list of programs, but comparable)
    Type of UseOTCYesSame
    Power SourceDC 3.7V lithium batteryYesSame
    Method of Line Current IsolationType BFYesSame
    Patient Leakage Current (Normal Condition)< 10μAYesSame
    Patient Leakage Current (Single Fault Condition)< 50μAYesSame
    Average DC current through electrodes< 0.01μAYesSame
    Number of Output Modes24SimilarMore treatment programs, all comparable to predicate's mode 3. No new safety/effectiveness issues.
    Number of Output Channels2YesSame
    Synchronous or Alternating Channel OutputAlternatingSimilarDifferent from predicate's synchronous, but same technology and treatment effect. No new safety/effectiveness issues.
    Method of Channel IsolationVoltage transformer IsolationYesSame
    Regulated Current or VoltageVoltage controlYesSame
    Software/Firmware/Microprocessor ControlSoftwareYesSame
    Automatic Overload Trip?NoYesSame
    Automatic No-Load Trip?NoYesSame
    Automatic Shut Off?YesYesSame
    User Override Control?YesYesSame
    Indicator (On/Off Status)YesYesSame
    Indicator (Low Battery)YesYesSame
    Indicator (Voltage/Current Level)YesYesSame
    Timer Range10 ~ 60 minutes, 10 min/stepYesSame
    BiocompatibilityAll user directly contacting materials comply with ISO10993-5 and ISO10993-10 requirements.YesSame
    AccessoriesSelf-adhesive electrodes, electrode wires, Battery charger, USB cableYesSame
    Physical Characteristics
    Weight102g±5gSimilarDifference in weight and dimensions noted, but deemed not to raise safety/effectiveness issues.
    Dimensions6012312mmSimilarDifference in weight and dimensions noted, but deemed not to raise safety/effectiveness issues.
    Housing Materials & ConstructionABSYesSame
    Output Specifications (Max Values from Table 5 for SM9079 vs. Predicate Mode 3/1/2)
    WaveformPulsed, symmetric, biphasicYesSame
    ShapeRectangularYesSame
    Max Output Voltage (@500Ω)42V±20%SimilarPredicate 42V±10%. Differences in pulse width, frequency, period allowed as parameters don't exceed safety limits and passed IEC 60601-2-10 tests.
    Max Output Current (@500Ω)84mA±20%SimilarPredicate 84mA±10%. Differences in pulse width, frequency, period allowed as parameters don't exceed safety limits and passed IEC 60601-2-10 tests.
    Pulse Width190µs (max)SimilarPredicate 100µs. Differences in pulse width, frequency, period allowed as parameters don't exceed safety limits and passed IEC 60601-2-10 tests.
    Pulse Period40-1070ms (range)SimilarPredicate 9.3~850ms. Differences in pulse width, frequency, period allowed as parameters don't exceed safety limits and passed IEC 60601-2-10 tests.
    Max Pulse Frequency0.935-25Hz±10% (range)SimilarPredicate 110Hz. Differences in pulse width, frequency, period allowed as parameters don't exceed safety limits and passed IEC 60601-2-10 tests.
    Net Charge (per pulse)0µC @500Ω; Balanced waveformYesSame
    Max Phase Charge (@500Ω)13.30µC@500ΩSimilarPredicate 16.8µC@500Ω. Differences in these parameters allowed as they don't exceed safety limits and passed IEC 60601-2-10 tests.
    Max Current Density (@500Ω)0.065mA/cm²YesMeets max current density <2mA/cm² (predicate value not given explicitly here, but stated in Note 4 for SM9126).
    Max Average Power Density (@500Ω)2.64mW/cm²YesMeets max average power density <0.25W/cm² (predicate value not given explicitly here, but stated in Note 4 for SM9126).
    Compliance with Standards
    ANSI AAMI ES60601-1YesYesSame
    IEC 60601-2-10YesYesSame
    ANSI AAMI IEC 60601-1-2YesYesSame
    IEC 62133YesYesSame
    ANSI AAMI HA60601-1-11YesYesSame
    21 CFR 898 (if applicable)YesYesSame

    2. Sample size used for the test set and the data provenance

    The document explicitly states that the new devices have "passed testings according to the following standards" and that the "conclusion drawn from the testings are that the new devices are substantially equivalent to the predicate device." These "testings" refer to compliance with various medical device standards (ANSI AAMI ES60601-1, IEC 60601-2-10, ANSI AAMI IEC 60601-1-2, IEC 62133, ANSI AAMI HA60601-1-11).

    However, the document does not provide details about a clinical study or a specific test set, its sample size, or data provenance (e.g., country of origin, retrospective/prospective). The submission is based on a comparison of technical specifications and performance against a predicate device and adherence to recognized standards, rather than a separate clinical trial with a defined sample size. The testing appears to be functional and electrical safety testing as per the listed standards.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Given that a clinical study or human-in-the-loop performance evaluation with a test set requiring expert ground truth is not described in the provided text, information on the number and qualifications of experts for establishing ground truth is not available. The evaluation is based on technical specifications and compliance with standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Since no clinical test set requiring human adjudication is mentioned, there is no information provided on any adjudication method.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The device described is a Transcutaneous Electrical Nerve Stimulator (TENS) and Powered Muscle Stimulator (PMS). It is a physical therapy device and not an AI-assisted diagnostic or imaging interpretation device. Therefore, a Multi Reader Multi Case (MRMC) comparative effectiveness study focusing on human readers improving with AI assistance would be irrelevant and is not mentioned in this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Again, this is a physical therapy device, not an algorithm, so no standalone algorithm performance evaluation is applicable or mentioned.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For this type of device, the "ground truth" for demonstrating substantial equivalence and safety/effectiveness relies on:

    • Compliance with recognized electrical safety and performance standards: This includes specific limits for voltage, current, pulse characteristics, leakage current, and electromagnetic compatibility.
    • Comparison of technical specifications with a legally marketed predicate device: The predicate device itself has established safety and effectiveness. The "ground truth" is that the predicate device is safe and effective when used as intended, and the new device demonstrates similar performance and characteristics to be considered equivalent.
    • Biocompatibility testing: Compliance with ISO10993-5 and ISO10993-10 for materials directly contacting the user.

    There is no mention of expert consensus, pathology, or outcomes data from a clinical trial for this specific submission. The "ground truth" is established by adherence to existing regulatory frameworks and engineering standards.

    8. The sample size for the training set

    As this is not an AI/machine learning device, the concept of a "training set" is not applicable and therefore, no sample size for a training set is provided.

    9. How the ground truth for the training set was established

    As the concept of a training set is not applicable, no information is provided on how its ground truth would be established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190009

    Validate with FDA (Live)

    Date Cleared
    2019-12-02

    (333 days)

    Product Code
    Regulation Number
    882.5890
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Model SM9187:

    TENS (Mode 1, 3, 4, 5, 6):

    To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities. PMS (Mode 1, 2, 3, 4, 5, 6):

    It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.

    Model SM9180L & SM9180S:

    TENS (Mode 1. 3. 4. 5. 6):

    To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, back due to strain from exercise or normal household work activities.

    PMS (Mode 1, 2, 3, 4, 5, 6):

    It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.

    Device Description

    Sunmas TENS & PMS (Model SM9187, SM9180L & SM9180S) is a portable battery powered device with a wireless feature for over-the-counter use, offering both Transcutaneous Electrical Nerve Stimulation (TENS) and Powered Muscle Stimulation (PMS) qualities in one device. Sunmas TENS & PMS has totally 6 modes, applying electrical current to electrodes on a patient's skin to relieve pain or build up muscle. Due to different electrode shapes and sizes, Model SM9187 can be used on the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (legs), while Model SM9180L and SM9180S can be used on the shoulder, back and abdomen. Model SM9187 is designed to be controlled by a mobile phone application program (APP) named “Dr.Stim”, through Bluetooth 4.0 technology (Frequency: 2402.0 - 2480.0MHz). The APP is developed by Hong Qiangxing (Shenzhen) electronics Limited. SM9187 consists of a round main unit, two sets of self-adhesive electrode pad, a mobile application (APP), an AC charger and a USB cable. Model SM9180L and SM9180S both have a butterfly-shaped main unit (Frequency: 2450MHz) and can be remotely controlled through wireless technology. SM9180L has a black remote control with an electronic display screen which is powered by two AAA alkaline batteries, while Model SM9180S has a white remote control powered by a CR2032 coin battery. The radio frequency of remote control (both black and white) is 2405.0 - 2470.0MHz. Model SM9180L and SM9180S consists of a unit, two pairs of electrode gels, a remote control (black or white), an AC charger and a USB cable.

    AI/ML Overview

    This document is a 510(k) Pre-Market Notification for a Transcutaneous Electrical Nerve Stimulator (TENS) and Powered Muscle Stimulator (PMS) device. The purpose of this notification is to demonstrate that the new device, Sunmas TENS & PMS (Models SM9187, SM9180L, SM9180S), is substantially equivalent to legally marketed predicate devices.

    Here's an analysis of the provided information regarding acceptance criteria and supporting studies, as requested:

    1. A table of acceptance criteria and the reported device performance

    The provided document doesn't explicitly present a table of "acceptance criteria" in the sense of predefined performance thresholds for a clinical study. Instead, it compares the technical specifications and safety standards compliance of the new device with those of its predicate devices to demonstrate substantial equivalence. The implicit "acceptance criteria" are that the new device's specifications and safety performance are comparable to or meet the standards of the predicate devices.

    Below is a table summarizing key comparative performance metrics, drawn directly from the "Substantial Equivalence to Predicate device" tables (Table 1 and Table 2) within the document. The "Reported Device Performance" for the new device is its own specifications, and the "Acceptance Criteria" are implied by the predicate device's specifications and the statement that the new device is "safe and effective as the predicate devices cited above."

    ParameterNew Device (SM9187) Reported PerformancePredicate Device 1 (K121719) Implied Acceptance CriteriaPredicate Device 2 (K143430) Implied Acceptance Criteria
    Number of Output Modes667
    Number of Output channels121
    WaveformBiphasic rectangular, Monophasic rectangularBiphasic rectangular, Monophasic rectangularBiphasic rectangular, Monophasic rectangular
    Patient Leakage Current (Normal Condition)< 10µA< 10µA< 10µA
    Patient Leakage Current (Single Fault Condition)< 50µA< 50µA< 50µA
    Average DC Current (No pulses)< 0.01µA< 0.01µA< 0.01µA
    Regulated ControlVoltage controlVoltage controlVoltage control
    Max Output Voltage @500Ω47.6V±10%42V±10%68V
    Max Output Current @500Ω95.2mA±10%84mA±10%133mA
    Pulse width100µs100µs50-250µs
    Max Pulse Frequency105.5Hz110Hz1-150Hz
    Net Charge (per pulse) @500Ω0µC / 9.6µC0µC0µC / max 10.01µC
    Max Phase Charge @500Ω19.04µC16.8µC20.02µC
    Max Average Current1.004mA0.924mA3.0375mA
    Max Current Density @500Ω0.11mA/cm² (19cm²)0.462mA/cm² (4cm²)0.066mA/cm²
    Max Average Power Density @500Ω2.52mW/cm² (19cm²)9.702mW/cm² (4cm²)2.66mW/cm²
    Wireless ControlYes (Bluetooth, APP)NoYes (Bluetooth, APP)

    Note: The table for models SM9180L & SM9180S showed similarly comparable specifications to their predicate devices, with particular attention to wireless control and power sources. For brevity, only the first new device model (SM9187) is fully detailed above, but the same principle of demonstrating comparable specifications applies across all models.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not describe a clinical test set or patient data for performance evaluation. The evaluation is based on a comparison of technical specifications and compliance with recognized safety standards. There is no mention of a "sample size" in terms of patients or data, nor the provenance of such data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    Since there was no clinical test set or patient data involved in a performance evaluation requiring expert interpretation or "ground truth" establishment, this information is not applicable and not provided.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    As no clinical test set or patient data requiring adjudication was used, this information is not applicable and not provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The device described is a Transcutaneous Electrical Nerve Stimulator (TENS) and Powered Muscle Stimulator (PMS). It is a physical therapy device and does not involve AI assistance for "human readers" or image interpretation. Therefore, an MRMC comparative effectiveness study in this context is not applicable, and no such study was reported.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This device is not an algorithm or AI system. It is a physical medical device that delivers electrical stimulation. Therefore, the concept of "standalone (algorithm only)" performance does not apply.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Given the nature of the device and the submission type (demonstrating substantial equivalence through technical specifications and safety standards compliance), no "ground truth" in the context of clinical disease manifestation or diagnostic interpretation was used. The "truth" in this submission lies in compliance with electrical safety standards and the device's physical and electrical properties matching established safe and effective parameters of predicate devices.

    8. The sample size for the training set

    As this is not an AI/ML device, there is no "training set" in the conventional machine learning sense.

    9. How the ground truth for the training set was established

    Since there is no training set, this information is not applicable.

    Study Proving Acceptance Criteria:

    The "study" demonstrating that the device meets "acceptance criteria" (understood as achieving substantial equivalence to predicate devices and complying with relevant safety standards) is a series of technical tests and evaluations against recognized voluntary standards. The document explicitly lists the following standards to which the new device has been tested and found compliant:

    • ANSI AAMI ES60601-1: 2005/(R)2012 And A1:2012, C1:2009/(R)2012 And A2:2010/(R)2012 (Consolidated Text) - Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance.
    • IEC 60601-2-10 Edition 2.1 2016-04 - Medical Electrical Equipment - Part 2-10: Particular Requirements For The Basic Safety And Essential Performance Of Nerve And Muscle Stimulators.
    • ANSI AAMI IEC 60601-1-2:2014 - Medical Electrical Equipment -- Part 1-2: General Requirements For Basic Safety And Essential Performance -- Collateral Standard: Electromagnetic Disturbances -- Requirements And Tests.
    • IEC 62133 Edition 2.0 2012-12 - Secondary Cells And Batteries Containing Alkaline Or Other Non-Acid Electrolytes - Safety Requirements For Portable Sealed Secondary Cells, And For Batteries Made From Them, For Use In Portable Applications.
    • IEC 60601-1-11 Edition 2.0 2015-01 - Medical Electrical Equipment - Part 1-11: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Requirements For Medical Electrical Equipment And Medical Electrical Systems Used In The Home Healthcare Environment.
    • FCC 47 CFR Part 15 - Radio Frequency Devices.
    • AAMI TIR69: 2017 - Technical Information Report Risk Management Of Radio-Frequency Wireless Coexistence For Medical Devices And Systems.
    • ANSI IEEE C63.27-2017 - American National Standard For Evaluation Of Wireless Coexistence.

    These tests likely involved engineering evaluations, bench testing, and potentially electromagnetic compatibility (EMC) testing to ensure the device's electrical safety, functional performance within specified parameters, and safe operation in its intended environment, similar to the predicate devices. The successful completion of these tests provides the evidence that the device meets the implied acceptance criteria for safety and performance as a TENS/PMS device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183154

    Validate with FDA (Live)

    Device Name
    SM Electrodes
    Date Cleared
    2019-03-26

    (132 days)

    Product Code
    Regulation Number
    882.1320
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SM Electrodes are intended to transmit electrical current to patient skin for use with transcutaneous electrical stimulation devices. Some common types of the stimulation devices include, but are not limit to TENS (Transcutaneous Electrical Nerve Stimulation) and EMS (Electrical Muscular Stimulation). The electrode is for OTC (Over-The -Counter) or Prescription use.

    Device Description

    SM Electrode pad is used as an accessory to the TENS or EMS device unit, which transmits electrical current to patient skin. The electrical current of electrode pad is first transmitted via the snap button then transmitted to the conductive gel which is adhered to patient skin.

    SM Electrode Pad is composed of a top cover material, connector snap button, conductive carbon film, conductive hydrogel media, and a carrier line.

    It is non-sterile and intended for single adult patient multiple application use. The SM Electrode has various shapes and sizes.

    To connect with a nerve or muscle stimulator, this SM Electrode Pad has male snap button. The electrical pad connecting two separated gels together through a non-woven fabric has a pair of male snap button and the diameter is 3mm. The electrical pad with only one male snap button are to be used in pairs and the diameter of snap button is 4mm.

    When not in use, the hydrogel face is covered by a PET(polyethylene terephthalate) carrier line.

    AI/ML Overview

    The provided text describes the submission for K183154, "SM Electrodes," a cutaneous electrode. It details non-clinical testing performed to establish substantial equivalence to predicate devices.

    Here's the information broken down as requested:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria / TestReported Device Performance
    Shelf life testing (ASTM F1980:2016)Device complies with the standard. (Specific results like duration or remaining strength are not detailed, but compliance implies meeting the established criteria for shelf life.)
    Biocompatibility (ANSI AAMI ISO 10993-5:2009/(R) 2014) - In Vitro CytotoxicityDevice complies with the standard, indicating an absence of unacceptable cytotoxic effects.
    Biocompatibility (ISO 10993-10 Third Edition 2010-08-01) - Irritation and Skin SensitizationDevice complies with the standard, indicating an absence of unacceptable irritation and skin sensitization potential.
    Contact Impedance Test (per ANSI AAMI IEC 60601-2-2:2009, 201.15.101.6)Test results demonstrated compliance with the standard (specific numerical values for impedance are not provided, but compliance is stated).
    Adhesion Test (per ANSI AAMI IEC 60601-2-2:2009, 201.15.101.7)Test results demonstrated compliance with the standard (specific numerical values for adhesion are not provided, but compliance is stated).
    Connector Retention Force (snap button with male snap connector)9.62 N (for the subject device, compared to 9.60 N and 9.61 N for predicates, implying this value falls within acceptable limits for substantial equivalence).
    Hydrogel thickness1.0mm ± 0.2mm (for the subject device, matching the predicates, implying it meets the specified thickness).
    Electrical Conductivity & Adhesive Property"Good electrical conductivity, good adhesive property" (Qualitative assessment, noted as equivalent to predicates).

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes used for each specific bench test (e.g., how many electrodes were tested for impedance, adhesion, or biocompatibility). It only states that "Bench tests were conducted on SM Electrodes."

    The data provenance is from non-clinical bench tests. The location where these tests were conducted is not specified, but the manufacturer is Hong Qiangxing (Shen Zhen) Electronics Limited in Guangdong, China.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This submission relies solely on non-clinical bench testing against established international and national standards, not on expert-driven ground truth for clinical performance. The "ground truth" here is the adherence to the defined parameters and limits within those standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. There was no clinical study with a test set requiring adjudication. The assessment was based on objective measurements against engineering and biocompatibility standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was performed or included in this submission. The device is a cutaneous electrode, an accessory for stimulation devices, and does not involve AI or human "readers" in the context of image interpretation or diagnostic tasks.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The device is a physical cutaneous electrode, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the non-clinical tests was established by recognized international and national standards. Compliance with these standards (ASTM, ISO, ANSI AAMI IEC) served as the basis for determining the device's acceptable performance and safety characteristics.

    8. The sample size for the training set

    Not applicable. This device is a physical medical device, not an AI/ML algorithm. There is no concept of a "training set" in this context.

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set for an AI/ML algorithm, no ground truth needed to be established for it.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121719

    Validate with FDA (Live)

    Device Name
    SM TENS & PMS
    Date Cleared
    2013-05-10

    (333 days)

    Product Code
    Regulation Number
    890.5850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    TENS: To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities. PMS: It is intended to be used to stimulate healthy muscles in order to improve and facilitate muscle performance.

    Device Description

    SM TENS & PMS is a portable and DC 3.7V battery powered multi function device, offering both Transcutancous Electrical Nerve Stimulator (TENS) and Powered Muscle Stimulator (PMS) qualities in one device. SM TENS & PMS has 6 operation modes, which can give certain electrical pulses through electrode adhesive pads to the suggested area of the body where the electrodes are placed. The electronic stimulatory module has the operating elements of ON/OFF Switch, Display screen, Mode Selection key, Intensity Modification keys, Timing key, Pause key, Output socket, and USB port for battery charging. The display screen can show battery power, selected mode, current intensity, time remaining of an application mode, and indication of a pause. The device is equipped with accessories of electrode pads, electrode cables, battery chargers, and USB cables. The electrode cables are used to connect the pads to the device; the USB cable is used to connect the charger and the built-in lithium battery. All accessories, including USB cables, electrode pads, electrode cables, chargers can only be changed by special person. The electrodes are interchangeable. The application area of electrode pads must be larger than 4cm². The electrode pads are provided by GMDASZ Manufacturing Co., Ltd. with 510(k) cleared Number K092546.

    AI/ML Overview

    This submission is for a Traditional 510(k) for a new device, the "SM TENS & PMS" from Hong Qiangxing (Shen Zhen) Electronics Limited. The device is a portable, battery-powered multi-function device offering both Transcutaneous Electrical Nerve Stimulator (TENS) and Powered Muscle Stimulator (PMS) functionalities. The 510(k) submission seeks to demonstrate substantial equivalence to a predicate device, the "Powered Muscle Stimulator, JQ-5C" (K102598).

    The acceptance criteria are established through a comparison of the new device's technical characteristics and performance against the predicate device and relevant safety standards. The study proving the device meets the acceptance criteria is a safety testing protocol, which demonstrates compliance with recognized electrical and medical device standards.

    Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria, for the purpose of a 510(k) for substantial equivalence, are primarily that the new device performs equivalently to the predicate device in its intended use and meets relevant safety standards, without raising new questions of safety or effectiveness. The reported device performance is based on the results of the safety testing.

    Acceptance Criteria CategorySpecific Acceptance Criteria (Derived from comparison to predicate and standards)Reported Device Performance
    Intended UseIdentical indications for use as the predicate device (TENS for temporary pain relief, PMS for muscle performance improvement).Met: Both SM TENS & PMS and the predicate (JQ-5C) have identical TENS and PMS indications for use.
    Technological CharacteristicsSimilar design, power source, number of output modes, number of output channels, synchronization, channel isolation method, timer range, waveform, and wave shape.Met: The new device is reported to have the same intended use, power supply, components, 6 modes, 2 channels, software control, standards compliance, same waveform and wave shape, same pulse width, and net charge.
    Electrical SafetyCompliance with IEC 60601-1 (General Safety), IEC 60601-2-10 (Nerve and Muscle Stimulators), and IEC 60601-1-2 (EMC).Met: The new device passed testing according to IEC 60601-1: 2005, IEC 60601-2-10: 2001, and IEC 60601-1-2: 2001.
    Battery SafetyCompliance with IEC 62133 (Secondary cells and batteries safety).Met: The new device passed testing according to IEC 62133: 2002.
    Electromagnetic Compatibility (EMC)Compliance with FCC 47 CFR Part 18 (Industrial, Scientific, and Medical Equipment - Conducted Emissions).Met: The new device passed testing according to FCC 47 CFR Part 18.
    Output Parameters (Key electrical characteristics - variations within acceptable limits)Maximum Output Voltage, Maximum Output Current, Pulse Width, Max. Pulse Frequency, Net Charge, Maximum Phase Charge, Maximum Average Current, Maximum Current Density, Maximum Average Power Density to be comparable or safer than predicate.Met/Compared: While some specific numerical values differ (e.g., Maximum Output Voltage and Current, Max. Pulse Frequency, Maximum Current Density, Maximum Average Power Density), the submission concludes these differences (and those attributable to different smallest surface area of electrodes) do not raise new safety or effectiveness issues. Net charge and pulse width are the same.
    Patient Leakage CurrentNormal Condition: ≤ 2μA; Single Fault Condition: < 10μA.Met: Reported as 2μA (Normal Condition) and < 10μA (Single Fault Condition).
    Average DC current through electrodes< 0.01 μA when device is on but no pulses are applied.Met: Reported as < 0.01 μA.
    Electrode PadsElectrode pads must be larger than 4cm². (Note: This is a design requirement rather than a performance criterion for the device itself).Met: The application area of electrode pads must be larger than 4cm². The electrode pads are provided by GMDASZ Manufacturing Co., Ltd. with 510(k) cleared Number K092546.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify a "test set" in the context of clinical data for the device. The evaluation is based on engineering testing of a single device (or set of devices sufficient for engineering tests) to verify compliance with standards and comparison to the predicate device's specifications. It is a technical submission, not a clinical trial.
    • Data Provenance: The data primarily comes from engineering and bench testing conducted by the manufacturer (Hong Qiangxing (Shen Zhen) Electronics Limited), likely performed in China (Shenzhen, Guangdong, China), where the manufacturer is located. This is a retrospective comparison to existing standards and predicate device specifications.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This type of information is not applicable to this 510(k) submission. A 510(k) for an electrical stimulator device like this does not typically involve expert review for ground truth in a clinical sense. The "ground truth" here is compliance with established electrical safety and performance standards, as well as functional equivalence to a legally marketed predicate device. This is assessed through objective engineering tests.

    4. Adjudication Method for the Test Set

    This is not applicable. There is no "test set" requiring adjudication by experts in the context of this 510(k) submission. The evaluation is against objective engineering standards and a comparative analysis of technical specifications.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is typically used for diagnostic imaging devices or other technologies where human interpretation or clinical effectiveness is a primary endpoint requiring comparison between different readers or technologies. For TENS/PMS devices seeking substantial equivalence, such studies are not usually required in a 510(k) submission; the focus is on safety and functional equivalence to the predicate.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This concept is not applicable to the "SM TENS & PMS" device. This is a direct electrical stimulation device; it does not involve algorithms that perform diagnostic or interpretative functions that would require a "standalone" performance evaluation in the way imaging AI algorithms do. Its performance is evaluated by its electrical output characteristics and safety.

    7. The Type of Ground Truth Used

    The "ground truth" in this context is compliance with recognized electrical safety and performance standards (IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2, IEC 62133, FCC 47 CFR Part 18) and technical specifications of the predicate device. There is no "pathology" or "outcomes data" ground truth as this is not a diagnostic or treatment outcome effectiveness study.

    8. The Sample Size for the Training Set

    Not applicable. This device does not use machine learning or AI that would require a "training set." The submission describes a hardware device not an AI algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set, there is no ground truth establishment for it.