510(k) Data Aggregation

    K Number
    K232709

    Validate with FDA (Live)

    Date Cleared
    2023-11-30

    (86 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Diode laser therapy device is intended for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.

    Device Description

    Diode Laser Hair Removal adopts Gold standard 808nm semi-conductor laser, based on selective photothermolysis principle, through specific wavelength, penetrate epidermis into dermis, optical energy was absorbed and translated into heat energy with restraining hair follicle tissues, and produce photothermal effect. It takes laser energy in hair follicle with rich in melanin, surrounding tissues absorb less even no energy, reach to restrained melanin of hair follicles and remove hair. Meanwhile, don't hurt around tissues and virtually pain-free remove surplus hair; eventually, reach to permanent Hair Removal.

    The Diode laser therapy device utilize a semiconductor diode with invisible infrared radiation as a laser source to emit 808nm wavelength laser which is absorbed by melanin. The laser power is delivered to the treatment area via laser handpiece. The emission laser is activated by a foot-switch.

    AI/ML Overview

    The provided documentation does not contain detailed acceptance criteria or a study proving the device meets specific performance criteria beyond substantial equivalence to a predicate device.

    Here's an analysis of the provided information based on your request:

    1. A table of acceptance criteria and the reported device performance

    No explicit acceptance criteria with numerical targets for performance (e.g., hair reduction percentage, adverse event rate) are presented for the proposed device itself. The document primarily focuses on demonstrating substantial equivalence to a predicate device (K210663). The "performance" reported is primarily a comparison of technical specifications to the predicate device.

    Acceptance Criteria (Explicitly Stated in Document)Reported Device Performance (as compared to Predicate Device K210663)
    Product Code is GEXSAME (GEX)
    Regulation Number is 21 CFR 878.4810SAME (21 CFR 878.4810)
    Class is 2SAME (2)
    Indication for Use is for hair removal, permanent hair reduction on all skin types (Fitzpatrick skin type I-VI), including tanned skin. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime.SAME
    Configuration includes Main Unit, Handpiece, Foot ControlSAME
    Principle of Operation is Diode LaserSAME
    Laser Type is Diode LaserSAME
    Laser Classification is Class IVSAME
    Laser wavelength is 808 nmSAME
    Fluence is 5-100 J/cm²SAME
    Frequency is 1-10 HzSAME
    Power Supply is AC 110V/60HzSAME
    Complies with IEC 60601-1-2: 2020Test results demonstrated compliance
    Complies with AAMI/ANSI ES 60601-1: 2012Test results demonstrated compliance
    Complies with IEC 60601-2-22: 2019Test results demonstrated compliance
    Complies with IEC 60825-1: 2014Test results demonstrated compliance
    Complies with ISO 10993-5: 2009Test results demonstrated compliance
    Complies with ISO 10993-10: 2010Test results demonstrated compliance
    Complies with ISO 14971: 2019Test results demonstrated compliance

    Differences and their analysis for safety and effectiveness:

    • Spot Size:
      • Proposed: 12mm × 35mm
      • Predicate: 10 × 30 mm
      • Analysis: "Spot size only affects the area of treatment, not affect the therapeutic effect. Therefore, this difference will not affect the safety and effectiveness."
    • Pulse Duration:
      • Proposed: 5-200ms
      • Predicate: 15-400ms
      • Analysis: "The frequency of the proposed device is within the range of that of the predicate device, which can justify that the difference in the parameter of Pulse duration will not raise new safety issues of the proposed device. In addition, we also found a reference device (K181019), which has the same intended use as the reference device. The Pulse Duration of the reference device is 30-200 ms, which is similar to the proposed device's pulse duration, so this difference will not affect the effectiveness of the device."
    • Dimension:
      • Proposed: 55cm × 55 cm × 123cm
      • Predicate: 65cm × 65cm × 123cm
      • Analysis: "the dimension and weight difference are just in physical specification and this difference will not raise any issues in safety and effectiveness. By complying with AAMI/ANSV/ES 60601-1, the mechanical performance of the proposed device is determined to be accepted."
    • Weight:
      • Proposed: 62kg
      • Predicate: 75kg
      • Analysis: (Same as Dimension)

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document explicitly states: "No clinical study is included in this submission."
    Therefore, there is no clinical test set, sample size, or specific data provenance (country of origin, retrospective/prospective) for device performance on human subjects.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable, as no clinical study was performed.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no clinical study was performed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable, as this device is a laser therapy device for hair removal, not an AI-assisted diagnostic tool involving human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This refers to the performance of a software algorithm. This device is a hardware laser therapy device, not a standalone algorithm. Therefore, this question is not applicable.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    Since no clinical study was conducted, there is no patient-level ground truth data (like expert consensus, pathology, or outcomes data) specifically for this proposed device. The basis for substantial equivalence relies on the technical specifications and indications for use being similar to a legally marketed predicate device, and the device complying with relevant electrical, laser, and biocompatibility standards through non-clinical testing.

    8. The sample size for the training set

    Not applicable, as no algorithm or machine learning model requiring a training set is mentioned for this device.

    9. How the ground truth for the training set was established

    Not applicable, as no algorithm or machine learning model requiring a training set is mentioned for this device.

    In summary: The provided FDA 510(k) clearance document for the Diode laser therapy device is based on demonstrating substantial equivalence to a predicate device (K210663) through non-clinical performance testing (compliance with standards) and a comparison of technical specifications, rather than a new clinical study. Therefore, most of the requested information regarding clinical study design, test sets, ground truth, and AI-related metrics is not available in this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K232716

    Validate with FDA (Live)

    Date Cleared
    2023-11-30

    (86 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Q-Switched Nd: Yag laser is indicated for the treatment of: benign cutaneous lessons, such as Warts, Scars, Striae and Psoriasis; benign pigmented lesions, such as Lentigines, nevus, and the removal of black or blue tattoos.

    Device Description

    The Q-switched laser therapy device is laser system which delivers laser at a wavelength 1064nm or 532nm and the pulse mode is single pulse. In laser handpiece, there is one optical cavity containing the Nd: YAG crystal. The laser beam is directed to the treatment zone. When the laser beam contacts human tissue, the energy in the beam is absorbed, resulting in a very rapidly, highly localized temperature increase to the target tissue. The instantaneous temperature increase (thermal effect) causes the cells change of target tissue to achieve laser treatment effect.

    AI/ML Overview

    The provided document is a 510(k) premarket notification letter from the FDA regarding a Q-Switched Nd:Yag laser device. This type of document establishes substantial equivalence to a predicate device and does not contain information about acceptance criteria or a study proving that the device meets specific performance criteria through clinical data, especially not for AI/machine learning devices where such studies would be critical.

    The document explicitly states:

    • "No clinical study is included in this submission." (Page 9)
    • The entire evaluation for substantial equivalence is based on "Non-Clinical Test Conclusion," showing compliance with various electrical safety, biocompatibility, and laser safety standards (ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1, ISO 10993-5, ISO 10993-10), and Software Validation & Verification Test.

    Therefore, I cannot provide the requested information for an acceptance criteria table, sample sizes, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, or training set details, as this information is not present in the provided FDA 510(k) summary for this laser device.

    The questions you've posed (acceptance criteria, sample size, expert ground truth, MRMC, etc.) are highly relevant for the validation of AI/ML-based medical devices or devices with a diagnostic component whose performance is assessed against human interpretation or clinical outcomes. The device in this document is a physical laser for treatment, and its FDA clearance is based on demonstrating equivalence in its technical specifications and adherence to relevant safety standards, not on diagnostic accuracy or AI performance.