Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K231389

    Validate with FDA (Live)

    Date Cleared
    2023-08-16

    (96 days)

    Product Code
    Regulation Number
    872.3760
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Harvest Dental HD Gum Strip is indicated for adding texture, form and color to the gingival area of a dental prosthesis which includes crowns, bridges and full and partial prostheses. This composite strip is bondable to polymethylmethacrylate (PMMA), metals and ceramics (zirconia).

    Device Description

    Harvest Dental HD Gum Strip is a bondable, light-cured composite for adding texture and color to the gingival area of a dental prosthesis which includes crowns, bridges and full and partial prostheses. This composite strip is bondable to polymethylmethacrylate (PMMA), metals and ceramics (zirconia). It is provided in a textured, malleable strip meant to be applied to a dentine/bone-like substructure. The strip is an alternative to the method of using a shaded composite paste to manually build-up and model visible gingival surfaces to look similar to natural gingival tissues. The Harvest Dental HD Gum Strip includes features such as gingival veins formed into its surface as well as being available in various shades of coloration. The strip can be easily cut, applied to the gingival surface of the restoration, then trimmed and finished with appropriate features such as gingival line and interdental papilla. Gum Strip Modifiers (paste versions of the same composite) are available in various shades and used for additional surface features and gingival repairs, and Liquid Modifiers (stains, which are a low-viscosity color-concentrated form of the same composite) are available in several shades and used for enhanced coloration resembling the natural gingiva. The Harvest Dental HD Gum Strip is bonded to the dental prosthesis utilizing a bonding adhesive provided with the strip and modifiers and, after any final detailing and coloring, is to be cured utilizing a typical light-curing device.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a dental device, Harvest Dental HD Gum Strip. It details the device's characteristics and its substantial equivalence to a predicate device, Gradia Gum. However, this document does not contain information about acceptance criteria or a study proving the device meets those criteria in the context of an AI/ML device. The document primarily focuses on biocompatibility and mechanical property testing for a physical dental material. Therefore, most of the requested information regarding AI/ML device evaluation is not present in this document.

    Here's the information that can be extracted or inferred based on the provided text, with many fields explicitly stated as "Not Applicable" for an AI/ML device, as this document is not about one.

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance (Harvest Dental HD Gum Strip)
    Biocompatibility (ISO 7405:2018, ISO 10993-1)Adequately demonstrated biocompatibility through Cytotoxicity, Sensitization, and Oral Mucosa Irritation tests.
    Flexural Strength (ISO 10477:2020)77 MPa (> 50 MPa required by standard)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
    Not applicable as this is not an AI/ML device studying a dataset. The testing refers to physical material property tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
    Not applicable as this is not an AI/ML device. Ground truth for material properties is established through standardized physical tests.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    Not applicable as this is not an AI/ML device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable as this is not an AI/ML device. No human-in-the-loop studies were conducted for this physical dental material.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    Not applicable as this is not an AI/ML device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
    For physical material properties, the "ground truth" is defined by the standards and methods used in the specified ISO tests (ISO 7405 and ISO 10477).

    8. The sample size for the training set
    Not applicable as this is not an AI/ML device.

    9. How the ground truth for the training set was established
    Not applicable as this is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222961

    Validate with FDA (Live)

    Date Cleared
    2023-06-02

    (248 days)

    Product Code
    Regulation Number
    872.6660
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Harvest Conceal ZR Block Out is indicated for burn-in zirconia shading. The applied liquid is sintered with the zirconia to adjust the restoration to match the natural color of the patient's teeth. The material is intended for professional dental work only.

    Device Description

    Harvest Conceal ZR Block Out is a water-based opacifying solution available in 5 shades. white, A1, B1, C1 and D2, for use with a translucent zirconia core. It is indicated for burn-in zirconia shading. The applied liquid is sintered with the zirconia to adjust the restoration to match the natural color of the patient's teeth. Fabrication using the Harvest Conceal ZR Block Out requires an appropriate drying oven and sintering furnace.

    AI/ML Overview

    The provided text describes a medical device, "Harvest Conceal ZR Block Out," and its substantial equivalence determination by the FDA. However, the document does not contain the specific information requested in the prompt regarding acceptance criteria, study details, sample sizes, expert qualifications, or ground truth establishment relevant to AI/ML device performance.

    The document states that:

    • No human clinical testing was conducted. This implies there would be no data on multi-reader multi-case studies or effect sizes of human readers with AI assistance.
    • The device is a "Liquid Stain for Dental Zirconia Restorations," which suggests it is a material used in the fabrication of dental products, not a diagnostic or prognostic AI/ML device.
    • Performance testing was "Bench Testing" for physical characteristics, and "Biocompatibility Testing" was conducted in accordance with ISO standards.

    Therefore, I cannot extract the requested information such as a table of acceptance criteria and reported device performance, sample sizes for test sets, number of experts, adjudication methods, MRMC studies, standalone performance, or details about training sets and their ground truth, because this information is not present in the provided text. The document is primarily focused on the device's substantial equivalence to a predicate device based on material properties and intended use, not on the performance of an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222489

    Validate with FDA (Live)

    Date Cleared
    2023-05-12

    (268 days)

    Product Code
    Regulation Number
    872.3760
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Harvest Printable Resin is indicated for the fabrication of dental bases for full removable dentures, artificial teeth, dental bite splints and guards and try-in devices, i.e., denture base and teeth. The material is an alternative to traditional heat-curable and auto-polymerizing resins and is intended for professional dental work only. This material is intended to be used by dental lab technicians and approved by licensed practitioners before being provided to the patient.

    Device Description

    Harvest Printable Resin is a light-cured resin developed for additive manufacturing (3D printing) of individual full denture bases, artificial teeth, bite splints and guards and try-in devices, i.e., denture base and teeth. After being utilized in a 3D stereolithographic printer to generate a dental device based on a solid model, the device is placed in a UV light curing unit for final polymerization. Fabrication using Harvest Printable Resin requires an appropriate computer-aided Design and Manufacturing (CAD/CAM) system, a digital light processing (DLP) printer, and post-processing light curing equipment. It has been optimized for use with Asiga (Max UV, Pro 4K65 and Pro 4K80) and Sprintray (Pro95 S and Pro55 S) printers and, therefore, may only be used in conjunction with these printers and their associated software systems. It is also only to be used in conjunction with Asiga (Max UV, Pro 4K65 and Pro 4K80) and Sprintray (Procure and Procure 2) curing units for post-processing. The 3D printer is not included with the device.

    AI/ML Overview

    The provided document describes a medical device, "Harvest Printable Resin," and its substantial equivalence to predicate devices based on technological characteristics and performance data, primarily through biocompatibility testing and ISO 20795-1 & -2 testing.

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Standard Reference)Reported Device Performance (Harvest Printable Resin)Predicate Device 1 (NextDent Denture/E-Denture) PerformancePredicate Device 2 (E-Guard) PerformanceComparison to Criteria
    Ultimate Flexural Strength (ISO 20795-1)≥ 65 MPa (reported: 81.6 MPa)≥ 65 MPa (reported: 84 MPa)N/APassed
    Ultimate Flexural Strength (ISO 20795-2)≥ 50 MPa (reported: 81.6 MPa)N/A≥ 50 MPa (reported: 79.4 - 85.1 MPa)Passed
    Flexural Modulus (ISO 20795-1)≥ 2,000 MPa (reported: 2,339 MPa)≥ 2,000 MPa (reported: 2,383 MPa)N/APassed
    Flexural Modulus (ISO 20795-2)≥ 1,500 MPa (reported: 2,339 MPa)N/A≥ 1,500 MPa (reported: 2050-2130 MPa)Passed
    Water Sorption (ISO 20795-1)≤ 32 µg/mm³ (reported: 24.3 µg/mm³)≤ 32 µg/mm³ (reported: 28 µg/mm³)N/APassed
    Water Sorption (ISO 20795-2)≤ 32 µg/mm³ (reported: 24.3 µg/mm³)N/A≤ 32 µg/mm³ (reported: 30-32 µg/mm³)Passed
    Water Solubility (ISO 20795-1)≤ 1.6 µg/mm³ (reported: 1.1 µg/mm³)≤ 1.6 µg/mm³ (reported: 0.1 µg/mm³)N/APassed
    Water Solubility (ISO 20795-2)≤ 5 µg/mm³ (reported: 1.1 µg/mm³)N/A≤ 5 µg/mm³ (reported: 0.5 µg/mm³)Passed
    Residual MonomerNone (no MMA monomers used for production)≤ 0.1% (w/w) (for some MMA) / ≤ 2.2% (w/w)UnknownImproved
    BiocompatibilityAdequate biocompatibilityN/A (implied by predicate status)N/A (implied by predicate status)Passed

    2. Sample size used for the test set and the data provenance

    The document does not explicitly state the sample sizes used for the ISO 20795-1 & -2 tests or the biocompatibility tests. It also does not specify the provenance (e.g., country of origin, retrospective or prospective) of the data. These tests are typically conducted in a laboratory setting on standardized test specimens.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable and not provided in the document. The "ground truth" for material property testing (like ISO 20795-1 & -2) and biocompatibility is established by adherence to recognized international standards and laboratory measurements, not by expert consensus on clinical cases. No human-in-the-loop performance is being evaluated here.

    4. Adjudication method for the test set

    Not applicable. As noted above, this is material property and biocompatibility testing, not a clinical assessment requiring adjudication of human expert opinions.

    5. If a multi-reader, multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was done. The document explicitly states: "No human clinical testing was conducted to support substantial equivalence." This device is a material, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, "standalone" performance was effectively done in the form of laboratory testing of the material properties described in section 1 and biocompatibility. The performance evaluation focuses solely on the inherent characteristics of the Harvest Printable Resin itself, without any human intervention in its measurement or any AI algorithms.

    7. The type of ground truth used

    The ground truth for the performance evaluation relies on:

    • Established International Standards: ISO 20795-1:2013 and ISO 20795-2:2013 for physical/mechanical properties. These standards define the methodology and acceptance criteria.
    • Recognized Biocompatibility Standards: ISO 7405:2018 and ISO 10993-1 as recognized by the FDA for biocompatibility. These standards outline tests to ensure the material is safe for biological contact.

    8. The sample size for the training set

    Not applicable. This device is a material, not an AI model, so there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180578

    Validate with FDA (Live)

    Date Cleared
    2018-12-12

    (282 days)

    Product Code
    Regulation Number
    872.3770
    Panel
    Dental
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For the fabrication of crowns, bridges and structures for implant supported provisional removable denture and appliance prosthetics.
    -Provisional anterior and posterior crowns & bridges

    • Implant and abutment supported prosthetics.
    • Partial, complete and hybrid denture prosthetics (base and teeth) - Removable appliances (splint)
    Device Description

    Harvest Dental Polymer Blocks are industrially polymerized, pre-colored or clear dental milling discs and blocks designed for milled fabrication of temporary anterior and posterior crowns and bridges (tooth-colored variants) or bite splints and dental CAD/CAM systems.

    AI/ML Overview

    The Harvest Dental Polymer Blocks device is a dental milling disc and block designed for the CAD/CAM fabrication of temporary anterior and posterior crowns and bridges, as well as provisional removable denture and appliance prosthetics. The submission demonstrates substantial equivalence to its predicate device, Idodentine (K150432), based on similar technological characteristics, indications for use, and performance testing.

    1. A table of acceptance criteria and the reported device performance:

    RequirementRequired Value (ISO 20795-1)Required Value (ISO 10477)Value (Submission Device)Passed/Failed
    Flexural Strength≥ 65 [MPa]≥ 50 [MPa]> 90 MpaPassed
    Water Absorption≤ 32 [µg/mm³]≤ 40 [µg/mm³]0.024 mg/mm³Passed
    Water Solubility≤ 1.6 [µg/mm³]≤ 7.5 [µg/mm³]0.0000 mg/mm³Passed
    Residual Monomer Content≤ 2.2 %-< 1.1 %Passed

    2. Sample size used for the test set and the data provenance:

    The document does not explicitly state the sample size used for each specific test or the specific provenance (e.g., country of origin, retrospective/prospective) of the testing data beyond stating it was performed in a laboratory.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable as the acceptance criteria and performance evaluation for this device are based on objective, standardized physical and chemical property testing rather than expert-derived ground truth from medical images or clinical observations.

    4. Adjudication method for the test set:

    Not applicable. The performance is assessed against established ISO and ADA standards through laboratory testing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a dental material, not an AI-assisted diagnostic tool requiring human reader studies.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    Not applicable. This device is a dental material, not an algorithm. The performance testing is analogous to "standalone" in that it evaluates the material properties directly without human interpretation as part of the primary outcome.

    7. The type of ground truth used:

    The ground truth used for evaluating the device's performance against its acceptance criteria is based on:

    • ISO Standards: Established international standards for dental materials (ISO 10477, ISO 20795-1).
    • Biocompatibility Standards: ISO 10993 for biological evaluation of medical devices.
    • ADA Standards: ADA 53 for Polymer-Based crown and Bridge Resin.

    These standards define the acceptable range or threshold for specific physical, chemical, and biological properties.

    8. The sample size for the training set:

    Not applicable. This device is a physical dental material with established chemical and physical properties, not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132937

    Validate with FDA (Live)

    Date Cleared
    2014-05-21

    (245 days)

    Product Code
    Regulation Number
    872.3770
    Panel
    Dental
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Harvest Dental Polymer Blocks (ZCAD) are PMMA blanks for manufacturing temporary crowns and bridges by a CAD/CAM system.

    Device Description

    Harvest Dental Polymer Blocks (ZCAD) are PMMA blanks.

    AI/ML Overview

    The provided text is a 510(k) premarket notification letter from the FDA regarding a dental product. It does not contain information about acceptance criteria for a device's performance through a study. Instead, it is a regulatory document confirming substantial equivalence based on prior predicate devices.

    Therefore, I cannot fulfill your request to describe the acceptance criteria and the study that proves the device meets them based on the provided text. The document does not discuss performance studies, sample sizes, expert ground truth, adjudication methods, MRMC studies, standalone performance, or training set details.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1