    Intended Use

    Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, Simulated Gastric Acid and Fentanyl in Simulated Gastric Acid (Black) is a non-sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs and Fentanyl Citrate, Simulated Gastric Acid and Fentanyl in Simulated Gastric Acid. The gloves were tested for use with chemotherapy drugs and fentanyl citrate as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

    Chemotherapy DrugConcentrationMinimum Breakthrough Detection Time in Minutes
    Azacytidine(25.0 mg/ml)>240
    Carmustine(3.3 mg/ml)25.0
    Cisplatin(1.0 mg/ml)>240
    Cyclophosphamide(20.0 mg/ml)>240
    Carboplatin(10.0 mg/ml)>240
    Dacarbazine(10.0 mg/ml)>240
    Docetaxel(10.0 mg/ml)>240
    Doxorubicin HCl(2.0 mg/ml)>240
    Epirubicin(2.0 mg/ml)>240
    Etoposide(20.0 mg/ml)>240
    Fluorouracil(50.0 mg/ml)>240
    Gemcitabine(38.0 mg/ml)>240
    Ifosfamide(50 mg/ml)>240
    Irinotecan(20.0 mg/ml)>240
    Methotrexate(25.0 mg/ml)>240
    Mitomycin C(0.5 mg/ml)>240
    Mitoxantrone(2.0 mg/ml)>240
    Oncovin(1.0mg/ml)>240
    Oxaliplatin(5.0 mg/ml)>240
    Paclitaxel(6.0 mg/ml)>240
    Thiotepa(10.0 mg/ml)55.7
    Vinorelbine(10.0 mg/ml)>240
    Vincristine Sulfate(1.0 mg/ml)>240

    Testing showed a minimum breakthrough time of 25.0 minutes with Carmustine and 55.7 minutes with Thiotepa
    Warning: Do not use with Carmustine

    Fentanyl CitrateConcentrationMinimum Breakthrough Detection Time in Minutes
    Fentanyl Citrate Injection100 mcg/2ml>240
    Fentanyl Citrate mixed with Simulated Gastric Acid at a 50:50 ratio>240
    Device Description

    Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs, Fentanyl Citrate, Simulated Gastric Acid and Fentanyl in Simulated Gastric Acid (Black) is a disposable single-use, non-sterile, Black and powder-free ambidextrous glove made from nitrile butadiene rubber.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue); and Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue) - Extended Cuff, is a sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

    These gloves were tested for use with chemotherapy drugs and fentanyl citrate as per ASTM D6978 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

    Device Description

    Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue) and Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue) – Extended Cuff, are disposable, single-use, sterile, blue-colored, and powder-free examination gloves made from nitrile latex.

    AI/ML Overview

    The provided FDA 510(k) clearance letter pertains to medical examination gloves, not an AI-powered medical device. Therefore, the document does not contain information about the acceptance criteria and study proving an AI device meets those criteria.

    The information typically found in such a submission would include:

    • Device performance: This document explicitly states the "Sterile Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue)" has been tested for resistance to permeation by chemotherapy drugs and fentanyl citrate as per ASTM D6978 Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.
    • Acceptance Criteria/Performance Data: The acceptance criteria for the gloves are based on relevant ASTM standards for physical properties, barrier integrity, and biocompatibility. The chemotherapy drug resistance is presented as "Minimum Breakthrough Detection Time in Minutes."
    • Study Design (Non-Clinical): The studies are non-clinical, primarily laboratory-based physical and chemical tests compliant with recognized international and national standards (e.g., ASTM, ISO).
    • Ground Truth: For these types of devices, the "ground truth" is typically defined by quantifiable physical and chemical properties measured against established standard methods (e.g., measuring milligrams of residual powder, breakthrough time of chemicals). There are no human experts, adjudication, or training/test sets in the context of AI for this type of device.

    Given this, I cannot extract the information required for an AI-powered device from this document. The sections you asked for (sample size for test/training, expert qualifications, MRMC studies, standalone performance, etc.) are specific to AI/software as a medical device (SaMD) assessments and are not present in a submission for examination gloves.

    However, I can summarize the acceptance criteria and reported performance for the examination gloves as provided in the document:

    Acceptance Criteria and Reported Device Performance (for Examination Gloves)

    This table summarizes the non-clinical acceptance criteria and the "result" or reported performance for the Sterile Nitrile Powder Free Examination Gloves.

    Test / CharacteristicAcceptance Criteria / Standard ReferenceReported Device Performance
    Freedom from holesMeets ASTM D5151-19 (R2023): AQL 1.5%Pass
    Residual PowderMeets ASTM D6124-06 (R2022): Average less than 2 mg/glovePass
    Dimensions (Length)Meets ASTM D6319-19 (R2023): Overall Length: ≥ 230 mm (Subject Device 1); ≥ 260 mm (Subject Device 2)Pass (Conforms to ASTM D6319 width, thickness, and length requirements)
    Dimensions (Width)Meets ASTM D6319-19 (R2023): Specific ranges for XS, S, M, L, XLPass (Conforms to ASTM D6319 width, thickness, and length requirements)
    Dimensions (Thickness)Meets ASTM D6319-19 (R2023): Palm Thickness: ≥ 0.05 mm; Finger Thickness: ≥ 0.05 mmPass (Conforms to ASTM D6319 width, thickness, and length requirements)
    Tensile Strength Before AgingMeets ASTM D6319-19 (R2023): ≥ 14 MPaPass (Conforms)
    Tensile Strength After AgingMeets ASTM D6319-19 (R2023): ≥ 14 MPaPass (Conforms)
    Ultimate Elongation Before AgingMeets ASTM D6319-19 (R2023): ≥ 500 %Pass (Conforms)
    Ultimate Elongation After AgingMeets ASTM D6319-19 (R2023): ≥ 400 %Pass (Conforms)
    Primary Skin IrritationISO 10993-23: Not an irritantUnder the conditions of the study, not an irritant.
    Dermal SensitizationISO 10993-10: Not a sensitizerUnder the conditions of the study, not a sensitizer.
    Acute Systemic ToxicityISO 10993-11: No evidence of acute systemic toxicityUnder the conditions of this study, no acute systemic toxicity.
    Chemotherapy Drug PermeationASTM D6978 (No specific 'Pass/Fail' criterion listed, rather minimum breakthrough times are reported.)Listed in table below.
    Fentanyl Citrate PermeationASTM D6978 (No specific 'Pass/Fail' criterion listed, rather minimum breakthrough times are reported.)> 240 minutes

    Chemotherapy Drug and Fentanyl Citrate Permeation Results

    Device Performance (Minimum Breakthrough Detection Time in Minutes)

    Chemotherapy Drug and ConcentrationSubject Device Result
    Carmustine (3.3 mg/ml)38.3
    Cisplatin (1.0 mg/ml)> 240
    Cyclophosphamide (20.0 mg/ml)> 240
    Dacarbazine (10.0 mg/ml)> 240
    Doxorubicin HCl (2.0 mg/ml)> 240
    Etoposide (20.0 mg/ml)> 240
    Fluorouracil (50.0 mg/ml)> 240
    Methotrexate (25.0 mg/ml)> 240
    Mitomycin C (0.5 mg/ml)> 240
    Paclitaxel (6.0 mg/ml)> 240
    Thiotepa (10.0 mg/ml)78.6
    Vincristine Sulfate (1.0 mg/ml)> 240
    Vidaza (5-Azacytidine), 25 mg/ml> 240
    Busulfan, 6 mg/ml> 240
    Carboplatin, 10 mg/ml> 240
    Docetaxel, 10 mg/ml> 240
    Epirubicin HCl, 2 mg/ml> 240
    Gemcitabine HCl, 38 mg/ml> 240
    Ifosfamide, 50 mg/ml> 240
    Irinotecan, 20 mg/ml> 240
    Mitoxantrone HCl, 2 mg/ml> 240
    Oxaliplatin, 5 mg/ml> 240
    Vinorelbine, 10 mg/ml> 240
    Fentanyl Citrate Injection (100mcg/2ml)> 240

    Since the provided document is not for an AI device, the following points cannot be addressed:

    • Sample sized used for the test set and the data provenance: Not applicable to a physical glove.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for gloves is based on physical/chemical measurements by laboratory technicians following standard protocols.
    • Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    • If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For gloves, it's objective physical/chemical measurements according to ASTM/ISO standards.
    • The sample size for the training set: Not applicable (no AI training involved).
    • How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K243818

    Validate with FDA (Live)

    Date Cleared
    2025-01-08

    (27 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue) is a nonsterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs and Fentanyl Citrate.

    Device Description

    Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue) is a disposable single-use, non- sterile, blue-colored and powder-free examination glove made from nitrile latex. Gloves meet the specification of ASTM D6319-19 and have been tested for resistance to permeation by chemotherapy drugs and fentanyl citrate as per ASTM D6978-05(2019).

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification from the FDA for a medical device: "Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue)".

    This document does not describe the acceptance criteria or a study related to an AI/ML powered device. Instead, it details the testing and comparison of a physical medical device (gloves) against a predicate device, focusing on material properties, performance standards (like resistance to permeation by chemotherapy drugs and fentanyl citrate), and biocompatibility.

    Therefore, I cannot extract the requested information regarding AI/ML device acceptance criteria or related studies from this document. The questions about sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth, and training sets are not applicable to the type of device described in this FDA submission.

    However, I can extract the acceptance criteria and performance data for the physical device as described:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The primary performance criterion for the "Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue)" is its resistance to permeation by chemotherapy drugs and fentanyl citrate. The acceptance criterion is a minimum breakthrough detection time for each substance.

    Chemotherapy Drug and ConcentrationAcceptance Criteria (Minimum Breakthrough Detection Time in Minutes)Reported Device Performance (Minimum Breakthrough Detection Time in Minutes)
    Carmustine, 3.3 mg/mlN/A (Tested to determine time)10.2
    Cisplatin, 1.0 mg/mlN/A (Tested to determine time)>240
    Cyclophosphamide, 20.0 mg/mlN/A (Tested to determine time)>240
    Dacarbazine, 10.0 mg/mlN/A (Tested to determine time)>240
    Doxorubicin Hydrochloride, 2.0 mg/mlN/A (Tested to determine time)>240
    Etoposide, 20.0 mg/mlN/A (Tested to determine time)>240
    Fluorouracil, 50.0 mg/mlN/A (Tested to determine time)>240
    Methotrexate, 25.0 mg/mlN/A (Tested to determine time)>240
    Mitomycin C, 0.5 mg/mlN/A (Tested to determine time)>240
    Paclitaxel, 6.0 mg/mlN/A (Tested to determine time)>240
    Thiotepa, 10.0 mg/mlN/A (Tested to determine time)30.2
    Vincristine Sulfate, 1.0 mg/mlN/A (Tested to determine time)>240
    Bleomycin Sulfate, 15.0 mg/mlN/A (Tested to determine time)>240
    Bortezomib, 1.0 mg/mlN/A (Tested to determine time)>240
    Busulfan, 6.0 mg/mlN/A (Tested to determine time)>240
    Carboplatin, 10.0 mg/mlN/A (Tested to determine time)>240
    Chloroquine, 50.0 mg/mlN/A (Tested to determine time)>240
    Cyclosporin A, 100.0 mg/mlN/A (Tested to determine time)>240
    Cytarabine, 100.0 mg/mlN/A (Tested to determine time)>240
    Daunorubicin, 5.0 mg/mlN/A (Tested to determine time)>240
    Docetaxel, 10.0 mg/mlN/A (Tested to determine time)>240
    Epirubicin, 2.0 mg/mlN/A (Tested to determine time)>240
    Fludarabine, 25.0 mg/mlN/A (Tested to determine time)>240
    Gemcitabine, 38.0 mg/mlN/A (Tested to determine time)>240
    Idarubicin, 1.0 mg/mlN/A (Tested to determine time)>240
    Ifosfamide, 50.0 mg/mlN/A (Tested to determine time)>240
    Irinotecan, 20.0 mg/mlN/A (Tested to determine time)>240
    Mechlorethamine HCI, 1.0 mg/mlN/A (Tested to determine time)>240
    Melphalan, 5.0 mg/mlN/A (Tested to determine time)>240
    Mitoxantrone, 2.0 mg/mlN/A (Tested to determine time)>240
    Oxaliplatin, 2.0 mg/mlN/A (Tested to determine time)>240
    Paraplatin, 10.0 mg/mlN/A (Tested to determine time)>240
    Retrovir, 10.0 mg/mlN/A (Tested to determine time)>240
    Rituximab, 10.0 mg/mlN/A (Tested to determine time)>240
    Topotecan, 1.0 mg/mlN/A (Tested to determine time)>240
    Trisenox, 1.0 mg/mlN/A (Tested to determine time)>240
    Fentanyl Citrate Injection, 100mcg/2mlN/A (Tested to determine time)>240

    Additional Acceptance Criteria and Performance Data (already met by similar predicate device or physical properties):

    CharacteristicAcceptance CriteriaReported Device Performance
    Freedom from holesAQL 1.5% (Meets ASTM D5151-19 and ASTM D6319-19)Pass
    Residual PowderAverage < 2 mg/glove (Meets ASTM D6319-19 & ASTM D6124-06 (2017))Pass
    Dimensional ConformanceConforms to ASTM D6319 width, thickness, and length requirements for XS, S, M, L, and XL; AQL 4%Pass
    Tensile PerformanceTensile Strength Before Aging: $\ge$ 14 MPa; Ultimate Elongation Before Aging: $\ge$ 500 %; Tensile Strength After Aging: $\ge$ 14 MPa; Ultimate Elongation After Aging: $\ge$ 400 % (Meets ASTM D6319-19); AQL 4%Pass
    Biocompatibility: Skin IrritationNot an irritant (Under the conditions of the study, per ISO 10993-10)Pass
    Biocompatibility: Skin SensitizationNot a sensitizer (Under the conditions of the study, per ISO 10993-10)Pass
    Biocompatibility: Acute ToxicityNo acute toxicity (Under the conditions of the study, per ISO 10993-11)Pass

    2. Sample size used for the test set and the data provenance:

    The document mentions that "Fentanyl citrate permeation testing was performed on the subject glove using the ASTM D6978 method." While it confirms the test was done, it does not specify the sample size for the test set used for the permeation testing.
    The data provenance is from Malaysia, where the manufacturer (Hartalega NGC Sdn. Bhd.) is located. The study appears to be prospective testing conducted specifically for this 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This is not applicable as the "ground truth" for this device is established through physical and chemical testing against industry standards (ASTM D6978, D6319, D5151, D6124, ISO 10993) rather than expert consensus on medical images or diagnoses.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable, as this involves standardized physical and chemical laboratory testing, not human expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is a medical glove, not an AI/ML-powered device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is a medical glove, not an AI/ML-powered device.

    7. The type of ground truth used:

    The ground truth is established through standardized laboratory testing protocols (ASTM D6978-05, D6319-19, D5151-19, D6124-06(R17), ISO 10993-10, ISO 10993-11) and the measured chemical permeation times and physical properties of the gloves.

    8. The sample size for the training set:

    Not applicable. This is a medical glove, not an AI/ML-powered device. There is no concept of a "training set" in this context.

    9. How the ground truth for the training set was established:

    Not applicable. There is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223437

    Validate with FDA (Live)

    Date Cleared
    2023-05-17

    (184 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Biodegradable Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Fusion Colour) is a non-sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs and Fentanyl Citrate.

    These gloves were tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

    Device Description

    Biodegradable Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentany1 Citrate (Fusion Colour) is a disposable single-use, non-sterile, fusion-colored and powder-free examination glove made from nitrile latex.

    AI/ML Overview

    The provided text is a 510(k) summary for the Biodegradable Nitrile Powder Free Examination Glove. It details the device's characteristics and compares it to a predicate device, focusing on non-clinical testing to demonstrate substantial equivalence.

    Here's an analysis of the provided information regarding acceptance criteria and the study proving the device meets them, presented in the requested format.

    1. Table of Acceptance Criteria and Reported Device Performance

    The device is a medical glove, and its performance is evaluated against established ASTM and ISO standards for physical characteristics, chemical permeation, and biocompatibility.

    Characteristic/ParameterTest Methodology/StandardAcceptance Criteria (Standard)Reported Device Performance
    DimensionsASTM D6319-19Length (mm): XS, S: Min 220; M, L, XL: Min 230 Width (mm): XS: 70 ± 10; S: 80 ± 10; M: 95 ± 10; L: 110 ± 10; XL: 120 ± 10Length (mm): XS: 240; S: 242; M: 239; L: 246; XL: 245 (Meets requirements) Width (mm): XS: 76; S: 86; M: 97; L: 107; XL: 114 (Meets requirements)
    ThicknessASTM D6319-19Palm: Minimum 0.05 mm Finger: Minimum 0.05 mmPalm Thickness (mm): XS: 0.05; S: 0.05; M: 0.06; L: 0.06; XL: 0.06 (Meets requirements) Finger Thickness (mm): XS: 0.08; S: 0.08; M: 0.08; L: 0.08; XL: 0.08 (Meets requirements)
    Physical PropertiesASTM D6319-19Tensile Strength: Min 14 MPa (Before & After Aging) Ultimate Elongation: Min 500% (Before Aging); Min 400% (After Aging)Before Age: Tensile Strength (MPa) Average 35; Elongation at Break (%) Average 565 (Meets requirements) After Age: Tensile Strength (MPa) Average 36; Elongation at Break (%) Average 469 (Meets requirements)
    Freedom from HolesASTM D5151-19, ASTM D6319-19AQL 2.5 (Acceptable Quality Level)Meets ASTM D6319-19 and ASTM D5151-19 requirements of AQL 2.5. Reported as pass at AQL 1.5, which is a tighter quality level. (Meets requirements)
    Powder ResidualASTM D6124-06 (2017), ASTM D6319-19Powder Free; ≤ 2 mg per gloveAverage 0.14 mg/glove (Meets requirements)
    Chemotherapy Drugs PermeationASTM D6978-05 (2019)Varies per drug; generally, high breakthrough times for most drugs, with specific values for Carmustine and Thiotepa (e.g., >240 minutes for many, as low as 12.1 for Carmustine)Carmustine (3.3 mg/ml): 12.1 minutes Thiotepa (10.0 mg/ml): 37.8 minutes Other 21 listed drugs: > 240 minutes (Meets specified breakthrough times; noted warnings about Carmustine usability)
    Fentanyl Citrate PermeationASTM D6978-05 (2019)>240 minutes>240 minutes (Meets requirements)
    In Vitro CytotoxicityISO 10993-5:2009Not cytotoxicThe neat extract was found to be cytotoxic. (This is a deviation from the acceptance criteria, but the device was still cleared, likely due to overall risk-benefit and other biocompatibility tests. The summary highlights this as a direct finding, not stating it "meets" the non-cytotoxic criterion.)
    Dermal SensitizationISO 10993-10:2021Not a sensitizerUnder the conditions of the study, the device is not a sensitizer. (Meets requirements)
    Primary Skin IrritationISO 10993-23:2021Not an irritantUnder the conditions of the study, the device is not an irritant. (Meets requirements)
    Acute Systemic ToxicityISO 10993-11:2017No systemic toxicity concernUnder the conditions of this study, the device showed no evidence of acute systemic toxicity. (Meets requirements)
    Shelf-Life ExpiryASTM D7160-16(Not explicitly listed as an acceptance criteria in the table, but reported as a characteristic)3 years (Identical to predicate device)

    (Note: The discrepancy for In Vitro Cytotoxicity is explicitly stated in the document as "found to be cytotoxic" while the acceptance criteria is "not cytotoxic." This indicates the device did not meet this specific acceptance criterion, but approval was still granted, implying the FDA found other evidence or risk mitigation sufficient.)

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not specify the exact sample sizes (e.g., number of gloves) used for each individual test (test set). It refers to the standards used (ASTM, ISO), which inherently define the testing methodology, including sample sizes for certain evaluations. However, the specific number of units tested for each parameter in this particular study is not provided in the summary.

    Data Provenance: The document does not explicitly state the country of origin of the data collectors or the testing laboratories, but the applicant is "Hartalega NGC Sdn. Bhd." located in Malaysia. The tests are non-clinical (laboratory testing of the physical device), not human subject data. The testing appears to be prospective in the sense that the tests were conducted specifically to support this 510(k) submission for this new device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This section is Not Applicable to this device and study type. The "ground truth" for this medical device (examination glove) is established through standardized, objective laboratory measurements as defined by ASTM and ISO standards for physical properties, chemical resistance, and biocompatibility. There is no human interpretative "ground truth" (e.g., expert consensus on images) involved, as there would be for an AI-based diagnostic device.

    4. Adjudication Method for the Test Set

    This is Not Applicable. Adjudication methods (e.g., 2+1, 3+1) are typically used in studies involving human interpretation or machine learning models where subjective assessment, diagnostic disagreement, or ground truth establishment based on expert consensus is required. The testing demonstrated here involves objective physical and chemical measurements.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    This is Not Applicable. An MRMC study assesses the performance of human readers, typically in a diagnostic setting, often with and without AI assistance. This submission describes a physical medical device (examination glove) and its non-clinical performance against established standards, not a diagnostic or AI-assisted product.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This is Not Applicable. This is not an algorithm or AI product, but a physical examination glove.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance is derived from:

    • Standardized Test Methods and Protocols: As defined by ASTM (American Society for Testing and Materials) and ISO (International Organization for Standardization) standards. These standards specify how to objectively measure various properties (e.g., dimensions, tensile strength, permeation time, cytotoxicity).
    • Laboratory Measurements: Direct, quantitative measurements of the glove's physical, chemical, and biological interactions according to the established standard methods.

    8. The Sample Size for the Training Set

    This is Not Applicable. There is no "training set" as this is not an AI/ML device. The testing described is for a physical product, not a statistical model that requires training data.

    9. How the Ground Truth for the Training Set was Established

    This is Not Applicable as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222225

    Validate with FDA (Live)

    Date Cleared
    2022-11-11

    (109 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs (Blue) is a non-sterile disposable device intended for medical purpose that is worn on the examination between patient and examiner. It is also tested to be used against Chemotherapy Drugs.

    These gloves were tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

    Device Description

    Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs (Blue) is a disposable singleuse, non-sterile, blue-colored and powder-free examination glove made from nitrile latex.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for a medical device: "Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs (Blue)". It describes the acceptance criteria and the study that proves the device meets those criteria, specifically for its resistance to chemotherapy drugs.

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Resistance to Permeation by Chemotherapy Drugs (ASTM D6978-05 (Reapproved 2019))For most chemotherapy drugs, a minimum breakthrough detection time (e.g., >240 minutes) is desired for adequate protection. Some drugs may have shorter accepted times based on their specific properties and risks. The predicate device (K151997) also established similar benchmark values.The device reports the following minimum breakthrough detection times: - Carmustine (3.3 mg/ml): 10.2 minutes - Thiotepa (10.0 mg/ml): 30.2 minutes - All other 31 listed chemotherapy drugs (e.g., Cisplatin, Cyclophosphamide, Doxorubicin, Fluorouracil, Methotrexate, Paclitaxel, etc.): >240 minutes Note: A caution explicitly warns against use with Carmustine and Thiotepa due to the shorter breakthrough times.
    Freedom from holes (ASTM D5151-19 and ASTM D6319-19)AQL 1.5%Pass
    Residual Powder (ASTM D6124-06 (2017))Average less than 2 mg/glovePass
    Dimensional Conformance (ASTM D6319-19)Conforms to ASTM D6319 width, thickness, and length requirements for XS, S, M, L, and XL (AQL 4%) - Length: ≥ 230 mm - Width (XS: 70 ± 10 mm, S: 80 ± 10 mm, M: 95 ± 10 mm, L: 110 ± 10 mm, XL: 120 ± 10 mm) - Palm Thickness: Min 0.05 mm - Finger Thickness: Min 0.05 mmPass (Meets specified length, width, palm thickness, and finger thickness requirements. The document specifically states for length "≥ 230 mm" for the subject device and "≥ 240 mm" for the predicate, but both are considered "Same" with a combined ASTM D6319-19 approval)
    Tensile Performance (ASTM D6319-19)Conforms to ASTM D6319 tensile strength of at least 14 MPa and ultimate elongation of at least 500% requirements prior to aging, and tensile strength of at least 14 MPa and ultimate strength of at least 400% after accelerated aging (AQL 4%)Pass
    Biocompatibility: Skin Irritation (ISO 10993-10)Under the conditions of the study, not an irritant.Pass (Under the conditions of the study, the device is not an irritant)
    Biocompatibility: Skin Sensitization (ISO 10993-10)Under the conditions of the study, not a sensitizer.Pass (Under the conditions of the study, the device is not a sensitizer)
    Biocompatibility: Acute Toxicity (ISO 10993-11)Under the conditions of the study, no acute systemic toxicity.Pass (Under the conditions of this study, the device showed no evidence of acute systemic toxicity)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes used for each test set within the provided tables (e.g., how many gloves were tested for holes, or how many individual glove samples were exposed to each chemotherapy drug). However, it indicates compliance with recognized standards (ASTM D5151-19, ASTM D6319-19, ASTM D6124-06, ASTM D6978-05). These standards typically specify the minimum sample sizes required for testing.

    The data provenance is implied to be prospective testing conducted by the manufacturer (Hartalega NGC Sdn. Bhd.) to meet the requirements for their 510(k) submission. Given the company's address, the testing was likely conducted in Malaysia or an affiliated testing facility.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable as the evaluation of the device's performance against physicochemical and biological standards (like chemical permeation, physical properties, and biocompatibility) does not typically involve human experts establishing ground truth in the way a clinical diagnostic study would. The "ground truth" here is defined by the objective metrics and thresholds established by the ASTM and ISO standards.

    4. Adjudication Method for the Test Set

    This is not applicable. Adjudication methods (like 2+1, 3+1) are typically used in clinical studies involving human interpretation or subjective assessments to resolve discrepancies. The tests described are objective, standardized laboratory tests.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. The device is an examination glove, not an AI-powered diagnostic or assistive tool for human readers. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is entirely irrelevant to this device.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. The device is a physical medical device (glove), not an algorithm or software. Standalone algorithm performance refers to the assessment of an AI system without human intervention, which does not apply here.

    7. The Type of Ground Truth Used

    The ground truth used for performance assessment of the Nitrile Powder Free Examination Glove is based on objective, standardized laboratory test results and their comparison against pre-defined acceptance criteria set by international standards:

    • Chemotherapy Drug Permeation: Measured breakthrough times (in minutes) as per ASTM D6978-05 (Reapproved 2019). The standard itself establishes the methodology and criteria for evaluating resistance.
    • Physical Properties: Measurements against specifications outlined in ASTM D5151-19 (freedom from holes), ASTM D6319-19 (dimensional conformance, tensile performance), and ASTM D6124-06 (R17) (residual powder).
    • Biocompatibility: Results of tests performed according to ISO 10993-10 (skin irritation, skin sensitization) and ISO 10993-11 (acute systemic toxicity), which define the methods and expected outcomes for non-toxic materials.

    8. The Sample Size for the Training Set

    This is not applicable. The device is a manufactured product undergoing standardized testing, not a machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K214074

    Validate with FDA (Live)

    Date Cleared
    2022-05-26

    (150 days)

    Product Code
    Regulation Number
    878.4460
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Latex Powder Free Surgical Glove with Protein Labeling Claim of 50 Microgram or Less per Gram of Glove and Tested for Use with Chemotherapy Drugs is made of natural rubber intended to be worn by operating personnel to protect a surgical wound from contamination. It is also tested to be used against Chemotherapy Drugs.

    These gloves were tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) premarket notification from the FDA, specifically for a Latex Powder Free Surgical Glove with Protein Labeling Claim of 50 Microgram or Less per Gram of Glove and Tested for Use with Chemotherapy Drugs.

    The request asks for information typically found in an AI/Software as a Medical Device (SaMD) study, such as acceptance criteria, sample sizes for training and testing, expert qualifications, and MRMC studies. However, this document describes a physical medical device (surgical gloves), not an AI/SaMD.

    Therefore, I cannot provide the requested information because it is not applicable to the provided document. The document details the device's intended use and performance against chemotherapy drugs, but it does not involve any AI/ML components or studies of human-in-the-loop performance with AI.

    To directly answer your questions based on the provided document:

    1. A table of acceptance criteria and the reported device performance:
      The acceptance criteria for chemotherapy drug permeation are implicitly defined by the "Minimum Breakthrough Detection Time in Minutes" columns, and the reported device performance is the value in that column. The standard used is ASTM D6978-05 (Reapproved 2019):

      • Chemotherapy Drug and Concentration | Minimum Breakthrough Detection Time in Minutes (Acceptance Criteria & Reported Performance)
      • Carmustine (BCNU) (3.3 mg/ml) | 12.5
      • Cisplatin (1 mg/ml) | > 240
      • Cyclophosphamide (Cytoxan) (20.0 mg/ml) | > 240
      • Dacarbazine (10 mg/ml) | > 240
      • Doxorubicin HCI (2.0 mg/ml) | > 240
      • Etoposide (20.0 mg/ml) | > 240
      • Fluorouracil (50.0 mg/ml) | > 240
      • Methotrexate (25 mg/ml) | > 240
      • Mitomycin C (0.5 mg/ml) | > 240
      • Paclitaxel (6.0 mg/ml) | > 240
      • Thiotepa (10.0 mg/ml) | 13.6
      • Vincristine Sulfate (1.0 mg/ml) | > 240
      • Note: The document also explicitly warns against using the gloves with Carmustine (BCNU) and Thiotepa due to low permeation times.
    2. Sample sizes used for the test set and the data provenance: Not applicable to this type of device. The testing methodology for ASTM D6978-05 would define the sample sizes for glove permeation testing, but this is not detailed in the FDA letter. Data provenance would be from laboratory testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for physical device performance is established through standardized laboratory testing, not human expert consensus.

    4. Adjudication method: Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done: Not applicable. This is for AI/SaMD, not physical gloves.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.

    7. The type of ground truth used: Laboratory test results based on ASTM D6978-05 for chemotherapy drug permeation and potentially other standards for protein labeling claims.

    8. The sample size for the training set: Not applicable. This is a manufactured product, not an AI model.

    9. How the ground truth for the training set was established: Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211758

    Validate with FDA (Live)

    Date Cleared
    2021-08-03

    (57 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Polychloroprene Powder Free Examination Glove Tested for Use with Chemotherapy Drugs (Fusion Colour) is a nonsterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs.

    These gloves were tested for use with chemotherapy drugs as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

    Chemotherapy Drug and ConcentrationMinimum Breakthrough Detection Time in Minutes
    Carmustine (3.3 mg/ml)14.9
    Cisplatin (1.0 mg/ml)>240
    Cyclophosphamide (20.0 mg/ml)>240
    Dacarbazine (10.0 mg/ml)>240
    Doxorubicin Hydrochloride (2.0 mg/ml)>240
    Etoposide (20.0 mg/ml)>240
    Fluorouracil (50.0 mg/ml)>240
    Methotrexate (25.0 mg/ml)>240
    Mitomycin C (0.5 mg/ml)>240
    Paclitaxel (6.0 mg/ml)>240
    Thiotepa (10.0 mg/ml)22.9
    Vincristine Sulfate (1.0 mg/ml)>240

    Please note that Carmustine and Thiotepa have extremely low permeation times of 14.9 minutes. Warning: Do not use with Carmustine Warning: Do not use with Thiotepa

    Device Description

    Polychloroprene Powder Free Examination Glove Tested for Use with Chemotherapy Drugs (Fusion Colour) is a nonsterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs.

    AI/ML Overview

    The provided document describes the FDA 510(k) clearance for a "Polychloroprene Powder Free Examination Glove Tested for Use with Chemotherapy Drugs (Fusion Colour)". This is not an AI/ML device but a physical medical device (examination gloves).

    Therefore, the requested information (acceptance criteria and study details related to AI/ML device performance, ground truth, expert adjudication, MRMC studies, etc.) is not applicable to this document.

    The document focuses on the glove's performance against chemotherapy drugs, as detailed in the "Indications for Use" section. It reports breakthrough times for various chemotherapy drugs based on ASTM D6978-05 (Reapproved 2019) standards.

    Here's the relevant information that can be extracted from the provided text, modified to fit the closest possible interpretation of your request, even though it's not an AI/ML device:

    1. Table of Acceptance Criteria and Reported Device Performance

    Chemotherapy Drug and ConcentrationAcceptance Criteria (Minimum Breakthrough Detection Time)Reported Device Performance (Breakthrough Detection Time in Minutes)
    Carmustine (3.3 mg/ml)(Implicitly, would ideally be high)14.9
    Cisplatin (1.0 mg/ml)(Implicitly, would ideally be high)>240
    Cyclophosphamide (20.0 mg/ml)(Implicitly, would ideally be high)>240
    Dacarbazine (10.0 mg/ml)(Implicitly, would ideally be high)>240
    Doxorubicin Hydrochloride (2.0 mg/ml)(Implicitly, would ideally be high)>240
    Etoposide (20.0 mg/ml)(Implicitly, would ideally be high)>240
    Fluorouracil (50.0 mg/ml)(Implicitly, would ideally be high)>240
    Methotrexate (25.0 mg/ml)(Implicitly, would ideally be high)>240
    Mitomycin C (0.5 mg/ml)(Implicitly, would ideally be high)>240
    Paclitaxel (6.0 mg/ml)(Implicitly, would ideally be high)>240
    Thiotepa (10.0 mg/ml)(Implicitly, would ideally be high)22.9
    Vincristine Sulfate (1.0 mg/ml)(Implicitly, would ideally be high)>240

    Note: The "acceptance criteria" here are implied by the standard ASTM D6978-05 for assessing permeation. The higher the breakthrough time, the better the performance. The document explicitly issues warnings for Carmustine and Thiotepa due to their low breakthrough times, suggesting these values did not meet an implicit desired criterion for safe use.

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not explicitly stated in the document. The ASTM D6978-05 standard would define the sample size requirements for testing gloves against chemotherapy drugs.
    • Data Provenance: The document does not specify the country of origin of the data. The testing was conducted according to ASTM D6978-05 (Reapproved 2019), which is an American Society for Testing and Materials standard. The study is prospective in the sense that the gloves were specifically tested for this clearance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This is not an AI/ML device where expert consensus for ground truth is typically required. The "ground truth" here is the physical measurement of chemical permeation according to a standardized test method.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. Adjudication methods are relevant for subjective interpretations, typically in AI/ML performance evaluation. The ASTM D6978-05 standard is a physical chemical test with objective measurements.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is not an AI/ML device; therefore, no MRMC study, human readers, or AI assistance is relevant.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an algorithm or AI system. The testing describes the standalone performance of the physical glove.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for this device's performance is established by objective chemical permeation measurements obtained through standardized laboratory testing as per ASTM D6978-05.

    8. The sample size for the training set:

    • Not applicable. This is a physical device, not an AI/ML model that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable. No training set is involved.
    Ask a Question

    Ask a specific question about this device

    K Number
    K201772

    Validate with FDA (Live)

    Date Cleared
    2020-08-25

    (57 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Nitrile Powder Free Examination Gloves with Low Dermatitis Potential Claim and Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue) is a non-sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs and Fentanyl Citrate.

    These gloves were tested for use with chemotherapy drugs and Fentanyl Citrate as per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

    Device Description

    Nitrile Powder Free Examination Gloves with Low Dermatitis Potential Claim and Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue) is a non-sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

    AI/ML Overview

    This document is an FDA clearance letter for Nitrile Powder Free Examination Gloves. It focuses on the device's substantial equivalence to predicate devices and its testing against chemotherapy drugs and fentanyl citrate. The document is primarily regulatory and does not contain information about an AI/ML device study or its performance. Therefore, I cannot provide the requested details regarding acceptance criteria and a study proving an AI/ML device meets them.

    The provided text pertains to a medical device clearance for examination gloves and does not contain any information about an AI/ML device, its acceptance criteria, or a study to evaluate its performance.

    Therefore, I cannot answer your request based on the provided input.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201530

    Validate with FDA (Live)

    Date Cleared
    2020-08-05

    (58 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Fusion Dark Grey) is a non-sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs and Fentanyl Citrate.

    These gloves were tested for use with Chemotherapy Drugs and Fentanyl Citrate as per ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

    Device Description

    Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Fusion Dark Grey) is a non-sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device: "Nitrile Powder Free Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl, Citrate (Fusion Dark Grey)". This document does not pertain to an AI/ML medical device and therefore does not include the typical information about acceptance criteria, study design, and ground truth establishment relevant to AI/ML products.

    Instead, this document focuses on the chemical permeation resistance of the examination gloves to various chemotherapy drugs and Fentanyl Citrate. The "acceptance criteria" here relate to the minimum breakthrough detection time for these chemicals, as per a specific ASTM standard.

    Here's a breakdown of the information that can be extracted from the provided text, adapted to the requested format where possible, and noting where the requested information is not applicable (N/A) for this type of device:


    1. Table of Acceptance Criteria and Reported Device Performance (Breakthrough Detection Time for Chemical Permeation)

    The study referenced is based on ASTM D6978-05 (Reapproved 2019) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. The acceptance criteria are implied by the standard, where a longer breakthrough time indicates better protection. While specific 'acceptance criteria' values (e.g., "must be > X minutes") are not explicitly stated as pass/fail thresholds in this summary, the results demonstrate the performance against the standard's methodology. The crucial findings for two drugs are highlighted as warnings.

    Chemotherapy Drug and ConcentrationAcceptance Criteria (Implied by standard and context of demonstrating protection)Reported Device Performance (Minimum Breakthrough Detection Time in Minutes)
    Carmustine (3.3 mg/ml)Longer breakthrough time indicates better performance.12.9
    Cisplatin (1.0 mg/ml)Longer breakthrough time indicates better performance.>240
    Cyclophosphamide (20.0 mg/ml)Longer breakthrough time indicates better performance.>240
    Dacarbazine (10.0 mg/ml)Longer breakthrough time indicates better performance.>240
    Doxorubicin Hydrochloride (2.0 mg/ml)Longer breakthrough time indicates better performance.>240
    Etoposide (20.0 mg/ml)Longer breakthrough time indicates better performance.>240
    Fluorouracil (50.0 mg/ml)Longer breakthrough time indicates better performance.>240
    Methotrexate (25.0 mg/ml)Longer breakthrough time indicates better performance.>240
    Mitomycin C (0.5 mg/ml)Longer breakthrough time indicates better performance.>240
    Paclitaxel (6.0 mg/ml)Longer breakthrough time indicates better performance.>240
    Thiotepa (10.0 mg/ml)Longer breakthrough time indicates better performance.45.0
    Vincristine Sulfate (1.0 mg/ml)Longer breakthrough time indicates better performance.>240
    Azacytidine (25.0 mg/ml)Longer breakthrough time indicates better performance.>240
    Carboplatin (10.0 mg/ml)Longer breakthrough time indicates better performance.>240
    Docetaxel (10 mg/ml)Longer breakthrough time indicates better performance.>240
    Epirubicin (2.0 mg/ml)Longer breakthrough time indicates better performance.>240
    Gemcitabine (38 mg/ml)Longer breakthrough time indicates better performance.>240
    Ifosfamide (50 mg/ml)Longer breakthrough time indicates better performance.>240
    Irinotecan (20 mg/ml)Longer breakthrough time indicates better performance.>240
    Mitoxantrone (2.0 mg/ml)Longer breakthrough time indicates better performance.>240
    Oncovin (1.0 mg/ml)Longer breakthrough time indicates better performance.>240
    Oxaliplatin (5 mg/ml)Longer breakthrough time indicates better performance.>240
    Vinorelbine (10 mg/ml)Longer breakthrough time indicates better performance.>240
    Fentanyl Citrate Injection (100 mcg/2ml)Longer breakthrough time indicates better performance.>240

    Warnings: The document explicitly states: "Please note that Carmustine and Thiotepa have extremely low permeation times of 12.9
    minutes and 45.0 minutes respectively. Warning: Do not use with Carmustine. Warning: Do not use with Thiotepa." This suggests an implicit acceptance criterion where these breakthrough times are deemed insufficient for safe use, leading to a contraindication.

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: Not explicitly stated in the provided document. The ASTM D6978-05 standard would specify the number of glove samples to be tested per drug.
    • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). The testing would have been conducted in a laboratory setting according to the ASTM standard.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This is N/A for the current device. The "ground truth" here is the measured chemical breakthrough time, determined by laboratory instruments and testing procedures defined in the ASTM standard, not by expert interpretation.

    4. Adjudication method for the test set:

    • This is N/A for the current device. Chemical permeation testing does not involve adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This is N/A for the current device. This is not an AI/ML device, and no human reader study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • This is N/A for the current device. This is not an AI/ML device. The performance reported is for the physical glove material itself.

    7. The type of ground truth used:

    • The "ground truth" for this study is direct laboratory measurement of chemical permeation, quantified as Minimum Breakthrough Detection Time, as determined by the ASTM D6978-05 standard.

    8. The sample size for the training set:

    • This is N/A for the current device. There is no AI/ML model involved, hence no training set. The gloves are manufactured based on established material science and manufacturing processes, with testing conducted to verify performance characteristics.

    9. How the ground truth for the training set was established:

    • This is N/A for the current device, as there is no training set for an AI/ML model.
    Ask a Question

    Ask a specific question about this device

    K Number
    K200581

    Validate with FDA (Live)

    Date Cleared
    2020-04-25

    (51 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Biodegradable Nitrile Powder Free Examination Gloves Tested for Use with Chemotheraty Drugs and Fentany Citrate (Blue) is a non-sterile disposable device intended for medical purpose that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Cliemotherapy Drugs and Fentanyl Citrate

    Device Description

    Biodegradable Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue)

    AI/ML Overview

    This document is a 510(k) clearance letter for "Biodegradable Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs and Fentanyl Citrate (Blue)". It details the device's indications for use and performance related to permeation by chemotherapy drugs and Fentanyl Citrate.

    Here's the breakdown of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria are implied by the standard ASTM D6978-05 (Reapproved 2013) and the reported minimum breakthrough detection times for various chemotherapy drugs and Fentanyl Citrate. The goal is to demonstrate that the gloves provide a protective barrier for a sufficient duration. The reported performance refers to the breakthrough times achieved during testing.

    Chemotherapy Drug and ConcentrationAcceptance Criteria (Implied by standard and intended use)Reported Device Performance (Minimum Breakthrough Detection Time in Minutes)
    Carmustine (3.3 mg/ml)Protection against permeation21.4
    Cisplatin (1.0 mg/ml)Protection against permeation>240
    Cyclophosphamide (20.0 mg/ml)Protection against permeation>240
    Dacarbazine (10.0 mg/ml)Protection against permeation>240
    Doxorubicin Hydrochloride (2.0 mg/ml)Protection against permeation>240
    Etoposide (20.0 mg/ml)Protection against permeation>240
    Fluorouracil (50.0 mg/ml)Protection against permeation>240
    Methotrexate (25.0 mg/ml)Protection against permeation>240
    Mitomycin C (0.5 mg/ml)Protection against permeation>240
    Paclitaxel (6.0 mg/ml)Protection against permeation>240
    Thiotepa (10.0 mg/ml)Protection against permeation67.2
    Vincristine Sulfate (1.0 mg/ml)Protection against permeation>240
    Azacytidine (25.0 mg/ml)Protection against permeation>240
    Carboplatin (10.0 mg/ml)Protection against permeation>240
    Docetaxel (10 mg/ml)Protection against permeation>240
    Epirubicin (2.0 mg/ml)Protection against permeation>240
    Gemcitabine (38 mg/ml)Protection against permeation>240
    Ifosfamide (50 mg/ml)Protection against permeation>240
    Irinotecan (20 mg/ml)Protection against permeation>240
    Mitoxantrone (2.0 mg/ml)Protection against permeation>240
    Oncovin (1.0 mg/ml)Protection against permeation>240
    Oxaliplatin (5 mg/ml)Protection against permeation>240
    Vinorelbine (10 mg/ml)Protection against permeation>240
    Fentanyl Citrate Injection (100 mcg/2ml)Protection against permeation>240

    Note: The document explicitly states: "Please note that Carmustine and Thiotepa have extremely low permeation times of 21.4 minutes and 67.2 minutes respectively. Warning: Do not use with Carmustine". This implies that while the device was tested against Carmustine, its low breakthrough time means it does not meet a practical "acceptance" for safe use with that particular drug, leading to a specific warning. For other drugs, a breakthrough time of >240 minutes indicates good performance.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the sample size used for the test set (number of gloves or repetitions per drug). It also does not explicitly state the data provenance or whether the study was retrospective or prospective. The testing was conducted "as per ASTM D6978-05 (Reapproved 2013) Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs," which is a standardized testing method.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable to this type of device and study. The "ground truth" here is the physical permeation of chemicals through the glove material, measured in a laboratory setting using analytical techniques as defined by the ASTM standard. It does not involve expert interpretation or consensus.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This study involves objective laboratory measurements, not subjective evaluations requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a study on the barrier effectiveness of gloves against chemical permeation, not an AI-assisted diagnostic study involving human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This document describes a physical performance test of a medical device (gloves), not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" is the empirically measured breakthrough time of various chemotherapy drugs and Fentanyl Citrate through the glove material, determined according to the methodologies outlined in ASTM D6978-05. This is a direct physical measurement.

    8. The sample size for the training set

    Not applicable. This refers to the testing of a physical medical device, not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for this type of device testing.