Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K231363

    Validate with FDA (Live)

    Date Cleared
    2023-09-27

    (139 days)

    Product Code
    Regulation Number
    880.5200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hantech Pre-filled 0.9% Normal Saline Flush Syringes, are in flushing compatible intravenous administration sets and indwelling intravenous access devices. Use according to the manufacturer for the appropriate device.

    Device Description

    The Hantech Pre-filled 0.9% Normal Saline Flush Syringes is a polypropylene plastic syringe filled with 0.9% sodium chloride for injection, USP, and capped with a polypropylene syringe tip cap. The device will be marketed as 10mL syringe with a 3mL, 5mL or10mL fill volume according to the market needs. It is terminally sterilized by steam sterilization. The product has a shelf life of 1 years.

    AI/ML Overview

    The provided document (K231363) is a 510(k) summary for Hantech Pre-filled 0.9% Normal Saline Flush Syringes. This type of device is a sterile medical product, not an AI/ML-driven diagnostic or image analysis tool.

    Therefore, the concepts of acceptance criteria, test sets, ground truth establishment by experts, MRMC studies, standalone algorithm performance, and training sets do not apply in the context of this device and the information provided in the 510(k) summary.

    The document describes non-clinical bench testing to demonstrate the substantial equivalence of the Hantech Pre-filled 0.9% Normal Saline Flush Syringes to a legally marketed predicate device. The "acceptance criteria" here refer to the predefined standards and limits for the physical, chemical, and biological properties of the syringe and its contents, as well as the sterilization process, which are based on recognized international and national standards (e.g., ISO, USP).

    Here's an interpretation of the relevant information provided:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document provides a table summarizing the non-clinical testing performed. The "Acceptance criteria" column lists the standards/criteria the device must meet, and the "Conclusion" column indicates "Pass" for all tests, meaning the device met these criteria.

    NO.TestMethodAcceptance criteriaConclusion
    1Physical testing of syringeISO7886-1, ISO80369-7ISO7886-1, ISO80369-7 (specific parameters for appearance, scale, sliding, dimension, dead space, acidity/alkalinity, chemical performance)Pass
    2Sodium Chloride Injection, USP TestingUSP40-<791>, USP-<191>, USP, USP6-471, USP40-<241>, USP<233>, USP<232>pH: 4.5-7.0, Positive identification for Sodium/Chlorine, 95% ~ 105% content, limits for oxidizable substances, iron, ammonium salt, calcium salt, carbonate, sulfate, and extractable metals (Cd≤2 μg/L, Pb≤5 μg/L, As≤15 μg/L, Hg≤3 μg/L)Pass
    3Particulate matterUSP<788>≥10um, ≤6000; ≥25um, ≤600Pass
    4Biocompatibility testingISO10993-11, ISO10993-10, ISO10993-5, ISO10993-4, ASTM F756-17No systemic toxicity, Non-irritant, Non-pyrogen, Non-sensitizer, Non-cytotoxic, Non-hemolyticPass
    5SterilizationUSP<85>, USP31-<71>Bacterial endotoxins ≤0.5EU/mL, Sterility (conformance to USP31-<71>)Pass

    2. Sample Size Used for the Test Set and Data Provenance:

    The document explicitly states "Summary of clinical Testing: N/A," meaning no clinical trials were conducted. The testing described is bench testing (laboratory-based, non-human, non-animal testing) of the physical product, its components, and its contents. The sample sizes for each specific test (e.g., how many syringes were tested for dimensions, or how many batches for chemical composition) are not provided in this summary document. Data provenance would be the manufacturer's internal testing facilities in China. The testing is prospective in the sense that it was conducted specifically for this 510(k) submission to demonstrate compliance.

    3. Number of Experts and Qualifications to Establish Ground Truth for the Test Set:

    This question is not applicable. The "ground truth" for a medical device like a syringe is established by validated measurement methods and recognized standards, not by expert consensus on clinical findings. For example, the pH of the saline is measured by a pH meter against a standard, not by an expert's opinion.

    4. Adjudication Method for the Test Set:

    Not applicable. This concept is relevant for studies involving human interpretation or subjective assessments, like image reading, not for objective bench testing of a physical product.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    No, an MRMC study was not done. This type of study is typically performed for diagnostic devices (e.g., AI in radiology) where the performance of human readers, with and without AI assistance, is compared. This device is a pre-filled saline syringe.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    No, this concept is entirely irrelevant for a medical device such as a pre-filled saline syringe. This applies to AI/ML algorithms.

    7. The Type of Ground Truth Used:

    The "ground truth" (or more accurately, the reference standard) for this device's performance is based on:

    • Recognized consensus standards: International Organization for Standardization (ISO) standards (e.g., ISO 7886-1, ISO 80369-7, ISO 10993 series)
    • United States Pharmacopeia (USP) monographs and general chapters: (e.g., USP<791> for pH, USP<191> for identification tests, USP<788> for particulate matter, USP<85> for Bacterial Endotoxins, USP31-<71> for Sterility tests, USP monographs for Sodium Chloride Injection)
    • Manufacturer's internal specifications: Which must meet or exceed these recognized standards.

    8. The Sample Size for the Training Set:

    Not applicable. This device is not an AI/ML algorithm that requires a "training set."

    9. How the Ground Truth for the Training Set was Established:

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222739

    Validate with FDA (Live)

    Date Cleared
    2023-04-06

    (209 days)

    Product Code
    Regulation Number
    880.5570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Insulin Pen Needle is intended for use with pen injector device for subcutaneous injection of insulin. It is sterile and for single use only. It can be used by the patient at home or healthcare professionals at medical/health care centers.

    Device Description

    The Disposable Insulin Pen Needle consists of a needle tube, a needle hub, a needle container, a needle shield and a seal. The needle tube is a double-ended needle that can be assembled into the needle hub using UV glue. The needle hub has the means of needle assembly attachment to allow it to be screwed onto the pen-injector device. This allows the Non-Patient (NP) end of the needle to penetrate through the rubber septum of the pen injector cartridge. The Patient and NP ends of the needle are lubricated using silicon oil for ease of injection and rubber septum penetration. The needle shield is assembled over the Patient end of the needle to protect the point from damage and accidental needle sticks. This needle assembly is inserted into a needle container and sealed with a peel-away label to provide a sterile barrier and tamper evidence. The peelaway label is pre-printed with information, which includes the lot number and needle gauge / length. It is supplied with several models. Different models are distinguished by needle gauge and length. The Disposable Insulin Pen Needle is a single-use disposable device and is provided sterile (EO sterilization). It is non-toxic and nonpyrogenic.

    AI/ML Overview

    Here's a breakdown of the requested information based on the provided FDA 510(k) summary for the Hantech Medical Device Co., Ltd. Disposable Insulin Pen Needle (K222739).

    It's important to note that this document is a 510(k) premarket notification, which demonstrates substantial equivalence to a predicate device, rather than a full study proving a device meets novel acceptance criteria. Therefore, many of the typical AI/ML study components (like expert consensus, adjudication, MRMC studies, training set details) are not applicable here. This submission focuses on demonstrating that the new device performs similarly to an already approved device through established engineering and biocompatibility testing.


    1. Table of Acceptance Criteria and Reported Device Performance

    For this type of device, "acceptance criteria" are generally derived from recognized international and national standards, and "reported device performance" indicates conformity to these standards.

    Acceptance Criteria CategorySpecific Criteria (from referenced standards)Reported Device Performance
    BiocompatibilityAbsence of cytotoxicity (ISO 10993-5:2009)No cytotoxicity
    Absence of skin sensitization (ISO 10993-10:2010)No evidence of sensitization
    Absence of hemolysis (ISO 10993-4:2017)No evidence of hemolysis
    Acceptable intradermal reactivity (ISO 10993-10:2010)Not explicitly detailed as "acceptable" but implied by conformity. "No evidence of skin irritation" is noted.
    Absence of acute systemic toxicity (ISO 10993-11:2017)No systemic toxicity
    Non-pyrogenic (ISO 10993-11:2017)Non-pyrogenic
    Absence of subacute systemic toxicity (ISO 10993-11:2017)No Subacute Systemic Toxicity
    Acceptable particulate levels (USP 788)Implied by conformity, no specific value given.
    SterilizationSterility Assurance Level (SAL) of $10^{-6}$ (ISO 11135:2014)Sterilized by ethylene oxide gas SAL = $10^{-6}$
    Compliance with ethylene oxide sterilization residuals (ISO 10993-7:2008)Implied by conformity to ISO 10993-7.
    Packaging integrity and sterile barrier system requirements (ISO 11607-1:2019, ISO 11607-2:2019)Implied by conformity to these standards.
    Shelf life stability (based on stability study and accelerated aging - ASTM F1980-16)5 years shelf life
    PerformanceRequirements for hypodermic needles (ISO 7864:2016)Complied with ISO 7864
    Requirements for stainless steel needle tubing (ISO 9626:2016)Complied with ISO 9626
    Requirements for needle-based injection systems (ISO 11608-2:2012)Complied with ISO 11608-2
    LabelingCompliance with 21 CFR Part 801Meet the requirements of 21 CFR Part 801

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not specify a "test set" in the context of clinical data or AI model evaluation. Instead, testing refers to product compliance with international standards for medical devices. The sample sizes for these specific engineering and biocompatibility tests are not detailed in this summary, but would typically be defined by the relevant ISO standards (e.g., a certain number of units for sterility testing or mechanical strength tests).

    • Data Provenance: The tests were conducted by the manufacturer, Hantech Medical Device Co., Ltd., which is located in Ningbo, Zhejiang, China. The testing would be prospective in the sense that physical devices were manufactured and then tested to demonstrate compliance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Not applicable. For this type of medical device 510(k) submission, "ground truth" is not established by human experts interpreting clinical data. Instead, it's established by objective measurements against predefined engineering standards and biocompatibility requirements. The "experts" involved would be qualified laboratory technicians, engineers, and biocompatibility specialists carrying out the specified tests following established protocols.

    4. Adjudication Method for the Test Set

    Not applicable. There is no qualitative assessment by human readers that would require an adjudication method. Test results are quantitative measurements or assessments against scientific criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    Not applicable. This device is a physical medical product (insulin pen needle), not an AI/ML diagnostic or assistive technology. Therefore, no MRMC study or AI assistance evaluation was performed.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. As noted above, this is a physical medical device, not an algorithm or software.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this substantially equivalent device is the demonstrated compliance with recognized international and national standards for mechanical performance, sterility, and biocompatibility, as well as comparison to the characteristics of a legally marketed predicate device. This is based on objective laboratory and quality control testing, not clinical outcomes data, pathology, or expert consensus on interpretations.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/ML device that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. This is not an AI/ML device, so no training set or its associated ground truth establishment is relevant.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222672

    Validate with FDA (Live)

    Date Cleared
    2023-03-06

    (181 days)

    Product Code
    Regulation Number
    880.5860
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The product is intended for subcutaneous injection of U-40 insulin in the treatment of diabetes.
    The product is intended for subcutaneous injection of U-40 and U-100 insulin in the treatment of diabetes.

    Device Description

    The Disposable Insulin Syringe is designed for subcutaneous injection of a desired dose of U-100 and U-40 insulin and has capacity of 0.3mL, 0.5mL, and 1.0mL. The product is a single use, sterile, non-toxic, non-pyrogenic, disposable syringe that consists of a needle, barrel, needle cap, plunger, piston, and plunger cap(only type A). The needle is a single ended lubricated stainless-steel cannula. The needle is covered by a polypropylene needle cap which is placed on the barrel. The barrel is molded from polypropylene with the exterior permanently marked with the proper scale markings. The piston is attached to the plunger and a Polypropylene plunger cap is placed over the plunger rod(only type A). All products are EO sterilized. The subject device operates on the principle of a piston syringe and has a shelf life of 5 years.

    AI/ML Overview

    The provided submission is for a medical device (Disposable Insulin Syringe), not an AI/ML powered device. Therefore, a table of acceptance criteria and the study that proves the device meets the acceptance criteria in the context of AI/ML is not applicable. The information requested (multi reader multi case study, standalone performance, ground truth establishment, training set sample size, etc.) pertains to AI/ML device evaluations and cannot be extracted from this document, which describes the testing and comparison of a traditional medical device against a predicate device.

    The document discusses various performance data, including biocompatibility, sterilization, shelf life, and physical performance testing according to ISO and ASTM standards. These tests demonstrate the device's safety and effectiveness as a disposable insulin syringe, but not as an AI/ML system.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221038

    Validate with FDA (Live)

    Date Cleared
    2022-09-01

    (147 days)

    Product Code
    Regulation Number
    878.4040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Medical Masks are intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. They are in infection control practices to reduce the potential exposure to blood and body fluids.

    Device Description

    The Disposable Medical Masks are single use, three-layer, flat-folded masks with ear loops and nose clamp. The Disposable Medical Masks are manufactured with three layers, the inner and outer layers are made of non-woven fabric, and the middle layer is made of melt blown Fabric. The ear loops are held in place over the users' mouth and nose by two ear loops welded to the face mask. The ear loops are made of Polyamide & Spandex. The nose clamp in the layers of face mask is to allow the user to fit the face mask around their nose, which is made of galvanized iron wire wrapped with PE material. The Disposable Medical Masks will be provided in white, blue, and purple which are sold non- sterile and are intended to be single use, disposable devices.

    AI/ML Overview

    This document describes the acceptance criteria and the results of a study demonstrating that the "Disposable Medical Masks" device, manufactured by Hantech Medical Device Co., Ltd., meets these criteria.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test MethodologyPurposeAcceptance Criteria: ASTM F2100 Level 3Result
    Fluid Resistance, Synthetic Blood Penetration ASTM F1862Evaluate Resistance to penetration by synthetic blood (Minimum pressure in mmHg)≥ 29 samples out of 32 pass (AQL 4%) at 160 mmHg for level 3PASS (32 out of 32 passes at 160 mmHg for level 3)
    Particulate Filtration Efficiency ASTM F2299Evaluate Sub-micron particulate filtration efficiency at 0.1 micron (%) (PFE)≥ 98% (29 out of 32 pass)PASS
    Bacterial Filtration Efficiency ASTM F2101Evaluate Bacterial filtration efficiency (BFE) (%)≥ 98% (29 out of 32 pass)PASS
    Differential Pressure (Delta P) EN 14683 Annex CEvaluate Differential pressure (Delta-P)< 6.0 mmH2O/cm² (29 out of 32 pass)PASS
    Flammability, 16 CFR 1610Evaluate Flame spreadClass 1PASS
    Cytotoxicity, ISO 10993-5Demonstrate biocompatibility safety of the subject deviceNon-cytotoxicPASS
    Irritation, ISO 10993-10Demonstrate biocompatibility safety of the subject deviceNon-irritatingPASS
    Sensitization, ISO 10993-10Demonstrate biocompatibility safety of the subject deviceNon-sensitizingPASS

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: 3 nonconsecutive lots were used for testing, with 32 samples for each model of surgical mask for the non-clinical tests.
    • Data Provenance: The document does not explicitly state the country of origin or whether the data was retrospective or prospective. However, Hantech Medical Device Co., Ltd. is based in the PEOPLE'S REPUBLIC OF CHINA. It can be inferred that the testing was conducted for the purpose of this premarket notification, making it prospective in nature relative to the submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    This document describes performance testing for medical masks against established standards. The nature of these tests (e.g., fluid resistance, filtration efficiency, flammability) does not typically involve human expert consensus for "ground truth" establishment in the same way an AI diagnostic algorithm would. The results are obtained through objective physical and biological tests conducted in laboratories. Therefore, information about "experts used to establish ground truth" with specific qualifications is not applicable in this context. The "ground truth" is defined by the technical specifications and criteria within the referenced standards (e.g., ASTM F2100, ISO 10993).

    4. Adjudication Method for the Test Set

    Given that the tests involve objective measurements against predefined criteria and standards, an adjudication method like "2+1" or "3+1" (typically used for resolving discrepancies in expert interpretations) is not applicable. The results are quantitative and pass/fail based on the physical properties of the masks.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done. This type of study is typically performed for diagnostic devices where human readers interpret medical images or data, and their performance with and without AI assistance is evaluated. This submission is for medical masks, which do not involve human diagnostic interpretation.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    No standalone (algorithm-only) performance study was done. The device is a physical medical mask, not an algorithm or AI software.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance of these medical masks is defined by established international and national standards and their specific criteria. These include:

    • ASTM F2100 (Standard Specification for Performance of Materials Used in Medical Face Masks)
    • ASTM F1862 (Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood)
    • EN 14683 (Medical Face Masks - Requirements and Test Methods)
    • ASTM F2101 (Standard Test Method for Evaluating the Bacterial Filtration Efficiency)
    • ASTM F2299 (Standard test method for determining the initial efficiency of materials used in medical face masks to penetration by particulates)
    • 16 CFR 1610 (Standard for the Flammability of clothing textiles)
    • ISO 10993-5 (Biological Evaluation of Medical Devices - Tests For In Vitro Cytotoxicity)
    • ISO 10993-10 (Biological Evaluation of Medical Devices - Tests for Irritation and Skin Sensitization)

    8. The Sample Size for the Training Set

    This submission pertains to a physical medical device (disposable medical masks). There is no "training set" in the context of machine learning or AI algorithms for this type of device. The production of such devices involves manufacturing processes that are validated, but not "trained" in the computational sense.

    9. How the Ground Truth for the Training Set Was Established

    As there is no training set for a physical medical device in the context of an AI/ML algorithm, this question is not applicable. The "ground truth" for manufacturing quality relies on process validation and quality control measures against established product specifications.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220603

    Validate with FDA (Live)

    Date Cleared
    2022-08-23

    (174 days)

    Product Code
    Regulation Number
    880.5570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Medical Safety Hypodermic Needle is intended for use in the aspiration and injection of fluids for medical purposes. The Disposable Medical Safety Hypodermic Needle is compatible for use with standard luer slip and luer lock syringes.

    Additionally, after withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needle-stick.

    Device Description

    The Disposable Medical Safety Hypodermic Needle is composed of a needle hub, a needle tube, a needle cap and a sharp injury protection feature. The sharp injury protection feature is simultaneously activated when manually pressed over the needle after use and prior to disposal to minimize the possibility of sharps injury. The sharp injury protection feature is activated with one-hand operation by pressing the sharp injury protection feature either with the finger or thumb, or by surface activation. The locking mechanism is positioned within the center and proximal end of the sharp injury protection feature. The hinge feature(sharp injury protection feature) allows the medical practitioner the flexibility to adjust the sharp injury protection feature to its desired position for use.

    AI/ML Overview

    This document describes the FDA's 510(k) clearance for the Hantech Medical Device Co., Ltd.'s Disposable Medical Safety Hypodermic Needle. The clearance is based on the device being substantially equivalent to a legally marketed predicate device (TK Safety Needle, K191644).

    Here's the breakdown of the acceptance criteria and the study that proves the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The device's performance is demonstrated through adherence to various international ISO standards, indicating it meets the acceptance criteria defined within those standards. The document doesn't provide specific numerical targets for "acceptance criteria" but rather states conformity to the clauses of these standards, which implicitly define the acceptance thresholds.

    Acceptance Criteria (Standard & Clause)Reported Device Performance (Conformity)Comments
    Biocompatibility:
    Cytotoxicity (ISO 10993-5: 2009)Performed in accordance with ISO 10993-1:2018 (body contact: "External communication device - Blood path indirect", < 24 hours)Device evaluated according to recommended tests for biocompatibility.
    Skin sensitization (ISO 10993-10: 2010)Performed in accordance with ISO 10993-1:2018 (body contact: "External communication device - Blood path indirect", < 24 hours)
    Hemolysis (ISO 10993-4: 2017)Performed in accordance with ISO 10993-1:2018 (body contact: "External communication device - Blood path indirect", < 24 hours)
    Intradermal reactivity (ISO 10993-10: 2010)Performed in accordance with ISO 10993-1:2018 (body contact: "External communication device - Blood path indirect", < 24 hours)
    Acute systemic toxicity (ISO 10993-11: 2017)Performed in accordance with ISO 10993-1:2018 (body contact: "External communication device - Blood path indirect", < 24 hours)
    Pyrogenicity (ISO 10993-11: 2017)Performed in accordance with ISO 10993-1:2018 (body contact: "External communication device - Blood path indirect", < 24 hours)
    Sterilization & Shelf Life:
    Sterilization (ISO 11135:2014)Validated to ISO 11135, routine control and monitoring parameters determinedDemonstrates effective sterilization.
    Ethylene oxide residuals (ISO 10993-7:2008)ConformsEnsures safe residual levels of sterilant.
    Packaging materials (ISO 11607-1: 2019)PerformedEnsures packaging maintains sterility.
    Packaging validation (ISO 11607-2: 2019)PerformedEnsures packaging integrity.
    Accelerated Aging (ASTM F1980-16)Used for shelf life determinationSupports shelf life claims.
    Performance Testing (ISO 7864: 2016 - Sterile Hypodermic Needles):
    Cleanliness (Clause 4.3)ConformsEnsures product is free from contaminants.
    Limits for acidity or alkalinity (Clause 4.4)ConformsEnsures biocompatibility and stability.
    Limits for extractable metals (Clause 4.5)ConformsEnsures patient safety.
    Tubular needle designation (Clause 4.6)ConformsEnsures correct labeling and identification.
    Colour coding (Clause 4.7)ConformsFollows industry standards for easy identification.
    Needle hub (Clause 4.8, ISO 80369-7 & ISO 6009)ConformsEnsures proper connection and compatibility.
    Needle cap (Clause 4.9)ConformsEnsures protection of the needle.
    Needle tube (Tolerance on length, Freedom from defects, Lubricant) (Clause 4.10)ConformsEnsures proper function and ease of use.
    Needle Point (Clause 4.11)ConformsEnsures sharpness and smooth penetration.
    Bond between Tube and Hub (Clause 4.12)ConformsEnsures structural integrity.
    Patency of Lumen (Clause 4.13)ConformsEnsures unobstructed fluid flow.
    Performance Testing (ISO 9626:2016 - Stainless steel needle tubing):
    Surface finish and visual appearance (Clause 5.2)ConformsEnsures quality and usability.
    Cleanliness (Clause 5.3)Conforms
    Limits for acidity and alkalinity (Clause 5.4)Conforms
    Size designation (Clause 5.5)Conforms
    Dimensions (Clause 5.6)Conforms
    Stiffness (Clause 5.8)ConformsEnsures needle rigidity.
    Resistance to breakage (Clause 5.9)ConformsEnsures needle integrity during use.
    Resistance to corrosion (Clause 5.10)ConformsEnsures material durability and patient safety.
    Performance Testing (ISO 80369-7:2016 - Luer connectors):
    Dimensional requirements for luer connectors (Clause 5)Conforms (2021 version mentioned implies adherence to latest standards)Ensures universal compatibility.
    Fluid leakage (Positive pressure liquid leakage) (Clause 6.1.3)Conforms (2021 version mentioned implies adherence to latest standards)Prevents leakage during aspiration/injection.
    Sub-atmospheric pressure air leakage (Clause 6.2)Conforms (2021 version mentioned implies adherence to latest standards)Prevents air aspiration.
    Stress cracking (Clause 6.3)Conforms (2021 version mentioned implies adherence to latest standards)Ensures material resilience.
    Resistance to separation from axial load (Clause 6.4)Conforms (2021 version mentioned implies adherence to latest standards)Ensures secure connection.
    Resistance to separation from unscrewing (Clause 6.5)Conforms (2021 version mentioned implies adherence to latest standards)Ensures secure connection.
    Resistance to overriding (Clause 6.6)Conforms (2021 version mentioned implies adherence to latest standards)Ensures proper connection.
    Performance Testing (ISO 23908 & Sharps Injury Prevention Guidance):Specific to the safety feature.
    Activation torquePerformedEnsures proper function of safety mechanism.
    Protection features separating forcePerformedEnsures integrity of safety mechanism.
    Resistance to breakagePerformedEnsures durability of safety mechanism.
    Simulated Clinical Use Testing ReportPerformedSimulates real-world use to confirm safety and effectiveness.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the specific sample sizes used for each test. Instead, it indicates that testing was performed according to various ISO standards. These standards typically define appropriate sample sizes and methodologies for testing.

    The data provenance is from China, as the submitting company, Hantech Medical Device Co., Ltd., is located in Ningbo, People's Republic of China. The studies are pre-market studies conducted to support the 510(k) submission, indicating a controlled, laboratory or simulated environment, rather than prospective clinical trials in humans.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not provided in the document. The studies primarily involve objective performance testing against established international standards rather than expert-driven ground truth establishment for a diagnostic or AI device.

    4. Adjudication Method for the Test Set

    This information is not applicable as the studies described are for the physical and functional performance of a hypodermic needle, not for diagnostic interpretation or a device requiring expert adjudication of results.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is typically conducted for diagnostic or imaging devices to assess the impact of AI on human reader performance. The device in question is a medical needle, not an AI or diagnostic device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    No, a standalone performance evaluation in the context of an algorithm or AI was not done. The device is a physical medical instrument.

    7. The Type of Ground Truth Used

    For the performance testing, the "ground truth" is defined by the specifications and requirements set forth in the referenced ISO and ASTM standards. These standards prescribe acceptable ranges, physical properties, and functional outcomes. For example, the cleanliness clause of ISO 7864 dictates the "ground truth" for acceptable levels of contamination.

    8. The Sample Size for the Training Set

    This information is not applicable as the device is a physical medical product and not an AI/machine learning model that would require a "training set."

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the same reason as above; there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1