510(k) Data Aggregation

    K Number
    K220549

    Validate with FDA (Live)

    Date Cleared
    2023-03-07

    (375 days)

    Product Code
    Regulation Number
    878.4840
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MINT Product Family (Including MINT Lift, and MINT-I Sutures) are comprised of PDO and are indicated for use in soft tissue approximation where use of a barbed absorbable suture is appropriate.

    These sutures are also indicated for use in face suspension surgery to temporarily fixate the cheek sub dermis in an elevated position.

    Device Description

    MINT Product Family Sutures are synthetic, absorbable, sterile, surgical sutures comprised of polydioxanone. These sutures are available in a range of lengths and thicknesses and have bi-directional barbs which hold the suture in position and exert tension throughout the length of the suture. These sutures are provided sterile. They degrade and reabsorb over time.

    AI/ML Overview

    The provided text is a 510(k) summary for the MINT Product Family of absorbable polydioxanone surgical sutures. It describes the device, its intended use, and argues for substantial equivalence to a predicate device. However, it does not contain the type of AI/ML device performance data, acceptance criteria, or study details (sample sizes, expert consensus, MRMC studies, etc.) that would typically be found for an AI/ML powered medical device.

    The document is for surgical sutures, a physical medical device, not an AI/ML software device. Therefore, the requested information about acceptance criteria for an AI/ML device, details of diagnostic performance studies, expert involvement in ground truth, etc., are not present in this document.

    The document discusses "non-clinical performance data" and "clinical performance data" related to the physical sutures, but these are for the material properties and safety/effectiveness of the suture itself, not for an AI/ML algorithm's diagnostic performance.

    Therefore, I cannot provide the requested information because the provided text is about a physical surgical suture device, not an AI/ML software device.

    The document does not describe:

    • Acceptance criteria related to AI/ML performance (e.g., sensitivity, specificity, AUC)
    • A test set size for an AI/ML model
    • Data provenance for AI/ML training/testing
    • Number/qualifications of experts for AI/ML ground truth
    • Adjudication methods for AI/ML ground truth
    • MRMC studies for AI/ML assistance
    • Standalone AI algorithm performance
    • Type of ground truth for an AI/ML model
    • Training set size for an AI/ML model
    • How ground truth for a training set was established for an AI/ML model.

    What the document does discuss (related to the physical suture device):

    • Non-Clinical Performance Data: Bench testing confirming compliance with USP standards for tensile strength, barb holding strength, suture-needle attachment strength, needle corrosion resistance, needle flexural stress, cannula pull testing, and needle penetration testing. Biocompatibility testing per ISO 10993.
    • Clinical Performance Data: A review of published clinical literature (10 relevant articles) from regions where the product is available. This literature involved over 500 patients with a minimum of 3 months follow-up (some 22+ months) supporting the safety and effectiveness of the sutures for face suspension surgery. The literature addressed complications associated with the sutures and procedure, not an AI algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K192423

    Validate with FDA (Live)

    Date Cleared
    2020-09-09

    (371 days)

    Product Code
    Regulation Number
    878.4840
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    MINT™ is indicated for use in mid-face suspension surgery to temporarily fixate the cheek subcutaneous fat layer and SMAS layer in an elevated position for the treatment of moderate to severe nasolabial folds.

    Device Description

    The MINT™ device is a sterile synthetic absorbable surgical suture comprised of polydioxanone, (GHoOJ)n. Polydioxanone has been found to be nonantigenic and to elicit only a slight tissue reaction during absorption. The pigment of the violet dye is D&C Violet No.2 (21CFR §74.3602). MINT™ is available in a range of gauge sizes and lengths. Each suture has bi-directional barbs along the long axis of the suture monofilament. The MINT™ Synthetic Absorbable PDO suture approximates tissues, without the need to tie surgical knots, by using the opposing barbs on the suture surface to embed in the tissues after the surgeon precisely places the suture within the tissues. Barbed suture lifting is a minimally invasive surgical technique for facial rejuvenation. The MINT™ Product Family includes models which are supplied with needles. For those models supplied with a needle, designated as "Lifting Thread Combined with needle" in the MINT™ Product Family instructions for use, the needle is used to make the insertion point and for threading of the suture in the patient's dermis.

    AI/ML Overview

    Here's an analysis of the provided text to extract the requested information about device acceptance criteria and the supporting study:

    Device: MINT™ Product Family (Absorbable Polydioxanone Surgical Suture) for mid-face suspension surgery.
    Indication: Temporarily fixate the cheek subcutaneous fat layer and SMAS layer in an elevated position for the treatment of moderate to severe nasolabial folds.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a "table of acceptance criteria" with quantifiable targets for the clinical study. However, the primary effectiveness endpoint and several secondary effectiveness evaluations serve as the de-facto acceptance criteria for clinical efficacy. The reported performance for these criteria is extracted below.

    Acceptance Criteria (Effectiveness Endpoint)Reported Device Performance
    Primary Effectiveness Endpoint: Ratio of subjects showing wrinkle improvement (WSRS: below -1 point) by independent evaluation at 12 weeks post-surgery.59 subjects (96.72%) showed improvement (WSRS: below -1 point). The lowest confidence level of 20.95% was greater than 0, indicating superiority.
    Secondary Effectiveness Evaluation (Independent Evaluators): Ratio of subjects showing improvement (WSRS: below -1 point) at 4, 8, and 24 weeks.Week 4: 100% (61 subjects) Week 8: 96.72% (59 subjects) Week 24: 90.16% (55 subjects)
    Secondary Effectiveness Evaluation (Testers): Ratio of subjects showing improvement (WSRS: below -1 point) at 4, 8, 12, and 24 weeks.Week 4: 100% (61 subjects) Week 8: 100% (61 subjects) Week 12: 100% (61 subjects) Week 24: 100% (61 subjects)
    Mean WSRS Change from Baseline (Independent Evaluators): Statistical significance of decrease in WSRS at 4, 8, 12, and 24 weeks compared to baseline.Baseline: 3.30 ± 0.45 points. Week 4: -1.56 ± 0.43 points Week 8: -1.41 ± 0.48 points Week 12: -1.23 ± 0.41 points Week 24: -1.20 ± 0.45 points. Statistically significant difference at all points (p<0.001).
    Mean WSRS Change from Baseline (Testers): Statistical significance of decrease in WSRS at 4, 8, 12, and 24 weeks compared to baseline.Baseline: 3.21 ± 0.41 points. Week 4: -1.23 ± 0.42 points Week 8: -1.25 ± 0.43 points Week 12: -1.18 ± 0.39 points Week 24: -1.16 ± 0.37 points. Statistically significant difference at all points (p<0.0001).
    Mean GAIS Assessed by Evaluators: Satisfaction level (e.g., above a certain threshold) at 12 and 24 weeks.Week 12: 1.93 ± 0.36 points Week 24: 1.85 ± 0.44 points. Distribution at Week 12: 86.89% (53 subjects) with 2 points ("very much improved"). Distribution at Week 24: 78.69% (48 subjects) with 2 points ("very much improved").
    Mean GAIS Assessed by Subjects: Satisfaction level (e.g., above a certain threshold) at 12 and 24 weeks.Week 12 and 24: 1.92 ± 0.63 points. Distribution at Week 12: 57.38% (35 subjects) with 2 points. Distribution at Week 24: 52.46% (32 subjects) with 2 points.
    Safety: Device is safe and effective as used according to the instructions for use.Clinical investigation supports that the device is safe and effective. (No specific quantifiable safety endpoints are detailed in this summary, but the conclusion states it.)
    Non-Clinical Performance: Compliance with USP standards for Suture diameter, Suture-Needle attachment, Suture tensile strength, and barb holding strength.Complies with USP <861> Suture diameter, USP <871> Suture-Needle attachment, and USP <881> Suture tensile strength. Displays superior barb holding strength (20.94 ± 3.93 N) compared to predicate (9.90 ± 0.93 N).
    Animal Performance (Absorption): Absorption timeframe.In vivo absorption occurs between 180-220 days post-implantation (on Monosorb, which is identical to the sutures).
    Animal Performance (Mechanical Strength): Tensile strength retention.Tensile strength retention of MINT™ (USP 1) was 44.5% at 6 weeks. Considered approximately 50% at 6 weeks. Unable to be measured at 10 and 12 weeks due to severe degradation.

    2. Sample Size for the Test Set and Data Provenance

    • Sample Size (Clinical Study): 62 male and female subjects (referred to as "FA set" in results, with 61 subjects for some analyses).
    • Data Provenance: Prospective clinical study. The country of origin of the data is not explicitly stated, but the sponsor, Hans Biomed Corporation, is located in the Republic of Korea, suggesting the study may have been conducted there.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Number of Experts: Unspecified, but referred to as "independent, blinded assessors" and "blinded evaluators." Also, "testers" (who may or may not be the same as independent evaluators) were involved.
    • Qualifications of Experts: Not specified. They are generally referred to as "evaluators" or "assessors."

    4. Adjudication Method for the Test Set

    The document mentions "Independent, blinded assessors" and that the analysis was based on "blinded evaluator Global Aesthetic Improvement Scale (GAIS) ratings" and "comparing the photos... by blinded evaluators." It also refers to "evaluation of independent evaluators" and "evaluation of testers."

    However, no specific adjudication method (e.g., 2+1, 3+1, majority vote, etc.) for resolving disagreements between multiple evaluators is described. It implies individual independent ratings were collected and analyzed (e.g., averaging or using individual scores).


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    • No, a MRMC comparative effectiveness study was not done in the sense of comparing human readers' performance with and without AI assistance, as this is a medical device (suture) and not an AI/CADe system for image analysis.
    • The clinical study evaluates the device's effectiveness through expert assessment of clinical outcomes (WSRS, GAIS) performed directly on subjects or their photos, not through a reader-AI interaction for diagnosis or interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    • Yes, a standalone performance was done for the device in the context of its mechanical and biological properties.
    • The "non-clinical performance data" (USP compliance, barb holding strength) and "animal performance data" (absorption, mechanical strength in vivo) represent standalone performance assessments of the physical characteristics of the suture itself, without human intervention in the device's inherent function.
    • The clinical study then assessed its performance in vivo in human subjects, which is the ultimate "standalone" performance for a medical implant like a suture.

    7. The Type of Ground Truth Used (Clinical Study)

    • Expert Consensus and Subjective Scales: The primary ground truth for the clinical study was established using validated subjective rating scales like the 5-grade Wrinkle Severity Rating Scale (WSRS) and the Global Aesthetic Improvement Scale (GAIS). These were applied by:
      • Independent, blinded assessors/evaluators.
      • Testers (likely clinicians involved in the study).
      • The subjects themselves (for GAIS).
    • This represents a form of expert assessment/consensus based on observable clinical outcomes, rather than objective pathology or hard outcomes data like mortality.

    8. The Sample Size for the Training Set

    • Not Applicable. This document describes a medical device (surgical suture), not an AI/Machine Learning model. Therefore, there is no "training set" in the context of machine learning. The clinical study served as the primary evidence for the expanded indication.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable, as there is no training set for an AI/ML model.
    Ask a Question

    Ask a specific question about this device

    K Number
    K171568

    Validate with FDA (Live)

    Date Cleared
    2018-02-22

    (268 days)

    Product Code
    Regulation Number
    888.3045
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SurFuse™ Gel, SurFuse™ Putty, ExFuse™ Gel, ExFuse™ Putty products are indicated for bony voids or gaps that are not intrinsic to the stability of the bone structure. They are intended to be gently packed into bony voids or gaps of the skeletal system as a bone void filler in the extremities and pelvis. These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. These products resorb and are replaced with bone during the healing process.

    Device Description

    The submitted devices are resorbable bone void filler, combining Human Demineralized Bone Matrix (DBM) with cancellous bone powder and carboxymethylcellulose (CMC). The primary component of SurFuse™ and ExFuse™ is demineralized particle bone which is derived from human donor cortical bone. The additional bone powder in the ExFuse™ is derived from human donor cancellous bone. The CMC is added to enhance the cohesiveness of the composition. The submitted devices are provided in several volumes ranging from 0.3cc to 10 cc. The devices are supplied sterile for single patient use.

    AI/ML Overview

    The provided text describes a 510(k) summary for medical devices (SurFuse™ Gel, SurFuse™ Putty, ExFuse™ Gel, ExFuse™ Putty), which are resorbable calcium salt bone void fillers. This type of submission focuses on demonstrating substantial equivalence to a predicate device, rather than detailed performance comparisons against acceptance criteria for an AI/software device.

    Therefore, much of the requested information (acceptance criteria, sample sizes for test/training sets, experts for ground truth, adjudication methods, MRMC studies, standalone performance, training set details) is not applicable or not present in this document, as it pertains to a different type of device evaluation (e.g., AI/ML-based diagnostic software).

    However, I can extract information related to the device's performance studies that were conducted to establish its substantial equivalence.

    Here's the breakdown of what can be extracted from the document:

    1. A table of acceptance criteria and the reported device performance

    This document does not specify quantitative acceptance criteria in the manner an AI study would (e.g., sensitivity, specificity thresholds). Instead, the performance is demonstrated qualitatively through in vivo studies to show osteoconductive and osteoinductive potential, indicating equivalence to predicate devices.

    Performance CharacteristicReported Device Performance
    OsteoconductionSuccessfully demonstrated bone growth in the in vivo rabbit unicortical defect model.
    Osteoinductive PotentialDemonstrated in the athymic (nude) rat muscle pouch model, showing new bone growth within muscle tissue. Correlated with in vitro BMP-2 ELISA assay results.

    2. Sample size used for the test set and the data provenance

    • Rabbit unicortical defect model: The text states "in vivo rabbit unicortical defect model," but does not explicitly provide the number of rabbits or defects used.
    • Athymic (nude) rat muscle pouch model: The text states "in vivo in the athymic (nude) rat muscle pouch model," but does not explicitly provide the number of rats used.
    • Data Provenance: The in vivo animal studies are experimental, not retrospective or prospective human clinical data. The bone material itself is derived from human donors in the United States and screened.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. These are in vivo animal studies (rabbit, rat) and in vitro assays, not studies requiring expert human interpretation of medical images or clinical outcomes. The "ground truth" would be the biological outcome observed and quantified in these animal models (e.g., new bone formation assessed histologically or radiographically by researchers).

    4. Adjudication method for the test set

    Not applicable. This is not a study involving human readers or clinical adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document describes bone void filler devices, not AI-powered diagnostic software.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a medical device, not an algorithm. The in vitro BMP-2 ELISA assay is an in vitro test, which is a standalone lab test, but not an algorithm.

    7. The type of ground truth used

    • For Osteoconduction and Osteoinductive Potential: The "ground truth" is established through direct observation and measurement of biological outcomes (new bone formation) in animal models. This would typically involve histological analysis, imaging (e.g., X-ray, micro-CT) and potentially biochemical markers.

    8. The sample size for the training set

    Not applicable. These are in vivo and in vitro performance studies for a medical device, not an AI/ML algorithm requiring a training set in the computational sense. Each lot of the device will be evaluated using the in vitro assay, implying ongoing quality control rather than a training set for an algorithm.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set for an AI/ML algorithm described.