Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K251053

    Validate with FDA (Live)

    Date Cleared
    2025-07-15

    (102 days)

    Product Code
    Regulation Number
    862.1155
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Shinetell Plus™ Digital Early Pregnancy Test is intended for the qualitative detection of human chorionic gonadotropin (hCG) in urine, as an aid in early detection of pregnancy, in some cases as early as five (5) days before the expected period, i.e., as early as six (6) days before the day of the missed period.

    Device Description

    Shinetell Plus™ Digital Early Pregnancy Test is used for in vitro qualitative detection of Human Chorionic Gonadotropin (HCG) in human urine, and is designed to be tested in dip or midstream mode. The test device consists of a single test strip assembled in a plastic housing, with an absorbent tip. The device is in a ready-to-use format.

    Shinetell Plus™ Digital Early Pregnancy Test uses lateral flow immunoassay and light reflection for the detection of the HCG in urine specimens. The test would detect the light intensity by using the LED as the light source. After that, the result can be displayed on the display screen.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the Shinetell Plus™ Digital Early Pregnancy Test, based on the provided 510(k) clearance letter:

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA clearance letter does not explicitly state "acceptance criteria" as a separate, quantitative table. However, we can infer the acceptance criteria from the performance studies presented, particularly the sensitivity and agreement rates. The primary criterion is demonstrating a sensitivity of 10 mIU/mL and high agreement with professional results and the predicate device.

    Performance MetricAcceptance Criteria (Implied)Reported Device Performance
    Analytical Sensitivity (LOD)10 mIU/mL (based on predicate and stated sensitivity)Midstream & Dip Testing: 100% positive at 10 mIU/mL hCG.
    Cross-ReactivityNo false positives with related hormones (hLH, hFSH, hTSH)No cross-reactivity observed with 1000 mIU/mL hLH, 1000 mIU/mL hFSH, 1000 μIU/mL hTSH. Not affected by hCG β-core fragment up to 500,000 pmol/L.
    InterferenceNo interference from common substances or urine propertiesNo interference observed at specified concentrations for various substances (Acetaminophen, Acetylsalicylic acid, Ascorbic acid, Atropine, Caffeine, Gentisic acid, Glucose, Hemoglobin, Tetracycline, Ampicillin, Albumin, β-hydroxybutyrate, Ephedrine, Phenylpropanolamine, Phenothiazine, EDTA, Salicyclic Acid, Benzoylecgonine, Cannabinol, Codeine, Ethanol, Bilirubin, Pregnanediol, Thiophene, Ketone). No effect from urine pH (4-9) or density (1.000-1.035).
    Hook EffectNo hook effect up to clinically relevant high concentrationsNo hook effect observed up to 500,000 mIU/mL hCG.
    Clinical Agreement (vs. Predicate)High agreement (e.g., >95%) with predicate deviceDip Testing: 100% conformity (49 Pos, 51 Neg vs Predicate). Midstream Testing: 100% conformity (48 Pos, 52 Neg vs Predicate).
    Lay Person Readability & InterpretationHigh agreement (e.g., >95%) between lay user and professional results; ease of use; clear labelingIn-Stream Method: 100% positive and 100% negative conformity with professional results (48 Pos, 52 Neg). Dip Method: 100% positive and 100% negative conformity with professional results (49 Pos, 51 Neg). Lay person results for spiked samples showed 98-100% agreement with professional results. Questionnaires indicated ease of use and clear labeling.
    Early Detection PerformanceDetect pregnancy early (e.g., 5 days before EMP)Detected 67.7% positive hCG five days before the Expected Menstrual Period (EMP). Detected 100% positive hCG on the day of EMP.
    StabilityStable for a reasonable shelf life24 months at 35.6-86°F (based on real-time stability study).

    2. Sample Sizes and Data Provenance

    • Analytical Performance (Precision/Reproducibility/Sensitivity):

      • Test Set Sample Size: For each hCG concentration (0, 3, 5, 7.5, 8, 10, 25, 50 mIU/mL), 10 replicates were tested per day for 5 days for each of 3 device lots, by 3 different operators. This sums to (8 concentrations * 10 replicates/day * 5 days * 3 lots) = 1200 tests for Midstream and another 1200 tests for Dip Testing.
      • Data Provenance: Not explicitly stated, but typically these are laboratory-controlled, prospective studies with spiked samples. The hCG standard is traceable to the 5th WHO.
    • Specificity (Non-pregnant females):

      • Test Set Sample Size: 300 urine samples (100 from pre-menopausal, 100 from peri-menopausal, 100 from post-menopausal women).
      • Data Provenance: Not explicitly stated, but implies prospectively collected urine samples from healthy non-pregnant females.
    • Method Comparison Study (vs. Predicate):

      • Test Set Sample Size: Urine samples from 200 women suspected to be pregnant in the early stage (<5 weeks).
      • Data Provenance: Collected from women presenting to test for pregnancy, implying a prospective, clinical setting, likely in an unspecified country (as no country of origin is mentioned).
    • Lay Person Study:

      • Test Set Sample Size: 200 women for self-testing. For spiked samples, 200 lay persons tested the samples.
      • Data Provenance: Individuals with varying educational and occupational backgrounds from three sites. This is a prospective, user study.
    • Early Pregnancy Test Study:

      • Test Set Sample Size: 650 urine samples from 65 pregnant women (65 characterized cycle segments of conceptive cycles).
      • Data Provenance: Collected from pregnant women. This is a prospective, clinical study.

    3. Number of Experts and Qualifications for Ground Truth for Test Set

    • Analytical Performance: The ground truth for hCG concentrations was established by spiking negative urine with a known hCG standard (Traceable to the 5th WHO). No human experts were involved in establishing this ground truth, as it's an analytical measurement.
    • Specificity (Non-pregnant females): The ground truth was based on the healthy, non-pregnant status of the female participants. No specific experts or their qualifications for this judgment are mentioned, but it's assumed to be based on medical history or clinical assessment.
    • Method Comparison Study: The ground truth was the result obtained from the predicate device (Wondfo One Step HCG Urine Pregnancy Test Midstream/Strip/Cassette, K150022). No human experts were involved in establishing a separate ground truth.
    • Lay Person Study (vs. Professional Results): The "Professional Result" served as the ground truth. The qualifications of these "professionals" are not specified.
    • Early Pregnancy Test Study: The "Expected Menstrual Period (EMP)" and "pregnant women" served as the ground truth, implying clinical determination of pregnancy and cycle dating. No specific experts or their qualifications for this determination are mentioned.

    4. Adjudication Method for the Test Set

    • No explicit adjudication method (like 2+1, 3+1) is mentioned for any of the studies.
    • For analytical performance, the ground truth was established by precise spiking of known concentrations.
    • For method comparison and lay person studies, direct comparison was made to the predicate device or "professional results" without specifying expert adjudication process if discrepancies arose.
    • For the early pregnancy test study, the ground truth seems to be clinical data (EMP, confirmed pregnancy) rather than expert review of the test results themselves.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC comparative effectiveness study was done to assess how much human readers improve with AI vs. without AI assistance.
    • This device is an Over-The-Counter (OTC) pregnancy test, where the "reader" is typically the lay user interpreting a digital display. The studies focused on the device's standalone performance, comparison to a predicate, and lay user comprehension/accuracy.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • Yes, a standalone study was done. The entire analytical performance section (sensitivity, hook effect, cross-reactivity, interfering substances) and the early pregnancy test study evaluated the device's technical performance to detect hCG and display a result ("Pregnant" or "Not Pregnant") without human interpretation beyond reading the digital display.
    • The "Professional Results" in the Lay Person study also reflect the device's standalone performance when used by trained individuals.

    7. Type of Ground Truth Used

    • Analytical Performance: Known concentrations of hCG spiked into negative urine (traceable to WHO International Standard). This is an analytical gold standard.
    • Specificity: Healthy, non-pregnant status of study participants. This is an outcome/status-based ground truth related to the absence of the target analyte.
    • Method Comparison: Results from a legally marketed predicate device.
    • Lay Person Study: "Professional Results" using the Shinetell Plus™ device. This indicates a reference test ground truth, where the "professionals" are presumably highly accurate users.
    • Early Pregnancy Test Study: Clinically determined pregnancy and cycle dating relative to the Expected Menstrual Period (EMP). This is outcomes data/clinical status ground truth.

    8. Sample Size for the Training Set

    • The document does not explicitly state a training set size or methodology for the Shinetell Plus™ Digital Early Pregnancy Test. This is common for this type of immunoassay device, which relies on established chemical and optical detection principles rather than machine learning algorithms that typically require large training datasets. The device's "learning" is inherent in its design and manufacturing.
    • The testing described is primarily for validation (verification and validation), not for training a predictive model.

    9. How the Ground Truth for the Training Set Was Established

    • As a training set is not explicitly mentioned or implied for an AI/machine learning model, this question is not directly applicable based on the provided document. The device operates on an immunoassay principle with "light reflection" and "photodiode" detection, suggesting a pre-programmed algorithm based on known HCG detection thresholds, rather than a learned model.
    Ask a Question

    Ask a specific question about this device

    K Number
    K241978

    Validate with FDA (Live)

    Date Cleared
    2024-08-13

    (39 days)

    Product Code
    Regulation Number
    862.1155
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ShinetellTM Digital Pregnancy Test is a pregnancy test. It is used for the qualitative detection of hCG in human urine as an aid in early detection of pregnancy. For in vitro diagnostic use, for over the counter use.

    Device Description

    Shinetell™ Digital Pregnancy Test is used for in vitro qualitative detection of Human Chorionic Gonadotropin (HCG) in human urine, and is designed to be tested in dip or midstream mode. The test device consists of a single test strip assembled in a plastic housing, with an absorbent tip. The device is in a ready-to-use format. Shinetell™ Digital Pregnancy Test uses lateral flow immunoassay and light reflection for the detection of the HCG in urine specimens. The test would detect the light intensity by using the LED as the light source. After that, the result can be displayed on the display screen.

    AI/ML Overview

    The provided document describes the Shinetell™ Digital Pregnancy Test, a device for the qualitative detection of hCG in human urine as an aid in early detection of pregnancy, for over-the-counter use. The document focuses on demonstrating the device's performance characteristics and its substantial equivalence to a legally marketed predicate device.

    Here's a breakdown of the acceptance criteria and the study that proves the device meets these criteria:

    1. Acceptance Criteria and Reported Device Performance

    The document doesn't explicitly list "acceptance criteria" in a separate table, but the performance characteristics evaluated serve as the de facto criteria for demonstrating substantial equivalence. The key performance metrics and their reported results are:

    Performance CharacteristicAcceptance Criteria (Implicit)Reported Device Performance
    SensitivityDetection of hCG at specified concentrations (e.g., 25 mIU/mL) with high positive rates and low false positive rates at lower concentrations.25 mIU/mL: 100% positive detection rate for both midstream and dip testing (overall 300/300 positive).12.5 mIU/mL: 0% positive detection rate (overall 0/300 positive).0 mIU/mL: 0% positive detection rate (overall 0/300 positive).The device demonstrated reproducible results, with operators and lots showing similar performance.
    Hook EffectNo false negative results at very high hCG concentrations.All tested concentrations (up to 500,000 mIU/mL) gave a positive result, demonstrating no hook effect.
    Specificity/Cross-ReactivityNo false positive results from healthy non-pregnant females or cross-reactive substances. No interference from HCG ß-core fragment.Non-pregnant females: 300 samples (100 from pre-menopausal, peri-menopausal, post-menopausal) tested by laypersons across both methods showed no false positives (100% negative).Cross-reactants: No cross-reactivity observed with 500 mIU/mL hLH, 1000 mIU/mL hFSH, 1000 µIU/mL hTSH at 0, 5, and 25 mIU/mL hCG spiked samples.hCG ß-core fragment: Performance not affected by concentrations up to 500,000 pmol/L.
    Interfering SubstancesPerformance of the device is not affected by common interfering substances found in urine.No interference effect observed with various substances at specified concentrations (e.g., Glucose 2000 mg/dL, Albumin 2000 mg/dL, Acetaminophen 20 mg/dL, etc.).Urine pH: Performance unaffected for pH 4-9.Urine Density: Performance unaffected for relative density 1.000-1.035.
    Method Comparison (vs. predicate)High conformity (agreement) between the new device and the predicate device.Midstream method: 100% conformity (agreement) between the candidate and predicate devices (52 positive, 48 negative, total 100 cases).Dip method: 100% conformity (agreement) between the candidate and predicate devices (41 positive, 59 negative, total 100 cases).
    Lay Person StudyHigh agreement between layperson results and professional results, and ease of use/understanding for lay users.First Study (self-testing): 100% positive and 100% negative conformity between layperson self-tests and professional results for both midstream (N=100) and dip (N=100) methods.Second Study (blinded spiked samples): 100% correct results for laypersons testing 5 mIU/mL hCG (negative) and 25 mIU/mL hCG (positive) spiked samples. Questionnaire results indicated ease of use and understanding.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Analytical Performance (Precision/Sensitivity):

      • Sample Size: For each hCG concentration (0, 12.5, 15, 18.75, 22.5, 25, 50, 100, 200 mIU/mL), 10 replicates were tested per day for 5 days for each of 3 device lots. With 3 operators, this resulted in 50 tests per operator per lot per concentration (10 replicates/day * 5 days), totaling 150 tests per concentration per lot.
      • Overall Sensitivity Sample Size: For the sensitivity conclusions, the most critical concentrations (0, 12.5, 25, 50, 100, 200 mIU/mL) each had 300 tests (3 lots * 50 replicates * 2 methods, or summarized directly as 300 total results).
      • Data Provenance: The document does not specify the country of origin for these spiked urine samples. It implies a laboratory setting for the spiking and testing, rather than patient samples. The study is retrospective in the sense that controlled samples were created for testing.
    • Specificity and Cross-Reactivity:

      • Non-pregnant females: 300 samples (100 from each age group: pre-menopausal, peri-menopausal, post-menopausal).
      • Cross-reactive substances/hCG ß-core fragment: Number of samples not explicitly stated per substance/concentration, but it involved "negative and positive female urine samples" spiked with the substances.
      • Data Provenance: Not specified, but implied to be collected samples.
    • Interfering Substances:

      • Number of samples not explicitly stated per substance/concentration, but involved "urine samples containing 0, 5 and 25 mIU/mL hCG" spiked with the substances.
      • Data Provenance: Not specified, implied to be collected urine samples spiked in a lab setting.
    • Method Comparison Study:

      • Sample Size: 200 women presenting to test for pregnancy, suspected to be pregnant (early stage, less than 5 weeks).
      • Data Provenance: Retrospective, collected from patients at three Point-of-Care (POC) sites. Country of origin not specified.
    • Lay Person Study:

      • First Study (self-testing): 200 women (individual pregnancy status self-tested).
      • Second Study (blinded spiked samples): 100 laypersons for 5 mIU/mL hCG aliquots and 100 laypersons for 25 mIU/mL hCG aliquots.
      • Data Provenance: "Individuals with varying educational and occupational backgrounds from three sites." Country of origin not specified. The first study used patient samples (their own urine); the second used controlled, spiked samples. Both are prospective in terms of the layperson testing event.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Analytical Performance (Precision/Sensitivity): Ground truth was established by the known concentration of hCG spiked into the negative urine samples, traceable to the 5th WHO International Standard. No human experts were used to interpret these results beyond confirming the digital readout.
    • Specificity, Cross-reactivity, Interfering Substances: Ground truth was established by the known negative status of the samples or the known spiked concentrations.
    • Method Comparison Study: The predicate device served as the "ground truth" for comparison. The study states "3 different professionals using the candidate device and 1 professional using the predicate device at each site." Their qualifications are not specified beyond "professionals."
    • Lay Person Study:
      • First Study (self-testing): The "professional results" served as ground truth for comparison with layperson results. The qualifications of these "professionals" are not specified.
      • Second Study (blinded spiked samples): The known spiked concentration (5 mIU/mL hCG meant to be negative, 25 mIU/mL hCG meant to be positive) acted as the ground truth. A "study administrator" was present to observe; their role was not to establish ground truth but to monitor the study.

    4. Adjudication Method for the Test Set

    The document does not describe a formal "adjudication method" involving multiple readers/experts to resolve discrepancies for the test sets.

    • For analytical performance, ground truth was by spiking, so no adjudication needed.
    • For method comparison and lay person studies, comparison was made against the predicate device results or professional results (in the case of actual patient samples). In the second layperson study, ground truth was by known concentration. Discrepancies (if any arose) or how they were handled (e.g., re-testing, exclusion) are not explicitly detailed.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. This is a digital pregnancy test with a clear "Pregnant" or "Not Pregnant" readout, not an imaging device requiring human interpretation for complex diagnostic decisions. Therefore, a multi-reader multi-case (MRMC) comparative effectiveness study to assess how human readers improve with AI vs. without AI assistance is not applicable to this device. The "human readers" (laypersons) are using the device to get a direct digital output, not making an interpretation with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, in essence, the "Analytical Performance" section (Precision/Sensitivity, Hook Effect, Specificity, Interfering Substances) represents the standalone performance of the device, as it evaluates the device's ability to detect hCG at various concentrations and in the presence of challenging substances, independent of human interpretation variability. The digital output is the algorithm's direct result.

    7. The Type of Ground Truth Used

    The types of ground truth used varied depending on the performance characteristic being evaluated:

    • Known Spiked Concentrations: For sensitivity, hook effect, cross-reactivity (with defined interferers), and the second layperson study. This is a highly controlled, artificial ground truth.
    • Clinical (Patient) Samples with Known Status (or compared to predicate/professional): For method comparison study and the first layperson study. Here, the "ground truth" was either the result from the legally marketed predicate device or the result from a "professional" using the candidate device, implying a clinical assessment of the patient's pregnancy status (e.g., confirmatory lab tests, clinical history).

    8. The Sample Size for the Training Set

    The document does not provide information on the training set size because this is not an AI/ML device that requires a separate "training set" in the conventional sense. This is an immunoassay device with a digital readout mechanism based on light intensity detection. Its "design" is based on biochemical and optical principles, not on learned parameters from a large dataset. Therefore, there's no "training set" as would be seen for, say, an image-recognition AI algorithm.

    9. How the Ground Truth for the Training Set Was Established

    As noted in point 8, there is no "training set" for this type of device in the context of an AI/ML algorithm. The device's operational parameters (e.g., thresholds for light intensity to display "Pregnant" or "Not Pregnant") are likely established during the device's development and internal validation processes using controlled, known concentrations of hCG, similar to the analytical performance testing described.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231698

    Validate with FDA (Live)

    Date Cleared
    2023-07-31

    (49 days)

    Product Code
    Regulation Number
    862.3650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AllTest Fentanyl Rapid Test (Urine) is an immunoassay intended for the qualitative detection of fentanyl in human urine at a cutoff concentration of 1.0 ng/mL. This in vitro diagnostic device is for prescription use only.

    The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS or LC/MS is the preferred confirmatory method.

    Device Description

    AllTest Fentanyl Rapid Test (Urine) is an immunoassay intended for the qualitative detection of fentanyl in human urine. Each AllTest Fentanyl Rapid Test (Urine) device consists of a Test Cassette and a package insert. Each Test Cassette is sealed with sachets of desiccant in an aluminum pouch.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the AllTest Fentanyl Rapid Test (Urine), based on the provided document:

    Acceptance Criteria and Device Performance

    The acceptance criteria are generally inferred from the positive and negative agreement with the LC/MS reference method, especially around the cutoff concentration. While explicit numerical acceptance criteria are not stated as "acceptance criteria," the performance in the method comparison study demonstrates the device's ability to classify samples correctly.

    Acceptance Criteria (Inferred)Reported Device Performance (Aggregate from 3 Sites)
    For Negative Samples (LC/MS < Cutoff): High agreement for definitively negative samples and samples below -50% of cutoff.Negative Samples: 90/90 (100%) correctly identified as negative (Low Negative) 36/45 (80%) correctly identified as negative (Near Cutoff Negative)
    For Positive Samples (LC/MS >= Cutoff): High agreement for definitively positive samples and samples above +50% of cutoff.Positive Samples: 63/63 (100%) correctly identified as positive (High Positive) 65/69 (94.2%) correctly identified as positive (Near Cutoff Positive)
    Overall Agreement with LC/MS: Demonstrate substantial equivalence to the predicate and accurate qualitative detection of fentanyl.Overall Performance: See detailed breakdown in "Discordant Results" for samples around the cutoff. The device shows strong performance for samples well above or below the cutoff.
    Precision: Consistent results across different lots and concentrations around the cutoff.For -25% Cutoff: - Lot 1: 58-/2+ - Lot 2: 59-/1+ - Lot 3: 60-/0+ For Cutoff: - Lot 1: 35+/25- - Lot 2: 28+/32- - Lot 3: 26+/34- For +25% Cutoff: - All lots 60+/0-
    Interference: No interference from common endogenous or exogenous substances at specified concentrations.Over 80 tested compounds showed no interference at 100 µg/mL or specified concentrations (except Trazodone with 0.1% cross-reactivity).
    Specificity (Cross-reactivity): Limited cross-reactivity with structurally similar compounds. Other opioids show no cross-reactivity.Detailed cross-reactivity provided for 17 fentanyl analogs; 21 other opioid compounds showed no cross-reactivity.
    Effect of Urine Specific Gravity and pH: Accurate results across a range of urine specific gravities and pH levels.Consistent results (positive for +50% Cut-Off and above, negative for -50% Cut-Off and below) over specific gravity 1.000-1.035 and pH 4-9.

    Study Details

    2. Sample Size for the Test Set and Data Provenance

    • Sample Size for Test Set: Operators ran 80 unaltered clinical samples at each of the three testing sites. This means a total of 240 tests were performed on the clinical samples (80 samples * 3 sites). Each sample was characterized by LC/MS results (40 negative and 40 positive for fentanyl).
    • Data Provenance: The document states "unaltered clinical samples." The country of origin is not specified, but the device manufacturer is based in Hangzhou, Zhejiang, China. The study appears to be prospective in the sense that the device was used to test these clinical samples, and the results were then compared to LC/MS.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • Ground Truth Establishment: The ground truth for the test set was established by LC/MS (Liquid Chromatography/Mass Spectrometry), which is the preferred confirmatory method for drug testing, rather than human experts.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable, as the ground truth was industrial standard LC/MS. Discordant results were identified by direct comparison to the LC/MS reference.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • MRMC Study: No, an MRMC comparative effectiveness study was not explicitly done. The study compares the device's performance against a gold standard (LC/MS), not against human readers with and without AI assistance. The device itself is an immunoassay, not an AI-assisted diagnostic tool.

    6. Standalone Performance Study

    • Standalone Performance: Yes, the described method comparison study is a standalone (algorithm only without human-in-the-loop performance) study, as it evaluates the device's ability to detect fentanyl in urine samples independently and compares its results to LC/MS. Human operators ran the tests, but their interpretation was based on the test line development, which is an intrinsic function of the device's chemistry.

    7. Type of Ground Truth Used

    • Type of Ground Truth: The ground truth used was established by LC/MS/MS results, specifically for fentanyl concentration, with a cutoff of 1.0 ng/mL. LC/MS is a highly accurate and sensitive analytical chemistry technique considered the gold standard for confirming the presence and quantity of drugs in biological samples.

    8. Sample Size for the Training Set

    • Sample Size for Training Set: The document does not explicitly mention a "training set" in the context of machine learning or AI. This device is an immunoassay and relies on chemical reactions, not on an algorithm that requires a training set. The various analytical performance studies (precision, interference, specificity) are internal validation studies that support the device's design and analytical capability, but these are not typically referred to as "training sets" in the same way as in AI/ML development.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set: Not applicable, as there is no explicitly defined "training set" for this immunoassay device in the context of AI/ML. All samples used in the analytical performance studies (precision, interference, specificity) involved prepared samples with known concentrations of fentanyl or interfering substances, often confirmed by LC/MS, or drug-free urine.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1