Found 2 results

510(k) Data Aggregation

    K Number
    K051430

    Validate with FDA (Live)

    Date Cleared
    2005-07-18

    (47 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Extreme H2O® 54% (hioxifilcon D) soft contact lens for daily wear is indicated for the correction of visual acuity in aphakic or not-aphakic persons with non-diseased eyes that are myopic or hyperopic. The spherical lens may be worn by persons who exhibit astigmatism of 0.75 Diopters or less that does not interfere with visual acuity. The toric lens may be worn by persons who exhibit astigmatism of up to 10.00 Diopters.

    Eye care practitioners may prescribe the lens for frequent/planned replacement wear, with cleaning disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfection system.

    Device Description

    Extreme H2O® 54% (hioxifilcon D) soft contact lenses are hemispherical shells and are available as spherical (G54 13.6 and G54 14.2) or toric (G54 Toric) lens designs. The Extreme H2O® 54% (hioxifilcon D) soft contact lens is fabricated from hioxifilcon D, which is a non-ionic, copolymer of 2-hydroxyethyl methacrylate (2-HEMA) and 2,3-Dihydroxypropyl Methacrylate (Glycerol Methacrylate, GMA) and cross-linked with ethylene glycol dimethacrylate (EGDMA). It consists of 46% hioxifilcon D and 54% water by weight when immersed in normal saline solution buffered with either sodium bicarbonate or sodium borate. The lens is available with a blue visibility handling tint. phthalocyanato (2) - (copper).

    In the hydrated state, the lens conforms to the curvature of the eye covering the cornea and extending slightly beyond the limbus forming a colorless, transparent optical surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard and appear somewhat warped. However, it will return to its proper configuration when completely rehydrated in the proper storage solution.

    AI/ML Overview

    Here is the information about the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Device Name: Extreme H2O® 54% (hioxifilcon D) Soft Contact Lens for Daily Wear

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission relies on a comparison to a predicate device (59% Extreme H2O® (hioxifilcon A) soft contact lens) and chemical/physical/toxicological testing to establish substantial equivalence. Direct "acceptance criteria" and "reported performance" for specific clinical outcomes are not explicitly stated as they would be for a novel device. Instead, the "acceptance criteria" are implied by demonstrating that the new device's material properties fall within the range of, or are comparable to, already cleared materials and predicate devices.

    PropertyAcceptance Criteria (Implied by Predicate/Cleared Materials)Reported Device Performance (Extreme H2O® 54%)Meets Criteria?
    Material ClassificationHydrophilic Lens Group IIHydrophilic Lens Group IIYes
    Method of ManufactureCast MoldedCast MoldedYes
    Water Content59% ± 2% (Predicate) / Bracketed by cleared materials54% ± 2%Yes
    Refractive Index1.404 hydrated (Predicate) / Bracketed by cleared materials1.414 hydratedYes
    Light Transmission> 95%> 95%Yes
    Specific Gravity1.308 (dry) (Predicate) / Bracketed by cleared materials1.300 (dry)Yes
    Oxygen Permeability (Dk units)28 (Predicate) / Bracketed by cleared materials21Yes
    Systemic Injection Test (Toxicology)No biological reaction (for cleared materials)No biological reaction within miceYes
    Agar Diffusion Test (Toxicology)Non-cytotoxic (for cleared materials)Non-cytotoxicYes
    Primary Ocular Irritation (Toxicology)Non-irritant (for cleared materials)Non-irritant to rabbitsYes
    Clinical Performance (Safety & Effectiveness)Substantially equivalent to predicate deviceNo clinical studies performed (justified)Yes

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not applicable in the context of a clinical study. The "test set" here refers to the new device (Extreme H2O® 54%), which underwent material characterization and toxicology testing.
    • Data Provenance: The toxicology tests were performed on the Extreme H2O® 54% material (hioxifilcon D). The document implies these were standard ISO tests. The country of origin and whether the data was retrospective or prospective is not specified for these specific tests, but they are generally carried out prospectively as part of product development.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    Not applicable. Ground truth, in the sense of expert consensus on clinical findings, was not established for this submission. The "ground truth" for the material properties and toxicology was based on established scientific and regulatory standards for material characterization and pre-clinical safety testing (e.g., ISO standards for toxicology).

    4. Adjudication Method for the Test Set

    Not applicable. There was no clinical test set requiring adjudication in the context of expert consensus or diagnostic accuracy.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. A multi-reader multi-case (MRMC) comparative effectiveness study was not done.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Not applicable. This device is a contact lens, not an AI algorithm. Its performance is inherent to the physical and chemical properties of the lens, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for this submission was based on:

    • Material Characterization: Scientific measurements of properties like water content, refractive index, light transmission, specific gravity, and oxygen permeability.
    • Toxicology Data: Results from standardized ISO safety tests (Systemic Injection Test, Agar Diffusion Test, Primary Ocular Irritation).
    • Substantial Equivalence to Predicate Device: The established safety and effectiveness of the legally marketed predicate device(s) (59% Extreme H2O® (hioxifilcon A) and Benz-G 3X (hioxifilcon B)).

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/machine learning device that uses a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K040303

    Validate with FDA (Live)

    Date Cleared
    2005-02-02

    (359 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The 59% Extreme H2O (hioxifilcon A) toric soft contact lens for daily wear is indicated for the correction of visual acuity in aphakic or not-aphakic persons with non-diseased eyes that are myopic or hyperopic. The lens may be worn by persons who exhibit astigmatism of 10.00 Diopters or less.

    Eye care practitioners may prescribe the lens for frequent/planned replacement wear, with cleaning disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfection system.

    Device Description

    59% Extreme H2O (hioxifilcon A) soft contact lenses are hemispherical shells and are available as spherical (G59 S-Thin and G59 S-Xtra) or toric (G59 Toric) lens designs. The 59% Extreme H2O (hioxifilcon A) soft contact lens is fabricated from hioxifilcon A, which is a non-ionic, copolymer of 2-hydroxyethyl methacrylate (2-HEMA) and 2,3-Dihydroxypropyl Methacrylate (Glycerol Methacrylate, GMA) and cross-linked with ethylene glycol dimethacrylate (EGDMA). It consists of 41% hioxifilcon A and 59% water by weight when immersed in normal saline solution buffered with either sodium bicarbonate or sodium perforate. The lens is available with a blue visibility handling tint, phthalocyanato (2) - (copper).

    AI/ML Overview

    The provided text is a 510(k) Premarket Notification for a soft contact lens and does not contain information about acceptance criteria, device performance metrics, or a study evaluating such performance.

    Instead, it focuses on demonstrating substantial equivalence to a previously cleared device. Therefore, I cannot provide the requested information.

    Here's why and what's missing:

    • No Acceptance Criteria or Performance Metrics: The document describes the device, its intended use, and argues for its substantial equivalence to another cleared product. It does not list specific performance criteria (e.g., visual acuity targets, comfort scores, lens fit parameters) that the new toric lens must meet, nor does it report data on how the device performed against such criteria.
    • No Study Described: There is no mention of a clinical trial or a formal study designed to assess the performance of the 59% Extreme H2O (hioxifilcon A) toric soft contact lens. The clearance is based on its similarity to an existing spherical lens, assuming similar safety and effectiveness.
    • No Information on Experts, Sample Sizes, or Ground Truth: Since no performance study is described, there is no information regarding sample sizes for test or training sets, data provenance, the number or qualifications of experts, adjudication methods, or the type of ground truth used.
    • No MRMC or Standalone Performance: These terms are relevant to assessing the performance of AI/CAD systems, which is not the subject of this 510(k) submission.

    Summary of what can be extracted related to the device:

    • Device Name: 59% Extreme H2O (hioxifilcon A) toric soft contact lens
    • Intended Use: Correction of visual acuity in aphakic or not-aphakic persons with non-diseased eyes that are myopic or hyperopic, who exhibit astigmatism of 10.00 Diopters or less.
    • Predicate Device: 59% Extreme H2O (hioxifilcon A) spherical lens (K992692)
    • Basis for Clearance: Substantial Equivalence to the predicate device, differing only in design (toric vs. spherical).

    To answer the questions posed, a document detailing a clinical or non-clinical performance study of the 59% Extreme H2O toric soft contact lens would be required.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1