    Device Name
    LIPCOLLECTOR II
    Manufacturer
    Date Cleared
    2010-06-29

    (11 days)

    Product Code
    Regulation Number
    878.5040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indications for use: Aesthetic body contouring
    Intended use: Harvesting, filtering and transferring of autologous fat tissue

    Device Description

    The LipoCollector II complete set consists of an autoclavable polymeric tissue collection container with a stainless steel lid and pre-fillering basket, and a disposable set including mesh filters. O-rings and inner tubing. The LipoCollector II complete set also includes a stainless steel extraction cannula (single-use) for withdrawal of the collection container. From the extraction cannula, the fat can be directly transferred to a Luer-Locksyringe (not part of the LipoCollector !! complete set) for autologous fat transfer. The intended use of the LipoCollector II complete set is to filter and collect the aspirated fat tissue for subsequent transfer of the autologous fat into subcutaneous fat tissues during the same procedure.

    AI/ML Overview

    The provided text describes the LipoCollector II complete set, a device for harvesting, filtering, and transferring autologous fat tissue. The submission focuses on demonstrating substantial equivalence to predicate devices and adherence to recognized standards rather than presenting a study to prove performance against specific acceptance criteria for a novel algorithm or diagnostic output.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not specify quantitative acceptance criteria for device performance (e.g., accuracy, sensitivity, specificity for a diagnostic device, or a specific metric for fat viability/purity). Instead, it relies on demonstrating compliance with recognized standards and substantial equivalence to predicate devices. The "acceptance criteria" discussed are related to design specifications, product function, safety, instructions for use, packaging, and risk management, which are general quality system requirements rather than specific performance metrics.

    Acceptance Criteria (Type)Reported Device Performance/Compliance
    Product Function, Safety, Instructions for Use, Packaging"Acceptance criteria for verification and validation testing of the device are based upon the approved product design specifications, including any additional design requirements imposed by Risk Management Process. Acceptance criteria for the LipoCollector II complete set include factors addressing product function, product safety, instructions for use, and packaging. All requirements of ISO 14971:2007 were met." (General statement of compliance)
    Biocompatibility- Cytotoxicity test (EN ISO 10993-5): No cytotoxic effects found. - Irritation and delayed-type hypersensitivity (EN ISO 10993-10): No negative effects observed in an epicutaneous test with 10 volunteers. - Endotoxin detection (EN ISO 10993-11): No endotoxins detected (LAL-Test). -> Device considered biocompatible for intended use.
    Sterilization (ETO)- Validated by physical and microbiological methods (ISO 11135-1:2007). - Sterilization safety level of 10^-6 guaranteed. - ETO residuals comply with ISO 10993-7.
    Sterile Packaging- Validated in accordance with ISO 11607-1, 11607-2, DIN 58953, EN 868-5. - Packaging is impervious to microorganisms.
    Reprocessing (Reusable Parts)- Reprocessing of reusable parts successfully validated according to AAMI/ANSI ST81:2004.
    Quality Management System- Implemented in accordance with DIN EN ISO 13485 and Council Directive 93/42/EEC.
    Design and Manufacturing Controls- In accordance with DIN EN ISO 13485 (Section 7.3) and 21 CFR 820.30. - Complies with DIN EN ISO 10079-1:2009, CSA Z 168.11-94, ISO 2768-1:1989, ISO 2768-2:1989, ISO 7153-1:1991/Amd.1:1999, and ISO 9626:1991/Amd.1:2001.
    Labeling- Complies with ISO 15223-1, FDA 21 CFR 801.1, and Canadian MDR labeling guidance.
    Risk Management- All requirements of ISO 14971:2007 were met.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set:
      • For the biocompatibility test (Irritation and delayed-type hypersensitivity - Epicutane test), 10 volunteers were used.
      • For other tests (cytotoxicity, endotoxin, sterilization, packaging, reprocessing), the "sample size" refers to tested components/devices, but specific numbers are not provided beyond the general statement that tests were conducted according to standards. The document doesn't define a "test set" in the context of device performance in humans or on clinical data, as it's not an AI/diagnostic device.
    • Data Provenance: Not explicitly stated for all tests. For the epicutaneous test, it's a prospective human study. For other tests, they are laboratory-based, not clinical data provenance. The company is based in Germany.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable as the device is not an AI/diagnostic system requiring expert readers for ground truth establishment on a performance evaluation test set. The "ground truth" for compliance testing is defined by the standards themselves (e.g., cytotoxicity present/absent, sterility achieved/not achieved).

    4. Adjudication Method for the Test Set

    This information is not applicable as there is no diagnostic output or interpretation by experts requiring adjudication for a test set.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This device is a surgical tool for fat processing, not an AI diagnostic assistant tool for human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    This information is not applicable. The LipoCollector II is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" used for this device's validation is primarily defined by:

    • Standard Compliance: Adherence to defined parameters and methodologies outlined in recognized international and national standards (e.g., ISO, EN, AAMI/ANSI, CFR).
    • Laboratory Test Results: Factual outcomes from specific biocompatibility (cytotoxicity, endotoxin, irritation), sterilization, and packaging tests.
    • Design Specifications: The device's ability to meet its pre-defined design and functional requirements.

    8. The Sample Size for the Training Set

    This information is not applicable as the device is a physical medical device, not an AI or machine learning model that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K082025

    Validate with FDA (Live)

    Device Name
    BODY-JET
    Manufacturer
    Date Cleared
    2008-08-13

    (27 days)

    Product Code
    Regulation Number
    878.5040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Aesthetic body contouring

    Device Description

    The body-jet®, a suction lipoplasty system, is a device intended for aesthetic body contouring. The device consists of a powered vacuum pump (containing an overflow/bacterial filter in the tube between the vacuum pump and suction bag), a collection bag with integrated overflow/bacterial filter, cannulae, and a connecting tube. The suction/collection bags including bacterial filters, the tubing, and the cannulae are to be changed between patients. The powered vacuum pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 26.5 inches of mercury, vacuum control valve to regulate the vacuum with accompanying vacuum gauge, a double piston, and a safety trap. The powered vacuum pump generates a negative pressure for the removal of fat or adipose tissue from distinct body-fat deposits for aesthetic body-contouring.

    The body-jet® includes a sterile, single-use tubing set with a small infiltration pump (the WAL Applicator), which transports the saline for the infiltration of the tissue area to be treated, and infiltration and irrigation cannulae. The single-use tubing set (WAL Applicator) and the reusable cannulae have been designed as proprietary products.

    AI/ML Overview

    The provided document is a 510(k) summary for the "body-jet®" suction lipoplasty system. It focuses on demonstrating substantial equivalence to predicate devices for aesthetic body contouring. The document does not describe acceptance criteria and a study proving the device meets those criteria in the typical format of a diagnostic or AI/machine learning device. Instead, it demonstrates safety and effectiveness through comparisons to existing devices and in-vitro studies.

    However, I can extract the relevant information and present it in a structured way that aligns with your request, interpreting "acceptance criteria" as the demonstration of "substantial equivalence" and "performance" as the findings supporting this equivalence for the described device, though it's not a direct match to a standard AI/diagnostic performance study.

    Here's an attempt to answer your questions based on the provided document, acknowledging that some fields may not be directly applicable or fully detailed for this type of device submission:


    1. Table of acceptance criteria and the reported device performance

    For a device like the body-jet®, the "acceptance criteria" are generally based on demonstrating substantial equivalence in terms of intended use, technological characteristics, safety, and effectiveness compared to legally marketed predicate devices. The performance data presented are primarily to support these claims.

    Acceptance Criteria (Demonstrated Equivalency to Predicate Devices)Reported Device Performance (Summary of Findings)
    Intended Use: Aesthetic body contouringThe body-jet® is intended for aesthetic body contouring, matching the intended use of predicate devices.
    Technological Characteristics:The vacuum pump system, infiltration pump system, and cannulae of the body-jet® are substantially equivalent to those of its predicate devices in terms of intended use, design, operating principles, and materials.
    Safety: No damage to vital structuresHistopathological/Immunohistological Performance Data (In-vitro): Two in-vitro laboratory investigations on human fat tissue samples (from corpses and "dead" abdominal fat excisates) showed that the fan-shaped saline spray produces no damage to vital structures such as connective tissue, muscle tissue, blood vessels, and nerves. This supports safety.
    Effectiveness: Loosens adipose tissue, separates/aspirates adipocytesHistopathological/Immunohistological Performance Data (In-vitro): The in-vitro investigations demonstrated that the water-assisted lipoplasty loosens adipose tissue and separates and aspirates adipocytes without damaging vital structures. Clinical Performance Data: Evaluation of published clinical performance data demonstrates that the body-jet® device is as safe, as effective, and performs as well as its predicate devices and other referenced lipoplasty techniques currently cleared for aesthetic body contouring. This supports its effectiveness in achieving aesthetic body contouring.
    Overall substantial equivalenceBased on all performance data, the body-jet® is concluded to be substantially equivalent to its predicate devices, and safe and effective for the intended use of aesthetic body contouring.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set:
      • In-vitro studies (Histopathological/Immunohistological Performance Data): "Two in-vitro laboratory investigations have been carried out. on fat tissue samples from human corpses (Investigation 1), and from "dead" abdominal fat excisates (Investigation 2)." The specific number of samples or excisates is not provided.
      • Clinical Performance Data: This refers to "published clinical performance data," indicating a review of existing literature. The specific number of patients or studies reviewed is not mentioned.
    • Data Provenance: Not explicitly stated for either the in-vitro studies or the published clinical data. "Human corpses" and "abdominal fat excisates" are mentioned for in-vitro. For clinical data, it's a review of "published" data, suggesting existing studies that could be from various countries and collected retrospectively or prospectively.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • The document does not specify the number of experts or their qualifications for evaluating the histopathological slides or for the clinical data review. The analysis summarized in "Section 18" (histopathology) and "Section 20" (clinical) would have been performed by qualified personnel (e.g., pathologists for histopathology, clinicians for clinical data review), but details are absent.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • This information is not provided in the document. It's unlikely to be applicable in the same way it would be for ground truth establishment in a diagnostic imaging study.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No MRMC comparative effectiveness study was done. This type of study (human readers with/without AI assistance) is typically relevant for diagnostic AI devices, not for a lipoplasty system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This is not applicable to the body-jet® device, which is a physical medical device (suction lipoplasty system), not an algorithm or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • For the in-vitro studies: Histopathological analysis and immunohistological performance data on tissue structures (adipose, connective, muscle, blood vessels, nerves). This implies assessment by pathologists.
    • For the clinical performance: A review of "published clinical performance data" to assess safety and effectiveness. This would involve synthesis of reported outcomes data from existing studies.

    8. The sample size for the training set

    • This device does not involve a "training set" in the context of machine learning or algorithms. The device's design and performance are based on engineering principles and pre-clinical/clinical validation, not on algorithm training.

    9. How the ground truth for the training set was established

    • This question is not applicable as there is no training set for this device.