510(k) Data Aggregation

    K Number
    K131866

    Validate with FDA (Live)

    Device Name
    SUSTAIN
    Date Cleared
    2014-05-01

    (311 days)

    Product Code
    Regulation Number
    884.5300
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Sustain Tailored Fit, Ultra Thin and Comfort Fit condoms are used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted diseases).

    Device Description

    The SUSTAIN Natural Rubber Latex Condoms are made of a natural rubber latex sheath, which completely covers the penis with a closely fitted membrane. These condoms conform to ISO 4074 Natural Rubber Latex Condoms - Requirements and Test Methods, ASTM D3492-08 Standard Specification for Rubber Contraceptives (Male Condom), and ISO 10993 Biological Evaluation of Medical Devices.

    AI/ML Overview

    This document is a 510(k) Summary for SUSTAIN Natural Rubber Latex Condoms. It describes the device, its intended use, and compares it to predicate devices to establish substantial equivalence.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document refers to established standards and specifications as acceptance criteria. The device performance is deemed to meet these criteria by conforming to them.

    Acceptance Criteria / StandardReported Device Performance (Conforms to)
    ASTM D3492-08 Standard Specification for Rubber Contraceptives (Male Condom)Yes
    ISO 4074 Natural Rubber Latex Condoms - Requirements and Test MethodsYes
    ISO 10993 Biological Evaluation of Medical DevicesYes
    ISO 10993-5 [2009] Tests for In Vitro CytotoxicityYes
    ISO 10993-10 [2010] Tests for Irritation and Skin Sensitization (Penile Irritation)Yes
    ISO 10993-10 [2010] Tests for Irritation and Skin Sensitization (Primary Irritation)Yes
    ISO 10993-10 [2010] Tests for Irritation and Skin Sensitization (Vaginal Irritation)Yes
    ISO 10993-10 [2010] Skin Sensitization Study (GPMT) of Refined Cotton Seed Oil Extract in Guinea PigsYes
    ISO 10993-10 [2010] Skin Sensitization Study (GPMT) of Normal Saline Extract in Guinea PigsYes
    ISO 10993-11 [2009] Tests for Systemic ToxicityYes
    Specific Product Specifications:
    Sustain Tailored Fit:
    Nominal Length186 ± 4 mm (meets ASTM D3492-08)
    Nominal Width49 ± 2 mm (meets ASTM D3492-08)
    Nominal Thickness0.065 ± 0.005 mm (meets ASTM D3492-08)
    Sustain Ultra Thin:
    Nominal Length186 ± 4 mm (meets ASTM D3492-08)
    Nominal Width52 ± 2 mm (meets ASTM D3492-08)
    Nominal Thickness0.055 ± 0.005 mm (meets ASTM D3492-08)
    Sustain Comfort Fit:
    Nominal Length186 ± 4 mm (meets ASTM D3492-08)
    Nominal Width52 ± 2 mm (meets ASTM D3492-08)
    Average Thickness0.065 ± 0.005 mm (meets ASTM D3492-08)

    Note: The document focuses on demonstrating substantial equivalence to predicate devices and conformance to recognized standards, rather than reporting specific numerical performance metrics from independent studies for each acceptance criterion with associated confidence intervals or statistical power. The "reported device performance" is essentially the statement that the devices "conform to" these standards.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the sample sizes used for each specific test described under "Nonclinical Testing," nor does it mention the data provenance (e.g., country of origin, retrospective or prospective). It simply lists the standards and the types of tests performed. For biological evaluation, the ISO 10993 series refers to standardized testing protocols, which would inherently include specified sample sizes for in vitro and in vivo tests (like guinea pig sensitization). However, these specific sample sizes are not detailed in this submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not applicable to this type of device submission. The "ground truth" for male latex condoms is established by their adherence to international and national standards (ASTM, ISO), which define performance characteristics for efficacy (e.g., prevention of pregnancy, STI transmission) and safety (e.g., biocompatibility, physical properties). These standards are developed by expert committees, but the individual "experts" establishing ground truth for a specific test set are not described in this document in the context of human interpretation of data for diagnostic purposes.

    4. Adjudication Method for the Test Set:

    Not applicable. This device is a physical medical device (condom) and its performance is evaluated against engineering/material standards and biological safety tests, not through human interpretation of clinical data requiring adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No MRMC comparative effectiveness study was done. This type of study is typically performed for diagnostic or imaging devices where human readers interpret medical images or data. The device in question is a physical barrier contraceptive.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    Not applicable. This device is not an algorithm or AI system. Its performance is inherent in its physical properties and biocompatibility.

    7. Type of Ground Truth Used:

    The "ground truth" used for this device's evaluation is based on established international and national consensus standards and specifications (ASTM D3492-08, ISO 4074, ISO 10993 series). These standards define the acceptable physical properties (length, width, thickness), burst strength, leakage, and biocompatibility requirements for male latex condoms to be considered safe and effective for their intended use. This is a form of standards-based ground truth.

    8. Sample Size for the Training Set:

    Not applicable. This device is not an AI or machine learning model, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for this device.