Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K113054

    Validate with FDA (Live)

    Device Name
    EQUIVITAL
    Manufacturer
    Date Cleared
    2012-02-17

    (127 days)

    Product Code
    Regulation Number
    870.1025
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Equivital™ Vital Signs Physiological Monitor is an ambulatory multiparameter vital signs telemetry device intended for monitoring of adults (16 - 65years) in hospital in the mital in the mater vital care facilities, the home, workplace for monitoning of addits (11) facilities, the home, workplace and alternate care settings.

    The device consists of a chest belt harness and a body worn electronics module (SEM) supported by the chest belt.

    The device collects and transmits ECG data and rate, respiration data and rate, skin temperature, body orientation and motion temperature, body orientation and motion.

    The monitor is indicated for use as a general patient monitor, to provide physiological information as part of an occupational welfare monitor, to provide physiology research and performance measuroment purpos research and performance measurement purposes.

    Device Description

    The Equivital™ Vital Signs Physiological Monitor comprises two components:

    i) A chest belt containing skin electrodes and an expansion sensor

    ii) A battery powered electronics module which connects directly to the chest belt and which acts to record, digitise and transmit the physiological information wirelessly to a receiving display device.

    The device offers continuous monitoring of two views of the user's heart electrical activity (ECG) and respiratory breathing frequency inferred from thoracic cavity movement and uses this data to derive a Heart Rate and a Breathing Effort Rate.

    The sensor also provides the following information:

    • ECG and Respiration physiological waveforms .
    • An indication of the user's activity level derived from a movement detection sensor .
    • Body orientation and indications for change in orientation .
    • Skin surface temperature .
    • Alternate secondary measurement of heart rate based on the detection of the user's R . wave using a separate hardware processing function
    • . Indications and alerts if physiology exceeds predefined boundaries
    • Alternate secondary skin temperature measurement using infrared thermometer .

    The device offers two variants for the wireless interface: a low power radio interface and a general purpose interface using Bluetooth™ technology.

    The device offers two battery power options for the user, rechargeable or primary disposable cells.

    The wireless data provided by the sensor may be viewed using a standalone PC based viewing application, or integrated into third party monitoring applications.

    AI/ML Overview

    This document describes the Equivital™ Vital Signs Physiological Monitor EQ02, but it does not contain a detailed study demonstrating its performance against specific acceptance criteria for each measured parameter.

    Instead, the submission focuses on demonstrating substantial equivalence to a previously cleared predicate device (Equivital EQ01) and compliance with relevant safety and electromagnetic compatibility standards.

    Here's a breakdown of the information requested, based on the provided text:


    1. Table of Acceptance Criteria and Reported Device Performance

    This table cannot be fully populated as specific performance acceptance criteria (e.g., accuracy ranges for heart rate, respiration rate, temperature) and the numerical results of studies proving these criteria were met are not explicitly provided in the given text.

    The document states that "Performance measurement and review to the applicable sections of the following standards has been undertaken and successfully demonstrated as recommended by available guidance from the agency." This implies that performance was verified against the requirements of these standards, but the specific numerical acceptance criteria and the device's measured performance against them are not detailed.

    ParameterAcceptance Criteria (from Standards, e.g., ANSI/AAMI EC13)Reported Device Performance (numerical results of testing)
    ECG Heart RateNot explicitly stated in document"successfully demonstrated as recommended by available guidance from the agency" against ANSI/AAMI EC13
    Respiration RateNot explicitly stated in document"successfully demonstrated as recommended by available guidance from the agency"
    Skin TemperatureNot explicitly stated in document"successfully demonstrated as recommended by available guidance from the agency" against ASTM E1965 98 (2009)
    Activity LevelNot explicitly stated in document"successfully demonstrated as recommended by available guidance from the agency"
    Body OrientationNot explicitly stated in document"successfully demonstrated as recommended by available guidance from the agency"
    Physiological AlertsNot explicitly stated in document"successfully demonstrated as recommended by available guidance from the agency" against ANSI/AAMI EC13
    SafetyEN60601-1"successfully demonstrated"
    EMCIEC60601-1-2"successfully demonstrated"
    BiocompatibilityISO 10993-1"successfully demonstrated"
    SoftwareEN60601-1-4"meets the requirements"
    EnvironmentalIP67"successfully demonstrated"

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The submission references "Performance measurement and review" and "testing undertaken" but does not detail the methodology, sample sizes, or data provenance for these tests.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the document. The document describes a medical device for physiological monitoring, not a diagnostic imaging or AI device requiring expert ground truth for image interpretation. The "ground truth" for vital signs monitoring would typically come from comparison to a highly accurate reference device or direct physiological measurements, not expert consensus in the same way an AI diagnostic model would.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided in the document. Adjudication methods like 2+1 or 3+1 are typically used in studies where human readers are making subjective interpretations (e.g., radiology studies). For a vital signs monitor, the comparison would be against a reference standard or objective measurements, not subjective human interpretations requiring adjudication.


    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted or reported for this device. This is a vital signs monitor, not an AI-powered diagnostic tool, so such a study would not be applicable in this context. The device's function is to collect and transmit physiological data, not to assist human readers in interpreting complex cases.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device itself is a "standalone" physiological monitor in the sense that it collects and processes data to derive heart rate, respiration rate, and other parameters directly from sensor inputs. The performance claimed (e.g., accurate heart rate calculation) is inherent to its algorithms. However, the document does not report a specific "standalone" performance study in the context of an AI algorithm being evaluated against human performance or ground truth. It refers to performance measurement against standards, which implies the device's accuracy in measuring various parameters.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The document does not explicitly state the specific "ground truth" methods used for the performance evaluations. However, based on the nature of medical device testing for vital signs monitors, the ground truth would typically be established by:

    • Comparison to a gold-standard reference device: For parameters like heart rate, respiration rate, and temperature, laboratories use highly accurate and calibrated reference instruments to provide the "true" value against which the device under test is compared.
    • Direct physiological measurements: In some cases, direct measurements (e.g., manual respiration counts, core body temperature probes) might be used as a reference.

    It would not typically involve expert consensus, pathology, or outcomes data for the primary performance assessment of vital sign accuracy.


    8. The sample size for the training set

    This information is not applicable and not provided. The Equivital™ Vital Signs Physiological Monitor EQ02 is described as a telemetry device and monitor, not an AI/machine learning model that undergoes "training" in the conventional sense with a distinct training set. Its functionality is based on established signal processing and sensor technologies, not adaptive learning from a large dataset.


    9. How the ground truth for the training set was established

    This information is not applicable and not provided. As stated above, the device does not use a "training set" in the context of machine learning. Its algorithms are likely based on fixed rules and signal processing techniques designed and validated through engineering principles.

    Ask a Question

    Ask a specific question about this device

    K Number
    K061993

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-11-03

    (112 days)

    Product Code
    Regulation Number
    870.1025
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Equivital™ Vital Signs Physiological Monitor is an ambulatory multiparameter vital signs telemetry device intended for monitoring of adults (16 - 65years) in hospital care facilities, the home, workplace, and alternate care settings.

    The device consists of a chest belt harness and a body worn electronics module (SEM) supported by the chest belt).

    The device collects and transmits ECG data and rate, respiration data and rate, skin temperature, body orientation and motion.

    The monitor is indicated for use as a general patient monitor, to provide physiological information as part of an occupational welfare monitoring system, and for general research and performance measurement purposes.

    Device Description

    The Equivital™ Vital Signs Physiological Monitor comprises two components:

    i) A chest belt containing skin electrodes and an expansion sensor

    ii) A battery powered electronics module which connects directly to the chest belt and which acts to record, digitise and transmit the physiological information wirelessly to a receiving display device.

    The device offers continuous monitoring of two views of the user's heart electrical activity (ECG) and respiratory breathing frequency inferred from thoracic cavity movement and uses this data to derive a Heart Rate and a Breathing Effort Rate.

    The sensor also provides the following information:

    • ECG and Respiration physiological waveforms. .
    • an indication of the user's activity level (none, low or high) derived from a movement . detection sensor.
    • . Body orientation.
    • Chest skin surface temperature. .
    • alternate secondary measurement of heart rate based on the detection of the user's R . wave using a separate hardware processing function.
    • indications and alerts if physiology exceeds predefined boundaries. .

    The device offers two variants for the wireless interface: a low power radio interface designed for military applications and a general purpose interface using Bluetooth™ technology.

    The device offers two battery power options for the user, rechargeable or primary disposable cells.

    The wireless data provided by the sensor may be viewed using a standalone PC based viewing application, or integrated into third party monitoring applications.

    AI/ML Overview

    The provided text describes the Equivital™ Vital Signs Physiological Monitor and its 510(k) summary. However, it does not contain specific acceptance criteria, detailed study designs, or performance results in the format requested. The document lists the standards against which the device was evaluated but does not present the actual results or how those results met specific criteria.

    Therefore, I cannot populate the table or provide detailed answers to many of the questions based solely on the provided text. I will indicate what information is available and what is missing.


    1. Table of Acceptance Criteria and Reported Device Performance

    Parameter / MetricAcceptance CriteriaReported Device Performance
    ECG Heart Rate PerformanceNot specified in documentNot specified in document
    Breathing Rate PerformanceNot specified in documentNot specified in document
    Temperature PerformanceNot specified in documentNot specified in document
    General Physical Properties PerformanceNot specified in documentNot specified in document
    Chest Expansion Respiration PerformanceNot specified in documentNot specified in document
    Activity/Motion Detection PerformanceNot specified in documentNot specified in document
    Bluetooth Telemetry PerformanceNot specified in documentNot specified in document
    Low Power 40.68MHz PAN PerformanceNot specified in documentNot specified in document
    Electrical Safety (EN60601-1)Compliance with standard"successfully demonstrated"
    EMC (IEC60601-1-2)Compliance with standard"successfully demonstrated"
    Biocompatibility (ISO 10993-1)Compliance with standard"successfully demonstrated"
    Cardiac Monitor Standards (ANSI/AAMI EC13)Compliance with applicable sections (not defibrillation/pacemaker immunity)"successfully demonstrated"
    Intermittent Temp. Spec. (ASTM E 112-00)Compliance with standard"successfully demonstrated"
    General Functional TestNot specified in document"undertaken"
    Environmental Performance TestNot specified in document"undertaken"
    Software Lifecycle (EN60601-1-4)Compliance with standard"meets the requirements"

    Study that Proves the Device Meets the Acceptance Criteria:

    The document states that "Performance measurement and review to the applicable sections of the following standards has been undertaken and successfully demonstrated as recommended by available guidance from the agency." This implies a series of verification and validation tests were conducted to ensure compliance with the listed standards (EN60601-1, IEC60601-1-2, ISO 10993-1, ANSI/AAMI EC13, ASTM E 112-00, and EN60601-1-4 for software). Additionally, "Additional voluntary testing has been undertaken to the following: General functional test, Environmental Performance Test."

    However, none of the specific study results or methodologies for these tests are described in the provided text. The document acts as a summary for a 510(k) submission, confirming that such testing occurred and was successful, rather than detailing the studies themselves.


    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: Not specified for any test set.
    • Data Provenance: Not specified. The manufacturer is based in the United Kingdom, but the origin of any test data is not mentioned within the summary.
    • Retrospective or Prospective: Not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not specified. The document does not describe the use of human experts for ground truth establishment for performance testing. Performance is noted to be measured against standards, implying instrument-based or standardized reference methods rather than expert consensus on a test set.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable/Not specified. The document does not describe a scenario requiring expert adjudication or consensus for ground truth on a test set.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study is not mentioned. The device is a "Vital Signs Physiological Monitor" and does not involve AI assistance for human readers/clinicians, but rather directly measures physiological parameters.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • The device itself is a standalone monitor, meaning its core function is to collect, digitize, and transmit physiological information. The performance measured would inherently be the algorithm/device only for parameters like heart rate, breathing rate, and temperature. The document states "Performance measurement and review to the applicable sections of the following standards has been undertaken and successfully demonstrated," suggesting performance evaluations of the device's output against established standards.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The implicit ground truth is likely reference measurements from established, validated medical devices or standardized calibration methods as per the requirements of the listed performance standards (e.g., ANSI/AAMI EC13 for cardiac monitors, ASTM E 112-00 for temperature). The document does not mention expert consensus, pathology, or outcomes data as ground truth.

    8. The sample size for the training set

    • Not applicable/Not specified. The document does not provide information about a "training set" as it describes a hardware physiological monitor, not typically a machine learning or AI-driven diagnostic device that undergoes explicit training on a dataset in the way a software algorithm might.

    9. How the ground truth for the training set was established

    • Not applicable/Not specified, as no training set is mentioned for this device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1