Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Heinrich Single-use non-stick bipolar forceps and single-use non-stick irrigating bipolar forceps are intended for use by a physician familiar with electrosurgery for bipolar coagulation and irrigation of tissue for general open surgery where coagulation of soft tissue is needed. Heinrich Single-use non-stick bipolar forceps and single-use non-stick irrigating bipolar forceps must be operated within the following parameters: -Frequency range between 300 kHz-1,000 kHZ; -Maximum generator operating voltage 600Vp. Heinrich Single-use non-stick bipolar forceps and single-use non-stick irrigating bipolar forceps have not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures and should not be used for these procedures. The types of surgery intended include: -ENT -Gynecology -Urology -General Surgery -Neurosurgery -Laryngeal Surgery -Orthopedic Surgery -Thoracic Surgery Heinrich Single-use non-stick bipolar forceps and Heinrich single-use non-stick irrigating bipolar forceps are connected through a suitable bipolar cable with the bipolar output of an electrosurgical generator. Heinrich single-use non-stick bipolar forceps and Heinrich single-use non-stick irrigating bipolar forceps must only be used with a bipolar coagulation current. HG Innovations, as a manufacturer, does not recommend a specific procedure. Use of bipolar techniques may be desirable in order to avoid unwanted tissue damage for surgical procedures where HF current could flow through relatively small cross-sectional area of body.

    Device Description

    HG Innovations Ltd.'s electrical surgical instruments encompass a wide variety of instruments that are widely used in the medical industry. Electrosurgical forceps and electrodes have been used in surgery for many years. The single-use, non-stick bipolar forceps and non-stick bipolar irrigating forceps (various sizes and tip configurations), with preattached cables are designed to grasp, manipulate, coagulate and irrigate soft tissues and are intended for use by a physician familiar with electrosurgery in bipolar coagulation for general open surgery where coagulation of soft tissue is needed. The blood vessel or tissue is grasped between the forceps tines, each of which acts as an electrode, and current passes to desiccate and coagulate the tissue. Bipolar forceps are connected through a suitable bipolar cable with the bipolar output of an electrosurgical generator. Bipolar forceps must only be used with bipolar coaqulation current. The bipolar forceps must be operated with the following parameters: -Frequency range between 300 kHz-1,000 kHZ: -Maximum generator operating voltage 600Vp.

    AI/ML Overview

    This is not a medical device that utilizes AI and therefore does not have acceptance criteria, study data, or ground truth information as typically found in AI/ML device submissions.

    The provided text is a 510(k) summary for Heinrich Single Use Non-Stick Bayonet Bipolar Forceps. This is a traditional medical device (bipolar forceps) and its clearance is based on substantial equivalence to predicate devices, not on performance metrics of an AI/ML algorithm.

    Therefore, I cannot provide the requested information. The document focuses on comparing the new device's features and materials to existing cleared devices (predicates) and ensuring it meets safety and effectiveness standards for such devices (e.g., electrical safety, biocompatibility, sterility).

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    McPherson Single-use non-stick bipolar forceps and single-use non-stick irrigating bipolar forceps are intended for use by a physician familiar with electrosurgery for bipolar coagulation and irrigation of tissue for general open surgery where coagulation of soft tissue is needed. Single-use non-stick bipolar forceps and single-use non-stick irrigating bipolar forceps must be operated within the following parameters:

    -Frequency range between 300 kHz-1,000 kHZ;

    -Maximum generator operating voltage 600Vp.

    Single-use non-stick bipolar forceps and single-use non-stick irrigating bipolar forceps have not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures and should not be used for these procedures. The types of surgery intended include:

    -ENT

    -Gynecology

    • -Urology
    • -General Surgery
    • -Neurosurgery
    • -Laryngeal Surgery
    • -Orthopedic Surgery
    • -Thoracic Surgery

    Single-use non-stick bipolar forceps and single-use non-stick irrigating bipolar forceps are connected through a suitable bipolar cable with the bipolar output of an electrosurgical generator. Single-use non-stick bipolar forceps and single-use non-stick irrigating bipolar forceps must only be used with a bipolar coagulation current.

    HG Innovations, as a manufacturer, does not recommend a specific procedure.

    Use of bipolar techniques may be desirable in order to avoid unwanted tissue damage for surgical procedures where HF current could flow through relatively small cross-sectional area of body.

    Device Description

    The single-use, non-stick bipolar forceps and single-use, non-stick irrigating bipolar forceps (various sizes, designs and tip configurations), with preattached cables are designed to grasp, manipulate, coagulate and irrigate soft tissues and are intended for use by a physician familiar with electrosurgery in bipolar coagulation for general open surgery where coagulation of soft tissue is needed. The blood vessel or tissue is grasped between the forceps tines, each of which acts as an electrode, and current passes to desiccate and coagulate the tissue. Bipolar forceps are connected through a suitable bipolar cable with the bipolar output of an electrosurgical generator. Bipolar forceps must only be used with bipolar coaqulation current. The bipolar forceps must be operated with the following parameters:

    -Frequency range between 300 kHz-1,000 kHZ: -Maximum generator operating voltage 600Vp.

    AI/ML Overview

    The provided document describes the FDA's "Substantial Equivalence" determination for a device, not a study to prove acceptance criteria for a new AI/software medical device.

    The document discusses various performance tests and comparisons to predicate devices for Single Use Non-Stick Bipolar Forceps and Single Use Non-Stick Irrigating Bipolar Forceps, which are mechanical electrosurgical accessories.

    Therefore, the requested information about acceptance criteria, study design (sample size, data provenance, ground truth, experts, adjudication, MRMC, standalone), and training set details for an AI/software device cannot be extracted from this document, as it pertains to a different type of medical device and a different regulatory review process.

    The document consistently states "N/A" for "Software Verification and Validation Testing," "Animal Study," and "Human Clinical Performance Testing," further confirming that this is not an AI/software medical device submission requiring such studies.

    However, I can provide a summary of the performance criteria and reported performance for the mechanical device as presented in the document:


    1. Table of Acceptance Criteria and Reported Device Performance

    Performance Test / FeatureAcceptance Criteria (Implied by "Pass" or "Same")Reported Device Performance (Subject Device)
    Biocompatibility (ISO10993-1)Demonstrably non-cytotoxic, non-irritating, systemically non-toxic, non-sensitizing, and passes pyrogenicity testing (as inferred from the justification for differences between subject and predicate where predicate did not report these tests, implying these are required for substantial equivalence). Meets the requirements of ISO 10993-5 (Cytotoxicity), ISO 10993-10 (Irritation & Sensitization), USP151 (Pyrogenicity), and ISO 10993-11 (Systemic Toxicity).Pass
    Electrical Safety & EMC (AAMI/ANSI IEC 60601-1, IEC 60601-2-2)Meets the specified standards for high-frequency leakage current.Pass
    High-frequency dielectric strengthMeets the specified standards for high-frequency dielectric strength.Pass
    Mains frequency dielectric strengthMeets the specified standards for mains frequency dielectric strength.Pass
    Active Accessory Insulation (for irrigating forceps)Meets the specified standards for active accessory insulation.Pass
    Mechanical TestingMeets the specified standards for mechanical performance.Pass
    Sterility Testing (ISO-10993-7)Sterilized by Ethylene Oxide (EO) and meeting the requirements of ISO-10993-7 for sterility. (Implied by the method used and regulatory acceptance of EO sterilization).Ethylene Oxide (EO)
    Shelf-Life TestingDemonstrates a product shelf life of at least 3 years through real-time aging studies.Shows product shelf life up to 3 years

    Study Proving Acceptance Criteria (General):

    The document indicates that the device's performance was evaluated through various bench performance tests. These tests covered biocompatibility, electrical safety and electromagnetic compatibility (EMC), mechanical testing, sterility testing, and shelf-life testing. The report concludes that "HG Innovation's bipolar forceps are identical to the predicate device in their intended use/indications for use and materials used for the arms, and tips. No further testing necessary," implying that the performance data from these bench tests demonstrated substantial equivalence to the predicate devices. The details of these studies are not provided in this summary, but the results are reported as "Pass" for the relevant tests.


    Information NOT Applicable/Available (as per the document's content for AI/Software specific questions):

    1. Sample size used for the test set and the data provenance: Not applicable. This is not an AI/software device. The document mentions "bench performance testing" which are laboratory-based physical tests, not data-driven tests on patient sets.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth and expert review are irrelevant for this type of mechanical device testing.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable.
    6. The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not applicable.
    7. The sample size for the training set: Not applicable. This is not an AI/software device.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K230432

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-05-18

    (90 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Single-use, bipolar forceps are intended for bipolar coagulation to remove tissue and control bleeding by use of high frequency electrical current.

    The bipolar forceps must be operated with the following parameters:

    -Frequency range between 300 kHz-1,000 kHZ;
    -Maximum generator operating voltage 600Vp.

    Device Description

    HG Innovations Ltd.'s Electrical Surqical Instruments encompass a wide variety of instruments that are widely used in the medical industry. Electrosurgical forceps and electrodes have been used in surgery for many years. The single-use, bipolar forceps (various lengths and designs), with preattached cables are designed to grasp, manipulate and coagulate soft tissues and are intended for use by a physician familiar with electrosurgery in bipolar coagulation for general open surgery where coagulation of soft tissue is needed. The blood vessel or tissue is grasped between the forceps tines, each of which acts as an electrode, and current passes to desiccate and coagulate the tissue. Bipolar forceps are connected through a suitable bipolar cable with the bipolar output of an electrosurgical generator. Bipolar forceps must only be used with bipolar coagulation current.

    AI/ML Overview

    This FDA 510(k) summary provides information for a Single Use Bipolar Forceps (K230432).

    Here's an analysis of the provided text in relation to acceptance criteria and supporting studies:

    Important Note: This document is a 510(k) Summary for a medical device (bipolar forceps), not a diagnostic AI software. Therefore, many of the typical AI/ML study components like multi-reader multi-case (MRMC) studies, expert ground truth adjudication for complex interpretations, and training set details are not applicable here. The "acceptance criteria" discussed are largely related to engineering performance and biological safety, not diagnostic accuracy.


    1. Table of Acceptance Criteria and Reported Device Performance:

    Criteria/TestAcceptance Criteria (Implied/Defined)Reported Device Performance/Result
    Biocompatibility (ISO 10993-1)
    Cytotoxicity (ISO 10993-5)Non-CytotoxicNon-Cytotoxic (Same as predicate)
    Irritation (ISO 10993-23)Non-IrritatingNon-Irritating (Same as predicate)
    Sensitization (ISO 10993-10)Non-SensitizingNon-Sensitizing (Same as predicate)
    Pyrogenicity (USP151)<10 EU/sample (Pass)Pass (Same as predicate)
    Systemic Toxicity (ISO 10993-11)Non-ToxicNon-Toxic (Same as predicate)
    Selection of Tests for Interactions with Blood(Not required for intended use)N/A (Not required for intended use)
    Electrical Safety (IEC-60601-2-2)
    HiPot & Continuity Test (Dielectric strength insulation; insulated safety plug)Pass performance for dielectric strength and insulationPass (Same as predicate)
    Compatibility with general electrosurgical generators with 4mm outletsCompatiblePass (Same as predicate)
    Drop TestWithstand specified drop conditions without functional impairmentPass (Same as predicate)
    Sterilization Method & SALSAL = 10^-6 (Sterility Assurance Level)Ethylene Oxide (EO), provided sterile to user, SAL= 10^-6 (Equivalent to predicate's steam sterilization in meeting SAL)
    Shelf-LifeTo demonstrate product integrity for a specified durationReal-time aging study shows product shelf life up to 3 years
    PackagingMaintain sterility and integrityPaper/film (Tyvek/Film) pouch (Equivalent to predicate's packaging in meeting requirements for package integrity)
    Operating ParametersFrequency range: 300 kHz-1,000 kHz; Maximum generator operating voltage: 600VpThe device is intended to be operated within these parameters, and its design is evaluated based on its consistency with these parameters. The performance tests (electrical safety) confirm its suitability for these conditions.
    Tip SizesWithin a range equivalent to the predicate device (0.2mm-2mm)0.5mm-2mm (Range falls within predicate's range)
    LengthsWithin a range equivalent to the predicate device (89mm-273mm)102mm-240mm (Range falls within predicate's range)

    2. Sample Size Used for the Test Set and the Data Provenance:

    • For a medical device like this, the "test set" refers to the physical units of the device that underwent various engineering, electrical, and biological safety tests.
    • The exact sample sizes for each specific test (e.g., how many devices were subjected to the drop test, how many samples for cytotoxicity) are not explicitly stated in this 510(k) summary. These details would typically be found in the full test reports referenced in the submission.
    • Data Provenance: The tests are standard performance and safety tests for medical devices, performed by the manufacturer (HG Innovations Ltd.) or contracted testing facilities. The provenance is internal to the manufacturer's quality system and likely conducted in the UK where the manufacturer is located, or by certified labs. These are prospective tests performed on newly manufactured devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

    • This question is not applicable to this type of device and submission. "Ground truth" in the context of diagnostic AI typically refers to an expert interpretation of medical images or data.
    • For the performance tests described, "ground truth" is established by standard engineering principles, calibrated instruments, and internationally recognized test methods. The "experts" involved are likely engineers, quality assurance personnel, and certified lab technicians who conduct and interpret the results of these physical and chemical tests, rather than clinical experts establishing a diagnostic ground truth.

    4. Adjudication Method for the Test Set:

    • This question is not applicable. Adjudication methods like 2+1 or 3+1 are used to resolve disagreements among multiple human readers when establishing a "ground truth" for diagnostic studies.
    • For engineering and safety tests, the outcome is typically objective (e.g., pass/fail based on a measurement against a defined tolerance).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • This question is not applicable. This device is a surgical instrument, not an AI diagnostic tool. No human reader involvement or AI assistance is relevant to its direct function, nor were MRMC studies performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • This question is not applicable. This device is a physical surgical instrument and does not involve an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For this device, the "ground truth" for its performance and safety is derived from:
      • International Standards: e.g., ISO 10993 series for biocompatibility, IEC 60601-2-2 for electrical safety.
      • Engineering Specifications: The device's design specifications (e.g., tip sizes, lengths, materials) are compared against functional requirements and predicate device characteristics.
      • Regulatory Requirements: Adherence to FDA regulations and guidance.
    • There's no "expert consensus" on diagnostic findings, pathology, or outcomes data in the context of this device's performance demonstration.

    8. The sample size for the training set:

    • This question is not applicable. This device is not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    • This question is not applicable for the same reason as above; there is no training set for a physical medical device.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1