Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K094003

    Validate with FDA (Live)

    Device Name
    REVIVATOR
    Manufacturer
    Date Cleared
    2010-09-17

    (263 days)

    Product Code
    Regulation Number
    868.5915
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Revivator resuscitator is intended for pulmonary resuscitation.
    The adult resuscitator is for people with a body weight of more than 30 kg (66 lbs)
    The child resuscitator is for children with a body weight between 7 and 30 kg (15-66 lbs)
    The infant resuscitator is for infants with a body weight up to 7 kg (15 lbs)

    Device Description

    A resuscitator is a device using positive pressure to inflate the lungs of an unconscious person who is not breathing, in order to keep him oxygenated and alive. The manual resuscitators basically consist of a mask and a large hand-squeezed plastic bulb using ambient air, or supplemental oxygen. The air and oxygen flows into the resuscitator bag. And compressing the bag, the patient is ventilated. In order not to apply a high pressure to the patient's airways, the resuscitators include a pressure-limiting valve.

    AI/ML Overview

    1. Acceptance Criteria and Reported Device Performance

    The device, "Revivator" manual resuscitator, claims substantial equivalence to predicate devices based on compliance with the ISO 10651-4:2002 standard for lung ventilators. The acceptance criteria are derived from this standard. The reported device performance is compared against these criteria and the performance of predicate devices.

    Acceptance Criteria (ISO 10651-4:2002)Revivator Performance (Adult, Child, Infant)Predicate Device Performance (Ambu & Mercury Medical Resuscitators)
    Pressure-limiting valve for <10 kg body weight: Not higher than 4.5 kPa (45 cm H2O)Child/Infant: 4.5 kPa (45 cm H2O)Child/Infant: 4.0 kPa (Ambu); 4.0 ± 0.5 kPa (Mercury Med)
    Pressure-limiting valve for >10 kg body weight: Not higher than 6 kPa (60 cm H2O) (if included, optional)Adult: 6 kPa (60 cm H2O)Adult: 4.0 kPa (Ambu Spur II); Optional for small adults (Mercury Med)
    Max. tidal volume for <10 kg body weight: At least 150 mLInfant: 200 mLMeets standard
    Max. tidal volume for 10-40 kg body weight: At least 450 mLChild: 500 mLMeets standard
    Max. tidal volume for >40 kg body weight: At least 600 mLAdult: 1000 mLMeets standard
    Delivered oxygen concentration (with O2 source) : At least 85% (V/V) (with attachment)Without reservoir: 39-88% O2 (at 2-15 L/min O2 flow); With reservoir: 98-100% O2 (at 2-15 L/min O2 flow)Claimed to meet standard (reservoir volume contributes)
    Apparatus dead space: Shall not exceed 5 mL + 10 % of the minimum delivered volumeLower than 7 mLBetween 5 and 7 mL
    Expiratory resistance: Shall not exceed 5 cm H2O0.09 cm H2O (at 5 L/min); 1.76 cm H2O (at 50 L/min)Values much lower than admissible
    Inspiratory resistance: Shall not exceed 5 cm H2O below atmospheric pressure0.37 cm H2O (at 5 L/min); 3.48 cm H2O (at 50 L/min)Values much lower than admissible (Ambu Spur II adult/pediatric 5 cm H2O at 50 L/min)
    Patient connectors: 22/15 mm22mm male / 15mm femaleIn conformity (Ambu)
    Expiratory connectors: 30 mm male30mm male (PEEP/AGSS)In conformity (Ambu)
    Bag refill valve connectors: 32 mm female conical32 mm femaleIn conformity (Predicate devices)
    Storage temperatures: -40°C to +60°C (with 40%-95% r.h.)-40°C to +70°CAll resuscitators meet; Revivator & Ambu reusable can store at 70°C
    Operating temperatures: -18°C to +50°C (with 40%-95% r.h.)-20°C to +60°CAll resuscitators meet; Revivator & Ambu reusable can operate at -20°C
    Dismantling & reassembly: Functional test after reassemblyIncludes functional test in labelingIncludes functional test in labeling
    Patient Valve function after contamination with vomitus: In conformity with standardIn conformity with standardIn conformity with standard
    Drop test: In conformity with standardIn conformity with standardIn conformity with standard
    Immersion in water: In conformity with standardIn conformity with standardIn conformity with standard
    End-expiratory pressure: If > 0.2 kPa, value in instructions; pressure gradient < 0.6 kPaNot an increase in pressure gradient at end of expiration at 30 L/minExpected to be lower than 0.2 kPa

    2. Sample Size and Data Provenance

    The provided document describes a device comparison study for market clearance (510(k) submission) against predicate devices and an international standard (ISO 10651-4:2002). This type of submission generally relies on bench testing and engineering analysis rather than clinical trials with human subjects.

    • Test Set Sample Size: The document does not explicitly state a "sample size" in terms of number of devices for specific tests. Instead, it refers to the device models (Adult, Child, Infant for both REVIVATOR PLUS and REVIVATOR RES-Q) and the inherent design specifications of these models. The testing implicitly covers these models to ensure compliance with the stated ISO standard.
    • Data Provenance: The data provenance is from bench testing conducted to evaluate compliance with ISO 10651-4:2002. The company, HERSILL, S.L., is located in Móstoles, Madrid, Spain, suggesting the testing was performed in Spain or by a contracted laboratory. The nature of the comparison is retrospective in the sense that the device's characteristics are compared against established standards and existing predicate devices' specifications, rather than a prospective clinical study.

    3. Number of Experts and Qualifications

    The document does not mention the use of experts to establish a "ground truth" for a test set in the way one might for an AI diagnostic device. Instead, the "ground truth" is the ISO 10651-4:2002 standard itself and the specifications of the legally marketed predicate devices. The individuals involved in the testing would be qualified engineers and technicians proficient in medical device testing and quality assurance, likely with expertise in respiratory devices and relevant ISO standards.

    4. Adjudication Method

    Not applicable. This is a technical performance comparison against a standard and predicate devices, not a clinical study requiring expert adjudication of cases.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. This document describes a 510(k) submission for a manual resuscitator, which involves demonstrating substantial equivalence through technical performance comparisons, not comparative effectiveness studies with human readers or AI assistance.

    6. Standalone Performance Study (Algorithm Only)

    The device is a manual resuscitator, a physical medical device. It does not involve an algorithm or software requiring a standalone "algorithm only" performance study in the context of AI. The performance metrics reported are for the physical device itself.

    7. Type of Ground Truth Used

    The ground truth used for this study is:

    • International Standard Compliance: Primarily, the requirements and specifications outlined in ISO 10651-4:2002: Lung ventilators - Part 4: Particular requirements for operator-powered resuscitators.
    • Predicate Device Specifications: The technical characteristics and performance data of previously cleared manual resuscitators (Ambu and Mercury Medical resuscitators).

    8. Sample Size for the Training Set

    Not applicable. This is not a machine learning or AI-based device, so there is no concept of a "training set" in the context of device performance evaluation described here.

    9. How Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set mentioned or implied for this mechanical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K093241

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-04-27

    (194 days)

    Product Code
    Regulation Number
    878.4780
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1