510(k) Data Aggregation

    K Number
    K141869

    Validate with FDA (Live)

    Date Cleared
    2014-10-10

    (91 days)

    Product Code
    Regulation Number
    882.1620
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hemedex Cranial Bolt is designed to achieve cranial access and to introduce and secure one to four sensors in place for intracranial monitoring.

    Device Description

    The QFlow 500™ Titanium Bolt Quad Lumen Kit is used to achieve cranial access and to introduce and secure up to four sensors in place for intracranial monitoring. The kit contains a cranial bolt, drill bit, scalpel, sensor introducer with stylet, Touhy Borst fittings and connectors. The cranial bolt contains four lumens with its primary materials being titanium, polyphenylsulfone, polyvinyl chloride and polycarbonate. The primary materials of the sensor introducer are polyether block amide and polycarbonate. The primary materials of the other components in the kit are stainless steel, titanium nitride, acrylonitrile butadiene styrene, polyphenylsulfone, polyvinyl chloride and polycarbonate.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    This device is not an AI/ML powered device, The current response contains information derived from the provided document.

    TestAcceptance Criteria (Predetermined Specification)Reported Device Performance (Hemedex QFlow 500™ Titanium Bolt Kit, Quad Lumen)
    Torque required to insert the cranial boltNot explicitly stated (implied to be easy)7 - 8 in-lbs maximum torque required. In comparative testing, required 38% to 64% less in-lbs of torque to fully insert compared to predicate/reference devices.
    Axial tension required to pull the cranial bolt outGreater than 40 lbfGreater than 40 lbf. In comparative testing, the Hemedex Quad Lumen Bolt pull out force was greater than 40 lbf, whereas predicate and reference devices could not reach this threshold without incurring damage.
    Torque required to tighten the Touhy Borst compression cap to seal the sensorsLess than 1 in-lb torque requiredLess than 1 in-lb torque required.
    Tension required to pull the sensors out of the Touhy BorstNot explicitly stated (implied to be secure)Approximately 1.3 lbf.
    Leak testingNo leaking at 185 mm Hg pressure for 14 daysNo leaking at 185 mm Hg pressure for 14 days. In comparative testing, the Hemedex Quad Lumen Bolts incurred less leaking at increasing intervals of insertion replicating different skull thicknesses at 185 mm Hg pressure.
    Axial tensile strength of the "pig tails"Not explicitly stated (implied to be strong)19 - 37 lbf depending on pigtail diameter.
    Axial tensile strength of the introducerNot explicitly stated (implied to be acceptable)3.7 - 4.4 lbf.
    Axial tensile strength of the pigtail extensionNot explicitly stated (implied to be strong)16.6 - 20.6 lbf.
    Drill bit dimensional and cutting functionalityBurr hole diameter 0.209"Burr hole diameter 0.209". Easily cut through bone. Depth collars remained secure. In comparative testing, the Hemedex drill bit meets specification. Predicate (K002765) and reference device specifications are unknown for diameter. All tested drill bits easily cut through bone; depth collars remained secure.

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes bench testing, which typically involves laboratory-controlled experiments rather than human or animal subjects. Therefore, the concept of "sample size used for the test set" in the context of human data or "data provenance" (country of origin, retrospective/prospective) as it relates to clinical data is not applicable here. The tests were performed on the device components themselves.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. The ground truth for this type of performance testing (mechanical and functional bench tests) is based on engineering specifications and physical measurements, not expert human interpretation.

    4. Adjudication Method for the Test Set

    Not applicable. This is not a study involving human adjudication of results; it's a series of physical and mechanical bench tests performed according to engineering protocols.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This device is a mechanical cranial bolt, not an AI/ML system that would involve human readers interpreting output.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This device is a physical medical device, not an algorithm, so "standalone performance" in the context of AI is not relevant. The performance tests described are standalone in the sense that they evaluate the device's physical properties and function without human interaction beyond operating the test equipment.

    7. The Type of Ground Truth Used

    The ground truth for the bench testing was based on:

    • Predetermined Engineering Specifications: These are quantifiable targets for mechanical properties (e.g., torque, tension, leak rates, dimensions).
    • Measurable Physical Properties: Direct measurements of the device's behavior under specific test conditions (e.g., actual torque required, actual pull-out force, actual leakage).
    • Comparison to Predicate Devices: Performance was also evaluated relative to the performance of existing legally marketed predicate devices (K032337, K002765) and a reference device (K992591) to demonstrate substantial equivalence or improvement.

    8. The Sample Size for the Training Set

    Not applicable. This device is a physical medical device, not an AI/ML algorithm that requires a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set, there is no ground truth establishment for it.

    Ask a Question

    Ask a specific question about this device

    K Number
    K032127

    Validate with FDA (Live)

    Date Cleared
    2003-12-30

    (173 days)

    Product Code
    Regulation Number
    876.5130
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hemedex Fixation Device is intended to be used on the patient's skin to secure the percutaneous placement of the QFlow Perfusion Probe.

    Device Description

    The Hemedex Fixation Device is a compact soft silicone disk with a securing clamp and adhesive dressing for securing a probe or catheter to the patients skin. The clear pressure sensitive adhesive dressing allows for easy visual inspection of catheter entry site.

    AI/ML Overview

    I am sorry, but based on the provided text, there is no detailed information available regarding acceptance criteria, device performance metrics, or the actual study results that would allow me to populate the requested table and answer questions 2 through 9.

    The document is a 510(k) premarket notification letter from the FDA to Hemedex, Inc. for their Fixation Device. It primarily states that the device has been found substantially equivalent to a predicate device (UreSil Fixation Device K914699).

    Here's what the document does state in relation to performance testing:

    • Section 7. Performance Testing: "Bench testing was conducted to demonstrate that this device meets the requirements of its intended use and meets the specified performance criteria."

    However, this statement provides no specifics about:

    • What those "specified performance criteria" were.
    • The actual results of the "bench testing."
    • The methodology, sample sizes, or ground truth establishment for any such testing.

    Therefore, I cannot provide the requested information. The document only confirms that some bench testing was done to demonstrate the device meets its intended use and performance criteria, but it does not elaborate on these aspects in any way that fulfills the detailed questions asked.

    Ask a Question

    Ask a specific question about this device

    K Number
    K032337

    Validate with FDA (Live)

    Date Cleared
    2003-10-21

    (84 days)

    Product Code
    Regulation Number
    882.1620
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hemedex Single and Double Lumen Bolts are designed to achieve cranial access and to introduce and secure a sensor in place for intracranial monitoring.

    Device Description

    The Hemedex Cranial Bolts are stainless steel threaded cranial access devices with standard tuohy borst compression fittings.

    AI/ML Overview

    The Hemedex Cranial Bolts are designed to achieve cranial access and to introduce and secure a sensor in place for intracranial monitoring. Bench testing was conducted to demonstrate that this device meets the requirements of its intended use and meets the specified performance criteria. The Hemedex Cranial Bolts and the predicate devices use identical technology and materials.

    Acceptance Criteria and Device Performance:

    Acceptance CriteriaReported Device Performance
    Meet requirements of intended useBench testing demonstrated the device meets these requirements
    Meet specified performance criteriaBench testing demonstrated the device meets these criteria
    Identical technology to predicate devicesConfirmed as identical
    Identical materials to predicate devicesConfirmed as identical

    Study Information:

    1. Sample size used for the test set and the data provenance: Not specified.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as this was bench testing.
    3. Adjudication method: Not applicable.
    4. Multi-reader multi-case (MRMC) comparative effectiveness study: No.
    5. Standalone performance study: Yes, bench testing was conducted.
    6. Type of ground truth used: Bench testing results against specified performance criteria and intended use requirements.
    7. Sample size for the training set: Not applicable, as this was bench testing for a physical device.
    8. How the ground truth for the training set was established: Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K013376

    Validate with FDA (Live)

    Date Cleared
    2002-05-08

    (209 days)

    Product Code
    Regulation Number
    870.2100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The QFlow™ 500 is intended for extravascular monitoring of microcirculation blood flow in buried tissues. Example of this application included (but are not limited to) 1) the monitoring of buried muscle or esophagus following free muscle transfer or esophageal reconstruction, 2) monitoring soft tissue microcirculation following reconstructive surgery, such as oral and facial reconstruction, 3) monitoring cerebral blood flow during and following neurosurgery for head trauma.

    Device Description

    The QFlow™ 500 has application where real-time determination of any tissue perfusion is desired, whether it is for measurement or monitoring utilization. The system may be used in the Intensive Care Unit, Operating Room, Radiology Suite or in a standard patient care room. However, the system is not designed for use during patient transport.

    The QFlow™ 500 consists of a thermal diffusion probe and a monitor.

    The QFlow™ 500 Perfusion System Probe thermally interacts with tissue. It contains two thermistors embedded at the distal tip of the probe. It is available in a range of lengths from 60 cm to 120 cm and less than 1 mm in diameter. The probe contains an EEPROM for the storage of probespecific calibration data, data on the cumulative time the probe has been in use, and a verification key to insure the origin of the probe. The probe connector physically connects the probe to the umbilical cord which connects to the monitor.

    The QFlow™ 500 Monitor controls the thermistors in the Probe.

    Perfusion data are displayed on the monitor. The current perfusion value, temperature and thermal conductivity are be displayed in numeric form. The data are also displayed in the form of a graph as a function of time.

    The monitor is capable of storing data for up to 10 days. All of this stored data may be reviewed by the user at any time during this period. Printing of data is also an option from the monitor. There is also the capability of uploading data from the monitor to an external computer for further analysis.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Hemedex, Inc. QFlow$^{\text{TM}}$ 500 Perfusion Monitoring System. It outlines the device, its intended use, comparison to predicate devices, and the performance testing performed to demonstrate substantial equivalence.

    However, the document focuses on compliance with electrical safety, thermal safety, electromagnetic compatibility, electromagnetic emissions, and probe integrity standards. It does not present acceptance criteria or reported device performance in terms of microcirculation blood flow measurement accuracy or clinical efficacy. The "Performance Testing" section lists various standards the device was tested for compliance with, but it does not detail the specific performance metrics or acceptance thresholds related to its core function of monitoring microcirculation blood flow.

    Therefore, many of the requested items cannot be extracted from the provided text.

    Here is an attempt to address the request based on the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Based on the provided text, the acceptance criteria and reported device performance are related to compliance with general safety and electrical standards, not the direct performance of blood flow monitoring.

    Acceptance Criterion (Type)Reported Device Performance (Reference)
    Electrical SafetyCompliance with IEC 60601-1 (1988) 2nd Edition, + A1: 1991 + A2: 1995. UL2601, UL standard for Safety, Medical Electrical Equipment, Part 1: General Requirements for Safety, First Edition.
    Thermal SafetyCompliance with IEC 60601-1 (1988) 2nd Edition, + A1: 1991 + A2: 1995.
    Electromagnetic Compatibility (EMC)Compliance with IEC60601-1-2: 1993.
    Electromagnetic EmissionsCompliance with EN 55011.
    Probe IntegrityCompliance with ISO 10555 Sterile, single-use intravascular catheters Part 1: General Requirements.
    BiocompatibilityCompliance with ISO 10993: 1997 Part 1: Guidance on selection of tests.

    Regarding the study that proves the device meets the acceptance criteria, the document states:

    "Bench testing, animal studies and biocompatibility testing was performed on the QFlow 500."
    "The QFlow™ 500 is tested for compliance with the following standards for electrical safety, thermal safety, electromagnetic compatibility, electromaqnetic emissions and probe integrity."

    This indicates that these tests were conducted, and the device was found to comply with the listed standards. No further details about the specific studies, their methodology, or results are provided in this summary.


    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    This information is not provided in the text. The document mentions "bench testing," "animal studies," and "biocompatibility testing" but does not detail the sample sizes, study design (retrospective/prospective), or data provenance for these tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This information is not provided in the text. The testing mentioned (compliance with standards) does not typically involve expert review for ground truth in the same way clinical diagnostic studies do.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This information is not provided in the text. Adjudication methods are typically relevant for clinical studies involving human interpretation or subjective assessments, which are not detailed here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study is not mentioned in the text. The device is a monitoring system, not an AI-assisted diagnostic tool that requires human reader improvement evaluation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The QFlow™ 500 is a monitoring system that directly measures and displays perfusion data. Its "standalone" performance would refer to its accuracy and reliability in measuring the intended physiological parameter. While "bench testing" and "animal studies" were performed, the text does not explicitly describe a standalone performance study with specific metrics for blood flow measurement accuracy (e.g., compared to a gold standard measurement technique). It only states compliance with safety and integrity standards.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the compliance tests mentioned (electrical safety, thermal safety, etc.), the "ground truth" would be the specifications and requirements defined by the international and national standards themselves. For the stated purpose of "extravascular monitoring of microcirculation blood flow," the document does not specify the ground truth used to validate the device's accuracy in measuring blood flow.

    8. The sample size for the training set

    This information is not applicable/not provided. The QFlow™ 500 is a physical measurement device based on thermal diffusion principles, not a machine learning model that requires a "training set" in the context of AI.

    9. How the ground truth for the training set was established

    This information is not applicable/not provided for the reasons stated in point 8.