510(k) Data Aggregation

    K Number
    K131550

    Validate with FDA (Live)

    Device Name
    HEART SYNC
    Manufacturer
    Date Cleared
    2013-12-06

    (191 days)

    Product Code
    Regulation Number
    870.5310
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Heart Sync Sterile Adult Radio Transparent Multi Function Electrodes are indicated for use in external pacing, defibrillation and monitoring applications as a sterile, disposable device for single patient use only. The electrodes provide the conductive interface between the defibrillator and/or the external transcutaneous (noninvasive) cardiac pacemaker and the patient's skin. The electrode is intended for use on adult patients. When a patient requires defibrillation, cardioversion or external pacing, these electrodes will be applied to the patient and connected to the instrument. This device is intended for use on defibrillators whose output is classified as low power (360 joule maximum). Compatible with Physio Control, Zoll, and "Anderson" adapted models of monophasic and bi-phasic defibrillators.

    Device Description

    These are single use, sterile, self stick defibrillator electrodes packaged in pairs. Effective electrode area = Conductive Area: 195.98 Sq.cm. They are radiotransparent. They come in three connector styles to match the specific defibrillator. The construction and materials employed are identical to the predicate. The patient contact material is a conductive adhesive hydrogel identical to the material used in the predicate devices. These pads meet the AAMI Standard ANSI/AAMI DF80:2003 and the connectors meet the FDA performance standard for touch proof ECG connectors.

    AI/ML Overview

    The provided 510(k) summary for K131550 describes Sterile Defibrillator Pads. This submission focuses on demonstrating substantial equivalence to a predicate device (K080421, Heart Sync C-100/T-100) for a sterile version of an already marketed product. Therefore, the "study" described is primarily a non-clinical bench testing and biocompatibility study to confirm that the new sterile version performs equivalently to the non-sterile predicate and meets relevant standards.

    Here's an analysis of the provided information, addressing your specific questions:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Standard / Test)Reported Device Performance
    AAMI Standard ANSI/AAMI DF80:2003"All units passed these testing routine was applied to a 30 month accelerated age shelf life test. All units passed the tests."
    - DC offset instabilityPassed (implied by above)
    - AC small signal impedancePassed (implied by above)
    - AC large signal impedancePassed (implied by above)
    - Defibrillator overloadPassed (implied by above)
    FDA Performance Standard for Touch-Proof ECG Connectors"Compliance with the FDA performance standard was verified by inspection of the connectors."
    Biocompatibility Testing (on patient contact material)"Biocompatibility testing was performed on the patient contact material passed biocompatibility testing."
    Sterilization Validation (ANSI/AAMI/ISO 11137-2)"Testing was conducted post-sterilization was validated per ANSI/AAMI/ISO ISO 11137-2 ANSI/AAMI/ISO Sterilization of Heath Care Products - Radiation - Establishing the sterilization dose"
    Performance Testing on Attenuator System (compared to predicate)"Performance testing on the attenuator system for both types of pads was performed and testing compared performance directly with the attenuator predicates showing excellent correlation of results."
    Accelerated Age Shelf Life (30 months)"All units passed these testing routine was applied to a 30 month accelerated age shelf life test. All units passed the tests."

    2. Sample Size Used for the Test Set and Data Provenance

    The document states, "Various different lots containing multiple samples each were subjected to the AAMI tests..." It does not specify an exact numerical sample size for the test set.

    • Data Provenance: The testing appears to be prospective bench testing conducted by Heart Sync, Inc. The country of origin for the data is not explicitly stated but can be inferred as the United States, given the company's location and FDA submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Not applicable (N/A). This submission pertains to a medical device (defibrillator pads), not an AI or diagnostic imaging device that requires expert ground truth establishment for a test set. The "ground truth" here is defined by objective engineering standards (AAMI, FDA, ISO) and laboratory measurements, not human interpretation.

    4. Adjudication Method for the Test Set

    Not applicable (N/A). Since the testing involves objective measurements against engineering standards, there is no need for adjudication by multiple human experts. The pass/fail criteria are defined by the standards themselves.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, a MRMC comparative effectiveness study was not done. This type of study is typically conducted for AI-powered diagnostic tools to assess the impact of AI on human reader performance. Defibrillator pads are a therapeutic device, not a diagnostic one that involves human image or data interpretation.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Not applicable (N/A). This device does not involve an algorithm. Its performance is assessed through physical and electrical bench testing, and its function is to provide a conductive interface for defibrillation and pacing, not to perform an algorithmic function.

    7. The Type of Ground Truth Used

    The "ground truth" for this device is based on objective engineering and biocompatibility standards.

    • AAMI Standard ANSI/AAMI DF80:2003: Performance criteria for defibrillator electrodes.
    • FDA Performance Standard for Touch-Proof ECG Connectors: Safety and performance criteria for connectors.
    • ANSI/AAMI/ISO 11137-2: Sterilization validation standard.
    • Biocompatibility testing against established biological safety criteria.
    • Direct comparison to the predicate device for attenuator system performance.

    8. The Sample Size for the Training Set

    Not applicable (N/A). This device is not an AI or machine learning algorithm, so there is no concept of a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable (N/A). As there is no training set for an algorithm, there is no ground truth to establish for it.

    Ask a Question

    Ask a specific question about this device

    K Number
    K131494

    Validate with FDA (Live)

    Device Name
    HEART SYNC
    Manufacturer
    Date Cleared
    2013-09-16

    (116 days)

    Product Code
    Regulation Number
    870.5310
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indicated for use to treat patients in cardiopulmonary arrest who are unconscious, without a pulse and not breathing spontaneously. Heart Sync T-Cardiac Science AED pads are single use and intended to be used in conjunction with Cardiac Science automated external defibrillators (AED) to monitor and deliver defibrillation energy to the patient. The T-Cardiac Science AED pads are used on patients over 8 years of age or greater than 55 pounds. The electrodes are intended for short term use (less than 8 hours)

    Device Description

    These are single use. non-sterile, self-stick defibrillator electrodes packaged in pairs. Effective electrode area = 170 Sq. cm. They are Radiotranslucent. They come in an Anderson connector to match the specific defibrillator. The construction and materials employed are similar to the predicates. The patient contact material is a conductive adhesive hydrogel identical to the material used in the predicate device. These pads meet the AAMI Standard ANSI/AAMI DF80:2003 and the connectors meet the FDA performance standard for touch proof ECG connectors.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the "T-CardiacSci-AED™ Pads." This submission focuses on demonstrating substantial equivalence to a predicate device through non-clinical bench testing, rather than clinical studies involving human patients or complex AI algorithms. Therefore, many of the requested elements for a study proving device meets acceptance criteria (e.g., sample size for test set, expert ground truth, MRMC study, training set details) are not applicable to this specific submission.

    The "acceptance criteria" here relate to engineering and performance standards for defibrillator pads, as well as demonstrating equivalence in design and materials to a legally marketed predicate device.

    Here's an analysis based on the provided text, addressing the points where applicable:

    Acceptance Criteria and Reported Device Performance

    Given that this is a 510(k) for defibrillator pads, the acceptance criteria are primarily related to meeting established performance standards for such devices and demonstrating material and functional equivalence to a predicate. The "study" proving this is the non-clinical bench testing.

    Acceptance Criteria (from text)Reported Device Performance (from text)
    Compliance with AAMI Standard ANSI/AAMI DF80:2003"These pads meet the AAMI Standard ANSI/AAMI DF80:2003"
    Connectors meet FDA performance standard for touch-proof ECG connectors"the connectors meet the FDA performance standard for touch proof ECG connectors." "Compliance with the FDA performance standard was verified by inspection of the connectors. (Connectors must be "touch-proof.")"
    DC offset instability (AAMI tests)"Various different lots containing multiple samples each were subjected to the AAMI tests for DC offset instability. All units passed these testing routine was applied to a 42 month accelerated age shelf life test. All units passed the tests."
    AC small signal impedance (AAMI tests)"Various different lots containing multiple samples each were subjected to the AAMI tests for ... AC small signal impedance. All units passed these testing routine was applied to a 42 month accelerated age shelf life test. All units passed the tests."
    AC large signal impedance (AAMI tests)"Various different lots containing multiple samples each were subjected to the AAMI tests for ... AC large signal impedance. All units passed these testing routine was applied to a 42 month accelerated age shelf life test. All units passed the tests."
    Defibrillator overload (AAMI tests)"Various different lots containing multiple samples each were subjected to the AAMI tests for ... defibrillator overload. All units passed these testing routine was applied to a 42 month accelerated age shelf life test. All units passed the tests."
    Biocompatibility of patient contact material"Biocompatibility testing was performed on the patient contact material passed biocompatibility testing."
    Performance of attenuator system (comparison to predicate)"Performance testing on the attenuator system for both types of pads was performed and testing compared performance directly with the attenuator predicates showing excellent of results." (Implies demonstration of equivalent or better performance compared to predicate's attenuator system, which would be an implicit acceptance criterion for substantial equivalence).
    Shelf Life (accelerated aging)"All units passed these testing routine [AAMI tests] was applied to a 42 month accelerated age shelf life test. All units passed the tests."
    Identical technological characteristics to predicate"This device has the same technological characteristics (i.e., design, material, chemical composition, energy source) as the predicate device." "Furthermore, the materials and construction methods are identical to the predicate." (This is a design characteristic claimed to be met, rather than a quantifiable performance metric).

    Study Details (Bench Testing)

    1. Sample size used for the test set and the data provenance:

      • Sample Size: "Various different lots containing multiple samples each" were subjected to AAMI tests and accelerated shelf-life testing. Specific numbers are not provided in the text.
      • Data Provenance: The testing was conducted internally by the manufacturer (Heart Sync, Inc.) or a third-party lab contracted by them. The country of origin for the data is implicitly the USA, as Heart Sync is located in Ann Arbor, MI. The data is prospective in the sense that it was generated specifically for this submission.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This is Not Applicable (N/A). The "ground truth" for this device is based on objective engineering and material standards (AAMI, FDA performance standards) and direct comparison to a predicate device's characteristics. There's no subjective expert interpretation of, for example, medical images.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • N/A. Adjudication methods are typically used in clinical studies involving subjective assessments (e.g., image interpretation). For bench testing against objective standards, the outcome is pass/fail based on measured parameters.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • N/A. This device is a defibrillator pad, not an AI-assisted diagnostic tool. No MRMC study was performed or required. The text explicitly states: "Clinical tests are not required to establish substantial equivalence."
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • N/A. This refers to AI algorithm performance. This device is not an AI algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for device performance is defined by recognized industry standards (AAMI), regulatory performance standards (FDA for connectors), and direct comparison/equivalence to the established characteristics of the predicate device. For biocompatibility, the ground truth is established by recognized biocompatibility testing protocols (e.g., ISO 10993).
    7. The sample size for the training set:

      • N/A. This concept applies to machine learning models, but this submission is for a physical medical device. There is no "training set" in the AI sense.
    8. How the ground truth for the training set was established:

      • N/A. As above, there is no training set. The "ground truth" for the device's design and manufacturing processes is established through engineering design specifications, material selections, and manufacturing quality control, all benchmarked against the predicate device and relevant standards.
    Ask a Question

    Ask a specific question about this device

    K Number
    K120536

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-08-06

    (166 days)

    Product Code
    Regulation Number
    870.5310
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Indicated for use to treat patients in cardiopulmonary arrest who are unconscious, without a pulse and not breathing spontaneously. They should only be used by personnel who have been trained in its operation. Used to deliver lower-energy therapy to children from 1 year of age to 8 years or up to 55 Ibs (25kg). DO NOT DELAY THERAPY IF YOU ARE NOT SURE OF EXACT AGE OR WEIGHT.

    Device Description

    These are single use, non-sterile, self stick defibrillator electrodes packaged in pairs. Effective electrode area = Conductive Area: 92 Sq. cm. They are radiotranslucent. They come in two connector styles to match the specific defibrillator. The construction and materials employed are similar to the predicates. The patient contact material is a conductive adhesive hydrogel identical to the material used in the predicate devices. These pads meet the AAMI Standard ANSI/AAMI DF80:2003 and the connectors meet the FDA performance standard for touch proof ECG connectors.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Heart Sync Pediatric AED Pad, formatted to describe acceptance criteria and study details.

    Please note: The provided text is a 510(k) Summary, which typically focuses on demonstrating substantial equivalence to predicate devices through non-clinical (bench) testing. It does not contain information about clinical trials, human reader studies, or AI performance, as the device is a medical accessory (defibrillator pads) and not an AI algorithm. Therefore, many sections of your requested output will be marked as "Not Applicable" (N/A) or "Not Provided in Document."


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    AAMI Standard ANSI/AAMI DF80:2003 ComplianceAll units passed tests for DC offset, offset instability, AC small signal impedance, AC large signal impedance, and defibrillator overload.
    FDA Performance Standard for Touch-Proof ECG ConnectorsCompliance verified by inspection of the connectors.
    Biocompatibility Testing (patient contact material Hydrogel)The material passed biocompatibility testing.
    Accelerated Age Shelf Life Test (42 months)All units passed relevant AAMI tests after accelerated aging.
    Attenuator System Performance (compared to predicates)Testing showed excellent correlation of results with attenuator predicates.

    2. Sample Size Used for the Test Set and the Data Provenance

    • Sample Size: "Various different lots containing multiple samples each" were subjected to AAMI tests. No specific numerical sample size is provided.
    • Data Provenance: Not explicitly stated, but assumed to be from manufacturing and laboratory testing conducted by Heart Sync, Inc. or its contracted labs. This is retrospective in nature (testing manufactured units). No country of origin of data is specified beyond the submitter's location (Ann Arbor, MI, USA).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • Not Applicable. The "ground truth" for this device's performance is based on established engineering and medical device standards (AAMI, FDA performance standards) and direct measurements from bench tests, not expert interpretation of cases or data.

    4. Adjudication Method for the Test Set

    • Not Applicable. As the testing is bench-top and objective against defined standards, there is no need for expert adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This device is a passive accessory (defibrillator pad) and does not involve AI or human interpretation for its function. Therefore, no MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This device is a medical accessory, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • The "ground truth" for this device's performance is based on defined technical specifications and validated test methods from standards bodies (AAMI) and regulatory agencies (FDA). For instance, the "ground truth" for impedance is a specific numerical range defined by the standard.

    8. The sample size for the training set

    • Not Applicable. This device does not use machine learning or AI, and therefore does not have a "training set."

    9. How the ground truth for the training set was established

    • Not Applicable. See point 8.