510(k) Data Aggregation

    K Number
    K102692

    Validate with FDA (Live)

    Date Cleared
    2011-03-11

    (172 days)

    Product Code
    Regulation Number
    878.4040
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Jincheng Surgical Gown is a single use, sterile or none sterile (sterile before use) item that is intended to be used in operation room as a protective covering, for operating room staff, from the transferring of body fluids and particulates.

    Gowns provided as sterile and non-sterile.

    Non-sterile surgical gowns are to be sold to OEMs for EtO sterilization according to ISO 11135. Sterile Surgical Gowns are to be sold directly to users after EtO sterilization validation to ISO 11135.

    Device Description

    The Jincheng Surgical Gowns are an open back gown manufactured from nonwoven fabric that the fibres are mechanically bonded together. They are full length, constructed with raglan sleeves, hook and loop neck closures, and tie waist closures. They come with three different types of materials (SMS polypropylene fabric, Spunlace fabric, and SFT fabric) and four different design styles (regular, reinforced, regular with guider, and reinforced with guider). Therefore there are total eight different models involved in this submission. They are: Jincheng SMS Gowns (SMS Reinforced Gown, SMS Reinforced Gown with Guider), Jincheng Spunlace Gowns (Spunlace Regular Gown, Spunlace Regular Gown with Guider, Spunlace Reinforced Gown, Spunlace Reinforced Gown with Guider), Jincheng SFT Gowns (SFT Gown, SFT Gown with Guider). Each design has four different sizes: Medium (M), Large (L), Extra Large (XL), and Extra Extra Large (XXL). All gowns come with sterile and non-sterile surgical gowns are to be sold to OEMs for EtO sterilization according to ISO 11135. Sterile Surgical Gowns are to be sold directly to users after EtO sterilization validation to ISO 11135.

    AI/ML Overview

    The provided text describes the acceptance criteria and study for a Surgical Gown, not a medical device in the typical sense of AI/software or diagnostic tools. Therefore, several of the requested categories (e.g., sample size for test set, data provenance, number of experts, adjudication method, MRMC study, standalone performance, training set details) are not applicable to this submission.

    Here's the information extracted from the provided 510(k) summary for the "Jincheng Surgical Gowns":

    1. Table of Acceptance Criteria and Reported Device Performance

    DescriptionAcceptance Criteria (Standard)Reported Device Performance
    CytotoxicityISO10993-5 (No Toxic Effect)No Toxic Effect
    Skin Irritation and SensitizationISO 10993-10 (No Effect)No Effect
    Hydrostatic Pressure: Water ResistanceAATCC Test Method 127: 1998 (Hydrostatic Pressure Test)Met requirements per AATCC Test Method 127: 1998
    Impact Penetration Test : Water ResistanceAATCC 42: 2007 (Water Resistance: Impact Penetration Test)Met requirements per AATCC 42: 2007
    Breaking StrengthASTM D5034: 2008 (Standard Test Method for Breaking Strength and Elongation)Met requirements per ASTM D5034
    Elmendorf TearASTM - D5734-95:2001 (Standard Test Method for Tearing Strength of Nonwoven Fabrics)In-house method similar to ASTM - D5734-95:2001, met requirements
    Flammability16 CFR 1610 (Flammability Test Method - CPSC CS-191-53)Met requirements per 16 CFR 1610
    LintISO 9073-10 (Lint and Other Particles Generation in the Dry State)Met requirements per ISO 9073-10
    Sterilization (for non-sterile gowns sold to OEMs)ISO 11135 (EtO sterilization)For OEMs to perform; validation to ISO 11135
    Sterilization (for sterile gowns sold directly)ISO 11135 (EtO sterilization validation)Validation to ISO 11135
    Overall Safety and EffectivenessMeets requirements of predefined acceptance criteria and intended uses"Safe and effective, and its performance meets the requirements of its predefined acceptance criteria and intended uses."

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not explicitly stated for each test. The submission implies that sufficient samples were tested to meet the standards.
    • Data Provenance: The tests were conducted by Hangzhou Jincheng Medical Supplier Manufacture Co., Ltd in China. The data would be considered retrospective as it was collected before the 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The "ground truth" for surgical gowns is established through adherence to recognized international and national consensus standards for materials and performance, not expert human interpretation.

    4. Adjudication method for the test set

    Not applicable. The performance is objectively measured against established standards, not through human adjudication of results.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a physical product (surgical gown), not an AI/software device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical product (surgical gown), not an algorithm or AI device.

    7. The type of ground truth used

    The ground truth used is based on consensus standards and regulations (e.g., ISO, AATCC, ASTM, CFR) which define the acceptable physical, mechanical, and biological properties for surgical gown materials and performance.

    8. The sample size for the training set

    Not applicable. This is a physical product (surgical gown), not an AI/machine learning model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is a physical product (surgical gown), not an AI/machine learning model.