Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K132744

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-05-27

    (266 days)

    Product Code
    Regulation Number
    884.4160
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Acessa Guidance System is indicated for enhancing the ultrasonic image of the Acessa Handpiece and for predicting its future path on a computer monitor screen which also shows the image of a B-scan (or similar display) of a medical ultrasound imaging system. The device is intended as an optional accessory for use during the Acessa System procedure.

    Device Description

    The Acessa Guidance System consists of the following components: Guidance Controller (Model Number 5100), Guidance Ultrasound Transducer Sleeve (Model Number 5500), Guidance Field Generator (Model Number 5200), Guidance Handpiece Cable (Model Number 5400), and Power Cord (Model Number 4110). The Acessa Guidance System uses electromagnetic tracking technology to track the positions of the Guidance Ultrasound Transducer Sleeve and the Guidance Handpiece shaft and draws virtual representations of them in their spatial relationship, so that a physician can predict the Guidance Handpiece shaft's future path in relation to the features in the ultrasound slice. The Acessa Guidance System is considered an optional accessory to procedures where ultrasound is currently used for visualization, such as the Acessa System procedure.

    AI/ML Overview

    The Acessa Guidance System includes the following performance testing:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test CategoryAcceptance CriteriaReported Device Performance
    Electrical and SafetyCompliance with ANSI/AAMI ES 60601-1:2005 with A2:2010 (3rd edition), EN/IEC 60601-1-2:2007 (3rd edition), and IEC 60601-1-4:2000 (1st edition)Device passed all testing.
    BiocompatibilityCompliance with ISO 10993-1:2009 (4th edition), ISO 10993-5:2009 (3rd edition), ISO 10993-10:2010 (2nd edition), and ISO 10993-11:2006 (2nd edition)Device demonstrated safety, suitability, and appropriateness for intended use based on additional materials.
    Mechanical (Guidance Sleeve end cap tensile testing)Not explicitly stated, but implies meeting a predefined threshold for tensile strength.Device met all criteria.
    Accuracy (System worse-case accuracy comparison)Not explicitly stated, but implies meeting a predefined accuracy threshold.Device met all criteria in a simulated clinical environment.
    Software ValidationNot explicitly stated, but implies successful validation of software functionality.Software validations were conducted and presumably successful.

    2. Sample Size Used for the Test Set and Data Provenance:

    • The document does not specify sample sizes for any of the listed performance tests.
    • The provenance of the data (e.g., country of origin, retrospective or prospective) is not provided for any of the tests.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    • This information is not provided in the document. The tests described are primarily engineering, biocompatibility, and simulated environment tests, rather than clinical studies requiring expert ground truth for interpretation of patient data.

    4. Adjudication Method for the Test Set:

    • This information is not applicable as the provided testing is not a clinical study involving human reader interpretation requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

    • No, an MRMC comparative effectiveness study was not conducted or described in this document. The device is a guidance system, and the testing focuses on its engineering and functional performance rather than its impact on human reader performance in a diagnostic context.

    6. If a Standalone (algorithm only without human-in-the-loop performance) was done:

    • The document describes performance testing that largely functions as "standalone" performance for the device's components (electrical, safety, biocompatibility, mechanical). The "system worse-case accuracy comparison testing in a simulated clinical environment" evaluates the algorithm's accuracy in a controlled setting without explicit mention of human-in-the-loop performance influencing the measurement. However, the device is an "optional accessory" for use during a procedure and is meant to enhance ultrasonic images and predict tool paths, implying human interaction during its intended use. Therefore, while some tests are algorithmic/system performance, it's not a standalone diagnostic algorithm.

    7. The Type of Ground Truth Used:

    • Electrical and Safety Testing: Ground truth is established by the specified international standards (e.g., ANSI/AAMI ES 60601-1).
    • Biocompatibility Testing: Ground truth is established by the specified international standards (e.g., ISO 10993 series).
    • Mechanical (Guidance Sleeve end cap tensile testing): Ground truth would be engineered specifications or established industry standards for material strength.
    • Accuracy (System worse-case accuracy comparison): Ground truth would be defined reference points or known positional values within the simulated clinical environment.
    • Software Validation: Ground truth would be the software requirements specifications and expected functional behavior.

    8. The Sample Size for the Training Set:

    • The document does not mention a "training set" in the context of machine learning or AI models. The software is noted to be from the manufacturer of the predicate device and based on their cleared device, implying reuse or adaptation of existing software rather than a de novo development requiring a distinct training phase.

    9. How the Ground Truth for the Training Set was Established:

    • This information is not applicable as no training set for an AI model is described.
    Ask a Question

    Ask a specific question about this device

    K Number
    K132184

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-04-28

    (287 days)

    Product Code
    Regulation Number
    884.4160
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Acessa Guidance Handpiece is indicated for use in percutaneous laparoscopic coagulation and ablation of soft tissue, including treatment of symptomatic uterine fibroids under laparoscopic ultrasound guidance.

    Device Description

    Acessa Guidance Handpiece (Model Number 5300):

    The Acessa Guidance Handpiece, when used with the tracking system, is meant to guide the tip of the Handpiece up to the uterine serosa. Once the device is advanced into the uterus, standard ultrasound views are used to guide the tip to the fibroid. Ultrasound visualization for fibroid penetration and treatment remain unchanged.

    The Acessa Guidance Handpiece is a single-use only, sterile, radiofrequency (RF) applicator that connects to the Acessa Generator by a custom Cable. The Handpiece consists of a handle with a trocar shaft, active needle tip, and seven deployable needle electrodes each with embedded thermocouple sensors for real-time temperature feedback. The Acessa Guidance Handpiece also has control buttons that allow the user to access the User Interface.

    This Acessa Guidance Handpiece also contains the magnetic guidance sensor used to determine spatial location. The embedded sensor and SROM can be used with the electromagnetic tracking technologies to track the positions of an Ultrasound Transducer and the Handpiece shaft, and draws virtual representations of them in their spatial relationship, so that a physician can predict the Handpiece shaft's future path in relation to the features in the ultrasound slice. These electromagnetic tracking technologies are considered optional accessories to Acessa System procedure.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study information for the Acessa Guidance Handpiece based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test CategoryAcceptance CriteriaReported Device Performance
    Electrical and Safety TestingCompliance with EN/IEC 60601-1:2005/2006 (3rd edition) Medical Electrical Equipment Part 1 General Requirements for Safety; EN/IEC 60601-1-2:2007 (3rd edition) Medical Electrical Equipment General Requirements for Basic Safety and Essential Performance Collateral Standard Electromagnetic Compatibility Requirements and Tests; EN/IEC 60601-2-2:2009 (5th edition) Medical Electrical Equipment Particular Requirements for the Safety of High Frequency Surgical Equipment.Passed all testing.
    Biocompatibility TestingCompliance with ISO 10993-1:2009 (4th edition) Biological Evaluation of Medical Devices Part 1 Evaluation and Testing within a Risk Management Process; ISO 10993-5:2009 (3rd edition) Biological Evaluation of Medical Devices Tests for In Vitro Cytotoxicity; ISO 10993-10:2010 (2nd edition) Biological Evaluation of Medical Devices Tests for Irritation and Delayed-Type Hypersensitivity; and ISO 10993-11:2006 (2nd edition) Biological Evaluation of Medical Devices Tests for Systemic Toxicity.Demonstrated safety, suitability, and appropriateness for intended use.
    Flexural Strength TestingNot explicitly stated in terms of specific numeric criteria, but evaluated against established criteria.Met all criteria.
    Compatibility with EM TrackingNot explicitly stated in terms of specific numeric criteria, but evaluated against established criteria for functionality with an electromagnetic tracking system.Met all criteria.

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document describes performance testing of the device itself (electrical, safety, biocompatibility, mechanical, and compatibility with a tracking system), not a clinical study on patient data. Therefore, there isn't a "test set" in the context of patient data, nor is there data provenance in terms of country of origin or retrospective/prospective nature for such a clinical study. The testing was technical in nature, conducted on the device hardware and materials.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. As noted above, this was not a study evaluating performance against a clinical ground truth established by experts.

    4. Adjudication Method for the Test Set

    Not applicable. This was not a study involving expert adjudication of clinical cases.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document describes the 510(k) submission for a medical device (Acessa Guidance Handpiece) which is a physical instrument, not an AI or software algorithm. Therefore, an MRMC study comparing human readers with and without AI assistance is not relevant here.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is not an algorithm or AI device.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's performance testing was based on:

    • Established national and international standards for electrical safety (EN/IEC 60601 series).
    • Established national and international standards for biocompatibility (ISO 10993 series).
    • Internal criteria for flexural strength and compatibility with the electromagnetic tracking system.

    8. The Sample Size for the Training Set

    Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121858

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-11-05

    (132 days)

    Product Code
    Regulation Number
    884.4160
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Acessa System is indicated for use in percutaneous, laparoscopic coagulation and ablation of soft tissue, including treatment of symptomatic uterine fibroids under laparoscopic ultrasound guidance.

    Device Description

    The Acessa System includes the following system components: Generator (Model Number 1000): Provides RF energy to the Handpiece through the Handpiece Cable. Handpiece (Model Number 2000): Consisting of a disposable handle with a trocar-pointed shaft and 7 deployable needle electrodes. For use only disposable electrosurgical devices provided by Halt Medical, Inc. Handpiece Cable (Model Number 4200): Connects Handpiece to the Generator. This extension cable is provided with the Generator. Pads (Model Number 3000): A disposable set of 2 units, providing the return path for the RF energy applied by the Handpiece. Use only the Pads provided by Halt Medical, Inc. Pad Cable (Model Number 4300): Connects the Pads to the Generator. This extension cable is provided with the RF Generator. Power Cord (Model Number 4110): A medical grade power cord that provides AC power to the Generator. The Power cord is provided with the Generator. Foot Pedal (Model Number 4100): Pneumatic foot pedal with tubing used to turn RF energy on and off. The Foot Pedal with tubing is provided with the Generator. The Acessa System is designed to deliver up to 200 W of RF power at 460 kHz in three operational modes: Temperature Control, Manual Control and Coagulation Mode. A touch screen with a graphical user interface (GUI) enables selection of operational parameters such as the mode of operation, the ablation time, the target temperature, and the power delivery level. With the Handpiece placed in the tissue to be ablated and its electrodes deployed, RF power can be turned on. The system parameters are continuously monitored and displayed on the Generator. If the measured parameters are outside the acceptable limits, the RF energy delivery automatically stops and a message appears on the graphical user interface. RF energy during an ablation or coagulation can also be stopped at any time by the user by pressing the foot pedal. NOTE: The Acessa System must be used under laparoscopic ultrasound guidance. Laparoscopic ultrasound equipment is not included with the Acessa System.

    AI/ML Overview

    The provided document describes the Acessa System, a device for laparoscopic coagulation and ablation of soft tissue, specifically for symptomatic uterine fibroids. Here's a breakdown of the acceptance criteria and the studies conducted:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" in a tabulated format with specific thresholds. Instead, it describes various performance testing and clinical study outcomes that demonstrate the device's safety and effectiveness. Based on the provided information, the closest approximation of acceptance criteria and reported performance is:

    Acceptance Criteria (Inferred)Reported Device Performance (as demonstrated by studies)
    Safety - Electrical, Mechanical, Software, and BiocompatibilityThe Acessa System passed all electrical, mechanical, and software validation testing, as well as applicable safety requirements (EN/IEC 60601-1, EN/IEC 60601-1-2, EN/IEC 60601-2-2). EMC and safety testing were completed on the entire device and per component. Biocompatibility testing demonstrated materials are safe and in compliance with ISO 10993-1, ISO 10993-5, and ISO 10993-10. Device-related adverse events were less than 4% in the pivotal study, with only 2 serious adverse events (a pelvic abscess and a serosal colon injury) considered related to the device among 10 reported serious events. Only one (1.4%) serious adverse event (abdominal wall hematoma) was considered procedure-related in the Phase II studies.
    Functional Performance - Ablation Capability & ControlAnimal and bench ablation testing successfully demonstrated the Acessa System performs as intended and per specifications. Ablation capability was confirmed, providing a reproducible, discretely demarcated zone of tissue necrosis. The peri-hysterectomy feasibility study showed that ex vivo bench testing ablation results are predictive of ablations in fibroid tissue in vivo, with no device-related adverse events. The system continuously monitors parameters, automatically stops RF energy if outside acceptable limits, and allows user-initiated stops.
    Effectiveness - Reduction in Menstrual Blood Loss (for symptomatic uterine fibroids)Phase II Studies: At 12 months post-treatment, 3.5% (p<0.0001) of subjects reported heavy to very heavy bleeding, down from 77% at baseline. Pivotal Study: Mean reduction in menstrual blood flow at 12 months post-treatment was 103.6 ml. 83.9% of subjects experienced a reduction in blood loss. 40.2% (95% CI 31.6% - 48.7%) met the bleeding relief criterion (≥50% reduction in menstrual bleeding at 12 months), though this was below the prespecified study hypothesis of ≥45% lower bound.
    Effectiveness - Symptom Severity Reduction (for symptomatic uterine fibroids)Phase II Studies: Mean percent improvement in Symptom Severity Scores (UFS-QOL) was 59.6% at 3 months, 65.0% at 6 months, and 79.0% at 12 months. Overall, 94.2% reported improved Symptom Severity Scores. Pivotal Study: Continuous decrease in Symptom Severity Scores (SSS) from a high 61 at baseline to a low 24 at 24 months post-treatment. At least 94% of subjects responded that the treatment had been somewhat, moderately, and very effective in eliminating their symptoms.
    Effectiveness - Improvement in Health-Related Quality of Life (for symptomatic uterine fibroids)Phase II Studies: Mean percent change in HRQL Scores from baseline was 41.6% at 3 months, 39.7% at 6 months, and 42.9% at 12 months. Overall, 88.4% reported improved HRQL Scores. Pivotal Study: Steady improvement in mean Health-Related Quality of Life (HRQL) scores (from a low 37 at baseline to 80 after 24 months post-treatment). Subjects reported a decrease in disease burden and improved health state with mean EQ-5D scores starting at 71 at baseline and rising to 85 after 24 months.
    Effectiveness - Reduction in Uterine and Fibroid VolumePhase II Studies: Mean uterine volume significantly reduced to 155.0 cm³ (p=0.012) at 6 months and 151.4 cm³ (p=0.009) at 12 months, from 204.4 cm³ at baseline. 79.3% of subjects showed reduced uterine volume at 12 months. Pivotal Study: At 12 months post-treatment, mean reduction in uterine volume was 25.1%, and mean reduction in fibroid volume was 44.3% (assessed by contrast-enhanced MRI).
    Effectiveness - Low Reintervention RatePivotal Study: Within 24 months, a cumulative reintervention rate for fibroid-related bleeding was <6% (6 out of 107 subjects).
    Outpatient Minimally Invasive Option with Rapid Recovery and High Patient SatisfactionTreatment with this device provides an outpatient minimally invasive option that allows rapid recovery. 96% of subjects in the pivotal study were treated on an outpatient basis. Subjects returned to work in a median of 5 days, and all subjects reported returning to normal in a median of 9 days. 94% of subjects were "very, moderately, or somewhat satisfied" with the treatment. 98% of subjects would "probably or definitely recommend" the procedure to friends with the same health problem at 12 months.
    Feasibility in Varied Cases (e.g., fibroid types/locations, calcified fibroids) & Predictability of AblationThe peri-hysterectomy feasibility study encountered various fibroid locations and types. The Phase II and pivotal studies also treated various fibroid types and locations. Seven subjects (5.1%) with calcified fibroids were treated, and calcified fibroids did not appear to impact the physician's ability to insert the Handpiece tip. Bench testing was predictive of in vivo ablation.

    2. Sample Sizes Used for the Test Set and Data Provenance

    The document details three main clinical studies that serve as the "test set" for the device's performance:

    • Peri-hysterectomy feasibility study:
      • Sample Size: Not explicitly stated, but implies a small cohort of patients undergoing hysterectomy where the Acessa System was used.
      • Data Provenance: Not explicitly stated, but likely prospective and from a clinical setting where hysterectomies were being performed.
    • Prospective, non-randomized, longitudinal Phase II studies:
      • Sample Size: A total of 69 subjects (implied from the phrase "sixty-five subjects (94.2%)" and "sixty-one subjects (88.4%)" which point to a denominator around 69).
      • Data Provenance: Prospective, non-randomized, conducted at "two separate centers," likely in the US (not explicitly stated country, but common for IDE trials).
    • Device pivotal (Phase III) study:
      • Sample Size: 137 women with symptomatic fibroids.
      • Data Provenance: Prospective, conducted at 11 centers in the United States (9 centers) and Latin America (2 centers).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not describe a ground truth established by a specific number of experts for image interpretation or diagnosis in the traditional sense of AI studies. Instead, the "ground truth" for the clinical effectiveness studies was derived from:

    • Patient-reported outcomes: Validated questionnaires (UFS-QOL, EQ-5D), menstrual pad counts, patient surveys on satisfaction and symptom effectiveness, and self-reported return to normal activities.
    • Clinical measurements: Uterine volume, fibroid volume (assessed by contrast-enhanced MRI).
    • Pathology reports: From the peri-hysterectomy study, correlated with ablation zones.
    • Adverse event reporting: By investigators.
    • Investigators: 13 investigators participated in the pivotal study. Their specific qualifications (e.g., number of years of experience) are not detailed but they are presumed to be qualified clinicians (e.g., gynecologists) experienced in treating uterine fibroids.

    4. Adjudication Method for the Test Set

    The document does not mention an explicit adjudication method (like 2+1 or 3+1 consensus) for establishing a "ground truth" for the test set. The clinical trial data relies on patient reporting, objective clinical measurements, and investigator reporting of adverse events and outcomes based on standard clinical practice.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. The Acessa System is a therapeutic device (for ablation), not a diagnostic device that involves "human readers" interpreting images. Therefore, an MRMC study comparing human readers with and without AI assistance is not applicable to this device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    No. The Acessa System is a physical medical device (generator, handpiece, etc.) used by a physician during a laparoscopic procedure. It is not an algorithm or AI system that operates in a standalone capacity without human interaction for its primary function. While it has internal software for control and monitoring, its performance is inherently tied to human use and guidance.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for evaluating the Acessa System's performance was primarily a combination of:

    • Clinical Outcomes Data:
      • Patient-reported outcomes: From validated questionnaires (Uterine Fibroid Symptom Quality of Life (UFS-QOL), EuroQol 5 Dimensions (EQ-5D)), patient satisfaction surveys, symptom severity scales, and menstrual pad counts.
      • Objective clinical measurements: Reduction in uterine and fibroid volume (via contrast-enhanced MRI), reduction in menstrual blood loss, and reintervention rates.
      • Adverse event reporting: Clinical events reported by investigators and medical staff.
    • Pathology: In the peri-hysterectomy feasibility study, pathology reports of ablation zones were correlated with procedure documentation.
    • Device specifications: For electrical, mechanical, and software performance.

    8. The Sample Size for the Training Set

    The document does not describe a "training set" in the context of an AI/machine learning model. The studies described are clinical trials and engineering validations for a physical medical device. The "training" in this context would be the development and iterative testing of the device during its engineering phase, which is not quantified by a patient sample size but by laboratory and bench testing efforts.

    9. How the Ground Truth for the Training Set Was Established

    Since there is no "training set" in the AI/ML sense, this question is not directly applicable. For the device development itself, "ground truth" for design and engineering was established through:

    • Internal specifications and standards: Adherence to regulatory standards (EN/IEC 60601-1, EN/IEC 60601-1-2, EN/IEC 60601-2-2, ISO 10993).
    • Bench testing: To qualify and quantify mechanical and functional properties, ablation output, and safety features.
    • Animal studies: To demonstrate performance as intended and per specifications prior to human trials.
    Ask a Question

    Ask a specific question about this device

    K Number
    K094009

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-06-14

    (168 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Halt 2000GI™ Electrosurgical Radiofrequency Ablation System is indicated for use in percutaneous, laparoscopic, and intraoperative coagulation and ablation of soft tissue.

    Device Description

    The Halt 2000GI™ Electrosurgical Radiofrequency Ablation System includes the following system components:
    • Halt 2000GI Radiofrequency Generator (RF Generator): that provides RF energy to the RF Probe
    • Disposable RF Probe (Tulip™, TU 1000), a hand piece with a trocar-pointed shaft and 7 deployable RF needle electrodes.
    • RF Probe Extension Cable that connects the RF Probe to the RF Generator.
    • Dispersive Electrode Pads (TSP 115) Set that provides the return path for the RF energy applied by the RF Probe. (Use only Dispersive Electrode Pads provided by Halt Medical, Inc.)
    • Dispersive Electrode Pad Set Extension Cable (TSPCBL) that connects the Dispersive Electrode Pad Set to the RF Generator.
    • Power Cord that is a medical grade power cord providing AC power to the RF Generator.
    • Foot Pedal that is a pneumatic foot pedal with tubing used to turn RF energy on and off.

    The Halt 2000GI™ Electrosurgical Radiofrequency Ablation System is designed to deliver up to 200 W of RF power at 460 kHz in three operational modes: Temperature Control. Manual Control and Coagulation Mode. A touch screen with a graphical user interface (GUI) enables selection of operational parameters such as the mode of operation, the ablation time, the target temperature, and the power delivery level. With the RF probe placed in the tissue to be ablated and its electrodes deployed, RF power can be turned on. The system parameters are continuously monitored and displayed on the RF Generator. If the measured parameters are outside the acceptable limits, the RF energy delivery automatically stops and a message appears on the graphical user interface. RF energy during an ablation or coagulation can also be stopped at any time by the user by pressing the foot pedal.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Halt 2000GI™ Electrosurgical Radiofrequency Ablation System. This document focuses on demonstrating substantial equivalence to predicate devices rather than providing detailed acceptance criteria and a study report as would be found for a novel device. Therefore, much of the requested information, particularly regarding specific performance metrics, sample sizes for test and training sets, expert qualifications, and adjudication methods, is not explicitly available in the provided text.

    However, I can extract and infer some information based on the content.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Inferred from testing types)Reported Device Performance
    Electrical Safety Standards: EN/IEC 60601-1 (General medical electrical equipment safety) EN/IEC 60601-1-2 (Electromagnetic compatibility) EN/IEC 60601-2-2 (High-frequency surgical equipment) ANSI/AAMI HF-18 (Electrosurgical devices safety)The system passed all testing for applicable safety requirements.
    Mechanical Validation Testing:The system passed all testing.
    Software Validation Testing:The system passed all testing.
    Ablation Capability: (Performs as intended and per specifications)Animal and bench testing successfully demonstrated that the Halt RF Ablation System performs as intended and per specifications. The ablation capability was confirmed, providing a reproducible, discretely demarcated zone of tissue necrosis surrounded by normal tissue perfused with blood.
    Biocompatibility: (Safe, suitable, and appropriate for intended use, compliant with ISO 10993-1, ISO 10993-5, ISO 10993-10)A series of biocompatibility testing demonstrated that the device materials are safe, suitable, and appropriate for their intended use and in compliance with the specified ISO standards.

    2. Sample size used for the test set and the data provenance

    The document does not specify a distinct "test set" in the context of clinical or image-based studies. The testing mentioned includes:

    • Electrical, mechanical, and software validation.
    • Animal testing.
    • Bench testing.
    • Biocompatibility testing.

    The size of these "test sets" (e.g., number of animals, number of bench tests) is not provided. Data provenance (country of origin, retrospective/prospective) is also not specified for these types of engineering and pre-clinical tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided. The ground truth for the performance testing appears to be based on the outcomes of engineering tests (e.g., meeting electrical safety standards) and observations from animal and bench studies (e.g., "reproducible, discretely demarcated zone of tissue necrosis"). There is no mention of expert consensus for establishing ground truth in the context of a diagnostic or interpretive device.

    4. Adjudication method for the test set

    Not applicable to the types of testing described. Adjudication methods are typically relevant for human-in-the-loop studies or clinical trials involving subjective assessment, which are not detailed here.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not conducted. This device is an electrosurgical ablation system, not a diagnostic imaging device utilizing AI that would involve human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This device does not involve an "algorithm only" component in the sense of AI or image interpretation. Its performance is related to its ability to deliver RF energy and achieve tissue ablation, which was evaluated through animal and bench testing, as well as engineering validation.

    7. The type of ground truth used

    The ground truth used for performance validation appears to be:

    • Engineering specifications and regulatory standards for electrical, mechanical, and software validation.
    • Histopathological or visual assessment of tissue necrosis in animal and bench testing.
    • Biocompatibility standards (ISO 10993 series) for material safety.

    8. The sample size for the training set

    Not applicable. This document describes an electrosurgical device, not a machine learning model, so there is no concept of a "training set" in this context.

    9. How the ground truth for the training set was established

    Not applicable, as no training set is mentioned for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1