Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K092240

    Validate with FDA (Live)

    Date Cleared
    2010-04-21

    (272 days)

    Product Code
    Regulation Number
    880.5580
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended to pierce the skin in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states.

    Device Description

    The Acupuncture Needles are sterile, hair thin needles which are inserted into specific points on the skin, called "acupuncture points." The Acupuncture Needles are manufactured from stainless steel and sterilized with gamma irradiation. The Acupuncture Needles are available in eleven diameters (0.14 to 0.50 mm), nine needle lengths (15-60mm), two stick lengths (20 and 30 mm), and nine tube lengths (30-145 mm).

    AI/ML Overview

    Acupuncture Needles (K092240) - Acceptance Criteria and Performance Study

    The K092240 submission concerns Acupuncture Needles manufactured by HaengLim SeoWon Medical Co. The primary study presented aims to demonstrate substantial equivalence to predicate devices rather than establishing novel safety and efficacy criteria through a clinical trial.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implicit from Substantial Equivalence Claim)Reported Device Performance (K092240)
    Surface Smoothness & Freedom from Visible DefectsSurfaces found to be smooth and free of visible defects at 200-300X magnification.
    Pull-out ForceQuantified pull-out force fell within the average values reported for the predicate devices.
    Elasticity Properties (No Permanent Deformation after Cantilever-Type Loading)Returned to original shape after cantilever-type loading, showing no permanent deformation, similar to predicate devices.
    SterilitySterilized with gamma irradiation (implied to meet relevant standards).
    Material CompositionManufactured from stainless steel (consistent with predicate devices).
    Available Sizes (Diameters, Lengths)Available in eleven diameters (0.14 to 0.50 mm), nine needle lengths (15-60mm), two stick lengths (20 and 30 mm), and nine tube lengths (30-145 mm) - implies offering a comparable range to predicate devices.
    Intended UseIntended to pierce the skin in the practice of acupuncture by qualified practitioners of acupuncture (identical to predicate devices).

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a sample size in terms of a "test set" in the traditional sense of a clinical or in vivo study. The evaluation appears to be a bench-top comparison between the subject device and two predicate devices.

    • Test Set: The "test set" consisted of the Acupuncture Needle (subject device) and two predicate devices (Dong Bang Acupuncture Needle, K972659 and C & G Acupuncture Needle, K043277).
    • Data Provenance: The study was a retrospective bench-top comparison performed on the devices themselves. No details on the origin of the specific samples (e.g., country of origin for the tested devices) are provided, beyond the manufacturing location of the subject device (Korea) and the listed applicants of the predicate devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    The study does not involve human observers or interpretation of medical images, thus the concept of "experts" establishing ground truth for a test set is not applicable in the traditional sense. The "ground truth" for the mechanical and microscopic properties was based on direct measurement and observation by the individuals conducting the testing. Their qualifications are not specified, but it's presumed they were technicians or engineers competent in microscopy and mechanical testing.

    4. Adjudication Method for the Test Set

    No adjudication method was used, as the study involved direct physical and mechanical testing rather than human interpretation requiring consensus.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was conducted. This device is an acupuncture needle, and the study focuses on its physical and mechanical properties compared to legally marketed predicate devices, not on the improvement of human readers with AI assistance.

    6. Standalone (Algorithm Only) Performance Study

    Not applicable. The device is a physical medical instrument (acupuncture needle), not an algorithm or AI-powered system.

    7. Type of Ground Truth Used

    The ground truth used was based on direct physical and mechanical measurements and microscopic observations. This includes:

    • Visual assessment under magnification for surface defects.
    • Quantitative measurement of pull-out force.
    • Qualitative assessment of elasticity (return to original shape after loading).

    8. Sample Size for the Training Set

    Not applicable. There was no "training set" as this was a physical device comparison study, not a machine learning or AI context.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there was no training set, there was no ground truth for a training set to be established.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1