Search Filters

Search Results

Found 8 results

510(k) Data Aggregation

    K Number
    K252697

    Validate with FDA (Live)

    Device Name
    3D Laparoscope
    Date Cleared
    2025-10-29

    (63 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Stellar Imaging System is intended to provide real-time visible and near-infrared imaging in minimally invasive procedures.

    Upon intravenous administration and use of an ICG agent consistent with its approved label, Stellar Imaging System is used to perform fluorescence imaging visualization of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts.

    Upon interstitial administration and use of an ICG agent consistent with its approved label, the Stellar Imaging System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.

    Device Description

    Stellar Imaging System comprises Image Processing Unit, Light Source, Camera Head, Laparoscope and relevant accessories. During surgical procedures, Stellar Imaging System is used to provide real-time 3D visible and fluorescence imaging in endoscopic surgery.

    The Stellar Imaging System including the laparoscope is designed to work with an approved infrared dye (Indocyanine green (ICG). ICG may be excited at excitation at either 785 or 805 nm. The System provides excitation light to the surgical field to excite the dye molecules and captures emission from the dye via the camera head. Fluorescence imaging with ICG permits the system to visualize blood flow and related tissue circulation, of lymphatic flow, etc.

    The System allows the capture of normal (white) light image in parallel with the fluorescence image and displays both to the surgeon to provide a view of the anatomy. In addition, the System permits recording surgical procedures, storing them on removable storage devices, and replaying the procedures.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K242513

    Validate with FDA (Live)

    Date Cleared
    2024-10-23

    (61 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Upon intravenous administration and use of an ICG consistent with its approved label, the FloNavi Endoscopic Fluorescence Imaging System is intended to provide real time endoscopic visible and near-infrared fluorescence imaging. During minimally invasive surgery, the FloNavi Endoscopic Fluorescence Imaging System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts, using near-infrared imaging.

    Upon interstitial administration and use of an ICG consistent with its approved label, the FloNavi Endoscopic Fluorescence Imaging System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.

    Device Description

    Not Found

    AI/ML Overview

    This document is an FDA 510(k) clearance letter for the FloNavi Endoscopic Fluorescence Imaging System. It does not contain information about the acceptance criteria, device performance, study design, or validation results. Therefore, I cannot provide the requested information based on the provided text.

    The letter primarily covers:

    • The FDA's determination of substantial equivalence.
    • Regulatory classifications and product codes.
    • Applicable regulations for the device.
    • Guidance on when a new 510(k) might be required for changes.
    • General controls and other requirements such as the Quality System (QS) regulation and Unique Device Identification (UDI) Rule.
    • Contact information for the FDA.
    • Indications for Use for the device.

    To answer your questions, I would need a different document, such as a summary of safety and effectiveness, or the 510(k) submission itself, which would detail the performance testing and validation efforts.

    Ask a Question

    Ask a specific question about this device

    K Number
    K241530

    Validate with FDA (Live)

    Date Cleared
    2024-08-07

    (69 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Stellar Imaging System is intended to provide real-time visible and near-infrared imaging in minimally invasive procedures.

    Upon intravenous administration and use of an ICG agent consistent with its approved label, Stellar Imaging System is used to perform fluorescence imaging and visualization of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts.

    Upon interstitial administration and use of an ICG agent consistent with its approved label, the Stellar Imaging System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.

    Device Description

    Stellar Imaging System comprises Image Processing Unit, Light Source, Camera Head, Laparoscope, Exoscope and relevant accessories.

    During surgical procedures, Stellar Imaging System is used to provide real-time visible and fluorescence imaging similar to that provided by conventional imaging system used in endoscopic surgery and innovative 3D visible and fluorescence imaging in endoscopic surgery. Besides, Stellar Imaging System is also used to provide real-time fluorescence confirmation in open field.

    The Stellar Imaging System including the laparoscope and the exoscope is designed to work with an approved infrared dye (Indocyanine green (ICG). ICG may be excited at excitation at either 785 or 805 nm. The System provides excitation light to the surgical field to excite the dye molecules and captures emission from the dye using a camera head. Fluorescence imaging with ICG permits the system to visualize blood flow and related tissue circulation, of lymphatic flow, etc.

    The System allows the capture of normal (white) light image in parallel with the fluorescence image and displays both to the surgeon to provide a view of the anatomy. In addition, the System permits recording surgical procedures, storing them on removable storage devices, and replaying the procedures.

    AI/ML Overview

    The provided text is a 510(k) Summary for the Stellar Imaging System. It focuses on demonstrating substantial equivalence to a predicate device based on technological characteristics and safety standards. It does not contain information about the acceptance criteria for a study, the results of such a study, sample sizes, data provenance, ground truth establishment, expert qualifications, adjudication methods, or MRMC comparative effectiveness studies.

    The section titled "VII. Performance data" only lists the non-clinical tests conducted to verify compliance with various safety and performance standards. It states: "Performance comparison testing are also conducted on the subject device and predicate/reference device, demonstrate that the proposed system performs according to specifications and functions as intended." However, no specific details or results of this performance comparison testing are provided within this document.

    Therefore, I cannot fulfill your request for the specific information you asked for, as it is not present in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231342

    Validate with FDA (Live)

    Date Cleared
    2023-09-20

    (134 days)

    Product Code
    Regulation Number
    884.1730
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Insufflator (OPTO-IFL1000) is intended to generate and maintain pneumoperitoneum by filling the abdominal cavity with gas to distend it during diagnostic or therapeutic laparoscopic procedures.

    Device Description

    Insufflator (OPTO-IFL1000) is a CO2 insufflation device for creating and maintaining a pneumoperitoneum during laparoscopic examinations and operations. It is capable of establishing the surgical field of view and operating space. CO2 gas can be injected into abdominal cavity by the device, and the gas separates the abdominal wall from the internal organs of the abdominal cavity, forming a space for the operation and visual field. The device is to be used with the following insufflation tubes:

      1. OPTO-T1000H (with heating function)
      1. OPTO-T1000 (without heating function)
    AI/ML Overview

    The provided document is a 510(k) summary for the Insufflator (OPTO-IFL1000) from Guangdong OptoMedic Technologies, Inc. It describes the device, its indications for use, and a comparison to a predicate device. However, it does not contain the detailed information required to answer many of the specific questions about the device's acceptance criteria and the study proving it meets those criteria, particularly for an AI/ML-based medical device.

    The document states: "Performance testing were also conducted and demonstrate that the proposed system performs according to specifications and functions as intended. And the test result shows that the preset acceptance criteria are met." It then lists several performance tests, but it does not specify what those "preset acceptance criteria" are numerically, nor does it provide the reported device performance values against these criteria. Furthermore, it does not describe a study involving "human readers" or "AI assistance," as this is an insufflator, not an imaging analysis AI.

    Therefore, many of the questions cannot be answered from the provided text. I will answer what is available and indicate where information is missing.


    Acceptance Criteria and Study for Insufflator (OPTO-IFL1000)

    Based on the provided document, the device is an Insufflator (OPTO-IFL1000), which is a physical medical device designed to create and maintain pneumoperitoneum during laparoscopic procedures, not an AI/ML-based diagnostic or imaging device. Therefore, questions related to AI performance, human readers, ground truth establishment by experts, and training/test set sample sizes for AI models are not applicable to the information presented for this specific device.

    The document discusses "Performance data" and lists various tests conducted to verify the device met all design specifications and is substantially equivalent to a predicate device. The "acceptance criteria" are generally implied to be satisfaction of these test specifications and compliance with relevant standards.

    1. A table of acceptance criteria and the reported device performance

    The document lists performance tests but does not provide a table with specific numerical acceptance criteria or the numerically reported device performance for these tests. It only states that "the preset acceptance criteria are met."

    Below is a table of the performance tests mentioned, with the understanding that specific numerical criteria and results are not detailed in the provided text.

    Performance Test ItemAcceptance Criteria (Not Detailed)Reported Device Performance (Not Detailed)
    1. Gas Supply IndicationMet specificationsMet specifications
    2. Accuracy of the PressureMet specificationsMet specifications
    3. Accuracy of the Pressure-Under Leak Condition (Continuous leakage compensation testing)Met specificationsMet specifications
    4. Overpressure AlarmAlarm triggers as specified (e.g., >4 mmHg beyond nominal)Alarm triggers as specified
    5. Overpressure ReductionSystem reduces pressure effectively as specifiedSystem reduces pressure effectively
    6. Under-pressure Replenishment (Transient leakage compensation testing)System replenishes gas effectively as specifiedSystem replenishes gas effectively
    7. Accuracy of the FlowMet specificationsMet specifications
    8. Heating FunctionFunctions within specified temperature rangeFunctions within specified range
    9. Overheating AlarmAlarm triggers as specified (e.g., at >41°C)Alarm triggers as specified
    10. Accuracy of Gas Consumption DisplayMet specificationsMet specifications

    2. Sample size used for the test set and the data provenance

    For a physical device like an insufflator, "test set" and "data provenance" (country of origin, retrospective/prospective) are typically not relevant in the same way they are for AI/ML models using patient data. The "test set" would refer to the specific Insufflator (OPTO-IFL1000) unit(s) used for the described performance and safety testing. The document does not specify the number of units tested. The provenance of the testing itself is implied to be conducted by the manufacturer, Guangdong OptoMedic Technologies, Inc., in China.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is applicable to AI/ML devices that interpret data (e.g., medical images) where expert consensus is used to establish ground truth. For a physical device like an insufflator, "ground truth" pertains to its functional performance characteristics, which are measured using validated test methods and equipment, not by human expert interpretation of device output in the same way. Therefore, this information is not applicable and not provided.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable for a physical device performance test.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an insufflator, not an AI-assisted diagnostic tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is not an algorithm. Its performance is inherent to its physical and software functionality.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For this physical device, "ground truth" would be established by:

    • Compliance with recognized standards: Such as ANSI/AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-1-8, AAMI TIR 30, AAMI TIR 12.
    • Engineering specifications and measurements: Direct measurement of pressure, flow, temperature, alarm triggers, etc., against predefined engineering tolerances and clinical requirements.
    • Software verification and validation: Testing to ensure the software functions as intended to control the device and its safety features.

    The document states that "The software verification and validation testing were conducted and the test results demonstrated the software function met the requirements. The software for this device was considered a 'Major' level of concern." This indicates that software functionality, which dictates much of the device's "ground truth" operation, was rigorously tested.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device that requires a training set of data.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230407

    Validate with FDA (Live)

    Date Cleared
    2023-07-17

    (152 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Upon intravenous administration and use of an ICG consistent with its approved labeling, the FloNavi Open Field Fluorescence Imaging System is used to perform intraoperative fluorescence angiography.

    The FloNavi Open Field Fluorescence Imaging System is indicated for fluorescence imaging of blood flow and tissue perfusion before, during, and after: vascular, gastrointestinal, organ transplant, and plastic, micro- and reconstructive surgical procedures.

    Upon interstitial administration and use of an ICG consistent with its approved labeling, the FloNavi Open Field Fluorescence Imaging System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.

    Device Description

    The FloNavi Open Field Fluorescence Imaging System is an imaging system used in hospitals for capturing and viewing fluorescent images for the visual assessment of blood flow, as an adjunctive method for the evaluation of tissue perfusion, and related tissue transfer circulation for use in imaging during various surgical procedures.

    The FloNavi Open Field Fluorescence Imaging System has two system models: HD system and 4K system. Each system consists of the following main components: Imaging Head, Image Processing Unit, power supply cord and video cables.

    The Imaging Head may be either handheld or attached to a mechanical arm and provides illumination of the regions of a patient's body to be observed with near infrared light to excite ICG fluorescence. Alternatively, the Imaging Head provides white light illumination of the regions of a patient's body to be observed for color imaging. The cameras in the Imaging Head capture the fluorescent image under near infrared illumination or a color image under white light illumination. The Image Processing Unit receives the video signal from the Imaging Head and processes and outputs the video image to a medical grade video monitor and/or video recorder. Adjustments to the operation of the FloNavi Open Field Fluorescence Imaging System are possible through switches at either the Imaging Head or the Image Processing Unit.

    The FloNavi Open Field Fluorescence Imaging System is a reusable device and provided nonsterile. Its components should be cleaned and low-level disinfected prior to the first use and after every subsequent use.

    AI/ML Overview

    The FloNavi Open Field Fluorescence Imaging System is indicated for fluorescence imaging of blood flow and tissue perfusion and for visualization of the lymphatic system. It is a Class II device.

    Here's an analysis of the acceptance criteria and supporting study details based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document describes the device as being "Substantially Equivalent (SE) to the predicate device" (SPY Portable Handheld Imaging (SPY-PHI) System, K200737). The acceptance criteria are implicitly linked to the predicate device's performance and the subject device's ability to meet its own design specifications and relevant safety standards.

    Acceptance Criteria CategorySpecific Criteria (Implicit for SE, based on predicate)Reported Device Performance
    SafetyCompliance with relevant electrical, EMC, and photobiological safety standards.The device complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-57, and IEC 62471.
    PerformanceAbility to capture and display fluorescence and white light images; functional operation as intended.Performance testing demonstrated that the proposed system performs according to specifications and functions as intended.
    Software ValidationSoftware meets "Moderate level of concern" requirements per FDA guidance.Software was validated as "Moderate level of concern" in accordance with FDA guidance documents.
    Biocompatibility(Not explicitly mentioned, but typically required for devices with patient contact)(Not explicitly mentioned in the provided text.)
    Sterility(Not explicitly mentioned, but the device is stated as reusable and non-sterile, implying reprocessing procedures are defined.)The device is provided nonsterile. Components should be cleaned and low-level disinfected prior to first use and after every subsequent use.

    2. Sample Size and Data Provenance for Test Set

    The document mentions "Performance testing and animal testing" were conducted.

    • Sample Size:
      • Performance Testing: Not specified.
      • Animal Testing: Not specified.
    • Data Provenance: Not specified (e.g., country of origin, retrospective or prospective).

    3. Number of Experts and Qualifications for Ground Truth - Test Set

    Not applicable/Not specified. The document does not describe a clinical study involving human readers or expert-established ground truth for the "test set" in the context of diagnostic accuracy. The testing primarily focuses on the device's functional and safety performance, as well as animal studies.

    4. Adjudication Method for Test Set

    Not applicable/Not specified. No such adjudication method is mentioned as the studies described appear to be non-clinical performance and animal studies, not reader studies for diagnostic accuracy.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. The document does not mention a MRMC comparative effectiveness study where human readers' performance with and without AI assistance was evaluated. The device appears to be an imaging system, not an AI-powered diagnostic algorithm enhancing human reading.

    6. Standalone (Algorithm Only) Performance Study

    No. The document presents the FloNavi Open Field Fluorescence Imaging System as a standalone imaging device, not an AI algorithm. The performance testing and animal testing evaluate the device itself.

    7. Type of Ground Truth Used

    For the performance and animal testing mentioned, the "ground truth" would likely be based on:

    • Engineering Specifications: For the technical performance aspects (e.g., image resolution, light output).
    • Physiological Response: In animal testing, the "ground truth" for fluorescence imaging would be the actual blood flow, tissue perfusion, or lymphatic system visualization as observed and validated within the animal model, potentially through direct observation or comparison with established methods.

    8. Sample Size for Training Set

    Not applicable. This device is presented as an imaging system, not an AI/machine learning algorithm requiring a separate training set.

    9. How Ground Truth for Training Set was Established

    Not applicable, as there is no mention of an AI/machine learning component requiring a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231003

    Validate with FDA (Live)

    Date Cleared
    2023-07-06

    (90 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OptoMedic Laparoscopes are intended to be used for endoscopy and endoscopic surgery within the thoracic and peritoneal cavities including the female reproductive organs. The OptoMedic Laparoscopes are intended to be used by trained healthcare professional in diagnostic and therapeutic procedures.

    Device Description

    The OptoMedic laparoscopes are rigid endoscopes that are used for endoscopy and endoscopic surgery within the thoracic and peritoneal cavities including the female reproductive organs. An endoscope is a slender, tubular optical instrument used as a viewing system for examining an inner part of the body. The inside of the endoscope contains a series of lenses that transmit the endoscopic image, which is illuminated by an external light source. The Laparoscope has 8 models which are available in two insertion widths (5.6 mm and 10.1mm), two working lengths (320mm and 310mm) and two different directions of view (0° and 30°).

    AI/ML Overview

    This document is a 510(k) Premarket Notification for the OptoMedic Laparoscope, signifying its clearance by the FDA based on substantial equivalence to predicate devices. The information provided is primarily focused on the device's technical specifications and comparisons, rather than a clinical study demonstrating performance against specific acceptance criteria in the context of AI/ML.

    Therefore, for the request regarding acceptance criteria and a study proving the device meets them, it's important to understand that this submission is for a medical device (a laparoscope), not an AI/ML diagnostic tool. As such, the typical metrics and study designs for AI/ML performance (like those related to sensitivity, specificity, reader studies, etc.) are not directly applicable or provided in this document.

    However, I can extract the information related to the device's performance testing as described for regulatory clearance, which serves a similar purpose to demonstrating that the device meets its design specifications and is safe and effective for its intended use.

    Here's a breakdown based on the provided document, addressing the closest analogues to your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    For this medical device, "acceptance criteria" are generally derived from recognized industry standards and regulatory requirements for safety, performance, and functionality. The "reported device performance" is the demonstration that the device complies with these standards.

    Acceptance Criteria Category (Derived from Standards)Reported Device Performance (Compliance Statement)
    Electrical SafetyComplies with ANSI/AAMI ES 60601-1: 2005+A2 (R2012) +A1 (General requirements for basic safety and essential performance).
    Endoscopic Equipment SafetyComplies with IEC 60601-2-18:2009 (Particular requirements for the basic safety and essential performance of endoscopic equipment).
    General Endoscope RequirementsComplies with ISO 8600-1-2015 (General requirements for endoscopes).
    Optical Characteristics (Field of View & Direction)Complies with ISO 8600-3-2019 (Determination of field of view and direction of view of endoscopes with optics).
    Physical Characteristics (Maximum Width)Complies with ISO 8600-4-2014 (Determination of maximum width of insertion portion).
    Optical Characteristics (Resolution)Complies with ISO 8600-5-2020 (Determination of optical resolution of rigid endoscopes with optics).
    Biocompatibility (Cytotoxicity)Complies with ISO 10993-5:2009 (Tests for in vitro cytotoxicity).
    Biocompatibility (Sensitization)Complies with ISO 10993-10:2021 (Tests for skin sensitization).
    Biocompatibility (Systemic Toxicity)Complies with ISO 10993-11:2017 (Tests for systemic toxicity).
    Biocompatibility (Irritation)Complies with ISO 10993-23:2021 (Tests for irritation).
    Reprocessing (Cleaning)Complies with AAMI TIR 30:2016 (Cleaning reusable medical devices) and AAMI TIR 12:2020 (Designing, testing, and labeling).
    Overall Performance & Intended Functionality"Performance testing were also conducted on the subject device and demonstrate that the proposed system performs according to specifications and functions as intended."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not detail specific "sample sizes" in the context of clinical or AI/ML test sets. The performance evaluation primarily involved non-clinical tests conducted to verify compliance with national and international standards. These tests typically involve physical and functional evaluations of the device itself rather than patient data.

    • Sample Size: Not applicable in the context of AI/ML test sets. For device testing, it would involve a representative number of device units.
    • Data Provenance: Not applicable in the context of clinical data. The testing involves the physical device and its components, complying with engineering and safety standards.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable to this type of device clearance. Ground truth, in the context of image analysis or diagnostic performance, is not established for a rigid laparoscope in this document. The "ground truth" for this device's performance is its adherence to the technical specifications defined by the relevant standards (e.g., whether its field of view is indeed 68°, its resolution meets the specified criteria, or its materials are biocompatible). These are verified through engineering and laboratory testing.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods like 2+1 or 3+1 are used in clinical studies or expert reviews of data (often medical images) to establish a consensus ground truth. This document describes non-clinical engineering and bench testing, not clinical data review.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No. This document does not mention any MRMC comparative effectiveness study. These studies are typically conducted for AI/ML diagnostic tools to evaluate how AI assistance impacts human reader performance. The OptoMedic Laparoscope is a physical medical instrument, not an AI system.

    6. If a Standalone (Algorithm Only) Performance Study was Done

    No. This is not an AI/ML algorithm. Therefore, a standalone algorithm performance study was not conducted or reported.

    7. Type of Ground Truth Used

    The "ground truth" used for this device's regulatory clearance is compliance with established engineering, safety, and performance standards (e.g., ISO, IEC, AAMI standards) as verified through non-clinical laboratory testing. This is distinct from clinical ground truth based on pathology, expert consensus, or outcomes data, which would be relevant for diagnostic devices or AI algorithms.

    8. Sample Size for the Training Set

    Not applicable. This is not an AI/ML device that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no AI/ML training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221861

    Validate with FDA (Live)

    Date Cleared
    2023-02-18

    (236 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Upon intravenous administration and use of an ICG consistent with its approved label, the FloNavi Endoscopic Fluorescence Imaging System is intended to provide real-time endoscopic visible and near-infrared fluorescence imaging. During minimally invasive surgery, the FloNavi Endoscopic Fluorescence Imaging System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts, using near-infrared imaging.

    Upon interstitial administration and use of an ICG consistent with its approved label, the FloNavi Endoscopic Fluorescence Imaging System is used to perform intraoperative fluorescence imaging and visualization of the lymphatic system, including lymphatic vessels and lymph nodes.

    Device Description

    The proposed system, FloNavi Endoscopic Fluorescence Imaging System (FloNavi System) is comprised of an image processing unit, a camera head, and a light source (including a flexible light guide cable).

    There are two models of the proposed system. The primary components of each model are provided in Table 1.

    Model 1: Image Processing Unit: OPTO-CAM214K, Camera Head: OPTO-CHD214KE/OPTO-CHD214KH, Light Source: OPTO-LED214K
    Model 2: Image Processing Unit: OPTO-CAM2100, Camera Head: OPTO-CHD2100, Light Source: OPTO-LED2100

    FloNavi Endoscopic Fluorescence Imaging System is capable of providing real-time endoscopic visible and near-infrared fluorescence imaging.

    During surgical procedures, FloNavi may be operated to provide visualization similar to that provided by conventional imaging systems used in surgical endoscopy. The area of interest is illuminated with visible light from the light source and the resulting reflecting light is imaged by the camera and displayed on the video monitor. When used with the VIS-only laparoscopes, the System is only capable of the conventional mode of visualization described herein.

    To provide NIR fluorescence imaging, FloNavi is used with the imaging agent, indocyanine green (ICG). The patient is injected with ICG imaging agent. The ICG fluoresces when illuminated through the laparoscope with NIR excitation light from the light source, and the fluorescence response is then imaged with the camera, processed and displayed on a video monitor.

    The proposed system is designed to be used with rigid endoscopes, monitors and other ancillary equipment.

    AI/ML Overview

    The requested information about the acceptance criteria and study proving device performance is not available in the provided text. The document is an FDA 510(k) clearance letter and summary, which confirms substantial equivalence to a predicate device based on non-clinical testing, but does not detail acceptance criteria or specific performance study results of the type requested.

    The document explicitly states:
    "No clinical tests were performed." and "Performance testing was conducted on imaging performance and light source performance to support the marketing claims and to confirm that safety and effectiveness of the FloNavi Endoscopic Fluorescence Imaging System is at least equivalent to the predicate device."

    Therefore, I cannot provide:

    1. A table of acceptance criteria and the reported device performance.
    2. Sample size used for the test set and data provenance.
    3. Number of experts used to establish ground truth and their qualifications.
    4. Adjudication method for the test set.
    5. Effect size of human readers with vs. without AI assistance (as no clinical AI assistance study is mentioned).
    6. Standalone performance (algorithm only) was done. (While non-clinical performance was done, it's not described in terms of specific standalone algorithm performance, but rather overall system performance.)
    7. Type of ground truth used.
    8. Sample size for the training set.
    9. How the ground truth for the training set was established.
    Ask a Question

    Ask a specific question about this device

    K Number
    K221781

    Validate with FDA (Live)

    Date Cleared
    2023-02-02

    (226 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Image Processing Unit is intended to be used with compatible laparoscopes, camera heads, light sources, monitors, and other ancillary equipment for visualization, image recording and documentation during laparoscopic procedures.

    Endoscopic Camera Head is intended to be used with compatible laparoscopes, image processing unit, and other ancillary equipment for visualization, image recording and documentation during laparoscopic procedures.

    Device Description

    The Endoscopic Camera System comprises the Image Processing Unit, the Camera Head (including the objective lens), power supply cord and video cables. The Camera Head with integrated cable can be connected to the Image Processing Unit through the LEMO interface.

    During surgical procedures, the Endoscopic Camera System is used to provide real-time visible imaging similar to that provided by conventional imaging system used in surgical endoscopy. The area of interest is illuminated with visible light from the light source and the resulting reflected light is collected by a variety of laparoscopes that are attached to the camera head via equipment coupler. The camera head receives optical image from the connected laparoscope and converts optical image into electronic signal using the incorporated image sensor. The Image Processing Unit serves as the control center of the system and should be connected to a compatible light source in order to provide visualization. It can be connected to a 4K video monitor via one of the video cables provided to display image in real time: DVI cable, HDMI cable, 12G-SDI cable, and 3G-SDI cable.

    The proposed system is a reusable device and provided non-sterile. The Camera Head should be cleaned and high-level disinfected prior to the first use and after every subsequent use. The recommended compatible Light Source is OPTO-LED104K (510k exempted, product code: NTN).

    AI/ML Overview

    This FDA 510(k) summary describes the "Image Processing Unit, Model: OPTO-CAM104K; Camera Head, Model: OPTO-CHD104KH" device. However, it does not contain specific acceptance criteria, detailed study designs, or performance results from any clinical or standalone studies conducted to prove the device meets acceptance criteria. The sumamry focuses on non-clinical testing for safety and electrical compatibilty, and the statement of substantial equivalence to a predicate device.

    Therefore, many of the requested details about acceptance criteria, study sizes, ground truth, and expert involvement cannot be extracted from the provided text.

    Here's what can be gathered:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly list acceptance criteria in a table or directly report device performance against defined criteria in a way that would be typical for clinical effectiveness studies. It primarily focuses on compliance with general safety and EMC standards.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The document refers to "Performance testing" but does not detail the nature of these tests in terms of sample size or data collection methodology relevant to clinical performance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the document. The filing is for an endoscopic camera system, not an AI-powered diagnostic device that typically relies on expert-established ground truth for performance evaluation in the context of diagnostic accuracy.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided in the document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC comparative effectiveness study was not performed or reported as part of this 510(k) submission. This is not an AI-assisted diagnostic device, but rather an imaging system.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable as the device is an imaging system (hardware), not an algorithm or AI that operates standalone for diagnostic purposes. Its function is to provide visualization for humans during laparoscopic procedures.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This information is not provided and is not directly relevant for a device whose primary function is visualization during procedures rather than automated diagnosis. The "performance testing" mentioned likely refers to technical specifications of image quality, resolution, latency, etc., compared to design specifications.

    8. The sample size for the training set

    This information is not provided and is not applicable. The device is a hardware imaging system, not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established

    This information is not provided and is not applicable for the reasons mentioned above.


    Summary of what is provided:

    The document primarily focuses on demonstrating substantial equivalence to a predicate device (Image1 SPIES, K131953) through:

    • Non-clinical tests: Verifying compliance with various electrical safety, electromagnetic compatibility, and equipment-specific standards (e.g., ANSI/AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-2-18, ASTM D4169-16, AAMI TIR 30:2016, AAMI TIR 12:2020).
    • Software Validation: The software of the proposed device was validated as having a "Moderate level of concern (LoC)" according to FDA guidance.
    • Performance testing: Stated that performance testing was conducted to demonstrate the system "performs according to specifications and functions as intended," but specific details of these tests, including acceptance criteria and results, are not elaborated.

    The provided information establishes the device's technical safety and fundamental functionality in accordance with relevant standards, rather than clinical performance against specific diagnostic or intervention-related acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1