Search Filters

Search Results

Found 19 results

510(k) Data Aggregation

    K Number
    K232444

    Validate with FDA (Live)

    Date Cleared
    2023-12-20

    (128 days)

    Product Code
    Regulation Number
    878.4460
    Age Range
    0 - 65
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The glove is intended to be worn by operating room personnel to protect a surgical wound from contamination. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978.

    Device Description

    Sterile Powder Free Latex Surgical Gloves, Tested for Use with Chemotherapy Drugs is a disposable single-use, sterile, natural-colored and powder-free surgical glove made from rubber latex. They are in different sizes - 5.5, 6.0, 6.5, 7.0, 7.5, 8.0, 8.5, 9.0. Gloves meet the specification of ASTM D3577-19 and have been tested for resistance to permeation by chemotherapy drugs as per ASTM D6978-05(2019).

    AI/ML Overview

    The provided text describes the acceptance criteria and the study that proves a medical device (Sterile Powder Free Latex Surgical Gloves, Tested for Use with Chemotherapy Drugs) meets those criteria.

    Here's the breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The device's performance is primarily assessed against ASTM standards and biocompatibility tests. The table below summarizes these, focusing on the key performance characteristics relevant to general surgical gloves and their resistance to chemotherapy drugs.

    Acceptance Criteria / Performance MetricAcceptance CriteriaReported Device Performance
    Physical Properties (per ASTM D3577-19)
    Length (size 5.5)≥ 245 mm283-308 mm
    Length (size 6-9)≥ 265 mm283-308 mm
    Palm Width (e.g., size 5.5, 6.0, etc.)Specific ranges (e.g., 5.5: 70±6 mm, 6.0: 76±6 mm, etc.)Average values within specified ranges (e.g., 5.5: 72 mm, 6.0: 76 mm)
    Finger ThicknessMinimum 0.10 mm0.20-0.26 mm
    Palm ThicknessMinimum 0.10 mm0.19-0.24 mm
    Cuff ThicknessMinimum 0.10 mm0.16-0.19 mm
    Tensile Strength, Before Aging≥ 24 MPaAverage 24-29 MPa
    Ultimate Elongation, Before Aging≥ 750 %Average 750-851%
    Stress at 500% Elongation≤ 5.5 MPaAverage 2.1-4.5 MPa
    Tensile Strength, After Accelerated Aging≥ 18 MPaAverage 18-25 MPa
    Ultimate Elongation, After Accelerated Aging≥ 560 %Average 610-796 %
    Other Performance Metrics
    Freedom from holes (per ASTM D3577-19, ASTM D 5151-19)AQL 1.5 requirementsMeets AQL 1.5
    Powder-Free (per ASTM D3577-19, ASTM D6124-06 (2022))≤ 2 mg per glove0.09-0.12 mg per glove
    Aqueous Extractable Protein Content (per ASTM D3577-19, ASTM D5712-15)≤ 200 ug/dm²0.07-0.34 ug/dm²
    Sterility Assurance Level10-6 SAL10-6 SAL
    Chemotherapy Drug Permeation (per ASTM D6978-05 (2019))
    Bleomycin Sulfate 15mg/ml (15000 ppm)Minimum Breakthrough Detection Time (BDT) based on predicate/reference devices (typically >240 minutes for most drugs, except Carmustine and Thiotepa)>240 minutes
    Busulfan 6mg/ml (6,000 ppm)->240 minutes
    Carboplatin 10mg/ml (10,000 ppm)->240 minutes
    Carmustine (BCNU)3.3 mg/ml (3,300 ppm)Comparative to predicate/reference (e.g., predicate 13.2 min, reference 12.6 min)14.0 minutes (Note: Warning: Do not use with Carmustine)
    Cisplatin 1mg/ml (1,000 ppm)->240 minutes
    Cyclophosphamide 20mg/ml (20,000 ppm)->240 minutes
    Cytarabine HCL, 100 mg/ml (100,000 ppm)->240 minutes
    Dacarbazine 10 mg/ml (10,000 ppm)->240 minutes
    Doxorubicin HCL, 2 mg/ml (2,000 ppm)->240 minutes
    Etoposide, 20 mg/ml (20,000 ppm)->240 minutes
    Epirubicin HCL, 2 mg/ml (2,000 ppm)->240 minutes
    Fludarabine, 25 mg/ml (25,000 ppm)->240 minutes
    Fluorouracil, 50mg/ml (50,000ppm)->240 minutes
    Idarubicin HCL, 1mg/ml (1,000ppm)->240 minutes
    Ifosfamide, 50mg/ml (50,000ppm)->240 minutes
    Mechlorethamine HCI, 1mg/ml (1,000ppm)->240 minutes
    Melphalan, 5mg/ml (5,000ppm)->240 minutes
    Methotrexate, 25mg/ml (25,000ppm)->240 minutes
    Mitomycin C, 0.5mg/ml (500ppm)->240 minutes
    Mitoxantrone HCL, 2mg/ml (2,000ppm)->240 minutes
    Paclitaxel, 6mg/ml (6,000ppm)->240 minutes
    Paraplatin, 10mg/ml (10,000ppm)->240 minutes
    Rituximab, 10mg/ml (10,000ppm)->240 minutes
    Thio Tepa, 10mg/ml (10,000ppm)Comparative to predicate/reference (e.g., predicate 12.0 min, reference 22.4 min)16.1 minutes (Note: Warning: Do not use with Thiotepa)
    Vincristine Sulfate, 1mg/ml (1,000ppm)->240 minutes
    Biocompatibility (per ISO 10993 series)
    Skin Sensitization (ISO 10993-10)No significant evidence of causing skin sensitizationNo significant evidence of causing skin sensitization
    Intracutaneous Reactivity (ISO 10993-23)Polar and non-polar extract score less 1.0Test met requirements (score less 1.0)
    Cytotoxicity (ISO 10993-5)Acceptable level of cytotoxicity (comparison to predicate for "grade 2 at 1:16 dilution, grade 0 at 1:32 and 1:64 dilutions")Showed potential toxicity to L929 cells (Note: This is presented as meeting the acceptance criteria by the submitter, implying it's acceptable for the device type)
    Acute Systemic Toxicity (ISO 10993-11)No evidence of acute systemic toxicityNo evidence of acute systemic toxicity
    Material-mediated Pyrogenicity (ISO 10993-11)No rabbit shows an individual rise in temperature of 0.5°C or moreTest met requirements (no rabbit showed individual rise in temperature of 0.5°C or more)
    Endotoxin<20 EU/glove<20 EU/glove

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not explicitly state the specific sample sizes for each individual test. However, it indicates that "Non-clinical testing was performed to verify that the subject device meets the acceptance criteria of the performance test and all design specifications." The testing was conducted in accordance with recognized international and national standards (ASTM, ISO, U.S. Pharmacopeia).

    • Sample Size: Not explicitly stated for each test, but implied to be sufficient to meet the statistical requirements of the listed ASTM and ISO standards for each test (e.g., AQL 1.5 for freedom from holes implies a certain sample size for inspection).
    • Data Provenance: The manufacturer is "Grand Work Plastic Products Co., Ltd." in "Zanhuang, Hebei, 050000 China." The tests would have been performed by or for this company. The document does not specify if the testing was performed in China or by an external laboratory.
    • Retrospective or Prospective: The testing described is prospective, as it involves evaluating a new device against predefined standards to support its market clearance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This type of information (number and qualifications of experts) is generally relevant for studies involving subjective interpretation, especially in AI/imaging devices. For the physical and chemical testing of surgical gloves, "ground truth" is established by adherence to standardized test methods (e.g., ASTM D3577 for physical properties, ASTM D6978 for chemotherapy permeation, ISO 10993 for biocompatibility). The "experts" in this context would be the technicians and scientists performing and interpreting these standardized tests in a laboratory setting, who are qualified to execute these specific methodologies. Their specific number or individual qualifications are not detailed in this type of regulatory submission but are assumed to meet the requirements of the testing laboratories.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Adjudication methods like "2+1" or "3+1" are typically used in clinical studies or expert review processes where there is subjective interpretation involved, especially for ground truth establishment in medical imaging or diagnostic devices. For the non-clinical physical, chemical, and biocompatibility testing of surgical gloves, such adjudication methods are not applicable. The results are obtained through objective measurements and standardized protocols with predetermined pass/fail criteria.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. An MRMC comparative effectiveness study is not relevant for this device. This is a physical medical device (surgical glove), not an AI-powered diagnostic or assistive technology that involves human readers interpreting cases.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No. This concept is not applicable to a physical surgical glove. "Standalone" performance testing for this device refers to its physical and chemical properties and its barrier properties, measured objectively according to specified standards, without any "human-in-the-loop" component in its performance itself (though human labor is involved in its use).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device is based on established industry standards and objective physical/chemical measurements. Specifically:

    • ASTM D3577-19: Standard Specification for Rubber Surgical Gloves (defines dimensions, physical properties, freedom from holes).
    • ASTM D5151-19: Standard Test Method for Detection of Holes in Medical Gloves.
    • ASTM D6124-06 (2022): Standard Test Method for Residual Powder on Medical Gloves.
    • ASTM D6978-05 (2019): Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs (defines the permeation testing methodology).
    • ASTM D5712-15 (2020): Standard Test Method for the Analysis of Aqueous Extractable Protein in Natural Rubber and Its Products Using the Modified Lowry Method.
    • ISO 10993 series: Biological evaluation of medical devices (for biocompatibility tests like cytotoxicity, sensitization, systemic toxicity, pyrogenicity).
    • U.S. Pharmacopeia: Sterility Test.

    The pass/fail criteria and metric values within these standards serve as the "ground truth."

    8. The sample size for the training set

    This is not applicable. The device is a physical product (surgical glove), not a software algorithm that requires a training set.

    9. How the ground truth for the training set was established

    This is not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K232039

    Validate with FDA (Live)

    Date Cleared
    2023-08-09

    (30 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition, these gloves were tested for use with Chemotherapy drugs and fentanyl citrate in accordance with ASTM D6978.

    Device Description

    The subject device is Nitrile Examination Gloves Sterile with Chemotherapy drugs and fentanyl citrate claims. The subject device is a patient examination glove made from nitrile compound, blue color, powder free and sterile. They are ambidextrous and come in different sizes - Extra Small, Medium, Large, Extra Large and XXL. The device meets all the specifications in ASTM D6319-19, Standard specification for Nitrile Examination Gloves. Additionally, the gloves have been tested for permeability to chemotherapy drugs and fentanyl citrate per ASTM D6978-05(2019).

    AI/ML Overview

    The document provided is a 510(k) Premarket Notification from the FDA for "Sterile Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs and Fentanyl Citrate." This document primarily focuses on demonstrating substantial equivalence to a predicate device, specifically by adding a claim for fentanyl citrate permeation. It performs non-clinical testing to support this claim, rather than a study involving AI or human readers for diagnostic purposes.

    Therefore, many of the requested details regarding AI device acceptance criteria, human reader studies, and AI-specific ground truth establishment are not applicable to this submission. This is a clearance for a physical medical device (gloves), not a diagnostic AI/ML device.

    However, I can extract the relevant information regarding the device's acceptance criteria and the non-clinical study conducted to prove it meets those criteria.


    Acceptance Criteria and Study for Sterile Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs and Fentanyl Citrate

    This medical device is a physical product (examination gloves), not an AI/ML diagnostic tool. The "acceptance criteria" here refer to the physical and chemical performance standards for the gloves, and the "study" refers to non-clinical bench testing.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the re-submitted device (K232039) are based on relevant ASTM and ISO standards, and for chemotherapy and opioid drug permeation, specific minimum breakthrough times (BDT). The reported device performance indicates that the gloves passed all these criteria.

    Referenced Standards / Test PerformedAcceptance CriteriaReported Device Performance
    Physical Dimensions (ASTM D6319-19)Passed
    LengthMinimum 220mm for size XS and S, 230mm for size M, L, XLPass (All size ≥290mm reported in 510(k) Summary, page 5)
    Palm WidthXS: 70±10mm; S: 80±10mm; M: 95±10mm; L: 110±10mm; XL: 120±10mm; XXL: 130±10mmPass (Reported ranges: XS: 76-78mm; S: 86-88 mm; M: 96-98mm; L: 106-108 mm; XL: 116-118 mm; XXL: 126-128 mm)
    Thickness (Finger & Palm)Finger: 0.05mm (min); Palm: 0.05mm (min)Pass (Reported ranges: Palm: 0.09-0.11mm; Finger: 0.13-0.14mm)
    Physical Properties (ASTM D6319-19, ASTM D412-16 (2021))Passed
    Tensile Strength (Before Aging)Min 14 MPaPass (Reported range: 15.8-20.8 MPa)
    Tensile Strength (After Accelerated Aging)Min 14 MPaPass (Reported range: 15.6-19.8 MPa)
    Ultimate Elongation (Before Aging)Min 500%Pass (Reported range: 500-560%)
    Ultimate Elongation (After Accelerated Aging)Min 400%Pass (Reported range: 400-500%)
    Freedom from holes (ASTM D6319-19, ASTM D5151-19)G-I, AQL 2.5Pass (Meet AQL2.5 requirements)
    Powder Residue (ASTM D6319-19, ASTM D6124-06 (2017))Max 2 mg/glovePass (≤ 2 mg, meet requirements)
    Chemical Permeation (ASTM D6978-05 (2019))Passed (Refer to specific drug breakdown for values)
    Chemotherapy DrugsSee "Indications for Use" table on pages 3 and 6, and "Chemotherapy Permeation Comparison Claim" table on pages 7-8 for specific minimum BDTs for 15 listed drugs.All reported BDTs met or exceeded the minimum requirements. For example, Carmustine (BCNU) had a BDT of 54.3 min (meeting criteria), and many others had >240 min.
    Opioid Drug (Fentanyl Citrate Injection)Minimum BDT >240 minutes>240 minutes
    Biocompatibility (ISO 10993-10:2010)No Skin sensitization and Skin irritationIs non-sensitization and Non-irritation
    Biocompatibility (ISO 10993-11:2017)Subject showed no adverse biological reactionno evidence of acute systemic toxicity.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the exact sample sizes used for each physical or chemical test. However, for tests like "Freedom from holes," it refers to AQL (Acceptable Quality Level) 2.5, which implies a sampling plan according to ASTM D5151-19. Similarly, other ASTM and ISO standards followed for physical properties and biocompatibility would specify sample sizes.

    The data for these tests would be generated prospectively as part of product development and quality control for the submission. The country of origin for the data generation is implied to be China, given the manufacturer's location.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This concept is not applicable here. Ground truth in the context of this submission refers to the established standards and methodologies of the ASTM and ISO tests (e.g., how to measure tensile strength, how to test for permeation). These are standardized laboratory tests, not subjective expert interpretations of images or clinical outcomes. The "experts" are the laboratory personnel who perform these tests according to the published standards.

    4. Adjudication Method for the Test Set

    Not applicable. This is not a study involving human readers or subjective interpretations requiring adjudication. The tests have predefined, objective measurement protocols.

    5. If a Multi Reader Multi Case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a physical medical device (gloves), not an AI-assisted diagnostic tool. No MRMC study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. There is no algorithm or AI component in this device. Performance is measured directly through bench testing.

    7. The type of ground truth used

    The "ground truth" for this device's performance is established by objective, standardized laboratory measurements against predefined performance specifications stated in ASTM (American Society for Testing and Materials) and ISO (International Organization for Standardization) standards. Specifically:

    • Physical Properties: Measured values (e.g., length, width, thickness, tensile strength, elongation) are compared directly against minimum/maximum thresholds defined in ASTM D6319-19 and ASTM D412-16.
    • Freedom from holes: Determined by water leak testing according to ASTM D5151-19, with acceptance based on AQL 2.5.
    • Powder Residue: Quantified according to ASTM D6124-06 (2017) against a maximum limit.
    • Chemical Permeation: Breakthrough Time (BDT) is measured for specific chemotherapy drugs and fentanyl citrate according to ASTM D6978-05 (2019) and compared against specified minimum times.
    • Biocompatibility: Assessed through in-vitro/in-vivo tests for irritation, sensitization (ISO 10993-10:2010), and acute systemic toxicity (ISO 10993-11:2017) to ensure no adverse biological reactions.

    8. The Sample Size for the Training Set

    Not applicable. As this is not an AI/ML device, there is no "training set."

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K230217

    Validate with FDA (Live)

    Date Cleared
    2023-05-13

    (107 days)

    Product Code
    Regulation Number
    878.4460
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This glove is intended to be worn by operating room personnel to protect a surgical wound from contamination. In addition, these gloves were tested for use with chemotherapy drugs in accordance with ASTM D6978-05(Reapproved 2019).

    Device Description

    Sterile Polyisoprene Powder Free Surgical Glove, tested for Use with Chemotherapy Drugs is a disposable singleuse, sterile, Cream-colored and powder-free surgical glove made from synthetic polyisoprene latex.

    AI/ML Overview

    This document describes the premarket notification (510(k)) for a medical device – "Sterile Polyisoprene Powder Free Surgical Glove, Tested for Use with Chemotherapy Drugs." This device is a Class I, reserved product, meaning it's subject to general controls but does not require extensive clinical data for approval if it demonstrates substantial equivalence to a legally marketed predicate device.

    The provided text focuses on the device's technical specifications and non-clinical testing to demonstrate substantial equivalence to a predicate device, not on proving an AI algorithm's performance in a diagnostic context. Therefore, many of the requested elements for describing an AI acceptance criteria study (e.g., sample size for test set, data provenance, expert ground truth, MRMC study, effect size, standalone performance, training set details) are not applicable to this submission.

    The "acceptance criteria" here refer to the product's physical, chemical, and performance standards as determined by established industry and regulatory test methods, not diagnostic performance metrics of an AI.

    Here's an interpretation of the "acceptance criteria" and "study" based on the provided document:

    Acceptance Criteria and Reported Device Performance

    The device's performance is demonstrated through adherence to various ASTM and ISO standards. The acceptance criteria are the minimum or maximum values specified by these standards, and the reported device performance indicates whether the tested product met these thresholds.

    CharacteristicAcceptance Criteria (Standard/Guidance)Reported Device PerformanceConclusion
    DimensionsASTM D3577-19Meets ASTM D3577 requirements for length, width, and thicknessPass
    LengthMinimum 265mm280-305mmPass
    Palm Width (size)Various (70±6mm to 114±6mm)Average values within range (e.g., 73mm for size 5.5, 114mm for size 9.0)Pass
    Thickness (Finger, Palm, Cuff)Minimum 0.10 mmFinger: 0.22-0.24 mm, Palm: 0.22 mm, Cuff: 0.17-0.18 mmPass
    Physical PropertiesASTM D3577-19Meets ASTM D3577-19Pass
    Tensile Strength, Before Aging≥ 17 MPaAverage 20-21 MPaPass
    Ultimate Elongation, Before Aging≥ 650 %Average 780-823 %Pass
    Stress at 500% Elongation7.0 MPa, maxAverage 2.5-3.1 MPaPass
    Tensile Strength, After Accelerated Aging≥ 12 MPaAverage 18-20 MPaPass
    Ultimate Elongation, After Accelerated Aging≥ 490 %Average 747-809 %Pass
    Freedom from holesASTM D3577-19, ASTM D 5151-19 (AQL 1.5)Meets AQL 1.5Pass
    Powder-FreeASTM D3577-19, ASTM D 6124-06(2022) (≤ 2 mg per glove)0.10-0.27 mg per glovePass
    Aqueous Extractable Protein ContentASTM D3577-19, ASTM D5712-15 (≤ 200 ug/dm2)≤ 50 ug/dm2Pass
    Sterility10-6 SAL10-6 SALPass
    BiocompatibilityISO 10993 standards (10, 23, 5, 11)Results indicate compliance or acceptable levelsPass
    Skin Sensitization TestNo significant evidence of causing skin sensitizationNo significant evidence of causing skin sensitizationPass
    Intracutaneous Reactivity TestScore < 1.0 (requirements met)Score < 1.0, requirements metPass
    Cytotoxicity Test(Acceptance criteria implied by predicate comparison)Potential toxicity to L929 cells (Similar to predicate)Pass
    Acute Systemic Toxicity TestNo evidence of Acute systemic toxicityNo evidence of Acute systemic toxicityPass
    Pyrogen TestNo rabbit shows individual rise in temperature of 0.5°C or moreMeets requirements of pyrogen testPass
    Chemotherapy Drug PermeationASTM D6978-05 (Reapproved 2019) (Minimum Breakthrough Detection Time)Varies per drug, compared to predicateSimilar

    Study Proving Device Meets Acceptance Criteria

    The "study" refers to a series of non-clinical tests conducted according to recognized international and industry standards.

    1. Sample Size Used for the Test Set and Data Provenance:

      • Sample Size: Not explicitly stated for each specific test, but the standards (e.g., ASTM D3577, ASTM D5151) would dictate the minimum sample sizes required for statistically valid testing (e.g., AQL levels for freedom from holes).
      • Data Provenance: The tests were performed to verify that the subject device meets the acceptance criteria. The location of the testing laboratories is not specified, but the manufacturing company is based in China. The data would be considered prospective as it was generated specifically for this 510(k) submission to demonstrate compliance.
    2. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

      • Not Applicable. This is a physical product testing scenario, not an AI diagnostic study. "Ground truth" is established by the standardized test methods themselves and the precise measurement of physical properties (e.g., length, thickness, tensile strength) or chemical properties (e.g., breakdown time, residual powder). The "experts" are the technicians and scientists performing these standardized tests in accredited laboratories. Their qualifications would involve training and expertise in material science, chemistry, and quality control relative to medical device testing.
    3. Adjudication Method for the Test Set:

      • Not Applicable. As this is not an AI diagnostic study, there is no human "adjudication" in the sense of reconciling expert opinions. The test results are typically quantitative measurements, and compliance is determined by whether these measurements fall within the specified acceptance criteria. Any discrepancies in test results would be handled through laboratory quality control procedures and repeat testing.
    4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

      • No. This type of study is relevant for diagnostic AI algorithms where human readers interpret medical images or data. It is not applicable to the testing of surgical gloves.
    5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Not Applicable. This product is a physical medical device, not an AI algorithm.
    6. The Type of Ground Truth Used:

      • Standardized Test Results and Established Benchmarks: The "ground truth" for the performance claims are the outcomes of tests conducted in accordance with well-established ASTM and ISO standards (e.g., ASTM D3577-19 for physical properties, ASTM D6978-05 for chemotherapy drug permeation, ISO 10993 series for biocompatibility). These standards define the methodologies and the "correct" or "acceptable" results for the device's characteristics.
    7. The Sample Size for the Training Set:

      • Not Applicable. There is no "training set" in the context of a physical product's performance validation as described here. This is not an AI/machine learning development.
    8. How the Ground Truth for the Training Set Was Established:

      • Not Applicable. (See point 7)

    In summary, the provided document details a non-clinical performance evaluation of a medical device against established industry standards and regulatory requirements, not a study of an AI algorithm's diagnostic performance. The "acceptance criteria" are the thresholds defined by these standards, and the "study" is the battery of tests conducted to demonstrate compliance.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222892

    Validate with FDA (Live)

    Date Cleared
    2022-12-15

    (83 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The glove is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. Gloves have been tested for use with chemotherapy drugs using ASTM D6978-05(2019)

    Device Description

    Powder Free Polyurethane Examination Gloves, Blue color, tested for use with Chemotherapy Drugs are Patient Examination Gloves, Disposable, single use only and non-sterile. The gloves are made of nitrile rubber materials and are powder free. They are ambidextrous and come in different sizes-XS, S, M, L, XL and XXL. Gloves meet the specification of ASTM D6319-19 and have been tested for resistance to permeation by chemotherapy drugs as per ASTM D6978-05(2019).

    AI/ML Overview

    The provided text describes the acceptance criteria and performance data for "Powder Free Polyurethane Examination Gloves, Blue color, Tested for use with Chemotherapy Drugs." This is a medical device, and the data pertains to its physical properties, biocompatibility, and resistance to chemotherapy drugs, all tested against established ASTM and ISO standards.

    It is important to note that this device is a physical product (gloves) and not an AI/ML powered medical device. Therefore, many of the requested criteria (e.g., sample size for AI test/training sets, data provenance, expert adjudication, MRMC studies) are not applicable in this context. The information provided focuses on material science and medical device performance standards for physical safety and efficacy.

    Here's a breakdown of the applicable information based on your request, along with an explanation for why certain criteria are not relevant:


    Acceptance Criteria and Device Performance for "Powder Free Polyurethane Examination Gloves"

    This device is not an AI/ML powered medical device. The acceptance criteria and performance are established through adherence to recognized international standards for medical examination gloves and specific testing for chemotherapy drug permeation.

    1. Table of Acceptance Criteria and Reported Device Performance

    MethodologyTest PerformedAcceptance CriteriaReported Device Performance
    ASTM D6319-19Physical Dimensions: LengthMinimum 230mm for all sizesPass
    ASTM D6319-19Physical Dimensions: Palm WidthXS: 70±10mm, S: 80±10mm, M: 95±10mm, L:110±10mm, XL: 120±10mm, XXL: 130±10mmPass
    ASTM D6319-19Physical Dimensions: ThicknessFinger: 0.05mm (min), Palm: 0.05mm (min)Pass
    ASTM D6319-19, ASTM D412-16(2021)Physical Properties: Tensile Strength and ElongationTensile Strength (Min 14 MPa) and Elongation (Before Aging 500% and After Aging 400%) MinPass
    ASTM D6319-19, ASTM D5151-19Water leak testAQL 2.5Pass
    ASTM D6319-19, ASTM D6124-06 (2017)Powder ResidueMax 2mg/glovePass
    ASTM D6978-05 (2019)Permeation by Chemotherapy DrugsSee detailed table below. Note: "Do not use with Carmustine or Thiotepa" is a critical warning based on the low breakthrough times measured for these drugs.Pass (as per specific BDTs)
    ISO 10993-10 & 23:2021Irritation And Skin SensitizationNo Skin sensitization and No Skin irritationIs non-sensitization and Non-irritation
    ISO 10993-5:2009CytotoxicityNot Cytotoxicshowed potential toxicity to L929 cells. (This indicates a finding, but the overall conclusion for biocompatibility still states "not sensitizers, irritants or systemic toxicity," implying this was deemed acceptable within the context of the device's use and other biocompatibility findings).
    ISO 10993-11:2017Acute systemic toxicity studySubject showed no adverse biological reactionno evidence of acute systemic toxicity.

    Chemotherapy Drug Permeation Data (from ASTM D6978-05 (2019) testing):

    Chemotherapy DrugMinimum Breakthrough Detection Time (BDT) in Minutes
    Carmustine 3.3 mg/ml (3,300 ppm)24.0
    Cisplatin 1mg/ml (1,000 ppm)>240
    Cyclophosphamide 20mg/ml (20,000 ppm)>240
    Dacarbazine 10 mg/ml (10,000 ppm)>240
    Doxorubicin HCL, 2 mg/ml (2,000 ppm)>240
    Etoposide, 20 mg/ml (20,000 ppm)>240
    Fluorouracil, 50mg/ml (50,000ppm)>240
    Methotrexate, 25mg/ml (25,000ppm)>240
    Paclitaxel, 6mg/ml (6,000ppm)>240
    Thiotepa, 10mg/ml (10,000ppm)0.6
    • Warning: Do not use with Carmustine or Thiotepa due to extremely low permeation times.

    2. Sample size used for the test set and the data provenance

    • Sample Size: The document does not explicitly state the sample size (number of gloves or test replicates) for each specific test (e.g., physical dimensions, watertightness, chemotherapy permeation). However, testing is performed in accordance with the methodologies described in the cited ASTM and ISO standards, which will specify appropriate sample sizes for validation.
    • Data Provenance: The document indicates the manufacturer is "Grand Work Plastic Products Co., Ltd" with a China address (Donggao Industrial Zone, Zanhuang, Hebei, 050000, China). This implies the testing would have been conducted or overseen by the manufacturer, likely at their facilities or accredited contract labs. The data is retrospective, as it was collected as part of the premarket notification (510(k)) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. For a physical device like examination gloves, "ground truth" is established through standardized laboratory testing protocols (e.g., tensile strength, leak tests, chemical permeation) as defined by ASTM and ISO organizations. These tests yield objective, quantitative measurements rather than subjective expert interpretations. Therefore, there are no "experts" in the sense of human readers/annotators establishing ground truth for a test set. The "experts" involved are the scientists and engineers who developed and validated these international testing standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable. Adjudication methods are relevant for studies where subjective human interpretation or labeling is involved (e.g., medical image analysis, clinical assessments). For physical product testing, the results are typically quantitative and objective, and thus do not require an adjudication process.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This is a physical medical device (gloves), not an AI-powered diagnostic or assistive tool. Therefore, MRMC studies are not relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • Ground Truth: The "ground truth" for this device's performance is established by the objective, quantitative measurements obtained through adherence to the specified ASTM and ISO standard test methods. These standards define the accepted methodologies for determining physical properties, biocompatibility, and chemical resistance. For example:
      • Physical properties (length, thickness, tensile strength, elongation): Measured directly using calibrated equipment according to ASTM D6319-19 and ASTM D412-16.
      • Watertightness: Determined by the "water leak test" (ASTM D5151-19) which is a pass/fail visual inspection after standardized water filling.
      • Powder Residue: Measured gravimetrically according to ASTM D6124-06.
      • Chemotherapy Drug Permeation (BDT): Determined by measuring the time it takes for a chemical to permeate the glove material using a detector, as per ASTM D6978-05.
      • Biocompatibility (Irritation, Sensitization, Cytotoxicity, Systemic Toxicity): Assessed through laboratory tests (e.g., cell cultures, animal models) conducted according to ISO 10993 standards, yielding specific biological responses.

    8. The sample size for the training set

    • Not Applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    • Not Applicable. This is not an AI/ML device.

    Summary of Study Proving Device Meets Acceptance Criteria:

    The study proving the device meets its acceptance criteria is a compilation of non-clinical performance tests conducted in accordance with well-established international standards (ASTM and ISO). These tests systematically evaluate the physical properties, chemical resistance, and biocompatibility of the gloves.

    The "Pass" results for each test (as shown in the table under "Non-Clinical Performance Data") demonstrate that the "Powder Free Polyurethane Examination Gloves" meet the pre-defined acceptance criteria for a Class I medical device of this type. The specifics of the testing (e.g., detailed methodology, equipment calibration, environmental controls) would be documented in the full test reports that support the 510(k) submission. The FDA's clearance (K222892) indicates that they have reviewed this data and found it sufficient to demonstrate substantial equivalence to legally marketed predicate devices, confirming the device's safety and effectiveness for its intended use. The critical warning regarding Carmustine and Thiotepa permeation times is a direct outcome of these tests and is vital for safe labeling and use.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213574

    Validate with FDA (Live)

    Date Cleared
    2022-07-12

    (244 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Nitrile Vinyl Blend Powder Free Examination Gloves (Black) is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    Nitrile Vinvl Blend Powder Free Examination Gloves (Black) are Patient Examination Gloves. Disposable, single use only and non-sterile. The gloves are made of vinyl and oil-based liquid nitrile rubber materials and are powder free. They are ambidextrous and come in different sizes-XS, S, M, L and XL. The physical and performance characteristics of the devices meet all requirements of ASTM D6319-19 Standard Specification for Nitrile Examination Gloves for Medical Application.

    AI/ML Overview

    The provided text is a 510(k) premarket notification for Nitrile Vinyl Blend Powder Free Examination Gloves (Black). This document focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing and does not involve AI or algorithms, human readers, or clinical performance data in the context of medical image analysis or similar diagnostic applications.

    Therefore, many of the requested categories are not applicable to this type of device and submission. Based on the provided text, here's the relevant information:

    1. Table of acceptance criteria and the reported device performance

    Test methodPurposeAcceptance CriteriaReported Device Performance
    ASTM D6319-19LengthAll size: Minimum 230mmXS: 230-245mmS: 230-245mmM: 230-245mmL: 230-245mmXL: 230-247 mm
    ASTM D6319-19Palm WidthXS: 70±10mmS: 80±10mmM: 95±10mmL: 110±10mmXL: 120±10mmXS: 76-78mmS: 86-88 mmM: 96-98mmL: 106-110 mmXL: 116-118 mm
    ASTM D6319-19ThicknessFinger: 0.05mm (min)Palm: 0.05mm (min)Finger: 0.06-0.08mmPalm: 0.07-0.09mm
    ASTM D6319-19ASTM D412-16Tensile Strength, Before Aging14MPa, min14.2-16.9 MPa
    ASTM D6319-19ASTM D412-16Tensile Strength, After Accelerated Aging14MPa, min14.1-16.2MPa
    ASTM D6319-19ASTM D412-16Ultimate Elongation, Before Aging500%, min500-520%
    ASTM D6319-19ASTM D412-16Ultimate Elongation, After Accelerated Aging400%, min400-480%
    ASTM D5151-19ASTM D6319-19Freedom from holesG-I, AQL 2.5XS: 0/125S:0/125M: 0/125L:1/125XL: 1/125
    ASTM D6124-06(2017)ASTM D6319-19Powder-Content≤ 2 mg per gloveXS:0.35mgS: 0.41mgM: 0.47mgL: 0.55mgXL: 0.66mg
    ISO 10993-10:2010Skin Irritation StudyUnder the conditions of the study, not an irritantUnder the conditions of the study, not an irritant
    ISO 10993-10:2010Maximization Sensitization StudyUnder the conditions of the study, not a sensitizerUnder the conditions of the study, not a sensitizer
    ISO 10993-5:2009Cytotoxicity TestUnder the conditions of this study, the test article extract showed potential toxicityUnder the conditions of this study, the test article extract showed potential toxicity
    ISO 10993-11:2017Systemic toxicityUnder the conditions of this study, there was no evidence of systemic toxicity.Under the conditions of this study, there was no evidence of systemic toxicity.

    2. Sample size used for the test set and the data provenance:

    • Freedom from holes: For each size (XS, S, M, L, XL), a sample size of 125 gloves was used.
    • Powder-Content: Not explicitly stated, but results are given for each size.
    • For other physical and biocompatibility tests, specific sample sizes are not detailed in the provided summary, but the tests were conducted.
    • Data provenance: Not specified, but the product owner is Grand Work Plastic Products Co., Ltd. from China, suggesting the tests were likely conducted on products manufactured in China. The study is retrospective as it refers to completed tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This is not applicable. The device is an examination glove, and its performance is evaluated against established material and biocompatibility standards (e.g., ASTM, ISO), not by expert interpretation of clinical images or data requiring ground truth establishment by medical experts.

    4. Adjudication method for the test set:

    Not applicable, as this is for the evaluation of a physical medical device against a set of standards, not an AI or diagnostic algorithm requiring adjudication of interpretations.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This submission is for a physical medical device (examination gloves), not an AI-powered diagnostic tool, and therefore no MRMC study or AI assistance evaluation was performed.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is not an algorithmic device.

    7. The type of ground truth used:

    The "ground truth" for this device is derived from established international standards for physical properties (e.g., tensile strength, dimensions, freedom from holes, powder content) and biological safety (biocompatibility tests for irritation, sensitization, cytotoxicity, systemic toxicity). These standards define the acceptable range or criteria for the device's performance.

    8. The sample size for the training set:

    Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established:

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220249

    Validate with FDA (Live)

    Date Cleared
    2022-05-26

    (115 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition these gloves were tested for use with Chemotherapy drugs in accordance with ASTM D6978 standards. Gloves is direct contact and not to be worn for more than 24 hours.

    Device Description

    The subject device is Sterile Nitrile Powder Free Examination Gloves Tested for Use with Chemotherapy Drugs And Fentanyl Citrate (Blue). The subject device is a patient examination glove made from nitrile latex compound, blue color, powder free and sterile (Per 21 CFR 880.6250, class I). The device is direct contact and not to be worn for more than 24 hours. The device meets all the specifications in ASTM D6319-19, Standard specification for Nitrile Examination Gloves. Additionally, the gloves have been tested for biocompatibility per 10993-11 and permeability to chemotherapy drugs and Fentanyl Citrate per ASTM D6978-05(2019).

    AI/ML Overview

    This document is a 510(k) premarket notification for "Sterile Nitrile Powder Free Examination Gloves Tested For Use With Chemotherapy Drugs And Fentanyl Citrate (Blue)". It details the device's characteristics, performance, and comparison to a predicate device.

    Here's the breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The device's performance is tested against various physical properties and chemical permeation criteria.

    Physical Properties (based on ASTM D6319-19):

    ItemsAcceptance CriteriaReported Device Performance (Results)
    LengthMinimum 230mmAll size ≥230
    Palm WidthXS: 70±10mm76-78mm
    S: 80±10mm86-88 mm
    M: 95±10mm96-98mm
    L: 110±10mm106-108 mm
    XL: 120±10mm116-118 mm
    XXL: 130±10mm126-128 mm
    ThicknessPalm: 0.05mm (min)0.07-0.08mm
    Finger: 0.05mm (min)0.09-0.11mm
    Tensile Strength, Before Aging14MPa, min15.7-20.6 MPa
    Tensile Strength, After Accelerated Aging14MPa, min15.5-19.5 MPa
    Ultimate Elongation, Before Aging500%, min500-550%
    Ultimate Elongation, After Accelerated Aging400%, min410-510%
    Freedom from holesG-I, AQL 2.5Meet and above AQL2.5 requirements
    Powder-Content≤ 2 mg per glove≤ 2 mg, meet requirements

    Chemotherapy Drug and Fentanyl Citrate Permeation (based on ASTM D6978 standards):

    Chemotherapy Drug / Fentanyl Citrate ConcentrationReported Minimum Breakthrough Detection Time (Minutes)
    Bleomycin Sulfate 15mg/ml (15000 ppm)>240
    Busulfan 6mg/ml (6,000 ppm)>240
    Carboplatin 10mg/ml (10,000 ppm)>240
    Carmustine 3.3 mg/ml (3,300 ppm)22.4
    Chloroquine 50mg/ml (50,000ppm)>240
    Cisplatin 1mg/ml (1,000 ppm)>240
    Cyclophosphamide 20mg/ml (20,000 ppm)>240
    Cyclosporin 100 mg/ml (100,000 ppm)>240
    Cytarabine HCL, 100 mg/ml (100,000 ppm)>240
    Dacarbazine 10 mg/ml (10,000 ppm)>240
    Daunorubicin HCL, 5 mg/ml (5,000 ppm)>240
    Docetaxel HCL, 10 mg/ml (10,000 ppm)>240
    Doxorubicin HCL, 2 mg/ml (2,000 ppm)>240
    Epirubicin HCL, 2 mg/ml (2,000 ppm)>240
    Etoposide, 20 mg/ml (20,000 ppm)>240
    Fludarabine, 25 mg/ml (25,000 ppm)>240
    Fluorouracil, 50mg/ml (50,000ppm)>240
    Gemcitabine, 38mg/ml (38,000ppm)>240
    Idarubicin HCL, 1mg/ml (1,000ppm)>240
    Ifosfamide, 50mg/ml (50,000ppm)>240
    Irinotecan, 20mg/ml (20,000ppm)>240
    Mechlorethamine HCI, 1mg/ml (1,000ppm)>240
    Melphalan, 5mg/ml (5,000ppm)>240
    Methotrexate, 25mg/ml (25,000ppm)>240
    Mitomycin C, 0.5mg/ml (500ppm)>240
    Mitoxantrone HCL, 2mg/ml (2,000ppm)>240
    Oxaliplatin, 5mg/ml (5,000ppm)>240
    Paclitaxel, 6mg/ml (6,000ppm)>240
    Paraplatin, 10mg/ml (10,000ppm)>240
    Retrovir, 10mg/ml (10,000ppm)>240
    Rituximab, 10mg/ml (10,000ppm)>240
    Thiotepa, 10mg/ml (10,000ppm)46.3
    Topotecan, 1mg/ml (1,000ppm)>240
    Trisenox, 1mg/ml (1,000ppm)>240
    Velcade, 1mg/ml (1,000ppm)>240
    Vincristine Sulfate, 1mg/ml (1,000ppm)>240
    Fentanyl Citrate Injection (100 mcg/2ml)>240

    Note: For Carmustine and Thiotepa, the document explicitly states warnings due to low permeation times, even though they have a reported BDT.

    Biocompatibility Studies (ISO 10993-10, ISO 10993-11, ISO 10993-5):

    Technological CharacteristicsStandard / Test / FDA GuidanceResult Summary (Acceptance Criteria & Performance)
    Primary Skin Irritation TestISO 10993-10:2010Meets (Under the conditions of the study, not an irritant)
    Dermal Sensitization AssayISO 10993-10:2010Meets (Under the conditions of the study, not a sensitizer)
    In Vitro Cytotoxicity TestISO 10993-5:2009The test article extract showed potential toxicity to L929 cells. Cytotoxicity concern was addressed by acute systematic toxicity testing.
    Acute Systemic Toxicity TestISO 10993-11:2017Meets (Under the conditions of this study, there was no evidence of systemic toxicity.)

    2. Sample size used for the test set and the data provenance

    The document specifies the standards used for testing, which implicitly defines the sample sizes and testing methodologies. However, it does not explicitly state the exact sample sizes used for each specific test (e.g., number of gloves tested for each chemotherapy drug).

    • Standards Used: ASTM D6319-19 (for dimensions, physical properties, freedom from holes, powder content) and ASTM D6978-05(2019) (for permeability to chemotherapy drugs and Fentanyl Citrate). Biocompatibility testing was done per ISO 10993-10, ISO 10993-11, and ISO 10993-5.
    • Data Provenance: Not explicitly stated, but based on the owner's identification (Grand Work Plastic Products Co., Ltd. in Hebei, China) and the contact information (Hongray USA Medical Products Inc. in Chino, California), the testing was likely conducted by or for the manufacturer, potentially in China or a contracted lab. The data is retrospective for the purpose of this 510(k) submission, as the tests have already been performed.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable as the document describes the testing of physical and chemical properties of gloves against established industry standards. The "ground truth" here is the performance requirement defined by these standards, not expert consensus on medical images or clinical outcomes. The tests are objective measurements, not subjective expert interpretations.

    4. Adjudication method for the test set

    Not applicable. The tests performed are objective, quantitative measurements against predefined specifications (e.g., minimum length, minimum tensile strength, breakthrough detection time). There is no "adjudication" in the sense of reconciling differing expert opinions.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a submission for sterile nitrile examination gloves, not an AI-powered diagnostic device. No human readers or AI assistance are involved in the performance evaluation of the gloves.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device (gloves), not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the device's performance is objective, laboratory-measured data based on established international standards:

    • Physical properties (dimensions, strength, elongation, hole detection, powder content) are measured quantitatively according to ASTM D6319-19, ASTM D5151-19, and ASTM D6124-06(2017).
    • Chemical permeation (breakthrough detection time for chemotherapy drugs and Fentanyl Citrate) is measured according to ASTM D6978-05(2019).
    • Biocompatibility (skin irritation, sensitization, systemic toxicity, cytotoxicity) is assessed using in vitro and in vivo tests guided by ISO 10993 standards.

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is a physical medical device, not a machine learning model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220250

    Validate with FDA (Live)

    Date Cleared
    2022-05-11

    (100 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sterile Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. In addition these gloves were tested for use with Chemotherapy drugs in accordance with ASTM D6978 standards.

    Device Description

    The subject device is Nitrile Examination Gloves Sterile with Chemotherapy drugs claims. The subject device is a patient examination glove made from nitrile latex compound, blue color, powder free and sterile (Per 21 CFR 880.6250, class I). The device is direct contact and not to be worn for more than 24 hours. The device meets all the specifications in ASTM D6319-19, Standard specification for Nitrile Examination Gloves. Additionally, the gloves have been tested for biocompatibility per 10993-10, 10993-11 and permeability to chemotherapy drugs per ASTM D6978-05(2019).

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA regarding Sterile Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs.

    This document does not describe the acceptance criteria and study proving a software algorithm device meets these criteria, but rather the acceptance criteria and test results for a physical medical device (examination gloves).

    Therefore, many of the requested categories in your prompt (such as sample size for test/training sets, data provenance, number of experts for ground truth, adjudication method, MRMC study, standalone performance, etc.) are not applicable to this type of device and the information provided in the 510(k) summary.

    However, I can extract the relevant information about the acceptance criteria and the device's performance based on the provided document.


    Device: Sterile Powder Free Nitrile Examination Gloves (Blue), Tested for Use with Chemotherapy Drugs

    Type of Device: Physical Medical Device (Examination Gloves), Class I, reserved

    Study Purpose: To demonstrate substantial equivalence to a predicate device (Riverstone Resources SDN BHD. Nitrile Examination Glove, Sterile Tested for use with Chemotherapy Drugs Claim, White (K190725)) by verifying that the subject device meets specific performance standards and has similar technological characteristics.

    1. Table of Acceptance Criteria and Reported Device Performance

    The document presents several tables detailing the acceptance criteria (standards/specifications) and the results indicating that the device meets these criteria.

    Item/TestAcceptance Criteria (Standard/Specification)Reported Device Performance (Results)
    Dimensions
    LengthMinimum 230mmAll size ≥290mm
    Palm Width (XS)70 ± 10mm76-78mm
    Palm Width (S)80 ± 10mm86-88mm
    Palm Width (M)95 ± 10mm96-98mm
    Palm Width (L)110 ± 10mm106-108mm
    Palm Width (XL)120 ± 10mm116-118mm
    Palm Width (XXL)130 ± 10mm126-128mm
    Physical Properties (Thickness)
    Thickness (Palm)0.05mm (min)0.09-0.11mm
    Thickness (Finger)0.05mm (min)0.13-0.14mm
    Physical Properties (Tensile Strength)
    Tensile Strength, Before Aging14MPa, min15.8-20.8 MPa
    Tensile Strength, After Accelerated Aging14MPa, min15.6-19.8 MPa
    Physical Properties (Ultimate Elongation)
    Ultimate Elongation, Before Aging500%, min500-560%
    Ultimate Elongation, After Accelerated Aging400%, min400-500%
    Integrity
    Freedom from holesG-I, AQL 2.5 (per ASTM D5151-19, ASTM D6319-19)Meet and above AQL 2.5 requirements
    Powder Content≤ 2 mg per glove (per ASTM D6124-06(2017), ASTM D6319-19)≤ 2 mg, meet requirements
    Biocompatibility
    Primary Skin Irritation TestNot an irritant (per ISO 10993-10:2010)Not an irritant
    Dermal Sensitization AssayNot a sensitizer (per ISO 10993-10:2010)Not a sensitizer
    Acute Systemic Toxicity TestNo evidence of systemic toxicity (per ISO 10993-11:2017)No evidence of systemic toxicity
    Chemotherapy Drug Permeation (Minimum Breakthrough Time - BDT)
    Carmustine (BCNU) (3.3mg/ml)- (Predicate: 39 min)54.3 minutes
    Cyclophosphamide (20mg/ml)- (Predicate: >240 min)>240 minutes
    Doxorubicin Hydrochloride (2 mg/ml)- (Predicate: >240 min)>240 minutes
    Etoposide (20mg/ml)- (Predicate: >240 min)>240 minutes
    Fluorouracil (50mg/ml)- (Predicate: >240 min)>240 minutes
    Paclitaxel (6mg/ml)- (Predicate: >240 min)>240 minutes
    Thiotepa (10mg/ml)- (Predicate: 97.6 min)196.7 minutes
    Methotrexate (25mg/ml)- (Predicate: >240 min)>240 minutes
    Cisplatin (1mg/ml)- (Predicate: N/A)>240 minutes
    Vincristine Sulfate (1.0 mg/ml)- (Predicate: N/A)>240 minutes
    Cytarabine HCL (100mg/ml)- (Predicate: N/A)>240 minutes
    Mitoxantrone (2.0 mg/ml)- (Predicate: N/A)>240 minutes
    Mitomycin C (0.5 mg/ml)- (Predicate: N/A)>240 minutes
    Ifosfamide (50.0 mg/ml)- (Predicate: N/A)>240 minutes
    Dacarbazine (DTIC) (10.0 mg/ml)- (Predicate: N/A)>240 minutes

    (Note: For chemotherapy drug permeation, the "Acceptance Criteria" are implied by comparison to the predicate device and showing equivalent or better performance. The exact criteria are not explicitly stated for each drug beyond the measured BDT.)

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not explicitly stated as a number of "cases" or "samples" in the way a software algorithm test set would be. The testing is based on batches of gloves or representative samples according to the referenced ASTM/ISO standards. For example, "Freedom from holes" testing follows an AQL (Acceptance Quality Limit) which implies a sampling plan.
    • Data Provenance: The tests are non-clinical and conducted based on international standards (ASTM, ISO). The manufacturer is Grand Work Plastic Products Co., Ltd, located in Donggao Industrial Zone, Zanhuang, Hebei, China. The testing would have been performed there or at certified labs used by the manufacturer. The data is retrospective in the sense that it was collected as part of the device's development and qualification prior to the 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The "ground truth" for this physical device's performance (e.g., tensile strength, breakthrough time) is established by adherence to standardized physical and chemical testing methods (ASTM, ISO standards), not by human expert consensus or clinical judgment in the traditional sense of diagnostic AI. These standards define the test procedures and acceptance limits.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not a study requiring human reader adjudication. Performance is measured by laboratory equipment and standard protocols.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-based diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI-based diagnostic device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" is established through standardized physical, chemical, and biological testing methods as defined by:

    • ASTM D6319-19 (Standard Specification for Nitrile Examination Gloves)
    • ASTM D5151-19 (Standard Test Method for Detection of Holes in Medical Gloves)
    • ASTM D6124-06(2017) (Standard Test Method for Residual Powder on Medical Gloves)
    • ASTM D6978-05(2019) (Standard Practice for Assessment of Resistance of Medical Gloves to Permeation by Chemotherapy Drugs)
    • ISO 10993-10:2010 (Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization)
    • ISO 10993-11:2017 (Biological evaluation of medical devices - Part 11: Tests for systemic toxicity)

    8. The sample size for the training set

    Not applicable. This is not a machine learning device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is not a machine learning device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220240

    Validate with FDA (Live)

    Date Cleared
    2022-03-31

    (62 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Vinyl Co-Polymer Powder-free Examination Gloves, Black is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner.

    Device Description

    Vinyl Co-Polymer Powder-free Examination Gloves, Black are Patient Examination Gloves, Disposable, single use only and non-sterile. The gloves are made of vinyl and oil-based liquid nitrile rubber materials and are powder free. The physical and performance characteristics of the devices meet all requirements of ASTM D5250-19 Standard Specification for Vinyl Examination Gloves for Medical Application.

    AI/ML Overview

    The provided text is related to the FDA 510(k) premarket notification for "Vinyl Co-Polymer Powder-free Examination Gloves, Black". This document describes the device's technical specifications and how it compares to a predicate device. It also includes non-clinical performance data to demonstrate that the device meets established standards.

    However, the request asks for specific information regarding acceptance criteria and a study proving a medical device meets those criteria, particularly in the context of AI/ML-assisted diagnostic tools (implied by questions about MRMC studies, human readers, ground truth establishment for training/test sets, etc.). The provided document is for examination gloves, which are Class I medical devices and do not involve AI, image analysis, or diagnostic capabilities in the way the questions are framed. Therefore, most of the requested information is not applicable to this submission.

    I will fill in the relevant information that is present in the document and explicitly state when the requested information is not applicable due to the nature of the device.

    Here's the breakdown based on the provided document:


    1. A table of acceptance criteria and the reported device performance

    Test MethodPurposeAcceptance CriteriaReported Device Performance (Results)
    ASTM D5250-19LengthMinimum 230mmAll size ≥230mm
    ASTM D5250-19Palm Width (XS)75 ± 5 mm76-78mm
    ASTM D5250-19Palm Width (S)85 ± 5 mm86-88mm
    ASTM D5250-19Palm Width (M)95 ± 5 mm96-98mm
    ASTM D5250-19Palm Width (L)105 ± 5 mm106-108mm
    ASTM D5250-19Palm Width (XL)115 ± 5 mm116-118mm
    ASTM D5250-19Thickness (Finger)0.08mm (min)0.08mm
    ASTM D5250-19Thickness (Palm)0.08mm (min)0.08-0.10mm
    ASTM D5250-19, ASTM D412-16Tensile Strength, Before Aging11MPa, min13.5-17.8 MPa
    ASTM D5250-19, ASTM D412-16Tensile Strength, After Accelerated Aging11MPa, min13.1-17.6 MPa
    ASTM D5250-19, ASTM D412-16Ultimate Elongation, Before Aging300%, min320-430%
    ASTM D5250-19, ASTM D412-16Ultimate Elongation, After Accelerated Aging300%, min310-420%
    ASTM D5151-19, ASTM D65250-19Freedom from holesG-I, AQL 2.5Meets AQL 2.5 requirements
    ASTM D6124-06(2017), ASTM D5250-19Powder-Content≤2 mg per glove0.35-0.68 mg
    10993-10:2010Skin Irritation StudyUnder the conditions of the study, not an irritantUnder the conditions of the study, not an irritant
    10993-10:2010Maximization Sensitization StudyUnder the conditions of the study, not a sensitizerUnder the conditions of the study, not a sensitizer
    10993-5:2009In Vitro Cytotoxicity TestNo potential cytotoxicity to L929 cellsNo potential cytotoxicity to L929 cells
    ISO 10993 Part 11Acute Systemic Toxicity TestNo systemic toxicityNo systemic toxicity

    2. Sample sizes used for the test set and the data provenance

    The document refers to "non-clinical tests" and "performance test data" but does not specify the exact sample sizes (e.g., number of gloves tested for each individual parameter). The tests conducted are standard physical and biocompatibility tests for examination gloves as per ASTM and ISO standards.

    • Sample Size: Not explicitly stated for each test, but implied to be sufficient to meet the specifications of the referenced ASTM and ISO standards.
    • Data Provenance: The manufacturing entity is Grand Work Plastic Products Co., Ltd in Donggao Industrial Zone, Zanhuang, Hebei, China. The testing was presumably conducted by or for this manufacturer. The document doesn't specify if the data is retrospective or prospective, but for device development and regulatory submission, it would reflect tests performed on batches of the device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This device is an examination glove, not an AI/ML-based diagnostic system that requires expert interpretation for ground truth establishment. The "ground truth" for these tests is based on objective physical, chemical, and biological measurements against established international standards (ASTM, ISO).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not an AI/ML diagnostic system requiring expert adjudication of results. Testing is based on laboratory measurements and adherence to specified tolerances.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an examination glove and does not involve human readers, AI assistance, or diagnostic interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is an examination glove and does not involve algorithms or standalone performance testing in the context of AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device is based on:

    • Physical and Mechanical Standards: Measurements of length, palm width, thickness, tensile strength, and elongation against specifications in ASTM D5250-19 and ASTM D412-16.
    • Performance Standards: Freedom from holes (ASTM D5151-19, ASTM D65250-19) and powder content (ASTM D6124-06(2017), ASTM D5250-19).
    • Biocompatibility Standards: Results from tests for skin irritation, sensitization (ISO 10993-10:2010), in vitro cytotoxicity (ISO 10993-5:2009), and acute systemic toxicity (ISO 10993 Part 11).

    8. The sample size for the training set

    Not applicable. This device is an examination glove and does not involve AI/ML models that require training sets.

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set for an AI/ML model, no ground truth establishment method for a training set is relevant.

    Ask a Question

    Ask a specific question about this device

    K Number
    K142409

    Validate with FDA (Live)

    Date Cleared
    2014-12-05

    (100 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is disposable non-sterile device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    Vinyl Co-Polymer Powder Free Examination Gloves, Blue Color are Patient Examination Gloves, Disposable, single use only and non-sterile. The gloves are made of vinyl and oil-based liquid nitrile rubber materials and are powder free. The physical and performance characteristics of the devices meet all requirements of ASTM D5250-06 (Reapproved 2011) Standard Specification For Poly (Vinyl Chloride) Gloves For Medical Application.

    AI/ML Overview

    The provided document describes the Grand Work Plastic Products Co., Ltd.'s Vinyl Co-Polymer Powder Free Examination Gloves, Blue Color (K142409) and seeks to prove its substantial equivalence to a predicate device (K061562). The study conducted is a non-clinical performance study comparing the subject device against the predicate device based on established standards and FDA requirements.

    Here's the breakdown of the acceptance criteria and study details:

    1. A table of acceptance criteria and the reported device performance:

    CharacteristicAcceptance Criteria (Stated Standard ASTM D5250-06(2011) or other)Reported Device Performance (Subject Device K142409)Result of Comparison to Predicate & Acceptance
    MaterialVinyl and oil-based liquid Nitrile compound (plasticizer type different)Vinyl and oil-based liquid Nitrile compound (plasticizer type different)Similar
    ColorBlueBlueSimilar
    Tensile Strength (Mpa)
    - Before aging11 Mpa min15.2 Mpa minPass (Exceeds standard and similar to predicate)
    - After aging11 Mpa min14.6 Mpa minPass (Exceeds standard and similar to predicate)
    Ultimate Elongation (%)
    - Before aging300% min390% minPass (Exceeds standard and similar to predicate)
    - After aging300% min360% minPass (Exceeds standard and similar to predicate)
    Freedom from PinholesAQL 2.5 (G-I)AQL 1.5 (G-I)Pass (Exceeds standard and similar to predicate, lower AQL is better)
    Overall Length (mm)230 for all sizes min (S-2, AQL4.0)239 mm (for all sizes)Pass (Exceeds standard and similar to predicate)
    Width (mm)S: 85±5, M: 95±5, L: 105±5, XL: 115±5 (S-2, AQL4.0)S: 87, M: 97, L: 107, XL: 118Pass (Within range and similar to predicate)
    Palm Thickness (mm)0.08 mm min (S-2, AQL4.0)0.08 mmPass (Meets standard and similar to predicate)
    Finger Thickness (mm)0.05 mm min (S-2, AQL4.0)0.10 mmPass (Exceeds standard and similar to predicate)
    Residual PowderNot more than 2mg per glove (N=5)Not more than 2mg per glovePass (Meets standard and similar to predicate)
    Primary Skin Irritation TestUnder conditions of the study, not an irritant (ISO 10993-10)Under conditions of the study, not an irritantSimilar (Pass)
    Dermal Sensitization AssayUnder conditions of the study, not an irritant (ISO 10993-10)Under conditions of the study, not an irritantSimilar (Pass)

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Sample Size for Test Set: The sample sizes for each test are indicated by the "Inspection Level and AQL" or "N" values.
      • For physical dimensions (Overall Length, Width, Palm Thickness, Finger Thickness), Tensile Strength, and Ultimate Elongation: S-2, AQL4.0.
      • For Pinhole: G-I, AQL1.5.
      • For Residual Powder: N=5 (5 gloves tested).
      • For Biocompatibility (Primary Skin Irritation, Dermal Sensitization): The specific number of samples (e.g., animals, human subjects, in-vitro replicates) is not explicitly stated, but the tests were performed "Under conditions of the study".
    • Data Provenance: The study was conducted by Grand Work Plastic Products Co., Ltd., which is located in Zanhuang, Hebei, China. The data originates from this manufacturer. It is a prospective study as the tests were performed on the subject device to demonstrate compliance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    This study is a physical and chemical performance test of a medical device (examination gloves), not an AI/clinical diagnostic device. Therefore, it does not involve experts establishing "ground truth" in the clinical sense (e.g., radiologists, pathologists). The "ground truth" is established by the pre-defined technical standards (e.g., ASTM D5250-06(2011), ISO 10993-10) and FDA requirements. The tests are performed by technicians in a laboratory setting according to specified protocols.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable. This is not a study involving human readers or clinical adjudication. The results are quantitative measurements or qualitative observations against a set standard.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This is a performance study for a physical device (examination gloves), not a comparative effectiveness study involving human readers or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is a performance study for a physical device (examination gloves), not an algorithm or AI system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    The "ground truth" for this study is based on established industry standards and regulatory requirements. Specifically:

    • ASTM D5250-06 (Reapproved 2011) Standard Specification For Poly (Vinyl Chloride) Gloves For Medical Application.
    • ISO 10993-10 for biocompatibility (Primary Skin Irritation and Dermal Sensitization).
    • FDA requirements (e.g., for pinhole AQL).

    8. The sample size for the training set:

    Not applicable. This is not a machine learning or AI study; there is no training set involved.

    9. How the ground truth for the training set was established:

    Not applicable. There is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K102089

    Validate with FDA (Live)

    Date Cleared
    2010-10-06

    (62 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A patient examination glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

    Device Description

    Classified by FDA's General Hospital and Personal Use Device panel as Class I, 21 CFR 880.6250, Power Free Vinyl Patient Examination Glove, 80 LYZ, and meets all requirements of ASTM standard D-5250-06e1.

    AI/ML Overview

    This is not a medical device that utilizes AI/ML. The provided text is a 510(k) summary for "Powder Free Vinyl Patient Examination Gloves," which is a Class I medical device. Therefore, information related to AI/ML specific criteria such as multi-reader multi-case studies, effect size of human readers with AI assistance, standalone algorithm performance, or training set details is not applicable and cannot be extracted from this document.

    However, I can provide the acceptance criteria and the studies that prove the device meets these criteria based on the provided text.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific CriteriaReported Device PerformanceStudy Reference Section
    Physical and DimensionsConformance to ASTM D-5250-06e1"All testing meets requirements for physical and dimensions testing conducted on gloves."Section 7
    Freedom from Pinhole DefectsFDA 1000 ml. Water Fill Test requirements (AQL 2.5, inspection level I)"The FDA 1000 ml. Water Fill Test was also conducted with samplings of AQL 2.5, inspection level I, meeting these requirements."Section 7
    Biocompatibility - Primary Skin IrritationNo primary skin irritant reactions"Primary Skin Irritation...testing was conducted with results showing no primary skin irritant...reactions."Section 7
    Biocompatibility - Skin SensitizationNo sensitization reactions (allergic contact dermatitis)"...Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no...sensitization reactions."Section 7
    Residual PowderNo more than 2 mg powder per glove (based on ASTM D-6124-06)"...our gloves meet our 'powder-free' claims (contain no more than 2 mg powder per glove)."Section 7
    General EquivalenceAs safe, as effective, and performs as well as or better than the predicate device (Better Care Plastic Products Co., Ltd Powder Free Vinyl Patient Examination Glove)"Grand Work Plastic Products Co., Ltd.'s Powder Free Vinyl Patient Examination Gloves conform fully to ASTM D-5250-06e1 standard as well as applicable 21 CFR references, and meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the 'substantial equivalence' products cited. The device herein mentioned is as safe, as effective, and performs as well as or better than the predicate device."Section 10

    2. Sample size used for the test set and the data provenance

    • Physical and Dimensions Testing: "Inspection level S-2, AQL 4.0." (Specific numerical sample size not provided, but AQL and inspection level are specified which determine the sample size. The data provenance is internal testing by the manufacturer.)
    • FDA 1000 ml. Water Fill Test: "samplings of AQL 2.5, inspection level I." (Specific numerical sample size not provided, but AQL and inspection level are specified. The data provenance is internal testing by the manufacturer.)
    • Primary Skin Irritation and Skin Sensitization: The document does not specify the sample size for human or animal subjects used in these tests. The testing was conducted, implying it's a prospective study for the purpose of this submission.
    • Residual Powder Test: The document does not specify the sample size. The data provenance is internal testing by the manufacturer.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This device is a physical product (examination gloves), and the "ground truth" is established through standardized physical, chemical, and biological testing rather than expert-based interpretations.

    4. Adjudication method for the test set

    Not applicable. The testing methods for physical properties, pinholes, and biocompatibility are objective measurements against established standards, not subjective interpretations requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/ML powered device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm-based device.

    7. The type of ground truth used

    The "ground truth" for the device's performance is based on objective measurements against established national and international standards for medical gloves:

    • ASTM D-5250-06e1 (Standard Specification for Vinyl Patient Examination Gloves) for physical and dimensional properties.
    • FDA 1000 ml. Water Fill Test for barrier integrity (pinholes).
    • Primary Skin Irritation and Skin Sensitization testing for biocompatibility.
    • ASTM D-6124-06 for Residual Powder.

    8. The sample size for the training set

    Not applicable. There is no AI/ML algorithm or training set for this type of device. The "training" for manufacturing would involve adherence to quality control processes and standards.

    9. How the ground truth for the training set was established

    Not applicable. As there is no AI/ML algorithm or training set, this question is not relevant.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 2