510(k) Data Aggregation

    K Number
    K251784

    Validate with FDA (Live)

    Device Name
    Hushd Pro Z-Link
    Date Cleared
    2025-10-24

    (135 days)

    Product Code
    Regulation Number
    872.5570
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hushd Pro Z-Link device is intended to reduce or alleviate snoring and mild to moderate obstructive sleep apnea (OSA) in adults.

    Device Description

    The Hushd Pro Z-Link device series are removable intraoral patient-specific device intended for the treatment of snoring and mild to moderate obstructive sleep apnea. The device consists of custom-made maxillary and mandibular splints that fit over the upper and lower teeth, and are connected via a pair of detachable rigid connectors. When linked, the splints function as a mandibular repositioner by holding the lower jaw in a forward position during sleep, thereby preventing the tongue and soft tissues from collapsing into the airway.

    Each device consists of one maxillary and one mandibular splint, designed and manufactured using 3D printing technology. The splints incorporate a pair of integrated studs that are located around the canine area for the maxillary splint and around the first molar area for the mandibular splint. These studs interface with slots on the detachable connectors to securely link the two splints together. This configuration permits the mandible to advance from the baseline position, offering improved patient comfort.

    Titration is achieved by replacing the connectors with another pair of a different size. The connectors are available in 1.0 mm incremental lengths, allowing progressive mandibular advancement according to patient tolerance and clinical need. A standard set includes six pairs of connectors, ranging from -1 mm to +5 mm protrusion, while an extended set ranging from +6 mm to +10 mm is available upon request, providing a maximum mandibular protrusion of 10 mm.

    The connectors are not elastic and do not contain adjustment mechanisms such as pistons, screws, straps, or repositioning elastics. The rigid connectors are manufactured with medical grade nylon PA12 using commercial injection molding process and are marked for easy identification.

    The splints of this device are manufactured using selective laser sintering (SLS) 3D printing with EOS PA2200 polyamide (nylon) PA12 material. This material offers excellent mechanical properties, long-term stability, and has a history of safe use in oral medical devices. The manufacturing process includes mechanical or chemical polishing, cleaning, and rigorous quality control prior to packaging.

    The Hushd Pro Z-Link is available for prescription only and designed based on optical or physical impressions taken by a licensed healthcare provider. The splints and connectors are designed exclusively by the device manufacturer and its authorized partners to ensure consistent quality and performance.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K232025

    Validate with FDA (Live)

    Device Name
    Hushd Pro Avera
    Date Cleared
    2024-05-09

    (307 days)

    Product Code
    Regulation Number
    872.5570
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hushd Pro Avera device is intended to reduce or alleviate snoring and mild to moderate obstructive sleep apnea (OSA) in adults.

    Device Description

    The Hushd Pro Avera device series are a set of removable intraoral patient specific device used for treating snoring and mild to moderate obstructive sleep apnea. The device consists of number of custom made maxillary and mandibular splints fit over the upper and lower teeth. when interfaced together, functions as a mandibular repositioner holding the mandible forward during sleep thus preventing the togue and soft tissues of the throat from collapsing into the airway.

    The Hushd Pro Avera device series are patient-specific devices, consisting of one or more maxillary device(s) and one or more mandibular device(s) that mates together. The sequential devices are CAD/CAM designed and manufactured according to the requested mandibular advancements position prescribed by the dentist or physician to provide a selection of gentle advancements according to patient comfort and need. As such, prescribed advancements can be achieved by simply removing current maxillary or mandibular device and inserting the next maxillary or mandibular device in the advancements series. The maximum protrusion of the device is 10mm.The Hushd Pro Avera device does not have any adjustment mechanisms to modify or maintain the mandibular position such as pistons, straps or repositioning elastics.

    AI/ML Overview

    Based on the provided text, the device in question, Hushd Pro Avera, is an intraoral device for snoring and mild to moderate obstructive sleep apnea. The text does not describe acceptance criteria or a study that proves the device meets those criteria from a clinical performance perspective. Instead, it explicitly states that no clinical performance data was deemed necessary or performed for this submission.

    The submission relies on a demonstration of substantial equivalence to predicate and reference devices, supported by non-clinical performance data (material properties and biocompatibility).

    Therefore, I cannot provide a table of acceptance criteria and reported device performance from a clinical study, nor specific details about sample size, expert ground truth establishment, or MRMC studies, as these types of studies were not conducted or reported in this 510(k) summary.

    However, I can extract the information provided regarding the non-clinical testing and the rationale for not performing clinical studies.


    Summary regarding Acceptance Criteria and Performance (based on provided text):

    The provided 510(k) summary for the Hushd Pro Avera device focuses on demonstrating substantial equivalence to legally marketed predicate devices, rather than establishing clinical performance against pre-defined acceptance criteria through a clinical study. The FDA's clearance (K232025) is based on this equivalence, along with successful non-clinical performance data and biocompatibility testing.

    Here's an analysis of the requested information based on the document:

    1. A table of acceptance criteria and the reported device performance

      • Clinical Performance: No clinical performance acceptance criteria or reported clinical performance data is provided in the document. The submission explicitly states: "The clinical performance of the subject device is deemed not necessary. The Hushd Pro Avera device has equivalent indication and method of use to its predicated and reference devices, therefore there was no clinical testing to support this device."
      • Non-Clinical Performance: The acceptance criteria for non-clinical testing are implied by the successful completion of the tests and the conclusion that the device "meets all design specification." The reported performance is that the tests were conducted and the device passed.
      Acceptance Criteria (Non-Clinical)Reported Device Performance (Non-Clinical)
      Compliance with ISO 14971:2007 (Risk Analysis)All identified risks have been addressed through device design, biocompatibility, and bench testing or through labeling provided to the consumer.
      Material Mechanical PropertiesTests conducted by the manufacturer for Flexural Modulus, Flexural Strength, and Impact Strength. (Details of specific values or thresholds are not provided, only that tests were done to "evaluate its ability to achieve its intended use.")
      Biocompatibility (ISO 10993-1)Final finished device tested for: Cytotoxicity, Irritation, Sensitization, System Toxicity, Pyrogen. (Specific results are not provided, but the overall statement implies successful completion: "The subject device is able to achieve its intended use based on the same materials of construction, similar manufactory process and the identical application to the predicate device.")
      Design SpecificationsThe device "meets all design specification". (Specific design specifications are not enumerated.)
    2. Sample sizes used for the test set and the data provenance

      • Clinical Test Set: Not applicable, as no clinical test set was used for performance evaluation (no clinical study was conducted).
      • Non-Clinical Test Set: The sample sizes for the non-clinical materials testing and biocompatibility testing are not specified in the document. However, these are generally standard tests conducted on representative samples of the device and its materials. The provenance is internal testing by the manufacturer.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

      • Not applicable, as no clinical test set with human-established ground truth was used for performance evaluation.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      • Not applicable, as no clinical test set requiring adjudication was used.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      • Not applicable, as this device is a physical intraoral device, not an AI/software product, and no clinical studies (MRMC or otherwise) were conducted.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

      • Not applicable, as this is a physical device, not an algorithm.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

      • Clinical Ground Truth: Not applicable, as no clinical studies were performed.
      • Non-Clinical Ground Truth: For mechanical properties and biocompatibility, the ground truth is established by validated laboratory testing methods and compliance with international standards (ISO 10993-1, ISO 14971:2007).
    8. The sample size for the training set

      • Not applicable, as this is a physical device, not a machine learning model requiring a training set.
    9. How the ground truth for the training set was established

      • Not applicable, as there was no training set.

    In conclusion, the core of this 510(k) submission and FDA clearance is based on the principle of substantial equivalence to existing devices, supported by non-clinical testing, rather than new clinical performance data or studies defining and meeting specific clinical acceptance criteria.