Search Results
Found 15 results
510(k) Data Aggregation
(256 days)
The Gold Standard Diagnostics (GSD) AIX1000 Rapid Plasma Reagin (RPR) Automated Test System is a non-treponemal flocculation test that can qualitatively determine the presence of reagin antibodies in human serum, automated on the AIX1000 Analyzer. It may be used to aid in the diagnosis of syphilis when used in conjunction with supplemental treponemal laboratory tests and other clinical information. This test may also be used to detect non-treponemal antibodies in samples serially diluted to establish titer information. This test is not intended for screening blood or tissue donors.
The Gold Standard Diagnostics (GSD) AIX1000 Rapid Plasma Reagin (RPR) Automated Test System in a non-treponemal test for the qualitative determination of reagin antibodies in human serum to aid in the diagnosis of syphilis. This test is also used to detect non-treponemal antibodies in samples serially diluted to establish titer information. The system consists of the Gold Standard Diagnostics RPR test reagents and the Gold Standard Diagnostics AIX1000 Agglutination Analyzer. The AIX1000 Analyzer delivers serum from collection tubes into test wells. After antigen suspension is added, the test wells are then incubated while being shaken. An onboard camera creates a high resolution image. The image is then analyzed by the proprietary software algorithm to produce a result.
The RPR test reagents consist of a reactive control, a non-reactive control, and the antigen suspended in a carbon solution. When the antigen is mixed with sera, if antibodies are present, they will bind to the antigen and form black flocculants due to the presence of carbon particles. If no antibodies are present, then the carbon particles remain evenly distributed.
The antigen used in the Gold Standard Diagnostics AIX1000 Rapid Plasma Reagin (RPR) Automated Test System is a modified VDRL carbon antigen. The formulation is the same as that established by the Center for Disease Control (CDC)¹ containing 0.03% cardiolipin, 0.9% cholesterol, and 0.21% lecithin.
The kit also includes untreated, 48 well reaction plates a reactive control and a non-reactive control.
N/A
Ask a specific question about this device
(133 days)
The Gold Standard Diagnostics Borrelia burgdorferi VISE-OspC IgG/IgM ELISA Test Kit is intended as a qualitative test for the detection of IgG and IgM class antibodies to VIsE and OspC antigens from Borrelia burgdorferi sensu stricto in human serum from symptomatic patients or people suspected of having Lyme disease. When used as the first-tier screening test, positive and equivocal results must be confirmed through additional testing by one of the following methods:
• Standard two-tier test methodology (STTT) using an IgG and/or IgM blot testing following current interpretation guidelines. OR
• Modified two-tier test methodology (MTTT) using one or more of the following three ELISA based assays:
a. Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test
b. Gold Standard Diagnostics Borrelia burgdorferi IgG ELISA Test
c. Gold Standard Diagnostics Borrelia burgdorferi IgM ELISA Test
The assay can also be used as a second-tier confirmation test using the MTTT methodology when used with one or more of the following three ELISA based assays:
a. Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test
b. Gold Standard Diagnostics Borrelia burgdorferi IgG ELISA Test
c. Gold Standard Diagnostics Borrelia burgdorferi IgM ELISA Test
Positive test results by either the STTT or MTTT methodology are supportive evidence for the presence of antibodies and exposure to Borrelia burgdorferi, the cause of Lyme disease. A diagnosis of Lyme disease should be made based on the presence of Borrelia burgdorferi antibodies. history, symptoms, and other laboratory findings.
Not Found
This document is a 510(k) clearance letter for the Gold Standard Diagnostics Borrelia burgdorferi VlsE-OspC IgG/IgM ELISA Test Kit. It states that the device is substantially equivalent to legally marketed predicate devices. However, this document does not contain the acceptance criteria for the device, nor does it provide details of a study proving the device meets acceptance criteria, or information regarding sample sizes, ground truth establishment, or expert involvement.
The letter explicitly states: "We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976..."
Therefore, I cannot extract the requested information from the provided text. To answer your questions, the actual 510(k) summary or the full submission (which is usually publicly available on the FDA website) would be required.
Ask a specific question about this device
(133 days)
The Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test Kit is intended as a qualitative test for the detection of IgG and IgM antibodies to Borrelia burgdorferi sensu stricto in human serum from symptomatic patients or people suspected of infection. When used as the first-tier screening test, positive and equivocal results must be confirmed through additional testing by one of the following methods:
• Standard two-tier test methodology (STTT) using an IgG and/or IgM blot testing following current interpretation guidelines, OR
• Modified two-tier test methodology (MTTT) using the Gold Standard Diagnostics Borrelia burgdorferi VlsE-OspC IgG/ IgM ELISA Test.
The assay can also be used as a second-tier confirmation test using the MTTT methodology when used with the Gold Standard Diagnostics Borrelia burgdorferi VlsE-OspC IgG/IgM ELISA Test as the first-tier screening test.
Positive test results by either the STTT or MTTT methodology are supportive evidence for the presence of antibodies and exposure to Borrelia burgdorferi, the cause of Lyme disease. A diagnosis of Lyme disease should be made based on the presence of Borrelia burgdorferi antibodies, history, symptoms, and other laboratory findings.
The kit includes 12 x 8 well Antigen Coated strips, Conjugate, Substrate, Stop Solution, Wash Buffer, Diluent, Negative Control, Positive Control, and Cutoff Control. The controls are provided to determine if the assay is functioning properly and to determine the antibody level. The reagents are sufficient for 96 determinations.
During the test procedure, antibodies to B. burgdorferi (sensu stricto) if present in the human serum sample will bind to the antigens coated onto the wells forming antigen-antibody complexes. Excess antibodies are removed by washing. A conjugate of goat anti-human IgG/IgM antibodies conjugated with horseradish peroxidase are then added, which binds to the antigen-antibody complexes. Excess conjugate is removed by washing. This is followed by the addition of a chromogenic substrate, tetramethylbenzidine (TMB). If specific antibodies to the antigen are present in the patients' serum, a blue color will develop. The enzymatic reaction is then stopped with a stopping solution causing the contents of the well to turn yellow. The wells are read photometrically with a microplate reader at 450nm.
The antigens used in the Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test kit is a combination of B. burgdorferi sensu stricto strain B31 lysate, B. burgdorferi sensu stricto strain 2591 lysate, and a recombinant VlsE from B. burgdorferi sensu stricto strain B31. The lysates use spirochetes growing in BSK-H complete medium until mid-exponential phase. The recombinant VlsE protein is produced in E. coli SURE2 cells and purified by affinity chromatography. The purity of each antigen is assayed by SDS-PAGE followed by Coomassie staining and/or Western blotting.
The provided document describes the Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test Kit and its performance in various studies. Here's a breakdown of the acceptance criteria and study details:
1. Table of Acceptance Criteria and Reported Device Performance (STTT comparison):
The document does not explicitly state pre-defined acceptance criteria for the comparative studies. Instead, it presents observed performance metrics (Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA)) against a predicate device. For the purpose of this response, I will highlight the reported performance metrics from the comparative study.
| Metric | Reported Device Performance (STTT) | Basis of Comparison |
|---|---|---|
| Positive Percent Agreement (PPA) | 96.0% (72/75) with 95% CI (88.8% - 99.2%) | Predicate IgG/IgM ELISA |
| Negative Percent Agreement (NPA) | 97.5% (434/445) with 95% CI (95.6% - 98.8%) | Predicate IgG/IgM ELISA |
1. Table of Acceptance Criteria and Reported Device Performance (MTTT Comparison):
Similar to the STTT comparison, the document reports observed performance metrics for the MTTT protocol against the STTT predicate.
| Metric | Reported Device Performance (MTTT) (Second-tier test of MTTT vs. STTT for first-tier positive/equivocal samples) | Reported Device Performance (MTTT) (MTTT vs. STTT for all prospective study samples) | Basis of Comparison |
|---|---|---|---|
| Positive Percent Agreement (PPA) | 100.0% (95% CI (90.3% - 100.0%)) | 100.0% (95% CI (90.3% - 100.0%)) | Predicate Immunoblots (STTT) |
| Negative Percent Agreement (NPA) | 27.8% (95% CI (9.7% - 53.5%)) | 97.1% (95% CI (95.1% - 98.4%)) | Predicate Immunoblots (STTT) |
2. Sample Size Used for the Test Set and Data Provenance:
For Comparison with Predicate Device (STTT):
- Sample Size: 520 serum samples.
- Data Provenance: Prospective samples submitted for Lyme serology testing. Collected from three sites (one internal and two external reference laboratories).
For MTTT Comparison (Prospective Study):
- Sample Size: 481 serum samples for the initial screening. Of these, 54 positive or equivocal samples were further tested.
- Data Provenance: Prospective samples submitted for Lyme serology testing. Collected from three sites (one internal and two external reference laboratories).
For Sensitivity Study (STTT and MTTT):
- Sample Size: 89 clinically characterized samples for STTT and 125 clinically characterized samples for MTTT.
- Data Provenance: Clinically characterized samples encompassing early, disseminated, and late stages of Lyme disease. The document doesn't specify geographical origin but implies a clinical context.
For CDC Panel (STTT and MTTT):
- Sample Size: 40 samples for STTT and 280 samples for MTTT.
- Data Provenance: Acquired from the Centers for Disease Control (CDC).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:
The document does not provide specific details on the number or qualifications of experts used to establish the ground truth for the test sets. The ground truth appears to be based on:
- "Clinically characterized samples" (for sensitivity studies).
- "Patients diagnosed with Lyme disease" and "healthy individuals" from the CDC panel.
- Comparison with predicate devices (for comparative studies).
4. Adjudication Method for the Test Set:
The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for resolving discrepancies or establishing ground truth within the test sets. For comparative studies, the predicate device results largely serve as the reference, and for sensitivity studies, samples are clinically characterized, implying pre-existing diagnoses.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
This device is an in-vitro diagnostic (IVD) ELISA test kit, not an AI-powered image analysis or diagnostic tool that involves human readers. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and was not performed.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:
The device is an ELISA test kit designed to provide a qualitative result (positive, equivocal, negative) based on optical density readings. It functions as a standalone diagnostic assay without a human-in-the-loop component in its primary function. The interpretation of the results by laboratory personnel is part of standard lab practice, but the "standalone performance" in this context refers to the assay's accuracy in detecting antibodies. The reported PPA, NPA, and sensitivity studies reflect this standalone performance.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc):
The ground truth used for different studies varies:
- Comparison Studies with Predicate Device: The results of the legally marketed predicate devices (Trinity Biotech Captia™ Borrelia burgdorferi IgG/IgM ELISA Test Kit and Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM Blot Tests) served as the reference standard.
- Sensitivity Studies: "Clinically characterized samples" and "patients diagnosed with Lyme disease" from the CDC panel. This implies a combination of clinical assessment, symptomology, and potentially other laboratory findings, which can be seen as a form of expert consensus or aggregated clinical data.
- Analytical Specificity and Cross-Reactivity: For analytical specificity, asymptomatic individuals from endemic and non-endemic regions were used. For cross-reactivity, samples confirmed positive for specific disease markers (from serum vendors) were used.
8. The Sample Size for the Training Set:
The document describes the determination of the assay cutoff but does not explicitly mention a "training set" in the context of machine learning or AI. For the cutoff determination:
- Assay Cutoff: 200 normal sera (100 from endemic, 100 from non-endemic regions).
- Verification: 125 characterized Lyme disease samples were used with the 200 normal samples for ROC analysis to verify the chosen cutoff.
9. How the Ground Truth for the Training Set Was Established:
As noted above, there isn't a "training set" in the typical AI sense. However, for the determination of the assay cutoff:
- Normal Sera: These were identified as "normal" based on their origin from non-diseased individuals in endemic and non-endemic regions.
- Characterized Lyme Disease Samples: These 125 samples were "characterized" for Lyme disease, implying a pre-established clinical diagnosis or a well-defined status as positive Lyme disease samples.
Ask a specific question about this device
(133 days)
The Gold Standard Diagnostics Borrelia burgdorferi IgM ELISA Test Kit is intended as a qualitative test for the detection of IgM antibodies to B. burgdorferi sensu stricto in human serum from symptomatic patients or people suspected of infection. When used as the first-tier screening test, positive and equivocal results must be supplemented through additional testing by one of the following methods:
· Standard two-tier test methodology (STTT) using an IgM blot test following current interpretation guidelines, OR
• Modified two-tier test methodology (MTTT) using the Gold Standard Diagnostics Borrelia burgdorferi VlsE-OspC IgG/ IgM ELISA Test.
The assay can also be used as a second-tier confirmation test using the MTTT methodology when used with the Gold Standard Diagnostics Borrelia burgdorferi VlsE-OspC IgG/IgM ELISA Test as the first-tier screening test.
Positive test results by either the STTT or MTTT methodology are supportive evidence for the presence of antibodies and exposure to Borrelia burgdorferi, the cause of Lyme disease. A diagnosis of Lyme disease should be made based on the presence of Borrelia burgdorferi antibodies, history, symptoms, and other laboratory findings.
The kit includes 12 x 8 well Antigen Coated strips. Conjugate. Substrate, Stop Solution, Wash Buffer, Diluent, Negative Control, Positive Control, and Cutoff Control. The controls are provided to determine if the assay is functioning properly and to determine the antibody level. The reagents are sufficient for 96 determinations.
During the test procedure, antibodies to B. burgdorferi (sensu stricto) if present in the human serum sample will bind to the antigens coated onto the wells forming antigen-antibody complexes. Excess antibodies are removed by washing. A conjugate of goat anti-human IgM antibodies conjugated with horseradish peroxidase is then added, which binds to the antigenantibody complexes. Excess conjugate is removed by washing. This is followed by the addition of a chromogenic substrate, tetramethylbenzidine (TMB). If specific antibodies to the antigen are present in the patients' serum, a blue color will develop. The enzymatic reaction is then stopped with a stopping solution causing the contents of the well to turn yellow. The wells are read photometrically with a microplate reader at 450nm.
The antigens used in the Gold Standard Diagnostics Borrelia burgdorferi IgM ELISA Test kit is a combination of B. burgdorferi sensu stricto strain B31 lysate, B. burgdorferi sensu stricto strain 2591 lysate, and a recombinant VlsE from B. burgdorferi sensu stricto strain B31. The lysates use spirochetes growing in BSK-H complete medium until mid-exponential phase. The recombinant VlsE protein is produced in E. coli SURE2 cells and purified by affinity chromatography. The purity of each antigen is assayed by SDS-PAGE followed by Coomassie staining and/or Western blotting.
Here's an analysis of the provided text to extract the acceptance criteria and study details for the Gold Standard Diagnostics Borrelia burgdorferi IgM ELISA Test Kit.
It's important to note that this document describes an In Vitro Diagnostic (IVD) device, not an AI/ML-based medical device. Therefore, many of the typical acceptance criteria and study designs associated with AI/ML, such as MRMC studies, expert adjudication, separate training/test sets with established ground truth protocols, and specific ground truth types (pathology, outcomes data, expert consensus), are not applicable or described in the same manner.
Instead, the acceptance criteria for IVDs typically revolve around analytical performance (sensitivity, specificity, precision, reproducibility, cross-reactivity, interference) and clinical concordance with a predicate device or clinical diagnosis.
Acceptance Criteria and Device Performance (IVD Device)
1. Table of Acceptance Criteria and Reported Device Performance
Since this is an IVD device, not an AI/ML device, the acceptance criteria are generally based on meeting certain performance metrics (e.g., specificity, sensitivity, precision, agreement with a predicate device) that demonstrate its substantial equivalence to a legally marketed device. The document does not explicitly state numerical acceptance thresholds for each metric, but rather presents the performance observed and implies that these results were deemed acceptable for market clearance.
Here's a table summarizing the reported device performance, which implicitly represents the met acceptance criteria for this IVD:
| Metric (Implicit Acceptance Criteria) | Reported Device Performance (Gold Standard Diagnostics Borrelia burgdorferi IgM ELISA) | Context/Notes |
|---|---|---|
| Primary Comparison (STTT) | Comparison with predicate device Trinity Biotech MarDx Borrelia burgdorferi EIA IgM Test Kit (K894293) for Standard Two-Tier Test (STTT) methodology. | |
| Positive Percent Agreement (PPA) | 90.3% (93/103) with 95% CI (82.9% - 95.5%) | Against the Predicate IgM ELISA from prospective samples. |
| Negative Percent Agreement (NPA) | 99.6% (460/462) with 95% CI (98.5% - 99.9%) | Against the Predicate IgM ELISA from prospective samples. |
| Second-Tier PPA (against IgM blot) | 98.3% (58/59) with 95% CI (90.9% - 99.9%) | When Gold Standard Diagnostics Borrelia burgdorferi IgM ELISA Test positive/equivocal samples were re-tested with an FDA-cleared IgM blot assay (as the second-tier for STTT). |
| Clinical Sensitivity (STTT) | Early: 75.9% (44/58)Disseminated: 100.0% (17/17)Late: 89.7% (35/39) | Compared to predicate IgM ELISA in clinically characterized samples. |
| CDC Panel Agreement (STTT) | Healthy: 93.0%Early Lyme: 76.7%Cardiac Lyme: 66.7%Neurological Lyme: 100.0%Late: 85.0%Look-alike disease: 81.1% | Agreement with clinical diagnosis for CDC panel. |
| Analytical Specificity | Endemic Region: 96.1%Non-endemic Region: 90.5% | Determined by testing 208 asymptomatic individuals. |
| Precision (Within-lab) | CV ranges from 1.3% to 13.3% depending on sample and parameter (SD or CV) | For various control and sample types across 12 days, duplicates per day. Lowest CV (positive control, SD) 1.3%, highest CV (negative sample, SD) 13.3%. |
| Reproducibility (Multi-site) | CV ranges from 2.5% to 18.6% depending on sample and parameter (SD or CV) | For various control and sample types tested at 3 sites, triplicates, twice per day for 5 days. Lowest CV (moderate positive, between run) 2.5%, highest CV (negative sample, total) 18.6%. |
| Cross Reactivity | Varies by condition (e.g., Tick-borne Relapsing Fever: 0% positive, Ehrlichiosis: 60% positive) | Some cross-reactivity noted for certain conditions (Ehrlichiosis, Babesiosis, Leptospirosis, Varicella Zoster Virus). This is expected for serological assays and acknowledged by the need for second-tier testing. |
| Interfering Substances | None detected performance effect for tested substances (Albumin, Bilirubin, Cholesterol, Hemoglobin, Triglycerides) | Tested at recommended concentrations; no effect on performance. |
| Secondary Comparison (MTTT) | Comparison of Modified Two-Tier Test (MTTT) methodology (using this device + VlsE-OspC IgG/IgM ELISA) against Standard Two-Tier Test (STTT) using the predicate IgM blot. | |
| Positive Percent Agreement (PPA) | 100.0% with 95% CI (83.2% - 100.0%) | For overall MTTT vs. STTT results for all prospective samples. |
| Negative Percent Agreement (NPA) | 95.7% with 95% CI (93.4% - 97.3%) | For overall MTTT vs. STTT results for all prospective samples. |
| Clinical Sensitivity (MTTT) | Early: 62.9% (39/62)Disseminated: 100.0% (22/22)Late: 82.9% (34/41) | Compared to predicate STTT-IgM in clinically characterized samples. Notably higher for Late Lyme than predicate STTT-IgM (82.9% vs 22.0%). |
| CDC Panel Agreement (MTTT) | Healthy: 100.0%Early Lyme: 76.7%Cardiac Lyme: 66.7%Neurological Lyme: 100.0%Late: 80.0%Look-alike disease: 96.7% | Agreement with clinical diagnosis for CDC panel. Notably higher for Late Lyme than predicate STTT-IgM (80.0% vs 35.0%) and for Look-alike Disease (96.7% vs 97.8%). |
2. Sample Sampled Used for the Test Set and Data Provenance
- Determination of Assay Cutoff: 208 normal sera (103 from endemic region, 105 from non-endemic region). Provenance not explicitly stated but implies a mix of endemic/non-endemic populations, likely within the US.
- Precision (Test Set): A panel of 4 samples (negative, high negative, low positive, moderate positive) + kit controls. Tested in-house.
- Reproducibility (Test Set): A panel of 4 samples (negative, high negative, low positive, moderate positive) + kit controls. Tested at three different sites.
- Analytical Specificity (Test Set): 208 asymptomatic individuals' samples from endemic and non-endemic regions.
- Cross Reactivity (Test Set): 277 samples from serum vendors, confirmed positive for specific markers (e.g., Tick-borne Relapsing Fever IgM, Treponemal Infections, etc.).
- Interfering Substances (Test Set): 3 samples (high negative, equivocal, low positive) spiked with interferents.
- Clinical Studies (Comparison with Predicate Device - STTT):
- Prospective Samples: 531 serum samples.
- Provenance: Collected from three sites (one internal, two external reference laboratories) "using prospective samples submitted for Lyme serology testing". This implies real-world, prospectively collected, likely US-based clinical samples.
- Clinical Studies (Sensitivity Study - STTT): 114 clinically characterized samples (early, disseminated, late stages of Lyme disease). Provenance not explicitly stated.
- Clinical Studies (CDC Panel - STTT): 280 positive and negative specimens from the Center of Disease Control (CDC). These are a characterized reference panel. Provenance is CDC, likely multi-site or pooled.
- Clinical Studies (Method Comparison – MTTT IgM, Prospective Study):
- Prospective Samples: 481 serum samples for the initial screening. 68 positive and equivocal samples carried forward for second-tier testing.
- Provenance: Collected from three sites (one internal, two external reference laboratories) "using prospective samples submitted for Lyme serology testing". Implies real-world, prospectively collected, likely US-based clinical samples.
- Clinical Studies (Sensitivity Study - MTTT): 125 clinically characterized samples (early, disseminated, late stages of Lyme disease). Provenance not explicitly stated.
- Clinical Studies (CDC Reference Panel - MTTT): 280 positive and negative specimens from the Centers of Disease Control (CDC).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
For IVD devices like this one, ground truth is typically established by:
- Consensus of clinical diagnosis for disease stages (clinically characterized samples).
- Confirmed status from reference laboratories or biological repositories (e.g., CDC panel, serum vendors for cross-reactivity).
- Comparison to a legally marketed predicate device (as done for the main comparative studies).
The document does not specify a number of experts or their qualifications for establishing ground truth for the test set. Ground truth is derived from the established clinical status of the samples (e.g., clinically characterized samples, CDC panel, confirmed positive from serum vendors). This is standard for IVD submissions, where the "ground truth" often relies on a combination of patient history, symptoms, other laboratory findings, and established reference panels or predicate device results, rather than a panel of independent human readers interpreting medical images.
4. Adjudication Method for the Test Set
Not applicable in the usual sense for an IVD kit. There's no human "reader" adjudication described. The "adjudication" is inherent in the established clinical status of the samples used as ground truth or the comparison to a predicate device. For instances where samples were "clinically characterized," this implies a clinical diagnosis that served as the reference standard, rather than an adjudication process of human interpretations of imaging/data.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not Applicable. This is an IVD (laboratory diagnostic test kit), not an AI/ML-based medical imaging and interpretation device. MRMC studies are used for evaluating AI/ML systems that assist human readers in tasks like image interpretation.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
Yes, in essence. The performance metrics (PPA, NPA, sensitivity, specificity, etc.) presented in the tables are the standalone performance of the Gold Standard Diagnostics Borrelia burgdorferi IgM ELISA Test Kit itself, analogous to an "algorithm only" performance for an IVD. There is no "human-in-the-loop" component for the performance of this diagnostic test kit once it is run according to its instructions. The interpretation of the overall Lyme disease diagnosis, however, is intended to be made by a clinician based on multiple factors, as stated in the Indications for Use.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc)
The ground truth for this IVD was established using a combination of:
- Clinical Diagnosis / Clinically Characterized Samples: For the "Sensitivity Study" and "CDC Panel," samples were derived from patients with known clinical diagnoses of Lyme disease stages (early, disseminated, late) or from healthy individuals.
- Reference Panels: Specifically, the CDC Panel which consists of "positive and negative specimens... for Lyme disease detection" that are "masked characterized serum panel." This implies a very high confidence in the true status of these samples.
- Confirmed Status from Vendors: For the "Cross Reactivity" study, samples were "obtained from serum vendors who confirmed their positivity for each respective marker."
- Predicate Device/Established Methodology: For the "Comparison with Predicate Device" and "MTTT Comparison" studies, the performance was measured against results obtained from a legally marketed predicate device (Trinity Biotech MarDx Borrelia burgdorferi EIA IgM Test Kit) or an established "Standard two-tier test methodology (STTT)" using the predicate IgM blot test. While the predicate is not "ground truth" in the pathological sense, it serves as the reference standard for substantial equivalence studies for IVDs.
8. The Sample Size for the Training Set
Not applicable for this type of IVD device. This is a laboratory immunoassay kit, not an AI/ML algorithm that undergoes a distinct "training" phase with a large dataset. The "training" in the context of IVD development would be the R&D and optimization process to develop the assay components and parameters (e.g., antigen formulation, antibody concentrations, incubation times, cutoff determination), which are iterative and not typically reported as a "training set" size in an FDA submission. The determination of the assay cutoff involved 208 normal sera, which could be considered part of the assay's "calibration" or optimization, but not a "training set" in the sense of supervised machine learning.
9. How the Ground Truth for the Training Set was Established
Not applicable. As explained in point 8, there isn't a "training set" in the AI/ML context for this IVD device. The assay development and cutoff determination process is a different methodology. The "ground truth" for establishing the assay cutoff was based on 208 normal sera (from endemic and non-endemic regions) and a subsequent ROC analysis to optimize sensitivity and specificity.
Ask a specific question about this device
(133 days)
The Gold Standard Diagnostics Borrelia burgdorferi IgG ELISA Test Kit is intended as a qualitative test for the detection of IgG antibodies to B. burgdorferi sensu stricto in human serum from symptomatic patients or people suspected of infection. When used as the first-tier screening test, positive and equivocal results must be supplemented through additional testing by one of the following methods:
· Standard two-tier test methodology (STTT) using an IgG blot test following current interpretation guidelines, OR
• Modified two-tier test methodology (MTT) using the Gold Standard Diagnostics Borrelia burgdorferi VlsE-OspC IgG/ IgM ELISA Test.
The assay can also be used as a second-tier confirmation test using the MTTT methodology when used with the Gold Standard Diagnostics Borrelia burgdorferi VlsE-OspC IgG/IgM ELISA Test as the first-tier screening test.
Positive test results by either the STTT or MTTT methodology are supportive evidence for the presence of antibodies and exposure to Borrelia burgdorferi, the cause of Lyme disease. A diagnosis of Lyme disease should be made based on the presence of Borrelia burgdorferi antibodies, history, symptoms, and other laboratory findings.
The kit includes 12 x 8 well Antigen Coated strips. Conjugate. Substrate, Stop Solution, Wash Buffer, Diluent, Negative Control, Positive Control, and Cutoff Control. The controls are provided to determine if the assay is functioning properly and to determine the antibody level. The reagents are sufficient for 96 determinations.
During the test procedure, antibodies to B. burgdorferi (sensu stricto) if present in the human serum sample will bind to the antigens coated onto the wells forming antigen-antibody complexes. Excess antibodies are removed by washing. A conjugate of goat anti-human IgG antibodies conjugated with horseradish peroxidase is then added, which binds to the antigenantibody complexes. Excess conjugate is removed by washing. This is followed by the addition of a chromogenic substrate, tetramethylbenzidine (TMB). If specific antibodies to the antigen are present in the patients' serum, a blue color will develop. The enzymatic reaction is then stopped with a stopping solution causing the contents of the well to turn yellow. The wells are read photometrically with a microplate reader at 450nm.
The antigens used in the Gold Standard Diagnostics Borrelia burgdorferi IgG ELISA Test kit is a combination of B. burgdorferi sensu stricto strain B31 lysate, B. burgdorferi sensu stricto strain 2591 lysate, and a recombinant VlsE from B. burgdorferi sensu stricto strain B31. The lysates use spirochetes growing in BSK-H complete medium until mid-exponential phase. The recombinant VlsE protein is produced in E. coli SURE2 cells and purified by affinity chromatography. The purity of each antigen is assayed by SDS-PAGE followed by Coomassie staining and/or Western blotting.
This document describes the validation of the Gold Standard Diagnostics Borrelia burgdorferi IgG ELISA Test Kit, including a new proposed Modified Two-tier Testing (MTTT) methodology. The provided text primarily focuses on comparative studies against predicate devices and existing standard methods, rather than defining explicit acceptance criteria in a quantitative table format. However, implied acceptance criteria can be derived from the performance percentages presented.
Based on the provided information, here's a structured response addressing the requested points:
1. Table of Acceptance Criteria (Implied) and Reported Device Performance
The acceptance criteria are not explicitly stated as numerical cutoffs in the document. However, they can be inferred from the reported "Percent Agreement" values in the comparison studies. The studies aim to demonstrate substantial equivalence to established predicate devices and methodologies.
| Study Type / Performance Metric | Implied Acceptance Criteria (Goal: High Agreement/Sensitivity/Specificity) | Reported Device Performance (Gold Standard Diagnostics Borrelia burgdorferi IgG ELISA Test Kit) |
|---|---|---|
| Comparison with Predicate Device (STTT) | ||
| Positive Percent Agreement (PPA) with Predicate IgG ELISA | High (> 90% typically desired for equivalence) | 90.2% (55/61) [95% CI: 79.8% - 96.3%] |
| Negative Percent Agreement (NPA) with Predicate IgG ELISA | High (> 95% typically desired for equivalence) | 99.6% (460/462) [95% CI: 98.5% - 99.9%] |
| Second-Tier Testing (STTT) - Agreement with FDA Cleared IgG Blot | ||
| 2nd Tier PPA (Tier 1 Pos/Equiv samples) | 100.0% (as reported) | 100.0% (37/37) [95% CI: 92.2% - 100.0%] |
| Clinical Sensitivity (STTT) | ||
| Early Lyme | Improvement over predicate | 46.6% (27/58) vs. Predicate: 27.6% (16/58) |
| Disseminated Lyme | Improvement over predicate | 82.4% (14/17) vs. Predicate: 52.9% (9/17) |
| Late Lyme | High agreement with predicate | 97.4% (38/39) vs. Predicate: 97.4% (38/39) |
| CDC Panel (STTT) | ||
| Healthy | High (> 99% agreement, implying high specificity) | 99.0% (1/100 positive/equivocal) |
| Early Lyme | Improvement over predicate | 68.3% (41/60 positive/equivocal) vs. Predicate: 35.0% (21/60) |
| Cardiac Lyme | Comparable/Improvement | 66.7% (2/3 positive/equivocal) vs. Predicate: 100.0% (3/3) |
| Neurological Lyme | Improvement over predicate | 85.7% (6/7 positive/equivocal) vs. Predicate: 42.9% (3/7) |
| Late Lyme | High agreement with predicate | 100.0% (20/20 positive/equivocal) vs. Predicate: 100.0% (20/20) |
| Look-alike Disease | High (> 85% agreement, implying high specificity) | 87.8% (11/90 positive/equivocal) vs. Predicate: 88.9% (10/90) |
| Method Comparison MTTT – IgG (vs. STTT) | ||
| PPA (second-tier test of MTTT vs. STTT) | 100.0% (as reported) | 100.0% (23/23) [95% CI: 85.2% - 100.0%] |
| NPA (second-tier test of MTTT vs. STTT) | High (implied > 95%) | 0.0% (0/15) [95% CI: 0.0% - 21.8%] - Note: This NPA is for a subset of samples that were negative by IgG Immunoblot, not overall. The overall NPA is 96.7% |
| PPA (Overall MTTT vs. STTT) | 100.0% (as reported) | 100.0% (23/23) [95% CI: 85.2% - 100.0%] |
| NPA (Overall MTTT vs. STTT) | High (implied > 95%) | 96.7% (443/458) [95% CI: 94.7% - 98.2%] |
| Clinical Sensitivity (MTTT) | ||
| Early Lyme | Improvement over predicate STTT | 48.4% (30/62) vs. Predicate STTT: 8.1% (5/62) |
| Disseminated Lyme | Improvement over predicate STTT | 81.8% (18/22) vs. Predicate STTT: 22.7% (5/22) |
| Late Lyme | Comparable/Improvement | 97.6% (40/41) vs. Predicate STTT: 95.1% (39/41) |
| CDC Reference Panel (MTTT) | ||
| Healthy | 100.0% (as reported) | 100.0% (0/100 positive/equivocal) |
| Early Lyme | Improvement over predicate STTT | 60.0% (36/60 positive/equivocal) vs. Predicate STTT: 33.3% (12/60) |
| Cardiac Lyme | Improvement over predicate STTT | 66.7% (2/3 positive/equivocal) vs. Predicate STTT: 33.3% (1/3) |
| Neurological Lyme | Improvement over predicate STTT | 85.7% (6/7 positive/equivocal) vs. Predicate STTT: 14.3% (1/7) |
| Late Lyme | 100.0% (as reported) | 100.0% (20/20 positive/equivocal) vs. Predicate STTT: 100.0% (20/20) |
| Look-alike Disease | 100.0% (as reported) | 100.0% (0/90 positive/equivocal) vs. Predicate STTT: 100.0% (0/90) |
2. Sample Size Used for the Test Set and Data Provenance
-
Determination of Assay Cutoff (Nonclinical):
- Sample Size: 210 normal sera (105 endemic, 105 non-endemic) + 194 samples (114 Lyme disease stages, 8 healthy negative, 72 negative Lyme disease with other conditions). Total = 404 samples.
- Data Provenance: Not explicitly stated, but "endemic region of Lyme disease" and "non-endemic region of Lyme disease" suggest geographical diversity. Retrospective, as these are "tested" samples for cutoff determination.
-
Precision (Nonclinical):
- Sample Size: 48 runs for each of 4 panel members (negative, high negative, low positive, moderate positive) + kit controls. This is a repetitive testing design, not individual patient samples.
-
Reproducibility (Nonclinical):
- Sample Size: 90 runs (3 sites x 2 runs/day x 5 days x 3 replicates) for each of 4 panel members (negative, high negative, low positive, moderate positive) + kit controls (30 runs for Pos/Neg control, 60 runs for Cutoff control).
-
Analytical Specificity:
- Sample Size: 208 asymptomatic individuals (103 from endemic regions, 105 from non-endemic regions).
- Data Provenance: Endemic and non-endemic regions. Retrospective (asymptomatic individuals' samples).
-
Cross Reactivity:
- Sample Size: 377 samples.
- Data Provenance: Samples obtained from "serum vendors who confirmed their positivity for each respective marker," implying retrospective, controlled samples.
-
Comparison/Prospective Study (Clinical):
- STTT Comparison: 523 serum samples.
- MTTT Comparison: 481 serum samples for the initial screening. 38 positive/equivocal samples were carried forward for second-tier testing comparison.
- Data Provenance: "Prospective samples submitted for Lyme serology testing" from "three sites (one internal and two external reference laboratories)." This indicates a prospective study design with samples from potentially diverse geographical locations (clinical labs).
-
Sensitivity Study (Clinical):
- STTT: 114 clinically characterized samples.
- MTTT: 125 clinically characterized samples.
- Data Provenance: "Clinically characterized samples" from unspecified sources; likely retrospective collection of known patient cohorts.
-
CDC Panel (Clinical):
- Sample Size: 280 positive and negative specimens.
- Data Provenance: From the "Center of Disease Control (CDC)." These are reference panels, typically collected and characterized over time, so retrospective.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
The document does not explicitly state the number or qualifications of experts used to establish the ground truth for the test sets.
- For "clinically characterized samples" and "CDC panel" samples, the implication is that these samples have a pre-defined and reliable "clinical diagnosis" or "reference characterization" based on established medical criteria. This often involves consensus from clinical experts (e.g., infectious disease specialists) and/or a combination of clinical presentation, symptoms, and other laboratory findings, but the specifics are not detailed here.
- For the "Second-Tier Testing" comparison, the "FDA cleared IgG blot assay" and "predicate B. burgdorferi IgG blot test" serve as the ground truth. These are approved diagnostic methods, and their interpretation would follow established guidelines.
4. Adjudication Method for the Test Set
The document does not describe any specific adjudication method (e.g., 2+1, 3+1) for resolving discrepancies in the test results or for establishing the initial ground truth diagnoses. The comparisons are based on the results of the different assays and against established clinical characterizations or CDC panels.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC study was performed or described. This device is an ELISA test kit, a laboratory diagnostic assay for detecting antibodies, not an AI or imaging diagnostic tool that involves human readers interpreting images. Therefore, the concept of "human readers improving with AI assistance" is not applicable.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, the primary performance evaluation is standalone. The Gold Standard Diagnostics Borrelia burgdorferi IgG ELISA Test Kit operates as a standalone diagnostic assay. Its performance (sensitivity, specificity, agreement) is measured as the output of the kit itself (optical density readings converted to units and qualitative results) without direct human interpretation of complex visual patterns or AI assistance. The results are quantitative and then interpreted qualitatively (positive, equivocal, negative) based on predefined cutoffs.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
The ground truth used varies by study:
- Predicate Devices/Methodologies: For the STTT and MTTT comparison studies, the ground truth for the device's performance is a comparison against the results from legally marketed predicate devices (e.g., Trinity Biotech MarDx Borrelia burgdorferi EIA IgG Test Kit) and/or a FDA cleared IgG blot assay. This is a form of comparative ground truth against established methods.
- Clinical Diagnosis/Characterization: For the Sensitivity Study and CDC Panel, the ground truth is "clinically characterized samples" and "positive and negative specimens from the Centers of Disease Control (CDC)." This implies a clinical diagnosis or consensus diagnosis based on comprehensive patient data (history, symptoms, other laboratory findings) or a reference panel with established disease status.
- Analytical Ground Truth: For non-clinical studies like analytical specificity and cross-reactivity, the ground truth is often the confirmed absence or presence of specific conditions in the tested samples, usually based on prior laboratory testing or sample vendor claims.
8. The Sample Size for the Training Set
The document does not explicitly mention a "training set" in the context of machine learning. The term "training set" is usually associated with the development of AI algorithms. For an ELISA kit, development involves:
- Antigen Selection and Optimization: Which antigens to use (e.g., B31 lysate, 2591 lysate, recombinant VlsE).
- Reagent Formulation: Optimizing conjugate, substrate, buffers, etc.
- Cutoff Determination: "The cutoff was determined by testing a total of 210 normal sera..." and "An additional 194 samples..." This process of determining the cutoff could be considered analogous to a 'calibration' or 'training' phase for a traditional diagnostic test, where a dataset is used to define the diagnostic thresholds. In this sense, a total of 404 samples (210 normal + 194 other known samples) were used for cutoff determination.
9. How the Ground Truth for the Training Set Was Established
As discussed above, for establishing the cutoff (analogous to training/calibration):
- Normal Sera: The 210 normal sera were likely verified as "normal" through standard clinical practice or donor screening, indicating the absence of target antibodies/disease in a healthy population.
- Known Positive/Negative/Other Disease Samples: The additional 194 samples included "different phases of Lyme disease," "negative healthy samples," and "negative Lyme disease samples but do have other diseases that may cause serologic cross-reactivity." The "ground truth" for these samples would be their known clinical status or disease classification, typically established through comprehensive clinical evaluation, follow-up, and/or panels from disease banks. The ROC analysis was used to confirm the chosen cutoff provides the best compromise between sensitivity and specificity, indicating an analytical optimization process based on these known samples.
Ask a specific question about this device
(91 days)
The Gold Standard Diagnostics Borrelia burgdorferi IgM ELISA Test kit is intended as a qualitative presumptive (first step) test for the detection of IgM antibodies to B. burgdorferi sensu stricto in human serum from symptomatic patients or people suspected of infection. Positive and equivocal results must be supplemented by testing with a second-step Western blot assay.
The kit includes 12 x 8 well Antigen Coated strips, Conjugate, Substrate, Stop Solution, Wash Buffer, Diluent, Negative Control, Positive Control, and Cutoff Control. The controls are provided to determine if the assay is functioning properly and to determine the antibody level. The reagents are sufficient for 96 determinations.
During the test procedure, antibodies to B. burgdorferi (sensu stricto) if present in the human serum sample will bind to the antigens coated onto the wells forming antigen-antibody complexes. Excess antibodies are removed by washing. A conjugate of goat anti-human IgM antibodies conjugated with horseradish peroxidase is then added, which binds to the antigenantibody complexes. Excess conjugate is removed by washing. This is followed by the addition of a chromogenic substrate, tetramethylbenzidine (TMB). If specific antibodies to the antigen are present in the patients' serum, a blue color will develop. The enzymatic reaction is then stopped with a stopping solution causing the contents of the well to turn yellow. The wells are read photometrically with a microplate reader at 450mm.
The antigens used in the Gold Standard Diagnostics Borrelia burgdorferi IgM ELISA Test kit is a combination of B. burgdorferi sensu stricto strain B31 lysate, B. burgdorferi sensu stricto strain 2591 lysate, and a recombinant VlsE protein from B. burgdorferi sensu stricto strain B31. The lysates use spirochetes growing in BSK-H complete medium until mid-exponential phase. The recombinant VlsE protein is produced in E. coli SURE2 cells and purified by affinity chromatography.
The Gold Standard Diagnostics Borrelia burgdorferi IgM ELISA Test Kit is intended as a qualitative presumptive (first step) test for the detection of IgM antibodies to B. burgdorferi sensu stricto in human serum from symptomatic patients or people suspected of infection. Positive and equivocal results must be supplemented by testing with a second-step Western blot assay.
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
The provided document does not explicitly state acceptance criteria in terms of specific thresholds for sensitivity, specificity, or agreement percentages for substantial equivalence. However, the performance metrics are compared against a legally marketed predicate device (Trinity Biotech MarDx Borrelia burgdorferi EIA IgM Test Kit). The demonstration of substantial equivalence relies on favorable comparisons to this predicate.
Approximate "Implied" Acceptance Criteria (Based on Predicate Performance and General Expectations) and Reported Performance:
| Performance Metric | Implied Acceptance Criteria (relative to predicate) | Reported Device Performance (Gold Standard Diagnostics IgM ELISA) |
|---|---|---|
| PPA (vs. Predicate) | High positive agreement (e.g., >80-90%) compared to the predicate. | 90.3% (93/103) with 95% CI (82.9% - 95.5%) |
| NPA (vs. Predicate) | High negative agreement (e.g., >95%) compared to the predicate. | 99.6% (460/462) with 95% CI (98.5% - 99.9%) |
| 2nd Tier PPA | High positive agreement with a Second Tier IgM Blot (e.g., >90%) | 98.3% (58/59) with 95% CI (90.9% - 99.9%) |
| Analytical Specificity (Endemic) | High, likely >90-95% | 96.1% |
| Analytical Specificity (Non-Endemic) | High, likely >90% | 90.5% |
| Precision (CV%) | Within acceptable ranges for ELISA assays (typically <15-20% for low values, lower for moderate/high) | Max CV reported for samples: Within-Run 18.1%, Total 17.9% (Negative control) |
| Reproducibility (CV%) | Within acceptable ranges for ELISA assays (similar to precision) | Max CV reported for samples: Within-Run 17.0%, Total 18.6% (Negative sample) |
| Cross-Reactivity | Minimal or explainable cross-reactivity with other conditions. | See table below for specific values. |
| Interfering Substances | No significant interference from common substances. | None detected |
| Clinical Sensitivity | Comparable to predicate across disease stages. | Early: 75.9%, Disseminated: 100.0%, Late: 89.7% |
| CDC Panel Agreement (Healthy) | High agreement for healthy individuals. | 93.0% |
| CDC Panel Agreement (Early Lyme) | Reasonable agreement for early Lyme. | 76.7% |
| CDC Panel Agreement (Other Lyme Stages) | Reasonable agreement for other Lyme stages. | Cardiac: 66.7%, Neurological: 100%, Late: 85.0% |
| CDC Panel Agreement (Look-alike Disease) | High agreement for look-alike diseases (i.e., low positivity expected). | 81.1% |
2. Sample Size Used for the Test Set and Data Provenance
- Comparison with Predicate Device (Clinical Study):
- Sample Size: 531 serum samples.
- Data Provenance: Prospective samples submitted for Lyme serology testing. Origin not explicitly stated by country, but implies clinical samples from typical patient populations where Lyme serology is performed.
- Analytical Specificity:
- Sample Size: 208 asymptomatic individual samples (103 from an endemic region, 105 from a non-endemic region).
- Data Provenance: Not explicitly stated, presumed retrospective acquisition from asymptomatic individuals. No country of origin is mentioned.
- Cross-Reactivity:
- Sample Size: 277 samples from serum vendors "who confirmed their positivity for each respective marker."
- Data Provenance: Retrospective acquisition from serum vendors with various infection/disease conditions. No country of origin is mentioned.
- Sensitivity Study (Clinically Characterized Samples):
- Sample Size: 114 clinically characterized samples (58 Early, 17 Disseminated, 39 Late stages of Lyme disease).
- Data Provenance: Not explicitly stated, but "clinically characterized" suggests retrospective collection with confirmed diagnoses. No country of origin is mentioned.
- CDC Panel:
- Sample Size: 280 positive and negative specimens.
- Data Provenance: Specimens from the Center for Disease Control (CDC). This is a well-characterized, presumably retrospective, panel. The U.S. CDC implies U.S. origin.
- Determination of Assay Cutoff:
- Sample Size: 208 normal sera (103 from an endemic region, 105 from a non-endemic region) plus an additional 194 samples (114 from different phases of Lyme disease, 8 negative healthy samples, 72 negative Lyme disease samples with other diseases causing cross-reactivity).
- Data Provenance: Normal sera from endemic/non-endemic regions, and samples representing various Lyme disease phases and other conditions. The provenance is mixed (healthy, disease stages, cross-reactivity).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The document does not specify the number or qualifications of experts used to establish the ground truth for any of the test sets.
- For the Comparison with Predicate Device and Analytical Specificity studies, the ground truth is indirectly established by the performance of the predicate device and the classification of samples as "asymptomatic" or "submitted for Lyme serology testing," rather than via an expert panel.
- For the Cross-Reactivity study, the samples were obtained from "serum vendors who confirmed their positivity for each respective marker." This implies that the ground truth for these samples was based on previous laboratory testing/diagnosis, not explicitly on expert adjudication for this specific study.
- For the Sensitivity Study, the samples were "clinically characterized," suggesting that the gold standard was clinical diagnosis, but the process and experts involved are not detailed.
- For the CDC Panel, the specimens are described as "masked characterized serum panel," implying the CDC has established the ground truth for these samples (e.g., through clinical diagnosis, Western blot, culture, or other validated methods), but the specific method or experts are not detailed.
- The Second Tier Testing uses an "FDA cleared IgM Western blot assay" as a reference standard, which is itself a laboratory test, not expert consensus.
4. Adjudication Method for the Test Set
No explicit adjudication method (e.g., 2+1, 3+1) is mentioned for establishing ground truth for any of the test sets. Ground truth appears to be based on:
- The performance of a predicate device for comparison studies.
- Clinical characterization or diagnosis for sensitivity studies.
- Confirmed positivity by serum vendors or CDC characterization for specificity and panel studies.
- FDA-cleared IgM Western blot for second-tier testing.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance
This device is an ELISA (Enzyme-Linked Immunosorbent Assay) Test Kit, which is a laboratory diagnostic assay, not an imaging device or AI-driven diagnostic system involving human readers. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and was not performed.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
This device is a standalone laboratory diagnostic kit. Its performance, as reported in the studies (e.g., PPA, NPA, sensitivity, specificity, precision, reproducibility), represents its standalone performance without a human-in-the-loop component beyond the standard laboratory technician performing the assay. The results are based on objective optical density measurements.
7. The Type of Ground Truth Used
The ground truth used for different studies varies:
- Comparison with Predicate Device: The predicate device's results are used as a reference for agreement, not necessarily a true "ground truth" per se but a comparative standard.
- Analytical Specificity: Clinical status (asymptomatic, endemic/non-endemic region) is used.
- Cross-Reactivity: Confirmed positivity for other markers by serum vendors.
- Second Tier Testing: Results from an "FDA cleared IgM Western blot assay."
- Sensitivity Study: "Clinically characterized samples," indicating clinical diagnosis based on symptoms, history, and potentially other lab data.
- CDC Panel: "Masked characterized serum panel" from the CDC, implying predefined positive/negative status likely based on a combination of clinical information and validated laboratory tests.
- Cutoff Determination: Normal sera from endemic/non-endemic regions and samples from different phases of Lyme disease (clinically defined).
8. The Sample Size for the Training Set
- Determination of Assay Cutoff:
- 208 normal sera: 103 from an endemic region and 105 from a non-endemic region were used to determine the initial cutoff value (mean plus two standard deviations).
- An additional 194 samples: 114 from different phases of Lyme disease, 8 negative healthy samples, and 72 negative Lyme disease samples with other conditions for cross-reactivity were used for ROC analysis to confirm the cutoff.
- While not explicitly called a "training set" in the context of machine learning, these samples were used to establish and validate the assay's interpretive criteria (cutoff value), which is analogous to how a training set informs a model's parameters.
- No other explicit training sets for the device's main performance characteristics are described, as ELISA kits do not typically involve iterative machine learning training in the same way an AI device would.
9. How the Ground Truth for the Training Set Was Established
For the samples used in cutoff determination (analogous to a training set):
- Normal sera: Presumed to be healthy controls based on their collection from "normal" individuals in endemic and non-endemic regions.
- Disease-phase samples: "Clinically characterized samples" and "samples from different phases of Lyme disease" suggest a ground truth established by clinical diagnosis.
- Cross-reactivity samples: Negative Lyme disease status with confirmed other conditions.
The establishment of ground truth for these samples aligns with the methods described in point 7.
Ask a specific question about this device
(91 days)
The Gold Standard Diagnostics Borrelia burgdorferi igG ELISA Test Kit is intended as a qualitative presumptive (first-step) test for the detection of IgM antibodies to B. burgdorferi sensu stricto in human serum from symptomatic patients of infection. Positive and equivocal results must be supplemented by testing with a second-step Western blot assay.
The kit includes 12 x 8 well Antigen Coated strips, Conjugate, Substrate, Stop Solution, Wash Buffer, Diluent, Negative Control, Positive Control, and Cutoff Control. The controls are provided to determine if the assay is functioning properly and to determine the antibody level. The reagents are sufficient for 96 determinations. During the test procedure, antibodies to B. burgdorferi (sensu stricto) if present in the human serum sample will bind to the antigens coated onto the wells forming antigen-antibody complexes. Excess antibodies are removed by washing. A conjugate of goat anti-human IgG antibodies conjugated with horseradish peroxidase is then added, which binds to the antigenantibody complexes. Excess conjugate is removed by washing. This is followed by the addition of a chromogenic substrate, tetramethylbenzidine (TMB). If specific antibodies to the antigen are present in the patients' serum, a blue color will develop. The enzymatic reaction is then stopped with a stopping solution causing the contents of the well to turn yellow. The wells are read photometrically with a microplate reader at 450mm. The antigens used in the Gold Standard Diagnostics Borrelia burgdorferi IgG ELISA Test kit is a combination of B. burgdorferi sensu stricto strain B31 lysate, B. burgdorferi sensu stricto strain 2591 lysate, and a recombinant VlsE protein from B. burgdorferi sensu stricto strain B31. The lysates use spirochetes growing in BSK-H complete medium until mid-exponential phase. The recombinant VlsE protein is produced in E. coli SURE2 cells and purified by affinity chromatography.
The provided document is a 510(k) summary for the Gold Standard Diagnostics Borrelia burgdorferi IgG ELISA Test Kit, a medical device for diagnosing Lyme disease. It describes the device's technical specifications and the nonclinical and clinical studies performed to demonstrate its substantial equivalence to a legally marketed predicate device.
Here's the breakdown of the acceptance criteria and the study that proves the device meets them:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly present acceptance criteria in a formal table with pass/fail thresholds. Instead, it demonstrates the device's performance through various studies and compares it to a predicate device. The implicit acceptance criteria are that the new device performs comparably to the predicate device and demonstrates acceptable precision, specificity, and sensitivity.
Here's a summary of the reported device performance for key metrics:
| Performance Metric | Reported Device Performance (Gold Standard Diagnostics Borrelia burgdorferi IgG ELISA Test Kit) | Predicate Device Performance (Trinity Biotech MarDx Borrelia burgdorferi EIA IgG Test Kit) | Notes |
|---|---|---|---|
| Comparison with Predicate Device (523 samples) | |||
| Positive Percent Agreement | 90.2% (55/61) with 95% CI (79.8% - 96.3%) | N/A (this is a comparative measure) | This compares the agreement of positive/equivocal results between the subject device and the predicate device. Equivocal samples were counted as positive for this calculation. |
| Negative Percent Agreement | 99.6% (460/462) with 95% CI (98.5% - 99.9%) | N/A (this is a comparative measure) | This compares the agreement of negative results between the subject device and the predicate device. |
| 2nd Tier Testing (IgG Western Blot) | |||
| 2nd Tier PPA | 100% (37/37) with 95% CI (92.2% - 100%) | N/A (this is a comparative measure) | This indicates that all samples positive or equivocal by both the subject device and the predicate device that also tested positive by Western Blot were identified by the subject device. |
| Clinical Sensitivity (114 clinically characterized samples) | |||
| Early Stage | 46.6% (27/58) | 27.6% (16/58) | The subject device showed higher sensitivity in the early stage compared to the predicate. |
| Disseminated Stage | 82.4% (14/17) | 52.9% (9/17) | The subject device showed significantly higher sensitivity in the disseminated stage compared to the predicate. |
| Late Stage | 97.4% (38/39) | 97.4% (38/39) | Sensitivity in the late stage was identical to the predicate device. |
| Analytical Specificity | |||
| Endemic Region (103 samples) | 96.1% | Not explicitly compared/stated for predicate | Determined by testing asymptomatic individuals from endemic regions. 4 out of 103 samples were positive/equivocal. |
| Non-Endemic Region (105 samples) | 100% | Not explicitly compared/stated for predicate | Determined by testing asymptomatic individuals from non-endemic regions. 0 out of 105 samples were positive/equivocal. |
| Cross-Reactivity (377 samples) | Varies by condition (e.g., Rickettsiosis IgG: 24%, Ehrlichiosis IgG: 20%, EBV IgG: 3%, VZV: 6%) | Not explicitly compared/stated for predicate | A detailed table is provided showing specific cross-reactivity rates with various other infections/conditions. The goal is to demonstrate that the device performs acceptably despite potential cross-reactivity. The implicit acceptance would be that the cross-reactivity profile is comparable to or better than the predicate, or at least clinically acceptable for the intended use. |
| Precision | Within-run CV: 2.7% - 14.2%Between-run CV: 1.1% - 10.6%Between-day CV: 2.8% - 11.0%Total CV: 2.6% - 14.8% (for various sample types) | Not explicitly compared/stated for predicate | A detailed table of SD and CV values for negative, high negative, low positive, moderate positive samples, and controls is provided, demonstrating good precision within their ranges. |
| Reproducibility | Within-run CV: 2.4% - 21.1%Between-run CV: 1.9% - 18.7%Between-day CV: 2.3% - 16.4%Between-site CV: 2.2% - 18.8%Total CV: 2.2% - 18.3% (for various sample types) | Not explicitly compared/stated for predicate | A detailed table of SD and CV values across three sites for negative, high negative, low positive, moderate positive samples, and controls, demonstrating good reproducibility. |
2. Sample sizes used for the test set and the data provenance
- Comparison with Predicate Device: 523 serum samples.
- Provenance: Prospective samples submitted for Lyme serology testing. Conducted at three sites (one internal and two external reference laboratories). The specific country of origin is not noted, but given the FDA submission, it's likely primarily US-based or at least includes US data.
- Cutoff Determination: 210 normal sera (105 from an endemic region, 105 from a non-endemic region). An additional 194 samples (114 Lyme disease, 8 negative healthy, 72 negative Lyme disease with other potential cross-reactive diseases).
- Provenance: Retrospective (these seem to be characterized samples). Specific regions/countries not detailed.
- Precision Study: 48 replicates for each of the 6 panel members (negative, high negative, low positive, moderate positive sample, positive control, cutoff control, negative control).
- Provenance: In-house study.
- Reproducibility Study: 90 replicates for each of 4 panel members (moderate positive, low positive, high negative, negative). 30 or 60 replicates for controls (positive, cutoff, negative).
- Provenance: Tested at three different sites.
- Analytical Specificity: 208 asymptomatic individual samples (103 from an endemic region, 105 from a non-endemic region).
- Provenance: Retrospective (asymptomatic individuals). Specific regions/countries not detailed.
- Cross-Reactivity: 377 samples from various infection/disease conditions.
- Provenance: Obtained from serum vendors, confirmed for positivity for respective markers or clinical diagnosis. Retrospective.
- Clinical Sensitivity (Clinically characterized samples): 114 clinically characterized samples (58 early, 17 disseminated, 39 late stages of Lyme disease).
- Provenance: Retrospective (clinically characterized samples).
- CDC Panel: 280 positive and negative specimens from the CDC.
- Provenance: Retrospective. The CDC panel implies a well-characterized, reference sample set.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
This is an ELISA diagnostic kit, not an AI/imaging device. Therefore, the concept of "experts establishing ground truth" in the sense of radiologists reading images is not directly applicable.
For this type of device:
- Ground Truth: The ground truth for the test samples (e.g., "Lyme positive/negative," "specific infection") is established through clinical characterization, confirmed diagnosis, or established reference panels (like the CDC panel), potentially involving a combination of clinical symptoms, other laboratory tests (e.g., Western blot, PCR), and expert clinical judgment.
- The document states that for the cross-reactivity study, samples were obtained from "serum vendors who confirmed their positivity for each respective marker or clinical diagnosis," implying that the ground truth for these samples was based on established laboratory methods or clinical records, not necessarily adjudicated by human experts in a reading study.
- For the "Clinically characterized samples" in the sensitivity study, the ground truth ("early, disseminated, and late stages of Lyme disease") would have been established by clinicians based on the patients' medical history, symptoms, and other diagnostic findings. The number and specific qualifications of these clinicians are not provided in this summary.
4. Adjudication method for the test set
Not applicable in the context of an ELISA kit where pre-characterized biological samples define the "ground truth." The "adjudication" is inherent in the characterization of the biological samples used for testing, often through a combination of clinical presentation and other established laboratory methods (e.g., a two-tiered testing algorithm for Lyme disease).
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is an in vitro diagnostic (IVD) kit, not an AI-assisted diagnostic imaging device that involves human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, this entire study is a "standalone" performance evaluation of the ELISA kit. The device itself (the ELISA kit) provides a quantitative and qualitative result (positive, equivocal, negative) based on the optical density reading, without direct human cognitive input shaping that specific result beyond the laboratory technician performing the assay according to instructions. The interpretation of "positive/equivocal/negative" is based on pre-defined cutoffs.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The ground truth for the various studies includes:
- Clinical Diagnosis/Characterization: For the sensitivity study (early, disseminated, late Lyme disease), the ground truth was based on the clinical status of the patients.
- Reference Panels: The CDC panel specimens serve as a well-characterized reference (positive and negative) for Lyme disease.
- Other Laboratory Confirmatory Tests: For the comparison study, all positive and equivocal samples from both the subject and predicate ELISA were tested by an FDA-cleared IgG Western blot assay, which serves as a second-tier confirmatory ground truth for these specific samples.
- Confirmed Positivity/Clinical Diagnosis from Serum Vendors: For cross-reactivity studies, samples were sourced with confirmed positivity for specific markers or clinical diagnoses.
- Healthy/Asymptomatic Status: For cutoff determination and analytical specificity, ground truth was derived from healthy, asymptomatic individuals from endemic and non-endemic regions.
8. The sample size for the training set
This document describes a premarket notification for a traditional ELISA diagnostic kit, not a machine learning/AI algorithm. Therefore, there is no explicit "training set" in the context of supervised machine learning. The term "training set" is usually associated with AI model development.
For an ELISA kit, development involves:
- Assay Design/Optimization: This is an iterative process using various internal samples to optimize antigen/antibody concentrations, incubation times, buffer compositions, etc.
- Cutoff Determination: As stated, 210 normal sera and 194 other characterized samples were used to determine and confirm the assay cutoff. This set functions somewhat like a "calibration" or "training" set for defining the interpretation thresholds.
9. How the ground truth for the training set was established
Again, this is not a machine learning model. For the samples used in cutoff determination (which is analogous to a "training" or "calibration" phase), the ground truth was established by:
- Normal Sera: Defined as "normal" based on clinical health and potentially geographic origin (endemic/non-endemic for Lyme disease).
- Lyme Disease Samples: "114 samples from different phases of Lyme disease" implies these were clinically characterized Lyme patient samples.
- Other Disease Samples: "72 negative Lyme disease samples but do have other diseases that may cause serologic cross-reactivity," implying a clinical diagnosis of the other disease and a confirmed negative status for Lyme.
The goal of this "cutoff determination" phase is to establish a threshold that optimizes sensitivity and specificity against a diverse set of known samples.
Ask a specific question about this device
(92 days)
The Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test Kit is intended as a qualitative presumptive (first-step) test for the detection of IgG and IgM antibodies to B. burgdorferi sensu stricto in human serum from symptomatic patients or people suspected of infection. Positive and equivocal results must be supplemented by testing with a second-step Western blot assay.
The kit includes 12 x 8 well Antigen Coated strips, Conjugate, Substrate, Stop Solution, Wash Buffer. Diluent. Negative Control. Positive Control. The control. The controls are provided to determine if the assay is functioning properly and to determine the antibody level. The reagents are sufficient for 96 determinations.
During the test procedure, antibodies to B. burgdorferi (sensu stricto) if present in the human serum sample will bind to the antigens coated onto the wells forming antigen-antibody complexes. Excess antibodies are removed by washing. A conjugate of goat anti-human IgG/IgM antibodies conjugated with horseradish peroxidase are then added, which binds to the antigen-antibody complexes. Excess conjugate is removed by washing. This is followed by the addition of a chromogenic substrate, tetramethylbenzidine (TMB). If specific antibodies to the antigen are present in the patients' serum, a blue color will develop. The enzymatic reaction is then stopped with a stopping solution causing the contents of the well to turn yellow. The wells are read photometrically with a microplate reader at 450nm.
The antigens used in the Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test kit is a combination of B. burgdorferi sensu strain B31 lysate, B. burgdorferi sensu stricto strain 2591 lysate, and a recombinant VIsE from B. burgdorferi sensu stricto strain B31. The lysates use spirochetes growing in BSK-H complete medium until mid-exponential phase. The recombinant VlsE protein is produced in E. coli SURE2 cells and purified by affinity chromatography. The purity of each antigen is assayed by SDS-PAGE followed by Coomassie staining and/or Western blotting.
Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are not explicitly stated as distinct numerical targets in the provided document. Instead, the document demonstrates substantial equivalence to a predicate device and provides performance metrics from various studies. Based on the studies performed, the "reported device performance" is presented in the context of sensitivity, specificity, agreement with the predicate device, precision, and reproducibility.
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Analytical Specificity (Non-endemic region) | 98.0% (2 cases positive/equivocal out of 100 samples) |
| Analytical Specificity (Endemic region) | 97.3% (3 cases positive/equivocal out of 110 samples) |
| Precision (Within-Run CV) | Moderate Positive: 4.2%; Low Positive: 2.2%; High Negative: 3.4%; Negative: 6.6% |
| Precision (Total CV) | Moderate Positive: 8.9%; Low Positive: 11.9%; High Negative: 11.4%; Negative: 11.6% |
| Reproducibility (Within-Run CV) | Moderate Positive: 5.3%; Low Positive: 4.3%; High Negative: 5.0%; Negative: 9.1% |
| Reproducibility (Total CV) | Moderate Positive: 9.0%; Low Positive: 9.4%; High Negative: 12.8%; Negative: 23.0% |
| Cross-Reactivity (Demonstrated absence of significant interference from listed conditions) | Low false positive rates observed for conditions like Tick-borne Relapsing Fever (7.7%), Treponemal Infections (8.7%), Fibromyalgia (10%); 0% for Rickettsia, Ehrlichiosis, Babesiosis, Leptospirosis, Parvovirus B19, Influenza A&B, Epstein-Barr Virus, Cytomegalovirus, H. pylori, Rheumatoid Arthritis, Herpes Simplex Virus, Varicella Zoster Virus, Autoimmune Disease. (Note: 2 Treponemal Infection samples were also positive on the predicate device). |
| Interfering Substances (Demonstrated no effect) | None detected for Albumin, Bilirubin, Cholesterol, Hemoglobin, Triglycerides at specified concentrations. |
| Positive Percent Agreement with Predicate Device | 96.0% (72/75), 95% CI (88.8% - 99.2%) |
| Negative Percent Agreement with Predicate Device | 97.5% (434/445), 95% CI (95.6% - 98.8%) |
| Clinical Sensitivity (Overall from Sensitivity Study) | Early: 76.3%; Disseminated: 100.0%; Late: 97.2% |
| Clinical Sensitivity (CDC Panel) | Healthy: 100%; Early (0-2m): 86.7%; Convalescent (3-12m): 53.8%; Late (>1yr): 100% |
| Equivalence to Predicate Device | Demonstrated substantial equivalence through comparative performance in clinical and analytical studies. |
2. Sample Sizes and Data Provenance
- Test Set (Clinical Studies):
- Comparison with Predicate Device: 520 serum samples (prospective samples).
- Sensitivity Study (Clinically Characterized): 89 clinically characterized samples.
- CDC Panel: 40 samples (5 healthy, 35 Lyme disease stratified by stage).
- Analytical Specificity: 210 asymptomatic individuals (110 from endemic region/Pennsylvania, 100 from non-endemic region/Arizona).
- Cross-Reactivity: 221 samples (obtained from serum vendors, confirmed positive for respective markers).
- Interfering Substances: 3 samples (negative, low positive, moderate positive, spiked).
- Data Provenance: The data appears to be from a mix of regions within the US (Pennsylvania, Arizona for specificity, CDC for panel), and some samples from "serum vendors" for cross-reactivity. The clinical comparison study used "prospective samples submitted for Lyme serology testing" from three sites (one internal, two external reference laboratories). The sensitivity study used "clinically characterized samples."
3. Number of Experts and Qualifications for Ground Truth (Test Set)
- The document does not specify the number or qualifications of experts used to establish ground truth for the test set in the comparative clinical studies or sensitivity studies.
- For the "clinically characterized samples" and "patients diagnosed with Lyme disease" (CDC panel), it is implied that a clinical diagnosis (made by medical professionals) served as the ground truth, but no details on the specific experts (e.g., infectious disease specialists, their experience) are provided.
- For cross-reactivity, samples were from "serum vendors who confirmed their positivity for each respective marker," which likely relied on established diagnostic methods but doesn't detail expert involvement.
4. Adjudication Method (Test Set)
- The document does not describe a specific adjudication method (e.g., 2+1, 3+1, none) for resolving discrepancies in the test set.
- For the "Comparison with Predicate Device" table, "Equivocal samples counted as positive" is an interpretation rule, not an adjudication method for ground truth.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. This device is an in-vitro diagnostic (IVD) ELISA test kit, which is a laboratory assay and does not involve human readers interpreting results in the same way an imaging AI would. Therefore, the concept of human readers improving with AI assistance is not applicable here.
6. Standalone Performance (Algorithm Only)
- Yes, the performance presented for the "Gold Standard Diagnostics Borrelia burgdorferi IgG/IgM ELISA Test Kit" throughout the document (e.g., Precision, Reproducibility, Analytical Specificity, Cross-Reactivity, and its results in the Clinical Sensitivity and CDC Panel studies) is its standalone performance. This test kit is a self-contained diagnostic assay.
7. Type of Ground Truth Used
- Clinical Diagnosis/Characterization: For the sensitivity study and the CDC panel, the ground truth was based on "clinically characterized samples" and "patients diagnosed with Lyme disease," implying a clinical diagnosis by medical professionals.
- Absence of Disease/Asymptomatic Status: For the analytical specificity study, ground truth was derived from "asymptomatic individuals" from endemic and non-endemic regions, presumed to be negative for Lyme disease.
- Confirmed Positivity: For the cross-reactivity study, samples were from "serum vendors who confirmed their positivity for each respective marker."
- ELISA Results (Predicate Device): For the "Comparison with Predicate Device" study, the predicate device's results served as a comparative reference point, but not necessarily the ultimate gold standard for true disease status.
8. Sample Size for the Training Set
- The document implicitly references a "training set" for the establishment of the assay cutoff: "The cutoff was determined by testing a total of 200 normal sera which consisted of 100 sera from an endemic region of Lyme disease and 100 sera from a non-endemic region of Lyme disease."
- Additionally, after initial cutoff determination, "125 characterized Lyme disease samples were tested" and an ROC analysis was generated with the 325 samples (200 normal and 125 characterized samples) "to verify the chosen cutoff." This combined set of 325 samples served as a "training" or calibration set for optimizing the assay's cutoff.
9. How the Ground Truth for the Training Set was Established
- For the "training set" used to determine the assay cutoff:
- Normal Sera: The 200 normal sera were obtained from "an endemic region of Lyme disease and a non-endemic region of Lyme disease." Their "normal" status implies they were considered negative for Lyme disease based on clinical assessment or lack of symptoms/exposure.
- Characterized Lyme Disease Samples: The 125 "characterized Lyme disease samples" likely had a confirmed clinical diagnosis of Lyme disease, although specific details of this characterization are not provided. The phrase "characterized Lyme disease samples" implies a definitive diagnosis beyond just a positive test.
Ask a specific question about this device
(273 days)
The Gold Standard Diagnostics AIX1000 Rapid Plasma Reagin (RPR) Automated Test System is a non-treponemal flocculation test that can qualitatively determine the presence of reagin antibodies in human serum. It may be used to aid in the diagnosis of syphilis when used in conjunction with supplemental treponemal laboratory tests and other clinical information. This test may also be used to detect non-treponemal antibodies in samples serially diluted to establish titer information. This test is not intended for screening blood or tissue donors.
The Gold Standard Diagnostics AIX1000 Rapid Plasma Reagin (RPR) Automated Test System in a non-treponemal test for the qualitative determination of reagin antibodies in human serum to aid in the diagnosis of syphilis. This test is also used to detect non-treponemal antibodies in samples serially diluted to establish titer information. The system consists of the Gold Standard Diagnostics RPR test reagents and the Gold Standard Diagnostics AIX1000 Agglutination Analyzer. The AIX1000 Analyzer delivers serum from collection tubes into test wells. After antigen suspension is added, the test wells are then incubated while being shaken. An onboard camera creates a high resolution image. The image is then analyzed by the proprietary software algorithm to produce a result.
Here's an analysis of the acceptance criteria and the studies that prove the device meets these criteria, based on the provided text:
Device Name: Gold Standard Diagnostics AIX1000 Rapid Plasma Reagin (RPR) Automated Test System
K Number: K150358
1. Table of Acceptance Criteria and Reported Device Performance
The document doesn't explicitly state "acceptance criteria" in a bulleted or numbered list with predefined thresholds for all studies. However, performance expectations are implied by the nature of the tests conducted and the reported agreement percentages. I will interpret the reported performance metrics as demonstrating the device meets an implied acceptance for its intended use, particularly through comparison to a predicate device.
| Study Type / Performance Metric | Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|---|
| Cross Reactivity | No observed cross-reactivity with common viral, bacterial, and autoimmune conditions. | Viral: Rubella (10), VZV (10), HIV (10), Hepatitis B (16), Hepatitis C (11), EBV (10), HSV Type 1 (10), HSV Type 2 (10), CMV (11), Heterophile antibodies (10) - All 0 Reactive. Bacterial: Toxoplasma gondii (10), Leptospira biflexa (10), Borrelia burgdorferi (10) - All 0 Reactive. Autoimmune: SLE (10), Rheumatoid Arthritis (10), Scleroderma (10), Primary Anti-phospholipid Syndrome (16) - All 0 Reactive. (Numbers in parentheses are 'Number Tested'). |
| Interfering Substances | No effect on performance by specified endogenous substances and prescription drugs. | Hemoglobin (20 g/dL), Bilirubin (unconjugated) (15 mg/dL), Cholesterol (250 mg/dL), Albumin (5 g/dL), Gamma Globulin (60 mg/dL), Glucose (120 mg/dL), Triglyceride (500 mg/dL), Antibiotic (Cephalexin) (337 umol/L), Antibiotic (Tetracycline) (34 umol/L). For all substances, RPR samples remained positive where expected; "None Observed" for interference. |
| Precision (Within-Lab) | High agreement (e.g., >90%) within ± 1 titer for various reactivity levels. | - Non-Reactive Serum: 100% agreement (93.6% - 100% C.I.) - Low RPR Reactivity (1:4): 100% agreement (93.6% - 100% C.I.) - Moderately Reactive (1:16): 100% agreement (93.6% - 100% C.I.) - Reactive (1:64): 97.8% agreement (88.2% - 99.9% C.I.) - Highly Reactive (1:128): 100% agreement (93.6% - 100% C.I.) - Reactive control: 100% agreement (54.9% - 100% C.I.) - Non-reactive control: 100% agreement (54.9% - 100% C.I.) |
| Reproducibility | High agreement (e.g., >90%) within ± 1 titer across operators, instruments, days, and runs. | - Non-Reactive Serum: 100% (Between-Runs, Between-Days, Between-Operators, Between-Instruments) - Low RPR Reactivity (1:4): 100% (all categories) - Moderately Reactive (1:16): 100% (all categories) - Reactive (1:64): 100% (Between-Runs, Between-Operators, Between-Instruments), 97.8% (Between-Days) - Highly Reactive (1:128): 100% (Between-Runs, Between-Days, Between-Instruments), 98.1% (Between-Operators) - Reactive control: 100% (90.5% - 100% C.I.) - Non-reactive control: 100% (90.5% - 100% C.I.) |
| Carry-over | No evidence of carry-over (no false positives in negative samples due to highly reactive samples). | All 480 replicates of the negative samples were reported as non-reactive when highly reactive samples were alternated with non-reactive samples. |
| Clinical Agreement (Prospective Samples vs. Predicate) | High Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) with predicate. | PPA: 95.5% (C.I. 77.2% - 99.9%) NPA: 99.9% (C.I. 99.3% - 100%) |
| Clinical Agreement (Retrospective Samples vs. Predicate) | High PPA and NPA with predicate. | PPA: 97.2% (C.I. 95.5% - 98.4%) NPA: 99.1% (C.I. 98.5% - 99.5%) |
| Special Populations (Pregnant Women vs. Predicate) | High PPA and NPA with predicate. | PPA: 100% (C.I. 90.5% - 100%) NPA: 100% (C.I. 98.8% - 100%) |
| Special Populations (HIV Positive Individuals vs. Predicate) | High PPA and NPA with predicate. | PPA: 100% (C.I. 90.5% - 100%) NPA: 100% (C.I. 98.8% - 100%) |
| Healthy Individuals | Low and expected reactivity in a healthy, no-risk population. | All 100 samples from apparently healthy individuals (not at risk, no syphilis test ordered) were non-reactive. |
| Clinical Stages Correlation | High agreement with clinically diagnosed syphilis cases across various stages. | Primary Treated (13/13 reactive), Primary Untreated (12/12 reactive), Secondary Treated (25/25 reactive), Secondary Untreated (25/25 reactive), Latent Treated (25/25 reactive), Latent Untreated (25/25 reactive). All showed 100% agreement with corresponding 95% C.I. ranges of 77.9%-100% to 88.7%-100%. |
2. Sample Sizes Used for the Test Set and Data Provenance
The document describes several test sets:
- Cross Reactivity: Panels of 10-16 individual patient samples per condition (totaling 17 conditions for viral, bacterial, and autoimmune).
- Interfering Substances: 5 samples (1 non-reactive, 4 reactive) per interfering substance.
- Precision: 5 samples (Non-reactive, Low RPR (1:4), Moderately Reactive (1:16), Reactive (1:64), Highly Reactive (1:128)). Each tested in 9 replicates.
- Reproducibility: The same 5 sample panels as precision, tested across 6 operators, 3 instruments, 2 runs over 5 days.
- Carry-over: 1 reactive (1:64), 1 highly reactive (1:128), and 2 non-reactive samples. Highly reactive samples alternated with non-reactive samples 96 times per run over 5 runs.
- Clinical Studies - Prospectively Collected Samples:
- Sample Size: 765 serum samples.
- Data Provenance: Collected prospectively from patient samples with a physician's order for syphilis testing. Collected at two geographically distinct reference laboratories (Southeastern and Western United States).
- Clinical Studies - Retrospectively Collected Samples:
- Sample Size: 2,246 samples.
- Data Provenance: Collected retrospectively from patients referred for syphilis testing. Obtained from sample brokers who collect from multiple sites across the United States. Collection dates: January 2005 - July 2014.
- Clinical Studies - Pregnant Women (Retrospective):
- Sample Size: 250 non-reactive, 30 reactive (spiked) samples. Total 280.
- Data Provenance: Retrospectively collected from pregnant women at one site (Southeastern United States). Collection dates: July 2012 - August 2013.
- Clinical Studies - HIV Positive Individuals (Retrospective):
- Sample Size: 250 non-reactive, 30 reactive samples. Total 280.
- Data Provenance: Retrospectively collected from HIV positive individuals at four sites (one Southeastern, one Mid-Western States). Collection dates: February 2012 - June 2015.
- Apparently Healthy Individuals (Prospective):
- Sample Size: 100 serum samples.
- Data Provenance: Prospectively collected from healthy individuals not at risk for syphilis and for whom a syphilis test had not been ordered (submitted for routine chemistry testing).
- Correlation with Clinically Diagnosed Syphilis Sera - Various Stages (Purchased Panel):
- Sample Size: Primary Treated (13), Primary Untreated (12), Secondary Treated (25), Secondary Untreated (25), Latent Treated (25), Latent Untreated (25).
- Data Provenance: Purchased from serum brokers, characterized by clinical diagnosis.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
The document does not mention the use of human experts to establish ground truth for the test sets in the context of adjudication or interpretation of the RPR results. Instead, it relies on:
- Reference laboratory results (for prospective samples).
- Predicate device results (for comparison in clinical studies).
- FDA-cleared treponemal (TP) assays (for further investigation of discrepant non-treponemal results).
- Clinical diagnosis (for the purchased syphilis stage panel).
- Serum broker confirmation of disease markers (for cross-reactivity samples).
Therefore, there is no direct information on the number or qualifications of experts involved in establishing ground truth through manual interpretation of RPR images or similar tasks for the "AIX1000" aspect. The 'AI' component is described as proprietary software interpreting images (standalone).
4. Adjudication Method for the Test Set
Adjudication methods (like 2+1, 3+1) are typically used when multiple human readers interpret the same image/case and disagreements need to be resolved. This device is an automated test system for a flocculation assay where the "AIX1000 Agglutination Analyzer" uses a proprietary software algorithm to produce a result.
For the discrepancies in the clinical studies where the GSD AIX1000 RPR Test System disagreed with the comparator device:
- Prospective Samples: The two discrepant samples were tested on a "third FDA cleared RPR assay."
- Retrospective Samples: The 31 discrepant samples were tested on a "third FDA cleared RPR assay."
This suggests a form of tie-breaking or external reference testing for discrepant results, rather than a human expert adjudication of visual interpretations by the device itself.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done
No, an MRMC comparative effectiveness study was not conducted. The device is an automated test system that interprets RPR assay results without direct human intervention in the interpretation process described. The comparisons were between the automated device and a predicate automated device or clinical outcomes, not between human readers with and without AI assistance.
6. If a Standalone Performance (i.e., algorithm only without human-in-the-loop performance) was Done
Yes, the studies reported represent standalone performance of the Gold Standard Diagnostics AIX1000 RPR Automated Test System. The device's proprietary software algorithm directly produces the results from images captured by its onboard camera. The clinical studies (prospective, retrospective, special populations) directly compare the device's output to a predicate device's output or clinical diagnosis, without human modification or override of the AIX1000's initial determination. The interpretation of the flocculation assay itself is automated.
7. The Type of Ground Truth Used
The ground truth or reference methods varied depending on the study:
- Clinical Studies (Prospective and Retrospective): The primary ground truth for comparison was the legally marketed predicate device (Arlington Scientific Inc. (ASI) RPR Card Test for syphilis on the ASiManager-AT Analyzer). For discrepant samples, a third FDA cleared RPR assay was used as a tie-breaker. Further investigation for NT+ samples used an FDA cleared treponemal (TP) assay.
- Correlation with Clinically Diagnosed Syphilis Sera: Clinical Diagnosis (Primary, Secondary, Latent stages as characterized by documented symptoms, dark field microscopy, and reactive treponemal/non-treponemal tests).
- Cross-Reactivity Study: Confirmation of disease markers by serum brokers.
- Special Populations (Pregnant Women, HIV Positive): Predicate device results, and also confirmed pregnancy status via HCG test and HIV positive status.
- Apparently Healthy Individuals: Lack of risk factors and no physician's order for syphilis testing defined this as "healthy."
8. The Sample Size for the Training Set
The document does not provide information on the sample size used for the training set for the proprietary software algorithm. This information is typically proprietary to the manufacturer and not usually detailed in 510(k) summaries unless specifically requested or deemed critical for substantial equivalence. The focus of the 510(k) is often on comparing the new device's performance to a predicate, rather than the internal development of its algorithm.
9. How the Ground Truth for the Training Set Was Established
Similar to the training set size, the document does not provide information on how the ground truth for the training set was established. Since the device uses a "proprietary software algorithm" to analyze images for flocculation, it can be inferred that this algorithm was developed and trained using a collection of RPR test results with established reactivity and titer values. However, the specifics of this ground truth (e.g., expert consensus on visual flocculation, correlation with other assays) are not detailed in this submission.
Ask a specific question about this device
(295 days)
The Gold Standard Diagnostics Herpes Simplex Virus Type 1 IgG ELISA Test Kit is intended for the qualitative detection of IgG antibodies to Herpes Simplex Virus Type 1 (HSV-1) in human serum. The test is indicated for sexually active individuals and expectant mothers as an aid for the presumptive diagnosis of HSV-1 infection.
The predictive value of positive or negative results depends on the population's prevalence and the pretest likelihood of HSV-1. The test is not intended for screening of blood and plasma donors. The performance of this assay has not been established for use in a pediatric population, neonates, or immunocompromised patients.
The Gold Standard Diagnostics Herpes Simplex Virus (HSV) Type I IgG ELISA Test is an enzyme linked immunosorbent assay for the qualitative detection of IgG antibodies to HSV-1 in human serum. The assay requires a total of 90 minutes incubation time. The test uses microtiter wells coated with a recombinant gG1 protein of HSV-1. Serum is added to each well and incubated for 30 minutes at 37℃. If antibodies are present they will bind to the antigen in the well. Unbound antibodies are removed by washing the wells three times. A Horse Radish Peroxidase (HRP) conjugated goat anti-human IgG (conjugate) is then added to each well and incubated for 30 minutes at 37°C. If antibodies are present in the patient's serum, the conjugate will bind to the antibody attached to the antigen on the wells are again washed to remove any unbound conjugate. In order to detect the bound conjugate a substrate containing tetramethylbenzidine (TMB) is added to each well and incubated for 30 minutes at 37°C. If conjugate is present, the HRP will react with the substrate to generate a colored product. After the incubation period, the reaction is stopped with a . Stop Solution and the color intensity is measured spectrophotometrically. The kit also includes a Wash Buffer, Diluent, a Negative Control, and a Cutoff Control. The cut-off control is used to determine the validity of the assay and subsequently to determine the result of the sample. Positive and Negative controls are provided to determine if the assay is functioning properly. The kit contains 12 x 8well antigen coated microtiter strips in a frame. The reagents are sufficient for 96 determinations.
Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for this device are implicitly defined by its comparison to a legally marketed predicate device (Focus Diagnostics HerpeSelect® 1 and 2 Immunoblot IgG) and by demonstrating acceptable precision, reproducibility, cross-reactivity, and interference performance. The document doesn't explicitly state numerical acceptance criteria for sensitivity and specificity a priori but lists the achieved performance values.
| Test Category | Acceptance Criteria (Implicit) | Reported Device Performance |
|---|---|---|
| Clinical Performance | Substantial equivalence to predicate device in intended use populations (pregnant women, sexually active adults, low prevalence). Comparable sensitivity and specificity to predicate. | Pregnant Women:Sensitivity = 96.0% (95% C.I. = 90.1% - 98.9%)Specificity = 96.5% (95% C.I. = 90.0% - 99.3%)Sexually Active Adults:Sensitivity = 92.3% (95% C.I. = 88.6% - 95.0%)Specificity = 95.5% (95% C.I. = 91.8% - 97.8%)Low Prevalence:Sensitivity = 93.8% (95% C.I. = 69.7% - 99.8%)Specificity = 92.9% (95% C.I. = 85.1% - 97.3%)CDC Seroconversion Panel:Sensitivity = 97.8% (95% C.I. = 88.5% - 99.9%)Specificity = 94.4% (95% C.I. = 84.6% - 98.8%) |
| Precision (Within-lab) | Demonstrable precision across various sample types (high positive, moderate positive, low positive, near cutoff, high negative, negative). CV values indicating low variability. | High Positive: SD 1.544, CV 2.6% (within-run); Total CV 13.1%Moderate Positive: SD 1.290, CV 5.2% (within-run); Total CV 14.6%Low Positive: SD 0.664, CV 3.7% (within-run); Total CV 15.0%Near Cutoff: SD 0.601, CV 4.4% (within-run); Total CV 13.7%High Negative: SD 0.388, CV 5.9% (within-run); Total CV 14.5%Negative: SD 0.153, CV 8.0% (within-run); Total CV 18.0% |
| Reproducibility | Reproducibility across multiple sites, days, runs, and technicians for various sample types. CV values indicating acceptable variability. | Overall (across 3 sites):High Positive: Overall Total CV 9.5%Low Positive: Overall Total CV 14.6%Positive Cutoff: Overall Total CV 5.2%Negative Cutoff: Overall Total CV 2.9%High Negative: Overall Total CV 10.9%Negative: Overall Total CV 14.6% (Site-specific data also provided with similar CVs) |
| Cross-Reactivity | Minimal or no cross-reactivity with common infectious diseases and conditions. If reactive, it should align with predicate device. | 10 tested for each with: CMV (0 reactive), EBV (1* reactive), VZV (1* reactive), Chlamydia trachomatis (0 reactive), Treponema pallidum (0 reactive), HPV (0 reactive), Rubella Virus (0 reactive), Toxoplasma gondii (3* reactive), Candida albicans (2* reactive), Neisseria gonorrhea (0 reactive), Rheumatoid Factor (0 reactive), ANA (0 reactive), Measles (0 reactive), HSV-2 (0 reactive), HIV (4* reactive), Bacteroides (0 reactive). *Samples also reactive with predicate device. |
| Interfering Substances | Performance unaffected (within ±20%) by specified high concentrations of common interfering substances (hemoglobin, bilirubin, cholesterol, triglycerides, albumin). | Performance not affected by: Hemoglobin (2 g/L), Bilirubin (342 µmol/L), Cholesterol (13 mmol/L), Triglycerides (37 mmol/L), Albumin (60 g/L). (Acceptance criterion of ±20% was used.) |
2. Sample Size Used for the Test Set and Data Provenance
- Clinical Performance Test Set:
- Total Samples: 703 samples.
- Pregnant Women: 185 samples.
- Sexually Active Adults: 518 samples.
- Low Prevalence Population: 96 samples (sera ages 16-19 years old).
- CDC Seroconversion Panel: 100 samples (46 positive, 54 negative - implied from counts).
- Data Provenance:
- Pregnant Women and Sexually Active Adults: Prospectively collected.
- Low Prevalence Population: Prospectively collected in a non-STD setting. Tested in-house.
- CDC Seroconversion Panel: Obtained from the CDC. Tested in-house.
- Cross-reactivity samples: Obtained from serum brokers.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
The ground truth for the clinical performance study (the primary test set) was established by comparison to a commercially available predicate device: Focus Diagnostics HerpeSelect® 1 and 2 Immunoblot IgG test (K000238). The document does not mention the use of human experts to establish ground truth for this comparison study, as the predicate device itself serves as the reference.
For the cross-reactivity study, ground truth for the "positive for" status of the samples was confirmed by "serum brokers who confirmed positivity for each respective disease and conditions using FDA cleared tests." The number and specific qualifications of these "serum brokers" or the experts validating the FDA cleared tests are not specified.
4. Adjudication Method for the Test Set
For the clinical performance studies comparing the Gold Standard Diagnostics ELISA to the Focus Diagnostics Line Blot, the adjudication method for discordant results (equivocal results) was explicitly stated: "Equivocal results were treated as the worst case results (disagreement)". This implies that equivocal results were counted as disagreements with the predicate device's result when calculating sensitivity and specificity.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, an MRMC comparative effectiveness study was not done. This device is an in-vitro diagnostic (IVD) ELISA test, which does not involve human readers interpreting images or data to the extent that an MRMC study for AI would typically describe. The "reading" is done by a spectrophotometer, and the comparison is between two IVD assay technologies.
6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, the device's performance as an algorithm-only system (an ELISA test measured by a spectrophotometer) was evaluated. The results presented for sensitivity, specificity, precision, reproducibility, cross-reactivity, and interfering substances all represent the standalone performance of the Gold Standard Diagnostics Herpes Simplex Virus Type 1 IgG ELISA Test Kit. There is no human-in-the-loop component described for the operation or interpretation of this particular assay's results.
7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)
The primary type of ground truth used was comparison to a legally marketed predicate device (Focus Diagnostics HerpeSelect® 1 and 2 Immunoblot IgG test).
Additionally:
- For the cross-reactivity panel, the ground truth for specific infections was established by "FDA cleared tests" as reported by serum brokers.
- For the CDC seroconversion panel, the ground truth was based on "well-characterized HSV I serum samples" obtained from the CDC, indicating a highly validated and accepted reference standard.
8. The Sample Size for the Training Set
The document does not mention a training set in the context of an algorithm or machine learning. This is an ELISA kit, which is a biochemical assay, not an AI/ML algorithm that requires training data in the traditional sense. The development and optimization of the ELISA kit would involve internal validation and batch testing, but this is distinct from an AI training set.
9. How the Ground Truth for the Training Set was Established
As explained above, there is no "training set" for this type of device in the AI/ML context. The assay's performance characteristics are established through analytical and clinical studies comparing it to known standards (predicate device, CDC panel) and characterized samples.
Ask a specific question about this device
Page 1 of 2