510(k) Data Aggregation

    K Number
    K172414

    Validate with FDA (Live)

    Date Cleared
    2017-10-12

    (63 days)

    Product Code
    Regulation Number
    884.5300
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The JUD custom fit condom is used for contraception and for prophylactic purposes (to help prevent pregnancy and the transmission of sexually transmitted diseases).

    Device Description

    The JUD custom fit condom is made of a natural rubber latex sheath, which completely covers the penis with a closely fitted membrane. The condom is shaped with a reservoir end and is flat with a cylindrical shape. JUD custom fit condoms are available in 23 sizes of different length/width combinations. There is a fitting kit and instructions included as part of safe and effective usage of the condom.

    JUD custom fit condoms are provided pre-lubricated with a silicone-based lubricant. JUD custom fit condoms are not provided with spermicide, and contain no flavors or color additives. The user must use a fitting kit to select the appropriate size of the JUD custom fit condom.

    AI/ML Overview

    The document describes the JUD custom fit condom and its clearance via a 510(k) premarket notification. The study primarily relies on nonclinical bench testing and biocompatibility testing to demonstrate substantial equivalence to predicate devices, rather than a clinical study evaluating human-AI interaction or diagnostic accuracy.

    Here's an breakdown of the requested information based on the provided text:

    1. A table of acceptance criteria and the reported device performance

    Test DescriptionAcceptance Criteria (Standard)Reported Device Performance
    DimensionalISO 4074:2015Acceptance criteria met; dimensions conform to standards.
    Lubricant QuantityISO 4074:2015Acceptance criteria met.
    Burst Volume and PressureISO 4074:2015Acceptance criteria met; conform to standards.
    Freedom from Holes and Visible DefectsISO 4074:2015Acceptance criteria met; no visible defects identified.
    Package Seal IntegrityISO 4074:2015Acceptance criteria met; no damaged package seal identified.
    Shelf LifeMaintain specifications for 5 yearsEach condom variant maintained specifications for 5 years.
    CytotoxicityISO 10993-5:2009No cytotoxic reactivity found.
    Acute Systemic ToxicityISO 10993-11:2006No mortality or evidence of systemic toxicity.
    Vaginal IrritationISO 10993-10:2010Non-irritating; no signs of vaginal irritation.
    Skin SensitizationISO 10993-10:2010Did not induce skin sensitization.

    While not explicitly called "acceptance criteria" in the text, the "Summary and Conclusion" column effectively reports that the device met the requirements of the specified ISO standards, which represent the acceptance criteria for these tests.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document refers to "tested samples" for bench testing and a "single condom variant" for biocompatibility testing. Specific sample sizes are not provided in the excerpt for any of the nonclinical tests. The data provenance (country of origin, retrospective/prospective) is also not specified, though the manufacturer is located in Malaysia.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable as the evaluation relies on objective bench testing and biocompatibility rather than expert interpretation of a test set related to medical imaging or diagnostics.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable for the type of testing described (bench and biocompatibility).

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study was performed or described. This device is a condom, not an AI-assisted diagnostic tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This is not applicable as the device is a medical device (condom) and not an algorithm or AI system. The testing performed was standalone for the device itself (e.g., burst pressure of the condom).

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The "ground truth" for the bench tests was adherence to established international standards (ISO 4074:2015 and WHO 2010) for physical properties. For biocompatibility, the ground truth was biological response in animal models or in-vitro tests as defined by ISO 10993 series standards.

    8. The sample size for the training set

    This is not applicable as the device is not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established

    This is not applicable as the device is not an AI/ML algorithm that requires a training set.