Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    K Number
    K180188

    Validate with FDA (Live)

    Date Cleared
    2018-06-07

    (135 days)

    Product Code
    Regulation Number
    884.6120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Geri Embryo Incubator is intended to provide an environment with controlled temperature and mixed gas (CO2 and other gases) for the development of embryos. The Geri Embryo Incubator has an integrated camera and optics for imaging and viewing embryos during incubation, for a maximum time of 120 hours.

    Geri Connect and Geri Assess are optional software accessories for the Geri Embryo Incubator. Geri Connect is intended for access and review of time-lapse data generated by the Geri Embryo Incubator on a local area network. Geri Assess is intended for viewing and recording embryo development events from images captured using the Geri Embryo Incubator. Geri Assess includes a user annotation for capturing information on embryo development parameters and a userdefined modeling function that allows the user to combine annotated information on embryo development parameters to aid in embryo selection. Geri Connect and Geri Assess do not control any hardware components in the Geri Embryo Incubator. Geri Connect and Geri Assess are combined in the same software package and must be used together.

    The Geri Dish is intended to be used for preparation, storage and imaging of human embryos. The Geri Dish is intended to be used only with the Geri Embryo Incubator.

    Device Description

    The Geri Embryo Incubator is a benchtop incubator with six modular incubation chambers each with its own temperature control capability and separate gas inlet feed. Each chamber contains heating elements in its lid and base, together with an orange light source (591 nm) and camera with integrated optics that take time-lapse images of embryos and allows operators to view embryos without removing them from the incubation chamber. Inside each chamber is a filter the gas mixture entering the incubation chamber from the gas supply. The gas is supplied from standard premixed medical gas cylinders. The Geri Embryo Incubator includes firmware to control the incubator settings, and software to control patient information and settings.

    The Geri Assess and Geri Connect are optional software accessories for the Geri Embryo Incubator. The Geri Connect allows the user to access the embryo data remotely, whereas the Geri Assess provides the user with a tool for analysis of embryo data. Using user defined parameters, the Geri Assess can score the embryos. However, the Geri Assess itself does not include any pre-loaded scoring assessments or perform any diagnostic functions. The Geri Connect and Geri Assess software package is provided with the Geri Embryo Incubator, but needs to be unlocked, when purchased by the end user.

    The Geri Dish is intended to be used for preparation, storage and imaging of human embryos. Up to 16 embryos can be stored in one Geri Dish. Geri Dishes are supplied sterile with a sterility assurance level (SAL) of 10 %, and have a shelf-life of 12 months. The Geri Dish is intended to be used only with the Geri Embryo Incubator. Other assisted reproduction culture dishes may be used in the Geri Embryo Incubator in separate non-time-lapse positions located within each chamber.

    AI/ML Overview

    The provided text is a 510(k) summary for the Geri Embryo Incubator with Geri Connect and Geri Assess Software. It explains the device, its intended use, and compares it to a predicate device. However, it does not contain any information about acceptance criteria or specific study results that prove the device meets those criteria, especially in relation to AI/software performance in aiding embryo selection.

    The document states:

    • "Software verification and validation testing was conducted on the subject device in accordance with the FDA guidance document, 'Guidance for the Content of Premarket Submissions for Software in Medical Devices' issued on May 11, 2005."
    • "The Geri Assess itself does not include any pre-loaded scoring assessments or perform any diagnostic functions."
    • "Geri Assess includes a user annotation function for capturing information on embryo development parameters and a user-defined modelling function that allows the user to combine annotated information on embryo development parameters to aid in embryo selection."
    • "The features added by this additional software do not represent a new intended use, and are seen in other cleared devices of this type."
    • "The differences in technological characteristics between subject and predicate devices do not raise different questions of safety and effectiveness."

    Based on these statements, it appears the Geri Assess software provides tools for users to define and apply their own models for embryo selection, rather than itself being an AI that performs diagnostic functions or makes automated predictions based on pre-loaded algorithms. The regulatory submission likely focused on verifying the software's functionality, usability, and safety as a tool, rather than validating its performance in an AI-assisted diagnostic capacity.

    Therefore, it is not possible to extract the requested information (acceptance criteria, specific study results, sample sizes, expert details, MRMC studies, AI effect size, etc.) from the provided text because these elements are not present. The document concludes that the device is substantially equivalent based on the software's functionality as a user-defined tool, rather than as a diagnostic AI requiring extensive clinical performance studies as described in the prompt.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180304

    Validate with FDA (Live)

    Date Cleared
    2018-04-20

    (77 days)

    Product Code
    Regulation Number
    884.6120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Geri Embryo Incubator is intended to provide an environment with controlled temperature and mixed gas (CO2 and other gases) for the development of embryos. The Geri Embryo Incubator has an integrated camera and optics for imaging and viewing embryos during incubation, for a maximum time of 120 hours.

    The Geri Dish is intended to be used for preparation, storage and imaging of human embryos. The Geri Dish is intended to be used only with the Geri Embryo Incubator.

    Device Description

    The Geri Embryo Incubator is a benchtop incubator with six modular incubation chambers each with its own temperature control capability and separate gas inlet feed. Each chamber contains heating elements in its lid and base, together with an orange light source (591 nm) and camera with integrated optics that take time-lapse images of embryos and allows operators to view embryos without removing them from the incubation chamber. Inside each chamber is a filter the gas mixture entering the incubation chamber from the gas supply. The gas is supplied from standard premixed medical gas cylinders. Each chamber can also contain an optional water bottle (Geri Water Bottle) to generate humidity. The Geri Water Bottle is supplied sterile with a sterility assurance level (SAL) of 10°, and has a shelf-life of three years. The Geri Water Bottle is designed to be used with the Geri Embryo Incubator only.

    The embryos are maintained in Geri Dishes that are supplied separately. The Geri Dish is an optically-clear polystyrene dish designed to be compatible with the Geri Embryo Incubator. Up to 16 embryos can be stored in one Geri Dish. Geri Dishes are supplied sterile with a sterility assurance level (SAL) of 10°, and have a shelflife of 12 months.

    The Geri Embryo Incubator and Geri Dish are intended to be used together for embryo imaging purposes. The Geri Dish is not compatible with other embryo time-lapse incubators. However, other assisted reproduction culture dishes may be used in the Geri Embryo Incubator in separate non-time-lapse positions located within each chamber.

    The Geri Embryo Incubator includes firmware to control the incubator settings, and software to control patient information and settings.

    AI/ML Overview

    This document, a 510(k) Summary for the Geri Embryo Incubator and Geri Dish, outlines the characteristics and testing for a medical device. It is not a document about an AI/ML-based medical device. Therefore, none of the specific questions regarding AI/ML device acceptance criteria, study design for AI/ML, sample sizes for AI/ML test and training sets, expert adjudication methods for AI/ML ground truth, or MRMC studies for AI assistance, can be answered from the provided text.

    The document focuses on the substantial equivalence of the Geri Embryo Incubator and Geri Dish to a predicate device, with the main change being the addition of a humidity control system. The studies described are traditional non-clinical performance tests for an incubator and culture dish, such as electrical safety, EMC, cleaning validation, gas maintenance, temperature control, time-lapse imaging function, embryo development testing (mouse embryo assay), and sterilization validation.

    Therefore, I cannot provide the requested information as the provided text relates to a medical device that does not appear to be an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171736

    Validate with FDA (Live)

    Date Cleared
    2017-11-15

    (156 days)

    Product Code
    Regulation Number
    884.6120
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Geri Embryo Incubator is intended to provide an environment with controlled temperature and mixed gas (CO2 and other gases) for the development of embryos. The Geri Embryo Incubator has an integrated camera and optics for imaging and viewing embryos during incubation, for a maximum time of 120 hours.

    The Geri Dish is intended to be used for preparation, storage and imaging of human embryos. The Geri Dish is intended to be used only with the Geri Embryo Incubator.

    Device Description

    The Geri Embryo Incubator is a benchtop incubator with six modular incubation chambers each with its own temperature control capability and separate gas inlet feed. Each chamber contains heating elements in its lid and base, together with an orange light source (591 nm) and camera with integrated optics that take time-lapse images of embryos and allows operators to view embryos without removing them from the incubation chamber. Inside each chamber is a filter the gas mixture entering the incubation chamber from the gas supply. The gas is supplied from standard premixed medical gas cylinders.

    The Geri Embryo Incubator includes firmware to control the incubator settings, and software to control patient information and settings.

    The embryos are maintained in Geri Dishes that are supplied separately. The Geri Dish is an optically-clear polystyrene dish designed to be compatible with the Geri Embryo Incubator. Up to 16 embryos can be stored in one Geri Dish. Geri Dishes are supplied sterile with a sterility assurance level (SAL) of 10-9, and have a shelf-life of 12 months.

    The Geri Embryo Incubator and Geri Dish are intended to be used together for embryo imaging purposes. The Geri Dish is not compatible with other embryo time-lapse incubators. However, other assisted reproduction culture dishes may be used in the Geri Embryo Incubator in separate non-time-lapse positions located within each chamber.

    AI/ML Overview

    The provided text is a 510(k) summary for the Geri Embryo Incubator and Geri Dish. It details the device, its intended use, comparison to predicate devices, and a summary of non-clinical performance testing. However, it does not include detailed acceptance criteria tables or specific reported device performance values for the Geri Embryo Incubator and Geri Dish.

    Specifically, while it mentions "Bench performance tests that met design specifications described in Section VII" and lists some parameters like "Gas maintenance testing," "Temperature control testing," and "Time-lapse testing," it generally states that these tests "met design specifications" without providing the actual acceptance criteria values (e.g., specific CO2 levels, temperature ranges, or image resolution thresholds) and the precise device performance results.

    Therefore, I cannot create a table of acceptance criteria and reported device performance from the provided text. The document describes the types of tests performed and generally states that the device met the specifications, but it does not quantify those specifications or the observed performance.

    Similarly, the document provides limited information on most of the other requested criteria for a study proving the device meets acceptance criteria, as it is a 510(k) summary, not a detailed study report.

    Here's an assessment based on the available information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    • Cannot be created from the provided text. The document states that "Bench performance tests that met design specifications described in Section VII" and lists categories like "Gas maintenance testing," "Temperature control testing," and "Time-lapse testing." It also provides a specific acceptance criterion for the Geri Dish's Mouse Embryo Assay: "1-cell MEA ≥80% embryos developed to blastocyst in 96 hours." However, the actual numerical design specifications (acceptance criteria) and the measured device performance for gas, temperature, and time-lapse functions are not provided.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample size for test set: Not explicitly stated for most tests. For the mouse embryo assays, it implies a comparative group ("in comparison with the control group"), but the number of embryos or dishes tested is not given.
    • Data Provenance: Not specified (e.g., country of origin, retrospective/prospective). This is typically lab-based bench testing.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    • Not applicable / Not specified. This document describes performance testing for an incubator and dish, which are physical devices with measurable parameters (temperature, gas, imaging). The "ground truth" for these tests would be established through calibrated instruments and established scientific principles, not typically by expert consensus in the way, for example, medical image analysis AI models use expert radiologists. The "Embryo development test" and "Mouse embryo assay" are biological assays, where the "ground truth" is typically the observed biological outcome (blastocyst development), assessed by trained laboratory personnel, but no expert qualifications or numbers are mentioned.

    4. Adjudication Method for the Test Set:

    • Not applicable / Not specified. Adjudication is relevant for subjective assessments, particularly in clinical studies involving interpretation. The tests described are primarily objective benchtop and biological assays.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No. An MRMC study is relevant for AI-assisted diagnostic devices where human readers interpret medical images. This document describes a medical device (incubator and dish) that does not directly involve human image interpretation in its primary function, nor does it describe AI assistance to human readers.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance):

    • Not applicable. This device is hardware with integrated software for control and imaging; it's not a standalone AI algorithm for interpretation. Its performance is inherent to the device's function.

    7. Type of Ground Truth Used:

    • For physical parameters (gas, temperature, imaging functions): Measured values against design specifications using calibrated instruments.
    • For biological assays (mouse embryo development): Observed biological outcomes (e.g., blastocyst formation).

    8. Sample Size for the Training Set:

    • Not applicable. This is a hardware device with integrated control/imaging software, not a machine learning model that requires a "training set" in the conventional AI sense for learning patterns from data. The software mentioned ("Software verification and validation testing") refers to traditional software engineering V&V, not AI model training.

    9. How Ground Truth for Training Set Was Established:

    • Not applicable. (As explained above, no separate training set for an AI model is described.)
    Ask a Question

    Ask a specific question about this device

    K Number
    K170498

    Validate with FDA (Live)

    Date Cleared
    2017-09-28

    (223 days)

    Product Code
    Regulation Number
    884.6180
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Gems Oocyte Retrieval Buffer is used in the removal of oocytes from ovarian follicles.

    Device Description

    The Gems Oocyte Retrieval Buffer is an assisted reproduction medium used for removal of oocytes from ovarian follicles. It consists of salts, energy substrates, amino acids buffering agents, stabilizers and antibiotics. This device has direct contact with the patients and oocytes.

    This product is a single-use device that is aseptically filled into sterilized bottles and has a sterility assurance level (SAL) of 10-3. It is tested for pH, osmolality, embryotoxicity, endotoxin and sterility before lot release and has a shelf-life of 30 weeks.

    AI/ML Overview

    The provided text does not contain information about an AI device or a study comparing AI with human readers. It describes a submission for a medical device called "Genea Biomedx Gems Oocyte Retrieval Buffer ORB-20, ORB-50," which is a reproductive medium used for oocyte retrieval.

    Therefore, I cannot provide the requested information regarding acceptance criteria and studies related to an AI device or human-in-the-loop performance. The document focuses on demonstrating the substantial equivalence of the Oocyte Retrieval Buffer to a legally marketed predicate device through non-clinical performance testing.

    However, I can extract the acceptance criteria and a summary of the non-clinical performance testing for the described Genea Biomedx Gems Oocyte Retrieval Buffer device.

    Acceptance Criteria and Reported Device Performance (Table for the Oocyte Retrieval Buffer)

    Acceptance CriteriaReported Device Performance (at release and end of shelf-life)
    pH7.3-7.5 (comparable to predicate)
    Osmolality280-290 mOsm/kg (comparable to predicate)
    1-cell MEA (Mouse Embryo Assay)≥80% developed to the blastocyst stage at 96 hours
    Endotoxin<0.4 EU/ml (LAL)
    SterilityNo microbiological growth (Sterility Assurance Level (SAL) of 10-3)
    CytotoxicityMet requirements per ISO 10993-5:2009
    SensitizationMet requirements per ISO 10993-10:2010 (Guinea Pig Maximization)
    Intracutaneous ReactivityMet requirements per ISO 10993-10:2010
    Aseptic ProcessingMet requirements in ISO 13408-2:2003

    Study Details (for the Oocyte Retrieval Buffer):

    The document describes non-clinical performance testing to support the substantial equivalence of the "Genea Biomedx Gems Oocyte Retrieval Buffer ORB-20, ORB-50" to a predicate device. This is not a study involving AI or human readers.

    1. Sample size used for the test set and the data provenance: The document does not specify a "test set" in the context of an AI algorithm. For the MEA (Mouse Embryo Assay), "one-cell mouse embryos" were used. The exact number of embryos or experimental replicates is not provided in this summary. The data provenance (country of origin, retrospective/prospective) for these non-clinical tests is not detailed, though Genea Biomedx Pty Ltd is based in Australia.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. This is not an AI study requiring expert ground truth for image interpretation or similar tasks.

    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.

    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. No AI component is described.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. No AI algorithm is described.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For the performance tests listed, the "ground truth" or reference was based on established scientific and regulatory standards/methods (e.g., USP <71> for sterility, USP <85> for endotoxin, ISO standards for biocompatibility, and a pre-defined blastocyst development rate for MEA).

    7. The sample size for the training set: Not applicable. This document pertains to a medical product, not an AI model requiring a training set.

    8. How the ground truth for the training set was established: Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162409

    Validate with FDA (Live)

    Date Cleared
    2017-05-24

    (268 days)

    Product Code
    Regulation Number
    884.6180
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Gems Vitrification Set is used for the vitrification of human blastocyst stage embryos for Assisted Reproductive Technology (ART) procedures.

    Gems Warming Set is used for the warming of human blastocyst stage embryos that have undergone vitrification.

    Device Description

    The Gems Vitrification Set and Gems Warming Set are intended for the vitrification and warming of human blastocysts as part of human ART procedures.

    The Gems Vitrification Set is designed to facilitate dehydration of blastocysts before vitrification via rapid cooling in liquid nitrogen. Dehydration of the blastocysts is achieved by the step-wise use of increasing concentrations of cryoprotectants in the Gems Vitrification Set, which results in water being withdrawn from the cell. The cryoprotectants also protect the blastocysts by reducing the potential for ice crystal formation during the vitrification process.

    The Gems Vitrification Set consists of three solutions (Vitsol 2, and Vitsol 3). The base formulation for Vitsol 1 and Vitsol 2 is a HEPES-buffered medium containing salts, energy substrates, amino acids, and human serum albumin. Vitsol 1 contains the cryoprotectant ethylene qlycol, while Vitsol 2 contains the cryoprotectants ethylene glycol and trehalose. Vitsol 3 consists of dimethyl sulphoxide (DMSO). The Gems Vitrification Set is not provided ready for use as Vitsol 3 must be added to the other vitrification solutions before use. Following DMSO addition to Vitsol 1 and 2, the final vitrification solutions for blastocyst vitrification procedures have the following properties:

    • . Vitrification Solution 1 (Vitsol 1 + DMSO) - 8% ethylene glycol and 8% DMSO
    • Vitrification Solution 2 (Vitsol 2 + DMSO) 16% ethylene glycol, 16% DMSO, and 0.57M trehalose ●

    The Gems Warming Set consists of three warming solutions (Warmsol 2, and Warmsol 3), which are designed to facilitate the re-hydration (warming) of vitrified blastocysts. In the warming process, trehalose in the media manages the inflow of water into blastocysts as concentrations of DMSO and ethylene glycol are reduced during the rehydration process.

    All three of the solutions in the Gems Warming Set consist of a HEPES-buffered medium containing salts, energy substrates, amino acids, and human serum albumin, with varying amounts of the cryoprotectant trehalose as described below:

    • . Warmsol 1 - 1.0 M trehalose
    • Warmsol 2 0.5 M trehalose ●
    • Warmsol 3 - No trehalose

    These media are single-use devices that are aseptically filled into sterilized bottles and have a sterility assurance level (SAL) of 10 3. The products are tested for pH, osmolality, embryotoxicity, endotoxin, and sterility before lot release.

    AI/ML Overview

    The provided document describes the acceptance criteria and performance testing for the "Gems Vitrification Set and Gems Warming Set" (subject device) in comparison to a predicate device, the "Cook Sydney IVF Blastocyst Vitrification Kit and Cook Sydney IVF Blastocyst Warming Kit" (K143724).

    Here's an analysis of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria / ParameterSubject Device - Gems Vitrification Set and Gems Warming Set (K162409) Reported PerformancePredicate Device - COOK Sydney IVF Blastocyst Vitrification Kit and COOK Sydney IVF Blastocyst Warming Kit (K143724)Comparison
    Intended UseVitrification and warming of human blastocyst stage embryos for ART procedures.Vitrification and warming of human blastocysts for ART procedures.Same
    FormulationHEPES buffered physiologic media containing ethylene glycol, DMSO, trehalose, and human serum albumin in addition to normal physiological salts. DMSO provided separately.HEPES buffered physiologic media containing ethylene glycol, DMSO, trehalose, human Serum albumin and gentamicin in addition to normal physiological salts. DMSO provided separately.Similar (differences in gentamicin and some concentrations, deemed not to raise different S&E questions)
    pH7.3-7.57.3-7.5Same
    Osmolarity (mOsm/kg)Warmsol 1: 1280-1320; Warmsol 2: 780-820; Warmsol 3: 295-305 (N/A for vitrification solutions)K-SIBW-SOL1: 657-683; K-SIBW-SOL2: 500-520; K-SIBW-SOL3: 285-295 (Other values for vitrification solutions not directly comparable)Different (higher in subject device for warming solutions due to cryoprotectant concentrations, deemed not to raise different S&E questions)
    Mouse Embryo Assay (MEA)1-Cell MEA: ≥80% developed to blastocysts at 96h2-Cell MEA: ≥80% blastocyst formation at 72hSimilar
    EndotoxinVitsol 1-2, Warmsol 1-3: < 0.40 EU/mL; DMSO: <0.05 EU/ml< 0.40 EU/mLSimilar
    Sterilization MethodAseptic Filtration, SAL 10⁻³Aseptic FiltrationSame
    Shelf-Life20 weeks20 weeksSame

    2. Sample size(s) used for the test set and the data provenance

    The document does not specify exact sample sizes for all non-clinical tests. However, for the Mouse Embryo Assay (MEA), it states: "One-cell mouse embryos were exposed sequentially to subject devices in the Vitrification and Warming Sets followed by culture at 37°C in an atmosphere containing 5% CO₂. The percent of embryos developed to the expanded blastocyst stage at 96 hours were assessed in comparison with the control group."

    The provenance of this data is a non-clinical study, likely conducted internally by Genea Biomedx Pty Ltd, or by a contract research organization. There is no mention of country of origin of the data or if it was retrospective or prospective in the context of human embryos. Given it's a non-clinical study using mouse embryos, these distinctions are less relevant than for human clinical trials.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. The studies are non-clinical (MEA, chemical analyses, sterility, endotoxin) and likely rely on laboratory measurements and protocols rather than expert consensus for ground truth on individual test items. For the MEA, the "ground truth" would be the observed development of mouse embryos to the blastocyst stage.

    4. Adjudication method for the test set

    This information is not provided. Given the nature of the non-clinical tests (quantitative measurements and adherence to specified limits), a formal adjudication method by experts is not typically applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This document describes the testing of reproductive media (Gems Vitrification Set and Gems Warming Set), which are medical devices used for in-vitro fertilization procedures, not imaging or diagnostic AI software that would typically involve human readers. Therefore, the concept of "human readers improve with AI vs without AI assistance" is not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable as the device is not an AI algorithm. It is a set of chemical media. The performance assessed is the biological effectiveness (embryo development) and chemical/physical characteristics of the media themselves in standalone non-clinical tests.

    7. The type of ground truth used

    The ground truth for the non-clinical performance testing involved:

    • Quantitative laboratory measurements: For pH, osmolality, endotoxin levels, and sterility. These are objective measurements against established scientific standards and limits.
    • Biological endpoint observation: For the Mouse Embryo Assay (MEA), the ground truth is the observed development of 1-cell mouse embryos to the expanded blastocyst stage within 96 hours. This is an objective biological outcome.
    • Validation against standards: Aseptic Processing Validation met requirements in ISO 13408-2:2003. Sterility testing was per USP <71> and Endotoxin testing per USP <85>.

    8. The sample size for the training set

    This information is not applicable. The device is not an AI algorithm that requires a training set. The descriptions relate to the manufacturing and quality control testing of physical chemical media.

    9. How the ground truth for the training set was established

    This information is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161261

    Validate with FDA (Live)

    Date Cleared
    2017-05-12

    (373 days)

    Product Code
    Regulation Number
    884.6180
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Gems Fertilisation Medium is used to provide a suitable environment for both oocytes and sperm and support in vitro fertilisation. This medium can be used for transfer of zygotes into the uterus.

    Gems Cleavage Medium is for in vitro culture of embryos following fertilization to the 4-8 cell stage. This medium can be used for transfer of cleavage stage embryos into the uterus.

    Gems Blastocyst Medium is for in vitro culture of embryos from the cleavage stage to the blastocyst stage of development. This medium can be used for transfer of blastocyst stage embryos into the uterus.

    Gems Geri Medium is for in vitro culture of embryos from fertilization to the blastocyst stage of development. This medium can be used for transfer of embryos into the uterus.

    Gems VitBase is used to maintain embryos for a short period of time in a non-gassed environment during embryo vitrification and warming procedures. The embryos can be placed in this medium for a maximum of 10 minutes.

    Device Description

    The subject devices are culture media consisting of salts, energy substrates, amino acids, buffering agents, nutrients supplements and antibiotics, with or without L-Carnitine and/or human serum albumin. These devices have different applications in Assisted Reproduction Technology (ART) procedures in a hospital environment. These media are single-use devices that are aseptically filled into the sterilized bottles and have a sterility assurance level (SAL) of 10-3. The products are tested for pH, osmolality, embryotoxicity, endotoxin, and sterility before lot release.

    AI/ML Overview

    This document describes the Genea Biomedx Genea Embryo Culture Media and its associated performance testing for substantial equivalence to predicate devices, as reviewed by the FDA.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for several key parameters are presented, primarily for shelf-life studies and comparison to predicate devices/literature.

    ParameterAcceptance Criteria (Internal/Comparative)Reported Device Performance
    pHVaries by medium (e.g., Gems Fertilisation: 7.5-7.7, Gems Cleavage: 7.4-7.6, Gems Blastocyst: 7.6-7.80, Gems Geri: 7.4-7.6, Gems VitBase: 7.3-7.5)Similar to predicate devices, met internal specifications. Included in shelf-life studies.
    OsmolalityVaries by medium (e.g., Gems Fertilisation: 295-305 mOsm/kg, Gems Cleavage: 285-295 mOsm/kg, Gems Blastocyst: 285-295 mOsm/kg, Gems Geri: 285-295 mOsm/kg, Gems VitBase: 295-305 mOsm/kg)Similar to predicate devices, met internal specifications. Included in shelf-life studies.
    1-cell MEA (Mouse Embryo Assay)≥80% developed to the blastocyst stage at 96 hoursMet acceptance criteria (implied by "performance data demonstrate that the subject devices are substantially equivalent"). Included in shelf-life studies.
    Endotoxin<0.4 EU/ml (LAL)Met acceptance criteria (implied). Included in shelf-life studies.
    SterilityNo microbiological growthMet acceptance criteria (implied). Included in shelf-life studies.
    Pregnancy Rate (Primary Clinical Endpoint)Comparable to CDC data for relevant age groups (comparator)Comparable to CDC data.
    Fertilization Rate (Secondary Clinical Endpoint)Comparable to published literature (comparator)Comparable to published literature.
    Cleavage Rate (Secondary Clinical Endpoint)Comparable to published literature (comparator)Comparable to published literature.
    Blastulation Rate (Secondary Clinical Endpoint)Comparable to published literature (comparator)Comparable to published literature.
    Live Birth Rate (Additional Clinical Endpoint)Comparable to CDC data for relevant age groups (comparator)Comparable to CDC data.
    BiocompatibilityPer ISO 10993 standards (e.g., cytotoxicity, sensitization, intracutaneous reactivity, genotoxicity)All tests met the requirements (implied by non-clinical testing summary).
    Aseptic ProcessingMet requirements in ISO 13408-2:2003Met requirements.

    2. Sample Sizes Used for the Test Set and Data Provenance

    A single-site clinical trial was conducted for performance testing.

    • Test Set Sample Size:
      • Embryos:
        • Test Group (Subject Devices): 641 embryos used
        • Control Group (Genea in-house media): 598 embryos used
      • Embryo Transfer Procedures:
        • Test Group: 569 procedures
        • Control Group: 531 procedures
      • The sample sizes for MEA, endotoxin, sterility, and biocompatibility tests are not explicitly stated, but standard procedures typically involve sufficient numbers for statistical validity.
    • Data Provenance: The clinical trial was conducted at the Genea Clinic (Sydney, Australia). The study was prospective as it was a controlled, double-blinded, single-site clinical trial designed to evaluate safety and effectiveness.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not specify the number or qualifications of experts involved in establishing the ground truth for the clinical trial endpoints (pregnancy rate, fertilization rate, cleavage rate, blastulation rate, live birth rate). These outcome measures are typically objectively determined through established medical protocols and laboratory analyses in an IVF clinic setting. The "ground truth" for clinical outcomes in this context would be the objective clinical data recorded (e.g., ultrasound confirmation of fetal heartbeat, successful fertilization observed microscopically, embryo development stages).

    4. Adjudication Method for the Test Set

    The document states it was a "controlled, double-blinded, single-site clinical trial." This implies that neither the patients nor the clinicians/researchers directly involved in assessing outcomes were aware of whether the specific subject device or the control media was used. This blinding inherently serves as a form of adjudication against bias in outcome assessment. However, a specific "adjudication method" (like 2+1 reads for imaging) is not applicable or described for these types of clinical outcomes.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. This type of study is typically relevant for interpretative diagnostic devices where human reader performance is a direct output. This document describes embryo culture media, and its effectiveness is measured by biological outcomes, not human interpretation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    This question is not applicable. The device is a "Reproductive Media and Supplements" – an in-vitro diagnostic/culture medium, not an algorithm or AI device. Its performance is inherent to its biochemical properties and biological effect on embryos, assayed through laboratory and clinical endpoints.

    7. The Type of Ground Truth Used

    The ground truth for the clinical study was based on:

    • Outcomes Data:
      • Primary Endpoint: Fetal heartbeat at 6-8 weeks post-embryo transfer (objective clinical outcome).
      • Secondary Endpoints: Fertilization rate, cleavage rate, and blastulation rate (observable biological outcomes assessed in the lab).
      • Additional Analysis: Live birth rate (objective clinical outcome).
    • Comparative Data: The study compared its clinical outcomes to:
      • Published literature for fertilization, cleavage, and blastulation rates.
      • Assisted Reproductive Technology National Summary Report published by the Centers for Disease Control and Prevention (CDC) for pregnancy and live birth rates.
      • Genea in-house media in the control group for direct comparison within the trial.

    8. The Sample Size for the Training Set

    This information is not provided. The document describes pre-market notification (510(k)) for substantial equivalence, which focuses on demonstrating safety and effectiveness for a device, typically through testing and comparison, rather than an "algorithm training" process. The media formulations and manufacturing processes would have been developed and optimized, but not necessarily through a formal "training set" in the sense of machine learning algorithms.

    9. How the Ground Truth for the Training Set Was Established

    This information is not provided. As mentioned above, the concept of a "training set" with established "ground truth" in the statistical learning sense is not directly applicable to the development of embryo culture media as described in this document. The development and optimization of such media would involve extensive R&D, experimentation, and quality control, ensuring optimal conditions for embryo development, but this is distinct from establishing ground truth for a machine learning model.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1