Search Results
Found 5 results
510(k) Data Aggregation
(39 days)
The Super Upper Limbs Versalock Plating System is intended for fracture fixation, arthrodesis, reconstruction, and osteotomy fixation of the hand, wrist and forearm. The use of locking plate/screw systems is suited for treatment of fractures in osteopenic bone.
The Super Upper Limbs Versalock Plating System consists of plates and screws in a variety of designs and sizes that are commonly used in trauma and reconstructive surgery. The bone plates and screws are made from commercial titanium alloy and pure titanium. The plates range in thickness from 0.6 to 4.0 mm, and the screws range in diameter from 2.0 to 3.5 mm.
Super Upper Limbs Versalock Plating System implants are manufactured with the following raw materials:
- Titanium Alloy Ti6Al4V according to ASTM F136
- Pure Titanium according to ASTM F67
N/A
Ask a specific question about this device
(58 days)
Meniscus Versaflex is indicated for repair of the meniscus.
Meniscus Versaflex - All Inside Meniscal Suture with Bendable Tip is an arthroscopy suture, non-absorbable, assembled in an insertion device. The inserter mechanism can be molded for better access to the posterior areas, medium region and anterior third of meniscus, with use of a bending tool and the options (models) of general use and reverse. Each device has two implants of non-absorbable polymer PEEK, previously connected with a UHMWPE suture and pre-assembled in an insertion system. The inserter is offered in a curved and reverse curved configuration. It is provided in sterile condition sterilized by Ethylene Oxide.
N/A
Ask a specific question about this device
(74 days)
The GMReis Suture Anchors is intended to be used for soft tissue reattachment in the elbow, shoulder, hand, wrist, foot, ankle, knee, and hip. Specific indications are listed below:
Elbow: Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction.
Shoulder: Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction.
Hand/Wrist: Scapholunate Ligament Reconstruction, Repair/Reconstruction of collateral ligaments, Repair of Flexor and Extensor Tendons at the PIP, DIP and MCP joints for all digits, tendon transfers, Carpal Ligament Reconstruction and Carpometacarpal joint arthroplasty (basal thumb joint arthroplasty).
Foot/Ankle: Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Metatarsal Ligament Repair, Hallux Valgus reconstruction, tendon transfers, Mid-foot reconstruction.
Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis.
Hip: Capsular repair, acetabular labral repair.
The Suture Anchors, manufactured by GM dos Reis Industria e Comercio Ltda, are anchor-type implants, and are manufactured in PEEK (polyetheretherketone). The anchors are made available preloaded in a disposable inserter device, composed of a metal rod and a polymeric handle. The implant and disposable inserter devices are unique products and cannot be sold separately. Implants of the Suture Anchors are classified as surgically invasive, non-absorbable, and non-active implantable medical devices. In addition, the implants, as well as the inserter device, are for single-use and sold in sterile form by Ethylene Oxide.
N/A
Ask a specific question about this device
(238 days)
Stitch Cerclage - Suture Tapes are indicated in orthopedic surgeries as bone fixation cerclage, specifically:
- Sternotomy indications including the "rewiring" of osteomized sternums;
- Repair of long bone fractures due to trauma or reconstruction.
Stitch Cerclage - Suture Tapes are indicated in orthopedic surgeries for internal fixation of discontinued bone structures. The product are not to be used in conjunction with other bone fixation implants, like plates. The surgical sutures and tapes are made of UHMWPE (ultra-high molecular weight polyethylene), the needles are made in AISI300 steel series, according to the standard ASTM F899 and cerclage is assembled on an ABS loader, additional instruments may be used as well. Those materials are identical to those used in other cleared GMReis devices like the reference device cleared under K223114. It is provided in sterile condition sterilized by Ethylene Oxide.
Based on the provided FDA 510(k) clearance letter for the Stitch Cerclage - Suture System, here's a breakdown of the acceptance criteria and the study that proves the device meets them:
Important Note: The provided document is an FDA 510(k) clearance letter, which means the device has already been cleared. This letter summarizes the regulatory decision and mentions that studies were performed, but it does not contain the detailed study protocols, results, or data itself. Therefore, I will extract all available information and explicitly state when information is not present in this document.
Acceptance Criteria and Device Performance Study Summary
The acceptance criteria for this device are largely implied through the comparison to a predicate device and the successful completion of specified mechanical tests. The performance study focused on demonstrating substantial equivalence to the predicate device, particularly in mechanical properties relevant to its intended use as a bone fixation cerclage.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Implied) | Reported Device Performance |
|---|---|
| Material Equivalence: Made of materials safe and equivalent to predicate devices. | The device is made of UHMWPE (ultra-high molecular weight polyethylene) for sutures/tapes and AISI300 steel series for needles, which are identical to those used in other cleared GMReis devices (Reference Device K223114) and are considered equivalent to the predicate device. |
| Design Equivalence: Shares equivalent design characteristics with predicate devices. | The device has “equivalent design characteristics as well as physical dimensions” compared to the predicate. Minor differences "do not raise new issues of safety or efficacy." |
| Mechanical Performance: Performance is comparable or superior to the predicate device in relevant mechanical tests (chemical composition, creep, tension static, fatigue). | Performance was demonstrated through mechanical testing (chemical composition, creep test, tension static, and fatigue tests). The specific results are not provided in this document, but the conclusion states: "We therefore consider the subject device equivalent to its predicate device in the mechanical tests performed." The methodology was based on the article "Cerclage Performance Analysis – a Biomechanical Comparison of Different Techniques and Materials", L. M. Hägerich, et al, Musculoskeletal Disorders (2022). |
| Sterility: Provided in a sterile condition via a validated method. | The device is provided in sterile condition, sterilized by Ethylene Oxide, which is a common and accepted sterilization method. |
| Biocompatibility: Materials are biocompatible for implantation. | While not explicitly detailed as a separate test, the use of UHMWPE and AISI300 steel, which are "identical to those used in other cleared GMReis devices" and are common in medical devices, implies that their biocompatibility has been previously established and met. |
| Substantial Equivalence: Overall device is substantially equivalent to legally marketed predicate devices in terms of indications for use, technological characteristics, and safety/effectiveness. | The submission concludes that the device is "substantially equivalent" to the predicate in "intended use, indications for use, material composition, anatomical region, multiple sizes, and basic design features." The mechanical testing supported this claim, with no new issues of safety or effectiveness identified. |
2. Sample Sizes Used for the Test Set and Data Provenance
- Sample Size: The document does not specify the exact sample sizes (e.g., number of test specimens, number of cycles for fatigue) used for the mechanical testing. It mentions "chemical composition, creep test, tension static and fatigue tests."
- Data Provenance: The studies were conducted by the manufacturer, GM Dos Reis Industria e Comercio Ltda, which is based in Campinas, Brazil. The data is based on mechanical testing, not clinical data, and therefore is not characterized as retrospective or prospective in the usual clinical sense. It's laboratory-based performance data.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
- For this type of mechanical device clearance, "ground truth" is established through engineering standards, validated test methodologies, and comparison to the known performance of predicate devices.
- The document does not mention the use of human experts (like radiologists) for "ground truth" establishment as would be common for AI/imaging device clearances. The "experts" involved would be engineering and quality control personnel at the manufacturer and potentially external testing laboratories, ensuring compliance with test methodologies.
4. Adjudication Method for the Test Set
- Since this is a mechanical device performance study, no explicit "adjudication method" in the sense of clinician consensus on medical images is applicable or mentioned. The "adjudication" is inherent in the adherence to established mechanical testing protocols and statistical analysis of the results to demonstrate equivalence to the predicate.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- No MRMC study was done. This type of study is specifically relevant for imaging devices or AI algorithms where human readers' performance is being evaluated or augmented. The Stitch Cerclage Suture System is a physical medical device, not an imaging or AI diagnosis system.
- The document explicitly states: "No clinical data were included in this submission."
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance)
- Not applicable. This concept pertains to AI algorithms. The Stitch Cerclage Suture System is a physical product and does not have an "algorithm-only" performance component.
7. Type of Ground Truth Used
- The "ground truth" for this device type is established through objective, quantifiable mechanical properties and material composition, compared against established industry standards (where available) or the performance of a legally marketed predicate device.
- Specifically, the study compared "chemical composition, creep test, tension static and fatigue tests" to information from a published article ("Cerclage Performance Analysis – a Biomechanical Comparison of Different Techniques and Materials", L. M. Hägerich, et al, Musculoskeletal Disorders (2022)) and to the predicate device.
8. Sample Size for the Training Set
- Not applicable / Not mentioned. This device is a physical product, not an AI algorithm. Therefore, there is no "training set" in the context of machine learning. The design and manufacturing process would involve internal validation and verification activities, but this is distinct from training an AI model.
9. How the Ground Truth for the Training Set Was Established
- Not applicable. As there is no AI "training set," there is no ground truth established for one. The "ground truth" for the device's design and manufacturing is derived from engineering principles, relevant material standards (e.g., ASTM F899 for steel), and performance expectations set by the predicate device.
Ask a specific question about this device
(88 days)
The EXPERT - Flexible Joint Fixation System is intended as an adjunct in fracture repair involving metaphyseal and periarticular bone fragments where screws are not indicated, and as an adjunct in external and intramedullary fixation systems involving plates and nails, with fracture braces and casting. The EXPERT - Flexible Joint Fixation System are intended to provide fixation during the healing process following:
EXPERT FAST - SYNDESMOSIS and EXPERT FAST DUAL:
Syndesmotic trauma, such as fixation of syndesmosis disruptions) in connection with Weber B and C ankle fractures.
EXPERT ACL and EXPERT ACL II:
Fixation of bone to bone or soft tissue to bone, and is intended as fixation posts, a distribution bridge, or for distributing suture tension over areas of ligament or tendon repair. Specifically, these models are offered for Anterior Cruciate Ligament (ACL) Repair.
EXPERT FAST - CORACOID PLATE and EXPERT FAST- AC:
Specifically, these models are intended to provide fixation during the healing process following a syndesmotic trauma, such as fixation of acromioclavicular separations due to coracoclavicular ligament disruptions.
The EXPERT - Flexible Joint Fixation System was designed for fixation, being used as a biomechanical structure to simulate a support pillar distributing the tensions in the suture wire or tape. The plates are manufactured in titanium alloy conforming to ASTM F136. The sutures are manufactured from UHMWPE. The devices are sold sterile and are single use. The EXPERT - Flexible Joint Fixation System is a line extension to the EXPERT - Joint Fixation System consisting of new models.
The provided text describes a medical device, the "EXPERT - Flexible Joint Fixation System," and its 510(k) summary for FDA clearance. However, the document does not contain information about acceptance criteria or a study that proves the device meets specific performance criteria in the way a clinical or algorithmic performance study would.
The document discusses equivalence to predicate devices based on "submitted testing data" but does not detail the nature of these tests, specific acceptance criteria, or the results. It mentions:
- Performance Data: "Based on submitted testing data, the proposed EXPERT - Flexible Joint Fixation System is equivalent to the Arthrex ACL TightRope (K100652) and TightRopeTM Acromioclavicular (K052776) predicate devices." (Page 6)
- Conclusion: "Any differences between the proposed device and the predicate device are considered minor and do not raise new or different questions concerning safety or effectiveness." (Page 6)
This refers to engineering testing or benchtop studies to demonstrate the mechanical equivalence or biocompatibility of the device's components to previously cleared predicate devices, rather than a study evaluating diagnostic or prognostic capabilities of an AI/ML algorithm.
Therefore, I cannot fulfill the request to provide a table of acceptance criteria and reported device performance related to a study proving the device meets these criteria in the context of AI/ML or clinical performance, nor can I provide information on sample sizes, ground truth establishment, expert qualifications, or MRMC studies for this particular device based on the provided text.
The information requested, such as sample size for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, type of ground truth, training set size, and how training set ground truth was established, is not present in the provided FDA 510(k) summary for the EXPERT - Flexible Joint Fixation System. This is typical for a 510(k) for a physical medical device like a fixation system, which focuses on substantial equivalence to predicate devices through technical, material, and mechanical comparisons, rather than AI/ML algorithm performance.
Ask a specific question about this device
Page 1 of 1