510(k) Data Aggregation

    K Number
    K191957

    Validate with FDA (Live)

    Date Cleared
    2019-10-21

    (90 days)

    Product Code
    Regulation Number
    866.3975
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BD MAX Vaginal Panel performed on the BD MAX System is an automated qualitative in vitro diagnostic test for the direction of DNA targets from bacteria associated with bacterial vaginosis (qualitative results reported based on detection and quantitation of targeted organism markers), Candida species associated with vulvovaginal candidiasis, and Trichomonas vaginalis from vaginal swabs in patients who are symptomatic for vaginitis/vaginosis. The test utilizes real-time polymerase chain reaction (PCR) for the amplification of specific DNA targets and utilizes fluorogenic target-specific hybridization probes to detect and differentiate DNA from:

    • . Bacterial vaginosis markers (Individual markers not reported)
      Lactobacillus spp. (L. crispatus and L. jensenii) Gardnerella vaginalis Atopobium vaginae Bacterial Vaginosis Associated Bacteria-2 (BVAB-2) Megasphaera-1
    • . Candida spp. (C. albicans, C. tropicalis, C. parapsilosis, C. dubliniensis)
    • Candida glabrata
    • Candida krusei
    • . Trichomonas vaginalis

    The BD MAX Vaginal Panel is intended to aid in the diagnosis of vaginal infections in women with a clinical presentation consistent with bacterial vaginosis, vulvovaginal candidiasis and trichomoniasis.

    Device Description

    The BD MAX System and the BD MAX Vaginal Panel are comprised of an instrument with associated hardware and accessories, disposable microfluidic cartridges, master mixes, unitized reagent strips, and extraction reagents. The instrument automates sample preparation including target lysis, DNA extraction and concentration, reagent rehydration, target nucleic acid amplification and detection using real-time PCR. The assay includes a Sample Processing Control (SPC) that is present in the Extraction Tube. The SPC monitors DNA extraction steps, thermal cycling steps, reagent integrity and the presence of inhibitory substances. The BD MAX System software automatically interprets test results. For the BD MAX Vaginal Panel, a test result may be called as POS, NEG or UNR (Unresolved) based on the amplification status of the targets and of the Sample Processing Control. IND (Indeterminate) or INC (Incomplete) results are due to BD MAX System failure.

    AI/ML Overview

    This is an FDA 510(k) summary for the BD MAX Vaginal Panel for detecting microorganisms associated with vaginitis and bacterial vaginosis. The document describes the device, its intended use, and its equivalence to a predicate device. While it mentions performance standards, it does not contain detailed acceptance criteria, device performance data, information on the study design (sample size, data provenance, ground truth establishment, expert qualifications, adjudication methods), or any multi-reader multi-case (MRMC) study results.

    Therefore, many of the requested fields cannot be filled from the provided text.

    Here's a breakdown of what can be extracted and what is missing:

    1. Table of acceptance criteria and the reported device performance:

    • Acceptance Criteria: Not explicitly stated in terms of specific performance metrics (e.g., sensitivity, specificity, accuracy thresholds). The document broadly refers to "Performance Standards" based on the Class II Special Controls Guideline for NAAT assays for Trichomonas vaginalis.
    • Reported Device Performance: Not provided in this document. This summary focuses on substantial equivalence based on technological characteristics and intended use.

    2. Sample size used for the test set and the data provenance: Not mentioned.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience): Not mentioned.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not mentioned.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is a molecular diagnostic test, not an AI-assisted imaging device or a test involving human readers in the interpretation loop.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: The device is a standalone molecular diagnostic assay. The results are "automatically interpreted" by the BD MAX System software, which indicates algorithm-only performance. However, specific standalone performance metrics (sensitivity, specificity etc.) are not provided.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not mentioned for the studies supporting this 510(k) submission. For molecular tests, ground truth typically involves culture or a validated composite reference method.

    8. The sample size for the training set: Not mentioned.

    9. How the ground truth for the training set was established: Not mentioned.

    Summary Table of Available Information:

    SectionInformation from Text
    1. Acceptance Criteria and Reported Device PerformanceAcceptance Criteria: Not explicitly stated as numerical thresholds. Refers to "Class II Special Controls Guideline: Nucleic Acid Amplification Assays for the Detection of Trichomonas vaginalis, August 4, 2015." Reported Device Performance: Not provided in this summary document.
    2. Sample size and data provenance for the test setNot mentioned.
    3. Number of experts and qualifications for ground truth (test set)Not mentioned.
    4. Adjudication method (test set)Not mentioned.
    5. MRMC Comparative Effectiveness Study (human with/without AI assist)Not applicable, as this is a molecular diagnostic test. It's an automated system, not an AI for image interpretation or decision support for human readers.
    6. Standalone (algorithm only) performance studyYes, the device itself is a standalone, automated system. The "BD MAX System software automatically interprets test results." However, specific performance metrics (e.g., sensitivity, specificity, PPV, NPV) from a standalone study are not provided in this document.
    7. Type of ground truth usedNot mentioned for the performance studies themselves. For molecular diagnostics, this typically refers to a gold standard like culture or a composite reference method.
    8. Sample size for the training setNot mentioned.
    9. How the ground truth for the training set was establishedNot mentioned.
    Ask a Question

    Ask a specific question about this device

    K Number
    K132822

    Validate with FDA (Live)

    Date Cleared
    2013-11-26

    (78 days)

    Product Code
    Regulation Number
    866.1640
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BD MAX™ StaphSR assay performed on the BD MAX™ System is an automated qualitative in vitro diagnostic test for the direct detection and differentiation of Staphylococcus aureus (SA) and methicillin-resistant Staphylococcus aureus (MRSA) DNA from nasal swabs in patients at risk for nasal colonization. The test utilizes realtime polymerase chain reaction (PCR) for the amplification of MRSA/SA DNA and fluorogenic target-specific hybridization probes for the detection of the amplified DNA. The BD MAX™ StaphSR assay is intended to aid in the prevention and control of MRSA and SA infections in healthcare settings. It is not intended to diagnose MRSA or SA infections nor guide or monitor treatment for MRSA/SA infections. A negative result does not preclude nasal colonization. Concomitant cultures are necessary to recover organisms for epidemiological typing or for further susceptibility testing.

    Device Description

    The BD MAX™ System and the BD MAX™ StaphSR assay are comprised of an instrument with associated hardware and accessories, disposable microfluidic cartridges, master mixes, unitized reagent strips, extraction reagents, and sample buffer tubes. The instrument automates sample preparation including target lysis, DNA extraction and concentration, reagent rehydration, and target nucleic acid amplification and detection using real-time PCR. The assay includes a Sample Processing Control (SPC) that is present in the Extraction Tube. The SPC monitors DNA extraction steps, thermal cycling steps, reagent integrity and the presence of inhibitory substances. The BD MAX™ System software automatically interprets test results. A test result may be called as [SA NEG MRSA NEG (negative)], [SA POS, MRSA POS (MRSA positive)], [SA POS, MRSA NEG (SA positive)] or [SA UNR, MRSA UNR (Unresolved)] based on the amplification status of the target and of the Sample Processing Control. IND (Indeterminate) or INC (Incomplete) results are due to BD MAX™ System failure.

    AI/ML Overview

    The BD MAX™ StaphSR assay is an automated qualitative in vitro diagnostic test for the direct detection and differentiation of Staphylococcus aureus (SA) and methicillin-resistant Staphylococcus aureus (MRSA) DNA from nasal swabs. The performance of the device was evaluated in both analytical and clinical studies.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the BD MAX™ StaphSR assay are primarily based on achieving certain levels of sensitivity and specificity when compared against a reference method (direct culture complemented by enriched culture). The document details the reported performance for MRSA and SA detection.

    Acceptance Criteria and Reported Performance for BD MAX™ StaphSR Assay

    MetricAcceptance Criteria (Implied)Reported Performance (Overall) - Compared to Reference Method
    MRSA DetectionHigh sensitivity and specificitySensitivity: 93.1% (95% CI: 88.1%, 96.1%)
    Specificity: 97.5% (95% CI: 96.8%, 98.1%)
    Positive Predictive Value (PPV): 73.2% (95% CI: 67.8%, 78.3%)
    Negative Predictive Value (NPV): 99.5% (95% CI: 99.1%, 99.7%)
    SA DetectionHigh sensitivity and specificitySensitivity: 92.0% (95% CI: 89.7%, 93.9%)
    Specificity: 93.1% (95% CI: 91.8%, 94.2%)
    Positive Predictive Value (PPV): 83.4% (95% CI: 81.9%, 85.8%)
    Negative Predictive Value (NPV): 96.8% (95% CI: 96.0%, 97.6%)
    Unresolved RateLow unresolved rateInitial Unresolved Rate: 0.6% (15/2399)
    Unresolved Rate after Repeat: 0.04% (1/2399)
    InclusivityDetection of various MRSA and MSSA strains/genotypes at low LoDDetected 11 MREJ genotypes, 5 SCCmec types, various PFGE types, VRSA, VISA, and 51 MSSA strains (at 2-3 x LoD).
    SpecificityNo cross-reactivity with non-target organismsNo interference from 57 non-staphylococcal species, 45 CoNS/CoPS strains, 50 MSSA strains, and 17 viruses.
    InterferenceNo significant interference from common nasal substancesNo reportable interference from 29 substances/microorganisms, except for Tobramycin at a high concentration.
    Carryover/Cross-ContaminationLow risk of carryover/cross-contamination3 false positive results (1.5%) out of 203 expected negative samples due to carry-over.
    ReproducibilityConsistent results across sites and lotsSite-to-Site & Lot-to-Lot agreement: 100% for MP MRSA & TN, 96.7-97.8% for LP MRSA & LP MSSA.

    2. Sample Size Used for the Test Set and Data Provenance

    For the clinical performance studies, a total of 2451 specimens were enrolled. After excluding non-compliant specimens, 2354 specimen results were used to determine the clinical performance against the Reference Method, and 2393 specimen results were used for comparison against Direct Culture.

    The data provenance for the clinical study is prospective, collected from three geographically diverse investigational centers. The specific countries of origin for these centers are not explicitly stated.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not explicitly state the number of experts used to establish the ground truth or their specific qualifications (e.g., radiologist with X years of experience). However, the ground truth was established using laboratory methods rather than expert interpretation of images or clinical assessments.

    4. Adjudication Method for the Test Set

    The adjudication method for the clinical test set involved a Comparative Reference Method consisting of direct culture complemented by enriched culture.

    • For specimens negative for MRSA or SA by direct culture, enriched culture analysis was performed.
    • Presumptive S. aureus colonies on chromogenic medium were subcultured onto Blood Agar (BA) for identification confirmation with an agglutination test.
    • Methicillin resistance was confirmed by Cefoxitin disk (30 µg) diffusion susceptibility testing.
    • Enrichment in Trypticase Soy Broth with 6.5% NaCl was carried out if MRSA or SA was not confirmed by initial direct culture.
    • Turbid TSB 6.5% NaCl broth was used to inoculate additional chromogenic medium and BA plates, and MRSA confirmation followed the same procedure.
    • Further investigation was performed on specimens with discordant results between the Reference Method and the BD MAX™ StaphSR assay. For MRSA, 12 of 54 false positive and 5 of 11 false negative BD MAX™ StaphSR specimens were re-evaluated by the Reference Method. For SA, 28 of 118 false positive and 23 of 52 false negative BD MAX™ StaphSR specimens were re-evaluated by the Reference Method.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, What was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This information is not applicable as the BD MAX™ StaphSR assay is an automated in vitro diagnostic test (molecular assay) for direct detection of DNA. It does not involve "human readers" or AI assistance in the interpretation of results in the context of an MRMC study. The device provides an automated interpretation of results.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a standalone performance study was done. The BD MAX™ StaphSR assay is an automated instrument-based test that directly provides a qualitative result ([SA NEG MRSA NEG], [SA POS, MRSA POS], [SA POS, MRSA NEG], or [SA UNR, MRSA UNR]). The clinical performance data presented in Tables 10-17 directly reflect the standalone performance of the assay compared to the reference microbiological methods.

    7. The Type of Ground Truth Used

    The ground truth used for both the analytical and clinical studies was based on microbiological culture methods:

    • Direct culture complemented by enriched culture for clinical performance evaluation.
    • Cultured bacterial suspensions and well-characterized clinical isolates for analytical studies (e.g., LoD, inclusivity, specificity, competitive inhibitory effect).

    8. The Sample Size for the Training Set

    The document describes premarket studies for a diagnostic device. For such devices, there isn't typically a "training set" in the machine learning sense. Instead, development and optimization are performed using various analytical samples and potentially some initial clinical samples, followed by a robust validation (clinical performance) on an independent set. The details of any specific samples used for internal development or optimization (analogous to a training set) are not provided in this 510(k) summary. The provided sample sizes relate to the analytical validation and the final clinical validation dataset.

    9. How the Ground Truth for the Training Set Was Established

    Given that this is a molecular diagnostic assay and not an AI/ML-based image interpretation product, the concept of a "training set" with ground truth established in that context is not directly applicable. If one were to consider the samples used for assay development and optimization as a "training set," their ground truth would have been established through known bacterial cultures (characterized strains, quantified suspensions) and potentially well-characterized clinical isolates with confirmed microbiological status (e.g., through traditional culture, biochemical tests, and susceptibility testing). However, specific details of such "pre-validation" ground truth establishment are not provided in this regulatory summary.

    Ask a Question

    Ask a specific question about this device

    K Number
    K130470

    Validate with FDA (Live)

    Date Cleared
    2013-04-02

    (36 days)

    Product Code
    Regulation Number
    866.3130
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The BD MAX™ Cdiff Assay performed on the BD MAX™ System is an automated in vitro diagnostic test for the direct, qualitative detection of the Clostridium difficile toxin B gene (tcdB) in human liquid or soft stool specimens from patients suspected of having C. difficile infection (CDI). The test. performed directly on the specimen, utilizes real-time polymerase chain reaction (PCR) for the amplification of C. difficile toxin B gene DNA and fluorogenic target-specific hybridization probes for the detection of the amplified DNA. The BD MAX ™ Cdiff Assay is intended to aid in the diagnosis of CDI.

    Device Description

    The BD MAX™ System and the BD MAX™ Cdiff Assay are comprised of an instrument with associated hardware and accessories, disposable microfluidic cartridges, master mixes, unitized reagent strips, extraction reagents, and sample buffer tubes. The instrument automates sample preparation including target Ivsis. DNA extraction and concentration, reagent rehydration, and target nucleic acid amplification and detection using real-time PCR. The assay includes a Sample Processing Control (SPC) that is present in the Extraction Tube. The SPC monitors DNA extraction steps, thermal cycling steps, reagent integrity and the presence of inhibitory substances. The BD MAX™ System software automatically interprets test result may be called as NEG (negative), POS (positive) or UNR (unresolved) based on the amplification status of the target and of the Sample Processing Control. IND (indeterminate) or INC (incomplete) results are due to BD MAX™ System failure.

    AI/ML Overview

    BD MAX™ Cdiff Assay - Acceptance Criteria and Study Findings

    The BD MAX™ Cdiff Assay is an automated in vitro diagnostic test for the direct, qualitative detection of the Clostridium difficile toxin B gene (tcdB) in human liquid or soft stool specimens. The assay is intended to aid in the diagnosis of C. difficile infection (CDI).

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document details various analytical and clinical performance studies, but explicit "acceptance criteria" for minimum performance thresholds (e.g., "Sensitivity must be >= 90%") are not stated in a direct, consolidated table. However, the reported performance metrics can be inferred as meeting the implicit acceptance criteria for the submission to be cleared.

    Here's a summary of the reported device performance from the provided text, focusing on key metrics:

    Performance MetricReported Device Performance (BD MAX™ Cdiff Assay)
    Clinical Performance (vs. Reference Method: Direct Culture + Enriched Culture)
    Sensitivity87.7% (265/302) (95% CI: 83.6%, 91.0%)
    Specificity96.8% (1469/1517) (95% CI: 95.8%, 97.6%)
    Positive Predictive Value (PPV)83.5% (95% CI: 79.4%, 87.1%)
    Negative Predictive Value (NPV)97.7% (95% CI: 97.0%, 98.4%)
    Clinical Performance (vs. Direct Culture)
    Positive Percent Agreement (PPA)96.5% (194/201) (95% CI: 93.0%, 98.3%)
    Negative Percent Agreement (NPA)92.7% (1507/1625) (95% CI: 91.4%, 93.9%)
    Comparison to FDA-cleared Molecular Assay 1
    Positive Percent Agreement (PPA)99.1% (95% CI: 94.9%-99.8%)
    Negative Percent Agreement (NPA)97.4% (95% CI: 95.7%-98.4%)
    Comparison to FDA-cleared Molecular Assay 2
    Positive Percent Agreement (PPA)95.5% (95% CI: 92.1%-97.5%)
    Negative Percent Agreement (NPA)98.8% (95% CI: 97.9%-99.3%)
    Analytical Sensitivity (LoD)LoD ranged from 125 to 265 CFU per loop for tested strains (lowest concentration at which 95% of replicates tested positive).
    Analytical InclusivityCorrectly identified 62 of 64 toxigenic C. difficile strains (~3xLoD). 2 strains produced low signal/false-negative in 1/5 replicates.
    Analytical SpecificityProduced negative results for 6/6 non-tcdB C. difficile strains, 30/30 other Clostridium strains, and 95/98 other bacterial strains.
    Reproducibility (Overall % Agreement for LP, MP, Neg samples)100% (both Lot-to-Lot and Site-to-Site)
    Unresolved Rate (after repeat testing)0.5% (10/1844) (95% CI: 0.3%, 1.0%)

    2. Sample Size and Data Provenance

    • Test Set (Clinical Performance Studies):

      • Clinical Performance (vs. Reference Method): 1819 compliant and reportable results from 2071 initial soft or liquid stool specimens.
      • Clinical Performance (vs. Direct Culture): 1826 compliant and reportable results.
      • Comparison Studies: 2013 specimens tested across two external sites (specific breakdown per comparator assay: 647 for one and 1366 for the other).
      • Data Provenance: Multi-site prospective investigational study from six investigational centers. The document does not specify countries of origin, but the submitting company is Canadian. The study was prospective, as specimens were from "patients suspected of having C. difficile infection for which diagnostic tests were indicated and ordered."
    • Training Set: The document does not provide specific details on a separate "training set" for algorithm development. This is common for molecular diagnostic assays where the "training" (e.g., selection of primers/probes, optimization of PCR conditions) is inherent in the assay's design and analytical validation, rather than a distinct machine learning training phase.

    3. Number of Experts and Qualifications for Ground Truth

    • Ground Truth for Clinical Performance (Reference Method): The ground truth was established by a "Comparative Reference Method consisting of direct culture complemented by enriched culture." The anaerobic culture was used to isolate C. difficile, followed by confirmation of isolate identification and a "Tissue Culture Cytotoxicity Assay to determine the toxigenicity of the isolate." This process typically involves trained microbiology laboratory personnel, but the specific number and qualifications of "experts" (e.g., clinical microbiologists, laboratory directors) involved in establishing the final ground truth from this multi-step process are not explicitly stated in the document.

    • Ground Truth for Analytical Studies (e.g., LoD, Inclusivity, Specificity): For analytical studies, ground truth was based on known concentrations of bacterial strains (CFU/mL or genomic copies/mL) or known presence/absence of certain organisms, established through standard laboratory methods of quantification and identification. No "experts" in the sense of clinical reviewers are typically involved in establishing these analytical truths.

    4. Adjudication Method for the Test Set

    The document does not describe a formal adjudication method (e.g., 2+1, 3+1 reviewer consensus) by human experts for the clinical test set results. The ground truth for the clinical performance was established by a "Comparative Reference Method" (direct culture + enriched culture + toxigenicity testing). Discordant results between the BD MAX™ Cdiff Assay and this Reference Method were further investigated:

    • "27 of 48 False Positive BD MAX™ Cdiff specimens were also found to be positive using another commercially available FDA-cleared RT-PCR assay targeting the C. difficile tcdB gene."
    • "32 of 37 False Negative BD MAX™ Cdiff specimens, were also found to be negative using another commercially available FDA-cleared RT-PCR assay targeting the C. difficile tcdB gene."
      This "further investigation" serves as a form of resolution for discordant cases but is not a traditional multi-reader "adjudication" process using human interpretation of images or clinical data.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was described. The device is an automated in vitro diagnostic test that provides a qualitative (positive/negative) result, meaning there is no human "reader" in the loop for interpreting the assay's output that would then be compared to human performance with or without AI assistance. The study compares the device's performance against a reference method and other FDA-cleared molecular assays.

    6. Standalone Performance

    Yes, a standalone (algorithm only without human-in-the-loop performance) study was done. The entire clinical and analytical performance evaluation describes the performance of the BD MAX™ Cdiff Assay as a standalone device. The assay automates sample preparation, nucleic acid amplification, and detection, with the BD MAX™ System software automatically interpreting test results as NEG (negative), POS (positive), or UNR (unresolved).

    7. Type of Ground Truth Used

    • Clinical Performance Studies: The ground truth was based on a "Comparative Reference Method" which primarily relied on culture-based methods (direct and enriched anaerobic culture to isolate C. difficile) combined with a Tissue Culture Cytotoxicity Assay to determine the toxigenicity of the isolated strains. This represents a combination of microbiological culture and a functional assay (outcome of bacterial toxin activity on cells).
    • Analytical Performance Studies (LoD, Inclusivity, Specificity): The ground truth was based on known characterized bacterial strains and their concentrations, determined by standard microbiological quantification methods.

    8. Sample Size for the Training Set

    The document does not specify a distinct "training set" sample size in the context of the device's development or a machine learning algorithm. For molecular diagnostic assays like this, the "training" involves the scientific development and optimization of the assay components (primers, probes, reaction conditions) performed by molecular biologists and R&D scientists, rather than a separate data-driven training phase for an AI model.

    9. How the Ground Truth for the Training Set Was Established

    As described above, no explicit "training set" with established ground truth in the machine learning sense is detailed. The assay's design and optimization (analogous to "training" in AI development) would have utilized known positive and negative controls, characterized bacterial strains, and various concentrations to perfect the real-time PCR parameters and ensure accurate detection of the tcdB gene. This process involves standard laboratory validation techniques rather than human expert annotation for ground truth.