510(k) Data Aggregation

    K Number
    K131264

    Validate with FDA (Live)

    Date Cleared
    2013-07-26

    (84 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Komet Medical Kirschner Wires, Steinmann Pins and Guide Pins and Screws are indicated for use in fixation of bone fractures, for bone reconstructions, as guide pins for insertion of other implants or implantation through the skin so that traction may be applied to the skeletal system.

    Device Description

    Komet Medical Kirschner Wires, Steinmann Pins and Guide Pins and Screws are available in several diameters and lengths, in both threaded and non-threaded designs with a variety of point geometries.

    AI/ML Overview

    The provided text does not contain information about acceptance criteria, device performance, or a study demonstrating that a device meets acceptance criteria. The document is a 510(k) summary for Komet Medical Kirschner Wires, Steinmann Pins and Guide Pins and Screws, outlining its indications for use, device description, and substantial equivalence to predicate devices. It also includes an FDA response confirming substantial equivalence.

    Therefore, I cannot provide the requested table or answer the specific questions about a study based on the given information.

    To answer your request, a document detailing device performance data, acceptance criteria, and specific study methodologies would be required.

    Ask a Question

    Ask a specific question about this device

    K Number
    K071352

    Validate with FDA (Live)

    Device Name
    CERADRILL
    Date Cleared
    2007-06-29

    (45 days)

    Product Code
    Regulation Number
    872.3240
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The drill is used to pre-drill into the maxilla or mandible to create an initial osteotomy for the later endosseous dental implant placement. The drills are used with a handpiece.
    CeraDrill is a ceramic drill for the dental implantology made of a modern high performance ceramic. The drills are intended to be used with a handpiece. The intended use for this device is to cut into the maxilla or mandible to create an osteotomy for endosseous dental implant placement.

    Device Description

    CeraDrills within the new "CeraLine" is intended for use in implantology. These are made of high efficiency ceramics featuring an excellent cutting performance. The new CeraDrill offers the possibility of working without any metal, thus guaranteeing a biocompatible operation. Damages, as they occur in particular when disinfecting and cleaning steel burs with inappropriate agents, are now a thing of the past. The multifunctional ceramic drills are suited for initial preparation of the implant site axis and depth. The drills feature high initial sharpness and optimal cutting efficiency, thus achieving an effective material reduction. Due to their high efficiency toothing and the distinct tip-transversing blade, the drill has very good axial drilling properties for precise cutting without the need for prior use of a centering bur. The special twist drill blade geometry ensures smooth and precise operation. Moreover, the depth markings, which are lasered on to the working part in 2 mm intervals guarantee precise and safe control of the penetration depth. The CeraDrill family is available in diverse dimensions and shapes for the individual preparation on the osteotomy.

    AI/ML Overview

    This is a submission for a 510(k) premarket notification for a dental device, specifically a bone-cutting instrument. These types of devices generally rely on demonstrating substantial equivalence to a predicate device rather than undergoing extensive clinical trials as typically seen for novel medical devices or diagnostic algorithms. Therefore, the information provided focuses on non-clinical performance and technological characteristics.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Aspects from "Summary of technological Characteristics" & "Mechanical Stability")Reported Device Performance (CeraDrill)
    Energy used and/or deliveredIntended to be used in electrical powered hand pieces (Same as predicate)
    DesignTwo/three cutting flutes, external irrigation (Predicate device had internal/external irrigation)
    Performance (Concenticity)Ordinary concentricity (Same as predicate)
    Corrosion ResistanceCorrosion resistive; not leading to decolourisation while reprocessing (Improved over predicate's limited resistance)
    MaterialsZirconium oxide (ZrO2/Al2O3/Y2O3); single material, no coating, limited probability of abrasion (Advantage over predicate's coated stainless steel)
    BiocompatibilityAccording ISO 7405 (Same as predicate)
    Ability to be reprocessedCan be used, cleaned, disinfected, and sterilized for at least 10 cycles (demonstrated in test report #06m1166)
    Mechanical stability after reprocessingReprocessed in 100 cycles without significant loss of mechanical stability (details in chapter 18 d)
    Thermal safetyImproved thermal safety; lower temperature rise observed with ceramic burs compared to steel burs (details in chapter 18 a)
    Resistiveness against bending forcesAssumable lower resistiveness against bending forces during operation is negligible (fracture would lead to unacceptable osteotomies/misuse)

    2. Sample Size Used for the Test Set and Data Provenance

    The submission does not specify a "test set" in the context of an AI/ML algorithm or a diagnostic test with patient data. Instead, it refers to non-clinical tests conducted on the CeraDrill devices themselves to evaluate their physical and biological characteristics. The data provenance is internal testing by the manufacturer (Gebr. Brasseler GmbH & Co. KG), and the studies are prospective in the sense that they were designed and executed to test the characteristics of the new device. No geographical origin of patient data is relevant here as it's not a clinical study involving patients.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This is not a study that involves expert review for ground truth in the context of diagnostic accuracy. The "ground truth" for the non-clinical tests would be the measured physical and biological properties of the device, assessed using established engineering and laboratory standards.

    4. Adjudication Method for the Test Set

    Not applicable. As noted above, this submission describes non-clinical performance data, not a study evaluating human interpretation or a diagnostic outcome requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

    No MRMC comparative effectiveness study was done. This device is a surgical instrument, not an AI-assisted diagnostic tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No standalone algorithm performance study was done. This is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the evaluation of the CeraDrill's performance is based on engineering and laboratory test results for mechanical stability, thermal safety, corrosion resistance, material composition, and biocompatibility, as determined by internal testing and adherence to relevant standards (e.g., ISO 7405 for biocompatibility).

    8. The Sample Size for the Training Set

    Not applicable. There is no training set as this is not an AI/ML device.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. There is no training set.