Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K240828

    Validate with FDA (Live)

    Device Name
    OEC One ASD
    Date Cleared
    2024-12-27

    (276 days)

    Product Code
    Regulation Number
    892.1650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OEC One ASD mobile C-arm system is designed to provide fluoroscopic and digital spot images of adult and pediatic patient populations during diagnostic, interventional, and surgical procedures. Examples of a clinical application may include: orthopedic, gastrointestinal, endoscopic, urologic, vascular, critical care, and emergency procedures.

    Device Description

    The OEC One ASD is a mobile C-arm X-ray system to provide fluoroscopic images of the patient during diagnostic, interventional, and surgical procedures such as orthopedic, gastrointestinal, endoscopic, urologic, neurologic, vascular, critical care, and emergency procedures. These images help the physician visualize the patient's anatomy and localize clinical regions of interest. The system consists of a mobile stand with an articulating arm attached to it to support an image display monitor (widescreen monitor) and a TechView tablet, and a "C" shaped apparatus that has a flat panel detector on the top of the C-arm and the X-ray Source assembly at the opposite end.

    The OEC One ASD is capable of performing linear motions (vertical, horizontal) and rotational motions (orbital, lateral, wig-wag) that allows the user to position the X-ray image chain at various angles and distances with respect to the patient anatomy to be imaged. The C- arm is mechanically balanced allowing for ease of movement and capable of being "locked" in place using a manually activated lock.

    The subject device is labelled as OEC One ASD.

    AI/ML Overview

    The provided document is a 510(k) Summary of Safety and Effectiveness for the GE Hualun Medical Systems Co. Ltd. OEC One ASD, a mobile C-arm X-ray system. The document focuses on demonstrating substantial equivalence to a predicate device, OEC One (K182626), rather than presenting a study with specific acceptance criteria and detailed device performance results for a new AI/CAD feature.

    The submission is for a modification of an existing device, primarily introducing an amorphous silicon (a-Si) flat panel detector as the image receptor and updating some hardware and software components. The changes are stated to enhance device performance and are discussed in terms of their impact on safety and effectiveness, concluding that no new hazards or concerns were raised.

    Therefore, the information required for a detailed description of acceptance criteria and a study proving device performance, especially for AI/CAD features, is largely not present in this document. The document centers on demonstrating that the modified device maintains safety and effectiveness and is substantially equivalent to the predicate, rather than detailing a study against specific acceptance criteria for a novel functionality.

    However, I can extract the available relevant information and highlight what is missing based on your request.


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly define "acceptance criteria" in the context of a study demonstrating novel AI/CAD feature performance. Instead, it presents a comparison table of technical specifications between the proposed device (OEC One ASD) and the predicate device (OEC One K182626) to demonstrate substantial equivalence. The "Acceptance Criteria" here are implicitly derived from the predicate's performance and safety profiles.

    Feature / Performance MetricPredicate Device (OEC One K182626)Subject Device (OEC One ASD)Discussion of Differences / Equivalence
    Image ReceptorImage Intensifier21cm Amorphous Silicon (a-Si) Flat Panel DetectorSubstantially Equivalent. Change to enhance device performance.
    DQE65%70% (0 lp/mm)Enhanced DQE, indicating improved image quality.
    MTF45%46% (1.0 lp/mm)Slightly enhanced MTF, indicating improved image quality.
    Field of View9 inch, 6 inch, 4.5 inch21 cm, 15 cm, 11 cmNo new hazards or hazard situations. Performance testing indicated effectiveness.
    Image Matrix Size1000x10001520x1520Substantially Equivalent. Driven by detector pixel matrix for higher resolution.
    Image ShapeCircleSquircleSubstantially Equivalent. Enhanced viewing area without typically unnecessary corner areas.
    Anti-scatter GridLine Rate: 60 L/cm, Ratio: 10:1, Focal Distance: 100 cmLine Rate: 74 L/cm, Ratio: 14:1, Focal Distance: 100 cmSubstantially Equivalent. Specification updated based on new image receptor.
    X-ray GeneratorFluoroscopy: 0.1-4.0 mAFluoroscopy: 0.1-8.0 mASubstantially Equivalent. mA range change for optimized image quality (ABS). No new safety/effectiveness concerns.
    Digital Spot: 0.2-10.0 mA (100-120V system)Digital Spot: 2-10.0 mA (100-120V system)Substantially Equivalent. mA range change for optimized image quality (increasing mA on thin anatomy). No new safety/effectiveness concerns.
    Imaging ModesDigital Spot: Normal Dose, Low DoseDigital Spot: Normal DoseLow Dose mode not provided for Digital Spot as high mA exposure ensures quality; similar functionality available via Fluoroscopy.
    Roadmap: Normal Dose, Low DoseRemovedRoadmap mode removed based on marketing; similar functionality via peak opacify function on cine.
    Imaging FeaturesZoom & RoamZoom (Live Zoom) & RoamImproved with Live Zoom during fluoro/cine.
    N/ADigital PenAdded for planning/educational purposes.
    Monitor DisplayResolution: 1920x1080Resolution: 3840 x 2160Substantially Equivalent. Updated to higher resolution due to IT advancement.
    8bit image display10bit image displaySubstantially Equivalent. Better display technology.
    Tech View TabletOS: Android 5.1OS: Android 11.0Substantially Equivalent. OS upgraded due to IT advancement.
    C-Arm Physical DimensionsOrbital Rotation: 120° (90° underscan /30° overscan)Orbital Rotation: 150° (95° underscan /55° overscan)Substantially Equivalent. Larger range for user convenience.
    Image Storage100,000 Images150,000 ImagesSubstantially Equivalent. Driven by IT advancement (more storage).
    Wireless Printing ModuleN/AWireless Printing ModuleSubstantially Equivalent. Not for diagnostic use or device control. No new risks.
    Video DistributorDVI, BNCDP, BNCSubstantially Equivalent. Driven by IT advancement.
    Laser AimerRed Laser, Class IIIa/3R, 650 nm, ≤ 5.0 mWGreen Laser, Class 2, 510-530nm, 1mWSubstantially Equivalent. Updated for green laser and convenience (tube side). Both meet laser product requirements.
    Image ProcessingADRO (based on CPU)ADRO (based on GPU)Substantially Equivalent. GPU for better calculation speed. All other listed image processing functions are the same.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the document. The submission states, "comparative clinical images were evaluated to demonstrate substantial equivalence for the OEC One ASD compared to the cleared predicate," but no details on the sample size, data provenance (e.g., country of origin, retrospective/prospective nature), or specific evaluation methodology for these clinical images are given.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not provided in the document. The document states that "comparative clinical images were evaluated," but it does not specify the number or qualifications of experts involved in this evaluation or the establishment of ground truth.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not provided in the document. The method used to resolve discrepancies or establish a consensus for the evaluation of comparative clinical images is not described.


    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study focusing on human reader improvement with AI assistance was not mentioned or described in this 510(k) submission. The document discusses device modifications and their impact on image quality and functionality, but not the comparative effectiveness of human readers utilizing AI.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This submission does not describe a standalone performance study for an AI algorithm. The device itself is an X-ray system, and while it has "Image Processing" features, these are not presented as standalone AI algorithms for diagnostic assistance but rather as integrated components affecting image generation and display characteristics. The update to ADRO from CPU to GPU based processing is noted for speed, but its standalone performance as an AI algorithm is not evaluated or presented.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The document mentions "comparative clinical images were evaluated," but it does not specify the type of ground truth against which these images were assessed. Since the primary focus is on demonstrating substantial equivalence of technical image characteristics rather than validating a diagnostic AI output, a traditional "ground truth" (like pathology or outcomes a specific AI would predict) is not explicitly detailed. The implicit ground truth would be the expected imaging performance and diagnostic utility comparable to the predicate device.


    8. The sample size for the training set

    This information is not provided in the document. The document describes modifications to an existing X-ray system, including software updates. It states, "Its software is based on the architecture, design and code base of the predicate device OEC One (K182626)," and underwent a standard software development lifecycle. There is no mention of a separate "training set" in the context of an AI/CAD algorithm as typically understood for deep learning models.


    9. How the ground truth for the training set was established

    Since no training set for an AI/CAD algorithm is mentioned (refer to point 8), the method for establishing its ground truth is not applicable/provided in this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231892

    Validate with FDA (Live)

    Date Cleared
    2023-09-12

    (76 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This product is intended to generate digital radiographic images of the skull, spinal column, extremities, and other body parts in patients of all ages. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position and the system is intended for use in all routine radiography exams. This device is not intended for mammographic applications.

    Device Description

    The Definium Pace Select is a Digital Radiographic System designed as a modular system with components that includes fixed table with tube-stand, wallstand, cleared wireless digital detector, X-ray tube, collimator, high kV generator and acquisition workstation in control room. The system generates diagnostic radiographic images which can be sent through a DICOM network for applications including printing, viewing, and storage. The components may be grouped into different configurations to meet specific customer needs.

    AI/ML Overview

    The FDA 510(k) summary for the Definium Pace Select, a Digital Radiographic System, indicates that clinical studies were not required to support substantial equivalence for this device. Therefore, a study proving the device meets specific acceptance criteria in terms of clinical performance (like sensitivity, specificity, or reader improvement with AI assistance) or a detailed clinical acceptance criteria table for such measures is not provided in this document.

    The submission focuses primarily on demonstrating substantial equivalence to its predicate device (Discovery XR656 HD) based on non-clinical tests, technological characteristics, and safety/effectiveness data. The device's safety and effectiveness were confirmed through design verification and validation testing.

    Here's a breakdown of the relevant information from the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Since clinical studies were not deemed necessary for this 510(k) submission, there isn't a table presenting clinical acceptance criteria (e.g., sensitivity, specificity) and corresponding device performance metrics in the document. The acceptance criteria focused on compliance with voluntary standards and successful completion of verification and validation testing.

    Acceptance Criteria CategoryReported Device Performance / Compliance
    Voluntary Standards (Non-Clinical)Device complies with listed standards:
    • ES 60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012
    • IEC 60601-1-2:2014[Including AMD 1:2021]
    • IEC 60601-1-3: 2021
    • IEC 60601-1-6: 2020
    • IEC 60601-2-54: 2018
    • IEC 62366: 2015 + AMD1:2020
    • ISO 10993-1: 2018
    • ISO 10993-5: 2009/(R)2014
    • ISO 10993-10: 2010/(R)2014
    • ISO 10993-18 Second edition 2020-01
    • PS 3.1 - 3.20: 2022d (DICOM set) |
      | Quality Assurance Measures | Applied: Risk Analysis, Requirements Reviews, Design Reviews, Unit level testing (Module verification), Integration testing (System verification), Performance testing (Verification), Safety testing (Verification), Simulated use testing (Validation). |
      | Risk Mitigation | Risks for new floor mounted positioners and image chain were evaluated, mitigated with design controls and labeling. Mitigations were verified and validated with acceptable results. |
      | Design Verification & Validation | Performed to confirm safety and effectiveness; test plans and results were executed with acceptable results. |

    2. Sample Size for Test Set and Data Provenance:

    No distinct "test set" in the context of clinical performance evaluation (e.g., images for diagnostic accuracy assessment) is mentioned because clinical studies were not required. The "testing" referred to in the document pertains to design verification and validation, which would involve hardware and software testing, rather than a clinical image dataset.

    3. Number of Experts for Ground Truth and Qualifications:

    Not applicable, as no clinical study requiring expert ground truth establishment for diagnostic accuracy was conducted for this submission.

    4. Adjudication Method:

    Not applicable, as no clinical study requiring adjudication of expert interpretations was conducted.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No MRMC study was conducted. The device is a digital radiographic system, and the submission emphasizes its substantial equivalence based on technological characteristics and safety, not on AI assistance to human readers.

    6. Standalone (Algorithm Only) Performance:

    Not explicitly detailed in terms of a standalone diagnostic performance study, as the device itself is an imaging system, not a standalone diagnostic algorithm. The "image processing with same algorithm" as the predicate device suggests the algorithms are part of the overall system functionality, not a separate AI diagnostic tool.

    7. Type of Ground Truth Used:

    For the non-clinical tests and design verification/validation, the "ground truth" would be established through engineering specifications, regulatory compliance standards, and functional requirements. For example, for radiation output, the ground truth would be the expected range defined by standards and the device's design. There is no mention of clinical ground truth (e.g., pathology, outcomes data) being used for this 510(k) submission.

    8. Sample Size for Training Set:

    Not applicable, as this device submission is for a digital radiographic system, not an AI/ML algorithm that typically requires a training set of data for development. The reference to "same algorithm" for image processing as the predicate implies existing, validated algorithms rather than newly trained ones.

    9. How Ground Truth for Training Set Was Established:

    Not applicable, as this device is not presented as an AI/ML system requiring a training set with established ground truth.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1