Search Filters

Search Results

Found 7 results

510(k) Data Aggregation

    K Number
    K242747

    Validate with FDA (Live)

    Device Name
    FAQ™ (302)
    Manufacturer
    Date Cleared
    2024-12-11

    (90 days)

    Product Code
    Regulation Number
    890.5500
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FAQ™ 302 is indicated to promote hair growth in males with androgenetic alopecia who have Hamilton-Norwood Classifications of Ila-V and females with androgenetic alopecia who have Ludwig-Savin Classifications of I-II and Fitzpatrick Classification of Skin Phototypes I to IV.

    Device Description

    FAQ™ 302 is a handheld rechargeable square shaped device designed for improving hair growth with 20 laser diodes (wavelength 650±10m) and 20 RED LED beads (640-660nm) on the inside surface.

    By pressing the universal button, the user can power the device on, start treatment and change the intensity by quick-pressing the button again. Once the treatment is complete, the device is turned off by pressing and holding the universal power button for 3 seconds.

    The device can also be connected wirelessly via Bluetooth to an FAQ App in order to manage treatment. After the treatment is completed, the device automatically powers off.

    The device uses a non-detachable polymer lithium battery for power supply in compliance with IEC 62133-2:2017+A1:2021 standard.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the FAQ™ 302 device, based on the provided FDA 510(k) summary:

    This device (FAQ™ 302) is an Infrared Lamp intended to promote hair growth. The 510(k) summary explicitly states that clinical testing was considered not to be needed for this 510(k). Therefore, the provided document does not contain information about a study proving the device meets acceptance criteria related to its clinical effectiveness (i.e., promoting hair growth). Instead, the studies presented are non-clinical tests to demonstrate safety and performance equivalence to predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance

    Since clinical efficacy was "considered not to be needed," the acceptance criteria are not for clinical effectiveness but rather for safety, electrical performance, software functionality, cybersecurity, and usability. The document focuses on demonstrating that the new device is substantially equivalent to predicate devices based on these non-clinical aspects.

    Acceptance Criteria CategorySpecific Standard/RequirementReported Device Performance
    BiocompatibilityISO 10993-5: In Vitro CytotoxicityDevice tested in compliance; no additional risks noted.
    ISO 10993-10: Irritation and Skin SensitizationDevice tested in compliance; no additional risks noted.
    ISO 10993-23: Tests for irritationDevice tested in compliance.
    Electrical SafetyIEC 60601-1:2005/AMD1:2012/AMD2:2020, CAN/CSA-C22.2 NO. 60601-1:14, AAMI 60601-1:2005 + AMD 1:2012 (General Req.)Device tested to applicable electrical performance standards.
    IEC 60601-1-6:2010+AMD1:2013+AMD2:2020 (Usability Collateral Standard)Device tested to applicable electrical performance standards.
    IEC 62366-1:2015+AMD1:2020 (Usability Engineering)Device tested to applicable electrical performance standards.
    IEC 60601-1-11:2015+AMD1:2020 (Home Healthcare Environment)Device tested to applicable electrical performance standards.
    IEC 60601-2-57: 2011 (Non-laser light source)Device tested to applicable electrical performance standards.
    EMCEN 60601-1-2: 2015, IEC 60601-1-2: 2014 (Edition 4) (Electromagnetic disturbances)Device tested to applicable electrical performance standards.
    IEC TR 60601-4-2:2016 (Electromagnetic immunity)Device tested to applicable electrical performance standards.
    Software FunctionalityMinor level of concern per FDA guidance "Content of Premarket Submissions for Device Software Functions"All software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.
    CybersecurityEvaluation of assets, threats, vulnerabilities, impact, likelihood, risk levels, mitigation, residual riskAll risks associated with network capable interfaces adequately mitigated. Cybersecurity risk analysis performed, controls identified, comparison with National Vulnerability Database, and labeling information detailed in Cybersecurity Assessment.
    UsabilityUser instruction provides adequate guidance for independent and appropriate home use for Laser and LED treatmentTest results in "Usability validation report" clearly indicate this.

    2. Sample Size Used for the Test Set and Data Provenance

    • Biocompatibility Testing: The document does not specify the sample size (e.g., number of test specimens or replicates) for the biocompatibility tests.
    • Electrical Safety and EMC Testing: The document does not specify the sample size used for these compliance tests, as these are typically conducted on a representative sample of the device itself rather than patient data.
    • Software Verification and Validation: The document does not specify the number of test cases or data points for software validation.
    • Cybersecurity: This involved an analysis, not a test set with patient data.
    • Usability Study:
      • Sample Size: 15 adults
      • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). It is implied to be prospective as it's a "validation testing" that "was conducted."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • For all non-clinical tests (biocompatibility, electrical safety, EMC, software, cybersecurity): "Ground truth" in the sense of clinical expert consensus is not applicable. These tests rely on established engineering and scientific standards. The "experts" involved would be the testing engineers and scientists responsible for conducting and interpreting these standard-based tests. Qualifications are not specified, but they would be presumed to be competent in the relevant fields.
    • For the Usability Study: The document does not mention experts establishing a "ground truth" for usability. The study assessed whether adult users could follow instructions.

    4. Adjudication Method for the Test Set

    • Not applicable as no clinical test set requiring expert adjudication was performed.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The document explicitly states: "Clinical testing was considered not to be needed for this 510(k)." The device's substantial equivalence is based on non-clinical performance and a comparison of its technical specifications and indications for use with legally marketed predicate devices.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • Not applicable. This device is a hardware device (infrared lamp) with embedded software, not a standalone AI algorithm. The software verification and validation section relates to its internal software functionality, not an AI or standalone diagnostic/treatment algorithm.

    7. Type of Ground Truth Used

    • Biocompatibility, Electrical Safety, EMC: Ground truth refers to compliance with established international and national standards (e.g., ISO, IEC, AAMI compliance).
    • Software: Ground truth refers to predefined software requirements and specifications, and the mitigation of identified software hazards.
    • Cybersecurity: Ground truth refers to industry best practices, identified risks and vulnerabilities (including comparison with the National Vulnerability Database), and the effectiveness of mitigation strategies.
    • Usability Study: Ground truth for this study was the ability of the users to follow the provided instructions to safely and appropriately use the device.

    8. Sample Size for the Training Set

    • Not applicable. This device is not an AI/ML algorithm that requires a "training set" in the context of learning from data to perform a diagnostic or treatment task. Software validation was performed but does not involve a training set as understood in AI/ML.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as no training set was used for an AI/ML algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K241102

    Validate with FDA (Live)

    Device Name
    Luna 4 plus
    Manufacturer
    Date Cleared
    2024-08-20

    (120 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Luna 4 plus is a Near-Infrared Heated Cleansing Device With Red LED/NIR light & Microcurrent Massage.

    Microcurrent: indicated for facial stimulation RED+NIR: indicated to treat periorbital wrinkles

    Device Description

    Luna 4 plus has multiple functions combined in a single device, like RED+NIR LED Treatment for wrinkle reduction. The device emits visible red/IR light (Red: 630nm +/- 10nm.) for elevating tissue temperature in improving appearance of facial wrinkles. Microcurrent to target facial stimulation. Two microcurrent pins at the back for a targeted toning massage, emitting microcurrent electrical therapy to deliver skin stimulation where maximum power delivered to the load is controlled and limited by integral software. Luna 4 plus is a single device powered by a rechargeable battery. A controller is integrated on the back to control the device, such as turn on/off the device, mode buttons. Rechargeable Li-battery contained in the controller with a USB cable accessible to target it. To user should use the controller to operate.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and supporting study for the Luna 4 plus device, based on the provided FDA 510(k) summary:

    The document provided primarily focuses on establishing "substantial equivalence" of the Luna 4 plus to existing predicate devices, rather than defining explicit acceptance criteria and then presenting a study to prove meeting those criteria in the traditional sense of a clinical trial for a novel device. The "acceptance criteria" here are implicitly related to demonstrating that the new device is as safe and effective as the predicate devices through conformance to recognized standards and comparative testing.

    The "study" that proves the device meets "acceptance criteria" (i.e., is substantially equivalent to predicates) is a combination of:

    • Non-clinical testing: Electrical safety, EMC, and biocompatibility, as per applicable standards.
    • Lab bench performance testing: Characterization of electrical stimulation output, including Anti-Shock System performance and overload conditions.
    • Usability studies: To demonstrate substantial equivalence given new indications.

    1. A table of acceptance criteria and the reported device performance

    Since explicit "acceptance criteria" in the format of specific numerical thresholds tied to a single study result are not provided, I will construct a table reflecting the key equivalency demonstrations and comparisons made against the predicate devices. The "reported device performance" will be the "Subject Device" values from the Substantial Equivalence Comparison Table, along with notes on safety and efficacy from the "Conclusion" and "Comparison in details" sections.

    Feature AreaImplicit Acceptance Criterion (Demonstrate Substantial Equivalence to Predicate)Reported Device Performance (Luna 4 plus)
    Regulatory Information
    Regulation NumberComparable regulatory classification for similar intended uses.878.4810 (Laser surgical instrument), 882.5890 (Transcutaneous Electrical Nerve Stimulator). Different from some predicates but within acceptable classifications for its functions.
    Intended UseIndications for facial stimulation and periorbital wrinkle treatment should be similar to predicates.1. Red + IR light intended to treat periorbital wrinkles. 2. Microcurrent intended for facial stimulation. Similar to predicates.
    Safety MechanismPossess safety mechanisms (e.g., thermal, electrical protection) equivalent to or better than predicates.Thermal sensor and heat stabilizer (prevents overheating above 42-43°C). Anti-Shock system (monitors voltage on feedback resistor to ensure stable current). Different from some predicates but detailed to demonstrate safety.
    Device Characteristics
    Application AreaFace.Face. Same as predicates.
    Use EnvironmentHome Use, lay users.Home Use, lay users. Same as predicates.
    Power SourceInternal rechargeable battery.Internal rechargeable Lithium battery. Same as predicates. Tested to IEC 62133-2.
    Software/Firmware/ControlPresence of control software.Yes. Same as predicates.
    Patient Override ControlUser control over device operation.Yes. Same as predicates.
    Microcurrent Function
    Electrical CharacteristicsMicrocurrent characteristics (e.g., waveform, pulse width, output voltage/current, frequency) should be safe and effective, and differences adequately justified/tested.Pulsed monophasic, alternating polarity, charge-balanced waveform; Modulated square shape; Pulse width 290 µs; Max RMS Output Voltage 0.28V @500Ω; Max RMS output Current 560μA @500Ω; Max Net Charge 0 µC (charge-balanced). Frequency Range 46Hz. Different from predicates but evaluated via lab bench performance testing to IEC 60601-2-10 to ensure safety and effectiveness, acknowledging lower output power density than one predicate.
    LED/NIR Light Function
    Optical CharacteristicsLED wavelength and power density suitable for stated indications and comparable to predicates.RED light: 633nm±10nm; Red+IR: 633±10nm / 850nm±10nm. Optical Power Density RED+IR: 63 mW/cm². Differences in wavelength and power density from predicates but supported by usability report and bench evaluation for safety and efficacy.
    Compliance & Standards
    BiocompatibilityCompliance with relevant ISO 10993 standards.ISO 10993-5, ISO 10993-10, ISO 10993-23. Similar to predicates.
    Electrical SafetyCompliance with relevant IEC 60601 standards.IEC 60601-1, IEC 60601-1-6, IEC 60601-1-11, IEC 60601-2-10, IEC 60601-2-57. Similar to predicates. Performance testing of Anti-Shock System and under overload conditions also performed.
    EMCCompliance with IEC 60601-1-2.IEC 60601-1-2. Same as predicates.
    21 CFR 898Compliance with 21 CFR 898 performance standards.Yes. Same as predicates.

    2. Sample size used for the test set and the data provenance

    The document does not specify a distinct "test set" in terms of a clinical or image dataset. The testing primarily involved non-clinical lab-based evaluations:

    • Electrical safety, EMC, and biocompatibility: Tested against applicable standards (e.g., IEC 60601-1, IEC 60601-1-2, ISO 10993 series). This typically involves a single or small number of device units.
    • Lab bench performance testing: Characterization of electrical stimulation, Anti-Shock System, and overload conditions. This would also involve a small number of physical devices.
    • Usability studies: These are mentioned, but no specifics on sample size (number of users) or study design (e.g., prospective, retrospective, country of origin) are provided.

    Therefore, there isn't a "sample size for the test set" in the context of an AI/algorithm performance study using patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not applicable/provided in the document as it does not describe a study involving expert-derived ground truth for a test set of patient data, which is typical for AI/algorithm performance evaluations. The evaluations performed are engineering-based and comparative.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable/provided as there is no described test set requiring expert adjudication for ground truth.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    There is no indication that a multi-reader multi-case (MRMC) comparative effectiveness study was done. The device (Luna 4 plus) is a physical device for topical aesthetic treatment and does not involve "human readers" interpreting medical cases or AI assistance in diagnostic tasks.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device is a physical, user-operated aesthetic device. While it contains integrated software ("integral software" to control microcurrent and the "Anti-Shock system"), this is part of the device's functional control, not a standalone diagnostic or interpretative algorithm. Therefore, a "standalone algorithm-only performance study" as typically understood for AI/ML devices is not applicable/provided.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The concept of "ground truth" as typically used for AI/ML models is not directly applicable here. The "truth" for this device's evaluation is primarily based on:

    • Reference standards: Adherence to national and international engineering and safety standards (e.g., IEC 60601 series, ISO 10993 series).
    • Benchmarking against predicate devices: Demonstrating similar or improved safety and functionality compared to legally marketed devices.
    • Physical measurements: E.g., optical power density, electrical output characteristics.

    For the usability studies mentioned, the "ground truth" would likely be user feedback or successful completion of tasks, but details are not provided.

    8. The sample size for the training set

    This information is not applicable/provided as the document does not describe the development of an AI/ML model that would require a "training set." The integral software is likely rule-based or control-loop driven rather than a machine learning algorithm.

    9. How the ground truth for the training set was established

    This information is not applicable/provided for the reasons stated in point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240616

    Validate with FDA (Live)

    Device Name
    FAQ™ (102)
    Manufacturer
    Date Cleared
    2024-07-15

    (132 days)

    Product Code
    Regulation Number
    878.4420
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FAQ™ 102 devices are over-the-counter home use devices intended for facial and neck skin stimulation, and non-invasive treatments of mild to moderate facial wrinkles for adult women with all skin tones.

    Device Description

    Not Found

    AI/ML Overview

    I am sorry, but the provided text does not contain any information about acceptance criteria or a study that proves a device meets such criteria. The document is a 510(k) clearance letter from the FDA for a device named FAQ™ (102), stating that the device is substantially equivalent to legally marketed predicate devices. It discusses regulatory requirements, classifications, and indications for use but does not detail performance studies or acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K232242

    Validate with FDA (Live)

    Device Name
    BEAR 2 Body
    Manufacturer
    Date Cleared
    2024-05-13

    (290 days)

    Product Code
    Regulation Number
    882.5890
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended for body skin stimulation and is indicated for over-the-counter cosmetic use.

    Device Description

    Not Found

    AI/ML Overview

    The provided document is an FDA 510(k) clearance letter for the FOREO Inc. BEAR 2 Body device. This letter indicates that the device has been found substantially equivalent to a predicate device and is intended for over-the-counter cosmetic use.

    However, the document does not contain any of the requested information regarding acceptance criteria, study details, sample sizes, ground truth establishment, expert qualifications, or comparative effectiveness studies. The FDA clearance letter itself does not typically include such detailed technical study reports.

    Therefore, I cannot provide the requested information based on the text provided.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240378

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-04-19

    (72 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FAO™ 201 and 202 devices are intended to use LED light for the treatment of wrinkles and mild to moderate acne.

    Device Description

    Not Found

    AI/ML Overview

    The provided text is an FDA 510(k) clearance letter for the Faq™ 201 and Faq™ 202 devices. This type of document determines substantial equivalence to a predicate device and includes the Indications for Use. However, it does not include detailed information about acceptance criteria, study methodologies, sample sizes, ground truth establishment, or multi-reader multi-case studies as requested in the prompt.

    Therefore, I cannot extract the requested information from the provided text. The document focuses on regulatory clearance based on substantial equivalence, not on the specific performance study details that would define and prove acceptance criteria.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231977

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-09-25

    (84 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PEACH 2, PEACH 2 Duo and PEACH 2 Go are over-the-counter device intended for the removal of unwanted body hair and/or facial hair.

    Device Description

    Not Found

    AI/ML Overview

    Based on the provided text, the document is an FDA 510(k) clearance letter for a medical device called PEACH 2, PEACH 2 Duo, and PEACH 2 Go. This letter does not contain information about acceptance criteria or a study proving the device meets those criteria.

    The letter explicitly states:

    • "We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices..."

    This means the clearance is based on a determination of substantial equivalence to existing devices, not necessarily on a new clinical study demonstrating specific performance metrics against acceptance criteria. For 510(k) submissions, manufacturers often rely on non-clinical performance data (e.g., electrical safety, electromagnetic compatibility, biocompatibility, mechanical testing) and comparison to a predicate device's performance to establish substantial equivalence. Clinical studies are often not required for 510(k) clearances unless the device raises new questions of safety or effectiveness.

    Therefore, I cannot provide the requested information for the following reasons:

    1. Acceptance Criteria Table: The document does not specify any performance-based acceptance criteria for the device, nor does it report any device performance metrics from a study.
    2. Study Details: There is no mention of a study (clinical or otherwise rigorous performance study) described in the provided text that would provide information on:
      • Sample size for a test set or training set.
      • Data provenance.
      • Number or qualifications of experts.
      • Adjudication method.
      • MRMC comparative effectiveness study or effect size.
      • Standalone algorithm performance.
      • Type of ground truth used or how it was established.

    The document is a regulatory approval letter, not a study report. To get the information you asked for, you would need to refer to the actual 510(k) submission (if it were publicly available and contained such a study) or the manufacturer's own internal documentation or publications.

    Ask a Question

    Ask a specific question about this device

    K Number
    K162450

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-11-30

    (90 days)

    Product Code
    Regulation Number
    878.4810
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ESPADA Acne-Clearing Blue Light Pen is intended for use as an at-home device for the treatment of mild to moderate inflammatory acne.

    Device Description

    Not Found

    AI/ML Overview

    This PDF does not contain the detailed study information required to answer your request regarding acceptance criteria and device performance for the Espada Acne-clearing Blue Light Pen. The document is an FDA 510(k) clearance letter, which states that the device is substantially equivalent to a legally marketed predicate device. It does not include specific clinical study results, acceptance criteria, or performance data that would typically be found in a clinical trial report or a more detailed submission.

    Therefore, I cannot provide the following information from the given text:

    1. A table of acceptance criteria and the reported device performance: This information is not present.
    2. Sample size used for the test set and the data provenance: Not available.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not available.
    4. Adjudication method for the test set: Not available.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance: Not available. This is also for an AI device, and the Espada Pen is a light therapy device.
    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not available. Again, this is for an AI device.
    7. The type of ground truth used: Not available.
    8. The sample size for the training set: Not available.
    9. How the ground truth for the training set was established: Not available.

    The document primarily focuses on the regulatory clearance of the device based on substantial equivalence. To find the acceptance criteria and study details, you would typically need to consult the original 510(k) submission summary or a clinical study report if one was conducted and made public.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1