510(k) Data Aggregation

    K Number
    K251114

    Validate with FDA (Live)

    Device Name
    FloPatch FP120
    Manufacturer
    Date Cleared
    2025-06-20

    (70 days)

    Product Code
    Regulation Number
    870.2100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FloPatch FP120 is indicated for use for the noninvasive assessment of blood flow in peripheral vasculature. FloPatch FP120 operates in a single mode, the Continuous Wave (CW) mode only.

    The device is intended to be used by medical professionals, such as physicians and nurses, in hospitals and professional environments. The device is intended for prescription use on adults only.

    Device Description

    The FloPatch (FP120) is a non-invasive blood flow detection device to be used in a medical/hospital setting for use by a medical professional. The device uses ultrasound and the Doppler effect to assess the flow of blood. The device consists of a signal processing unit and an adhesive strap. The device transmits ultrasonic waves from the ultrasonic transducer to a peripheral vessel. The Doppler shifted ultrasonic waves are reflected by moving blood cells back to the ultrasonic flow transducer. The reflected signal is received by the signal processing unit which outputs the Doppler signal wirelessly to a mobile medical application. The mobile medical application then processes the Doppler signal and displays a Max Velocity trace, Max VTI (Velocity Time Integral) and the Corrected Flow Time.

    AI/ML Overview

    I cannot provide a specific study that proves the device meets acceptance criteria, as the provided text is a 510(k) clearance letter and 510(k) summary, not a detailed clinical study report. The document describes the device, its intended use, comparison to predicates, and lists performance tests conducted. It asserts that these tests passed, but doesn't provide the detailed methodology, specific acceptance criteria values (beyond general descriptions like "≤720 mW/cm2"), or the results in a manner that would allow for a complete reconstruction of a clinical study or performance evaluation.

    However, I can extract and infer information based on the typical content of a 510(k) submission and the provided text to address your points regarding acceptance criteria and the "study" (referring to the performance testing described).

    It's crucial to understand that 510(k) clearance primarily relies on demonstrating substantial equivalence to a legally marketed predicate device, not necessarily on new clinical efficacy studies. The "performance data" section in this document describes a series of engineering and bench tests, along with usability testing, to support this claim of substantial equivalence for the modifications made to the device.

    Here's an attempt to answer your questions based on the provided FDA 510(k) clearance letter:


    Acceptance Criteria and Device Performance (Inferred from 510(k) Summary)

    The "acceptance criteria" for a 510(k) are generally defined by compliance with recognized standards (e.g., IEC, ISO) and demonstrating that the modified device performs as intended and is as safe and effective as the predicate. The document lists the specific tests and states that they "Pass." The table below summarizes the implied acceptance criteria based on industry standards and the reported (binary: Pass/Fail) device performance from the 510(k) summary.

    1. Table of Acceptance Criteria and Reported Device Performance

    Test CategoryImplied Acceptance Criteria (Based on Standard/Requirement)Reported Device Performance
    Acoustic OutputGlobal maximum derated ISPTA ≤ 720 mW/cm²; Global maximum MI ≤ 1.9Pass (Reported: Max ISPTA.3 – 53.58 mW/cm², Max MI – 3.09E-02)
    Ultrasound SafetyMeets IEC 60601-2-37Pass
    Ultrasound PerformanceMeets performance and ultrasound transducer requirementsPass
    Electrical and Basic Safety TestingMeets IEC 60601-1 requirementsPass
    EMCMeets IEC 60601-1-2:2014+A1:2020Pass
    Emergency Service Environment TestingMeets IEC 60601-1-12 requirementsPass
    Ingress Protection (IP) TestingMeets IP67 ingress requirementsPass
    Material and Shelf LifeAdhesive and device enclosure satisfy performance requirementsPass
    Usability TestingValidated device meets new indications for usePass
    Hardware TestingMeets mechanical and electrical requirementsPass
    Software TestingSoftware application satisfies functional requirements, performs as intended; Algorithms and calculations verified.Pass
    Biocompatibility TestingAdhesive and device enclosure met biocompatibility requirements (ISO 10993-1, -5, -10, -12, -21)Pass
    Simulated Use Cycles (Reprocessing)Device can be reprocessed (soiling, cleaning, disinfection) without adverse effects on physical integrity or battery runtime.Pass
    Low Level Disinfection (LLD) with Visual InspectionValidate a low level disinfection process during reprocessing.Pass
    Physical IntegrityRepeated reprocessing does not adversely affect physical integrity.Pass
    Cumulative RuntimeRepeated reprocessing does not adversely affect battery runtime.Pass

    Study Details (Inferred/Not Applicable)

    The document refers to "Performance Data" which are a series of tests to demonstrate substantial equivalence, rather than a single, comprehensive "study" with a defined clinical protocol to prove efficacy or diagnostic accuracy in the way a clinical trial would.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not specified in the provided text. These tests are typically bench and engineering tests, and possibly a limited usability test. They are not clinical trials with large patient cohorts. For example, "Usability Testing" likely involves a small number of representative users, not a large patient sample.
    • Data Provenance: The document does not specify the country of origin of the data for these performance tests. Given that Flosonics Medical is based in Canada and the submission is to the US FDA, the tests could have been conducted in Canada, the US, or other locations. These are primarily engineering/bench tests, not retrospective or prospective clinical data in the traditional sense, except possibly for the usability testing which would be prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • For bench and engineering tests (Acoustic Output, Electrical Safety, EMC, etc.), "ground truth" is typically established by calibrated measurement devices and compliance with international standards, not by human experts.
    • For "Usability Testing," it's implied that medical professionals (physicians and nurses) were involved as the intended users, but the specific number or their qualifications are not stated. They would provide feedback on the user interface and operation.

    4. Adjudication Method for the Test Set

    • For engineering and bench tests, adjudication methods (like 2+1, 3+1 consensus) are not typically applicable. The results are quantitative measurements against defined thresholds.
    • For Usability Testing, the "adjudication" would involve assessment against predefined usability goals and heuristics, likely by product designers and human factors specialists. It's not a consensus reading of medical images.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

    • No, an MRMC comparative effectiveness study was not done or reported in this 510(k) summary. This type of study is more common for AI/ML-driven diagnostic imaging devices where human reading performance with and without AI assistance is being evaluated.
    • The FloPatch FP120 is described as a "Cardiovascular blood flowmeter" that uses ultrasound and the Doppler effect to assess blood flow. While it has "Software including but not limited to: Algorithm changes," it's not presented as a device that assists human readers in interpreting complex medical images in the way an AI diagnostic algorithm would. Its purpose is to output calculated values (Max Velocity trace, Max VTI, Corrected Flow Time). Therefore, an MRMC study comparing human readers with AI assistance would not be applicable to this device's reported function.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • The software testing explicitly states: "Verified that the software application satisfies functional requirements and performs as intended. Algorithms and calculations were also verified." This constitutes a form of standalone performance evaluation for the algorithms and calculations embedded in the device/app.
    • The device outputs quantitative values (Max Velocity trace, Max VTI, Corrected Flow Time), which implies its core function is to produce quantitative readings based on its algorithms, rather than requiring subjective human interpretation of raw data.

    7. The Type of Ground Truth Used

    • Benchmarking/Reference Standards: For the technical performance tests (acoustic output, electrical safety, EMC, etc.), the ground truth is established by adherence to recognized international standards (e.g., IEC 60601 series, ISO 10993) and comparison to calibrated reference instruments.
    • Engineering Specifications: For hardware and software testing, the ground truth is adherence to the device's predefined engineering specifications and functional requirements.
    • User Feedback/Expert Opinion: For usability testing, the 'ground truth' comes from medical professionals' feedback on the device's ease of use and ability to perform its function in a simulated environment, validating that it "meets new indications for use."

    8. The Sample Size for the Training Set

    • Not Applicable / Not Provided. This document describes a 510(k) for a medical device that uses "Algorithm changes," but it does not specify if these algorithms are machine learning/AI algorithms that require a "training set" in the typical sense. Even if ML components are present, information on training sets is not commonly disclosed in the high-level 510(k) summary provided in a clearance letter. The focus here is on demonstrating the device's overall performance and safety after modifications, not on describing its underlying ML development process.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable / Not Provided. As the existence and nature of a "training set" are not specified, neither is the method for establishing its ground truth.
    Ask a Question

    Ask a specific question about this device

    K Number
    K223843

    Validate with FDA (Live)

    Device Name
    FloPatch FP120
    Manufacturer
    Date Cleared
    2023-05-03

    (132 days)

    Product Code
    Regulation Number
    870.2100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FloPatch FP120 is indicated for use for the noninvasive assessment of blood flow in the carotid artery. FloPatch FP120 operates in a single mode, the Continuous Wave (CW) mode, and is not capable of operating in any other mode.

    The device is intended to be used by medical professionals, such as physicians and nurses, in hospitals and professional environments. The device is intended for prescription use on adults only.

    Device Description

    The FloPatch (FP120) is a non-invasive blood flow detection device to be used in a medical/hospital setting for use by a medical professional. The device uses ultrasound and the Doppler effect to assess the flow of blood. The device consists of a signal processing unit and an adhesive strap. The device transmits ultrasonic waves from the ultrasonic transducer to a peripheral vessel such as the carotid artery. The Doppler shifted ultrasonic waves are reflected by moving blood cells back to the ultrasonic flow transducer. The reflected signal is received by the signal processing unit which outputs the Doppler signal wirelessly to a mobile medical application. The mobile medical application then processes the Doppler signal and displays a Max Velocity trace, Max VTI (Velocity Time Integral) and the Corrected Flow Time.

    AI/ML Overview

    The provided text is a 510(k) summary for the FloPatch FP120 device, focusing on modifications related to reprocessing of "Single-Use Only" devices. It refers to a previous 510(k) submission (K222242) for the original device. This document primarily details performance data for the reprocessing capabilities of the device, rather than the initial clinical performance or diagnostic accuracy of the device itself.

    Therefore, much of the requested information regarding acceptance criteria for device performance and the study proving it (e.g., sample sizes for test/training sets, expert qualifications, MRMC studies, standalone performance, type of ground truth) is not directly available in the provided document. The document focuses on demonstrating that reprocessing does not degrade the device's technical characteristics, assuming its original performance was already established in K222242.

    However, I can extract the information related to the acceptance criteria and the study for the reprocessing modifications.

    Here's a summary based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance (for reprocessing modifications):

    The acceptance criteria for the reprocessing modifications were that repeated reprocessing does not adversely affect specific device characteristics. The reported performance confirms that these criteria were met.

    TestBrief DescriptionAcceptance CriteriaReported Performance
    Simulated Use CyclesVerified that the device can be reprocessed up to six cycles (consisting of soiling, cleaning and disinfection).Device functions correctly after up to six reprocessing cycles.Pass
    Manual CleaningValidate a manual cleaning process during the reprocessing of the FloPatch FP120.Manual cleaning process is validated and effective.Pass
    Low Level DisinfectionValidate a low level disinfection process during the reprocessing of the FloPatch FP120.Low level disinfection process is validated and effective.Pass
    Acoustic OutputVerified that repeated reprocessing does not adversely affect the acoustic output of the transducer.Acoustic output unaffected by reprocessing.Pass
    Transducer PerformanceVerified that repeated reprocessing does not adversely affect transducer performance characteristics: • Velocity Accuracy • Flow Velocity and DepthTransducer performance characteristics (Velocity Accuracy, Flow Velocity & Depth) unaffected by reprocessing.Pass
    Physical IntegrityVerified that repeated reprocessing does not adversely affect the physical integrity of the FloPatch FP120.Physical integrity unaffected by reprocessing.Pass
    Cumulative RuntimeVerified that repeated reprocessing does not adversely affect the battery runtime.Battery runtime unaffected by reprocessing.Pass

    2. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: Not explicitly stated in this document. The description only mentions "repeated reprocessing" and "up to six cycles," implying a limited number of devices were tested for these engineering/validation tests.
    • Data Provenance: Not specified. These are likely internal validation studies conducted by Flosonics Medical.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable for these engineering/reprocessing validation tests. Ground truth would be based on objective measurements and established standards for cleaning, disinfection, and device performance specifications.

    4. Adjudication method for the test set:

    • Not applicable. The tests described are objective engineering and validation tests, not subjective expert reviews.

    5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done:

    • No, an MRMC comparative effectiveness study was not done, as this submission is for reprocessing modifications, not the initial clinical efficacy of the device's diagnostic capabilities.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No, this refers to the effects of reprocessing on the physical device and its components, not the performance of an algorithm. The device "outputs the Doppler signal wirelessly to a mobile medical application. The mobile medical application then processes the Doppler signal and displays a Max Velocity trace, Max VTI (Velocity Time Integral) and the Corrected Flow Time." The performance of this application and its underlying algorithm would have been assessed in the K222242 submission, not in this document.

    7. The type of ground truth used:

    • For these reprocessing tests, the "ground truth" would be the pre-reprocessing performance specifications of the device and established engineering/biomedical standards for cleaning, disinfection, acoustic output, and physical integrity.

    8. The sample size for the training set:

    • Not applicable. These are validation tests for reprocessing, not machine learning model training.

    9. How the ground truth for the training set was established:

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K222242

    Validate with FDA (Live)

    Device Name
    FloPatch FP120
    Manufacturer
    Date Cleared
    2022-12-09

    (136 days)

    Product Code
    Regulation Number
    870.2100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FloPatch FP120 is indicated for use for the noninvasive assessment of blood flow in the carotid artery. FloPatch FP120 operates in a single mode, the Continuous Wave (CW) mode, and is not capable of operating in any other mode.

    The device is intended to be used by medical professionals, such as physicians and nurses, in hospitals and professional environments. The device is intended for prescription use on adults only.

    Device Description

    The FloPatch (FP120) is a non-invasive blood flow detection device to be used in a medical/hospital setting for use by a medical professional. The device uses ultrasound and the Doppler effect to assess the flow of blood. The device consists of a signal processing unit and an adhesive strap. The device transmits ultrasonic waves from the ultrasonic transducer to a peripheral vessel such as the carotid artery. The Doppler shifted ultrasonic waves are reflected by moving blood cells back to the ultrasonic flow transducer. The reflected signal is received by the signal processing unit which outputs the Doppler signal wirelessly to a mobile medical application. The mobile medical application then processes the Doppler signal and displays a Max Velocity trace, Max VTI (Velocity Time Integral) and the Corrected Flow Time.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting studies for the FloPatch FP120 device, structured to answer your specific questions:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission describes product modifications to an already cleared device and demonstrates substantial equivalence through various tests. The acceptance criteria, while not explicitly stated as numerical values for all tests, are implied by a "Pass" result for each test performed.

    Acceptance Criteria (Implied)Reported Device Performance
    Device meets Class B EMC requirements (CISPR B).Pass
    Adhesive meets biocompatibility requirements.Pass
    Adhesive satisfies material and shelf life performance testing.Pass
    Packaging meets ISTA3 and performance testing.Pass
    Device meets applied part BF electrical safety requirements.Pass
    Software application satisfies functional requirements and performs as intended; algorithms and calculations were verified.Pass
    Device meets IP6X dust ingress requirements.Pass
    Device meets transport medical equipment requirements (shock, vibration, free fall).Pass
    Global maximum derated ISPTA ≤ 720 mW/cm² and global maximum MI ≤ 1.9 (Acoustic Output).Pass
    Device meets IEC 60601-2-37 for ultrasound safety.Pass

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not provide specific sample sizes for the test sets for the various engineering validation and verification tests. The data provenance (e.g., country of origin, retrospective/prospective) is also not specified in this summary. These tests appear to be laboratory-based engineering and software validation tests rather than clinical studies with patient data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not provided in the document. Given that the tests are primarily engineering and software verification, the "ground truth" would likely be established by technical specifications and standards rather than expert clinical consensus.

    4. Adjudication Method for the Test Set

    This information is not provided. For engineering and software tests, adjudication usually involves technical review against predefined specifications rather than a multi-expert clinical review process.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance was not mentioned or described in this document. The device, FloPatch FP120, is a cardiovascular blood flowmeter that directly assesses blood flow and does not appear to involve human interpretation of images or data that would typically necessitate such a study.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    The "Software Testing" mentioned includes verification that "Algorithms and calculations were also verified." This implies that the algorithms were tested for their accuracy in isolation (standalone), but the details of this testing (e.g., specific metrics, dataset) are not provided. The device itself is user-operated by medical professionals, so an entirely "standalone" clinical performance might not be applicable in the same way it would be for an AI-powered diagnostic image analysis tool.

    7. The Type of Ground Truth Used

    For the engineering tests (EMC, Biocompatibility, Material/Shelf Life, Packaging, Electrical Safety, Dust Ingress, Emergency Services Equipment, Acoustic Output, Ultrasound Safety), the ground truth is based on regulatory standards and recognized industry specifications (e.g., Class B requirements, ISTA3, applied part BF requirements, IP6X, IEC 60601-1, IEC 60601-2-37, global maximum derated ISPTA, global maximum MI limits).

    For "Software Testing," the ground truth for "Algorithms and calculations" would be based on mathematical correctness and predefined functional requirements of the device's calculations (Max Velocity trace, Max VTI, Corrected Flow Time).

    8. The Sample Size for the Training Set

    The document does not mention any training sets. This submission is for modifications to a previously cleared device, not for the development of a new AI model that typically requires a training set. The "Software Testing" focuses on verification of existing algorithms and calculations rather than training a new machine learning model.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is mentioned, this information is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200337

    Validate with FDA (Live)

    Device Name
    FloPatch (FP120)
    Date Cleared
    2020-03-24

    (42 days)

    Product Code
    Regulation Number
    870.2100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FloPatch FP120 is indicated for use for the noninvasive assessment of blood flow in the carotid artery. FloPatch FP120 operates in a single mode, the Continuous Wave (CW) mode, and is not capable of operating in any other mode.

    The device is intended to be used by medical professionals, such as physicians and nurses, in hospitals and professional environments. The device is intended for prescription use on adults only.

    Device Description

    The FloPatch (FP120) is a non-invasive blood flow detection device to be used in a medical/hospital setting for use by a medical professional. The device uses ultrasound and the Doppler effect to assess the flow of blood. The device consists of a signal processing unit and an adhesive strap. The device transmits ultrasonic waves from the ultrasonic transducer to a peripheral vessel such as the carotid artery. The Doppler shifted ultrasonic waves are reflected by moving blood cells back to the ultrasonic flow transducer. The reflected signal is received by the signal processing unit which outputs the Doppler signal wirelessly to a mobile medical application. The mobile medical application then processes the Doppler signal and displays a Max Velocity trace, Max VTI (Velocity Time Integral) and the Corrected Flow Time.

    AI/ML Overview

    The provided text is a 510(k) summary for the FloPatch (FP120) and mainly focuses on showing substantial equivalence to predicate devices based on intended use, technology, and compliance with recognized standards. It lacks detailed information about specific acceptance criteria for device performance and the study that proves those criteria are met.

    However, based on the information provided, here's what can be extracted and inferred:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state quantitative acceptance criteria for device performance (e.g., accuracy, precision) or provide a table with these criteria alongside reported performance data. It broadly mentions compliance with consensus standards and performance verification testing.

    It states:

    "The FloPatch has been evaluated to and found compliant with recognized consensus standards for EMC, electrical, thermal & mechanical safety. Additionally, the device has been evaluated to and complies with the requirements of the recognized consensus standard for basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment. Further testing was conducted to verify performance, labelling, packaging and shelf life. The results of the performance testing to recognized consensus standards demonstrate that the characteristics of the FloPatch FP120 are equivalent to the recognized predicate(s)."

    This indicates that acceptance criteria are tied to:

    • Compliance with electrical safety, EMC, thermal, and mechanical safety standards (e.g., ES60601-1, IEC 60601-1-2, IEC 60601-2-37).
    • Biocompatibility (ISO 10993 standards).
    • General performance verification (likely comparing its outputs like Max Velocity, VTI, and Corrected Flow Time to a reference method, though the specific targets and results are not detailed).

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    This information is not provided in the given text.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This information is not provided in the given text.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This information is not provided in the given text.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The FloPatch (FP120) is described as a "non-invasive blood flow detection device" that processes Doppler signals and displays metrics (Max Velocity trace, Max VTI, Corrected Flow Time) on a mobile medical application. It is used by medical professionals. The document does not describe an MRMC comparative effectiveness study comparing human readers with and without AI assistance or any AI components in its operation that would involve human "readers" or "interpretation" by AI beyond signal processing and data display. Its function appears to be a measurement device rather than an AI-assisted diagnostic interpretation tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    While the device operates as an "algorithm only" in the sense that it processes signals and calculates metrics, the document does not specifically describe a "standalone" performance study in the context typically used for AI/CADe devices, which would involve evaluating its agreement with a ground truth independently of human intervention. It states that "the mobile medical application then processes the Doppler signal and displays a Max Velocity trace, Max VTI (Velocity Time Integral) and the Corrected Flow Time," indicating algorithm-driven data presentation. Therefore, it implicitly operates in a standalone manner to generate these outputs, but a formal "standalone study" with detailed metrics is not described.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    This information is not explicitly stated in the provided text. For a blood flowmeter, ground truth would typically be established using a highly accurate, often invasive, or more complex reference method for blood flow measurement.

    8. The sample size for the training set

    This information is not provided in the given text.

    9. How the ground truth for the training set was established

    This information is not provided in the given text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191388

    Validate with FDA (Live)

    Device Name
    FloPatch (FP110)
    Date Cleared
    2019-06-21

    (28 days)

    Product Code
    Regulation Number
    870.2100
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FloPatch(FP110) is intended for the detection of blood flow in peripheral vasculature.

    The device is intended to be used by medical professionals, such as physicians and nurses, in hospitals and professional environments. The device is intended for prescription use only.

    Device Description

    The FloPatch (FP110) is a non-invasive blood flow detection device intended to be used in a medical/hospital setting for use by medical professionals. The device uses ultrasound and the Doppler effect to evaluate the flow of blood. The device consists of signal processing box (main unit) and an ultrasonic vascular flow transducer. The device transmits ultrasonic waves from the vascular flow transducer to a peripheral vessel such as the carotid artery. The Doppler shifted ultrasonic waves are reflected by moving blood cells back to the ultrasonic flow transducer. The reflected signal is received by the signal processing box (main unit) which outputs the doppler signal to the device speaker.

    AI/ML Overview

    I understand that you're asking for details about the acceptance criteria and the study that proves a device meets those criteria, specifically concerning the FloPatch (FP110) based on the provided FDA 510(k) summary.

    However, after carefully reviewing the provided text, I can determine the following:

    • Trade/Device Name: FloPatch (FP110)
    • Regulation Number: 21 CFR 870.2100
    • Regulation Name: Cardiovascular blood flowmeter
    • Regulatory Class: Class II
    • Product Code: DPW
    • Intended Use: The FloPatch (FP110) is intended for the detection of blood flow in peripheral vasculature. It is intended to be used by medical professionals (physicians and nurses) in hospitals and professional environments and is for prescription use only.

    The document primarily focuses on demonstrating substantial equivalence to a predicate device (Edan SD3 Vascular Doppler, K140579) based on design features, intended use, and compliance with various non-clinical performance standards (EMC, electrical, thermal & mechanical safety, biocompatibility, acoustic output, etc.).

    The provided text does not contain the following information:

    1. A table of acceptance criteria and the reported device performance. While it mentions "performance verification testing" and that the device "complies with the requirements of the recognized consensus standard for basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment," specific quantitative acceptance criteria and the FloPatch's performance against them are not detailed.
    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective).
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts.
    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, or the effect size of how much human readers improve with AI vs. without AI assistance. (This device does not appear to involve AI/algorithms for "human improvement" as it's a basic blood flow detection device).
    6. If a standalone study (i.e., algorithm only without human-in-the-loop performance) was done. (Again, not applicable to the type of device described).
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.) for any performance studies.
    8. The sample size for the training set.
    9. How the ground truth for the training set was established.

    Conclusion based on the provided text:

    The 510(k) summary indicates that the FloPatch (FP110) was deemed substantially equivalent to its predicate device (Edan SD3 Vascular Doppler) based on the following:

    • Non-clinical Performance Data: The device was tested and found compliant with relevant recognized consensus standards for:

      • ES60601-1:2005/(R)2012 (Medical Electrical Equipment - Basic Safety and Essential Performance)
      • IEC 60601-2-37 Edition 2.1 2015 (Ultrasonic Medical Diagnostic and Monitoring Equipment)
      • IEC 60601-1-2 Edition 4.0 2014-02 (Electromagnetic Disturbances)
      • ISO 10993-1, -5, -10 (Biocompatibility: cytotoxicity, sensitization, skin irritation)
      • IEC 62127-1 Edition 1.1 2013-02 (Ultrasonics - Measurement and Characterization of Medical Ultrasonic Fields)
      • IEC 61161:2013 (Ultrasonics - Power measurement)
      • IEC 62359 Edition 2.1 2017-09 (Ultrasonics - Field characterization - Thermal and mechanical indices)
      • IEC 62304:2006 (Software Life Cycle Process)
    • Performance Verification Testing: The summary generally states that "performance testing to compare the performance of the device to predicate was conducted besides verification of performance, packaging and labelling." It concludes that "The results of the performance testing and testing to recognized consensus standards demonstrate that the characteristics of the FloPatch FP110 are equivalent to the recognized predicate in terms of acoustic output and performance."

    Without further documentation (e.g., specific test reports linked in the 510(k) submission), the detailed acceptance criteria and the specific performance results from clinical or comprehensive bench performance studies are not available in this summary. The substantial equivalence argument relies heavily on adherence to recognized safety and performance standards for similar devices and a comparison of features to the predicate.

    Ask a Question