    Intended Use

    A-Model: The F&P Nova Micro Pillows Mask is intended to be used by adults weighing ≥ 66 lbs (30kgs) who have been prescribed non-invasive positive airway pressure therapy such as CPAP or bi-level by a physician. The F&P Nova Micro Pillows Mask is intended for single-patient use in the home.
    SLA-Model: The F&P Nova Micro Pillows Mask is intended to be used by adults weighing ≥ 66 lbs (30kgs) who have been prescribed non-invasive positive airway pressure therapy such as CPAP or bi-level by a physician. The F&P Nova Micro Pillows Mask is intended for single-patient use in the home and for multiple-patient use in the hospital or other clinical settings where proper disinfection of the device can occur between patient uses.

    Device Description

    The F&P Nova Micro Pillows Mask is a non-invasive, Positive Airway Pressure (PAP) therapy nasal pillows mask with a silicone seal that seals the nasal airway entrance of the Nova Micro Pillows Mask is designed to aid in the delivery of PAP by providing an interface between the flow generator and tubing, and the patient. The Nova Micro Pillows Mask features a pillows cushion with prongs that extrance of the patient's nasal nares, held in place by adjustable headgear straps. The Nova Micro Pillows Mask is a prescription-only device, provided in a non- sterile state. The F&P Nova Micro Pillows Maskrange is available in three cushion sizes – Small, Medium, and Large. The mask has two models: A-Model and Sleeplab (SL) A-Models are identical except for their Indications for Use, Operating Environment, and Reusability. The A-Model is intended to be single-patient use in the home, while the SL A-Model is intended to be used on multiple patients in a hospital or other clinical setting where proper disinfection of the device can occur between patient uses.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA regarding the F&P Nova Micro Pillows Mask. It primarily focuses on demonstrating substantial equivalence to a predicate device based on indications for use, technological characteristics, and non-clinical performance data.

    However, the provided text does not contain the specific information requested about acceptance criteria and a study proving a device meets these criteria in the context of AI/ML performance. This document describes a medical device (a CPAP mask) that is not an AI/ML device according to the information provided. Therefore, there is no discussion of performance metrics like accuracy, specificity, sensitivity, or the methodology of an AI/ML study (e.g., sample size for AI training/test sets, expert adjudication methods, MRMC studies, or ground truth establishment).

    The "Performance Data" section (Section VII) lists various non-clinical tests performed on the mask, such as Cleaning Validation, Leak, Dead Space Analysis, CO2 Rebreathing, Pressure-Flow Curve, Resistance to Flow, Vibration and Noise, Human Factors/Usability Engineering, Mechanical Integrity, and Shelf-Life. These are standard engineering and safety tests for a physical medical device, not performance evaluations for an AI/ML algorithm.

    Therefore, I cannot provide the requested table or details about an AI/ML study from the given text. The information is simply not present because the device described is not an AI/ML device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173060

    Validate with FDA (Live)

    Date Cleared
    2018-01-25

    (119 days)

    Product Code
    Regulation Number
    868.5895
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Fisher & Paykel Healthcare single patient use masks are intended for use as an accessory to ventilators to enable noninvasive positive pressure ventilation (NPV) therapy (CPAP or bi-level) to be delivered to spontaneously breathing adult patients (>30kg) with respiratory insufficiency or respiratory failure who have been prescribed NPPV. The masks are to be fitted and therapy maintained by trained medical practitioners in a hospital/institutional environment with patient monitoring in place.

    Device Description

    The Nivairo™ RT046 Non-Vented Hospital Full Face Mask, Standard Elbow Version (herein referred to as RT046) is a non-vented hospital full face mask with a standard elbow for use with dual limb circuits. The RT046 is a single use device intended for use as an accessory to deliver non-invasive positive pressure ventilation (NPPV) to a patient as part of a passively vented non-invasive ventilation system.

    The RT046 is an oronasal full face mask intended to be used in a hospital/institutional environment by trained medical staff with patient monitoring systems in place. It connects to a dual limb breathing circuit via a 22mm female swivel adaptor to receive pressurized breathing gases from an external flow source or ventilator (CPAP or Bi-Level).

    The RT046 mask is a prescription only device, provided in a non-sterile state.

    AI/ML Overview

    The provided document is a 510(k) premarket notification summary for the F&P Nivairo™ RT046 Non-Vented Full Face Mask. It details the device, its intended use, and its comparison to a predicate device, focusing on non-clinical performance data.

    Here's an analysis based on your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" in the format of a typical performance study with specific device metrics and target values. Instead, it describes general performance testing conducted to demonstrate "substantial equivalence" to a predicate device (RT041). The "reported device performance" is essentially that the RT046's performance was compared to the RT041's and found to be "substantially equivalent," raising no new safety or effectiveness concerns.

    However, it does mention the standards the device was evaluated against, which can be seen as implicit acceptance criteria for certain aspects:

    Aspect TestedAcceptance Criteria (Implied by standards)Reported Device Performance
    Mechanical/Physical PerformanceCompliance with ISO 5356-1:2004 (Conical connectors)Evaluated to the standard, implicitly passed.
    Compliance with Clauses 5.3 and 5.5 of ISO 17510-2:2007 (Masks and Application Accessories)Evaluated to the standard, implicitly passed.
    BiocompatibilityCompliance with ISO 10993-1:2009 (Biological Evaluation of medical devices)Materials passed all parameters.
    Irritation, Sensitization, Cytotoxicity, Genotoxicity, Muscle Implantation, Acute Systemic Toxicity, Particulate Matter, Chemical Characterization, Volatile Organic CompoundsAll tests passed.
    Overall EquivalencePerformance comparable to predicate device RT041Data demonstrate substantial equivalence to RT041. No new questions regarding safety or effectiveness.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "sample size" in terms of number of patients or measurement instances for the non-clinical performance tests. These are typically bench tests. The data provenance is Fisher & Paykel Healthcare Limited, New Zealand. The tests are "non-clinical," implying they were conducted in a lab/bench setting, not on human subjects.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. This is a medical device for respiratory support, and the performance data presented are non-clinical (bench testing) and not related to diagnostic accuracy or human interpretation. Therefore, there's no "ground truth" to be established by experts in this context.

    4. Adjudication Method for the Test Set

    Not applicable, as there is no expert ground truth or ambiguity to adjudicate for these non-clinical performance tests.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No. The document explicitly states: "Clinical performance testing was not required to demonstrate substantial equivalence for the RT046." This indicates that no MRMC study, or any clinical study involving human readers/users, was performed.

    6. Standalone Performance (Algorithm Only Without Human-in-the-Loop Performance)

    Yes, in the sense that the performance testing described is focused on the device itself (its mechanical properties, biocompatibility) rather than its interaction with a human user in a clinical effectiveness study. The tests are purely "algorithm only" in that they evaluate the device's inherent characteristics against standards and a predicate, without human variability in effectiveness being a primary measurement.

    7. Type of Ground Truth Used

    For the non-clinical performance data, the "ground truth" is established by:

    • Standards Compliance: Meeting the requirements of international standards like ISO 5356-1, ISO 10993-1, and ISO 17510-2.
    • Predicate Device Performance: The performance characteristics of the legally marketed predicate device (RT041) serve as a benchmark for substantial equivalence.

    8. Sample Size for the Training Set

    Not applicable. There is no mention of a "training set" as this is not an AI/machine learning device. The document describes performance testing of a physical medical device.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Fisher & Paykel Healthcare single patient use masks are intended for use as an accessory to ventilators to enable noninvasive positive pressure ventilation (NPPV) therapy (CPAP or bi-level) to be delivered to spontaneously breathing adult patients (>30 kg) with respiratory insufficiency or respiratory failure who have been prescribed NPPV. The masks are to be fitted and therapy maintained by trained medical practitioners in a hospital/institutional environment with patient monitoring in place.

    Device Description

    The Nivairo™ RT045 Non-Vented Hospital Full Face Mask Anti-Asphyxiation Valve (herein referred to as RT045) is a non-vented hospital full face mask with anti-asphyxiation valve for use with single limb circuits. The RT045 is a single use device intended to deliver noninvasive positive pressure ventilation (NPPV) to a patient as part of a passively vented noninvasive ventilation system.

    The RT045 is an oronasal full face mask intended to be used in a hospital/institutional environment by trained medical staff with patient monitoring systems in place. It connects to a single limb breathing circuit via a 22mm female swivel adaptor to receive pressurized breathing gases from an external flow source or ventilator (CPAP or Bi-Level).

    The RT045 mask is a prescription only device, provided in a non-sterile state.

    AI/ML Overview

    I am sorry, but the provided text is a 510(k) premarket notification for a medical device (Nivairo™ RT045 Non-Vented Hospital Full Face Mask Anti-Asphyxiation Valve). This document primarily focuses on establishing substantial equivalence to a predicate device based on technological characteristics and non-clinical performance data.

    The document states: "Clinical performance testing was not required to demonstrate substantial equivalence for the RT045."

    Therefore, I cannot provide information on:

    • Acceptance criteria and device performance based on a study of the device's effectiveness in clinical use. The document does not describe such a study.
    • Sample size used for the test set and data provenance.
    • Number of experts used to establish ground truth and their qualifications.
    • Adjudication method.
    • Multi-reader multi-case (MRMC) comparative effectiveness study results.
    • Standalone (algorithm-only) performance.
    • Type of ground truth used (expert consensus, pathology, outcomes data, etc.).
    • Sample size for the training set.
    • How the ground truth for the training set was established.

    The document does mention that non-clinical performance data was conducted to demonstrate substantial equivalence. It refers to:

    • Performance testing of the RT045 compared to the RT040 (predicate device).
    • Evaluations against standards:
      • ISO 5356-1:2004 Anaesthetic and respiratory equipment- Conical connectors: Part 1: Cones and sockets.
      • ISO 10993-1:2009 Biological Evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.
      • Clauses 5.3 and 5.5 of ISO 17510-2:2007 Sleep Apnoea Breathing Therapy- Part 2: Masks and Application Accessories.

    However, the specific "acceptance criteria" and "reported device performance" in relation to these non-clinical tests are not detailed in the provided text in a way that would allow for a table similar to what you requested for clinical studies. The document only states that "These data demonstrates substantial equivalence of the RT045 to the RT040. The results of the comparative bench testing do not raise any new questions regarding safety or effectiveness for the RT045."