Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K202467

    Validate with FDA (Live)

    Date Cleared
    2020-12-14

    (109 days)

    Product Code
    Regulation Number
    878.4750
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Powered Articulating Endoscopic Linear Cutter Stapler & Reloads are intended for transection, resection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general abdominal, gynecologic, thoracic, and pediatric surgical procedures.

    Device Description

    The Disposable Powered Articulating Endoscopic Linear Cutter Stapler & Reloads are sterile. single patient use instruments that simultaneously cut and staple tissue through a battery powered firing system. The Power Stapler and Reloads are sterilized by irradiation. The instruments deliver six staggered rows of staples, three on either side of the cut line. The instruments are available in three shaft lengths: compact (192±20mm), regular (252±20mm) and long (352±20mm). The shaft can rotate freely in both directions and incorporates an articulation mechanism, which enables the distal portion of the shaft to pivot to facilitate lateral access to the operative site.

    AI/ML Overview

    This document is a 510(k) Pre-market Notification for a medical device, specifically a "Disposable Powered Articulating Endoscopic Linear Cutter Stapler & Reloads." This type of document is filed with the FDA to demonstrate that a new device is substantially equivalent to a legally marketed predicate device, rather than to prove its safety and effectiveness through a comprehensive clinical study per se.

    Therefore, the requested information regarding "acceptance criteria" and "study that proves the device meets the acceptance criteria" in the context of AI/Software device performance is not applicable to this document. This document describes a physical surgical stapler and its reloads, not a software or AI medical device.

    The "Performance Data" section (Section 9) describes non-clinical performance tests conducted to demonstrate substantial equivalence to a predicate device, not to establish performance criteria for an AI or software algorithm.

    Here's why each of your requested points cannot be answered from this document:

    1. A table of acceptance criteria and the reported device performance: The document lists non-clinical tests (e.g., Visual Appearance, Sterility, Hardness, Anastomosis and Cutting Performance) and states, "During verification testing, all data meets pre-defined criteria." It does not provide specific numerical acceptance criteria or reported performance values for these physical device attributes. It also doesn't relate to AI/software performance.
    2. Sample sized used for the test set and the data provenance: Not applicable. The tests are for a physical device, not an algorithm's test set. There's no "data provenance" in the sense of clinical data (e.g., country of origin, retrospective/prospective).
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for an AI/software device is not relevant here.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-assisted device.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc): Not applicable.
    8. The sample size for the training set: Not applicable. There is no training set mentioned as this is not an AI/software device.
    9. How the ground truth for the training set was established: Not applicable.

    In summary, this 510(k) submission focuses on demonstrating the substantial equivalence of a physical surgical stapler to an existing one. It does not involve AI or software, and therefore, the questions about acceptance criteria and study design for such devices are not relevant to the information provided in this document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K182476

    Validate with FDA (Live)

    Date Cleared
    2019-12-18

    (464 days)

    Product Code
    Regulation Number
    878.4750
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Powered Articulating Endosopic Linear Cutter Stapler & Reloads are intended for transection, resection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general abdominal, gynecologic, thoracic, and pediatric surgical procedures.

    Device Description

    The Disposable Powered Articulating Endoscopic Linear Cutter Stapler & Reloads are sterile, single patient use instruments that simultaneously cut and staple tissue through a battery powered firing system. The Power Stapler and Reloads are sterilized by irradiation. The instruments deliver six staggered rows of staples, three on either side of the cut line. The instruments are available in three shaft lengths: compact (192±20mm), regular (252±20mm) and long (352±20mm). The shaft can rotate freely in both directions and incorporates an articulation mechanism, which enables the distal portion of the shaft to pivot to facilitate lateral access to the operative site.

    AI/ML Overview

    This document describes a 510(k) premarket notification for a medical device, the "Disposable Powered Articulating Endoscopic Linear Cutter Stapler & Reloads." The core of a 510(k) submission is to demonstrate that a new device is "substantially equivalent" to a legally marketed predicate device. This process primarily relies on non-clinical performance testing and comparison of technological characteristics, rather than extensive clinical efficacy studies in the same way a PMA (Pre-Market Approval) process would require.

    Therefore, the input document does not contain the information requested in your prompt regarding acceptance criteria and studies proving the device meets them in the context of diagnostic accuracy or comparative effectiveness with human readers. This is because the device described is a surgical instrument (a stapler), not a diagnostic AI system.

    However, I can extract information related to performance testing that serves as "acceptance criteria" for this type of device to demonstrate substantial equivalence.

    Here's the information derived from the document, tailored as much as possible to your prompt's structure, but with the understanding that it's for a surgical stapler and not an AI diagnostic tool:


    1. A table of acceptance criteria and the reported device performance

    The document lists "Non-Clinical Tests," "Biocompatibility Tests," "Electrical Safety Test," and "EMC Test" that the device "passed." For a 510(k) submission of a surgical stapler, "passing" these tests indicates that the device met predetermined performance specifications (which serve as acceptance criteria) established by the manufacturer and recognized standards. Specific numerical acceptance criteria are generally detailed in the full test reports, which are summarized here.

    Acceptance Criteria CategorySpecific TestReported Device Performance
    Non-Clinical PerformanceAppearancePassed
    FlexibilityPassed
    Assembling capacityPassed
    HardnessPassed
    Anastomosis and cutting performancePassed
    Resistance to PressurePassed
    Safety Device (e.g., empty-reload safety protection)Passed (and feature explicitly mentioned)
    SharpnessPassed
    Surface RoughnessPassed
    Package SealPassed (ASTM F 88/F-15)
    Rotation and ArticulationPassed
    Firing Electric CurrentPassed
    Battery VoltagePassed
    Motor Rotate SpeedPassed
    SizePassed
    Sterility TestPassed (Sterilized by Irradiation, SAL 10^-6)
    Accelerated Aging TestPassed (ASTM F1980-16)
    Dye Penetration TestPassed (ASTM F1929-15, ASTM F3039-15)
    Vacuum Leak TestPassed (ASTM D3078-2002(2013))
    Packaging Resistance Bacteria Performance TestPassed (DIN 58953-6-2010)
    BiocompatibilityIn Vitro Cytotoxicity TestPassed (ISO 10993-5: 2009)
    Skin Sensitization TestPassed (ISO 10993-10:2010)
    Intracutaneous reactivity TestPassed (ISO 10993-10:2010)
    Acute Systemic ToxicityPassed (ISO 10993-11:2006)
    Pyrogen TestPassed (ISO 10993-11:2006)
    Subchronic Systemic ToxicityPassed (ISO 10993-11:2006)
    Ames testPassed (ISO 10993-3: 2014)
    In Vitro Mammalian Chromosome Aberration TestPassed (ISO 10993-3:2014)
    In Vivo Mammalian Erythrocyte Micronucleus TestPassed (ISO 10993-3:2014)
    Muscle Implantation TestPassed (ISO 10993-6: 2016)
    Electrical SafetyElectrical Safety TestPassed (AAMI / ANSI ES60601-1:2005/(R)2012 and standards)
    Electromagnetic Comp.EMC TestPassed (IEC60601-1-2:2014)

    2. Sample size used for the test set and the data provenance

    The document does not specify exact sample sizes for each non-clinical test. For devices like this, sample sizes would typically be determined by statistical methods or industry standards for each specific test (e.g., number of stapler firings, number of animals for biocompatibility tests). There is no mention of "data provenance" in terms of country of origin or retrospective/prospective as this is not a diagnostic AI study. The "test set" here refers to the physical devices and materials used in the non-clinical and biocompatibility testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable to the provided document. The device is a surgical stapler, not a diagnostic AI system that requires expert interpretation for ground truth establishment. The "ground truth" for stapler performance would be objective measurements (e.g., staple formation integrity, seal strength, cutting effectiveness) against predefined engineering specifications and relevant ISO/ASTM standards.

    4. Adjudication method for the test set

    This question is not applicable to the provided document. Adjudication methods (like 2+1 or 3+1) are used in diagnostic studies to resolve discrepancies in expert interpretations to establish a ground truth. For a surgical stapler, performance is evaluated against objective, measurable criteria, not subjective interpretation.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable to the provided document. An MRMC study is relevant for diagnostic AI tools, not surgical instruments. The device's performance is not about human 'readers' or AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable to the provided document. This device is a powered surgical instrument; there is no "algorithm only" or "human-in-the-loop" performance in the context of diagnostic assessment. It functions as a tool operated by a surgeon.

    7. The type of ground truth used

    For this surgical stapler, the "ground truth" is established through:

    • Engineering Specifications and Performance Standards: The device's physical dimensions, electrical characteristics, firing mechanism, and output (staple form, cut line) are compared against pre-defined engineering requirements and relevant international standards (e.g., ISO, ASTM, AAMI).
    • Biocompatibility Standards: Materials are tested according to ISO 10993 series for their safety in contact with the body.
    • Sterility Assurance: Sterilization methods are validated to achieve a specified Sterility Assurance Level (SAL) (10^-6 in this case).
    • Functional Testing: "Anastomosis and cutting performance," "resistance to pressure," "sharpness," etc., are assessed through laboratory and potentially ex-vivo tissue testing against established benchmarks for similar devices.

    8. The sample size for the training set

    This question is not applicable to the provided document. "Training set" refers to data used to train an AI model. This document describes a physical surgical device, not an AI system.

    9. How the ground truth for the training set was established

    This question is not applicable to the provided document, as there is no "training set" for an AI model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180208

    Validate with FDA (Live)

    Date Cleared
    2018-04-09

    (75 days)

    Product Code
    Regulation Number
    876.1500
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Disposable Endoscopic Trocar has applications in abdominal, thoracic, and gynecologic minimally invasive surgical procedures to establish a path of entry for endoscopic instruments. The trocar may be used with or without visualization for primary and secondary insertions.

    Device Description

    The Fengh Medical Disposable Endoscopic Trocar has applications in abdominal, thoracic and gynecologic minimally invasive surgical procedures to establish a path of entry for endoscopic instruments. The trocar may be used with or without visualization for primary and secondary insertions. A Disposable Endoscopic Trocar is mainly composed of Obturator Handle, Outer Seal Release Lever. Trocar Stability Sleeve, Obturator Locking Button, Trocar Stability Sleeve, etc. Materials being used include Stainless 12Cr18Ni9, Stainless 06Cr19Ni10, ABS plastic, PC. All the components of Disposable Endoscopic Trocar are evaluated by cytotoxicity test, sensitization test, intracutaneous mucosa stimulation test, Systemic toxicity(acute) test and Pyrogenicity test. Therefore, it will not cause bad stimulations, allergic reactions or other harm to human tissue or mucosa. Furthermore, no toxic substances will be separated out during the use of Disposable Endoscopic Trocar. Endoscopic surgery is conducted through a cannula that extends across the abdominal wall and provides channels where instruments, such as scopes, retractors and staplers, can be inserted to perform surgery. As part of this procedure, the abdominal cavity is inflated with an insufflation gas, such carbon dioxide, to maintain the abdomen in a distended state and provide increased instrument maneuvering capability within the cavity. Valves are typically provided in the trocars to form seals around the instruments to prevent leakage of the insufflation gas.

    AI/ML Overview

    The provided text describes the acceptance criteria and the results of a study for the Fengh Medical Disposable Endoscopic Trocar.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document lists several "Testing" categories and provides specific criteria along with implied performance (meeting those criteria). However, the "Table 2" under "Substantial Equivalence" merges proposed device characteristics with general requirements. I've extracted the acceptance criteria from the "Testing" section and the detailed comparison table, and for performance, the document generally states that the device is "substantially equivalent" because "testing standards are identical" and "the differences between both devices are insignificant in terms of safety and effectiveness," implying the device met these criteria.

    Acceptance Criteria / Test NameCriterionReported Device Performance (Implied)
    Non-Clinical Testing
    Sterility Test(Implied: Device must be sterile)Met (Included in tests)
    Peel Strength(Implied: Must meet strength requirements)Met (Included in tests)
    Flexibility(Implied: Must meet flexibility requirements)Met (Included in tests)
    Fitness(Implied: Components fit correctly)Met (Included in tests)
    Leakage(Implied: No leakage or minimal leakage as specified)Met (Included in tests, further detailed below)
    RoughnessObturator roughness Ra must not be bigger than 0.4μm.Met (Included in tests)
    Dimension(Implied: Dimensions must be within specified tolerances)Met (Included in tests)
    Appearance(Implied: Must meet visual inspection standards)Met (Included in tests)
    Insertion & Cannula StabilityWhen the obturator pulled out Trocar sleeve, the friction should not exceed 5N.Met (Included in tests)
    Trocar sleeve retention forceApply a pressure of 5N ± 1N on both sides of the thread area, the pull-out force of the sleeve should not less than 10N.Met (Included in tests)
    Endoscope Visualization QualityAccommodate an appropriately sided 0° endoscope.Met (Included in tests)
    Obturator & Cannula CompatibilitySleeve and Obturator should fit well without any interference. Max clearance between sleeve and Obturator: D≤0.3mm. The tip of Obturator should be fully out of sleeve when fit.Met (Included in tests)
    Air Leakage as a whole deviceThere should be no leakage under 4kPa pressure between stopcock and sealing cap.Met (Included in tests)
    Air Leakage with obturator withdrawnChoke valve should have good gas blocking performance. Bubbles should be less than 20 under 4KPa pressure.Met (Included in tests)
    Biocompatibility Testing
    In Vitro Cytotoxicity Test(Implied: Must not show cytotoxicity)Met (Included in tests; document states "will not cause bad stimulations, allergic reactions or other harm")
    Skin Sensitization Test(Implied: Must not cause skin sensitization)Met (Included in tests; document states "will not cause bad stimulations, allergic reactions or other harm")
    Intracutaneous reactivity Test(Implied: Must not cause intracutaneous reactivity)Met (Included in tests; document states "will not cause bad stimulations, allergic reactions or other harm")
    Acute Systemic Toxicity(Implied: Must not cause acute systemic toxicity)Met (Included in tests; document states "will not cause bad stimulations, allergic reactions or other harm")
    Pyrogen Test(Implied: Must be non-pyrogenic)Met (Included in tests; document states "will not cause bad stimulations, allergic reactions or other harm")

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the sample sizes used for each specific test mentioned (e.g., Sterility Test, Peel Strength, Biocompatibility tests). It only lists "Testing" categories and the methods.
    The data provenance is not specified regarding country of origin or whether it was retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    This information is not provided in the document. The document describes non-clinical and biocompatibility tests, which typically rely on standardized laboratory protocols and measurements rather than expert consensus on medical images or patient outcomes.

    4. Adjudication Method for the Test Set:

    This information is not provided. Given the nature of the tests (non-clinical, biocompatibility), adjudication methods common in diagnostic device studies (like 2+1 or 3+1) are not typically applicable.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    No, an MRMC comparative effectiveness study was not done or reported in this document. The document focuses on performance testing of the device itself against established criteria and comparison to a predicate device based on technological characteristics and intended use.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    This question is not applicable as the device is a physical medical instrument (trocar) and not an algorithm or AI-based diagnostic tool.

    7. Type of Ground Truth Used:

    For the non-clinical and biocompatibility tests, the "ground truth" would be objective measurements and adherence to pre-defined physical, chemical, and biological standards (e.g., specific force measurements, absence of microbial growth, absence of toxic reactions). These are not expert consensus, pathology, or outcomes data in the typical sense for diagnostic devices.

    8. Sample Size for the Training Set:

    This question is not applicable as the device is a physical medical instrument and does not involve AI or machine learning models that require a training set.

    9. How the Ground Truth for the Training Set Was Established:

    This question is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1