510(k) Data Aggregation

    K Number
    K221264

    Validate with FDA (Live)

    Date Cleared
    2022-06-01

    (30 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The forceps valve is intended to facilitate passage of an endotherapy device, to prevent leak or back flow of air and/or fluids, and to enable the suction function of a FUJIFILM or Olympus gastrointestinal endoscope.

    Device Description

    The FROG forceps valve forms a tight seal with the biopsy port to prevent leakage of biomaterial, provide easy passage of endotherapy devices, and support the suction function.

    AI/ML Overview

    The provided document is a 510(k) premarket notification letter and a 510(k) summary for the FROG Forceps Valve (VAL1-F1-100). This device is an endoscope accessory and its clearance is based on substantial equivalence to a predicate device (JAZZ Forceps Valve (part of JAZZ Disposable Valve Kit) K210625).

    The document does not describe a study involving an AI/Machine Learning device or a multi-reader multi-case (MRMC) comparative effectiveness study. Instead, it focuses on bench testing, biocompatibility testing, and a comparison of technological characteristics to a predicate device.

    Therefore, the requested information regarding acceptance criteria, performance data, sample sizes, expert involvement, ground truth establishment, and MRMC studies related to an AI device cannot be extracted from this document, as it pertains to a different type of medical device clearance (forceps valve, not AI/ML).

    The information that can be extracted regarding the FROG Forceps Valve is as follows:

    FROG Forceps Valve (VAL1-F1-100) Analysis:

    This device is a physical accessory for endoscopes, not an AI/ML diagnostic or assistive tool. The performance evaluation is based on bench testing and biocompatibility, not on AI model performance metrics or human reader studies.

    1. Table of acceptance criteria and the reported device performance:

    The document states: "Additional performance specifications were evaluated against pre-defined acceptance criteria to demonstrate the effectiveness of the device over its stated 2-year shelf life." However, the specific acceptance criteria and the detailed reported performance values are NOT provided in this summary.

    2. Sample sizes used for the test set and the data provenance:

    • Bench Testing: The document states "Bench testing was conducted to confirm the compatibility of the subject device with Olympus scopes." and "Additional performance specifications were evaluated...". The specific sample sizes for these bench tests are not mentioned.
    • Biocompatibility Testing: The document refers to consensus standards (ISO 10993-1, -5, -7, -10) and FDA guidance for biocompatibility. Specific sample sizes for these tests are not provided.
    • Data Provenance: The tests are likely laboratory-based bench tests and biocompatibility assessments, not clinical data from patients. The manufacturer is FUJIFILM medwork GmbH, located in Höchstadt, Germany, implying the testing would likely be performed in a controlled laboratory environment. Retrospective or prospective classification is not applicable here as it's not a clinical data study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications:

    • This is not applicable as the device is a physical medical device (forceps valve), not an AI/ML diagnostic. Ground truth (e.g., for diagnostic accuracy) is not established by human experts in the context of this device's performance evaluation for FDA clearance. Performance is assessed via engineering and biocompatibility testing.

    4. Adjudication method for the test set:

    • Not applicable for this type of device and performance testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done:

    • No, an MRMC study was NOT done. This type of study is relevant for diagnostic AI tools where human readers' performance with and without AI assistance is evaluated. This document concerns a physical endoscope accessory.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. This is not an algorithm.

    7. The type of ground truth used:

    • For biocompatibility: Adherence to ISO standards and FDA guidance for biological evaluation.
    • For performance: Engineering specifications and functional requirements (e.g., tight seal, easy passage of devices, support suction function). The "ground truth" is based on the device's ability to meet these pre-defined engineering performance criteria.

    8. The sample size for the training set:

    • Not applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K210625

    Validate with FDA (Live)

    Date Cleared
    2021-09-10

    (192 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The JAZZ Disposable Valve Kit is comprised of the JAZZ Suction Valve, JAZZ Air/Water Valve, JAZZ Forceps Valve.

    The JAZZ Suction Valve is intended to control the suction (aspiration of air and fluids) of a FUIFILM gastrointestinal endoscope during endoscopy procedures.

    The JAZZ Air/Water Valve is intended to control air of a FUJIFILM gastrointestinal endoscope for insufflations and water delivery (for distal lens cleaning) during endoscopy procedures.

    The JAZZ Forceps Valve is intended to facilitate passage of an endotherapy device, to prevent a leak or backflow of air and/or fluids, and to enable the suction function of a FUJIFILM gastrointestinal endoscope.

    Device Description

    The JAZZ Disposable Valve Kit is comprised of the JAZZ Suction Valve, Air/Water Valve, and Forceps Valve. These devices are intended for single-use and are supplied sterile, eliminating the need for manual cleaning and reprocessing. These valves are designed to the corresponding port of a FUJIFILM G7 gastrointestinal endoscope. The air/water valve can be activated to control air and water flow, while the suction valve can be activated to aspirate excess fluids or other debris. The forceps valve forms a tight seal with the biopsy port to prevent leakage of biomaterial, provide easy passage of endotherapy devices, and support thesuction function.

    AI/ML Overview

    The provided text describes the JAZZ Disposable Valve Kit, a medical device for use with FUJIFILM gastrointestinal endoscopes. The document is a 510(k) summary submitted to the FDA, asserting substantial equivalence to predicate devices based on performance data.

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't provide explicit "acceptance criteria" in a quantitative format with thresholds for each test. Instead, it states that "The subject devices met performance specifications in the following additional non-clinical tests." and "Bench testing demonstrates that the subject device is as safe and effective as the predicate devices." Implied acceptance is that the device performs functionally and safely, similar to the predicate devices.

    Performance Test (Implied Acceptance Criteria: Met Specifications/Safe and Effective)Reported Device Performance
    BiocompatibilityMet consensus standards (ISO 10993-1, 10993-5, 10993-7, 10993-10) and FDA guidance.
    SterilityMet consensus standards (ISO 11135, ISO 11607-1).
    JAZZ Air/Water Valve - Backflow PreventionValidated for efficiency in preventing backflow to proximal irrigation system as defined by FDA Guidance.
    Spring Force TestingPerformed; met performance specifications (implied by overall conclusion).
    Suction Valve - Suction pressureMet performance specifications.
    Suction Valve - TightnessMet performance specifications.
    Suction Valve - Functional lifeMet performance specifications.
    Suction Valve - Compatibility with endoscopeMet performance specifications.
    Air/Water Valve - Insufflation pressureMet performance specifications.
    Air/Water Valve - Flushing pressureMet performance specifications.
    Air/Water Valve - TightnessMet performance specifications.
    Air/Water Valve - Functional lifeMet performance specifications.
    Air/Water Valve - Compatibility with endoscopeMet performance specifications.
    Forceps Valve - TightnessMet performance specifications.
    Forceps Valve - Compatibility with endoscopeMet performance specifications.
    Forceps Valve - Compatibility with endoscopic instrumentsMet performance specifications.
    Forceps Valve - Ease of useMet performance specifications.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document describes bench testing and validation studies. It does not specify the sample sizes (e.g., number of valves tested for each parameter) used for these tests. The data provenance is not explicitly stated beyond being "bench testing" and "validation" which implies a controlled laboratory environment. There is no indication of clinical data, animal studies, or human subject data. Therefore, the data can be considered prospective in the sense that the tests were conducted specifically for this submission. The "country of origin of the data" is not stated, but the manufacturer is FUJIFILM medwork GmbH (Germany).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:

    The document does not describe any studies involving human experts to establish "ground truth." The performance evaluations are based on non-clinical tests (bench testing) against established standards and internal specifications to demonstrate functional performance and safety.

    4. Adjudication Method for the Test Set:

    Not applicable, as no studies involving human interpretation or subjective adjudication are mentioned. The testing involves objective measurements rather than expert consensus on a "ground truth" derived from images or clinical cases.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

    No, an MRMC comparative effectiveness study was not conducted for this device. This type of study is typically used for diagnostic or screening devices where human readers interpret medical images or data. The JAZZ Disposable Valve Kit is an accessory for an endoscope, and its evaluation focuses on mechanical and functional performance, sterility, and biocompatibility.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done:

    This question is not applicable to the JAZZ Disposable Valve Kit. This device is a mechanical accessory, not an algorithm or AI-driven system. Its performance is inherent in its physical and functional design, not in a standalone algorithm's output.

    7. The Type of Ground Truth Used:

    The "ground truth" for the performance evaluation can be considered the established consensus standards (ISO standards), FDA guidance documents, and internal performance specifications for medical devices of this type. For example, for sterility, the ground truth is that the device must meet the requirements of ISO 11135 and 11607-1. For backflow prevention, the ground truth is compliance with the FDA guidance. For functional aspects like suction pressure or tightness, the ground truth would be predetermined functional thresholds defined by the manufacturer or industry standards.

    8. The Sample Size for the Training Set:

    Not applicable. As a physical medical device (valve kit), there is no "training set" in the context of machine learning or AI algorithms. The design and manufacturing processes are developed based on engineering principles and regulatory requirements, not through data training.

    9. How the Ground Truth for the Training Set was Established:

    Not applicable, as there is no training set for this type of device.