Search Results
Found 15 results
510(k) Data Aggregation
(15 days)
The Sonosite iLOOK Ultrasound System is a general purpose ultrasound system and non-continuous patient monitoring platform intended for use by qualified physicians and healthcare professionals for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include:
- Pediatric
- Peripheral vessel
- Needle Guidance
Modes of operation include: B Mode (B), Color Power Doppler, Color Velocity Doppler, Multi-beam imaging, and combined modes, including duplex imaging: B+C.
This device is indicated for Prescription Use Only.
The Sonosite iLOOK system is intended to be used in medical practices, clinical environments, including healthcare facilities, hospitals, clinics, and clinical point-of-care for diagnosis of patients. The system includes a transducer and a display connected together wirelessly. The transducer is powered by battery, while the display is powered either by battery or by AC electrical power. The clinician is positioned next to the patient and places the transducer onto the patient's body where needed to obtain the desired ultrasound image.
The Sonosite iLOOK Ultrasound System is a fully featured, general-purpose, software-controlled diagnostic ultrasound system utilizing an all-digital architecture. The system is a portable ultrasound device consisting of a display, adjustment handle, wireless L15-5 linear ultrasound transducer, and a transducer holder with a built-in wireless charger.
The system is designed to acquire and display high-resolution, real-time ultrasound data in 2D, Color Power Doppler (CPD), Color Velocity Doppler (CVD), or a combination of these modes. It provides measurement capabilities for anatomical structures and pediatric biometry, offering information for clinical diagnostic purposes.
The system includes a variety of accessories, including an optional stand. Additionally, it features Digital Imaging and Communications in Medicine (DICOM) capabilities, along with general computer communication functionalities, including wireless networking. These capabilities enable the acceptance, transfer, display, storage, and digital processing of ultrasound images and loops.
The system features a single USB port, allowing the use of compatible storage devices, such as memory sticks and hard drives. An AC adapter is included in the stand module, enabling manual connection to AC power.
N/A
Ask a specific question about this device
(32 days)
The Sonosite MT Ultrasound System is a general-purpose ultrasound system intended for use by qualified physicians and healthcare professionals for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include:
- Abdominal
- Adult Cephalic
- Cardiac Adult
- Cardiac Pediatric
- Fetal-OB/GYN
- Musculo-skeletal (Conventional)
- Musculo-skeletal (Superficial)
- Ophthalmic
- Pediatric
- Peripheral Vessel
- Small Organ (breast, thyroid, testicles, prostate)
- Transvaginal
- Needle Guidance
Modes of operation include: B Mode (B), M-Mode (M) (including simultaneous M-mode and anatomical M Mode), PW Doppler (PWD) (including High Pulse Repetition Frequency (HPRF) and simultaneous PWD for certain exam types), Tissue Doppler Imaging (TDI), Continuous Wave Doppler (CWD), Color Power Doppler, Velocity Color Doppler, Color Variance, Tissue Harmonic Imaging (THI), Multi-beam imaging, Steep Needle Profiling, Trapezoid, and combined modes, including duplex and triplex imaging: B+M, B+PWD, B+CWD, B+C, (B+C)+PWD, (B+C)+CWD.
This device is indicated for Prescription Use Only.
The Sonosite MT Ultrasound System is intended to be used in medical practices, clinical environments, including Healthcare facilities, Hospitals, Clinics and clinical point-of-care for diagnosis of patients.
The system is used with a transducer attached and is powered either by battery or by AC electrical power. The clinician is positioned next to the patient and places the transducer onto the patient's body where needed to obtain the desired ultrasound image.
The Sonosite MT is a high-resolution, portable and versatile ultrasound system designed for use on a stand and easy grab-and-go deployment. It is a fully featured, general purpose, software controlled, diagnostic ultrasound system using all digital architecture.
The system is used to acquire and display high-resolution, real-time ultrasound data in B Mode (B), M-Mode (M) (including simultaneous M-mode and anatomical M-Mode), PW Doppler (PWD) (including High Pulse Repetition Frequency (HPRF) and simultaneous PWD for certain exam types), Tissue Doppler Imaging (TDI), Continuous Wave Doppler (CWD), Color Power Doppler, Velocity Color Doppler, Color Variance, Tissue Harmonic Imaging (THI), Multi-beam imaging, Steep Needle Profiling, Trapezoid, and combined modes, including duplex and triplex imaging.
The system includes a variety of accessories including needle guide starter kits (optional). The system also includes an ECG-specific port to support the ECG feature. The non-diagnostic ECG module provides ECG tracing of the cardiac signal synchronized with the ultrasound image. ECG is available only with cardiac exam types with the P5-1 transducer. Sonosite MT also features a Triple Transducer Connect (TTC) that allows simultaneous connection of three transducers to the system. The system also includes an I/O hub to facilitate connection to a network using an Ethernet connection. The I/O hub includes three USB 2.0 ports to connect with the accessories and one USB 3.0 plug to connect the module with the system. Sonosite MT system features a battery charge indicator presented by four LED lights and a scan time remaining display that indicates the battery charge level.
N/A
Ask a specific question about this device
(27 days)
The Sonosite LX Ultrasound System is a general-purpose ultrasound system intended for use by qualified physicians and healthcare professionals for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include:
- Abdominal
- Adult Cephalic
- Neonatal Cephalic
- Cardiac Adult
- Cardiac Pediatric
- Dermatological
- Fetal OB/GYN
- Musculo-skeletal (Conventional)
- Musculo-skeletal (Superficial)
- Ophthalmic
- Pediatric
- Peripheral vessel
- Small Organ (breast, thyroid, testicles, prostate)
- Transesophageal (cardiac)
- Transrectal
- Transvaginal
- Needle Guidance
Modes of operation include: B Mode (B), M-Mode (M) (including simultaneous M-mode and anatomical M Mode), PW Doppler (PWD) (including Pulsed Wave Tissue Doppler), Continuous Wave Doppler (CWD), Color Power Doppler (including Velocity Color Doppler), Tissue Harmonic Imaging (THI), Tissue Doppler Imaging (TDI), and Combined modes, including Triplex imaging: B+M, B+PWD, B+CWD, B+C, (B+C)+PWD, (B+C)+CWD.
This device is indicated for Prescription Use Only.
The Sonosite LX Ultrasound System is intended to be used in medical practices, clinical environments, including Healthcare facilities, Hospitals, Clinics and clinical point-of-care for diagnosis of patients.
The Sonosite LX Ultrasound System is a full featured, general purpose, software controlled, diagnostic ultrasound system used to acquire and display high-resolution, real-time ultrasound data in 2D, M Mode, Pulsed Wave (PW) Doppler, Continuous Wave (CW) Doppler, Color Power Doppler (CPD), and color Doppler (Color) or in a combination of these modes. The Sonosite LX is designed as an integrated unit with its kiosk-style stand.
The system includes a variety of accessories including needle guide starter kits. The system includes USB host support for peripherals such as input devices, storage devices, and an Ethernet port. Input devices include wired and wireless devices. The system also includes an ECG-specific port to support the ECG feature. The non-diagnostic ECG module provides ECG tracing of the cardiac signal synchronized with the ultrasound image.
The provided FDA 510(k) clearance letter for the Sonosite LX Ultrasound System (K251830) describes a general-purpose ultrasound system. As such, the clearance does not detail specific acceptance criteria for diagnostic performance metrics (e.g., sensitivity, specificity, accuracy) that would typically be associated with AI-powered diagnostic devices. Instead, the submission focuses on demonstrating substantial equivalence to predicate devices through technical characteristics and compliance with general safety and performance standards.
However, based on the information provided, we can infer some "acceptance criteria" related to the device's technical specifications and safety, and how the study addresses them.
Here's an analysis based on your requested information:
1. Table of Acceptance Criteria and Reported Device Performance
Since this is a general-purpose ultrasound system and not an AI/diagnostic algorithm, the acceptance criteria are not in the form of diagnostic performance metrics. Instead, they are related to equivalence with predicate devices, safety, and general performance standards.
| Acceptance Criteria (Inferred from Substantial Equivalence and Safety Standards) | Reported Device Performance / Assessment Method |
|---|---|
| Intended Use Equivalence | Reported: "The Sonosite LX Ultrasound System (this submission) and the predicate device is diagnostic ultrasound imaging or fluid flow analysis of the human body."Evaluation: "The intended use of the Sonosite LX is identical to the predicate and reference devices." |
| Indications for Use Equivalence | Reported: Covers a broad range of clinical applications (Abdominal, Cardiac, OB/GYN, Musculoskeletal, etc.), with the addition of a "Dermatological" indication via the new UHF46-20 transducer. This new indication was previously cleared on the reference device (Vevo MD Imaging system K190476).Evaluation: "The Indications for Use are similar as what was cleared on Sonosite LX Ultrasound system (primary predicate- K233597). The dermatological indication introduced with Sonosite LX in this submission is previously cleared on the reference device, Vevo MD Imaging system (K190476). The dermatological indication on UHF46-20 supports vessel and superficial exam types and is an extension of the dermatological indication on UHF48 (K190476)." |
| Technological Characteristics Equivalence | Reported: Similar transducer types, modes of operation, DICOM support, transmit/receive channels, system characteristics (form factor, beamformer, screen, ports, dimensions, weight, battery, ratings, measurement packages). The primary differences noted are the addition of the UHF46-20 transducer, which is equivalent to the UHF48 transducer from the reference device.Evaluation: "The technological characteristics are unchanged from the primary predicate Sonosite LX (K233597) and reference device, Vevo MD imaging system (K190476). The Sonosite LX Ultrasound System employs the same fundamental scientific characteristics as the currently marketed predicate devices." |
| Acoustic Output within FDA Limits | Reported: Global Maximum Outputs including Ispta.3, TI Type, TI Value, MI, Ipa.3@MI Max values are provided (e.g., Ispta.3: 607 mW/cm^2, MI: 1.72). MI & TI are always displayed.Evaluation: "Acoustic output is within the FDA established limits and same as the primary predicate." and "MI & TI are always displayed, and a power management system ensures that they never exceed the derated FDA limits." |
| Patient Contact Materials Biocompatibility | Reported: Materials used for transducers and needle guides are listed. The new material Rexolite (for the UHF46-20 lens) is highlighted.Evaluation: "Rexolite is introduced as lens material for the UHF46-20 on the Sonosite LX system, This UHF48-20 has passed biocompatibility testing per ISO 10993-1. The rexolite material used for the lens is also the same as the one used on Vevo MD's UHF48 and UHF70 transducers (K190476). The rest of the ultrasound material is the same as the material used on the previously cleared transducers under (K233597).""All patient contact materials are biocompatible." |
| Electrical, Thermal, Mechanical, EMC Safety | Reported: The new transducer UHF46-20 was evaluated for these aspects.Evaluation: "The subject ultrasound system Sonosite LX Ultrasound System uses the new transducer UHF46-20, which has been evaluated for electrical, thermal, mechanical, and EMC safety." |
| Cleaning/Disinfection Effectiveness | Reported: Evaluated for Sonosite LX.Evaluation: "cleaning/disinfection... have been evaluated for Sonosite LX..." |
| Compliance with Recognized Standards | Reported: Listed compliance with ISO 10993-1, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-37, IEC 62304, ISO 14971, IEC 62359.Evaluation: "The Sonosite LX Ultrasound System is designed to comply with the following FDA recognized standards." |
| Quality System Compliance | Reported: Mention of Design Phase Reviews, Risk Assessment, Requirements Development, System and Software Verification, Hardware Verification, Safety Compliance Verification, and Clinical Validation.Evaluation: "Assurance of quality was established by employing the following elements of product development..." and "FUJIFILM SonoSite, Inc. believes that the Sonosite LX Ultrasound System is substantially equivalent with regards to safety and effectiveness to the predicate devices." (implies QS compliance supports this) |
| Performance in Clinical Environment | Reported: The device with the new transducer was tested.Evaluation: "The Sonosite LX with new transducer UHF46-20 has been successfully tested in a clinical environment and the device performed as expected." |
Regarding specific AI-related questions (2-9), this clearance letter does not pertain to an AI/ML device for diagnostic interpretation. It's for a general-purpose ultrasound system. Therefore, most of these questions are not applicable to the information provided.
However, I will address them to indicate their non-applicability or provide any remotely relevant information.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not Applicable. This submission is for a general-purpose ultrasound system, not an AI/ML algorithm requiring a specific test set for diagnostic performance. The evaluation was primarily based on engineering testing (electrical, thermal, mechanical, EMC safety, biocompatibility, acoustic output, standards compliance) and comparison to predicate devices, rather than a dataset of patient cases.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable. No ground truth (in the context of diagnostic accuracy) was established for a test set in this submission, as it's not an AI/ML diagnostic algorithm.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable. No test set requiring expert adjudication for diagnostic ground truth.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable. This is not an AI-assisted diagnostic device, so no MRMC study comparing human readers with and without AI assistance was performed or reported.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable. This is a hardware device (ultrasound system), not a standalone algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Not Applicable. The "ground truth" here relates to engineering specifications and safety standards, such as:
- Biocompatibility confirmed by ISO 10993-1 standards (chemical testing, etc.)
- Acoustic output confirmed by IEC 62359 standards.
- Electrical safety confirmed by IEC 60601-1 standards.
- Functional "ground truth" confirmed by successful testing in a clinical environment, meaning it operates as intended for imaging and fluid flow analysis.
8. The sample size for the training set
- Not Applicable. This clearance is not for an AI/ML device that requires a training set.
9. How the ground truth for the training set was established
- Not Applicable. No training set for an AI/ML device.
In summary, the FDA 510(k) clearance for the Sonosite LX Ultrasound System is focused on demonstrating substantial equivalence to pre-existing, legally marketed predicate ultrasound systems and compliance with international safety and performance standards. The "study" proving it meets acceptance criteria primarily involves comprehensive engineering validation and verification testing (electrical, thermal, mechanical, EMC, acoustic output, biocompatibility, cleaning/disinfection) and documented adherence to a robust Quality System, rather than a clinical study evaluating diagnostic accuracy with a patient dataset or the performance of an AI algorithm.
Ask a specific question about this device
(26 days)
The Sonosite PX Ultrasound System is a general purpose ultrasound system intended for use by qualified physicians and healthcare professionals for evaluation by ultrasound imaging of the human body. Specific clinical applications and exam types include: Abdominal, Adult cephalic, Cardiac Adult, Cardiac Pediatric, Fetal - OB/GYN, Musculo-skeletal (Conventional), Musculo-skeletal (Superficial), Ophthalmic, Pediatric, Peripheral Vessel, Small Organ (breast, thyroid, testicles, prostate), Transvaginal, Transesophageal (cardiac), Transrectal, Needle Guidance. Modes of operation include: B Mode (B), M-Mode (M) (including simultaneous M-mode and anatomical M-Mode), PW Doppler Mode (B), M-Mode (M) (Including simultaneo (PWD) (including Pulsed Wave Tissue Doppler), Continuous Wave Doppler (including Velocity Color Dopler), Tissue Harmonic Imaging (THI), Tissue Dopler Imaging (TDI), and Combined modes, including Triplex imaging: B+M, B+PWD, B+CWD, B+C, (B+C)+PWD, (B+C)+CWD. This device is indicated for Prescription Use Only. The Sonosite PX Ultrasound System is intended to be used in medical practices, clinical environments, including Healthcare facilities, Hospitals, Clinics and clinical point-of-care for diagnosis of patients.
The Sonosite PX Ultrasound System is a full featured, general purpose, software controlled, diagnostic ultrasound system used to acquire and display high-resolution, real-time ultrasound data in 2D, M-Mode (including Simultaneous M-Mode and anatomical M-Mode), Pulsed Wave (PW) Doppler (including Pulsed Wave Tissue Doppler), Continuous Wave (CW) Doppler, Color Power Doppler (including Velocity Color Doppler), Tissue Harmonic Imaging or in a combination of these modes. including Triplex imaging. The system includes a variety of accessories including optional needle guide starter kits. The system also includes an ECG-specific port to support the ECG feature. The non-diagnostic ECG module provides ECG tracing of the cardiac signal synchronized with the ultrasound image.
The provided text does not contain information about acceptance criteria for a medical device or a study proving its performance against these criteria. Instead, it is a 510(k) summary for the FUJIFILM Sonosite PX Ultrasound System, establishing its substantial equivalence to previously cleared predicate devices.
The document focuses on:
- Indications for Use: Listing the clinical applications and exam types for which the ultrasound system is intended.
- Technological Characteristics: Comparing the subject device (Sonosite PX Ultrasound System) with its predicate devices (Sonosite PX Ultrasound System K200964 and SonoSite Edge II Ultrasound System K162045) in terms of features like intended use, indications for use (subsets of predicates), transducer types and frequencies, acoustic output, modes of operation, DICOM support, and safety certifications.
- Non-Clinical Tests: Stating that the device underwent evaluations for electrical, thermal, mechanical, and EMC safety, as well as cleaning/disinfection, biocompatibility, and acoustic output, conforming to applicable medical device safety standards. It also mentions compliance with specific ISO and IEC standards.
- Clinical Tests: Explicitly stating that clinical studies were not required to support the determination of substantial equivalence.
Therefore, I cannot provide the requested information regarding acceptance criteria and a study proving performance against them, as it is not present in the given text.
Ask a specific question about this device
(28 days)
The Sonosite PX Ultrasound System is a general purpose ultrasound system intended for use by qualified physicians and healthcare professionals for evaluation by ultrasound imaging of the human body. Specific clinical applications and exam types include:
Abdominal Adult cephalic Cardiac Adult Cardiac Pediatric Fetal - OB/GYN Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Ophthalmic Pediatric Peripheral Vessel Small Organ (breast, thyroid, testicles, prostate) Transvaginal Needle Guidance
This device is indicated for Prescription Use Only.
The Sonosite PX Ultrasound System is intended to be used in medical practices, clinical environments, including Healthicare facilities, Hospitals, Clinics and clinical point-of-care for diagnosis of patients.
The Sonosite PX Ultrasound System is a full featured, general purpose, software controlled, diagnostic ultrasound system used to acquire and display high-resolution, real-time ultrasound data in 2D, M-Mode (including Simultaneous M-Mode), Pulsed Wave (PW) Doppler (including Pulsed Wave Tissue Doppler), Continuous Wave (CW) Doppler, Color Power Doppler (including Velocity Color Doppler), Tissue Harmonic Imaging or in a combination of these modes. The Sonosite PX Ultrasound System also includes needle guidance capability. The system is capable of working with two different types of needle guide brackets available, including 1) Fixed-angle, in-plane brackets where a pair of guidelines are generated that represent the path of the needle and 2) Transverse-angle, out-of-plane brackets (adjustable depths) where the quidelines appear down the center of the imaqe. Each Sonosite PX transducer has an appropriate needle quide bracket kit to support needle quidance with the system.
The provided text is a 510(k) Pre-Market Notification for the FUJIFILM SonoSite PX Ultrasound System. It focuses on demonstrating substantial equivalence to existing predicate devices (SonoSite Edge II Ultrasound System and SonoSite X-Porte Ultrasound System) rather than presenting a study to prove acceptance criteria for a new AI/algorithm-based diagnostic device.
Therefore, many of the requested details about acceptance criteria, device performance metrics, sample sizes, ground truth establishment, and comparative effectiveness studies are not applicable as they are specific to studies evaluating the performance of diagnostic algorithms.
However, I can extract and summarize the information that is present:
1. Table of Acceptance Criteria and Reported Device Performance:
The document does not provide a table of acceptance criteria with specific performance metrics (e.g., sensitivity, specificity, AUC) for an AI or algorithmic function. Instead, it demonstrates substantial equivalence to predicate devices based on technological characteristics and compliance with safety standards.
The "reported device performance" in this context refers to the device's ability to perform various ultrasound imaging modes across different clinical applications, which is considered equivalent to the predicate devices.
| Feature / Criterion | Sonosite PX Ultrasound System (This submission) | Predicate Devices (SonoSite Edge II & X-Porte) |
|---|---|---|
| Intended Use | Diagnostic ultrasound imaging or fluid flow analysis of the human body | Diagnostic ultrasound imaging or fluid flow analysis of the human body |
| Indications for Use (Clinical Applications) | Abdominal, Adult Cephalic, Cardiac Adult, Cardiac Pediatric, Fetal - OB/GYN, Musculo-skeletal (Conventional), Musculo-skeletal (Superficial), Ophthalmic, Pediatric, Peripheral vessel, Small Organ (breast, thyroid, testicles, prostate), Transvaginal, Needle Guidance. | Similar clinical applications (Abdominal, Adult Cephalic, Cardiac Adult, Cardiac Pediatric, Fetal - OB/GYN, Musculo-skeletal (Conventional), Musculo-skeletal (Superficial), Ophthalmic, Pediatric, Small Organ, Trans-vaginal, Peripheral vessel, Neonatal Cephalic, Trans-Rectal (Edge II), Trans-esophageal (Cardiac) (Edge II & X-Porte), Needle Guidance). Some minor differences exist (e.g., PX has "Adult Cephalic" but not "Neonatal Cephalic" explicitly listed in its primary indication table, though it is general purpose). |
| Modes of Operation | B-mode Grayscale Imaging, Tissue Harmonic Imaging, M-mode, Simultaneous M-Mode, Color Power Doppler, Zoom, Combination Modes, Pulsed Wave (PW) Doppler, Continuous Wave (CW) Doppler, Speckle reduction algorithm, SonoMB/MBe Image Compounding, CW Doppler, Velocity Color Doppler, Tissue Doppler Imaging (TDI). | Similar modes, including B-mode, M-mode, PW Doppler, CW Doppler, Color Power Doppler, Tissue Harmonic Imaging, SonoHD2 Noise Reduction (instead of speckle reduction algorithm, though similar function), SonoMB/MBe Image Compounding, Velocity Color Doppler, Tissue Doppler Imaging (TDI). |
| Transducer Frequency Range | 1.0-19.0 MHz | 1.0-15.0 MHz |
| Acoustic Output (Worst Case) | I_spta.3: 607 mW/cm^2 (L12-3); TI Value: 4.87 (P5-1); MI: 1.72 (L12-3); I_pa.3 @MI Max: 793 mW/cm^2 (L15-4). | I_spta.3: 598.9 (HFL50x) [Edge II], 629.3 (P21xp) [X-Porte]; TI Value: 4.98 (rP19x) [Edge II], 4.0 (P21xp) [X-Porte]; MI: 1.7 (rP19x) [Edge II], 1.7 (P21xp) [X-Porte]; I_pa.3 @MI Max: 776 (L38xi) [Edge II], 678 (L38xp) [X-Porte]. |
| Product Safety Certification | Complies with AAMI/ANSI ES60601-1, IEC 60601-2-37, CAN/CSA-C22.2 No. 60601-1, JSA JIS T 0601-1, IEC 61157, NEMA UD 2-2004, IEC 62359. | Complies with similar and relevant versions of safety standards (e.g., AAMI/ANSI ES60601-1, IEC 60601-2-37, NEMA UD2-2004, IEC 62359). |
| EMC Compliance | Complies with IEC 60601-1-2:2014, CISPR 11:2015+AMD1:2016+AMD2:2019, IEC 61000-4 series. | Complies with AAMI / ANSI / IEC 60601-1-2:2007 (R)2012 CISPR 11, IEC 61000-4 pt 2-5 (X-Porte). |
| Number of Transmit/Receive Channels | 128 digital channels (Transmit), 128 digital channels (Receive) | 128 digital channels (Transmit) [Edge II & X-Porte]; 64 digital channels (Receive) [Edge II & X-Porte], with X-Porte also noting "128 digital channels using Synthetic Aperture". |
| Biocompatibility | All patient contact materials are biocompatible (Silicone Rubber, Polysulfone, PolyVinylChloride (PVC), Silicone RTV Adhesive, Silicone). | Patient contact materials listed for transducers and needle guides, similar materials used (e.g., ABS, Polysulfone, Polyurethane, Silicone Rubber, PVC). |
2. Sample size used for the test set and data provenance:
- Not Applicable. The document explicitly states: "The Sonosite PX Ultrasound System and transducers, subject of this submission, did not require clinical studies to support the determination of substantial equivalence." Non-clinical tests were conducted for safety and technological characteristics, not for clinical performance evaluation comparable to an AI algorithm's test set.
3. Number of experts used to establish the ground truth for the test set and qualifications of those experts:
- Not Applicable. No clinical test set requiring expert ground truth was conducted for this submission. The device is a general-purpose ultrasound system, and its functionality relies on the interpretation of images by qualified healthcare professionals.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not Applicable. No clinical test set requiring adjudication was performed.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs. without AI assistance:
- Not Applicable. This is a submission for a general-purpose ultrasound imaging system, not an AI-assisted diagnostic algorithm. Therefore, no MRMC study comparing human readers with and without AI assistance was conducted or mentioned.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not Applicable. This device is an imaging system, not a standalone algorithm. Its intended use is for qualified physicians and healthcare professionals to perform evaluations using the ultrasound images.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not Applicable. As no clinical studies were performed to establish performance metrics for a diagnostic algorithm, no specific ground truth type was established for that purpose. The "ground truth" for demonstrating substantial equivalence was conformity to recognized safety standards and verification of technological functionality against predicate devices.
8. The sample size for the training set:
- Not Applicable. This device is an ultrasound system. While it includes "software controlled" elements and features like "Speckle reduction algorithm" and "SonoMB/MBe Image Compounding," these are traditional image processing techniques inherent to ultrasound technology, not machine learning algorithms that would require a distinct "training set" in the modern AI sense.
9. How the ground truth for the training set was established:
- Not Applicable. See point 8.
Ask a specific question about this device
(16 days)
The Vevo MD Imaging System is a general purpose imaging system intended for use by a qualified physician for evaluation by ultrasound imaging or fluid flow analysis of the human body.
Specific clinical applications and exam types include:
Abdominal
Dermatological
Musculoskeletal (conventional)
Musculoskeletal (superficial)
Neonatal Cephalic
Ophthalmic
Pediatric
Peripheral vessel
Small Organ (breast, thyroid, testicles, prostate)
The Vevo MD Imaging System is a high frequency general purpose, software controlled, diagnostic imaging system used to acquire ad display high-resolution, real-time ultrasound data in 2D, Color Doppler, Power Doppler, PW Doppler and M-Mode. The Vevo MD Imaging System is comprised of transducers responsible for ultrasound signal generation and recording, and a main unit that controls the transducers, processes the acoustic data, and processes and displays images.
This document, K190476, details the 510(k) premarket notification for the FUJIFILM SonoSite Vevo MD Imaging System.
Here's an analysis based on your requested information:
1. Table of Acceptance Criteria and Reported Device Performance
This 510(k) summary primarily focuses on demonstrating substantial equivalence to a predicate device rather than presenting specific acceptance criteria and performance metrics from device testing. The document states that the device was evaluated for safety and effectiveness against its predicate. The "acceptance criteria" for a 510(k) typically involve demonstrating that the new device is as safe and effective as a legally marketed predicate device.
The study presented does not specify quantitative acceptance criteria (e.g., specific accuracy, sensitivity, specificity thresholds) or report numerical device performance against such metrics. Instead, it relies on demonstrating equivalence through:
- Intended Use Alignment: Both the new device and the primary predicate (FUJIFILM SonoSite Vevo MD Imaging System K160674) share the same intended use.
- Technological Characteristics Comparison: The document provides a table comparing features, modes of operation, acoustic output, DICOM capabilities, and channels with the predicate devices. The new device and primary predicate are highly similar, particularly for "Track 3" devices.
- Compliance with Voluntary Standards: The device is designed to comply with several recognized standards related to biological evaluation, electrical safety, electromagnetic disturbances, ultrasonic medical diagnostic equipment, field characterization, software life cycle, and risk management. This compliance forms a key part of demonstrating safety.
Summary of "Acceptance Criteria" (Implicit) and Reported Performance:
| Category | Implicit Acceptance Criteria (for 510(k) equivalence) | Reported Device Performance (as demonstrated for equivalence) |
|---|---|---|
| Intended Use | Identical or highly similar to predicate device(s). | The Vevo MD Imaging System (this submission) has the same intended use as the predicate Vevo MD Imaging System (K160674): "Diagnostic ultrasound imaging or fluid flow analysis of the human body." Specific clinical applications for both include Abdominal, Dermatological, Musculoskeletal (conventional/superficial), Neonatal Cephalic, Ophthalmic, Pediatric, Peripheral vessel, and Small Organ. |
| Technological Characteristics | No significant differences in features or fundamental scientific technology compared to predicate that would raise new questions of safety or effectiveness. | The Vevo MD Imaging System (this submission) is a "Track 3" device and employs the same fundamental scientific technology as its primary predicate (Vevo MD K160674). Key common features include Linear Array transducer type, 15-49MHz transducer frequency, MI/TI output display, 2D Mode, Color Doppler, M-Mode, Power Doppler, PW Doppler operation modes, DICOM 3.0 support, manual IMT measurement, 64 digital transmit and receive channels. The comparison shows substantial similarity. |
| Safety Standards | Compliance with relevant mandatory and voluntary medical device safety standards. | Evaluated for electrical, thermal, mechanical, and EMC safety. Cleaning/disinfection and biocompatibility were also evaluated. Found to conform to applicable mandatory medical device safety standards. Specifically states compliance with AAMI/ANSI/ISO 10993-1, IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, IEC 62359, IEC 62304, ISO 14971, and NEMA UD 2. All patient contact materials are biocompatible and are either used in the predicate or meet ISO 10993. |
| Clinical Performance | No new clinical concerns or requirement for clinical studies if equivalence demonstrated through non-clinical means. | Clinical studies were not required to support the determination of substantial equivalence. |
2. Sample Size Used for the Test Set and Data Provenance
The document explicitly states: "The Vevo MD Imaging System and transducers, subject of this submission, did not require clinical studies to support the determination of substantial equivalence." This means there was no clinical "test set" in the traditional sense of patient data used to evaluate performance metrics like sensitivity or specificity. The evaluation was primarily non-clinical, comparing technical specifications and adherence to standards against a predicate device. Therefore, no information on sample size or data provenance for a clinical test set is provided.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
As no clinical study was conducted for performance evaluation in this 510(k) submission, there was no "ground truth" to be established by experts for a test set of patient data.
4. Adjudication Method for the Test Set
Not applicable, as no clinical test set for performance evaluation was used.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This device is an imaging system, not an AI-based diagnostic aid that would involve human readers improving with AI assistance. Clinical studies were not performed for this submission.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This device is an imaging system, not a standalone algorithm. Clinical studies were not performed for this submission.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
Not applicable, as no clinical study for performance evaluation was conducted where ground truth would be established. The "ground truth" for this submission revolves around the technical specifications and safety profile of the device aligning with regulated standards and the predicate device.
8. The sample size for the training set
Not applicable. As this 510(k) is for an ultrasound imaging system and not for an AI/ML-driven device or an algorithm requiring a training set, no information about a training set size is provided.
9. How the ground truth for the training set was established
Not applicable, as there was no training set.
Ask a specific question about this device
(79 days)
The SonoSite SII Ultrasound System is a general purpose ultrasound system intended for use by qualified physicians and healthcare professionals for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include:
Ophthalmic
Fetal - OB/GYN
Abdominal
Pediatric
Small Organ (breast, thyroid, testicle, prostate)
Neonatal Cephalic
Adult Cephalic
Trans-rectal
Trans-vaginal
Musculo-skeletal (Conventional)
Musculo-skeletal (Superficial)
Cardiac Adult
Cardiac Pediatric
Peripheral Vessel
The SonoSite SII Ultrasound System is a mountable style, full featured, general purpose, software controlled, diagnostic ultrasound system used to acquire and display high-resolution, real-time ultrasound data through multiple imaging modes. SII is a custom fabricated digital electronic design that readily lends itself to be configured for specific ultrasound imaging applications through different system feature selections. SII can operate on either battery or AC power.
This is an FDA 510(k) premarket notification for the FUJIFILM SonoSite, Inc. SonoSite SII Ultrasound System. The document focuses on establishing substantial equivalence to previously cleared predicate devices, rather than presenting a standalone study with specific acceptance criteria and performance metrics for a novel AI device.
Therefore, the information you're requesting regarding acceptance criteria, study details, sample sizes, expert qualifications, and comparison with human readers is not directly available in this type of regulatory submission for a standard ultrasound system. This document is a clearance for a medical device that does not appear to involve AI/ML components for diagnostic interpretation that would necessitate such detailed performance studies.
However, I can extract information relevant to the device's characteristics and how its substantial equivalence was determined based on what is provided:
1. A table of acceptance criteria and the reported device performance
The document doesn't present a table of specific acceptance criteria in numerical performance metrics typically seen with AI devices (e.g., sensitivity, specificity, AUC). Instead, substantial equivalence is claimed based on similar technological characteristics, intended uses, and compliance with recognized safety and performance standards to the predicate devices.
Here's a summary of the characteristics of the SonoSite SII Ultrasound System as reported in the submission, compared to its predicates (K162045 SonoSite SII Ultrasound System and K162045 SonoSite Edge II Ultrasound System):
| Feature | SonoSite SII Ultrasound System (This submission) | SonoSite SII Ultrasound System (K162045 - Predicate) | SonoSite Edge II Ultrasound System (K162045 - Predicate) |
|---|---|---|---|
| Intended Use | Diagnostic ultrasound imaging or fluid flow analysis of the human body | Diagnostic ultrasound imaging or fluid flow analysis of the human body | Diagnostic ultrasound imaging or fluid flow analysis of the human body |
| Indications for Use (Key Applications) | Ophthalmic, Fetal - OB/GYN, Abdominal, Pediatric, Small Organ (breast, thyroid, testicle, prostate), Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Musculo-skeletal (Conventional), Musculo-skeletal (Superficial), Cardiac Adult, Cardiac Pediatric, Peripheral Vessel, Needle guidance | Ophthalmic, Fetal - OB/GYN, Abdominal, Pediatric, Small Organ (breast, thyroid, testicle, prostate), Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Musculo-skeletal (Conventional), Musculo-skeletal (Superficial), Cardiac Adult, Cardiac Pediatric, Peripheral Vessel, Needle guidance | Ophthalmic, Fetal - OB/GYN, Abdominal, Pediatric, Small Organ (breast, thyroid, testicle, prostate), Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Musculo-skeletal (Conventional), Musculo-skeletal (Superficial), Cardiac Adult, Cardiac Pediatric, Trans-esophageal (cardiac), Peripheral Vessel, Needle guidance |
| Transducer Types | Linear Array, Curved Linear Array, Intracavitary, Phased Array | Linear Array, Curved Linear Array, Intracavitary, Phased Array | Linear Array, Curved Linear Array, Intracavitary, Phased Array, Trans-esophageal |
| Transducer Frequency | 1.0 — 15.0 MHz | 1.0 – 15.0 MHz | 1.0 – 15.0 MHz |
| Global Maximum Outputs/Worst Case Setting | Ispta.3: 598.9 (HFL50x), TI Type: TIB (rP19x), TI Value: 4.98 (rP19x), MI: 1.7 (rP19x), Ipa.3@MI Max: 776 (L38xi) | Ispta.3: 420.2 (L38xi), TI Type: TIC (rC60xi), TI Value: 3.05 (rC60xi), MI: 1.7 (rP19x), Ipa.3@MI Max: 776 (L38xi) | Ispta.3: 598.9 (HFL50x), TI Type: TIB (rP19x), TI Value: 4.98 (rP19x), MI: 1.7 (rP19x), Ipa.3@MI Max: 776 (L38xi) |
| Modes of Operation (Key differences) | Includes PW Doppler (available on all except P11x), CW Doppler (available on P10x and rP19x), Tissue Doppler Imaging (TDI) | No PW or CW Doppler listed for K162045 SII. Modes listed: B-mode Grayscale, THI, M-mode, Color M-Mode, Color Power Doppler, Zoom, Combination Modes, SonoHD2 Noise Reduction, SonoMB/MBe, Compounding, Velocity Color Doppler. | Includes PW and CW Doppler, Tissue Doppler Imaging (TDI) |
| ECG Feature | 3-lead ECG input, or ECG Slave Cable | Not available | 3-lead ECG input, or ECG Slave Cable |
| DICOM Version | DICOM 3.0 Store, Print, Modality Worklist, Perform Procedure Step (PPS), Storage Commitment (NEMA PS3.15 2011) | DICOM 3.0 Store, Print, Modality Worklist, Perform Procedure Step (PPS), Storage Commitment (NEMA PS3.15 2003) | DICOM 3.0 Store, Print, Modality Worklist, Perform Procedure Step (PPS), Storage Commitment (NEMA PS3.15 2011) |
| Product Safety Certification | AAMI/ANSI ES60601-1:2005 (R2012), IEC 60601-2-37 Ed. 2.1:2015, CAN/CSA C22.2 No. 60601-1:08, NEMA UD2-2004, IEC 62359 (2017 - although form states 2015 for 60601-2-37 and 2017 for 62359 in standard listing) | AAMI/ANSI ES60601-1:2005 (R2012), IEC 60601-2-37: 2007, CAN/CSA C22.2 No. 60601-1:08, NEMA UD2-2004, IEC 62359:2010 | AAMI/ANSI ES60601-1:2005 (R2012), IEC 60601-2-37 Ed. 2.1 :2015, CAN/CSA C22.2 No. 60601-1:08, NEMA UD2-2004, IEC 62359:2017 |
Acceptance Criteria for Substantial Equivalence (General, inferred from 510(k) process):
- Identical Intended Use: The device must have the same intended use as a legally marketed predicate device.
- Similar Technological Characteristics: The device must have similar technological characteristics to the predicate, or if there are differences, these differences must not raise new questions of safety or effectiveness.
- Equivalent Safety and Effectiveness: The device must be as safe and effective as the predicate device. This is often demonstrated through adherence to recognized standards for electrical, thermal, mechanical, EMC safety, cleaning/disinfection, biocompatibility, and acoustic output.
- Compliance with Standards: Conformance to relevant national and international medical device safety and performance standards (e.g., ISO, IEC, NEMA UD).
Reported Performance (Relevant to ultrasound system functionality, not AI-specific):
The performance of the device is assessed against the established standards and its predicate's performance. The submission states: "The SonoSite SII Ultrasound System and predicates meet FDA requirements for Track 3 devices, share indications for use, have biosafety equivalence, and conform to applicable electromedical device safety standards."
New "N" indications for specific transducers (found in Tables 1.3-1 to 1.3-14) for Pulsed Wave Doppler (PWD), and in some cases, Color Doppler modes show the scope of the device's capabilities are being expanded or specifically detailed for certain transducers.
2. Sample size used for the test set and the data provenance:
The document explicitly states: "The SonoSite SII Ultrasound System, subject of this submission, did not require clinical studies to support the determination of substantial equivalence."
Therefore, there is no test set sample size or data provenance related to clinical performance metrics for this device as it is not an AI/ML device requiring such validation for regulatory clearance. The assessment is based on technical, safety, and performance characteristic comparison.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
Not applicable, as no clinical study or test set with ground truth established by experts was conducted or required for this 510(k) submission.
4. Adjudication method for the test set:
Not applicable, as no clinical study or test set requiring adjudication was conducted or required for this 510(k) submission.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable, as this device is a general-purpose ultrasound system and does not appear to incorporate AI assistance for diagnostic interpretation. Hence, no MRMC study or AI assistance effect size is mentioned.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
Not applicable, as this is a medical imaging device as opposed to an AI algorithm.
7. The type of ground truth used:
Not applicable, as no clinical study requiring ground truth for clinical decision or diagnostic performance was conducted. The "ground truth" in this context refers to compliance with performance and safety standards, and equivalence to predicate devices, rather than diagnostic accuracy.
8. The sample size for the training set:
Not applicable, as this is a traditional ultrasound system and not an AI/ML device that requires a training set.
9. How the ground truth for the training set was established:
Not applicable, for the same reason as above.
Ask a specific question about this device
(25 days)
The SonoSite iViz Ultrasound System is a general purpose ultrasound system and non-continuous patient monitoring platform intended for use in clinical care by qualified physicians and healthcare professionals for evaluation by ultrasound imaging or fluid flow analysis. Specific clinical applications and exam types include:
Fetal - OB/GYN Abdominal Pediatric Small Organ (breast, thyroid, testicles, prostate) Musculo-skel. (Convent.) Musculo-skel. (Superfic.) Cardiac Adult Cardiac Pediatric Peripheral vessel Ophthalmic
The SonoSite iViz Ultrasound System is a highly featured, general purpose, software controlled, diagnostic ultrasound system used to acquire and display high-resolution, real-time ultrasound data through multiple imaging modes. iViz is a custom fabricated digital electronic handheld tablet that is highly portable, battery-operated, and consists of an active transducer that connects to and is controlled by the tablet. iViz supports Bluetooth and wireless network connectivity for image transfer and over-the-air (OTA) software updates.
The provided document is a 510(k) premarket notification for the FUJIFILM SonoSite iViz Ultrasound System. The purpose of this document is to demonstrate that the iViz Ultrasound System is substantially equivalent to legally marketed predicate devices.
Here's an analysis of the acceptance criteria and the study information as requested:
1. Table of Acceptance Criteria and Reported Device Performance:
The document does not explicitly state acceptance criteria in terms of specific performance metrics (e.g., accuracy, sensitivity, specificity, image quality scores) for the device's diagnostic capabilities. Instead, the "acceptance criteria" are implied through the comparisons to predicate devices and adherence to recognized standards.
The primary method to demonstrate acceptance is through substantial equivalence to existing, legally marketed predicate devices by showing comparable:
- Intended Use
- Indications for Use
- Technological Characteristics
- Safety and Effectiveness (via non-clinical testing)
Below is a table summarizing the general "acceptance criteria" (implicitly by comparison to predicates and standards) and the device's reported performance (by asserting compliance and equivalence).
| Acceptance Criteria (Implicit) | Reported Device Performance |
|---|---|
| Intended Use Equivalence: Device intended for diagnostic ultrasound imaging or fluid flow analysis of the human body, comparable to predicate devices. | The SonoSite iViz Ultrasound System has the same intended use as the predicate devices: "a general purpose ultrasound system and non-continuous patient monitoring platform intended for use in clinical care by qualified physicians and healthcare professionals for evaluation by ultrasound imaging or fluid flow analysis." |
| Indications for Use Equivalence: Device supported clinical applications align with or expand upon those of predicate devices. | The SonoSite iViz Ultrasound System and its various transducers (L38v/10-5 MHz, P21v/5-1 MHz, C60v/5-2 MHz, L25v/13-6 MHz) cover a range of clinical applications including Fetal, Abdominal, Pediatric, Small Organ, Musculo-skeletal, Cardiac Adult/Pediatric, Peripheral vessel, and Ophthalmic. Many of these are "P" (previously cleared) from identified predicate devices (K162288, K133454, K162045, K152983), and some are "N" (new indications) for specific transducers. The document asserts these are consistent with predicates. |
| Technological Characteristics Equivalence: Fundamental scientific technology, transducer types, frequency ranges, and modes of operation are comparable to predicate devices. | The SonoSite iViz Ultrasound System employs the "same fundamental scientific technology" as the SonoSite iViz, Edge, and Edge II Ultrasound systems (Track 3 devices). Comparison tables (Pg 9) detail comparable transducer types (Linear Array, Phased Array, Curved Array), frequency ranges (1.0 - 13.0 MHz), and modes of operation (B-mode, M-mode, Color Power Doppler, SonoHD3 Noise Reduction, etc.), with some differences noted (e.g., PW and CW Doppler not available on iViz but available on older Edge models). |
| Safety and Effectiveness: Compliance with recognized electrical, thermal, mechanical, EMC, cleaning/disinfection, biocompatibility, and acoustic output standards. | The iViz Ultrasound System was evaluated for electrical, thermal, mechanical, and EMC safety, cleaning/disinfection, biocompatibility, and acoustic output. It was found to conform to applicable mandatory medical device safety standards (e.g., ISO 10993-1, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-37, IEC 62304, IEC 62359, ISO 14971, NEMA UD 2-2004). Quality assurance processes were also followed. |
| No new questions of safety or effectiveness are raised. | The conclusion explicitly states that "FUJIFILM SonoSite, Inc. believes that the iViz Ultrasound System is substantially equivalent with regard to safety and effectiveness to the predicate devices." |
Study Proving Acceptance Criteria:
The document states: "The iViz Ultrasound System and transducers, subject of this submission, did not require clinical studies to support the determination of substantial equivalence."
Instead, the determination of substantial equivalence is based on non-clinical tests and a comparison of technological characteristics and intended uses with identified predicate devices.
2. Sample Size Used for the Test Set and Data Provenance:
No clinical test set was used, as no clinical studies were deemed necessary. The acceptance was based on non-clinical testing and comparison to predicates.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
Not applicable, as no clinical test set requiring expert ground truth was utilized.
4. Adjudication Method for the Test Set:
Not applicable, as no clinical test set requiring adjudication was utilized.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
Not applicable. This device is a diagnostic ultrasound system, not an AI-assisted diagnostic tool for interpretation by human readers. The document makes no mention of AI assistance.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
Not applicable. The device is a diagnostic ultrasound system, operated by a qualified physician or healthcare professional. It is not a standalone algorithm.
7. The Type of Ground Truth Used:
For the non-clinical tests (e.g., electrical, thermal, EMC, acoustic output), the "ground truth" would be the specifications and requirements defined by the referenced industry standards (e.g., IEC 60601 series, ISO 10993, NEMA UD 2-2004). The device's performance against these standards served as the "ground truth" for non-clinical acceptance.
8. The Sample Size for the Training Set:
Not applicable. The document does not describe the development of an AI model or algorithm that would require a "training set."
9. How the Ground Truth for the Training Set Was Established:
Not applicable, as there was no mention of a training set for an AI model.
Ask a specific question about this device
(27 days)
The SonoSite X-Porte Ultrasound System is a general purpose ultrasound system intended for use by qualified physicians and healthcare professionals for evaluation by ultrasound imaging of thid flow analysis of the human body. Specific clinical applications and exam types include:
Ophthalmic Fetal – OB/GYN Abdominal Pediatric Small Organ (breast, thyroid, testicles, prostate) Neonatal Cephalic Adult Cephalic Trans-vaginal Musculo-skel. (Convent.) Musculo-skel. (Superfic.) Cardiac Adult Cardiac Pediatric Trans-esophageal (card.) Peripheral Vessel
The SonoSite X-Porte Ultrasound System is a highly mobile, full featured, general purpostic ultrasound system used to acquire and display high-resolution, real-time ultrasound data through multiple imaging modes. X-Porte is a custom fabricated digital electronic design that readily lends itself to be configured for specific ultrasound imaging applications through different system feature selections. The system interface can be customized for the user and controlled using a backlit touchscreen much like what is used in consumer tablet products. X-Porte can be operated in two different configurations, standbased with AC power or battery, and desktop-based with AC power only. In desktop configuration the ultrasound engine can be removed from the stand and used by itself with a single transducer and external monitor.
The provided documentation is a 510(k) Premarket Notification for the FUJIFILM SonoSite, Inc. SonoSite X-Porte Ultrasound System. This document focuses on demonstrating substantial equivalence to existing legally marketed devices, rather than establishing performance criteria through novel clinical studies. Therefore, much of the requested information regarding algorithm performance, sample sizes for test/training sets, expert qualifications, and MRMC studies is not present.
However, based on the provided text, we can extract details regarding the general acceptance criteria (primarily safety and compliance with standards) and the "study" (non-clinical tests and comparisons to predicates) used to demonstrate adherence to these criteria.
Here's a breakdown of the available information:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are primarily conformance to various medical device safety and performance standards. The "reported device performance" is implicitly that the device meets these standards, enabling a substantial equivalence determination.
| Acceptance Criteria (Standards Adherence) | Reported Device Performance |
|---|---|
| Product Safety Certification: | Conforms to specified standards |
| AAMI/ANSI ES60601-1:2005 (R2012) | |
| IEC 60601-2-37: 2007 | |
| CAN/CSA C22.2 No. 601.1 | |
| JIS T 0601-1, JIS T 1507 | |
| CEI/IEC 61157 | |
| ANSI/AAMI EC53 | |
| NEMA UD2-2004 | |
| IEC 62359:2010 | |
| Biocompatibility ISO 10993-1:2009/(R)2013 | All patient contact materials are biocompatible (implicit) |
| Risk Management ISO 14971: 2007 | Employed in product development (implicit) |
| Electrical, thermal, mechanical safety evaluation | Evaluated (implicit) |
| Acoustic Output Measurement Standard (NEMA UD 2-2004) | Conforms, specific maximum outputs provided |
| EMC Compliance: | Conforms to specified standards |
| IEC 60601-1-2:2007 | |
| CISPR 11 | |
| IEC 61000-4 pt 2-5 | |
| DIOCOM Compliance: | Conforms to specified standard |
| NEMA PS3.15:2011 | |
| Airborne Equipment Standards: | Conforms to specified standard |
| RTCA/DO160D (section 21) | |
| Indications for Use: (e.g., Ophthalmic, Fetal, Abdominal, etc., across various modes for the system and for each transducer type) | Device is intended for these specific clinical applications and exam types |
| Technological Characteristics: (e.g., Transducer Types, Frequency, Maximum Outputs, Modes of Operation, PW/CW Doppler features, ECG, DICOM, Channels, Patient Contact Materials, System Characteristics) | The new device has comparable technological characteristics to predicate devices |
2. Sample size used for the test set and the data provenance
- Test Set Sample Size: Not applicable in the context of this 510(k) submission. The submission relies on non-clinical testing and comparison to predicates, not a clinical test set for AI performance.
- Data Provenance: Not applicable for performance evaluation in the usual sense. The "data" here refers to measurements and tests conducted to ensure compliance with standards, and these are typically internal engineering tests. The document states "Clinical Validation" was part of product development, but no details are provided about this "validation" with respect to a test set.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. "Ground truth" in the clinical performance sense (e.g., expert consensus on diagnoses) is not described or required for this type of 510(k) submission focused on substantial equivalence of an ultrasound system, not an algorithmic diagnostic device. The "truth" is established by adherence to engineering and safety standards.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not applicable. There is no mention of an adjudication method as no clinical test set for diagnostic accuracy is described.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a submission for an ultrasound imaging system, not an AI-assisted diagnostic algorithm. There is no MRMC study mentioned or required for this type of device.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
Not applicable. This is not an algorithmic diagnostic device, but rather a general-purpose ultrasound system.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" or basis for the acceptance of this device primarily stems from:
- Engineering and safety standards compliance (e.g., IEC 60601-1, NEMA UD2-2004 for acoustic output).
- Biocompatibility testing for patient contact materials.
- Risk assessment outcomes.
- Comparison of technological characteristics and intended uses to previously cleared predicate devices (K152209, K133454, K162045).
8. The sample size for the training set
Not applicable. This document describes an ultrasound hardware system, not an AI/algorithm that requires a training set.
9. How the ground truth for the training set was established
Not applicable, as no training set for an algorithm is described.
Ask a specific question about this device
(24 days)
The SonoSite iViz Ultrasound System is a general purpose ultrasound system and non-continuous patient monitoring platform intended in clinical care by qualified physicians and healthcare professionals for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include:
Fetal - OB
Abdominal
Pediatric
Small Organ (breast, thyroid, testicles, prostate)
Musculo-skel. (Convent.)
Musculo-skel. (Superfic.)
Cardiac Adult
Cardiac Pediatric
Peripheral vessel
The SonoSite iViz Ultrasound System is a highly featured, general purpose, software controlled, diagnostic ultrasound system used to acquire and display high-resolution, real-time ultrasound data through multiple imaging modes. iViz is a custom fabricated digital electronic handheld tablet that is highly portable, battery-operated, and consists of an active transducer that connects to and is controlled by the tablet. iViz supports Bluetooth and wireless network connectivity for image transfer and over-the-air (OTA) software updates.
The provided document is a 510(k) premarket notification for the FUJIFILM SonoSite iViz Ultrasound System. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than proving de novo safety and effectiveness through clinical trials with defined acceptance criteria and study designs.
Therefore, the document does not contain the information requested regarding acceptance criteria and a study proving the device meets those criteria, specifically concerning the performance of an AI or algorithm. This notification is for an ultrasound imaging system, which produces images for human interpretation, not an AI-powered diagnostic device that would have specific performance metrics like sensitivity, specificity, or reader improvement.
Here's why the requested information cannot be extracted from this document:
- No AI/Algorithm: The document describes a general-purpose ultrasound system. There is no mention of any integrated AI, machine learning algorithm, or software that performs automated diagnoses or assists in interpretation beyond standard imaging functionalities (e.g., SonoHD3 Imaging for speckle reduction, which is a signal processing technique, not an AI for diagnosis).
- Substantial Equivalence: The core of a 510(k) submission is to show that the new device is "substantially equivalent" to a legally marketed device (predicate) in terms of intended use, technological characteristics, and safety/effectiveness. This typically involves non-clinical tests (electrical safety, EMC, biocompatibility, acoustic output, and adherence to standards) rather than clinical performance studies for diagnostic accuracy against predefined acceptance criteria for an AI.
- No Clinical Performance Data on "AI": The document explicitly states: "The iViz Ultrasound System and transducers, subject of this submission, did not require clinical studies to support the determination of substantial equivalence." This further confirms that no clinical performance data (like sensitivity, specificity, or MRMC studies) related to an AI or algorithm's diagnostic performance was generated or presented in this submission.
- "Acceptance Criteria" for a PURE Imaging Device: For an ultrasound imaging system, acceptance criteria would revolve around image quality (resolution, contrast, penetration, etc.), safety (acoustic output, electrical safety), usability, and reliability, rather than diagnostic accuracy metrics of an AI. These are typically assessed through engineering tests and comparison to the predicate's performance specifications, not through studies with expert ground truth establishment for a test set of diagnoses.
In summary, as this document pertains to a general diagnostic ultrasound system and not an AI/ML-powered diagnostic device, the requested information (AI acceptance criteria, test set, experts, MRMC studies, etc.) is not applicable and therefore not present.
Ask a specific question about this device
Page 1 of 2