Found 2 results

510(k) Data Aggregation

    K Number
    K112944

    Validate with FDA (Live)

    Date Cleared
    2011-12-30

    (87 days)

    Product Code
    Regulation Number
    890.3110
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medical Treatment Chair, T688 Series is intended for use in medical procedures such The Medical Preatinent Chalysis to, and collecting blood from, patients in hospital departments or home use, under the supervision of trained medical staff. The Medical Treatment Chair is also intended for use in day surgery and nursing homes. T688 chair is I readined so that the occupant is accommodated in a seated position with the hips moved designed so that the occupant is against the back rest and the legs outstretched and supported by the seat and leg rests.

    T688 chair is also used to position patients for easy access by healthcare professionals. The I voo enair is designed so that the occupant is accommodated in a seated position with the hips enair is decired or that the occupant's back is against the back rest and the legs outstretched and supported by the seat and leg rests.

    This chair is intended to be used by patients with a weight not exceeding:

    • 200kg (T688 series) .
    Device Description

    T688 series chair is used to aid medical procedures such as renal dialysis, blood collection and chemotherapy and is called as Medical Treatment Chair.

    T688 series is electric-powered treatment chair with the following features:

    • Power-operated seat and leg rest and back rest reclines, with options offering one-. touch memorized positions
    • Swing Out Side/Arms .
    • Adjustable head/neck rest .
    • Adjustable height .
    • CPR support posts, gas spring assisted back rest and accessories. .
    • . Fold-down trays
    • Four individually locking castors .
    • Heavy duty washable vinyl .
    • Removable upholstery .
    • The chair will recline to Trendelenburg position .
    • Suitable for side transfer of patients .
    • No timber included in the frame .
    • Battery back-up-24V rechargeable .
    • 2 Actuators & 1 Column .
    • Compact Design .
    • Lighter weight .
    • Reduced components ●

    The device is classified as:

    • Class II (FKS, INO) electric-powered version. .
      Specific ancillary devices and accessories are not listed here due to the wide variety of procedures in which the chair is used.
    AI/ML Overview

    The Fresenius Medical Treatment Chair T688 Series is a medical treatment chair designed for use in medical procedures such as renal dialysis, blood collection, and chemotherapy, under the supervision of trained medical staff. It is also intended for use in day surgery and nursing homes.

    Here's an analysis of the acceptance criteria and the study proving the device meets these criteria:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Weight Capacity200 kg (T688 series)
    BiocompatibilityMaterials used that may come into contact with patients are biocompatible, conforming to ISO 10993: Biological Evaluation of Medical Devices, Part 1.
    Electrical SafetyMeets applicable requirements of IEC 60601-1 General Safety.
    Electromagnetic Compatibility (EMC)Meets applicable requirements of IEC 60601-1-2 EMC.
    Frame LoadingTested against AS 4688.2:2000 (Furniture - Fixed height chairs. Part 3: Determination of stability - Upright chairs, s8, s7.1)
    Flame RetardancyTested against California 117 sD p2, AS/NZS4088.1:1996, and AS 1530.3-1999 for various aspects of flame retardance.
    Quality System ComplianceISO 13485:2003 (Medical devices -- Quality management systems -- Requirements for regulatory purposes)
    Risk Analysis ComplianceISO 14971:2009 (Medical devices -- Application of risk management to medical devices)
    Castors and BrakesConforms to EN 12526 - 12533 Castors and wheels. Hospital bed castors.
    LabelingConforms to BS EN 1041:2008 and BS EN 980:2008.
    Dimensional Specifications- Length Upright: 930 mm- Length Tren/CPR: 1800 mm- Overall Width: 730 mm- Seat Width: 510 mm- Seat Height to Floor: 590 mm- Backrest Length: 800 mm

    2. Sample Size Used for the Test Set and the Data Provenance

    The document states that "Representative samples for the device underwent load and repeatability testing to verify functional and performance characteristics." It does not specify the exact sample size for these tests. There is no information provided regarding data provenance (e.g., country of origin of the data, retrospective or prospective) for any of the testing mentioned.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Not applicable. This device is a medical treatment chair, and its performance is evaluated against engineering standards and functional requirements, not against ground truth established by experts in a diagnostic context.

    4. Adjudication Method for the Test Set

    Not applicable, as the testing involves engineering and performance standards rather than subjective assessments requiring adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

    No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation of images or data, which is not the case for a medical treatment chair.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Not applicable. The device is a physical medical chair and does not involve an algorithm for standalone performance. Its performance is assessed through physical and electrical testing against established standards.

    7. The Type of Ground Truth Used

    The ground truth for this device is based on engineering standards, functional specifications, and safety requirements. This includes:

    • Specified weight capacity (200 kg).
    • Compliance with various ISO, IEC, AS/NZS, BS EN standards for biocompatibility, electrical safety, EMC, flame retardancy, quality management, risk analysis, and component specifications (e.g., castors).
    • Internal "self-imposed load and repeatability test requirements" to verify functional and performance characteristics.

    8. The Sample Size for the Training Set

    Not applicable. As a physical medical device, there is no "training set" in the context of machine learning or AI.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K091727

    Validate with FDA (Live)

    Date Cleared
    2009-07-06

    (25 days)

    Product Code
    Regulation Number
    890.3110
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Medical Treatment Chairs are intended for use in medical procedures such as the administration of renal dialysis to, and taking bloods from, patients in hospital departments or home use, under the supervision of trained medical staff. The Medical Treatment Chairs are also intended for use in day surgery and nursing homes. The chairs are designed so that the occupant is accommodated in a seated position with the hips moved back so that the occupant's back is against the back rest and the legs outstretched and supported by the seat and leg rests.

    The chairs are also used to position patients for easy access by healthcare professionals. The chairs are designed so that the occupant is accommodated in a seated position with the hips moved back so that the occupant's back is against the back rest and the legs outstretched and supported by the seat and leg rests.

    All models are intended to be used by patients with a weight not exceeding:

    • 130kg (T100 and T200 series)
    • 130kg (T300 and T400 series)
    • 200kg (T500, T550, T600 and T650 series)
    • 300kg (T500B, T600B)
    Device Description

    The range of chairs, used to aid medical procedures such as renal dialysis, blood collection and chemotherapy, are called Medical Treatment Chairs and are made up of models T-100, 200, 300, 400, 550, 500B, 550, 600, 600B, 650:

    The manually-operated range of treatment chairs (T100, 200, 300) typically have the following features:

    • Manually-operated recliner
    • Adjustable head/neck rest
    • Upholstered arms
    • Four locking braked castors
    • Vinyl-covered
    • The T100, T200 and T300 models offer optional CPR support posts, gas spring assisted back rest and accessories, whilst other models come standard with these features
    • Fold-down trays
    • Fold-out arms

    The electric-powered range of treatment chairs (T400 through to T600 series) offer the above features in addition to:

    • power-operated seat and leg rest and back rest reclines, with some models offering one-touch memorized positions
    • battery back-up 24V rechargeable
    • height-adjustable swing-out arm rests (except the T400)

    The electric-powered chairs, depending on the model, may have different numbers of actuators, i.e:

    • T400 2 actuators
    • T500 2 actuators
    • T500B 2 actuators, reinforced frame
    • T550 3 actuators
    • T600 3 actuators
    • T600B 3 actuators, 2 motors, reinforced frame, extra lift levers
    • T650 3 actuators

    The device is classified as:

    • Class I (FMR)
    • Class II (FKS, INO) electric-powered version.

    Specific ancillary devices and accessories are not listed here due to the wide variety of procedures in which the chair is used.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided document:

    Acceptance Criteria and Device Performance

    The document doesn't explicitly list "acceptance criteria" in a typical quantitative format (e.g., "Accuracy > 90%"). Instead, it describes compliance with various standards and self-imposed test requirements for functional and performance characteristics. The "reported device performance" is primarily stated as meeting these standards and requirements.

    Table of Acceptance Criteria and Reported Device Performance

    Critical Area/Acceptance Criteria (Implied)Standard Met/Reported Device Performance
    Quality SystemISO 13485:2003
    Risk AnalysisISO 14971:2000
    Design Control(Not specified, but generally implied by QS)
    Flame RetardancyCalifornia 117 sD p2, AS/NZS4088.1:1996, AS 1530.3-1989
    Frame LoadingAS 4688.2:2000 (Specifically for stability)
    Electrical Safety (Whole Chair)IEC 60601-1
    Electrical Safety (Linak actuators)EN 60601-1-2:2002 (EMC), EMC parts of EN 1970:200, EN 60601-2-38:1997, EN 60601-2-52:2007, IEC 60601-1:1988+A1:91+A2:95
    Electrical Safety (Dewert actuators)EN 60601-1/A2:1995, EN 60601-2-38:1996, EN 60335-1/A2:2006, EN 60601-1-2:2001, EN 1970:2000, EN 60529/A2:2000
    BiocompatibilityISO 10993-1:2003 (Evaluation and testing)
    Castors and BrakesEN 12526 - 12533
    LabelingBS EN 1041:2008, BS EN 980:2003
    Clinical InvestigationAS ISO 14155-1:2004, AS ISO 14155-2:2004
    Self-imposed Load and Repeatability"Representative samples for the device underwent load and repeatability testing to verify functional and performance characteristics." (Implied successful Verification)
    Weight Capacity- T100, T200, T300, T400 series: 130kg
    - T500, T550, T600, T650 series: 200kg
    - T500B, T600B: 300kg
    Substantial EquivalenceFound substantially equivalent to the Convertible I-Series Positioning and Transfer Chair in intended use, basic construction, principles of operation, electrical and mechanical characteristics, and general safety and EMC compliance.

    Study Details

    The document describes pre-market testing and adherence to various recognized standards rather than a typical "study" in the sense of an AI/software device performance evaluation. The "device" in question is a medical treatment chair, primarily a mechanical and electrical device, not an AI or diagnostic tool.

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test Set Sample Size: "Representative samples for the device" were used for load and repeatability testing. A specific number is not provided.
    • Data Provenance: Not explicitly stated, but the company (Fresenius Medical Care South Asia Pacific Pty Ltd) has its primary contact in Australia, and several cited standards are Australian (AS/NZS, AS). Testing would likely have been conducted in a relevant facility that adheres to these international and local standards, potentially in Australia or a subcontracted testing lab. The testing for substantial equivalence is implied to be prospective verification against defined standards.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • This is not applicable as the device is a medical chair, not a diagnostic or AI device requiring expert ground truth for output evaluation. Performance is assessed against engineering and safety standards.

    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable. Performance is determined by meeting specified technical requirements of the cited standards, not by human consensus or adjudication of results.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is not an AI-assisted diagnostic device.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a physical medical device (a chair).

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" for the device's performance is against established engineering, electrical, mechanical, safety, flame retardancy, and biocompatibility standards (e.g., ISO, IEC, AS, EN, California standards). These standards define acceptable performance metrics and test methods. The weight capacities are also a form of "ground truth" against which the chairs were tested.

    7. The sample size for the training set

    • Not applicable. This is not an AI/machine learning device.

    8. How the ground truth for the training set was established

    • Not applicable. This is not an AI/machine learning device.