Found 50 results
510(k) Data Aggregation
(189 days)
Ask a specific question about this device
(114 days)
NaturaLyte® Dry Bicarbonate Concentrate (08-4112-2) is indicated for use in patients undergoing extracorporeal bicarbonate hemodialysis for acute and chronic renal failure. NaturaLyte® Dry Bicarbonate Concentrate (08-4112-2) is intended to be used as one component in the preparation of dialysate in a 3-stream proportioning hemodialysis machine according to a physician's prescription.
The NaturaLyte Dry Bicarbonate Concentrate (08-4112-2) is composed of dry sodium bicarbonate. NaturaLyte Dry Bicarbonate Concentrate (08-4112-2) is available in a 7807 g presentation that yields 25.36 gallons (96 liters) of bicarbonate concentrate volume. The 7807 g NaturaLyte Dry Bicarbonate Concentrate (08-4112-2) is supplied in bag-style packaging.
The NaturaLyte Dry Bicarbonate Concentrate (08-4112-2) is a non-sterile, single use Class II medical device composed of United States Pharmacopeia (USP) grade sodium bicarbonate.
NaturaLyte Dry Bicarbonate Concentrate (08-4112-2) is composed of sodium bicarbonate powder and is used as one component in the preparation of dialysate in a 3-stream proportioning hemodialysis machine according to a physician's prescription. NaturaLyte Dry Bicarbonate Concentrate (08-4112-2) is formulated for use in 45X proportioning systems which proportion a nominal ratio of 1 : 1.72 : 42.28 (acid : bicarbonate : water) to generate dialysate. The dialysate is intended to be pumped through a dialyzer, creating an osmotic gradient across the dialyzer membrane to exchange solutes with blood during hemodialysis.
N/A
Ask a specific question about this device
(36 days)
The FX CorAL hemodialyzers are intended for single use only for extracorporeal blood purification during intermittent renal replacement therapies hemodialysis (HD), hemodiafiltration (HDF) using pre-, post or mixed-dilution modes, and isolated ultrafiltration for patients suffering from renal insufficiency, including pediatric patients.
Consider body and dialyzer surface area, blood flow, body weight and extracorporeal blood volume when selecting dialyzers for use with pediatric patients.
The FX CorAL dialyzers are high-flux, single-use, steam-sterilized hemodialyzers. The dialyzers are provided blood pathway sterile and non-pyrogenic. The dialyzers allow for the transfer of water and solutes between blood and dialysate using semipermeable, hollow fiber membranes.
The FX CorAL dialyzers are high-flux, sterile devices designed for single-use acute and chronic hemodialysis. The dialyzers are configured to connect to a bloodline set which connects to a patient's vascular access system when used with a hemodialysis machine equipped with ultrafiltration control. During hemodialysis, blood is pumped from the patient's body through an extracorporeal circuit, one component of which is the dialyzer. The dialyzers contain semipermeable membranes that allow for diffusion and/or ultrafiltration to transport toxins and excess fluid from the blood compartment (fiber lumen) to the dialysate compartment. Dialyzers utilize a counter-current flow in which dialysate and blood flow in opposite directions in the dialyzer during hemodialysis. The counter-current flow maintains the concentration gradient across the membrane for waste and fluid removal.
N/A
Ask a specific question about this device
(266 days)
The BCM2 is intended for use, under the direction of a physician, for the noninvasive intermittent measurement of fluid status in patients with end-stage kidney disease (ESKD) including those receiving maintenance dialysis.
The device does not generate any real-time alarms and outputs from the device should be used in the context of all clinical data to assess a patient's fluid status.
The device is intended for use in patients 18 years and older.
The BCM2 is a multifrequency bioimpedance device that can be used by a clinician for the noninvasive intermittent measurement of fluid status in patients with end-stage kidney disease, including those receiving maintenance dialysis. A physician or designated healthcare professional is responsible for interpreting data from the device to determine what action is required as a result of changes in the measurements.
To assess a patient's hydration status, the BCM2 measures impedance by applying a low-level signal (50–800 µA) for less than 10 seconds at 50 different frequencies from 5 kHz to 1,000 kHz. Measurements are performed by dialysis clinicians (e.g., dialysis nurses, patient care technicians) in a healthcare environment such as a dialysis clinic or hospital.
In the normal clinical workflow, the BCM2 provides the following output parameters:
- Overhydration (OH)
- Urea distribution volume (V)
The following calculated and derived parameters are also available:
- Extracellular water (ECW)
- Intracellular water (ICW)
- Total body water (TBW)
The BCM2 system consists of the BCM2 touchscreen console and the electrode set. The BCM2 console powers and measures the bioimpedance spectroscopy frequencies to assess fluid parameters. The device can also be operated in battery mode. Battery mode provides flexibility when moving between patients. The battery charge status is shown in the upper corner of the display regardless of which power source is being used. The power supply connection is located on the rear of the console.
The Calibration Box employs different resistors to calibrate the entire range of the BCM2's measurement (5 kHz to 1,000 kHz). Impedance measurements are performed for each resistor for all frequencies and the data is verified against an expected tolerance range. After the BCM2 has been calibrated, the Test Box is used to verify that the device is functioning properly.
The BCM2 will be available in one (1) configuration for sale within the U.S.
N/A
Ask a specific question about this device
(269 days)
The AquaC UNO H Portable Water Purification System is a reverse osmosis unit intended for use with hemodialysis systems to remove organic and inorganic substances and microbial contaminants from the water used for treating hemodialysis patients or related therapies.
The AquaC UNO H can be connected to hemodialysis equipment used in hospitals, clinics and in home environments. This device is intended to be a component in a complete water purification system, and is not a complete water treatment system. The reverse osmosis unit must be preceded by pre-treatment devices, and may need to be followed by post-treatment devices as well, to meet current AAMI/ANSI and FDA recognized U.S. standards.
The AquaC UNO H water purification system is a microcontroller-controlled, fully automatic reverse osmosis (RO) system with heat disinfection function, which uses pretreated soft water (hereinafter referred to as "feed water") to produce highly deionized water (hereinafter referred to as "dialysis water") for use by a hemodialysis (HD) device and preparation of dialysis concentrates.
Feed water is defined as the water supplied to a water treatment system or an individual component of a water treatment system per ISO 23500-1:2019 Preparation and quality management of fluids for haemodialysis and related therapies - Part 1: General requirements. The feed water must be of drinking water standard, filtered, free of iron and chlorine, and softened. Potentially critical limits must be monitored by regular checks. Bacterial growth in the system must be prevented by continuous operation of the system with a minimum of idle times and by preventive measures such as chemical or heat disinfection.
The dialysis water flows through the heater, across the dialysis water conductivity/temperature sensor and the dialysis water stop valve or outlet into the ring main. The ring main can be connected to a HD device or to a device used for the preparation of dialysis concentrates. Excess dialysis water flows through the check valve and back into the break tank. The concentrate flows through the concentrate drain restrictor and is either returned into the break tank or drained via the drain valve.
The system features the following operating modes:
- Standby
- Supply
- Rinse
- Module Heat Disinfection
- Ring Disinfection
- Chemical Disinfection
- Decalcification
- Isolated Standby
N/A
Ask a specific question about this device
(115 days)
The DIASAFE®plusUS is intended for the preparation of ultrapure dialysis fluid (for the 2008-series machines and 5008X machines) and sterile, non-pyrogenic substitution fluid (for the 5008X machines) from pretreated water and is not intended to provide primary purification. Attention must still be given to the chemical and microbiological quality of water and concentrates and the maintenance of supply systems (e.g. RO system, central delivery system). The microbiological quality (microbial count [CFU/mL] and endotoxin measurement [EU/mL]) of the incoming water should be < 200 CFU/mL and < 2 EU/mL, respectively.
The DIASAFE®plusUS can only be used with Fresenius Medical Care hemodialysis machines fitted for use with the DIASAFE®plusUS.
The DIASAFE®plusUS is a non-sterile dialysis fluid filter that produces ultrapure dialysis fluid (for the 2008-series machines and 5008X machines) and sterile, non-pyrogenic substitution fluid (for the 5008X machines) for use in chronic and acute hemodialysis and hemodiafiltration. The DIASAFE®plusUS reduces microbial contaminants including endotoxin in the dialysate during treatments. The DIASAFE®plusUS is available in one (1) configuration and is compatible for use with 2008 series and 5008X hemodialysis machines. It is installed on the hemodialysis machines using the DIAFIX™ Lock System. The DIAFIX™ Lock System is a standard feature on the hemodialysis machines and is installed during machine production.
The 2008 series hemodialysis machines use 1 DIASAFE®plusUS. Mixed dialysis fluid is forced through an open filter port across the fibers of the DIASAFE®plusUS. A closed bypass valve at the other end forces the dialysis fluid across the fiber membrane. Dialysis fluid passes through the porous membrane and into the filtrate compartment where it is allowed to flow through the uncapped side ports and into the hemodialyzer. Microorganisms and molecules too large to pass through the membrane are trapped in the fibers until they are flushed out, and/or disinfected during a routine cleaning cycle.
The 5008X hemodialysis system uses two (2) DIASAFE®plusUS for preparation of the dialysis and substitution fluid which can be used for up to 100 on-line HDF treatments. The dialysis fluid is filtered by the first DIASAFE®plusUS. A second DIASAFE®plusUS is used to filter the ultrafiltered dialysis fluid a second time to prepare the substitution fluid. The substitution fluid can be infused in the post-dilution mode (i.e., after the blood has passed through the dialyzer), in the pre-dilution mode (i.e., before the blood enters the dialyzer), or combined.
N/A
Ask a specific question about this device
(60 days)
Bicarby™ Dialysate solutions are indicated for use as a dialysis fluid for adult patients in acute care settings, requiring Kidney Replacement Therapies (KRT).
The Bicarby Dialysate solutions are single-use, sterile, ready-to-use dialysis solutions consisting of sodium chloride, potassium chloride, magnesium chloride, calcium chloride (select solutions), sodium bicarbonate, and glucose. One (1) new calcium-containing solution and seven (7) new calcium-free solutions are being introduced. Of the 7 calcium-free solutions, there are only four (4) unique formulations (Table 1).
Bicarby Dialysate solutions are sterile dialysates for use in KRT. The solutions are each provided in a 2-compartment bag. One compartment contains 4.75 L of a slightly alkaline bicarbonate solution and the other compartment contains 0.25 L of an acidic electrolyte, glucose solution. The 2 solutions are mixed before use by opening the peel seam between the compartments, yielding 5 L of a ready-to-use sterile solution. The dialysate bags are made of a gas barrier foil, a material manufactured without PVC, latex, or plasticizer. The calcium-containing solutions are designed for modalities using heparin anticoagulation, while the calcium-free solutions are intended for modalities using regional citrate anticoagulation (RCA). The calcium-free solutions sustain the anticoagulatory effect of citrate in the filter. Therefore, a separate infusion of calcium is mandatory when using these solutions.
N/A
Ask a specific question about this device
(246 days)
The 2008T BlueStar Hemodialysis Machine is indicated for acute and chronic dialysis therapy in a healthcare facility.
Additional therapy options for patients receiving hemodialysis include: Isolated Ultrafiltration, Sustained Low Efficiency Dialysis (SLED), and low volume hemodialysis (patients weighing ≥ 20kg and ≤ 40 kg). This machine accommodates the use of both low flux and high flux dialyzers. The SLED therapy option is not to be used for patients weighing ≤ 40 kg. The 2008T BlueStar Hemodialysis Machine is not to be used for plasma replacement therapies, for patients weighing less than 20 kg, or for renal therapies using substitution fluid.
The 2008T BlueStar Hemodialysis Machine is an electromechanical device. Software controls the machine during hemodialysis treatment, including fluid flow, mixing, heating, and alarms.
The 2008T BlueStar Hemodialysis Machine provides hemodialysis treatment by controlling and monitoring both the dialysate circuit and the extracorporeal blood circuit. The machine pumps blood from the patient's body through an extracorporeal circuit, one component of which is the dialyzer. The dialyzer contains a semi-permeable membrane that uses diffusion to transfer toxins and ultrafiltration to transport excess water from the blood into the dialysate circuit. In this separate dialysate circuit, the dialysate concentrates are mixed with purified water, heated, degassed, and delivered to the dialyzer. Balancing chambers control the incoming flow and outgoing flow of the dialysate fluid during ultrafiltration. During hemodialysis, the extracorporeal blood circuit is monitored for venous and arterial blood pressures as well as for the presence of air and blood.
The 2008T BlueStar Hemodialysis Machine has automation features (independent internal conductivity testing, auto priming, auto start, CDX auto on, assisted reinfusion) to minimize user workload and improve user experience. The 2008T BlueStar Hemodialysis Machine includes SLED (Sustained Low Efficiency Dialysis), Low Volume, and Isolated Ultrafiltration therapy options. The 2008T BlueStar Hemodialysis Machine also accommodates the use of the Patient Card system, a dialysis treatment information system.
Based on the provided FDA 510(k) Clearance Letter, the device in question is the 2008T BlueStar Hemodialysis Machine. This document is a Class II Medical Device submission, which means it relies on substantial equivalence to a predicate device, rather than a clinical trial proving new efficacy. Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context are focused on demonstrating substantial equivalence to a legally marketed predicate device, rather than establishing new clinical effectiveness benchmarks through a de novo study.
The primary study type proving the device meets the "acceptance criteria" for 510(k) clearance is a comparison to a predicate device and supporting non-clinical performance and safety testing.
Here's a breakdown of the requested information based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The "acceptance criteria" for a 510(k) submission are typically related to demonstrating that the new device meets the same performance specifications as the predicate and does not raise new questions of safety or effectiveness. The reported device performance is presented as its features and specifications.
| Feature / Acceptance Criteria (Derived from Predicate) | Reported Device Performance (2008T BlueStar Hemodialysis Machine) |
|---|---|
| Intended Use Equivalence | Identical to predicate (K231125): "The 2008T Machine is intended for use in acute and chronic hemodialysis therapy." |
| Indications for Use Equivalence | Identical to predicate. Includes acute and chronic dialysis therapy in a healthcare facility, with options for Isolated Ultrafiltration, SLED, and low volume hemodialysis (patients ≥20kg and ≤40kg). Accommodates low/high flux dialyzers. Restrictions: not for SLED on patients ≤40kg, plasma replacement, patients <20kg, or renal therapies using substitution fluid. |
| Design Specifications Equivalence | "Design Specifications" are listed as equivalent to the predicate. |
| Technological Characteristics Equivalence | "Technological Characteristics" are listed as equivalent to the predicate. |
| Principle of Operation Equivalence | "Principle of Operation" is listed as equivalent to the predicate. |
| Performance Requirements Equivalence | "Performance Requirements" are listed as equivalent to the predicate. |
| Biocompatibility | Tested per ISO 10993-1:2018 and FDA guidance. Evaluated for Cytotoxicity, Sensitization, Irritation, Pyrogenicity, Hemocompatibility, Chemical Characterization, Toxicological Risk Assessment. (Specific pass/fail criteria are implicit in compliance with standards). |
| Electrical Safety and EMC | Tested per IEC 60601-1-2 (2020). (Specific pass/fail criteria are implicit in compliance with standards). |
| Software V&V (if changes) | "No changes to the 2008T BlueStar Hemodialysis Machine's software." Therefore, software V&V was not required. (Implicit acceptance criteria: software stability/no new risks). |
| Key Performance Specifications (Specifics compared) | Blood Flow Rates: - 8 mm: 20–600 mL/min* (*Not available with Low Volume)- 6.35 mm: 20–465 mL/min- 4.8 mm: 10–274 mL/min- 2.6 mm: 6–86 mL/min- Accuracy: ±10% at -200 mmHg |
| Maximum Dialysate Flow Rate: - Selectable: (0)/100†‡/150†‡/200†‡/300†/400/500/600/700/800 mL/min- Auto Flow (based on Qb): Ranges from 300 mL/min (for Qb 0-165 mL/min) up to 800 mL/min (for Qb 481+ mL/min)- Dialysate flow adjusts if blood pump adjusted ~15-20 mL/min. | |
| Net Fluid Removal: 0–4000 mL/hr- Accuracy: ± (1% UF rate + 18 mL/hr) at 100 mL/min Qd- Accuracy: ± (1% UF rate + 30 mL/hr) at 500 mL/min Qd- Accuracy: ± (1% UF rate + 48 mL/hr) at 800 mL/min Qd | |
| Dialysis Time: - Hemodialysis: 0–9:59 hours (adjustable)- SLED: Fixed at 12 hours- Accuracy: + 1 second per hour | |
| Dialysis Fluid Composition: - Volumetric, selectable- Acid adjustment: 130–155 mEq/L Na+- Bicarbonate adjustment: 20–40 mEq/L Bicarbonate- Monitoring conductivity average accuracy: ±1.5% | |
| Dialysis Fluid Temperature: - Range: 35°C–39°C (alarm window ±2°C)- Accuracy: ±0.3°C | |
| Heparin Delivery Rate: - 0–9.9 mL/hr- Accuracy: ±5% | |
| Human Factors Impact | "Changes do not impact the usability of the device. Human Factors Validation Testing is not required." |
2. Sample Size Used for the Test Set and Data Provenance
For a 510(k) of this nature (a hemodialysis machine), the "test set" does not refer to a clinical data set in the same way it would for an AI/ML algorithm. Instead, it refers to the units of the device that underwent various engineering and biocompatibility tests.
- Sample size for test set: Not explicitly stated as a number of specific units in a formal statistical sense, but implies a sufficient number of devices were tested to demonstrate compliance with standards and specifications for various non-clinical tests (e.g., electrical safety, EMC, performance specifications). For biocompatibility, representative materials were tested.
- Data provenance: Not directly applicable in the sense of patient data. The provenance of the testing is internal company testing (Fresenius Medical Care Renal Therapies Group, LLC). All testing is prospective in the sense that the tests were conducted specifically for this submission. The origin of the testing would be within the manufacturer's facilities or accredited testing labs.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This question is not applicable to this type of 510(k) submission. This clearance is for a hardware device (hemodialysis machine) with embedded software, not an AI/ML diagnostic or prognostic device that relies on expert interpretation of medical images or patient data to establish ground truth for a test set. The "ground truth" for the device's performance is established by engineering specifications, validated sensors, and calibrated measurement systems.
4. Adjudication Method for the Test Set
Not applicable. There is no "test set" of patient cases requiring adjudication by multiple experts for a hemodialysis machine clearance. Engineering tests are pass/fail based on predefined thresholds and standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
Not applicable. This is not an AI/ML diagnostic or imaging device. There were no human readers or AI assistance studies conducted or required for this clearance.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
While the device contains software, it is an electromechanical device, not a standalone algorithm. Its performance is intrinsically linked to the physical machine. Therefore, a "standalone algorithm performance" study is not relevant or applicable. The software controls the machine's functions, and its performance is evaluated as part of the total system.
7. The Type of Ground Truth Used
The "ground truth" for the 2008T BlueStar Hemodialysis Machine's performance is established by:
- Engineering Specifications: The design and expected performance metrics (e.g., flow rates, temperatures, accuracies) are the primary ground truth for the device's functional integrity.
- International and Industry Standards: Compliance with standards like IEC 60601-1-2 for EMC and ISO 10993-1 for biocompatibility serves as the ground truth for safety and foundational performance aspects.
- Predicate Device Performance: The underlying assumption for a 510(k) is that the predicate device's performance is the established "ground truth" for safe and effective hemodialysis delivery. The new device demonstrates equivalence to this established standard.
8. The Sample Size for the Training Set
Not applicable. This is not an AI/ML device that requires a training set of data.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for an AI/ML model.
Ask a specific question about this device
(199 days)
The 5008X Hemodialysis System is intended for use in acute and chronic hemodialysis therapy. Therapy options include hemodialysis (HD), hemodiafiltration (HDF), hemofiltration (HF), and isolated ultrafiltration (ISO).
The 5008X™ system is indicated for intermittent hemodialysis treatment for patients with acute and chronic renal failure in a healthcare facility. Therapy options for patients weighing more than 40 kg include: Hemodiafiltration (HDF), Hemodialysis (HD), Hemofiltration (HF) and Isolated Ultrafiltration.
The 5008X™ Machine is equipped with a bibag system, but it may also be used with liquid bicarbonate using the connector included with the machine. The bibag system is indicated for use in patients undergoing extracorporeal bicarbonate hemodialysis for acute and chronic renal failure. The bibag system is intended to be used as one component in the preparation of dialysate according to a physician's prescription in a 3-stream proportioning hemodialysis machine equipped with the bibag module.
The 5008X™ Machine is available equipped with a Crit-Line Clip (CLiC) Monitor System.
The optional CLiC system is used to non-invasively measure hematocrit, oxygen saturation and percent change in blood volume in real time for application in the treatment of dialysis patients with the intended purpose of providing a more effective treatment for both the dialysis patient and the dialysis technician. Based on the data that the monitor provides, the clinician/nurse, under physician direction, intervenes (i.e. increases or decreases the rate at which fluid is removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common symptoms of dialysis which include nausea, cramping and vomiting.
The DIASAFE®plusUS is intended for the preparation of ultrapure dialysate and sterile, nonpyrogenic substitution fluid from pretreated water and is not intended to provide primary purification. Attention must still be given to the chemical and microbiological quality of water and concentrates and the maintenance of supply systems (e.g. RO system, central delivery system).
The microbiological quality (microbial count [CFU/mL] and endotoxin measurement [EU/mL]) of the incoming water should be < 200 CFU/mL and < 2 EU/mL, respectively.
The DIASAFE®plusUS can only be used with Fresenius Medical Care dialysis machines fitted for use with DIASAFE®plusUS.
The blood tubing sets are intended to be used only with the Fresenius Medical Care 5008X Hemodialysis Device, which is indicated for intermittent hemodialysis treatment for patients with acute kidney injury or chronic kidney disease in a healthcare facility. Therapy options include: hemodiafiltration (HDF), hemodialysis (HD), hemofiltration (HF), and isolated ultrafiltration.
The HDF blood tubing sets are indicated for use with a prescribed dialyzer. For use with a compatible dialyzer, as per the labeling.
The Fresenius Twister Access Flow Reversing Connector (AFRC) is indicated for use during hemodialysis to reverse the blood flow to and from the arterial and venous vascular access devices in order to obtain an access flow measurement.
The blood tubing set is intended to be used with the Crit-Line Clip Monitor System (CLiC) on the 5008X Hemodialysis Device.
The CLiC is used to non-invasively measure hematocrit, oxygen saturation, and percent change in blood volume in real time for application in the treatment of dialysis patients with the intended purpose of providing a more effective treatment for both the dialysis patient and the dialysis technician. Based on the data that the monitor provides, the clinician/nurse, under physician direction, intervenes (i.e., increases or decreases the rate at which fluid is removed from the blood) in order to remove the maximum amount of fluid from the dialysis patient without the patient experiencing the common symptoms of dialysis which include nausea, cramping and vomiting.
CitraSure is intended for thermal disinfection of the Fresenius Medical Care 5008X Hemodialysis System when delivered and diluted per the 5008X heat disinfection program.
The 5008X Hemodialysis System consists of the following:
- The 5008X Hemodialysis Machine
- Two (2) DIASAFE®plusUS Filters (hereinafter referred to as the "Diasafe filter(s)")
- The 5008X Bloodlines are available in five (5) configurations:
- 5008X HD/HDF – Standard Blood Tubing Set (hereinafter referred to as the "HD/HDF Standard Bloodline")
- 5008X HD/HDF – With Twister Blood Tubing Set (hereinafter referred to as the "HD/HDF Twister Bloodline")
- 5008X HD/HDF – With CLiC Blood Tubing Set (hereinafter referred to as the "HD/HDF CLiC Bloodline")
- 5008X HD Pre-Flush – Standard Blood Tubing Set (hereinafter referred to as the "HD Standard Bloodline")
- 5008X HD Pre-Flush – With Twister Blood Tubing Set (hereinafter HD Twister Bloodline")
- CitraSure™ Disinfectant
The 5008X Hemodialysis Machine is an electromechanical device. Software controls the machine during hemodialysis treatment, including fluid flow, mixing, heating, and alarms.
The Diasafe filter is a non-sterile dialysis fluid filter that produces ultrapure dialysate and sterile, non-pyrogenic substitution fluid as defined in ISO 23500-5. The filter reduces microbial contaminants including endotoxins in the dialysate during hemodialysis treatment.
The 5008X Bloodlines are single-use, ethylene oxide (EO) sterilized bloodlines.
CitraSure™ disinfectant is a liquid chemical disinfectant for chemical heat disinfection of the 5008X Hemodialysis Machine.
The provided FDA 510(k) clearance letter and summary for the 5008X Hemodialysis System pertains to a medical device for renal treatment. The document does not describe an AI/ML device, a comparative effectiveness study (MRMC), or a standalone AI algorithm performance study. Therefore, sections of your request related to those topics (e.g., sample size for test/training sets, data provenance, number of experts, adjudication methods, ground truth types, human reader improvement with AI) are not applicable and cannot be extracted from this document.
However, I can provide the acceptance criteria and a summary of the performance testing conducted to prove the device meets these criteria based on the provided text for the various components of the 5008X Hemodialysis System.
Acceptance Criteria and Device Performance Summary for 5008X Hemodialysis System
1. Table of Acceptance Criteria and Reported Device Performance
Note: The document lists "Key Performance Specifications/Characteristics" which serve as the acceptance criteria for each component. The "Reported Device Performance" column summarizes the claims made in the document regarding the device meeting these specifications through various tests.
| Feature/Test Objective | Acceptance Criteria (Specification/Characteristic) | Reported Device Performance (Summary of Test Results) |
|---|---|---|
| 5008X Machine | ||
| Maximum Blood Flow Rate | 600 mL/min | System Level Performance Testing and Functional Design Verification conducted. Results support substantial equivalence, safety, and efficacy. |
| Maximum Dialysate Flow Rate | 1000 mL/min | System Level Performance Testing and Functional Design Verification conducted. Results support substantial equivalence, safety, and efficacy. |
| Maximum Substitution Flow Rate | AutoSub plus: 400 mL/min; Manual substitution: 600 mL/min | System Level Performance Testing and Functional Design Verification conducted. Results support substantial equivalence, safety, and efficacy. |
| Net Fluid Removal | 0–4000 mL/hr; Accuracy: ± (1%UF + 0.15% of balanced fluid volume) | System Level Performance Testing and Functional Design Verification conducted. Results support substantial equivalence, safety, and efficacy. |
| Dialysis Time | Intermittent, typical time 4 hours | System Level Performance Testing and Functional Design Verification conducted. Results support substantial equivalence, safety, and efficacy. |
| Dialysis Fluid Composition | Volumetric, selectable: Acid 125–151 mEq/L Na+; Bicarbonate 25–40 mEq/L; Monitoring conductivity average accuracy: ± 1.5% | System Level Performance Testing and Functional Design Verification conducted. Results support substantial equivalence, safety, and efficacy. |
| Dialysis Fluid Temperature | Range 34°C–39°C. Fixed alarm window 33°C (or 32°C during BTM recirculation measurement and preparation) or above 40°C. | System Level Performance Testing and Functional Design Verification conducted. Results support substantial equivalence, safety, and efficacy. |
| Heparin Administration Rate | 0.1 to 9.9 mL/hr; Accuracy: ± 0.5mL or ± 6% | System Level Performance Testing and Functional Design Verification conducted. Results support substantial equivalence, safety, and efficacy. |
| Electrical Safety and EMC | IEC 60601-1-2 Edition 4.1 2020-09 compliance | EMC testing conducted in accordance with IEC 60601-1-2. |
| Software Verification & Validation | Functional and Performance Verification, Regression Testing, Code Reviews | Unit, integration, and system level software verification testing performed to demonstrate efficacy and confirm operation. |
| Human Factors Validation | Safe and effective use (per FDA guidance) | Human Factors Validation Testing conducted. |
| Diasafe Filter | ||
| Bacterial and Endotoxin Filtration | Produces sterile, non-pyrogenic substitution fluid from dialysis fluid with max incoming water quality < 200 CFU/mL (bacteria) and < 2 EU/mL (endotoxin) | Performance testing conducted. Results support substantial equivalence, safety, and efficacy. |
| Maximum Number Disinfection Cycles | 100 CitraSure cycles; 13 Pure Bright Bleach cycles | Performance testing conducted. Results support substantial equivalence, safety, and efficacy. |
| Use Life | 90 days maximum or if max disinfection cycles reached | Performance testing conducted. Results support substantial equivalence, safety, and efficacy. |
| 5008X Bloodlines | ||
| Maximum Blood Flow Rate | 600 mL/min | Blood Pump Endurance test performed. Results support substantial equivalence, safety, and efficacy. |
| Minimum Labeled Arterial Pressure | −300 mmHg | Structural Integrity and Positive/Negative Leak Test performed. Results show the bloodlines can withstand 1.5X labeled max positive/negative pressures and pressures from faulty conditions. |
| Maximum Labeled Venous Pressure | +500 mmHg | Structural Integrity and Positive/Negative Leak Test performed. Results show the bloodlines can withstand 1.5X labeled max positive/negative pressures and pressures from faulty conditions. |
| Arterial Pump Segment | 8.0 mm/12.0 mm (ID/OD) | Pump Segment Performance test evaluated characteristics over range of inlet pressures and flow rates. |
| Twister Component | Facilitates measurement of access flow by reversing blood flow while maintaining closed circuit | Twister Rotation Test (180° ± 1°) and Twister Torque Test (≤ 7 inch-pound) performed. |
| CLiC Chamber (Optional) – Hematocrit (HCT) | Standard deviation X2 ≤ 3%, average bias ≤ 1% (of average HCT) on control blood chambers; measurement range 10–60 HCT | Functional CLiC Chamber Test (Hematocrit) performed. |
| CLiC Chamber (Optional) – Oxygen Saturation (O2 Sat) | Standard deviation X2 ≤ 3%, average bias ≤ 2% (of average O2 Sat) on control blood chambers; measurement range 30%–100% (when HCT ≥ 10) | Functional CLiC Chamber Test (O2 Sat) performed. |
| Structural Integrity (Bloodlines & Accessories) | Withstand 1.5X max positive/negative pressures; Withstand pressures from faulty machine conditions | Testing performed. |
| Pump Segment Performance | Evaluate characteristics over range of inlet pressures (0 to -250 mmHg), flow rates, and treatment time | Testing performed. |
| Blood Pump Endurance | No break or detachment at max blood flow rates & pressures for 12 hours | Testing performed. |
| Simulated Use | No tubing failures (kinking, collapsing, disconnection) under simulated use for ≥ 4 hours; components removable without external leaks | Testing performed. |
| Needle Access Port | Withstand 1.5X recommended max positive/negative pressures after multiple accesses | Testing performed. |
| Luer Connectors (Performance) | Meet ISO 80369-7 2021, Section 6 | Testing performed. |
| Luer Connectors (Dimensions) | Meet ISO 80369-7: 2021, Section 5 | Testing performed. |
| Luer Connectors (Raw Material) | Nominal modulus of elasticity > 700 MPa | Testing performed. |
| Transparency of Transducer Protector | Clear to allow visual inspection of blood contamination | Testing performed. |
| Transducer Protector Leak Test | Maintain secure and leak-free connection when subjected to 2X max labeled pressure | Testing performed. |
| Arterial Pressure Dome Leak Test | Maintain secure and leak-free connection when subjected to 2X max labeled pressure | Testing performed. |
| CLiC Blood Chamber Lens Gap | 0.078 ± 0.005 in. | Testing performed to verify optical property. |
| Tubing Compliance Test | Capable of being occlusively clamped by venous line clamp of 5008X machine | Testing performed. |
| Clamp Occlusion Test | Air and liquid tightness occlusion at desired pressure and time | Testing performed. |
| Check Valve Cracking Pressure | Lower than 400 mbar | Testing performed. |
| Check Valve Pressure Drop | Specific range when substitution fluid pumped at certain flow rate | Testing performed. |
| Tube Transparency | Observe interface of air and liquid during passage of air bubbles | Testing performed. |
| Tube Resistance to Kinking | No kinking at worst-case bend radius when packaged (evaluate % flow reduction) | Testing performed. |
| Tube Clamping | Resist clamp occlusions by subjecting to desired pressure and cycles | Testing performed. |
| Tube Patency | Maintain patency after clamping and reopening (measure flow reduction) | Testing performed. |
| Air Bubble Trapping (Venous Chamber) | Able to trap incoming air bubbles larger than 20 µL | Testing performed. |
| Recirculation Connector Packaging | Burst strength, peel strength, dye penetration, microbial resistance | Testing performed. |
| Viral Retentiveness (Transducer Protector) | Prevent passage of bacteriophage (PhiX174) from patient to machine side up to 600 mmHg for 1 hr | Testing performed. |
| Sterility (Bloodlines) | Sterility assurance level (SAL) 10-6 | Sterilized by 100% ethylene oxide (EO). |
| EO Residuals (Bloodlines) | < 4.6 mg/device for EO and ECh | Tested per AAMI/ANSI/ISO 10993-7:2008/(R)2012. |
| Bacterial Endotoxin (Pyrogenicity) (Bloodlines) | Non-pyrogenic (< 20 EU/device) | Tested per ANSI/AAMI/ST72:2019. |
| Sterile Barrier Testing (Bloodlines) | Maintain sterility of fluid path | Structural integrity test adapted from ISO 8637-2:2018 performed on samples after aging and distribution. |
| CitraSure Disinfectant | ||
| Minimum Disinfection Temperature | 70ºC | Performance testing conducted. Results support substantial equivalence, safety, and efficacy. |
| Minimum Contact Time | 10 minutes | Performance testing conducted. Results support substantial equivalence, safety, and efficacy. |
| Use Life | 90 Days | Performance testing conducted. Results support substantial equivalence, safety, and efficacy. |
| Shelf Life | 2 years | Performance testing conducted. Results support substantial equivalence, safety, and efficacy. |
2. Sample Size for the Test Set and the Data Provenance
As this is not an AI/ML device, the concept of "test set" in the context of diagnostic AI algorithms (e.g., patient images or clinical data) is not directly applicable. The performance data presented refers to engineering and quality assurance tests on the physical devices/systems.
- Sample Size for Physical Device Testing: The document does not specify exact sample sizes for each individual engineering performance test (e.g., how many bloodlines were tested for structural integrity, how many machines for fluid flow accuracy). It generally states "Testing performed" or "Testing conducted."
- Data Provenance: Not applicable in the context of patient data provenance (e.g., country of origin, retrospective/prospective). These are laboratory and simulated use tests on hardware components.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of those Experts
Not applicable. The "ground truth" for the performance of a hemodialysis system is based on direct physical measurements against defined engineering specifications and regulatory standards (e.g., flow rates, pressure tolerances, material properties, sterility levels), not expert interpretation of complex data like medical images.
4. Adjudication Method for the Test Set
Not applicable for a medical device's engineering performance testing. Adjudication methods like 2+1 or 3+1 typically refer to resolving discrepancies in expert interpretations (e.g., radiologists reviewing medical images for ground truth labeling in AI studies).
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No. The document explicitly states: "No clinical studies were performed for the 5008X Hemodialysis System." MRMC studies inherently involve human readers (clinicians) assessing cases, often with or without AI assistance, to measure diagnostic performance and comparative effectiveness. This type of study was not conducted as this is not an AI-assisted diagnostic device.
6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done
No. The device described is a physical hemodialysis system, not a standalone AI algorithm. While it contains software, the performance evaluation is for the integrated system, and there is no mention of a standalone AI algorithm being assessed.
7. The Type of Ground Truth Used
For the various performance tests, the "ground truth" is defined by:
- Engineering Specifications and Standards: Numeric values for flow rates, pressures, temperatures, accuracies, dimensions, and material properties derived from established engineering principles and relevant international standards (e.g., ISO 8637-2:2018 for bloodlines, ISO 23500-5 for dialysate quality, IEC 60601-1-2 for EMC).
- Regulatory Requirements: Compliance with FDA guidance documents and regulations (e.g., for biocompatibility, sterility, software verification and validation, human factors).
- Absence of Failure: Demonstrating that components do not break, kink, leak, or fail under specified conditions (e.g., blood pump endurance, simulated use).
8. The Sample Size for the Training Set
Not applicable. This is not an AI/ML device that requires a training set. The software for the 5008X Hemodialysis Machine would have undergone development, verification, and validation, but this is distinct from AI model training.
9. How the Ground Truth for the Training Set Was Established
Not applicable. As there is no AI training set, there is no ground truth established in this context.
Ask a specific question about this device
(116 days)
Bicarby™ Dialysate solutions are indicated for use as a dialysis fluid for adult patients in acute care settings, requiring Kidney Replacement Therapies (KRT).
Ci-Ca Dialysate solutions are indicated for use as a dialysis fluid for adult patients in acute care settings, requiring Kidney Replacement Therapies (KRT).
The Bicarby and Ci-Ca solutions are single-use, sterile, ready-to-use dialysis solutions consisting of sodium chloride, potassium chloride, magnesium chloride, calcium chloride (select solutions), sodium bicarbonate, and glucose. The calcium-containing solutions are available in five (5) formulations, while the calcium-free solutions are available in three (3) formulations. The Bicarby and Ci-Ca solutions are single-use, steam sterilized dialysis solutions for use in Kidney Replacement Therapy (KRT). The solutions are each provided in a 2-compartment bag. One compartment contains 4.75 L of a slightly alkaline bicarbonate solution and the other compartment contains 0.25 L of an acidic electrolyte, glucose solution. The 2 solutions are mixed before use by opening the peel seam between the compartments, yielding 5 L of a ready-to-use sterile solution. The dialysate bags are made of a gas barrier foil, a material manufactured without PVC, latex, or plasticizer. The calcium-containing solutions are designed for modalities using heparin anticoagulation, while the calcium-free solutions are intended for modalities using regional citrate anticoagulation (RCA). The calcium-free solutions sustain the anticoagulatory effect of citrate in the filter. Therefore, a separate infusion of calcium is mandatory when using these solutions.
The provided FDA 510(k) clearance letter (K243786) is for Bicarby and Ci-Ca Dialysate solutions, which are dialysis fluids. This type of device is classified as a Class II medical device (Hemodialysis system and accessories, product code KPO).
The provided documentation describes the acceptance criteria and supporting studies for these dialysate solutions, not for a software algorithm or AI-powered medical device that would have performance metrics like sensitivity, specificity, or AUC. The studies conducted are primarily focused on the physical, chemical, and biological safety and performance of the solution and its container system, rather than an AI device's diagnostic or predictive capabilities.
Therefore, many of the requested categories (e.g., sample size for test set, data provenance, number of experts, adjudication method, MRMC study, standalone performance, ground truth types, training set sample size, ground truth for training set) are not applicable to this type of device and the information provided in the 510(k) summary.
Below is a summary of the available information relevant to acceptance criteria and supporting studies for the Bicarby and Ci-Ca Dialysate solutions:
Acceptance Criteria and Device Performance (Not Applicable in the traditional sense for an AI device)
For this type of device (dialysis solutions), "acceptance criteria" relate to meeting specifications for chemical composition, sterility, packaging integrity, and biocompatibility rather than diagnostic performance metrics. The document details the key performance specifications for the chemical components.
Table of Acceptance Criteria (Key Performance Specifications) and Reported Device Performance
| Chemical Component | Acceptance Criteria (Ionic Contribution, mEq/L, mixed) - as designed/specified | Reported Device Performance (Implied to meet specification) |
|---|---|---|
| Calcium-containing solutions | ||
| Sodium (Na+) | 140 | Implied to meet 140 |
| Potassium (K+) | 0, 2, 3, or 4 | Implied to meet 0, 2, 3, or 4 |
| Magnesium (Mg2+) | 1.0 or 1.5 | Implied to meet 1.0 or 1.5 |
| Calcium (Ca2+) | 2.5 or 3.0 | Implied to meet 2.5 or 3.0 |
| Chloride (Cl-) | 109, 111, 112, or 113 | Implied to meet 109, 111, 112, or 113 |
| Bicarbonate (HCO3-) | 35 | Implied to meet 35 |
| Glucose (g/L) | 1 | Implied to meet 1 |
| Calcium-free solutions | ||
| Sodium (Na+) | 133 or 140 | Implied to meet 133 or 140 |
| Potassium (K+) | 2 or 4 | Implied to meet 2 or 4 |
| Magnesium (Mg2+) | 1.5 | Implied to meet 1.5 |
| Calcium (Ca2+) | 0 | Implied to meet 0 |
| Chloride (Cl-) | 116.5, 118.5, or 120.5 | Implied to meet 116.5, 118.5, or 120.5 |
| Bicarbonate (HCO3-) | 20 or 25 | Implied to meet 20 or 25 |
| Glucose (g/L) | 1 | Implied to meet 1 |
Note: The document states these are the "Key Performance Specifications," meaning the solutions are manufactured to these specific ionic contributions. The "Reported Device Performance" is implied to meet these specifications as part of the manufacturing and quality control process, which are regulated under GMP but specific quantified "performance" results are not provided in this 510(k) summary.
Study Information (as applicable to a dialysis solution):
A. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
* Not Applicable in the context of an AI device's test set. The studies conducted are primarily engineering, chemical, and biological evaluations, not data-driven performance assessments.
* For Biocompatibility, "testing was conducted to evaluate the material changes in accordance with ISO 10993-1:2018 and FDA guidance document...". This implies lab-based testing.
* For Human Factors Validation Testing, it was stated that "The Bicarby and Ci-Ca solutions were found to be safe and effective for their intended users, uses, and use environments." The sample size for this human factors study is not provided in the summary.
B. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
* Not Applicable. No "ground truth" establishment by medical experts for a test set of images/data is relevant here. The ground truth for chemical composition is defined by the formula and analytical testing. Biocompatibility experts would assess the results of the specific biological tests.
C. Adjudication method (e.g. 2+1, 3+1, none) for the test set
* Not Applicable. This concept is for clinical or diagnostic studies involving human readers, not for chemical solutions or their packaging.
D. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
* No. This is a dialysis solution, not an AI diagnostic system.
E. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
* Not Applicable. This is a medical device (dialysate solution), not an algorithm.
F. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
* For chemical composition: Defined chemical formula and analytical testing standards.
* For physical properties (e.g., shipping, connector integrity): Engineering specifications and test standards.
* For biocompatibility: Reference to international standards (ISO 10993-1:2018) and established biological endpoints (Cytotoxicity, Sensitization, Irritation, Material Mediated Pyrogenicity, Hemocompatibility, Genotoxicity) with a toxicological risk assessment.
G. The sample size for the training set
* Not Applicable. This is a manufactured product, not a machine learning model.
H. How the ground truth for the training set was established
* Not Applicable. This is a manufactured product, not a machine learning model.
Summary of Performance Data Conducted:
The device's substantial equivalence to predicate devices was supported by the following performance data:
- Shipping and Distribution Testing: Demonstrated "the integrity and robustness of the bag system packaging within the distribution environment."
- HF-Connector Testing:
- Evaluated torque for removing protective cap and male/female connector torque.
- Evaluated breaking strength of the cone.
- Evaluated leakage and tightness of connectors.
- Evaluated pull-out force of connectors and injection ports.
- Biocompatibility Testing: Conducted in accordance with ISO 10993-1:2018 and FDA guidance. Assessed endpoints including Chemical Characterization, Cytotoxicity, Sensitization, Irritation, Material Mediated Pyrogenicity, Hemocompatibility, and Genotoxicity. A toxicological risk assessment was also performed. These tests confirm the safety of the materials used in the container closure system.
- Human Factors Validation Testing: Concluded the solutions were "safe and effective for their intended users, uses, and use environments."
- Not Applicable for this device type: Electrical Safety and Electromagnetic Compatibility (EMC), Software Verification and Validation Testing, Animal Studies, Clinical Studies.
Conclusion:
The 510(k) summary concludes that the Bicarby and Ci-Ca solutions are substantially equivalent to the predicate devices and are safe and effective for their intended use, based on the non-clinical performance testing summarized above, primarily focusing on materials, packaging integrity, and chemical specifications.
Ask a specific question about this device