510(k) Data Aggregation

    K Number
    K141106

    Validate with FDA (Live)

    Date Cleared
    2014-10-10

    (164 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FUSIO Screw Fuze System is intended for large bone fixation, including sacroiliac joint fusion for conditions including sacroiliac joint disruptions and degenerative sacroiliitis.

    Device Description

    The Frontier Devices FUSIO Screw Fuze System consists of titanium bone screws of 7mm, 9.5mm, and 11mm diameter of various lengths to accommodate patient anatomy. An optional serrated washer can be placed on the screw. The screws are made from titanium alloy, Ti-6Al-4V, that conforms to the ASTM F136 standard.

    AI/ML Overview

    The provided document describes the Frontier Devices FUSIO Screw Fuze System, a medical device for sacroiliac joint fusion. The document focuses on demonstrating substantial equivalence to predicate devices rather than providing a study proving performance against specific acceptance criteria in the typical sense of a clinical trial or algorithm performance study.

    Instead, the "acceptance criteria" here relate to demonstrating that the new device performs similarly in mechanical tests to legally marketed predicate devices, thereby meeting the FDA's requirements for 510(k) clearance.

    Here’s a breakdown based on the information provided:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Demonstrated Equivalence to Predicate)Reported Device Performance
    Insertion Torque:FUSIO Screw Fuze System:
    - Comparable insertion torque to predicate screws.- Tests showed "similar results as predicate screws" (Globus and X-Spine) into Grade 15 polyurethane foam blocks. Peak torque measured after 4 revolutions was recorded and compared using mean torque values with standard deviations. "The insertion torque measurements... support the fact that the Frontier Devices FUSIO Screw Fuze System screws are substantially equivalent to predicate screws."
    Mechanical Properties (Cross-Sectional):FUSIO Screw Fuze System:
    - Comparable ability to handle forces and stresses.- Engineering analysis (using SolidWorks) of the worst-case FUSIO Screw Fuze screw at the cross-section (middle of the screw) determined it could "handle forces and associated stresses similarly" to the legally marketed X-Spine Silex screw.
    Thread Properties (Projected Thread Engagement Area):FUSIO Screw Fuze System:
    - Comparable thread engagement area.- SolidWorks was used to calculate the projected threaded area, showing "substantial equivalence" to the predicate devices.
    Cyclic Bending Endurance:FUSIO Screw Fuze System:
    - Meet pass-fail criteria for endurance under cyclic loading.- Cyclic bending tests on the worst-case FUSIO Screw Fuze screw produced an endurance curve "which yielded acceptable factors of safety for the force and moment." The system "successfully passed the pass-fail criteria."
    Material Composition:FUSIO Screw Fuze System:
    - Made from titanium alloy (ASTM F136).- The screws are made from titanium alloy, Ti-6Al-4V, conforming to ASTM F136. This is consistent with predicate devices.
    Dimensions (Diameter and Length):FUSIO Screw Fuze System:
    - Similar diameters and lengths.- Available in 7mm, 9.5mm, and 11mm diameters of various lengths, which are "similar to the predicate devices."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated as a number of individual screws tested for each assessment (insertion torque, cyclic bending). The document mentions "representative screws" for insertion torque and "the worst-case FUSIO Screw Fuze screw" for engineering analysis and cyclic bending tests. This suggests a limited number of physical samples for mechanical testing, characteristic of device design verification.
    • Data Provenance: Not applicable in the sense of patient data. The "data" comes from laboratory mechanical testing of the FUSIO Screw Fuze System and comparative measurements/analyses against predicate devices.

    3. Number of Experts Used to Establish Ground Truth and Qualifications of Experts

    • This question is not applicable to this type of device submission. There is no "ground truth" to be established by medical experts for mechanical performance tests of bone screws. The "ground truth" here is the objective physical measurement of mechanical properties.

    4. Adjudication Method

    • Not applicable. There is no adjudication required for laboratory mechanical test results.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • No. This is a mechanical device, not an imaging or diagnostic AI device. MRMC studies are not relevant.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • Not applicable. This is a physical medical device (bone screw), not an algorithm or AI.

    7. The Type of Ground Truth Used

    • Objective Mechanical Measurements and Engineering Analysis: The "ground truth" for the performance tests consists of:
      • Direct physical measurements (e.g., peak insertion torque, force sustained during cyclic loading).
      • Engineering calculations and simulations (e.g., stress analysis using SolidWorks, projected threaded area).
      • Comparison against established ASTM standards (ASTM F136 for material).
      • Comparison against the known performance of legally marketed predicate devices.

    8. The Sample Size for the Training Set

    • Not applicable. This is a physical device, not a machine learning model. There is no concept of a "training set."

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As above, there is no training set. The design of the device would be informed by engineering principles, material science, and the existing predicate devices.
    Ask a Question

    Ask a specific question about this device

    K Number
    K100205

    Validate with FDA (Live)

    Date Cleared
    2010-12-13

    (322 days)

    Product Code
    Regulation Number
    882.5320
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Frontier Devices Neuro System is intended for use in selective trauma of the cranial skeleton and cranial surgery.

    Device Description

    The Frontier Devices Neuro System consists of titanium bone plates and screws of various shapes and sizes. The plates are made from unalloyed titanium that conforms to the ASTM F67 standard. Plates are 0.4 and 0.6 mm thick and come in a variety of shapes and sizes that are designed for cranial closure applications. The shapes and corresponding sizes include: Straight plates 9-12mm long (1x1, 2x2, 5 and 7 holes), Box plates 10mm x 10mm to 16mm x 16mm (4 hole), X-plate 8mm x 8mm (4 hole), Y-plate 16mm long (6 and 8 holes), Double Y-plate 18mm to 21mm long (6 and 8 hole), Strut plate 24 and 35mm long (6 and 8 hole). The burr hole covers are are made from unalloyed titanium that conforms to the ASTM F67 standard. They are 0.4 and 0.6 mm thick and are circular in shape with equi-spaced drill holes around the circumference. The burr hole covers have two basic shapes. The burr hole cover for shunt has a segment of the circular burr hole cover missing which is where a shunt device is placed. The other burr hole cover is a complete circle. The corresponding sizes include: Burr hole covers 12mm to 24mm diameter, Burr hole covers for shunt 12mm to 24mm diameter. The mesh is made from unalloyed titanium that conforms to ASTM F67 standard. They are 0.4 and 0.6mm thick and rectangular in shape, 90mm x 120mm. The screws are self-drilling screws that do not require a pre-drilled pilot hole and are made from a titanium alloy, Ti-6Al-4V, that conforms to the ASTM F136 standard. Screws are 1.5 to 1.8 mm in diameter and range in length from 3 to 6 mm.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Frontier Devices Neuro Closure System:

    Summary of Device Performance and Acceptance Criteria

    This 510(k) submission for the Frontier Devices Neuro Closure System (K100205) describes a medical device seeking substantial equivalence to predicate devices, rather than establishing entirely new performance criteria. Therefore, the "acceptance criteria" are primarily based on demonstrating similarity to existing, legally marketed devices.

    The performance testing focused on demonstrating that the new device performs similarly to the predicate devices and meets relevant safety standards for its materials.

    Acceptance Criteria CategoryReported Device Performance (Frontier Devices Neuro Closure System)
    Material CompositionPlates made from unalloyed titanium (ASTM F67). Screws made from titanium alloy, Ti-6Al-4V (ASTM F136).
    Material Biocompatibility/SafetyEndotoxin levels < 0.06 EU/ml for screws, plates, and mesh devices.
    Mechanical Performance (Screws)Screw insertion tests showed similar results to predicate screws.
    Dimensions & DesignPlates: 0.4 and 0.6 mm thick; various shapes and sizes for cranial closure. Burr hole covers: 0.4 and 0.6 mm thick, circular. Screws: 1.5 to 1.8 mm diameter, 3 to 6 mm length. Mesh: 0.4 and 0.6 mm thick, 90mm x 120mm.
    Intended UseIntended for use in selective trauma of the cranial skeleton and cranial surgery (same as predicate devices).
    Substantial Equivalence"All of the technical characteristics are substantially equivalent to the corresponding characteristics of the predicate devices." "information provided in this submission proves substantial equivalence to the predicate devices."

    Details of the Study/Testing:

    Based on the provided text, the "study" conducted for this device appears to be a series of bench tests and material characterization tests aimed at demonstrating "substantial equivalence" to predicate devices, rather than a clinical trial or a study involving human or animal subjects for efficacy or safety in a real-world scenario.

    1. Sample Size used for the test set and the data provenance:

      • Test Set Sample Size: Not explicitly stated for each test, but referred to as "representative screws" for insertion tests and "screws, plates and mesh devices" for endotoxin testing. No specific number for each item is provided.
      • Data Provenance: The tests were conducted internally by the manufacturer (Frontier Devices) or a contracted lab. The document does not specify the country of origin of the raw materials, but the company is based in Pelham, AL, USA. The testing is presumably prospective for this specific device.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not applicable. This device relies on physical and chemical testing against established material standards (ASTM F67, ASTM F136) and comparative bench testing against predicate devices. There is no "ground truth" established by human experts in the way clinical diagnostic devices require. The "ground truth" for material standards is the standard itself.
    3. Adjudication method for the test set:

      • Not applicable. As this involves objective physical and chemical measurements against standards and predicate device performance, there isn't a need for expert adjudication of results. The results are quantitative (e.g., endotoxin levels, screw insertion test results).
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not applicable. This is a physical implantable device (plates and screws), not a diagnostic or AI-driven system that would involve human readers or AI assistance.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is a physical medical device, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for the material components is adherence to established ASTM material standards (ASTM F67 for unalloyed titanium, ASTM F136 for Ti-6Al-4V titanium alloy) and meeting biological safety standards (LAL endotoxin limit of < 0.06 EU/ml). For mechanical performance (screw insertion), the ground truth is simply achieving results similar to the predicate devices.
    7. The sample size for the training set:

      • Not applicable. This device did not involve machine learning or AI algorithms, so there is no "training set."
    8. How the ground truth for the training set was established:

      • Not applicable, as there was no training set.