Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    K Number
    K121860

    Validate with FDA (Live)

    Device Name
    ESA615
    Manufacturer
    Date Cleared
    2013-01-25

    (213 days)

    Product Code
    Regulation Number
    870.2300
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ESA615 Electrical Safety Analyzer is an electronic signal source and measurement device for verifying the electrical safety of medical devices. The ESA615 also provides ECG simulation and performance waveforms to verify patient monitors are performing within their operating specifications. The ESA615 provides following function categories: ECG Functions, ECG-Performance Testing.

    Device Description

    Fluke Biomedical's ESA615 Electrical Safety Analyzer (hereafter referred to as the ESA615) provides a basis for verifying the electrical safety of medical devices. The Product also provides ECG simulation and performance waveforms to verify patient monitors are performing within their operating specifications. The ESA615 can be controlled by a suitably equipped computer to carry out tests, via either a cable or wirelessly. The ESA615 consists of the following components: 1) Printed Circuit Board Assemblies using surface mount components and firmware loaded in embedded processors. 2) Plastic injection molded case parts. 3) Liquid Crystal Display for user interface. 4) Power cord for powering the unit at 120V and 60Hz.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study information for the ESA615 Electrical Safety Analyzer, based on the provided text:

    Device: ESA615 Electrical Safety Analyzer

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission claims substantial equivalence to the predicate device (MPS450) based on similar technological characteristics and performance, not against specific, quantifiable acceptance criteria. The "differences" column below highlights how ESA615 performance differs from the predicate.

    FeatureAcceptance Criteria (Predicate: MPS450)Reported Device Performance (ESA615)Met Criteria? (Comparison to Predicate)
    Intended UseTest and verify basic operation of patient monitoring devices for ECG, Respiration, Invasive Blood Pressure, and Cardiac Output.Verify electrical safety of medical devices; provide ECG simulation and performance waveforms.Differs: Fewer functions (no Respiration, Invasive Blood Pressure, Cardiac Output), but adds Electrical Safety Analysis.
    ConstructionPlastic CasePlastic CaseYes
    Size5.5" wide x 7.5" deep x 1.8" high6.94" wide x 3.3" deep x 11.2" highBigger
    Weight2 lbs3.5 lbsHeavier
    Display4 line x 20-character super twist LCD display3" diagonal screen Monochrome STN transflectiveLarger screen
    Function KeySoftSoftYes
    ECG Leads10 binding posts5 applied parts jacks (with optional expander box)Fewer applied parts
    Communications PortRS232USBChange from RS232 to USB
    Power9V alkaline battery & battery eliminatorNo Battery - AC line powered onlyNo battery due to higher power demands
    Lead Configuration12 leads5 leadsFewer leads
    Amplitude Accuracy±2% of setting± 5% of 1mV settingLess accurate
    Rate Accuracy±1% setting± 2% of settingLess accurate
    Normal Sinus Rhythm30, 40, 45, 60, 80, 90, 100, 120, 140, 160, 180, 200, 220, 240, 260, 280, 300 bpm30, 60, 120, 180, 240 bpmFewer NSR waves
    Sine wave0.5, 5, 10, 40, 50, 60, 100 Hz10, 40, 50, 60, 100 HzFewer frequencies
    Square wave0.125, 2.0 Hz0.125, 2.0 HzYes
    Triangle wave2.0, 2.5 Hz2.0 HzFewer frequencies
    Pulse wave30, 60 bpm, 60 ms pulse width30, 60 bpm, 63 ms pulse widthWider pulse width
    Cable ConnectorECG leads, 10 binding postingsECG leads, 5 binding postingsFewer binding postings

    2. Sample size used for the test set and the data provenance

    The document mentions "Laboratory studies have been conducted with a representative patient monitor..." but does not specify the sample size (number of "representative patient monitors" or tests performed) or the data provenance (country of origin, retrospective/prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. The device is an analyzer, not a diagnostic tool requiring expert interpretation. The "ground truth" for its performance is its physical and electronic output measurements compared against its published specifications.

    4. Adjudication method for the test set

    Not applicable, as this is a device performance verification, not a clinical study requiring human interpretation and multi-expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is an electrical safety analyzer and ECG simulator, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Yes, the testing described appears to be standalone performance verification of the ESA615 itself, against its published specifications. The description "Laboratory studies have been conducted with a representative patient monitor to verify and validate the ESA615 will perform within its' published specifications" indicates this.

    7. The type of ground truth used

    The ground truth for the test set appears to be its own published specifications, as verified by "Laboratory studies... to verify and validate the ESA615 will perform within its' published specifications."

    8. The sample size for the training set

    Not applicable. This is a hardware device with firmware, not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K121722

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2012-10-11

    (121 days)

    Product Code
    Regulation Number
    870.2300
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ESA620 Electrical Safety Analyzer is an electronic signal source and measurement device for verifying the electrical safety of medical devices. The ESA620 also provides ECG simulation and performance waveforms to verify patient monitors are performing within their operating specifications.

    The ESA620 provides following function categories:

    • ECG Functions .
    • ECG-Performance Testing ●
    Device Description

    Fluke Biomedical's ESA620 Electrical Safety Analyzer (hereafter referred to as the ESA620) provides a basis for verifying the electrical safety of medical devices. The Product also provides ECG simulation and performance waveforms to verify patient monitors are performing within their operating specifications.

    The ESA620 consists of the following components:

    1. Printed Circuit Board Assemblies using surface mount components and firmware loaded in embedded processors.
    2. Plastic injection molded case parts.
    3. Liquid Crystal Display for user interface.
    4. Power cord for powering the unit at 120V and 60Hz.
    AI/ML Overview

    This document describes the regulatory submission for the Fluke Biomedical ESA620 Electrical Safety Analyzer. It focuses on demonstrating substantial equivalence to a predicate device (MPS450), rather than proving that a device meets and acceptance criteria via a study. As such, many of the requested details about a study to prove acceptance criteria are not present in the provided text.

    Here is an analysis based on the provided document:

    Study Type and Purpose:
    The document describes non-clinical testing performed to verify and validate the ESA620. The primary purpose of this testing was to demonstrate substantial equivalence to the predicate device (MPS450), not to meet predefined acceptance criteria in a robust clinical or standalone AI study.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly list "acceptance criteria" in the traditional sense of a performance study with specific target metrics for accuracy, sensitivity, specificity, etc. Instead, it compares the technological characteristics and intended use of the ESA620 against its predicate device, the MPS450, to establish substantial equivalence.

    The "differences" column in the provided table implicitly represents areas where the ESA620's performance or features might be considered against the predicate. However, these are presented as factual differences rather than "acceptance criteria" where a specific pass/fail threshold is defined for each.

    FeaturePredicate (MPS450) Performance/CharacteristicESA620 Performance/CharacteristicAcceptance Criteria/Target (Implied)Reported Device Performance (ESA620)Meets Criteria?
    Intended UseTest and verify basic operation of patient monitoring devices (ECG, Respiration, Invasive BP, Cardiac Output). Lab environment, not patient-connected, not for calibration, not OTC.Electronic signal source and measurement for electrical safety; ECG simulation, performance waveforms for patient monitors. Lab environment, not patient-connected, not for calibration, OTC.Similar intended use, with some functional differences (less functionality in some areas, added electrical safety).As described, fulfills the stated intended use.Yes (demonstrated substantial equivalence)
    ConstructionPlastic CasePlastic CaseSamePlastic CaseYes
    Size5.5" W x 7.5" D x 1.8" H12.5" W x 9.5" D x 5" HCan be largerLargerYes (accepted difference as not impacting safety/effectiveness for the stated use)
    Weight2 lbs9.5 lbsCan be heavierHeavierYes (accepted difference)
    Display4 line x 20-character super twist LCD5.2" diagonal screen Monochrome STN transflectiveLarger screenLarger screenYes (accepted difference)
    Function KeySoftSoftSameSoftYes
    ECG Leads10 binding posts; compatible with snaps, 3.2mm/4mm electrodes, banana plugs10 binding posts; compatible with snaps, 3.2mm/4mm electrodes, banana plugs (with/without adapter)SameAs describedYes
    Communications PortRS232USBAdvancement in technology (USB equivalent/better)USBYes (accepted as an advancement)
    Power9V alkaline battery or AC transformerAC line powered onlyNo battery due to higher power requirements (accepted)AC line powered onlyYes (accepted as a functional difference)
    ECG Lead Config12 leads12 leadsSame12 leadsYes
    Amplitude Accuracy$\pm$ 2% of setting (MPS450)$\pm$ 5% of 1mV setting (ESA620)Potentially less stringent, but acceptable for intended use.$\pm$ 5% of 1mV settingYes (difference noted, accepted for equivalence)
    Rate Accuracy$\pm$ 1% setting (MPS450)$\pm$ 2% of setting (ESA620)Potentially less stringent, but acceptable for intended use.$\pm$ 2% of settingYes (difference noted, accepted for equivalence)
    Normal Sinus RhythmMore options (30-300 bpm, 14 settings)Fewer options (30-240 bpm, 5 settings)Fewer options are acceptable for the device's purpose.Fewer NSR wavesYes (accepted difference)
    Sine WaveMore options (0.5-100 Hz, 7 settings)Fewer options (10-100 Hz, 5 settings)Fewer frequencies are acceptable.Fewer frequenciesYes (accepted difference)
    Square Wave0.125, 2.0 Hz0.125, 2.0 HzSame0.125, 2.0 HzYes
    Triangle Wave2.0, 2.5 Hz2.0 HzFewer frequencies are acceptable.Fewer frequenciesYes (accepted difference)
    Pulse Wave30, 60 bpm, 60 ms pulse width30, 60 bpm, 63 ms pulse widthSlight difference in pulse width, accepted.Wider pulse widthYes (accepted difference)
    Cable ConnectorECG leads, 10 binding postingsECG leads, 10 binding postingsSameECG leads, 10 binding postingsYes
    Electrical Safety AnalysisNot presentPresentAdded functionalityPresentN/A (new feature, not compared)
    Respiration, BP, Cardiac OutputPresentNot presentRemoved functionality, but still equivalent for overall purpose due to other new features.Not presentN/A (removed feature)

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: The document states "Laboratory studies have been conducted with a representative patient monitor to verify and validate the ESA620 will perform within its' published specifications -Document: NPI-05012012-00002". This indicates a very limited "sample size" of only one representative patient monitor for the non-clinical tests. This is typical for a device meant for testing other equipment, where the focus is on the device's output accuracy and stability, rather than diverse patient data.
    • Data Provenance: The studies were conducted in a laboratory setting ("Non-Clinical Test Data"). The country of origin for the data is not specified, but given the submitter (Fluke Biomedical) is US-based and the submission is to the FDA, it is highly likely to be U.S.-based. The data is prospective in the sense that it was generated specifically for this submission to validate the new device.

    3. Number of Experts Used to Establish Ground Truth and Qualifications:

    • This information is not provided in the document. For a technical device like an electrical safety analyzer, "ground truth" would typically refer to the accuracy of its measurements and simulations compared to established standards and calibrated reference equipment, rather than expert human interpretation. The "validation" likely refers to engineering design verification and validation by qualified engineers rather than clinical experts.

    4. Adjudication Method for the Test Set:

    • This information is not applicable/not provided as the validation described is non-clinical performance testing of measurement parameters against specifications, not an assessment requiring multiple human readers and adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, an MRMC study was not done. This type of study is specifically designed for assessing the impact of AI assistance on human reader performance, typically in diagnostic imaging. The ESA620 is a test and measurement device, not a diagnostic tool requiring human interpretation.

    6. Standalone (Algorithm Only) Performance Study:

    • The document implies a standalone performance assessment of the device, but not in the context of an "algorithm" as would be found in AI/ML devices. The "Non-Clinical Test Data" section states "The ESA620 software has been successfully validated to confirm the performance of the device." This confirms that the device's technical performance (measurements, simulations) was evaluated independently against its specifications.

    7. Type of Ground Truth Used:

    • The "ground truth" for this device would be established engineering specifications, calibrated reference standards, and potentially a "golden standard" or master device that has been proven accurate. The statement "to verify and validate the ESA620 will perform within its' published specifications" indicates that the ground truth was the published specifications themselves, which are derived from metrology standards.

    8. Sample Size for the Training Set:

    • This information is not applicable/not provided. The ESA620 is a hardware device with embedded firmware; it is not an AI/ML device that requires a "training set" in the context of machine learning. The validation described is traditional software and hardware verification.

    9. How the Ground Truth for the Training Set Was Established:

    • This information is not applicable. As it's not an AI/ML device, there's no "training set" or corresponding ground truth establishment process in that sense. The "ground truth" for the device's function (e.g., whether it accurately outputs a 1mV ECG signal) would be based on fundamental electrical engineering principles, national/international standards (e.g., IEC 60601 for medical electrical equipment safety), and calibration to a higher-level physical standard.
    Ask a Question

    Ask a specific question about this device

    K Number
    K110429

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-09-02

    (199 days)

    Product Code
    Regulation Number
    870.2300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended us of ProSim 4 is to test and verify the basic operation of patient monitoring devices or systems used to monitor various physiological parameters of a patient, including ECG, Respiration, Invasive blood pressure and Non-invasive blood pressure.

    The ProSim 4 Vital Signs Simulator provides electronic and pneumatic simulation of physiological parameters for determining that patient monitoring devices or systems are performing within their operating specifications. The device includes the following physiological simulations:

    • ECG adult or neonatal
    • Invasive and non-invasive blood pressure
    • Respiration

    The intended us of ProSim 6 and ProSim 8 is to test and verify the basic operation of patient monitoring devices or systems used to monitor various physiological parameters of a patient, including ECG, Respiration, Invasive blood pressure, Non-invasive blood pressure and Cardiac output. Additionally, the devices provide an optical signal to verify that the electronics within the pulse oximeter probe are functional.

    The ProSim 6 and ProSim 8 Vital Signs Simulators provide electronic and pneumatic simulation of physiological parameters for determining that patient monitoring devices or systems are performing within their operating specifications. The devices provide the following physiological simulations:

    • ECG adult or pediatric
    • Invasive and non-invasive blood pressure
    • Respiration
    • Temperature
    • Cardiac Output
    • Fetal Simulation includes fetal, maternal ECG, & uterine contractions (ProSim 8 only)
      Additionally, the devices provide an optical signal to verify that the electronics within the pulse oximeter probe are functional
    Device Description

    Fluke Biomedical's ProSim 4 (hereafter referred to as the ProSim) provides a basis to train, evaluate, and perform preventive maintenance for virtually all patient monitors found in the healthcare industry. This is accomplished with multiple physiological simulations for ECG electrical signals, respiration electrical signals, invasive blood pressure (IBP) electrical signals and non-invasive blood pressure pulses. The ProSim is a lightweight, battery powered unit that is portable enough to test a patient monitor anywhere the monitor is being used.

    ProSim vital signs simulator consists of the following components:

    1. Printed Circuit Board Assemblies using surface mount components and firmware loaded in embedded processors.
    2. Plastic iniection molded case parts.
    3. Stepper Motor and piston pump for pneumatic simulation that makes reliable pressure pulses.
    4. Liquid Crystal Touch Screen Display for user interface. User interface follows modern and ergonomic concepts.
    5. Lithium Ion rechargeable battery for portable operation, giving user flexibility and portability.

    Fluke Biomedical's ProSim 6 and ProSim 8 (hereafter referred to as the ProSim) provides a basis to train, evaluate, and perform preventive maintenance for virtually all patient monitors found in the healthcare industry. This is accomplished with multiple physiological simulations for ECG electrical signals, respiration electrical signals, invasive blood pressure (IBP) electrical signals, non-invasive blood pressure (NIBP) pressure pulses, temperature electrical signal, cardiac output electrical signal, and pulse oximetry SPO2 optical simulated light attenuation. The ProSim is a lightweight, battery powered unit that is portable enough to test a patient monitor anywhere the monitor is being used.

    ProSim vital signs simulator consists of the following components:
    l ) Printed Circuit Board Assemblies using surface mount components and firmware loaded in embedded processors.

    • 2) Plastic injection molded case parts.
      1. Stepper Motor and piston pump for pneumatic simulation that makes reliable pressure pulses.
      1. Liquid Crystal Display for user interface. User interface follows modern and ergonomic concepts.
      1. Lithium Ion rechargeable battery for portable operation, giving user flexibility and portability.
    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the Fluke Biomedical ProSim 4, ProSim 6, and ProSim 8 devices:

    Executive Summary:

    The Fluke Biomedical ProSim 4, ProSim 6, and ProSim 8 are vital signs simulators designed to test and verify the basic operation of patient monitoring devices. The devices function by providing electronic and pneumatic simulations of various physiological parameters. Extensive non-clinical bench testing was conducted to demonstrate that the ProSim devices perform within their published specifications and are substantially equivalent to their predicate devices (MedSim300B, Cufflink, and Index 2MF SPO2 for ProSim 6/8). Clinical testing was not performed, as the devices are intended for laboratory use for preventative maintenance and not for direct patient use or medical equipment calibration. The ground truth for the testing was the published specifications of the device itself, verified through laboratory measurements.


    1. Table of Acceptance Criteria and the Reported Device Performance:

    The document provides a detailed comparison table between the ProSim devices and their predicate devices, indicating various performance characteristics. Since the acceptance criteria are essentially the published specifications and the study aims to show the device performs within those specifications, the "reported device performance" is implicitly that the device met these specifications, thus demonstrating substantial equivalence.

    Below is a summarized table consolidating key features and their performance (or range of performance), with the acceptance criteria being that the ProSim meets or improves upon the predicate device's comparable specification or meets its own stated specifications if it's a new feature or different range.

    ProSim 4 (compared to MedSim 300B and Cufflink)

    FeaturePredicate Devices (Range/Specification)ProSim 4 (Reported Performance/Specification)Acceptance Criteria
    Display2x24 char LCD (MedSim); 8x20 char alpha & 64x240 graphics (Cufflink)¼ VGA graphic LCD Touch Color DisplayMore display, Touch Screen, and Color – Met/Exceeded
    IBP Channels4 channels (MedSim)1 channelReduced per market requirements and use – Met
    Comm PortRS232USBAdvancement in technology – Met
    Battery TestLimited low battery indication (MedSim); No battery (Cufflink)Multiple levels of battery life indicationPredicate checks at one level, ProSim checks at multiple levels – Met/Exceeded
    Power2x9V alkaline (MedSim); AC line (Cufflink)Li-Ion rechargeable battery w/ low battery indicator; battery eliminator (115VAC)Longer operating life with modern battery technology – Met/Exceeded
    Amplitude accuracy (ECG)+/- 5%, 2Hz @ 1.0 mV p-p SQ wave Lead II (MedSim)+/- 2% setting lead IIMore accurate on newer devices – Met/Exceeded
    NSR rates (ECG)30 to 300 BPM (MedSim)30 to 320 BPMWider range due to market preferences – Met/Exceeded
    NSR amplitudes (ECG)50 uV to 5.5mV (MedSim)1mVLimited amplitude for basic simulation – Met
    Pacemaker0.1 to 2.0 ms width, -700 to +700 mV (MedSim)1 ms width, 3mVBasic simulation for targeted market – Met
    Impedance variation (Resp)0 to 3 ohms (MedSim)1.0 ohmBasic simulation for targeted market – Met
    Respiration rates15, 20, 30, 40, 60, 120 brpm (MedSim)10, 20, 30, 40, 50, 60, 70, 80, 90 & 100 brpmMore Respiration rates – Met/Exceeded
    Apnea (Resp)Off, Continuous, momentary, 12 & 32 s (MedSim)Off & ContinuousLess apneas for basic simulation – Met
    Transducer Sensitivity (IBP)5 or 40 uV/V/mmHg (MedSim)5 uV/V/mmHgFewer selections for basic simulation – Met
    Static pres. Selection (IBP)Manual and automatic (MedSim)ManualLimited selection mode per target market and use – Met
    Dynamic BP selections (IBP)Arterial, left and right ventricle, pulmonary artery, pulmonary wedge, Swan-Ganz (MedSim)Arterial and left ventricleFewer selections for basic simulation – Met
    Static BP selections (IBP)-10, -5, 0, 20, 30, 40, 80, 100, 200, 250 & 300 mmHg (MedSim)0, 80, 160 & 250 mmHgFewer selections for basic simulation – Met
    Manometer (NIBP)N/A (MedSim); Max. 499.75 mmHg (Cufflink)0 to 400 mmHgLower range per market requirements – Met
    Simulation (NIBP)N/A (MedSim); Systolic/Diastolic simulations. Adult 60/30 to 255/195 (Cufflink)Systolic/Diastolic Adult – 60/30, 120/80, 150/100 & 200/150; Neonatal 35/15 & 70/40Fewer selections for basic simulation; Adult and Neonatal – Met
    Synchronization to ECG (NIBP)N/A (MedSim); 30 to 240 BPM (Cufflink)Up to 150 BPMLimited for basic simulation – Met

    ProSim 6 & 8 (compared to MedSim 300B, Index 2, and Cufflink)

    FeaturePredicate Devices (Range/Specification)ProSim 6 & 8 (Reported Performance/Specification)Acceptance Criteria
    Intended UseMedSim: ECG, Resp, BP, Temp, CO; Index 2: SpO2; Cufflink: NIBPECG, Resp, IBP, NIBP, Temp, CO, SpO2 (Fetal Sim for ProSim 8)Additional functions of noninvasive blood pressure and pulse oximetry simulation. – Met/Exceeded
    DisplayMedSim: 2x24 char LCD; Index 2: 2x24 char LCD; Cufflink: 8x20 char alpha & 64x240 graphics¼ VGA graphic LCD Color DisplayMore display, Color. – Met/Exceeded
    High level ECG¼" standard phone jack w/ lead II waveform at .2V/mV (MedSim)BNC jack for 0.5V/mV output into 50 Ohm impedanceOutput to oscilloscope via BNC is preferred by customers. – Met/Exceeded
    IBP Channels4 independent BP channels (MedSim)2 channelsNumber of channels reduced per market requirements and use. – Met
    Comm PortRS232 (all predicates)USBChange from RS232 to USB data port with advancement in technology. – Met/Exceeded
    Battery TestLimited low battery indication (MedSim, Index 2); No battery (Cufflink)Multiple levels of battery life indicationPredicate devices only check at one level. ProSim checks battery status at multiple charge levels. – Met/Exceeded
    Amplitude accuracy (ECG)+/- 5%, 2Hz @ 1.0 mV p-p SQ wave Lead II (MedSim)+/- 2% setting lead IIMore accurate on newer devices due to market preferences and technology improvements. – Met/Exceeded
    NSR rates (ECG)30 to 300 BPM (MedSim)30 to 360 BPMWider range due to market preferences. – Met/Exceeded
    Square and/or Pulse wavesSquare at 2 Hz and 0.125 Hz (MedSim)Square at 2.5, 2 & 0.125 Hz. Pulse at 60ms / 60 and 30 BPMMore choices. – Met/Exceeded
    Sine waves0.05 to 100 Hz (MedSim)0.05 to 150 HzMore choices. – Met/Exceeded
    Triangle wave2 Hz (MedSim)0.125, 2 and 2.5 HzMore choices. – Met/Exceeded
    R Wave detectorYes (MedSim)YesNone. – Met
    QRS Detection and Tall T-wave rejectionNo (MedSim)YesImproved capability. – Met/Exceeded
    Impedance variation (Resp)0 to 3 ohms (MedSim)0 to 5 ohmsExpanded capability to meet new market requirements. – Met/Exceeded
    Respiration rates15, 20, 30, 40, 60, 120 brpm (MedSim)15 to 120 brpm in incremental stepsMore Respiration rates due to market requirements and use. – Met/Exceeded
    Apnea (Resp)Off, Continuous, momentary, 12 & 32 s (MedSim)Off, Continuous, momentary, 12, 22 & 32 SAdditional apneas. – Met/Exceeded
    Blood temperatures (CO)36C to 38C and user programmable (MedSim)36C to 38C in incremental stepsUser programmable not included due to market requirements and use. – Met
    Injective temp (CO)Chilled (2C) (MedSim)Chilled (0C) or 24 CMore selections per market requirements. – Met/Exceeded
    Fixed blood flow rate (CO)3, 5, 7 L/min (MedSim)2.5, 5, 10 L/minDifferent selections per market requirements. – Met
    Output trend (CO)1 normal, 2 defective (MedSim)NoNot included due to market requirements and use. – Met
    Exciter range (IBP)2 to 16 V/DC to 4 kHz (MedSim)2 to 16 V/DC to 5kHzHigher frequency range driven by market trend and technology. – Met/Exceeded
    Static pres. Selection (IBP)Manual and automatic (MedSim)ManualLimited selection mode per target market and use. – Met
    Dynamic BP selections (IBP)Arterial, left and right ventricle, pulmonary artery, pulmonary wedge, Swan-Ganz (MedSim)Arterial, radial artery, left and right ventricle, pulmonary artery, pulmonary wedge, right atrium, left atrium and Swan-GanzMore selections. – Met/Exceeded
    Static BP selections (IBP)-10, -5, 0, 20, 30, 40, 80, 100, 200, 250 & 300 mmHg (MedSim)-10 to 300 mmHg in incremental stepsMore selections. – Met/Exceeded
    Temperature (Sim)0, 24, 37 and 40C (MedSim)30C to 42C in incremental stepsDifferent selections per market requirements. – Met
    SpO2 AccuracySelect 50% to 100% in 1% increments - accuracy: 75%to 100% ± 1% plus the accuracy of the pulse oximeter under test. 50%-75%, ± 2% plus accuracy of the oximeter under test. Under 50%, unspecified. (Index 2)Select 30% to 100% in 1% increments w/ oximeter manufacturer's R-curve or Fluke Biomedical R-curves (91-100% ±(3 counts + UUT accuracy), 81-90% ±(5 counts + UUT accuracy), 71-80% ±(7 counts + UUT accuracy), <70% monotonic unfilled accuracy)More selections per market requirements. – Met/Exceeded
    SpO2 Test featuresTransmission Light Control (TLC) feature for different light attenuation (Index 2)Transmission of light selected through selection of finger type: light finger, thick dark finger or neonate.Better user interface and understanding of transmission with ProSim selection of light transmission levels. – Met/Exceeded
    Magnetic HolderN/A (Index 2)YesIncludes tested magnetic holder for SPO2 module. – Met/Exceeded
    Simulation (NIBP)Systolic/Diastolic simulations. Adult 60/30 to 255/195 (Cufflink)Systolic/Diastolic Adult 60/30 to 255/195; Neonatal 35/15 to 255/195Adult and Neonatal available. – Met/Exceeded
    Arrhythmias (NIBP)Premature atrial contraction, Premature ventricular contraction, Atrial fibrillation, Missed Beat, aberrant Sinus conduction (Cufflink)Premature atrial contraction, Premature ventricular contraction, Atrial fibrillation, Missed Beat.No aberrant Sinus conduction. Not needed per market use. – Met
    Synchronization to ECG (NIBP)30 to 240 BPM (Cufflink)30 to 240 BPMNone. – Met

    2. Sample Size Used for the Test Set and the Data Provenance:

    • Sample Size: The document does not specify a numerical sample size for individual tests (e.g., how many measurements were taken for each parameter). It states that "Laboratory studies have been conducted with a representative patient monitor..." and "The ProSim [4/6/8] software has been successfully validated..." This implies that the testing involved a single representative device (the ProSim itself) undergoing various tests against its own specifications or in comparison to predicate devices.
    • Data Provenance: The studies were non-clinical laboratory studies conducted by Fluke Biomedical. The location of the laboratory is likely the manufacturer's facility in the USA (Fluke Biomedical, Everett, WA, or Solon, OH). The data is retrospective in the sense that it's a summary of completed bench tests demonstrating performance.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

    • Number of Experts: The document does not specify the number of experts.
    • Qualifications of Experts: The ground truth for this type of device, a vital signs simulator, is typically based on electrical and pneumatic engineering principles and established medical device standards for physiological parameter simulation (e.g., AAMI, IEC standards for ECG, BP). The testing would be performed by trained technicians or engineers proficient in metrology and test equipment, verifying the device's output against known, calibrated reference standards. While not explicitly stated, it's implied these are qualified personnel within Fluke Biomedical's R&D or Quality Assurance departments.

    4. Adjudication Method for the Test Set:

    • Adjudication Method: Not applicable in the traditional sense of human consensus for, say, image interpretation. The "adjudication" is based on objective measurement against defined performance specifications and comparison to predicate devices. The results are summarized by the manufacturer.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done:

    • No, an MRMC study was not done. This type of study is typically performed for diagnostic devices where human readers interpret medical images or data. The ProSim devices are test equipment for patient monitors, not diagnostic tools. Their effectiveness is measured by their ability to accurately simulate physiological signals for verification and maintenance, not by improving human diagnostic performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    • Yes, a standalone study was done. The entire "Non-Clinical Test Data" section describes rigorous bench testing of the ProSim devices themselves, which are algorithmic-driven instruments, to verify their performance within published specifications. The statement "The ProSim [4/6/8] software has been successfully validated to confirm the performance of the device" directly supports this. The device's function is standalone simulation; there isn't a "human-in-the-loop" once the simulation parameters are set.

    7. The Type of Ground Truth Used:

    • Engineering Specifications and Calibrated Reference Standards: The ground truth used is the device's own published specifications (e.g., amplitude accuracy, rate ranges, impedance values) and the measured performance of predicate devices. Performance was verified against these using calibrated test equipment and established metrology practices.

    8. The Sample Size for the Training Set:

    • Not applicable. The ProSim devices are simulators, not AI/machine learning models that require a training set. They are designed based on established physiological principles and engineering design.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable. As stated above, there is no "training set" in the context of an AI/ML model for these devices.
    Ask a Question

    Ask a specific question about this device

    K Number
    K103336

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-04-13

    (149 days)

    Product Code
    Regulation Number
    870.2300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MPS450 Multiparameter Simulator is an electronic signal source for determining that patient monitors are performing within their operating specifications. The MPS450 provides the following function categories: ECG Functions Arrhythmia Functions ECG-Performance Testing Respiration Blood Pressure Temperature Cardiac Output (Optional) Fetal/Maternal ECG & IUP (Optional)

    Device Description

    Fluke Biomedical's MPS450 Multiparameter Simulator (hereafter referred to as the MPS450) provides a basis to train, evaluate, and perform preventive maintenance for virtually all patient monitors found in the healthcare industry. This is accomplished with multiple physiological simulations for ECG, blood pressure, respiration, temperature, artifact, and arrhythmia conditions. The MPS450 is a lightweight, battery powered unit that is portable enough to test a patient monitor anywhere the monitor is being used.

    AI/ML Overview

    The provided document describes the Fluke Biomedical MPS450 Multiparameter Simulator, a device intended to test the operation of patient monitors by simulating various physiological parameters. The submission focuses on demonstrating substantial equivalence to a predicate device, the MedSim300B.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the MPS450 are implicitly defined by comparing its performance specifications to those of the predicate device, the MedSim300B. The differences highlight where the MPS450 either meets or exceeds the predicate's capabilities or where features were deemed "not needed by customers."

    Feature/ParameterAcceptance Criteria (Predicate - MedSim300B)Reported Device Performance (MPS450)Notes
    Physical Characteristics
    ConstructionAluminum casePlastic caseMPS450 is Lighter.
    Size10 L x 7 W x 3 H inches7.5 L x 6 W x 2 H inchesMPS450 is Smaller.
    Weight3.55 lbs1 lbs 8 ozMPS450 is Lighter.
    Display2 by 24 character LCD4 by 20 character LCDMPS450 has More lines on display.
    High level ECG connection¼" standard phone jack w/ lead II waveform at .2V/mV of ECG lead II signal0.2 V/mV +/- 5% of the ECG amplitude setting available on the BP3 connectorDifferent type of connector – same functionality.
    Battery testYes, 8.4V or 9V, low battery signal if < 6.2VNoNot needed by customers in MPS450 ("nice to have, but not Must have feature").
    Lead testYesNoNot a needed feature on MPS450.
    Functional Operation - ECG Simulation
    Lead configuration12 leads12 LeadsNone.
    Output impedances500 to 2000ohms to RL500 to 2000ohms to RLNone.
    Amplitude accuracy+/- 5%, 2Hz @ 1.0 mV p-p SQ wave Lead II+/- 2% setting lead IIMPS450 is More accurate.
    NSR rates30 to 300 bpm30 to 300 bpmNone.
    NSR amplitudes50 uV to 5.5mV50 uV to 5.5mVNone.
    ST Segments-0.8 to + 0.8 mV-0.8 to + 0.8 mVNone.
    Axis deviationNormal, horizontal and verticalNoFeature not needed by end-user customers.
    Pediatric ECGYes, R Wave width reduced to 40 msYes, R Wave width reduced to 40 msNone.
    Functional Operation - Performance Testing
    Square wave2 Hz2 Hz, 0.125 Hz2 selections rather than 1 in MPS450.
    Pulse4.0 sec.60msFaster pulse in MPS450.
    Sine wavesVarious -0.05 to 1000 HzVarious -0.5 to 100 HzOnly narrower range of sine waves needed in MPS450.
    Triangle wave2 Hz2 Hz and 2.5Hz2 selections in MPS450.
    R Wave detectorYesYesNone.
    Pacemaker0.1 to 2.0 msec pulse duration, -700 to +700 mV pulse amplitude0.1 to 2.0 msec pulse duration, -10 to +10 mV pulse amplitudePacer pulse amplitude lower for MPS450-testing at lower level than MedSim300.
    Premature beat insertionYes, PVC, PNC or PACYes, PVC, PNC or PACNone.
    Functional Operation - Respiration Simulation
    Normal baseline impedances500 to 2000 ohms ref. to RL500 to 2000 ohms ref. to RLNone.
    Lead selectionsI, II, and RL-LLLA or LLReference ECG leads only 2 choices in MPS450.
    Impedance variation0, 0.1, 0.2, 0.5, 1.0, and 3.0 ohms0.2, 0.5, 1.0, and 3.0 ohmsSimplified selection of Impedance variables in MPS450.
    Respiration rates15, 20, 30, 40, 60, 120 brpm15, 20, 30, 40, 60, 80, 100, 120 brpmAdded 80 and 100 brpm in MPS450.
    ApneaOff, continuous, momentary, 12 & 32 sOff, continuous, momentary, 12, 22 & 32 sAdded 22 second Apnea in MPS450.
    Functional Operation - Cardiac Output
    Catheter sizeFixed, 7F injectate vol. 10 ccFixed, 7F injectate vol. 10 ccNone.
    Blood temperatures36.0 to 38.0C; and user programmable37.0Single selection for temp. of blood in cardiac output.
    Injectate tempChilled (2C)Chilled (0C) or 24 C2 selections for injectate temp.
    Fixed blood flow rate3, 5, 7 L/min2.5, 5, 10 L/minSelection of blood flow rates widened in MPS450.
    CurvesNormal, interrupt, slow, L/R shuntNormal, faulty and L/R shuntDifferent curves for simulating different cardiac flow conditions in MPS450.
    Output trend1 normal, 2 defectiveNo.Feature not needed by customers in MPS450.
    Functional Operation - Blood Pressure
    I/O impedance300 ohms300 ohmsNone.
    Exciter range2 to 16 V/DC to 4 kHz2 to 16 V/DC to 5kHzHigher frequency in MPS450.
    Transducer Sensitivity5 or 40 uV/V/mmHg5 or 40 uV/V/mmHgNone.
    Level accuracy+/- 1% full scale; +/- 1mmHg+/- (2% setting + 2mmHg)Different ranges on MPS450 resulted in different accuracy specification.
    Static pres. selectionManual and automaticManualOnly Manual in MPS450.
    BP selections - DynamicArterial, left and right ventricle, pulmonary artery, pulmonary wedge, Swan-GanzArterial, Radial artery, left and right ventricle, pulmonary artery, pulmonary wedge, right atrium, left atrium and Swan-GanzMore BP simulation types in MPS450.
    BP selections - Static-10, -5, 0, 20, 30, 40, 80, 100, 200, 250, 300 mmHg-10, -5, 0, 20, 40, 50, 60, 80, 100, 150, 160, 200, 240, 320 and 400 mmHgMore static IBP pressures simulated in MPS450.
    BP cable connectorDIN styleMini-DINDifferent connector uses different set of IBP cables – mini-DIN is smaller, but DIN style is more rugged.
    Functional Operation - Temperature Simulation
    Temperature points34, 37, 40C0, 24, 37 and 40CAdditional temp simulation points in MPS450.
    Dynamic trendsHypothermia, spike and hyperthermiaNoFeature not implemented in MPS450 – not useful to end user customers.
    Probe compatibilitySeries 400 and 700Series 400 and 700None.
    Temp cable connectorDIN StyleMini-DinDifferent connector uses different set of IBP cables – mini-DIN is smaller, but DIN style is more rugged.

    Study Information:

    1. Sample size used for the test set and the data provenance:

      • The document states "Laboratory studies have been conducted with a representative patient monitor to verify and validate the MPS450 will perform within its' published specifications."
      • It also mentions "The MPS450 software has been successfully validated to confirm the performance of the device."
      • However, specific numbers for the sample size of "patient monitors" or detailed data provenance (e.g., country of origin, retrospective/prospective) are not provided. The studies are described as "Non-Clinical Test Data" and were conducted in a laboratory setting.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not provided. As "non-clinical test data" performed in a laboratory, the "ground truth" would likely be established by comparing the MPS450's output against known, highly accurate reference signals or measurements from calibrated laboratory equipment, rather than expert human interpretation.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • This information is not provided. Given the nature of a simulator, the "adjudication" would be a technical verification process against defined specifications, not a consensus-based method typically seen in image interpretation or diagnostic studies.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was not done. This type of study relates to AI-assisted human performance in diagnostic tasks, which is not applicable to a device like the MPS450, designed as a simulator for equipment testing. The MPS450 does not involve "human readers" or "AI assistance" in the diagnostic sense.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Yes, a standalone performance evaluation was done. The "Non-Clinical Test Data" section explicitly states: "Laboratory studies have been conducted with a representative patient monitor to verify and validate the MPS450 will perform within its' published specifications" and "The MPS450 software has been successfully validated to confirm the performance of the device." This describes the device's inherent functional performance without human intervention in the signal generation process.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The ground truth for this device would be established by known, precise electrical and physiological signal characteristics generated by the simulator itself and verified against highly accurate measurement equipment. It's a technical "ground truth" based on physics and engineering specifications, not clinical outcomes or expert labels.
    7. The sample size for the training set:

      • This device is a simulator, not an AI/machine learning algorithm that requires a "training set" in the conventional sense. Therefore, the concept of a training set sample size is not applicable. The device's "training" refers to its design, programming, and rigorous testing against engineering specifications.
    8. How the ground truth for the training set was established:

      • As explained above, there is no "training set" in the context of machine learning. The device's "ground truth" is based on the fundamental electronic and physiological parameters it is designed to simulate, which are derived from established medical and engineering standards. Its performance is verified against these standards in laboratory testing.
    Ask a Question

    Ask a specific question about this device

    K Number
    K083347

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-12-24

    (41 days)

    Product Code
    Regulation Number
    870.5325
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Impulse 7010 is used to determine that a defibrillator is performing within its operating specifications by providing multiple loads of 25, 50, 75, 100, 125, 150, 175 and 200 Ohms. The Impulse 7010 is used in conjunction with the Impulse 7000DP Defibrillator Analyzer.

    Device Description

    The Impulse 7010 Defibrillator Selectable Loads is an optional accessory to the Impulse 7000DP to simulate 25 to 200 Ohm thoracic impedance. Four 50 Ohm resistors are switched in combinations to make series or parallel circuits of 25, 50, 75, 100, 125, 150, 175 and 200 Ohms. Defibrillator output energy is measured by the Impulse 7000DP Defibrillator Tester.

    AI/ML Overview

    The provided text is a 510(k) summary for the Impulse 7010 Defibrillator Selectable Loads, which is an accessory to a defibrillator tester. The document focuses on establishing substantial equivalence to a predicate device for regulatory approval, rather than detailing a study with acceptance criteria and device performance evaluation in the manner typically found for AI/ML medical devices or diagnostic tools.

    Therefore, much of the requested information regarding acceptance criteria, study design, expert involvement, and ground truth establishment is not present in this regulatory submission for a simple accessory device. The device is a physical component (resistance box) for testing other medical devices, not a diagnostic or AI-powered system that would undergo extensive performance validation against a clinical ground truth.

    However, I can extract the relevant information where available and explain why other sections are not applicable.

    1. A table of acceptance criteria and the reported device performance

    This information is not explicitly provided in the 510(k) summary. For a device like the Impulse 7010, acceptance criteria would typically relate to the accuracy of the selectable resistance values it provides. The summary states that the device "simulates 25 to 200 Ohm thoracic impedance" and consists of "Four 50 Ohm resistors switched in combinations to make series or parallel circuits of 25, 50, 75, 100, 125, 150, 175 and 200 Ohms." The implication is that the accuracy of these resistance values is critical for its function as a defibrillator tester accessory, but specific numeric tolerances or performance data are not reported in this summary.

    Acceptance CriteriaReported Device Performance
    Not explicitly stated in the document (Likely relates to accuracy of selectable resistance values for 25, 50, 75, 100, 125, 150, 175, and 200 Ohms)"Four 50 Ohm resistors are switched in combinations to make series or parallel circuits of 25, 50, 75, 100, 125, 150, 175 and 200 Ohms." (Implies the device correctly provides these resistance values, but no specific performance metrics like tolerance or accuracy % are given.)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable. This device is an accessory that provides fixed resistance loads for testing other equipment. Its performance validation would involve electrical engineering tests (e.g., measuring the actual resistance output) rather than a clinical "test set" of data or patients. No clinical data or patient data is mentioned or relevant for this device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This device does not generate diagnostic outputs or interpretations that require a "ground truth" established by medical experts. Its "ground truth" is based on fundamental electrical engineering principles.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. No "test set" or clinical adjudication is described or relevant for this device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/ML device, nor is it a diagnostic tool that involves human readers interpreting cases.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a hardware accessory, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    Not applicable in the clinical sense. The "ground truth" for this device would be established by fundamental electrical measurement standards (e.g., a high-precision ohmmeter to verify the resistance values provided).

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K072114

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2008-01-28

    (180 days)

    Product Code
    Regulation Number
    870.5325
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Impulse 6000D/7000DP is used to determine that defibrillators and transcutancous pacemekers are performing within their performance specifications through the measurement of energy output.

    Device Description

    The Impulse 600013/7000DP is a portable, rechargeable, battery-operated defibrillator tester. The Impulse 7000DP also functions as a transcutaneous pacemaker tester. The device's defibrillator input is connected to the output of a defibrillator under test which provides a 50-Ohm test load, approximately the impedance of the human body. The Impulse 6000D/7000DP simulates patient electrocardiogram signals to trigger automated defibrillation when a ventricular fibrillation waveform is presented, and the device tests that the automated defibrillator does not advise shock when a normal sinus rhythm electrocardiogram is presented. The energy output delivered by the defibrillator under test is measured. The Impulse 7000DP also tests transcutaneous pacemaker outputs by presenting a low level electrocardiogram at various pulse rates and measures the response of the pacemaker under testing rate and amplitude. For pacemaker testing, the device incorporates inputs of 50 to 1500 Ohm impedance test loads. The Impulse 6000D is a defibrillator tester only without the pacemaker test option. Both models have 10 electrocardiogram outputs to simulate patient milli-volt level electrocardiogram signals to test combination patient monitors/defibrillators/pacemakers.

    The Impulse 7000DP has a USB type "B" interface to a PC to allow data download to a PC. It is electrically isolated from the measurement circuitry and allows remote control of the test from a PC. A BNC type connector on the rear panel also allows an oscilloscope to record the waveform output from the defibrillator under test, attenuated to a lower voltage level, and also electrically isolated from the measurement circuitry. Another BNC connector outputs a higher level signal (greater amplitude) to view the electrocardiogram signal on a second oscilloscope channel.

    The accessories for the Impulse 6000D/7000DP include an external power supply to operate and re-charge the internal battery. Optional adapters to connect defibrillators marketed by different manufacturers to the standard 4 mm banana style input jacks are available.

    Waveform analysis determines the characteristics of a defibrillator discharge pulse. Peak voltage amplitude, current, timing, overall energy and the refractory period of a pacemaker are recorded. Measurement is done by attenuating the high voltage signal to a lower voltage level, which is then input into an analog to digital converter. A digital signal processor calculates the measurements and corrects hardware error sources with mathematical calibration constants for any offset and gain errors.

    AI/ML Overview

    The provided text describes a medical device, the Impulse 6000D/7000DP Defibrillator Tester, and its FDA 510(k) clearance (K072114). However, it does not contain the specific details required to fully address your request.

    The document is a 510(k) summary and the FDA's clearance letter, which focuses on demonstrating substantial equivalence to a predicate device rather than presenting detailed performance study results with acceptance criteria.

    Therefore, I cannot provide a complete answer to your request. Here's specifically what is missing or cannot be inferred from the provided text:

    • Acceptance Criteria Table: No explicit acceptance criteria are mentioned for the device's performance. The document only states the device "is used to determine that defibrillators and transcutaneous pacemakers are performing within their performance specifications." These "performance specifications" are not detailed.
    • Study That Proves the Device Meets Acceptance Criteria: While a 510(k) submission requires performance data, the provided summary does not include the actual study design, results, or comparison to specific acceptance criteria. It mentions "waveform analysis determines the characteristics of a defibrillator discharge pulse," and "digital signal processor calculates the measurements and corrects hardware error sources," but doesn't elaborate on the studies or their outcomes.

    Here's what can be extracted/inferred from the provided text, acknowledging the significant gaps:

    Analysis of Acceptance Criteria and Performance Study:

    The provided document (a 510(k) summary and FDA clearance letter) does not detail specific acceptance criteria or a comprehensive study that proves the device meets those criteria. Instead, it focuses on establishing substantial equivalence to a predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance:

    • Acceptance Criteria: Not explicitly stated in the document. The device's purpose is to determine if defibrillators and pacemakers are performing "within their performance specifications." These specifications themselves are not provided.
    • Reported Device Performance: The document describes the device's functions (e.g., measuring energy output, simulating ECG signals, testing pacemaker outputs), but it does not provide quantitative performance metrics or results against any defined criteria.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified.
    • Data Provenance: Not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as no such expert-based ground truth establishment is described for a performance study in this document.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is a defibrillator and pacemaker tester, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • The device itself is a standalone tester. The document describes its functional capabilities (e.g., "Measurement is done by attenuating the high voltage signal to a lower voltage level, which is then input into an analog to digital converter. A digital signal processor calculates the measurements and corrects hardware error sources..."). This implies the device provides objective measurements independently. However, no specific details on standalone performance studies (e.g., accuracy, precision tests) are provided in this summary.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • For a device like this, the "ground truth" would likely be established through reference standards (e.g., highly accurate and calibrated physical measurement equipment, or known electrical signals) to which the device's measurements are compared. However, the document does not specify the ground truth methodology used for its internal validation or for its 510(k) submission package.

    8. The sample size for the training set:

    • Not applicable. This device is a measurement instrument, not a machine learning algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable (as above).

    In summary, the provided 510(k) documents focus on device description, intended use, and substantial equivalence to a predicate device, rather than detailed performance study reports with acceptance criteria, sample sizes, or ground truth methodologies. To obtain such information, one would typically need to review the full 510(k) submission package, which is not provided here.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1