Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K172661

    Validate with FDA (Live)

    Date Cleared
    2018-05-08

    (245 days)

    Product Code
    Regulation Number
    884.4530
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fetzer Medical Gynecological Forceps are hand-held instruments with dual blades that are indicated for pulling, grasping, holding, or compressing tissue during gynecological procedures.

    Device Description

    This 510(k) covers 31 devices a total of 132 variants representing three general designs (vulsellum, tenaculum, and hysterectomy). These products are reusable manual surgical instruments to be used in gynecological procedures. They are scissor-like, self-retaining devices with ring handles and two blades and made of high-grade stainless steel or titanium. They are available in various sizes with different blade designs at the distal end (e.g., straight or curved). Some of these devices have an additional grasping claw at the distal end. The blades are heavily serrated or may have a profiled longitudinal groove to provide extra grip of the organ.

    The subject devices vary in the jaw configurations and curvature to meet the surgeon's needs and preferences, based on individual, anatomical variations of the patients. The products are sold nonsterile and can be reused (cleaned and sterilized) according to the instructions for use.

    AI/ML Overview

    This document does not describe "acceptance criteria" for a device in the context of clinical performance metrics (e.g., accuracy, sensitivity, specificity). Instead, it focuses on demonstrating "substantial equivalence" of a new medical device to a predicate device for FDA clearance. The "performance criteria" described are engineering and biocompatibility tests rather than clinical study endpoints.

    Therefore, many of the requested sections (e.g., sample size for test set, number of experts, MRMC studies, standalone performance) are not applicable or cannot be extracted from this document, as it pertains to a 510(k) premarket notification for a manual surgical instrument, not an AI/software device that would typically have such performance criteria.

    Here's the information that can be extracted or inferred:

    1. A table of acceptance criteria and the reported device performance

    TestAcceptance Criteria (Implicit)Reported Device Performance
    BiocompatibilityMeet G.P. requirements for patient safetyCytotoxicity testing per ISO 10993-5:2009 was performed. Biocompatibility information from cleared devices (K160104, K150468, and K120492) was leveraged. (Implicitly, the device passed these tests to support substantial equivalence).
    Cleaning & Sterilization ValidationMeet requirements for reprocessingCleaning and sterilization validation study performed to meet requirements in: - FDA Guidance "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling" (March 17, 2015) - AAMI TIR12:2010. (Implicitly, the device's reprocessing instructions were validated and met these requirements).
    Boiling Water Test for CorrosionNo evidence of corrosionPerformed per ASTM F1089-10. (Implicitly, the device met the criteria and showed no unacceptable corrosion).
    Elasticity TestMaintain functional elasticityPerformed per DIN 96198-3. (Implicitly, the device met the criteria for elasticity).
    Functional EquivalencePerform as intended for its useThe device is demonstrated to have the "same intended use and fundamental technological characteristics" as the predicate device. Differences in material (titanium vs. stainless steel) or specific dimensions were deemed not to raise different questions of safety and effectiveness as titanium is widely used and both devices offer ranges of configurations.

    2. Sample sized used for the test set and the data provenance

    • Sample Size: Not specified for any of the non-clinical tests. This typically refers to the number of devices or material samples tested.
    • Data Provenance: The tests are described as "non-clinical performance testing" and involved laboratory studies, not patient data. Therefore, country of origin or retrospective/prospective is not applicable in the usual sense for clinical studies. The tests were performed in support of the device's substantial equivalence.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable. This device is a manual surgical instrument. "Ground truth" in the context of expert consensus is not relevant here; the ground truth is established by standardized material and performance testing.

    4. Adjudication method for the test set

    • Not applicable. This is for non-clinical, objective performance tests, not subjective interpretations requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is not an AI or software device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is not an algorithm or AI device.

    7. The type of ground truth used

    • For Biocompatibility: The "ground truth" is adherence to ISO standards and established biocompatibility principles.
    • For Cleaning & Sterilization Validation: The "ground truth" is the ability to render the device clean and sterile according to FDA guidance and AAMI standards.
    • For Bench Performance Tests (Corrosion, Elasticity): The "ground truth" is the physical properties of the materials and device and its performance against established engineering standards (ASTM, DIN).

    8. The sample size for the training set

    • Not applicable. There is no training set mentioned, as this is not an AI/ML device.

    9. How the ground truth for the training set was established

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K163524

    Validate with FDA (Live)

    Device Name
    Vessel Dilator
    Date Cleared
    2017-06-29

    (196 days)

    Product Code
    Regulation Number
    870.4475
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fetzer Medical vessel dilators are devices used to enlarge or calibrate a vessel during cardiovascular surgery.

    Device Description

    Fetzer Medical Vessel Dilators are reusable manual surgical instruments. They are rod-like devices with a long, slender, solid shaft and a distal tip of various shapes and sizes. They are sold unsterile and can be reused (cleaned and sterilized) according the instructions for use.

    There are several types of dilators made from stainless steel or aluminum:

    • Cooley Coronary Dilators
    • Cooley Vessel Dilators
    • Garrett Vascular Dilators
    • Hiebert Vascular Dilators
    • DeBakey Vascular Dilators
    • Sarns-Style Dilators
    • Myocardial Dilators

    These dilators are available in a range of different length and diameters.

    The handle of the DeBakey Vascular Dilator, the Cooley Vessel Dilator, the Cooley Coronary Dilator, the Hiebert Vascular Dilator and the Garrett Vascular Dilator is textured with one site flat. The handles of the Sarns and Myocardial Dilators are smooth. The Myocardial Dilator has a flat distal end for holding it and the Sarns Dilator has a hollow handle at the distal end.

    Apart from the Hiebert Vascular Dilator and the Myocardial Dilator, the handle is connected to a thin shaft. The diameter of the shaft of the different dilator types varies. The shaft of the Cooley Coronary Dilator, the Cooley Vessel Dilator, The Garrett Vascular Dilator and the DeBakey Vascular Dilator can be thin so that the shaft is flexible. The handle of the Myocardial Dilator has no shaft while the Hiebert Vascular Dilator is double-ended; it possesses two flexible shafts, with the handle located in between the two thin shafts.

    The tips of all dilators have the aim to enlarge a vessel. The design of the Sarns and Myocardial Dilator is similar to an elephant's tusk. The diameter increases steadily, moving towards the handle's direction. The tips of the other Dilators are olive-shaped. The Cooley Coronary Dilator possesses no formed tip. For all dilators, the penetrating diameter corresponds to the diameter of the shaft. The Hiebert Vascular Dilators working elements are each malleable wires with an olive tip in different diameters.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device: Fetzer Medical Vessel Dilators. This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, rather than proving the safety and effectiveness from scratch through extensive clinical trials for new medical technologies like AI/ML devices. As such, the information you're asking for, particularly regarding acceptance criteria, study design for proving performance, and expert-driven ground truth establishment (which is typical for AI/ML diagnostic devices), is not present in this document.

    The document focuses on non-clinical performance testing (reprocessing, sterilization, biocompatibility, mechanical tests) to demonstrate that the Fetzer Medical Vessel Dilators are as safe and effective as existing, legally marketed predicate devices.

    Here's a breakdown of why your specific questions cannot be answered by this document and what is provided:

    Why the document does not contain the requested information (related to AI/ML device performance studies):

    • Device Type: The Fetzer Medical Vessel Dilators are manual surgical instruments, specifically vessel dilators. They are not an AI/ML diagnostic or therapeutic device.
    • Regulatory Pathway: A 510(k) submission primarily relies on demonstrating substantial equivalence to a predicate device. This involves comparing indications for use, technological characteristics (design, materials, performance), and demonstrating that any differences do not raise new questions of safety or effectiveness. It does not typically involve defining acceptance criteria for diagnostic performance (e.g., sensitivity, specificity, AUC) or conducting large-scale clinical studies with human readers, ground truth consensus, or advanced statistical analyses like MRMC studies, which are common for AI/ML-driven diagnostic devices.
    • Performance Data Scope: The "performance data" mentioned in this document refers to physical testing of the device (e.g., cleanability, sterilizability, mechanical strength), not diagnostic accuracy or clinical outcomes in patients.

    What the document does provide regarding "Testing":

    The document details performance testing conducted to show the device's characteristics are similar to the predicate devices and that it meets relevant standards for a surgical instrument.

    1. A table of acceptance criteria and the reported device performance:

    • Acceptance Criteria (Implicit): For this type of device, the "acceptance criteria" are broad and relate to meeting industry standards and demonstrating comparable performance to the predicate device in specific functional aspects.

      • Automatic Reprocessing Validation according to ANSI/AAMI ST81: The implicit acceptance criterion is that "contamination...could be removed...by the given cleaning and disinfection instructions."
      • Sterilization testing: The implicit acceptance criterion is that "the device will be adequately sterilized under the validated sterilization parameters" to achieve a SAL (Sterility Assurance Level) of 10^-6.
      • Biocompatibility testing (ISO 10993-5): The implicit acceptance criterion is that materials are "non cytotoxic for the indicated contact duration."
      • Mechanical tests: The implicit acceptance criterion is that devices "were able to withstand the required pulling force without deformation or failure."
    • Reported Device Performance:

    TestReported Device Performance
    Automatic Reprocessing Validation"This validation provides evidence that contamination of the 'Vessel Dilators' could be removed in health care facilities by the given cleaning and disinfection instructions."
    Sterilization testing"A reduction of test bacteria was observed and provided assurance that the device will be adequately sterilized under the validated sterilization parameters."
    Biocompatibility testing (Cytotoxicity)"All patient contacting materials were from the same material as the equivalent predicate devices. They were considered non cytotoxic for the indicated contact duration."
    Mechanical tests"All devices were able to withstand the required pulling force without deformation or failure."

    2. Sample size used for the test set and the data provenance:

    • Test Sets: The document refers to "worst case devices" for reprocessing validation and "test specimens" for sterilization testing. The exact number of units or biological samples tested is not specified.
    • Data Provenance: This is not applicable in the context of clinical patient data. The tests are laboratory-based. No country of origin for clinical data is mentioned as no clinical studies were performed. The tests are retrospective in the sense they were done before submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications:

    • Not applicable. The "ground truth" for these tests relates to measurable physical/biological parameters (e.g., residual protein levels, bacterial reduction, material cytotoxicity, mechanical force applied), not expert interpretation of medical images or patient outcomes.

    4. Adjudication method for the test set:

    • Not applicable. This refers to consensus building among experts for subjective interpretations, which is not relevant for the objective laboratory tests performed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done:

    • No. This type of study is for evaluating observer performance, typically in diagnostic imaging. It was not conducted for this surgical instrument.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. This relates to AI/ML algorithm performance. The device is a manual surgical tool.

    7. The type of ground truth used:

    • Laboratory-based measurements against established standards:
      • Reprocessing: Measured contamination levels (red blood cells, residual protein, organisms, total organic carbon) against AAMI TIR 30 guidelines.
      • Sterilization: Reduction of bioindicators/challenge suspension to achieve SAL 10^-6.
      • Biocompatibility: Cytotoxicity results per ISO 10993-5.
      • Mechanical: Resistance to pulling forces.

    8. The sample size for the training set:

    • Not applicable. This is typically for machine learning models. No training set for an algorithm was used.

    9. How the ground truth for the training set was established:

    • Not applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K161004

    Validate with FDA (Live)

    Date Cleared
    2016-09-09

    (151 days)

    Product Code
    Regulation Number
    882.4800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fetzer Medical Self-Retaining Retractors are intended to hold the edges of a wound open during spinal surgery.

    Device Description

    Fetzer Medical Self-retaining Retractors are reusable manual instruments made from stainless steel, aluminum or titanium. They are sold unsterile and can be reprocessed (cleaned and sterilized) according the instructions for use. Devices are available with the following features: frame type retractors, Ring handle retractors, MIS Port retractor.

    AI/ML Overview

    The document describes the submission for a medical device called "Fetzer Medical Self-Retaining Retractors" for 510(k) clearance. The device is a self-retaining retractor for neurosurgery.

    Here's an analysis of the acceptance criteria and the study used to prove the device meets these criteria:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test (Acceptance Criteria Category)Test Method SummaryReported Device Performance (Results)
    Reprocessing ValidationVerification that devices can be cleaned with provided cleaning steps as per IFU. Contamination of accessible, interior, and exterior surfaces (worst-case) with visible contamination (red blood cells), residual protein, and organisms. Effectiveness evaluated by comparing contamination between control and test instruments. (Acc. ANSI/AAMI ST81 and AAMI TIR 30).This validation provides evidence that viable microbiological contamination as well as a soil contamination of the "Self-retaining retractors" are removed by the given cleaning and disinfection instructions.
    Sterilization TestingVerification that devices can be sterilized with the provided sterilization procedure described in the IFU. Test specimens contaminated with bioindicators or challenge suspension and tested for sterility after the sterilization process. Part-cycle sterilization performed to assure a SAL of 10^-6.A reduction of test bacteria was observed and assured showing that contaminated test specimens are free of viable/augmentable bacteria after sterilization.
    Biocompatibility TestingVerification of biocompatibility of all patient-contacting materials according to ISO 10993-1. Devices considered "tissue/bone contacting" with contact duration < 24 hours. Cytotoxic tests (ISO 10993-5) and sensitization/irritation tests (ISO 10993-10) performed for different materials and manufacturing lines.All tested devices were considered non cytotoxic and they did not cause any skin irritation reaction for the indicated contact duration.
    Mechanical TestsTesting for typical forces to hold a wound open during spinal surgery. Testing load was double to triple the maximum load assumed during surgery.All devices were able to withstand the required test load without deformation or failure.

    2. Sample Size for Test Set and Data Provenance

    The document does not explicitly state a numerical sample size for the "test set" in the context of device performance testing. Instead, it refers to "representative worst-case instruments" for reprocessing validation and "test specimens" for sterilization. It also performs tests on "all patient contacting materials" for biocompatibility and "all devices" (implied across the product line for mechanical tests).

    The data provenance is not specified. It is implied to be internal testing conducted by Fetzer Medical GmbH & Co. KG. No country of origin for the data is explicitly mentioned, nor is it stated whether the data is retrospective or prospective, though it's likely prospective laboratory testing.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable to this type of device and study. The ground truth for performance of a mechanical surgical retractor is established through objective engineering and biological testing (e.g., measuring forces, observing bacterial reduction, assessing cytotoxicity), not through expert consensus on interpretation.

    4. Adjudication Method for the Test Set

    This information is not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies or image-based diagnostic device evaluations where subjective human interpretation is involved. For the engineering and biological tests described, the 'results' are objective measurements or observations.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and effect size.

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging devices where human readers interpret medical images, and the AI's impact on their performance is being evaluated. The Fetzer Medical Self-Retaining Retractors are mechanical surgical instruments.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done.

    No, a standalone (algorithm only) performance study was not done. This concept is relevant for AI/ML-driven software as a medical device (SaMD) or diagnostic algorithms. The Fetzer Medical device is a conventional, mechanical surgical tool.

    7. The Type of Ground Truth Used

    The ground truth used for these tests is based on established scientific and engineering principles, and regulatory standards:

    • For reprocessing and sterilization: Reduction of microbial contamination to acceptable, defined levels (e.g., SAL of 10^-6), and removal of visual and chemical contaminants.
    • For biocompatibility: Absence of cytotoxic effects, sensitization, or irritation as defined by ISO 10993 standards.
    • For mechanical tests: Ability to withstand (without deformation or failure) loads at multiples of assumed surgical forces, meeting predetermined engineering specifications.

    8. The Sample Size for the Training Set

    This information is not applicable. "Training set" refers to data used to train machine learning algorithms. The Fetzer Medical device is a mechanical instrument, not an AI/ML device, and therefore does not have a training set.

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable as there is no training set for this mechanical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1