510(k) Data Aggregation

    K Number
    K041280

    Validate with FDA (Live)

    Date Cleared
    2004-06-24

    (42 days)

    Product Code
    Regulation Number
    880.5430
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This needle-free fluid injector system is indicated for the delivery of vaccines or medications to the subcutaneous tissue by penetrating the skin under high pressure. The device is to be used by a health care professional only.

    Device Description

    The Human Subcutaneous Injector system consists of a hand-held injector unit to which a multiple-dose vial of vaccine or medication can be attached. The unit is spring-powered and is connected to its hydraulic fluid power source, which is operated by a foot pump.

    AI/ML Overview

    The provided text describes a 510(k) summary for the Human Subcutaneous Injector system (HSI 500) and its substantial equivalence to predicate devices, but it does not include detailed studies or specific acceptance criteria with reported device performance numbers. The summary states:

    "The performance characteristics are substantially equivalent to the Bi-3M according to test protocols conducted for dosage accuracy, penetration, force, focus, sound, and stream quality. Safety tests performed for contamination were successful."

    This indicates that tests were performed, but the actual acceptance criteria (e.g., "dosage accuracy must be within X%") and the specific results (e.g., "dosage accuracy was Y%") are not provided in the document.

    Therefore, I cannot fully complete the requested table and many of the study-related questions based only on the provided text.

    However, I can extract what is implied and explicitly stated:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Acceptance Criteria (Implied/General)Reported Device Performance (Implied/General)
    Dosage AccuracySubstantially equivalent to predicate device (Bi-3M)Performance characteristics are substantially equivalent to Bi-3M
    PenetrationSubstantially equivalent to predicate device (Bi-3M)Performance characteristics are substantially equivalent to Bi-3M
    ForceSubstantially equivalent to predicate device (Bi-3M)Performance characteristics are substantially equivalent to Bi-3M
    FocusSubstantially equivalent to predicate device (Bi-3M)Performance characteristics are substantially equivalent to Bi-3M
    SoundSubstantially equivalent to predicate device (Bi-3M)Performance characteristics are substantially equivalent to Bi-3M
    Stream QualitySubstantially equivalent to predicate device (Bi-3M)Performance characteristics are substantially equivalent to Bi-3M
    ContaminationSuccessful, meeting safety standardsSafety tests for contamination were successful

    2. Sample size used for the test set and the data provenance

    • Sample size for test set: Not specified in the provided text.
    • Data provenance: Not specified, but implied to be from tests conducted by the submitter (Regulatory/Clinical Consultants, Inc. for Felton International). The country of origin and whether it was retrospective or prospective is not mentioned.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not applicable/Not specified. The assessment appears to be based on technical performance metrics against a predicate device, not expert human interpretation.

    4. Adjudication method for the test set

    • Not applicable/Not specified. Adjudication methods like 2+1 or 3+1 typically refer to expert consensus in image analysis or similar diagnostic contexts. This device's testing relates to physical performance.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is not an AI-assisted diagnostic device, nor does the document describe a human-reader study.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This device is a fluid injector system, not an algorithm. The testing described relates to the device's physical performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The "ground truth" for demonstrating substantial equivalence was the performance characteristics of the predicate device, Felton International's Bi-3M. The "test protocols" conducted would have measured the HSI 500's performance against these established characteristics for dosage accuracy, penetration, force, focus, sound, and stream quality.

    8. The sample size for the training set

    • Not applicable. This is a physical medical device, not a machine learning model. There is no concept of a "training set" in this context.

    9. How the ground truth for the training set was established

    • Not applicable, as there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K013256

    Validate with FDA (Live)

    Date Cleared
    2001-12-07

    (67 days)

    Product Code
    Regulation Number
    880.5430
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    This needle-free fluid injector system is indicated for the delivery of vaccines or medications to the subcutaneous tissue by penetrating the skin under high pressure. This device is to be used by a health care professional only.

    Device Description

    The Bi-3M Needle-Free Injector System consists of a hand-held injector unit to which a multiple-dose vial of vaccine or medication can be attached. The unit is spring-powered and is connected to its hydraulic fluid power source, which is operated by a foot pump.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Bi-3M Needle-Free Injector System, structured to answer your questions:


    Acceptance Criteria and Study for the Bi-3M Needle-Free Injector System

    The provided document describes performance and safety tests conducted to demonstrate substantial equivalence of the Bi-3M Needle-Free Injector System to predicate devices, particularly the Ped-O-Jet. The primary focus of the studies was on dosage accuracy and prevention of cross-contamination.


    1. Table of Acceptance Criteria and Reported Device Performance

    Note: The document does not explicitly state pre-defined "acceptance criteria" as pass/fail thresholds for the performance tests. Instead, it compares the Bi-3M's performance directly to the predicate device (Ped-O-Jet) to demonstrate substantial equivalence. For the cross-contamination study, it reports a specific clean sample rate.

    Performance ParameterAcceptance Criteria (Implied by comparison)Reported Bi-3M Performance
    Dosage AccuracyNo significant difference compared to Ped-O-JetNo significant difference between Bi-3M and Ped-O-Jet
    PenetrationSimilar to Ped-O-JetSimilar to Ped-O-Jet
    FocusSimilar to Ped-O-JetSimilar to Ped-O-Jet
    ForceSimilar to Ped-O-JetSimilar to Ped-O-Jet
    Stream QualitySimilar to Ped-O-JetSimilar to Ped-O-Jet
    Environmental StabilityConsistent functionality at temperature extremes and after free fallIndicated consistent functionality
    Noise LevelBelow OSHA exposure limitsBelow OSHA exposure limits
    Cross-Contamination PreventionDesigned to significantly reduce cross-contamination99.8% clean sample rate (vs. 54% for Ped-O-Jet in high-contamination model)

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for performance tests (Dosage Accuracy, Penetration, Focus, Force, Stream Quality): The document does not specify the exact sample sizes used for these tests. It generally states a "battery of performance testing was done."
    • Sample Size for Cross-Contamination Safety Test: The document does not explicitly state the number of injections or replicates performed in the cross-contamination study, but it refers to "the one Bi-3M sample for which fluorescein was reported in the downstream collection," implying multiple samples were tested.
    • Data Provenance: The document does not explicitly state the country of origin. The organization (Felton International, Inc.) is based in Lenexa, Kansas, USA, so it's likely the testing was conducted in the USA or supervised by this entity. The study appears to be prospective as it was designed to test the Bi-3M device.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    • The document does not mention the use of experts to establish ground truth for any of the reported tests. The performance and safety tests appear to be laboratory-based, objective measurements (e.g., measuring fluorescein, mechanical properties).

    4. Adjudication Method for the Test Set

    • None. Since expert-defined ground truth or human interpretation is not indicated, an adjudication method is not applicable.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No. The document does not describe an MRMC comparative effectiveness study where human readers' performance with and without AI assistance is evaluated. This device is a physical medical device (injector), not an AI diagnostic tool.

    6. Standalone Performance

    • Yes. The document details standalone performance tests for the Bi-3M device. These tests evaluate the device's inherent mechanical function (dosage accuracy, penetration, etc.) and its ability to prevent cross-contamination, independent of human interaction in the outcome assessment (beyond operating the device). The results for dosage accuracy, penetration, focus, force, stream quality, environmental testing, noise level, and the 99.8% clean sample rate in the cross-contamination study represent its standalone performance.

    7. Type of Ground Truth Used

    • Objective Laboratory Measurements: The ground truth for the performance tests (dosage accuracy, penetration, focus, force, stream quality, environmental, noise) was established through direct, quantifiable physical or mechanical measurements comparing the Bi-3M to the Ped-O-Jet.
    • Fluorescein Detection: For the cross-contamination study, the ground truth was the presence or absence of fluorescein (a chemical marker) in the downstream collection vessel, indicating contamination. This is an objective, laboratory-based detection method.

    8. Sample Size for the Training Set

    • Not applicable. This document describes the testing and 510(k) submission for a physical medical device. It does not involve a machine learning algorithm or an "AI" device that would typically have a training set in the computational sense. The "training" of the device would refer to its engineering and design, not an ML model.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As noted above, there is no "training set" in the context of machine learning for this physical device. The device's design and manufacturing are based on engineering principles and previous iterations (e.g., the Bi-3 injector).