510(k) Data Aggregation

    K Number
    K252722

    Validate with FDA (Live)

    Device Name
    Biopsy Forceps
    Date Cleared
    2025-09-30

    (33 days)

    Product Code
    Regulation Number
    870.4075
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FEHLING Biopsy Forceps are used to obtain endomyocardial biopsy specimens from the right and left ventricle via percutaneous arterial or venous approach.

    Device Description

    The Biopsy Forceps are designed to allow percutaneous access to the right or left ventricles of the heart to obtain diagnostic tissue samples. The forceps consist of three main components:

    1. an actuating handle ergonomically designed for comfortable use,
    2. a flexible shaft,
    3. and surgical stainless steel cutting jaws.

    At the distal end of the forceps is a pair of stainless steel jaws used to obtain the tissue samples. At the proximately end of the forceps is the actuation handle used to activate the jaws and steer the device.

    The product is provided single for use and sterile.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K170726

    Validate with FDA (Live)

    Device Name
    Biopsy Forceps
    Date Cleared
    2017-06-02

    (85 days)

    Product Code
    Regulation Number
    870.4075
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    FEHLING biopsy forceps are used to obtain endomyocardial biopsy specimens from the right and left ventricle via percutaneous arterial or venous approach.

    Device Description

    The Biopsy Forceps are designed to allow percutaneous access to the right or left ventricles of the heart to obtain diagnostic tissue samples. The forceps consist of three main components:

    • an actuating handle ergonomically designed for comfortable use, 1.
      1. a flexible shaft,
    • and surgical stainless steel cutting jaws. 3.
      At the distal end of the forceps is a pair of stainless steel jaws used to obtain the tissue samples. At the proximately end of the forceps is the actuation handle used to activate the jaws and steer the device.
      The product is provided single for use and sterile.
    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device called "Biopsy Forceps." It describes the device, its intended use, and its substantial equivalence to previously cleared predicate devices. The document includes non-clinical performance data to support its claims but does not include details about acceptance criteria or a study proving the device meets those criteria in the way typically associated with AI/ML devices (e.g., performance metrics like accuracy, sensitivity, specificity).

    Instead, the document focuses on demonstrating physical and biological compatibility and functionality for a mechanical medical device. Therefore, many of the requested items (e.g., sample size for AI test sets, number of experts for ground truth, MRMC study, standalone performance) are not applicable or cannot be extracted from this type of regulatory submission.

    Here's an attempt to answer the questions based only on the provided text, recognizing the limitations:


    1. A table of acceptance criteria and the reported device performance

    The document states that "All samples passed the test and met the acceptance criteria" for several non-clinical tests, but it does not explicitly define those acceptance criteria. The reported "performance" is simply that the device met these (unspecified) criteria.

    Test TypeAcceptance Criteria (Not Explicitly Defined)Reported Device Performance
    Dimensional Verification TestsNot explicitly definedAll samples passed the test
    Tensile TestsNot explicitly definedAll samples passed the test
    Simulation Test (Insertion & Tissue Samples)Not explicitly definedAll samples passed the test
    Shelf Life Testing & Transport SimulationNot explicitly definedAll samples passed the test
    Biocompatibility (ISO 10993-1)Not explicitly definedMeets ISO 10993-1 requirements

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document mentions "All samples" for the dimensional, tensile, and shelf life tests, but it does not specify the exact number of samples.

    For the Simulation Test - Insertion and Taking of Tissue Samples:

    • Sample Size: "The process was repeated 10 times per device." The number of devices used is not specified.
    • Data Provenance: No country of origin or retrospective/prospective nature is specified for the non-clinical tests. The tissue samples were "fresh porcine heart specimens."

    For Biocompatibility: "The full strength EMEM10 test article showed no cytotoxic potential to L-929 mouse fibroblast cells." No specific sample size (number of cells or test articles) is given, beyond indicating "the test article."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    Not applicable. This is a non-clinical, mechanical/biological performance study for a physical device, not an AI/ML study requiring expert consensus for ground truth.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. This is a non-clinical, mechanical/biological performance study for a physical device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a non-clinical, mechanical/biological performance study for a physical device, not an AI/ML study.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a non-clinical, mechanical/biological performance study for a physical device, not an AI/ML algorithm. The performance tests ("Dimensional Verification," "Tensile Tests," "Simulation Test - Insertion and Taking of Tissue Samples," "Shelf Life Testing") are effectively "standalone" tests of the device itself.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical performance tests:

    • Dimensional Verification: Comparison to engineering specifications/manufacturing tolerances (implied).
    • Tensile Tests: Comparison to mechanical strength specifications (implied).
    • Simulation Test: Successful introduction and harvesting of tissue samples from "fresh porcine heart specimens" (functional verification).
    • Shelf Life Testing: Maintenance of specified performance after accelerated aging/transport simulation.
    • Biocompatibility: Results from standardized cytotoxicity, chemical characterization, and previous toxicological data compared against ISO 10993-1 requirements.

    8. The sample size for the training set

    Not applicable. This is a non-clinical, mechanical/biological performance study for a physical device, not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. This is a non-clinical, mechanical/biological performance study for a physical device, not an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K153243

    Validate with FDA (Live)

    Device Name
    Fehling-Punches
    Date Cleared
    2016-08-04

    (269 days)

    Product Code
    Regulation Number
    882.4840
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fehling rongeurs (bone punches) are manually operated instruments indicated for cutting or biting bone during surgery involving the skull or spinal column.

    Device Description

    Fehling-Punches are reusable stainless steel instruments that are coated with CERAMO® that are sterilizable and packaged non-sterile. The instruments are offered in various sizes and with different features to accommodate the variations of patient anatomy and access. Instruments are available with the following features: 0.8-8 mm bite sizes; 9 – 19 mm jaw openings; 40° and 90° up/down cutting angles; up/down curved; reqular and thin profile footplates; standard and ejector tips; 110 - 400 mm shaft lengths; and various handle and shaft styles, detachable and non-detachable. Fehling-Punches are distributed under the following brand names: CERAMO® CONCEPT X, CERAMO® CONCEPT CLASSIC, CERAMO® CONCEPT APART, CERAMO® GENTLE, CERAMO® GENTLE "SINGER", CERAMO® TRADITION X, CERAMO® FIST. CERAMO® EJECTOR, CERAMO® SUBLAMINAR, CERAMO® APART, CERAMO® KERRISION. The instruments are made of stainless steel according to ASTM F899 and are coated with CERAMO®. CERAMO®is an AlTiN coating. A reusable screwdriver (item code TXW-9X) is offered with the Fehling Punches for the assembly and disassembly.

    AI/ML Overview

    The acceptance criteria table and device performance based on the provided document are as follows. Please note that the document does not describe a study involving an AI/algorithmic device. The device described is a manual surgical instrument (bone punches). Therefore, many of the requested categories related to AI/MRMC studies, ground truth methodology, and training/test set details are not applicable as they pertain to AI algorithms, not manual instruments.


    1. Table of Acceptance Criteria and Reported Device Performance

    Test/CharacteristicAcceptance Criteria (Implied/Stated)Reported Device Performance
    Biocompatibility
    CytotoxicityNo cytotoxic potentialNo cytotoxic potential to L-929 mouse fibroblast cells.
    SensitizationNot considered a sensitizer (per ISO 10993-10)Not considered a sensitizer in the guinea pig maximization model.
    HemolysisNon-hemolyticDirect contact and extract were non-hemolytic (mean hemolytic index < 2%).
    Acute Systemic ToxicityNo significant systemic toxicity or mortality (per ISO 10993-11)No evidence of significant systemic toxicity or mortality.
    Intracutaneous StudyMet requirements of Intracutaneous injection test (per ISO 10993-10)Met requirements of the Intracutaneous injection test.
    PyrogenicityNon-pyrogenic (for devices that have undergone validated cleaning and sterilization)Non-pyrogenic.
    LAL-Test (Bacterial Endotoxin)Non-pyrogenic (for devices that have undergone validated cleaning and sterilization)Non-pyrogenic.
    Performance Testing
    Dynamic (Fatigue Resistance)Maintained integrity after 12,500 cutting simulationsConfirmed integrity after 12,500 cutting simulations.
    Static (Cutting Force)Comparable mean force required to operate to predicate deviceMean force required comparable to the predicate device.
    Coating Thickness VerificationCoating thickness present on all relevant surface areas and within an advantageous rangeCoating thickness verified to be present and within an advantageous range.
    Dimensional VerificationConformed to actual specificationsDimensional and functional specifications verified.
    Performance TestsPassed defined parameters for worst-case and extended useDevices passed tests (one deviation noted but attributed to misplacement).
    Processing Validation
    Automated Cleaning ValidationAutomated cleaning validated (per AAMI TIR 30)Automated cleaning was validated.
    Sterilization ValidationSterilization process validated (per ISO 17665, AAMI TIR 30)Sterilization process was validated (full cycle 132°C and 4 min).

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes non-clinical laboratory testing of physical device properties and biocompatibility, not analysis of a "test set" of data in the context of an AI/algorithmic device. Therefore, typical metrics for data-driven test sets (e.g., country of origin, retrospective/prospective) are not applicable.

    • Biocompatibility Tests: Sample sizes for each specific biocompatibility test (cytotoxicity, sensitization, etc.) are not explicitly stated within the summary, but they were conducted on the "Fehling-Punches" device.
    • Performance Tests: The "dynamic test" involved 12,500 cutting simulations, implying a test set for fatigue evaluation. Other performance tests (static, coating, dimensional) were conducted on the device itself.
    • Data Provenance: The tests were conducted by the manufacturer or their contracted labs as part of the 510(k) submission process.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not applicable. The device is a manual surgical instrument, and its performance was evaluated through laboratory testing and engineering analysis against established standards and predicate devices, not by human experts establishing "ground truth" in the context of an AI algorithm's output.

    4. Adjudication Method for the Test Set

    This is not applicable. There was no "test set" requiring adjudication by multiple experts in the context of an AI device. Performance was determined by objective measurements and standardized testing protocols.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This is not applicable. The device is a manual surgical instrument and does not involve AI assistance or human readers.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    This is not applicable. The device is a physical, manual surgical instrument, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for this device's evaluation was based on:

    • Established industry standards: E.g., ISO 10993 for biocompatibility, AAMI TIR 30 and ISO 17665 for cleaning and sterilization.
    • Comparison to legally marketed predicate devices: Establishing substantial equivalence by demonstrating similar performance characteristics (e.g., cutting force, fatigue resistance) and safety profiles.
    • Dimensional and functional specifications: Internal engineering and design specifications for the device.

    8. The Sample Size for the Training Set

    This is not applicable. The device is a manually operated surgical instrument, not an AI algorithm requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable. There was no training set or AI algorithm involved.