Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K212741

    Validate with FDA (Live)

    Device Name
    EZYPOR
    Date Cleared
    2022-06-24

    (298 days)

    Product Code
    Regulation Number
    886.3320
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EZYPOR® orbital implants are designed to fill the orbital cavity following enucleation, evisceration or during secondary implantation.

    Device Description

    EZYPOR® orbital implants are high density polyethylene (UHMWPE) implants designed to fill the orbital cavity following enucleation, evisceration or during secondary implantation procedures. The polyethylene material has an open porosity structure of 40 to 60%. The implants are available in six diameter sizes, 12, 14, 16, 18, 20 and 22 mm. EZYPOR® orbital implants are supplied sterile and are sterilized by ethylene oxide.

    AI/ML Overview

    This is a 510(k) summary for the EZYPOR® orbital implant, a medical device. This document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a standalone study proving the device meets specific performance criteria through clinical trials. Therefore, much of the requested information about acceptance criteria for device performance in a clinical context, sample sizes for test sets, expert ground truth establishment, MRMC studies, and training set details are not applicable or available within this type of submission.

    The document discusses non-clinical bench testing to ensure the device meets predefined specifications for its physical characteristics and biocompatibility.

    Here's a breakdown of the information that can be extracted or inferred:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria CategorySpecific Criteria (Implicit for S.E. to Predicate)Reported Device Performance
    Non-Clinical Bench Testing
    Diameter and Tunnel DimensionsWithin specified toleranceMet acceptance criteria
    Implant MassWithin specified toleranceMet acceptance criteria
    SphericityWithin specified toleranceMet acceptance criteria
    Open Porosity40-60%Met acceptance criteria
    Resistance to CompressionSufficient to withstand expected forcesMet acceptance criteria
    Resistance to TractionSufficient to withstand expected forcesMet acceptance criteria
    Biocompatibility
    Chemical CharacterizationNo unacceptable leachablesMet acceptance criteria
    CytotoxicityNon-cytotoxicMet acceptance criteria
    SensitizationNon-sensitizingMet acceptance criteria
    IrritationNon-irritatingMet acceptance criteria
    SterilizationSterile (SAL 10^-6)Sterilized by EO
    EndotoxinMeets endotoxin limitsPerformed and met criteria
    Shelf Life5 years, maintained functional performance and package integrityEstablished at 5 years

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified for non-clinical bench testing. Standard engineering and material testing practices would imply a sufficient number of units were tested to ensure consistency and meet statistical requirements for demonstrating product specifications.
    • Data Provenance: Not explicitly stated, but assumed to be from internal testing conducted by FCI S.A.S. (France Chirurgie Instrumentation SAS). The testing is non-clinical/bench-level, so concepts like "country of origin of the data" in a clinical sense or "retrospective/prospective" studies do not apply.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. Ground truth for non-clinical bench testing is established by engineering specifications, material standards, and validated test methods.

    4. Adjudication method for the test set:

    • Not applicable. Adjudication methods are typically for clinical studies where subjective interpretations might be involved. For bench testing, results are objectively measured against predefined specifications.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This device is an orbital implant, not an AI-powered diagnostic tool requiring human reader studies.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used:

    • For non-clinical bench tests (diameter, mass, sphericity, porosity, resistance, biocompatibility, sterilization, shelf life), the "ground truth" is defined by pre-established engineering specifications, material standards (e.g., ISO standards), and validated test methodologies.

    8. The sample size for the training set:

    • Not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable. No training set is used for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K203569

    Validate with FDA (Live)

    Date Cleared
    2021-06-11

    (186 days)

    Product Code
    Regulation Number
    886.5700
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Gold and Platinum Tapered Weight Eyelid Implants are intended for the gravity-assisted treatment of protracted or permanent lagophthalmos, usually resulting from facial paralysis.

    Device Description

    The Gold and Platinum Tapered Weight Eyelid Implants are implantable devices for the gravityassisted treatment of lagophthalmos by the addition of weight to the upper eyelid. The device is made of either 99.99% purity gold or platinum. Platinum devices can be implanted in patients who are allergic to gold. The Gold and Platinum Tapered Weight Eyelid Implants come in two thicknesses (thin and normal profiles) and in seven different weights ranging from 0.6 g to 1.8 g (in 0.2 g increments). The Gold and Platinum Tapered Weight Eyelid Implants are each provided as a sterilized product.

    AI/ML Overview

    This document describes the premarket notification (510(k)) for Gold and Platinum Tapered Weight Eyelid Implants. It focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than providing a detailed clinical study for novel device acceptance criteria. Therefore, much of the requested information regarding clinical study design, ground truth, expert opinions, and sample sizes for diagnostic performance is not directly applicable or available in this document.

    However, based on the provided text, I can extract and infer information relevant to the non-clinical acceptance criteria and the studies performed to meet them.

    1. Table of Acceptance Criteria and Reported Device Performance

    Given that this is a 510(k) submission focused on substantial equivalence to an existing device, the "acceptance criteria" are primarily established by demonstrating that the new device shares fundamental technological characteristics and performance (or equivalent performance) with the predicate device, especially considering any differences. The key performance metrics evaluated are MRI safety, biocompatibility, and sterility/shelf-life.

    CategoryAcceptance Criteria (typically from established standards or predicate performance)Reported Device Performance (FCI Gold/Platinum Eyelid Implants)
    MRI SafetyDevice is "MR Conditional" under specified conditions (e.g., static field ≤ 3T, spatial gradient ≤ 4000 Gauss/cm, SAR ≤ 4 W/kg for 15 min).MR Conditional: meets these conditions.
    MRI-Related ForceNegligible deflection/torque in a 3T MR system. (e.g., 2° deflection or "0" on a 0-4 torque scale). Adherence to ASTM F2052-06.Gold: 2° deflection, Torque "0". Platinum: 2° deflection, Torque "0".
    MRI-Related HeatingMaximum temperature rise ≤ acceptable limits (e.g., usually <5°C for implants during 15 min scan). Adherence to ASTM F2182-11a.Gold (2.8g): Max temperature rise of +2.0°C after 15 min. Platinum (2.8g): Max temperature rise of +2.3°C after 15 min.
    MRI-Related ArtifactImage artifact extension to be reported. Adherence to ASTM F2119-07.Gold (2.8g): Artifact extends ~5mm from device with gradient echo, 3T MR. Platinum (2.8g): Artifact extends ~10mm from device with gradient echo, 3T MR.
    BiocompatibilityMeets requirements of ISO 10993 standards (e.g., no cytotoxicity, acceptable leachable levels, etc.).Established through review of existing data, chemical characterization (leachable test), and cytotoxicity testing per ISO 10993-1, -5, -18. (Results implied as favorable for clearance).
    SterilitySterility Assurance Level (SAL) of 10⁻⁶. Ethylene Oxide (EO) residuals within acceptable limits. Meets ISO 11135-1:2007. Bioburden specification.SAL 10⁻⁶ achieved via Ethylene Oxide sterilization. EO Residuals: EO 4.0µg/device, ECH 5.8 µg/device. Bioburden: < 100 CFU/device.
    Shelf-LifeFunctional performance maintained, and package integrity preserved over the claimed shelf life (e.g., 5 years) following accelerated aging. Adherence to ASTM standards for package integrity.5 years established via accelerated aging. Functional performance (dimensional, visual inspection) verified. Package integrity (visual, peel, dye, bubble leak) confirmed per ASTM F1929, F1886, F88/F88M, F2096.

    2. Sample Size for Test Set and Data Provenance

    • MRI Safety: The testing was conducted using the largest sizes available in the market (2.8g) for both Gold and Platinum Eyelid Weight Implants. While the specific number of individual implants tested for each condition (force, heating, artifact) is not explicitly stated, it implies at least one sample of each type and weight was used for the respective tests. The provenance of the data is from non-clinical laboratory testing specific to the device.
    • Biocompatibility: Not directly referenced as a "test set" in the traditional sense for diagnostic performance. It involved chemical characterization and cytotoxicity testing. The sample size for these tests (e.g., number of replicates, specimens) is not specified but would follow the requirements of the ISO 10993 standards. Data provenance is from laboratory testing.
    • Sterility and Shelf-Life: The testing involved validation protocols based on standards like ISO 11135-1 and ASTM standards for package integrity. The sample sizes for these validation activities are not explicitly detailed but are typically defined by the respective standards (e.g., for bioburden, sterility validation, accelerated aging studies). Data provenance is from laboratory testing (in-house or contract sterilizer).

    3. Number of Experts and Qualifications for Ground Truth

    This document does not pertain to the diagnostic performance of an AI/ML device or a device requiring expert interpretation for its primary function. Therefore, there were no "experts used to establish ground truth" in the context of clinical outcomes or diagnostic accuracy. The ground truth for the non-clinical tests is based on objective measurements and adherence to established engineering and material science standards and protocols.

    4. Adjudication Method

    Not applicable, as this is not a study requiring human adjudication of clinical or diagnostic outcomes.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable, as this is not a study assessing human reader performance.

    6. Standalone (Algorithm Only) Performance

    Not applicable, as this is a physical medical implant, not an AI/ML algorithm.

    7. Type of Ground Truth Used

    The ground truth for the non-clinical tests described is based on:

    • Objective Measurements: For MRI safety (deflection angles, temperature changes, artifact dimensions), sterility (SAL, residual levels), and shelf-life (dimensional stability, package integrity).
    • Established Standards and Protocols: Adherence to international standards (ISO, ASTM) for methods, acceptance criteria, and measurement techniques.
    • Material Characterization: For biocompatibility, the ground truth is based on the chemical composition of the materials and their known biological responses, validated by specific in-vitro tests (cytotoxicity, leachable tests).

    8. Sample Size for Training Set

    Not applicable, as this is not an AI/ML algorithm requiring a training set.

    9. How Ground Truth for Training Set Was Established

    Not applicable, as this is not an AI/ML algorithm requiring a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201892

    Validate with FDA (Live)

    Date Cleared
    2021-03-11

    (246 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RITLENG®+ products are indicated in the treatment of 12 months and older in cases of:

    • Canalicular pathologies (stenoses, obstructions, lacerations),

    • Congenital duct obstruction (Hasner valve stenosis),

    • Dacryocystorhinostomy (conventional or laser).

    Device Description

    The RITLENG®+ is a self-retaining nasal bicanalicular intubation for the lacrimal ducts. It consists of a silicone tube with two larger diameter portions, connected at each extremity with a PEEK thread guide. Each thread guide has a wide diameter part and a narrow diameter part. The silicone tube may be coated with polyvinylpyrrolidone (PVP) to improve its wettability. The central body of the intubation displays a marking intended for the intubation placement between the lacrimal puncta.

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification from the FDA for a medical device called RITLENG®+ and RITLENG®+ PVP. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than providing extensive clinical study data with acceptance criteria for device performance in the same way one might describe a novel diagnostic AI algorithm.

    Therefore, the information about acceptance criteria and a study proving the device meets them needs to be interpreted within the context of a 510(k) submission, where the "study" is primarily a non-clinical comparison and testing against established specifications and predicate device characteristics.

    Here's the breakdown based on the provided text:

    Description of Acceptance Criteria and the Study that Proves the Device Meets the Acceptance Criteria:

    The RITLENG®+ device is a bicanalicular lacrimal stent. For a 510(k) submission, the "acceptance criteria" are predominantly related to the device's technical specifications, safety, and functionality being comparable to a predicate device, as well as meeting relevant standards. The "study" refers to the non-clinical tests performed to demonstrate this equivalence and performance.

    1. Table of Acceptance Criteria and the Reported Device Performance:

    Acceptance Criteria CategorySpecific Criteria (Implicitly Derived)Reported Device Performance
    Intended UseIdentical to predicate device"RITLENG®+ is substantially equivalent to the RITLENG® Intubation as both devices have the identical intended use and intubation method, with similar indications for use statements."
    Material BiocompatibilityMeets ISO 10993 standards and comparable to predicate device material"The implant part of both devices is made from the identical grade silicone... The biocompatibility of the device is supported by cytotoxicity testing per ISO 10993-5, three sets of chemical characterization testing ... per ISO 10993-18, and toxicological risk assessment." Also supported by "biocompatibility information of the materials used in the subject device to the materials used in the cleared predicate."
    SterilizationAchieves SAL 10-6 via Ethylene Oxide (EO)"Both devices are packaged in a double Tyvek pouch and sterilized by Ethylene Oxide to a sterility level of SAL 10-6." "Ethylene oxide sterilization specifications... were performed to the applicable standards."
    Wettability (Optional Feature)Comparable to predicate device's coated option"The silicone tube of both devices may be coated with polyvinylpyrrolidone (PVP) to improve its wettability."
    Physical Design/FunctionalityMeets established specifications for new features (self-retaining) and overall function"RITLENG®+ has been designed and tested to the applicable standards. All nonclinical test results met the established specifications for the device." "The results from non-clinical testing demonstrate RITLENG®+ meets the established specifications for the device..."
    Packaging & Shelf-LifeMeets applicable standards"package integrity studies, and stability studies were performed to the applicable standards; and the test results support the shelf-life and storage conditions for the device."
    Manufacturing Quality ControlIn-process and final product controls assure manufacturing within specifications"In-process controls and final product quality controls, including finished product release testing and inspection, assure that RITLENG®+ is manufactured within specifications."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This document does not describe a "test set" in the context of clinical data for an AI/diagnostic device. The "tests" mentioned are non-clinical bench tests (e.g., cytotoxicity, chemical characterization, sterilization efficacy, material testing, package integrity, stability studies). Therefore, terms like "sample size," "data provenance," "country of origin," and "retrospective/prospective" are not applicable in the way they would be for a clinical study on human subjects or an AI product's performance. The "samples" for testing would be units of the device itself or its constituent materials. The provenance of this physical testing data is from the manufacturer (FCI SAS) and the testing laboratories they contracted.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable. The ground truth for non-clinical tests is based on scientific and engineering standards and validated test methodologies, not expert consensus in a medical diagnostic sense. For example, "ground truth" for cytotoxicity is determined by cell viability assays against established thresholds, not by a panel of experts.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This is not applicable as there is no "test set" requiring expert adjudication in the context of this 510(k) submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This is not applicable. This is a conventional medical device (lacrimal stent), not an AI-powered diagnostic tool. Therefore, an MRMC study comparing human readers with and without AI assistance was not performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This is not applicable. This is a conventional medical device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's evaluation is primarily based on established engineering and material science standards (e.g., ISO 10993 for biocompatibility), validated test methods, and comparison to the characteristics of the predicate device. For instance:

    • Biocompatibility: Results of cytotoxicity, chemical characterization tests against established safety thresholds.
    • Sterility: Absence of microorganisms to a sterility assurance level (SAL) of 10^-6, confirmed by validated sterilization processes.
    • Physical Properties: Measurement of dimensions, material strength, etc., against predefined specifications.
    • Intended Use: Direct comparison of the stated indications for use with the predicate device.

    8. The sample size for the training set

    This is not applicable. This is a conventional medical device, not an AI product that requires a "training set."

    9. How the ground truth for the training set was established

    This is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201606

    Validate with FDA (Live)

    Device Name
    LacriJet®
    Date Cleared
    2020-12-18

    (186 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    LACRIJET® from 30 to 50 mm are indicated in the treatment of congenital lacrimal duct obstructions (stenosis of the valve of Hasner) in patients 12 months and older.

    LACRIJET® with small size (15/20 mm) are indicated in the repair of canalicular lacerations.

    Device Description

    LacriJet® is a monocanalicular intubation device, with self-retaining punctal fixation for the treatment of monocanalicular lacerations and stenoses. The silicone stent acts as a conformer. The shape of the fixation head keeps the probe from migrating and from expulsing.

    AI/ML Overview

    The provided text is a 510(k) Pre-Market Notification from the FDA regarding the LacriJet® device. It details the device's indications for use, its comparison to predicate devices, and a summary of non-clinical tests.

    However, the document does not describe a study involving an AI/Machine Learning device that requires a test set, ground truth established by experts, MRMC studies, or standalone algorithm performance. The LacriJet® is a physical medical device (a monocanalicular lacrimal stent), not a software or AI-based diagnostic/assessment tool.

    Therefore, it is not possible to extract the requested information regarding acceptance criteria met by a study, sample size, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, or training set details, as these concepts are not applicable to the type of device and submission described in the FDA document.

    The document primarily focuses on establishing substantial equivalence based on:

    • Intended Use: Similar to predicate devices (MASTERKA, Eagle Vision Monocanalicular Stent, Mini-Monoka).
    • Technological Characteristics: Made of the same medical-grade silicone as a predicate, similar design for the stent and fixation head. The primary difference is in the delivery system, which the submission argues does not raise new safety or effectiveness concerns.
    • Non-Clinical Testing: Confirms the device meets established specifications related to biocompatibility, sterilization, package integrity, and shelf-life, demonstrating equivalence to the predicate device in these aspects.

    In summary, the provided text does not contain the information requested in your prompt as it pertains to a physical medical device clearance, not an AI/ML-driven device validation study.

    Ask a Question

    Ask a specific question about this device

    K Number
    K161373

    Validate with FDA (Live)

    Device Name
    Nunchaku
    Date Cleared
    2016-11-03

    (170 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Nunchaku stents are intended for use for nasolacrimal intubation in patients 12 months and older. Indications for nasolacrimal intubation performed with the Nunchaku are:

    • Canalicular pathologies (e.g., congenital or acquired stenosis, laceration)
    • During dacryocystorhinostomy
    • Congenital lacrimal duct obstruction
    Device Description

    The Nunchaku is a bicanalicular intubation device for the treatment of epiphora in patients 12 months and older. The device consists of two lateral silicone tubes connected to a silicone body that is delivered pre-mounted on two guides facilitate insertion of the Nunchaku and are completely removed once insertion of the device is complete. The Nunchaku comes in two different model lengths (90 or 105 mm depending on length of tube required) and is provided as a sterilized product.

    AI/ML Overview

    Based on the provided text, the device in question (Nunchaku) is a bicanalicular intubation device. The document describes non-clinical (bench-top) tests and results, but does not describe a study involving human readers or an AI algorithm, nor does it establish ground truth through expert consensus, pathology, or outcomes data for clinical performance. Therefore, I am unable to provide information on the following points:

    • Acceptance criteria related to clinical performance or human reader studies.
    • Sample sizes for test sets in a clinical study.
    • Number and qualifications of experts for ground truth.
    • Adjudication methods.
    • Multi-reader multi-case (MRMC) comparative effectiveness study.
    • Standalone algorithm performance.
    • Type of clinical ground truth used.
    • Sample size for a training set (as no AI algorithm is mentioned).
    • How ground truth for a training set was established.

    The document focuses on demonstrating substantial equivalence to predicate devices through technical characteristics, intended use, and non-clinical testing.

    Here's a summary of the provided information relevant to acceptance criteria and "studies" (non-clinical tests) that "prove" the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance (Non-Clinical)

    The document lists several non-clinical tests conducted to validate the device's mechanical characteristics and biocompatibility. The "acceptance criteria" are implied by the statement "All non-clinical test results met the established specifications for the device" and "The biocompatibility test results met the required specifications for all tests."

    Acceptance Criteria (Implied)Reported Device Performance
    Manufacturing process validatedCapacity to manufacture Nunchaku demonstrated
    Mechanical integrity during guide detachmentMet established specifications
    Mechanical integrity during guide insertion and removalMet established specifications
    Tensile strength of silicone tube (resistance to breakage/pull-out)Met established specifications; demonstrated sufficient tensile strength
    Biocompatibility (cytotoxicity, irritation, sensitization, systemic toxicity, muscular implantation, endotoxins)Met required specifications for all tests
    Sterilization effectivenessStudies performed according to applicable standards; results support shelf-life
    Shipping and package integrityStudies performed according to applicable standards; results support storage

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The document states "Bench top testing was performed on samples before and after sterilization." The exact number of samples tested for each non-clinical test is not specified in the provided text.
    • Data Provenance: Not applicable as these are bench-top tests, not a clinical study involving patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. The ground truth for these non-clinical tests is based on measurable physical properties and established ISO/USP standards, not expert clinical consensus.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable for non-clinical testing.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study was mentioned or performed. The device is a physical medical device (stent) for intubation, not an AI-powered diagnostic tool for image interpretation or a device requiring human reader assistance with AI.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. The device is a physical medical device. There is no mention of an algorithm involved.

    7. The type of ground truth used:

    • For non-clinical tests, the "ground truth" is defined by established engineering specifications, mechanical properties standards, and biocompatibility standards (e.g., ISO standards, USP standards for endotoxins).

    8. The sample size for the training set:

    • Not applicable. These are non-clinical engineering and materials tests, not an AI model.

    9. How the ground truth for the training set was established:

    • Not applicable.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1